EASYBus-Sensor module for temperature
H20.0.3X.6C-06
from version V3.2
Operating Manual
EBT – AP
WEEE-Reg.-Nr.: DE93889386
GHM GROUP – Greisinger
GHM Messtechnik GmbH | Hans-Sachs-Str. 26 | 93128 Regenstauf | GERMANY
Tel.: +49 9402 9383-0 | info@greisinger.de |
www.greisinger.de
Intended use
The device measures temperature.
Field of application
- Room climate monitoring
- Monitoring of storage rooms
etc…
The safety instructions (see chapter 3) have to be observed.
The device must not be used for purposes and under conditions for that the
device had not been designed.
The device must carefully dealt with and has to be used according to the
specifications (do not throw, knock, etc.). It has to be protected against
dirt.
Do not expose the sensor to aggressive gases (like ammonia) for longer time.
Avoid condensation, as after drying there may remain residues, which may
affect the precision negatively.
In dusty environment additional protection has to be applied (special
protection caps).
General advice
Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to- hand way in order to be able to look up in the case of doubt.
Safety instructions
This device has been designed and tested in accordance to the safety
regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed
unless the standard safety measures and special safety advises given in this
manual will be adhered to when using it.
-
Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under “Specification”.
Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start- up. -
General instructions and safety regulations for electric, light and heavy current plants, including domestic safety regulations (e.g. VDE), have to be observed.
-
If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully.
Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or destroying the device or another device connected. -
Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if:
– there is visible damage to the device
– the device is not working as specified
– the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance. -
Warning: Do not use this product as safety or emergency stop device or in any other application where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage. -
This device must not be used at potentially explosive areas! The usage of this device at potentially explosive areas increases danger of deflagration, explosion or fire due to sparking.
-
This device is not constructed for use in medical applications.
Disposal notes
This device must not be disposed as “residual waste”.
To dispose this device, please send it directly to us (adequately stamped).
We will dispose it appropriately and environmentally friendly.
Assignment of elbow-type plug
2-wire connection for EASYBus, no polarity, at terminals 1 and 2
General installation instructions:
To mount the connection cable (2-wire) the elbow-type plug screw has to be
loosened and the coupling insert has to be removed by means of a screw driver
at the position indicated (arrow).
Pull out connection cable through PG gland and connect to the loose coupling
insert as described in the wiring diagram. Replace loose coupling insert onto
the pins at the transducer housing and turn cover cap with PG gland in the
direction desired till it snaps on (4 different starting positions at 90°
intervals). Re-tighten the screw at the angle plug.
Design types, dimension
Display Functions (only available for devices with option …-VO)
8.1 Measuring display
During normal operation the display will show the temperature in [°C] or [°F].
8.2 Min/Max Value Memory
watch Min values (Lo):| press ▼ shortly once| display changes between ‘Lo‘ and
Min values
---|---|---
watch Max values (Hi):| press ▲ shortly once| display changes between ‘Hi‘
and Max values
restore current values:| press ▼ or ▲ once again| current values are displayed
clear Min-values:| press ▼ for 2 seconds| Min values are cleared. Display
shows shortly ‘CLr‘.
clear Max-values:| press ▲ for 2 seconds| Max values are cleared. Display
shows shortly ‘CLr‘.
After 10 seconds the currently measured values will be displayed again.
8.3 Min/Max Alarm Display
Whenever the measured value is exceeding or undershooting the alarm-values
that have been set, the alarm-warning and the measuring value will be
displayed alternating.
AL.Lo | the lower alarm boundary is reached or is undershot |
---|---|
AL.Hi | the upper alarm boundary is reached or is exceeded |
Error and system messages
Display | Description | Possible fault cause | Remedy |
---|---|---|---|
Err.1 | Measuring range | ||
exceeded | Wrong signal | Temperatures above the measuring range are not |
allowed.
Err.2| Measuring value below
measuring range| Wrong signal| Temperatures below the measuring range are not
allowed.
Err.7| System fault| Error in device| Disconnect from supply and reconnect.
If error remains: return to manufacturer
Err.9| Sensor error| Sensor or cable defective| Check sensors, cable and
connections, damages visible?
Er.11| Calculation not possibl| Calculation variable
missing or invalid| Check temperature
8.8.8.8| Segment test| The transducer performs a display test for 2 seconds
after power up. After that it will change to the display of the measuring.
Configuration of the device
10.1 Configuration via interface
The configuration of the device is done by means of the PC-software EASYBus-
Configurator or EBxKonfig.
The following parameters can be changed:
- Adjusting of temperature display (offset and scale correction)
- Setting of the alarm function for temperature
The adjusting by means of offset and scale is intended to be used to compensate errors of the measuring’s.
It is recommended to keep the scale correction deactivated. The display value is given by following formula: value = measured value – offset
With a scale correction (just for calibration laboratories, etc) the formula changes: value = (measured value – offset) * ( 1 + scale adjustment/100)
10.2 Configuration at the device (only available for device with option …-VO)
Note:
If EASYBus sensor modules are operated by a data acquisition software, there
can be problems if the configuration is changed during a running acquisition.
Therefore it is recommended not to change configuration values during a
running recording and furthermore to protectit against manipulation by
unauthorised persons. (please refer to right picture)
Follow these instructions to configure the functions of the device:
-
Press SET until the first parameter appears in the display
-
If a parameter should be changed, press ▼ or ▲,
The device changed to the setting – edit with ▼ or ▲ -
Confirm the value with SET
-
Jump to the next parameter with SET.
Parameter | value | information |
---|---|---|
SET | ▼ and ▲ | |
+ Temp arrow | Unit of temperature displays factory setting: °C |
°C
°F| Temperatures in °Celsius
Temperatures in °Fahrenheit
+ Temp arrow| Offset correction of temperature measuring *)
oFF
_2.0 … +2.0| deactivated (factory setting)
Selectable from -2.0 to +2.0 °C
+ Temp arrow| Scale correction of temperature measuring *)
oFF
-5.00 +5.00| deactivated (factory setting)
Selectable from -5.00 to +5.00 % scale correction
+ Temp arrow| Min. alarm-point for temperature measuring
Min.MB … AL.Hi| Selectable from: min. measuring range to AL.Hi
+ Temp arrow| Max. alarm-point for temperature measuring
AL.Lo … Max.MB| Selectable from: AL.Lo to max. measuring range
+ Temp arrow| Alarm-delay for temperature measuring
oFF
1 ••• 9999| deactivated (factory setting)
Selectable from 1 to 9999 sec
Pressing SET again stores the settings, the instruments restarts (segment
test)
Please note: If there is no key pressed within the menu mode within 2
minutes, the configuration will be cancelled, the entered settings are lost!
) if higher values are needed, please check sensor, if necessary return to manufacturer for inspection.
Calculation: corrected value = (measured value – Offset) (1+Scale/100)
Notes to the calibration services
Calibration certificates – DKD-certificates – other certificates:
If device should be certificated for its accuracy, it is the best solution to
return it with the referring sensors to the manufacturer.
Specification
Measuring range | please refer to type plate |
---|---|
EBT – AP1, AP3, AP4 | – 50,0 … 150,0 °C or – 58,0 … 302,0 °F |
EBT – AP2 | – 50,0 … 400,0 °C or – 58,0 … 752,0 °F |
EBT – AP5 | -199,9 … 650,0 °C or -199,9 … 999,9 °F |
Accuracy (at nominal temperature) Electronic:
Sensor:| ±0,2 % of measured value ±0,2 °C
please refer to type plate
Sensor| Pt1000 sensor, 2-wire
Meas. frequency| 1 per second
Adjusting| Digital offset and scale adjustment
Min-/Max-value memory| Min and max measured values are stored
Output signal
Connection
Busload| Easybus-protocol
2-wire Easybus, polarity free
1.5 Easybus-devices
Display (only with option VO)
Operating elements| approx. 10 mm high, 4-digit LCD-display 3 keys
Ambient conditions
Nom. temperature
Operating temperature
Relative humidity
Storage temperature| 25°C
-25 … 70 °C
0 … 95 %RH (not condensing)
-25 … 70 °C
Housing
Dimensions
Mounting
Mounting distance
Electrical connection
Design-types: EBT – AP1: EBT – AP2: EBT – AP3: EBT – AP4: EBT – AP5:| ABS
(IP65, except sensor head)
82 x 80 x 55 mm (without elbow-type plug and sensor tube)
Holes for wall mounting (in housing – accessible after cover has been
removed).
50 x 70 mm, max. shaft diameter of mounting screws is 4 mm
Elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm with
threaded stem for direct screw connection.
with threaded stem at a distance from housing (for higher temperature).
indoor /outdoor probe for direct wall mounting.
duct type probe with centrally mounted sensor tube arrangement with the shaft
of device pointing downwards at a 90° angle measuring transducer for external
Pt1000 sensors.
Insertion of sensor cable through PG7 screwing.
Directives / standards| The instruments confirm to following European
Directives:
2014/30/EU EMC Directive
2011/65/EU RoHS
Applied harmonized standards:
EN 61326-1 : 2013 emissions level: class Bemi immunity according to table 2
Additional fault: <1%
When connecting long leads adequate measures
against voltage surges have to be taken.
EN 50581 : 2012