midmark CMDT-45 Canis Major Wet Dental Treatment Lift Table User Guide
- June 16, 2024
- Midmark
Table of Contents
Canis Major® Wet Dental / Treatment
Lift Table
For models:
CMDT-45
CMDT-WS-45
CMDT-60
CMDT-WS-60
User’s Guide
StyleP
003-1725-00 Rev. AA4 3/9/21
Product Information
(The information below is required when calling for service)
Dealer [name / phone]: Date of Purchase:
Model / Serial Number:
Midmark Authorized Service Company [name / phone]:
- Model / Serial
- Number Locations
Important Information
Symbols
These symbols may appear on your equipment and/or in the manuals.
WARNING
Indicates a hazardous situation which could result in serious injury if not
avoided.
CAUTION
Indicates a potentially hazardous situation which could result in minor or
moderate injury if not avoided. It may also be used to alert against unsafe
practices.
Equipment Alert
Indicates a potentially hazardous situation which could result in equipment
damage if not avoided.
Note
Amplifies a procedure, practice, or condition.
Dangerous voltage / shock hazard
Catalogue Number
Manufacturer
Consult User Guide for important information.
**** Serial number
Transportation / Storage Conditions
Ambient Temperature Range: ………….. 5°C to +38°C (+41°F to 100°F)
Relative Humidity …………………………….. 10% to 90% (non-condensing)
Atmospheric Pressure ………………………. 500hPa to 1060hPa (0.49atm to 1.05atm)
Intended Use
This product is intended to be used as a table to provide positioning, support and cleaning of animals during examination or minor surgery procedures conducted by qualified professionals.
Electrical Precautions
WARNING
When using high frequency surgical devices or endocardial catheters:
- Use nonconductive material to insulate patient from metal portions of table.
- Consult operating instructions for the device before use.
Failure to comply may result in electric shock or burns to patient.
Safety Instructions
CAUTION
Be sure that all personnel and equipment are clear of the table before
activating any function. Failure to do so could result in personal injury.
At the table’s lowest height [9.5 in. (24.1 cm)] the clearence between the bottom of the table and the floor is approximately 4.5 in. (11.4 cm). Make sure patient and all objects are clear of this area before lowering table.
If table malfunctions, immediately release control, unplug the table power cord and assist the patient from the table. If table continues to malfuction, call for service.
Failure to comply with these instructions may result in injury to patient and/or damage to equipment.
VA19661
- Clearance
Approx. 4.5 in. (11.4 cm)
Operation
Table Up / Down
VA101400i
- UP
- DOWN
To move the table top UP or DOWN…
Press and hold the appropriate button.
Note
Movement automatically stops at the max. / min. height.
Water System
To acquire water pressure…
Turn water valve knob to side and utilize spray nozzle.
Equipment Alert
Always close shutoff valve when table is not in use.
VA101200i
VA101300i
- CAP SCREW
To adjust water temperature…
Pull table away from wall, loosen screw on cap, turn mixing valve knob until
desired temperature is achieved then tighten cap screw.
[factory setting: 70°F (21°C)] [adjustable range: 70° – 120°F (21° – 49°C)]
Scale (optional)
The 300 pound capacity platform scale utilizes four (4) built in load sensors. These electronic sensors are accurate within .67%, and are protected against impacts, moisture and liquid spills. The scale has been calibrated and all parameters are preset so it should never need calibration.
Equipment Alerts
- Prevent flammables and liquids from entering the scale head.
- Allow adequate clearance on all sides of scale platform to ensure accuracy.
- Do not drop large loads onto the scale platform.
- Never exceed the rated capacity of the scale.
- Do not operate in close proximity to RF transmitters.
There are two (2) controls located on the display:
UNITS Toggles the displayed weight between pounds (lbs) and kilograms (kg).
ZERO Resets the display to indicate zero (0.0).
Note
The scale will not respond to the ZERO control under these conditions:
Motion on Scale
An under load condition (less than 0.1lbs or 50g)
An over load condition (more than 300lbs or 1kg)
Weighing an animal without a cage:
- Zero scale (0.0), place animal on center of platform and read the weight displayed.
Note
A blank scale while weighing means excess motion is detected and weight
averaging is in process. Once completed, the scale will display and HOLD the
recorded weight. Adding additional weight while in the HOLD condition will NOT
change the display.
Weighing an animal with a cage:
- Place empty cage on center of platform, then zero scale (0.0).
- Remove cage from platform; scale will read negative weight (-.).
- Place cage with animal on center of platform and read/record weight.
- To continue to weigh animals in same sized cage, place next cage with animal on center of platform and read/record weight; do not zero scale.
Troubleshooting:
Symptom or Error | Possible Problem (and/or) Remedy |
---|---|
No Display | Power is OFF. / Table is unplugged from the power source. |
Incorrect weight | Interference with scale platform. / Ensure there is nothing |
in contact with the sides of the scale platform. Ensure the correct UNIT (of
measure) is selected.
Display segment (s) missing/illegible| Low or intermittent voltage. / Unplug
for one (1) minute to reset system then reconnect to original power source.
Disconnect power cord and use different power source.
ZERO key does not reset to [0.0]| Motion or activity on the scale. / Remove
source of motion. Ensure platform is level and supported evenly at corners.
Weight on scale may exceed the zero limit. Unplug for one (1) minute to reset
system.
Display “locked & will not change | System Error. / Unplug the unit for one
(1) minute to reset system. then reconnect to original power source.
Displays “HiLoAd”| Weight on scale exceeds capacity. / Remove excess load.
Displays “HiLdX” | The load placed on the scale is too heavy on a corner
[shown as X]. / Move load to center of the scale platform.
Displays “LoLdX” | The scale is under loaded / Ensure that tabletop cover pins
are properly seated on load cells.
If the problem cannot be resolved using the above corrective actions, please contact Midmark at 1-800-Midmark (1-800-643-6275) or 937-526-3662.
Maintenance
Cleaning
Debris Trap:
A debris trap exists beneath the drain of the stainless steel table tub. It
should be visually inspected daily and removed and cleaned weekly if table is
used only for examinations. The debris trap must be used and should be removed
and cleaned after each procedure.
VA101500i
-
Inspect debris trap…
If any debris is evident, remove cap. -
Clean debris trap…
Remove yellow trap, clean and replace.
Painted Surfaces:
Clean the table weekly, wiping the painted metal and plastic surfaces with a
clean, soft cloth, and mild cleaners.
Stainless Steel Surfaces:
For best results, stainless steel should be cleaned weekly or whenever films
or deposits become apparent. Use a mild cleaning agent and whenever possible
rinse thoroughly and towel dry. More stubborn deposits or tightly adhering
debris may require vigorous rubbing, or brushing. Always scrub with the grain
of the stainless steel. If a commercial cleaner needs to be used, make sure it
is approved for use on stainless steel surfaces.
Equipment Alert
Always use nylon or plastic bristle scrubbers. The use of steel wool or steel
brushes may cause steel particles to become embedded in the surfaces and will
cause rust.
Avoid prolonged surface exposure to chlorides, bromides, iodides and thiocyanates. Salty solutions allowed to evaporate on stainless steel surfaces may cause pitting, corrosion or metal discoloration.
Troubleshooting
If the table will not go up or down, check the following:
- Unit is plugged into a wall outlet
- Power to the outlet
- No visible damage to wiring
- The table is at its maximum or minimum height
- Hand control is fully inserted and making proper connections
If the table won’t drain, check the following:
- The filter is clogged
If the pump won’t start, check the following:
- The filter is clogged
- Unit is plugged into a wall outlet
- Power to the outlet
- The water sensor is out of adjustment (contact factory)
If the pump won’t stop running:
- The water sensor is out of adjustment (contact factory)
If the scale does not operate, check the following:
- Refer to scale section elsewhere in this manual
Preventive Maintenance
Periodically inspect the following areas:
- Power cord should be free of cuts or other visible damage.
- All fasteners should be present and tightened securely.
- All mechanical functions should operate properly.
Have an authorized service technician inspect your table on a regular basis.
Calling For Service
Note
Model / Serial Number information is required when calling for service.
If service is required, contact your Midmark dealer.
To contact Midmark directly:
1-800-Midmark (1-800-643-6275) or
937-526-3662
8:00 am until 5:00 pm. Monday through Friday (EST)
[excluding standard U.S. holidays]
Specifications
Weights, Dimensions, Electrical Specifications
Maximum Patient Weight: ………………….. 300 lbs (136 kg)
Weight of Table:
45″ tabletop ………………………………………………… 366 lbs (166 kg)
45″ tabletop [with packaging & skid] ………….. 435 lbs (197 kg)
60″ tabletop ………………………………………………… 395 lbs (179 kg)
60″ tabletop [with packaging & skid] ………….. 465 lbs (211 kg)
(Add 5 lbs (2.2 kg) for units equipped with a scale)
Range of Motion & Dimensions: ……………. (see illustration below)
Electrical Requirements: ………………………….. Dedicated Circuit, 115 VAC, 60Hz,
10A
Power Cord: ……………………………………………….. 4 ft. (121 cm) long
Duty Cycle
(actuator run time): ……………………………… Intermittent Operation [1 minute ON / 9
minutes OFF] (pump run time): ………………………………….. Intermittent Operation (thermal
protection) [90 minute ON @ < 230° case temp.] GFI Receptacle:
…………………………………….. Class A, 2P, 125V, 60HZ
Range of Motion & Dimensions
VA1699i
- Water Inlet Height
- Drain Outlet Height
Warranty Information
SCOPE OF WARRANTY Midmark Corporation (“Midmark”) warrants to the original retail purchaser that it will repair or replace components of the animal health products manufactured by Midmark (except for products and components not warranted under “Exclusions”) that are defective in material or workmanship under normal use and service. The sole remedy under this limited warranty is the repair or replacement, at Midmark’s option, of the applicable products or components. This limited warranty shall only apply to defects that: (i) are reported to Midmark within the applicable warranty period; and (ii) are determined to exist upon examination by Midmark. This limited warranty extends only to the original retail purchaser of a product, and is not transferable or assignable.
APPLICABLE WARRANTY PERIOD The applicable warranty period, measured from the date of delivery to the original user, shall be one (1) year for all warranted products and components.
OBTAINING WARRANTY SERVICE Warranty service must be obtained through either Midmark or an authorized dealer in the Midmark product line for which warranty service is requested. Midmark may be contacted for warranty service inquiries or issues via email at www.midmark.com, by mail to Midmark Corporation, 60 Vista Drive, Versailles, Ohio 45380, or by phone at: 1-800MIDMARK.
It is the retail purchaser’s obligation to arrange for delivery of a product to Midmark or one of its authorized dealers for warranty service, which delivery shall be at retail purchaser’s expense. It is also the retail purchaser’s obligation to comply with the warranty service instructions provided either by Midmark or its authorized dealer. The retail purchaser must provide Midmark with completed warranty registration information within thirty (30) days after purchase in order to obtain the benefits of this limited warranty.
EXCLUSIONS This limited warranty does not cover and Midmark shall not be liable for the following:
(1) defects, damage or other conditions caused, in whole or in part, by
misuse, abuse, negligence, alteration, accident (including animal acts of any
kind), freight damage, tampering or failure to seek and obtain repair or
replacement in a timely manner;
(2) matching of color, grain or texture except to commercially acceptable
standards;
(3) changes in color caused by natural or artificial light;
(4) products which are not installed, used and properly cleaned and maintained
as required in the installation and operation manuals for the applicable
product (imaging product must be installed by a certified Midmark installer);
(5) products considered to be of a consumable nature;
(6) accessories or parts not manufactured by Midmark;
(7) specially manufactured products;
(8) charges by anyone (including Midmark’s authorized dealers) for
adjustments, repairs, replacement parts, installation or other work performed
upon or in connection with such products which are not expressly authorized in
writing in advance by Midmark;
(9) costs and expenses of routine maintenance and cleaning;
(10) all sinks, faucets and plumbing accessories;
(11) representations and warranties made by any person or entity other than
Midmark; and
(12) with respect to software that is a product or a component thereof, that
the software will be error free, can be used without problems or
interruptions, or will be free from vulnerability to intrusion or attack by
viruses or other methods.
EXCLUSIVE REMEDY; CONSEQUENTIAL DAMAGES DISCLAIMER Midmark’s only obligation under this limited warranty is the repair or replacement of defective parts. Midmark shall not be liable for and hereby disclaims any direct, special, indirect, incidental, exemplary or consequential damages or delays including, but not limited to, damages for loss of profits or income, loss of use, downtime, cover, and employee or independent contractor wages, payments and benefits. This disclaimer shall survive any failure or asserted failure of the essential purpose of this limited warranty or its remedies specified herein.
NO AUTHORIZATION No person or firm is authorized to create or approve for Midmark any other obligation or liability in connection with Midmark products.
WARRANTY DISCLAIMER: THIS LIMITED WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
STATUTE OF LIMITATIONS No action may be brought against Midmark for breach of this limited warranty, an implied warranty, if any, or for any other claim arising out of or relating to the products, more than ninety (90) days following expiration of the warranty period. In the event multiple warranty periods exist with respect to a product, the ninety (90) day period provided for herein shall begin to run from expiration of the warranty period for the component to which the claim relates.
SEVERABILITY In the event any provision of this limited warranty is determined to be invalid or otherwise unenforceable: (i) the provision shall be enforced in a manner that closest holds to its intent, while at the same time curing the invalidity or unenforceability, and (ii) the balance of this limited warranty, being severable, shall not be affected in the event the provision that is invalid or unenforceable cannot be enforced in any respect.
Midmark Corporation
60 Vista Drive
Versailles, OH 45380-0286
1-800-643-6275
1-937-526-3662
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>