MELAG Vacuklav 24 B+Documentation Software MELAtrace User Guide
- June 16, 2024
- MELAG
Table of Contents
General information for installation
Pro-Class, Euroklav, Premium-Plus-Class, Vacuvet 23 B+
Vacuklav 24 B+Documentation Software MELAtrace
Devices with fixed connection
(Vacuklav 24 B+, 24 BL+, 30 B+, 40 B+, 44 B+)
Space requirements
The space requirements for the device corresponds to its dimensions plus at
least 5 cm on the side and above the device to provide a sufficient
ventilation.
When you incorporate the device into existing furnishings, be sure to maintain
5 cm of clearance space on all sides.
Steam egress can occur during operation. Do not set up the device in the
immediate proximity of a smoke detector. Maintain clearance from materials
which could suffer damage from steam.
Additional space for feed water supply
There is additionally required space for the MELAG storage container,
alternatively for the water treatment unit MELAdem 40, MELAdem 47 or other
containers such as the original container of the supplier.
Bearing capacity (without load and feed water)
Place the device on a level and horizontal surface.
Model name | Weight | Model name | Weight |
---|---|---|---|
Vacuklav 24 B+ | 48 kg | Vacuklav 40 B+ | 55 kg |
Vacuklav 24 BL+ | 53 kg | Vacuklav 44 B+ | 64 kg |
Vacuklav 30 B+ | 45 kg |
Mains supply
Model name| Vacuklav 24 B+, 24 BL+, 30 B+| Vacuklav 40 B+, 44
B+
---|---|---
Electrical connection| 220-240 V) , 50/60 Hz,separate fuse protection with 16
A, residual current device 30 mA| 220-240 V) , 50/60 Hz, separate fuse
protection with 16 A, residual current device 30 mA
Electrical power| 2100 W (from SW v4.07)| 3400 W
*) max. voltage range 207-253 V
Cold water connection (max. temperature. 15 °C)
The connection of the device to the water supply is comparable with the
connection of a washing machine in a domestic context. The water connection
must be done according to the European standard EN 1717. A separate 3/4” water
tap with integrated return flow inhibitor and pipe aerator is recommended
(nominal width of water pipe DN15 with 1/2” fitting).
NOTICE
Do not connect any further device to an existing angle valve.
Malfunctions in the cooling water system could be the consequence.
Flow pressure
Required flow pressure with a flow rate of 3 l/min > 1.2 bar.
Inlet hose
Length of the cooling water inlet hose is 2.5 m (DN16), optional 5 m available
Water stop
MELAG recommends the installation of a water stop (leak monitor with shut-off
valve and probe).
Wall drain or siphon drain
A wall drain, nominal width DN40 or a siphon drain (sink drain) is required to
drain off the cooling water.
Double chamber wastewater trap
MELAG delivers a special double chamber wastewater trap which can be
connected to an available siphon fixture for the quiet drainage of the cooling
water instead of the existing siphon of a sink. Alternatively, you can acquire
a surface-mounted siphon.
Outlet hose
In order to attach the device to the effluent pipe, MELAG delivers a
wastewater hose with a length of 2 m (optional 5 m). The drain must be located
at least 30 cm beneath the device and be installed dip-free with continuous
descent.
Stand-alone devices
(Vacuklav 23 B+, 31 B+, 41 B+, 43 B+, 23 S+,23 VS+, 29 Vs+, Vacuvet 23 B+)
Space requirements
The area above the steam sterilizer should be freely accessible in order to
enable easy filling of the storage tank and good ventilation. The steam
sterilizer works with a cooler on the rear of the device for the cooling
system (not with Euroklav). The function and life-span of the steam
sterilizer can be affected if the heat dissipation above the cooler is
restricted in any way. As such, MELAG advise against incorporating the steam
sterilizer and is only possible if sufficient air circulation is ensured.
Steam egress can occur during operation. Do not set up the device in the
immediate proximity of a smoke detector. Maintain clearance from materials
which could suffer damage from steam.
Requirements for the incorporation of a device
If the incorporation of the device is absolutely necessary, one of the
following measures must be implemented:
- The device must be able to be pulled out for operation.
- In the installation space, there must be an exhaust air duct in the rear area which diverts the warm air upwards or actively to the rear.
- For Vacuklav 41 B+ Evolution / 43 B+ Evolution: Switching off the DRYtelligence (menu Settings > In- telligent drying).
Bearing capacity (without load and feed water)
Model name | Weight | Model name | Weight |
---|---|---|---|
Vacuklav 23 B+ | 50 kg | Vacuklav 41 B+ | 60 kg |
Vacuklav 31 B+ | 45 kg | Vacuklav 43 B+ | 69 kg |
Vacuvet 23 B+ | 50 kg | Euroklav 23 VS+ | 45 kg |
Euroklav 23 S+ | 45 kg | Euroklav 29 VS+ | 42 kg |
Mains supply
Model name| Vacuklav 23 B+, 31 B+, Vacuvet 23 B+| Euroklav 23 S+,
23 VS+, 29 VS+| Vacuklav 41 B+, 43 B+
---|---|---|---
Electrical connection| 220-240 V ) , 50/60 Hz, separate fuse protection with
16 A, residual current device 30 mA| 220-240 V ) , 50/60 Hz, separate fuse
protection with 16 A, residual current device 30 mA| 30 mA 220-240 V ) ,
50/60 Hz, separate fuse protection with 16 A,
residual current device
Electrical power| 2100 W (from SW v4.07)| 2300 W (23 S+, 23 VS+)
2100 W (29 VS+)| 3400 W
) max. voltage range 207-253 V
PLEASE NOTE
You find more information regarding setting up and installation of the
device in the respective Technical Manual.
Rev. 3 – 22/3748
AS_093-11_Profi_PremPlus