MAXUS 50G0001G Digital Milligram Scale
INSTALLATION INSTRUCTION
Calibration
- As a milligram reloading scale, the scale should be calibrated before each use, or if the scale is moved to a new place.
- Take out the 20-gram calibration weight from the box.
- Find a flat and level surface to perform calibration and let the scale acclimate to room temperature.
- Be sure the scale is on, warmed up at least 60 seconds, then press the “TARE“ key to zero the display.
- Press and hold the “MODE“ key for 6+ seconds, until the display starts to flash 20.000.
- Gently put the 20g weight in the center of the platform. After a few seconds, the display will display 40.000, and then put another 20g weight on it. After a few seconds, the screen shows “PASS“, then you may remove the calibration weight from the platform.
- The calibration is complete and the weighing unit has been reset to gram.