UNITRONICS Z645 Series Zoom Stereo Microscope User Manual
- June 14, 2024
- UNITRONICS
Table of Contents
- SAFETY NOTES
- CARE AND MAINTENANCE
- INTRODUCTION
- UNPACKING AND COMPONENTS
- COMPONENTS DIAGRAM
- ASSEMBLY
- DETAILED ASSEMBLY PROCEDURE
- ADJUSTMENT AND OPERATION
- SPECIFICATIONS
- TROUBLESHOOTING
- MAINTENANCE
- SERVICE
- LIMITED MICROSCOPE WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Z645 SERIES
ZOOM STEREO MICROSCOPE
MANUAL
SAFETY NOTES
- Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from dropping and being damaged.
- Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
- If any specimen solutions or other liquids splash onto the stage, objective or any other component, disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be damaged.
- LAMP REPLACEMENT — CAUTION: the glass housing of the lamp may be extremely hot. DO NOT attempt to change the lamp before it is completely cooled or without wearing adequate skin protection.
- All electrical connectors (power cord) should be inserted into an electrical surge protector to prevent damage due to voltage fluctuations.
- FUSE REPLACEMENT — For safety when replacing the fuse (ONLY replace with the same size, type and rating of original fuse), be sure the main switch is in the off position, disconnect the power cord from outlet, and replace the fuse. Reconnect the power cord and turn unit on.
- Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a different input voltage other than indicated will cause severe damage to the microscope.
NOTE: Always plug the microscope power cord into a suitable grounded electrical outlet.
A grounded 3-wire cord is provided.
CARE AND MAINTENANCE
- Do not attempt to disassemble any component including eyepieces, objectives or the focusing assembly.
- Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap solution. Do not use organic solvents for cleansing.
- The outer surface of the optics should be inspected and cleaned periodically using an air bulb. If dirt remains on the optical surface, use a soft, lint free cloth or cotton swab dampened with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular motion. A small amount of absorbent cotton wound on the end of a tapered stick makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult.
- Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in use.
- UNITRON® microscopes are precision instruments which require periodic servicing to maintain proper performance and to compensate for normal wear. A regular schedule of preventative maintenance by qualified service personnel is highly recommended. Your authorized UNITRON ® distributor can arrange for this service.
INTRODUCTION
Congratulations on the purchase of your new UNITRON® microscope. UNITRON microscopes are engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if used and maintained properly. UNITRON microscopes are carefully assembled, inspected and tested by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the shipping carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form. Carefully remove the microscope from the shipping carton and place the microscope on a flat, vibration-free surface.
COMPONENTS DIAGRAM
NOTE: Stand ordered may not be as pictured above.
ASSEMBLY
The diagram below shows how to assemble the various modules. The numbers
indicate the order of assembly.
When assembling the microscope, make sure that all parts are free of dust and
dirt, and avoid scratching any parts or touching glass surfaces.
DETAILED ASSEMBLY PROCEDURE
Mounting the Binocular Head (Fig. 1)
Loosen the lock screw ③ on the focus mount ④ and insert the binocular head and
zoom body assembly ② into the bracket of the focus mount, then tighten the
lock screw ③.
Mounting the Eyepieces (Fig. 1, Fig. 2, Fig. 3)
Remove the eyepiece dust caps ① (Fig. 1) and gently insert the two adjustable
eyepieces ⑤ (Fig. 2) into the eyepiece sleeves until they stop, as shown in
Fig.3. Replace the dust caps when not in use.
IMPORTANT
ENVIRONMENT REQUIREMENTS
- Ambient temperature: 0°C to 40°C (32°F to 104°F) ,Maximum R.H :85% 。
- High temperatures and wide swings in temperature will result in moisture and mildew and could damage instrument.
- Avoid placing the instrument in a dusty environment. When not in use, cover the microscope with the dust cover.
- Keep the microscope on a flat and stable surface.
For Microscopes Purchased With Illuminated Stands
VOLTAGE CHECK
Confirm that the input voltage indicated on the rear label of the microscope
corresponds to your line voltage. The use of a different input voltage than
indicated will cause severe damage to your microscope.
Connecting the Power Cord (Fig. 4)
Turn the power switch ⑥ to the off position (O) before connecting the power
cord ⑦. Insert the power cord plug ⑦ into the power jack of the microscope;
make sure the connection is snug. Plug the power cord ⑧ into the power supply
receptacle.
Using the Glass Stage (Fig. 5)
Set the glass stage plate ⑨ on the center of the base, and secure the slide
with the two clips ⑩.
Video/Photography Accessories Assembly and Use (Fig. 6)
Loosen the lock screw ① of trinocular head with an Allen key and remove the
dust cover ②.
Thread the camera adapter ③ onto the bottom of the camera.
Insert the camera adapter ③ with camera into the camera port in the direction
shown in Fig. 6, and tighten the lock screw ①.
Focus on a sample using the eyepieces, the compare to the live image on the
camera software. If the camera image is out of focus, loosen the adjustment
screw ④ on the camera adapter ③, and move the adjustment screw ④ up and down
for focusing, then lock the adjustment screw ④ until the image is clear. (See
Fig. 6)
★ Rotate the camera tube, to make the camera display image and the observation
image in the same direction.
ADJUSTMENT AND OPERATION
Adjusting Binocular Viewing Head
Adjusting Interpapillary Distance (Fig. 7)
Different users have different interpapillary distances (this distance is
between the centers of the pupils of each eye). When the operator changes, it
will be necessary to adjust the interpapillary distance.
While looking through the eyepieces, hold the left and right myotubes of the
viewing head and adjust the eyetubes by opening or closing them until the left
and right fields of view coincide completely and you are able to see a
complete circle.
Adjusting Diopter Ring (Fig. 8)
Set the dioptor rings ② of both eyepieces to “0” position. (Do this when users
change, because different users will have different diopter settings.)
Place an easy-to-observe specimen on the stage plate, i.e., a coin.
Rotate the zoom knob ③ to the highest magnification 5x, then turn the focusing
knob ④ to focus the specimen.
Rotate the zoom knob ③ to the lowest magnification 0.8x, looking only into the
left eyepiece, adjust the diopter ring on left eyepiece to focus the specimen.
Then repeat procedure for the right eyepiece.
NOTE: The working distance (the distance between the microscope objective to
the top of the specimen) of the microscope is 115mm, the binocular head
bracket ⑤ can be mounted onto higher or lower positions according to your
needs by removing the clamping bolt with the Allen wrench provided.
Focus Adjustment
Adjusting the Rotation Tension of the Focus Adjustment Knob (Fig. 9, Fig. 10)
To adjust tension, hold both left and right focus adjustment knobs ① with both
hands, hold the left knob (to prevent it from turning), and rotate the right
knob clockwise to increase (tighten) or counterclockwise to decrease (loosen)
the focus knob tension.
After tension adjustment has been completed, always rotate both adjustment
knobs in the same direction.
Changing the Magnification
The zooming knobs located on both sides of the zooming body will change the
magnification of the specimen image.
Total Magnification = Magnification of zoom body x magnification of eyepiece
(i.e., 0.8 x 10=8x)
NOTE: If using any of the optional auxiliary objectives, then the above
magnification would be multiplied by auxiliary objective magnification. (i.e.,
auxiliary 0.5 x 8=4x)
Moving the Microscope (Fig. 11)
Before moving the microscope, be sure to remove any slide/samples. When moving
the microscope, hold the stand and base as shown in Fig. 10 to keep it level.
SPECIFICATIONS
Optical System | Greenough optical design; 0.67x – 4.5x zoom range |
---|---|
Viewing Body | 45° inclined binocular head with fixed eyepiece tube or |
trinocular head with 50:50 light split; Ø76mm body; interpapillary distance
52mm – 76mm; click stops at 0.67, 0.8, 1, 1.2, 1.5, 2, 2.5, 3, 3.5, 4, 4.5;
zoom range limits
Eyepieces| WF10x/22mm; WF10x/25mm; WF15x/16mm; WF20x/12.5mm; all with diopter
adjustment
Magnification| 6.7x – 45x magnification range with 10x eyepieces and included
1x glass lens shield
Auxiliary Objectives| 1x glass shield (included);
Optional: 0.3x, 0.4x, 0.5x, 0.7x, 1.5x, 2.0x
Working Distance| 110mm with included 1x;
With optional auxiliary objectives: 177mm with 0.5x; 120mm with 0.7x; 47mm
with 1.5x
Stands & Illumination| Variety of plain focusing stands, pole stands, diascopy
stands, boom stands, ball bearing boom stands, articulating arms; wide variety
of ring lights and fiber optic illuminators
Camera Adapters (trinocular model only)| Optional: 0.35x, 0.5x, 0.65x, 1.0x
C-mount adapters; photo tube
Digital Cameras| Excels™ and Teledyne Lumen era® digital cameras
OPTICAL DATA
EYEPIECE
Auxiliary Objective| Working Distance (mm)| WF10x/22mm| WF15x/16mm|
WF20x/12.5mm
Magnification range| FOV (mm)| Magnification range| FOV (mm)| Magnification
range| FOV (mm)
None or 1.0x| 110| 6.7x – 45x| 32.8 – 4.8| 10.1x – 67.5x| 23.8 – 3.5|
13.4x – 90x| 18.6 – 2.7
0.3x| 287| 2x – 13.5x| 109.3 – 16| 3x – 20.3x| 79.3 – 11.6| 4x – 27x| 62
– 9
0.4x| 217| 2.7x – 18x| 82 – 12| 4x – 27x| 59.5 – 8.7| 5.4x – 36x| 46.5 –
6.7
0.5x| 177| 3.4x – 22.5x| 65.6 – 9.7| 5.1x – 33.8x| 47.7 – 7.1| 6.8x –
45x| 37.3 – 5.5
0.75x| 120| 5x – 33.8x| 43.7 – 6.5| 7.5x – 50.7x| 31.8 – 4.7| 10x –
67.6x| 24.8 – 3.7
1.5x| 47| 10.1x – 67.5x| 21.8 – 3.2| 15.2x – 101.3x| 15.9 – 2.3| 202.2x
– 135x| 12.4 – 1.8
2.0x| 26| 13.4x – 90x| 16.4 – 2.4| 20.1x – 135x| 11.9 – 1.7| 26.8x –
180x| 9.3 – 1.3
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local dealer for assistance.
Trouble | Cause | Remedy |
---|---|---|
Double images | Interpapillary distance is not correct | Readjust it |
Diopter adjustment is not correct | Readjust it | |
Dirt appears in the view field | Dirt on the specimen | Clean specimen |
Dirt on the surfaces of eyepieces | Clean eyepieces | |
Image is not clear | Dirt on the surface of objectives | Clear objectives |
Image is not clear while focusing change | Diopter adjustment is not correct |
Readjust diopter
Focusing is not correct| Readjust it
The focusing knob is not smooth| The focusing knob is too tight| Loosen it to
a suitable position
The image is obscure because of the head slipping down during observation| The
focusing knob is to o loosen| Tighten it to a suitable position
MAINTENANCE
Please remember to never leave the microscope with eyepieces removed and always protect the microscope with the dust cover when not in use.
SERVICE
UNITRON® microscopes are precision instruments which require periodic servicing to keep them performing properly and to compensate for normal wear. A regular schedule of preventative maintenance by qualified service personnel is highly recommended. Your authorized UNITRON distributor can arrange for this service. Should unexpected problems be experienced with your instrument, proceed as follows:
- Contact the UNITRON distributor from whom you purchased the microscope. Some problems can be resolved simply over the telephone.
- If it is determined that the microscope should be returned to your UNITRON distributor or to UNITRON for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or component should be shipped prepaid and insured.
LIMITED MICROSCOPE WARRANTY
This microscope is warranted to be free from defects in material and workmanship for a period of five (5) years for mechanical and optical components and one (1) year for electrical components from the date of invoice to the original (end user) purchaser. This warranty does not cover damage caused in-transit, misuse, neglect, abuse or damage resulting from improper servicing or modification by other than UNITRON approved service personnel. This warranty does not cover any routine maintenance work or any other work, which is reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty. No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the control of Unitron Ltd. This warranty expressly excludes any liability by Unitron Ltd. for consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or electronic component occur under this warranty contact your UNITRON distributor or UNITRON at 631-543-2000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to Unitron Ltd., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America. For all foreign warranty repairs, return freight charges are the responsibility of the individual/company who returned the merchandise for repair.
73 Mall Drive, Commack, NY 11725
631-543-2000 (P)
631-589-6975 (F)
www.unitronusa.com
info@unitronusa.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>