PCE Instruments PCE-MA 110 Absolute Moisture Meter User Manual
- June 14, 2024
- PCE Instruments
Table of Contents
PCE Instruments PCE-MA 110 Absolute Moisture Meter User Manual
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture. Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- The measuring device should be transported as horizontally as possible, with the bottom side down and well padded. Make sure that the loose attachments are removed from the measuring chamber to avoid damage. It is recommended to keep the original packaging for transport.
- Only use cold-device plugs with a PE contact.
- Do not touch the halogen bulbs or the lid during or after the measuring process as this can cause severe burns.
- After measuring, let the sample cool down before touching it.
- Never operate the device with the lid open.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of business.
Safety symbols
Safety-relevant instructions the non-observance of which may cause damage to the instrument or to injuries are additionally marked with a safety symbol.
Symbol | Designation / description |
---|---|
General warning sign Non-observance can cause damage to the device and |
injuries to the user.
| Warning: hot surface Non-observance can cause burns.
Specifications
Technical data
Model | PCE-MA 100 | PCE-MA 110 | PCE-MA 200 | PCE-MA 202 |
---|---|---|---|---|
Weighing range* | 110 g | 110 g | 200 g | 200 g |
Readability [g] | 1 mg / 0.001 g | 10 mg / 0.01 g | 1 mg / 0.001 g | 10 mg / 0.01 |
g
Repeatability| ±0.003 g| ±0.03 g| ±0.003 g| ±0.03 g
Linearity| ±0.003 g| ±0.03 g| ±0.003 g| ±0.03 g
Heating element| halogen lamp / round / approx. Ø 90 mm
Moisture measurement range| 0…100 %
Readability [%]| 0.01 %| 0.1 %| 0.01 %| 0.1 %
Drying temperature| +40…+199°C; adjustable
Drying time| 1…199 minutes; adjustable (timing mode)
Settling time| <4 seconds| <3 seconds| <4 seconds| <3 seconds
Memory| 15 drying programmes (parameters)
Display| LCD, 17 mm digit size (white on black)
Power supply| 220 V / 50 Hz
Dimensions| 200 x 180 x 380 mm
Weight| approx. 4.4 kg
Shipping dimensions| 500 x 350 x 360 mm / approx. 7 kg
Delivery scope
1 x moisture balance
1 x windscreen
1 x sample pan holder/handle
1 x sample pans
1 x mains cable
1 x user manua
Optional accessories
PCE-MA-PS; Spare sample pans (50 pieces)
PCE-MA-SL1: Replacement lamp for PCE-MA 110/202 (for 220 V version)
PCE-MA-SL2 ; Replacement lamp for PCE-MA 100/200 (for 220 V version)
PCE-BP1; Thermal printer
PCE-BP1-EP ; Spare paper rolls for PCE-BP1 (10 pieces)
Device description
**Device overview
**
- Heat protection plate
- Windscreen
- Insertion aid
- Sample pan holder
- Sample pa
Assemble the individual parts as shown in the picture above:
-
Start with the heat protection plate (1).
-
The cut-out in the side wall of the windscreen (2) points in the direction of the control panel.
-
The handle of the insertion aid (3) is inserted into this cut-out.
-
Place the sample pan holder (4) in the centre of the measuring chamber. Make sure that the cone of the sample pan holder is properly seated in the guide.
-
The sample pan (5) is inserted last.
Display
- Status display
- Heating mode
- Termination criterion
- Temperature display
- Time display
- Stability indicator
- Measured value
- Unit
Overview control panel
Key description
Key | Function |
---|
| Switching the moisture
meter on and off
| Start/stop measurement
| ** Tare / exit settings
| Menu / open settings
| ** Unit / switch between
measurement results (%MC / g / %)
| Confirm current settings / switch to the next setting
| Return to weighing mode (short key press) / calibration (long
keystroke)
| Change the settings (in the menu) /switch on the backlight (only in
weighing mode)
****| Change the settings (in the menu) /switch off the backlight (only in
weighing mode)
Preparing the measurement
First steps
Allow the instrument to acclimatize to the new environment for at least 30
minutes after unpacking
Adjust the height of the feet at
the front of the meter so that the scales are plane. There is a spirit level
on the back; you can use this for orientation.
Connect the scales to mains power by connecting the power cable to the scales*
power connector on one side and to a wall socket on the other side.
The location of the scales should be chosen so that disturbing influences are
reduced to a minimum. There should be enough space to work and the scales
should be placed on a stable desk the material of which cannot have any
magnetic effect on the scales. Strong wind, vibration, dust, strongly
fluctuating temperatures or a humidity of more than 75 % should be avoided
when using the scales. The same applies to strong heat sources and devices
that generate a strong electromagnetic or magnetic field.
Preparing the sample
Always prepare only one sample for measurement. This avoids that the sample
can exchange moisture with the environment. If several samples have to be
taken at the same time, they should be packed in airtight containers so that
their properties do not change during storage. The containers should not be
exposed to high temperature fluctuations as this may lead to condensation and
can affect the sample. Before starting the measurement, place the sample pan
and, if applicable, the glass fibre round filter on the pan stand and tare so
that only the weight of your sample is evaluated. Spread a thin layer of the
sample evenly on the sample pan to obtain reproducible results. Uneven
application will cause inhomogeneous heat distribution in the sample to be
dried, resulting in incomplete drying or an increased drying time. Due to an
accumulation of the sample, the upper layers are heated up more strongly,
which consequently leads to burns or incrustations. An excessive layer
thickness or possible incrustation prevents the moisture from escaping from
the sample. This residual moisture has the consequence that measurement
results determined in this way are not comprehensible and reproducible. In
addition, a change in the state of aggregation can lead to a change in weight,
which can falsify the result. Never re-use the aluminum sample pans. If you
use the warm/hot sample pan for the next measurement, the heat of the pan can
influence the sample even before drying. When you take several measurements in
a row, bear in mind that some heat from the previous measurement still exists
and evaporation already takes place when closing the lid, which can lead to a
measurement deviation. If you do not let the drying chamber cool down, you
will have to accept a measurement deviation.
Tools for sample preparation
The tools and instruments used for sample preparation are crucial for the accuracy and reliability of the measurement. Tools that are thermally conductive in their properties, i. e. they can transfer heat to the sample, should be avoided. Improper handling and preparation of the sample will falsify the final result of the measurement.
**Sample distribution
ATTENTION !
**
- The samples should not exceed the 8 mm height and 85 mm diameter of the sample pan.
- Exceeding the prescribed max. height of the sample can cause the sample To burn or encrust.
- The sample should be applied as evenly as possible to the sample pan.
- Make sure that the temperature sensor, which is located in the lid, does not touch the sample, as this could falsify the result.
Solids
- Distribute powdery and granular samples evenly on the sample pan.
- Grind coarse-grained samples with a mortar, chopping knife or similar. Avoid applying heat to the sample when crushing it as this will lead to moisture loss.
Liquids
The glass fibre round filter has the following advantages:
- Even distribution due to capillary effect
- no drop formation
- faster evaporation due to larger surface area
Note:
For the PCE-MA series, glass fibre round filters are not available from PCE
Instruments. Filters
with a diameter of approx. 80 mm are required.
Settings
After switching it off and back on, the moisture analyser has saved the
last drying parameters
used. These flash in the display.
Selection of the heating programme
Press the menu key to access the settings. Here you can call up existing
modes as well as
create and save new modes.
A flashing number appears on the left of the display to represent the memory location. Here you can retriebve up to 15 different memory locations using the arrow keys.
[The numbers 1 … 9 and the letters A … F are available].
ttention! “0” cannot be selected. When confirming this, two signal tones sound
in quick
succession and the scales automatically exit the menu.
The saved drying parameters of the individual memory locations flash in the
display together
with the memory location. If you want to save the memory location with its
settings, confirm
this with the tare key. The scales beep twice and switch to drying mode.
If you want to change the settings, confirm the memory location with the enter key and you can change the parameters. The settings are described in the next chapter of the instructions.
Setting the drying parameters
After you have selected a memory location in the selection menu and confirmed it with the enter key, you can now set the individual parameters one after the other.
Heating mode selection
Here you can choose between three heating speeds with the arrow keys. The
current mode flashes.
STANDARD: This mode is the factory setting and is suitable for most samples.
FAST: Within the first few seconds, the instrument heats beyond the set temperature to generate a base heat in the heating chamber. With the help of this, the set temperature can be reached more quickly.
SLOW: In this mode, the temperature is increased slowly to protect the
sample. Thus,
this mode is particularly suitable for relatively dry materials.
Termination criterion / stop mode
The arrow keys can be used to select between three abort criteria for the
measurement. The current mode flashes.
AUTOMATIC: This mode ends the measurement automatically when the measured
value is
constant over a certain period of time.
The following chart shows the conditions that must be met to end the measurement.
Stop conditions <period (weight change)>
| Fast heating mode| Standard heating mode| Slow heating
mode
PCE-MA 110| 36 s (<20 mg)| 45 s (<20 mg)| 60 s (<20 mg)
PCE-MA 202| 36 s (<20 mg)| 45 s (<20 mg)| 60 s (<20 mg)
PCE-MA 100| 36 s (<3 mg)| 45 s (<2 mg)| 60 s (<2 mg)
PCE-MA 200| 36 s (<3 mg)| 45 s (<2 mg)| 60 s (<2 mg)
ATTENTION: In this mode, no time setting for the measurement is possible.
The stop conditions cannot be changed.
MING: This mode ends the measurement as soon as the set time has elapsed.
is.
ATTENTION: The maximum measuring / drying time that can be set is
199 minutes. MANUAL In this mode, the measurement can be stopped by the
user. All you have to do is press the START/STOP key at the desired time.
ATTENTION: Depending on the selected temperature, combustion of the sample
is very likely to occur. This mode should only be used under constant
observation. The advantage of this measurement is that the sample can be left
to dry for longer in order to be able to remove trapped moisture from the
sample.
To confirm the selected heating mode, press the enter key and the scales switch to the next parameter.
Setting the drying temperature
The drying temperature can be entered here using the arrow keys.
First, the placeholder for the tens digit will flash. This can be set from 0
to 19 and it also covers the hundreds digit. Press the enter key to confirm
the entry and the device switches to the units digit. This can be set from 0
to 9 and is confirmed again by pressing the enter key.
ATTENTION: Set temperatures below +40 °C will cause an error message
“LOtEMP” when the device starts drying. When you return to the setting mode to
change the temperature, the instrument will show you the temperature +40 °C as
an indication of the minimum temperature.
Setting the drying time
This item can only be set if you have selected the TIMING termination
criterion in advance.
Again, the setting is made here by using the arrow keys. First, the
placeholder for the tens digit
will flash. This can be set from 0 to 19 and it also covers the hundreds
digit. Press the enter key to confirm the entry and the device switches to the
tens step of the seconds. The seconds can only be set in steps of ten up to
50. Thus, the maximum time that can be set is 199 min and 50 s. By pressing
the enter key, the meter now enters the measuring mode.
Operation
Notes on the selection of settings
If you are working with a moisture analyzer for the first time, it is
advisable to familiarize yourself with the measuring procedure by carrying out
some test measurements. You can use a moistened piece of paper for this
purpose, for example.
Since the moisture analyses do not work with characteristic curves, almost all
materials can be
dried in the device. However, the properties of the sample must be taken into
account. For the correct settings, you must therefore think about the
properties of the sample to be measured.
- How does the sample react to the effects of heat?
- Can the material ash, burn or melt?
- Does the material have a specific ignition / flash point?
- What is the grain size / sample size?
The settings should also be chosen depending on the material properties. It is advisable to make some test measurements with different settings, if necessary, to specify the appropriate ones for the different materials. This means, for example, starting with a low temperature to test how the sample reacts and then slowly increasing to minimize the drying time, for example.
Furthermore, it may also be necessary to take additional measures. If the
sample can evaporate
substances that may be toxic to humans, an appropriate exhaust air system
should be available.
For comparability of measurements, care should be taken to always use the same
sample
quantity and settings and to follow the same sample preparation procedure. The
environmental
conditions can already influence the sample during sample preparation, which
can lead to
deviations in the measurement results. Please observe the specifications for
sample preparation in chapter 4.
**Measurement
**
Preparation of a measurement:
- Place the sample pan on the sample pan holder with the insertion aid. Make sure that the sample pan lies flat on the sample pan holder.
- Tare the sample pan with the tare key.
- Now put the desired amount of your sample into the sample pan.
- Make sure that the sample does not weigh less than 1 g. Depending on the material, a sample weight of 5 to 10 g is advisable.
- Do not close the lid again until the “ο” icon disappears.
- Start the drying process by pressing the START/STOP key. The ongoing drying process is indicated by the flashing of the status indicator “TEST”.
- Depending on the set stop criterion, an audible signal sounds when the drying process is completed.
- The display shows you the moisture content of the sample in %.
- Open the lid and make sure to remove the sample with the insertion aid to avoid burns.
- If you want to make the next measurement directly, press the RESET key to remove the last measured values and prepare a new measurement.
You have the option of switching back and forth between the units during and
after the
measurement with the unit key. If you do this during the measurement, the
current values at the time of the measurement are displayed.
% MC = moisture content in %
g = weight of the sample in g
% = dry matter in %
Attention!
Do not touch the lid during the measurement to avoid burns and scalds!
The hot water vapour escapes through the ventilation slots at the top and
the lid heats up during the measurement.
Advanced settings
Setting the baud rate
Switch on the scales with the on/off key and immediately press the tare key
while the entire display is still lit up. The baud rate stored is now
displayed.
You can now change this with the arrow keys. You can choose between 1200 /
2400 / 4800 /
9600. To accept the desired selection, press the RESET key and the meter
switches to measuring mode
Print measured values
You can print out the measured values with the optional printer PCE-BP1. To do
this, proceed as follows:
-
Check that the printer is correctly set and connected to the scales via the cable.
-
After successfully completing the measurement, keep the menu key pressed until “PrInt” appears on the display.
-
Then the option “L-C” (printed in Chinese) and “L-E” (printed in English) appears in
the display. Use the arrow keys to select the option “L-E” and confirm it with the enter key. -
Now the numbers 1 to 5 appear on the display, representing the last 5 measurements. Use the arrow keys to select the desired measurement and confirm it with the enter key.
-
The data is now sent to the printer via the RS-232 interface.
**Example of a printout:
**
Setting continuous data transfer
Press and hold the menu key until you see a “t” in the display.
Adjustment (calibration)
In the event that the internal balance does not work correctly and / or
noticeable deviations in the measured values occur, you can carry out an
adjustment with appropriate weights. In order to maintain the accuracy of the
balance, it is necessary to use a weight of accuracy class F1 for the
adjustment. A 100 g weight is required for the PCE-MA 110 / 100 and a 200 g
weight for the
PCEMA 202 / 200.
ATTENTION: It is not possible to change the adjustment weight in the
programme!
NOTE! The weight readings in the following illustrations belong to the
calibration of a balance of the model PCE-MA 100.
Press and hold the “RESET” key in weighing mode until – CAL – appears in the
display
The display now changes to show the weight to be used. It flashes until you
have put the weight on
While the display is showing the lines, the scales must not be touched or
exposed to vibration as the weight must stabilise internally.
After successful adjustment, the display shows the weight again.
When you lift the calibration weight, the scales still perform a zero setting.
Again, do not touch the scales or expose them to vibration as the weight must
stabilise internally.
After the zero point setting has also been successfully completed, the display switches to the weighing mode and shows 0.000 g. This completes the adjustment.
Data transfer to the PC
The drying data can also be transferred live to the optionally available PCE-
SOFT-MA software
via the RS-232 interface. For this purpose, the scales must be set to
continuous data transfer (see item 7.3).
The following optionally available cables are required for this: PCE-CK- RS232 + RS232-USB
The necessary system requirements for the PC are:
- Operating system Windows 10 or higher
- USB port (2.0 or higher)
- An installed .NET Framework 4.6
- A minimum resolution of 1200×600 pixels
- Processor with 2 GHz
- 4 GB RAM memory recommended
NOTE!
The exact description for the correct set-up of the data transfer can be found
in the software manual.
Printer connection
The RS-232 interface can be used to connect the optional PCE-BP1 printer. To
do this, the baud rate of the scales must be adjusted on the printer:
**Baud rate 1200
Baud rate 9600
**
1200-2400-4800-9600
SW1 -ON-OFF-ON-OFF
SW2- ON-ON-OFF-OFF
SW3 -ON-8 bit data (no paraty)
-OFF -7 bit data
SW 4 -ON-parity checked
-OFF-ignore parity
SW5 -ON-even parity
-OFF – odd parity
SW 6- ON-80 characters/line
-OFF – 40 characters/line SW7-ON-CR->CR+LF
OFF- CR->CR
SW8- ON double parallel print
-OFF normal print
Pin assignment of the connection cable on the scales:
SUB-D 9 connector for RS-232 interface
on the PCE-MA series
MAS60 930 017-517 / DIN 45 322
connector for interface on printer
Troubleshooting
**Error message** | **Cause** | **Possible solution** |
---|---|---|
**** | Weight of the sample below 1 g | Use a samplequantity of approx. 5 … 10 |
g
| Temperature setting below+40 °C| Change the settings
| Measuring time below 30 seconds| Change the settings / restart the
scales
| Time setting for slowheating mode “SOFT” below 3 minutes| Change the
settings / restart the scales
| Problem with the halogen lamp| Contact PCE Instruments
**** Black display| Power supply interrupted ordown arrow key was pressed in
measuring mode.| Check the power supply,press up arrow key in measuring mode
Contact
If you have any questions, suggestions or technical problems, please do not
hesitate to contact us. You will find the relevant contact information at the
end of this user manual.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in
accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.
User manuals in various languages can be found by using our product search
on: www.pce-instruments.com
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- Industrial Measurement Products and Solutions | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>