APERA WS100 Fluoride/pH Portable Meter Kit Instruction Manual

June 3, 2024
APERA

WS100 Fluoride/pH Portable Meter Kit

WS100 Fluoride/pH Portable Meter Kit
Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
– 1 –

Inhalt:
1 Overview ………………………………………………………………………………………………………………- 3 1.1 Basic Functions ……………………………………………………………………………………………….- 3 1.2 Features of Fluoride Measurement……………………………………………………………………..- 3 1.3 Features of pH measurement …………………………………………………………………………….- 3 –
2 Technical Parameters ………………………………………………………………………………………….- 3 2.1 Fluoride ………………………………………………………………………………………………………….- 3 2.2 pH …………………………………………………………………………………………………………………- 4 2.3 mV…………………………………………………………………………………………………………………- 4 2.4 Other technical parameters: ………………………………………………………………………………- 4 –
3 Instrument Description ……………………………………………………………………………………………- 4 3.1 LCD Display…………………………………………………………………………………………………- 4 3.2 Operation keys ………………………………………………………………………………………………..- 5 3.3 Data Storage, Recall Data, and Clear Data: …………………………………………………………- 5 –
4 Fluoride concentration measurement ………………………………………………………………………..- 6 4.1 Preparation……………………………………………………………………………………………………..- 6 4.2 Calibration ………………………………………………………………………………………………………- 6 4.3 Sample Measurement ………………………………………………………………………………………- 6 4.4 Parameter Settings…………………………………………………………………………………………..- 7 4.5 Notes……………………………………………………………………………………………………………..- 7 –
5 pH Measurement……………………………………………………………………………………………………- 8 5.1 Preparing the Meter………………………………………………………………………………………….- 8 5.2 Meter Calibration ……………………………………………………………………………………………..- 8 5.3 Sample Measurement ………………………………………………………………………………………- 9 5.4 Parameter Settings…………………………………………………………………………………………..- 9 5.5 Notes…………………………………………………………………………………………………………… – 11 5.6 Self-diagnose Information………………………………………………………………………………..- 13 –
6 ORP Measurement……………………………………………………………………………………………….- 13 6.1 Sample Test…………………………………………………………………………………………………..- 13 –
7 Complete Kit ……………………………………………………………………………………………………….- 13 8 Warranty …………………………………………………………………………………………………………….- 14 9 Appendix 1. Fluoride Concentration Conversion Table ……………………………………………….- 14 –
– 2 –

1 Overview
Thank you for purchasing the Apera Instruments WS100 Fluoride/pH Portable Meter Kit (Hereon after referred as the meter). This fluoride/pH combo meter is widely used for drinking water, surface water and other water treatment applications, ideal for both in-field and laboratory testing. Before using the product, please read this manual carefully before use to help you properly use and maintain the instrument.
For technical support, please contact us at info@aperainst.de or +49 202 51988998 1.1 Basic Functions
· Automatic calibration and temperature compensation (ATC), data storage, self-diagnostics, automatic shutdown, low-battery reminder
· Adopting digital filtering and intelligence to improve the instrument’s responsiveness and accuracy of measurement data, measured value display when stable ” ” Icon.
· Meets IP57 Waterproof and dustproof rating, ideal for use under harsh environment.
1.2 Features of Fluoride Measurement · Patented structure of the 3-in-1 ion electrode contains: ion-selective electrode, reference electrode, and temperature electrode. This design generates quick stability, fast response, and is easy to use. · no need for stirring or added reagents, directly measuring the ion concentration of fluoride in a quick and simple manner, directly measure the fluoride concentration with three different units: mg/L, ppm and pF.

1.3 Features of pH measurement · 1-3 points of automatic calibration, with calibration icons and self-diagnosis function. · Automatic recognition of up to 12 different pH standard buffer solutions, there are three series of standard buffer solutions to choose from: U.S.A, NIST, and CH.

2 Technical Parameters

2.1 Fluoride Measuring range Display units Accuracy

0.02 ppm to 1900 ppm mg/L, ppm, pF
±0.02 ppm or ±5% of reading (which ever is greater)

Temperature Compensation Range

( 0-80 ) ( Automatic)

– 3 –

2.2 pH

Measuring range Resolution Accuracy

Temperature Compensation Range

( -2.00 – 19.99) pH 0.1/0.01 pH
± 0.01 pH ± 1 digit
(0 ­ 100) °C (Automatic)

2.3 mV Measuring range Resolution Accuracy

-1999mV ~ 0 ~ 1999mV 1mV
±0.1% FS

2.4 Other technical parameters:

Data Storage
Storage Content
Power Supply
Size and Weight Quality and Safety
Certification IP Class

300 Groups Numberings, measurements, units of measurement and a
temperature value AA Battery (1.5VX2) Instrument: (65X120X31) mm/180g Case: (360X270X76) mm/1.5Kg ISO9001:2015, CE
IP57 Dustproof and waterproof

3 Instrument Description
3.1 LCD Display —- Parameter mode —- Measuring value —- Measured value storage and display of numbers, symbols indicate: M+ — Store Measured value RM – Recall measurement values Icon —- Measurement unit —- Temperature measurement and unit —- Smiley face; icon indicates stable readings —- Instrument calibration indicator —- Low battery reminder
– 4 –

3.2 Operation keys

3.2.1

— Switch on / off key

3.2.2

— Calibration key

(a) When in the measurement mode, press this key to enter calibration mode. (b) When in the parameter setting mode, press this key to change the number or ON/OFF state.

3.2.3

Function key

(a) In pH measurement mode, short press to change measurement parameters, long press this key to enter parameter setting mode P1, P2, P3 ……
(c) In Fluoride measurement mode, press this key to enter parameter setting mode P1, P2, P3 ……

3.2.4

Backlight and enter key

(a) In measurement mode, short press (< 1.5s) to turn the backlight display on or off; (b) In calibration mode or the parameter setting mode, press this key to confirm, and enter
measurement (c) In pH measurement mode, press this key to change resolution simultaneously: (d) 0.010.1 pH (e) In ion measurement mode, press this key to change parameter unit: mg/L ppm pX

3.2.5

The combination key of memory storage and recall memory.

(a) In the measurement mode, short press this button (< 1.5s) to save the measuring data. Long press this button(2s) to display the stored measuring data.
(b) In the parameter setting mode, press this button to change the unit % mol g/cm3 g/L or turn the meter ON/OFF.

3.3 Data Storage, Recall Data, and Clear Data:

3.3.1 Data Storage

While in the measurement mode, when the measuring data is stable and the smiley face icon

appears, short press

key(1.5s), LCD will display “M+” icon and a storage serial, and

meanwhile storing all of the measuring information. It can store up to 300 groups. 3.3.2 Display measuring information

(a) While in the measurement mode, long press

(2s), meter will display the last stored

information, and the storage number and “RM” icon will appear in the lower right corner of

the LCD, and the complete measuring information. Again short press

or

meter

will in turn display all the measuring data, Hold

or

to quickly check the

measuring data. (b) In the data displaying mode (there are “RM” and storage number in the upper right corner of

the LCD), press

key to return to the measuring mode.

– 5 –

(c) Clear the stored measuring information: In the data displaying mode, press the

key

for 5 seconds, LCD will appear CLr for 2 seconds. It means that all the storage has been eliminated and returns to measuring mode.

4 Fluoride concentration measurement

4.1 Preparation

4.1.1 Press

key to start connect F501-S fluoride ion electrode; (pay attention to inserting

direction), the meter automatically switches to the fluoride concentration measurement mode.

4.1.2 Hold

to change fluoride concentration unit from mg/l, ppm to pX, release the button

after selection. 4.1.3 If it is a new electrode or if it is your first time use, activate the electrode by soaking it in
19.00 mg/L calibration solution for 1 hour. For regular usage, soak 10 to 20 minutes to activate

it. When the reading is fully stabilized ( stays on screen), the electrode activation is

finished. Then you can start calibration and measurement. (Please note that the reading during activation is random).

4.2 Calibration

4.2.1 Press

, LCD displays ” 0.19 mg/L “, and LCD top right corner CAL and the bottom

left L will be flashing, indicating that the meter has entered the 1th point calibration mode and

the calibration solution is 0.19 mg/l; Rinse the fluoride-ion electrode in distilled or deionized

water and shake off the excess water; dip the electrode into 0.19 mg/L calibration solution, stir

gently and let it stand still.

4.2.2 When the reading is stable and

icon stays on screen, then press

key again to

finish 1st point calibration. End and 19.0 mg/L will be displayed after a few seconds. CAL and H will be flickering on the screen, indicating the meter has entered the 2nd point calibration mode and the calibration solution is 19.00 mg/L. 4.2.3 Take out the electrode and shake off excess solution, then dip it into 19.00 mg/L calibration

solution, stir gently and let it stand still. When the reading is stabile and stays on screen,

then press

key again to finish 2nd point calibration. After a few seconds End shows up,

indicating the end of calibration; LCD’ left corner shows two-point calibration icons L and H; The

instrument automatically enters measurement mode.

4.3 Sample Measurement 4.3.1 Rinse the fluoride-ion electrode in distilled or deionized water and shake off excess water or
bolt dry with filter paper. Soak the electrode into your sample solution, stir gently and let it stand
still; get the measurement when the reading is stable and stays on screen.

4.3.2 Hold

to show Fluoride concentrations in different units, release after selection.
– 6 –

4.3.3 Short press

key, the meter enters the mV mode and displays the mV value of the

fluoride ion concentration. 4.3.4 After test, rinse off the electrode with distilled water, shake off excess water, and store dry
with the cap on.

4.4 Parameter Settings

Table (4-3)

Symbol

Parameter sets the project

Code

P1

Temperature unit selection

P2

Backlight display time setting

P3

Auto power off time setting

P4

Restore to factory default settings

Parameters °C / °F
0-1-3-6 min 0-10-20 min
OFF-ON

4.4.1 Temperature units °C /°F selection (P1)

Long Press

to enter P1remaining operations see section 5.4.5

4.4.2 Backlight display time setting (P2)

In P1, long press

to enter P2remaining operations see section 5.4.6

4.4.3 Auto-off time setting (P3)

In P2, press

to enter P3, remaining operations see section 5.4.7

4.4.4 Restore factory default (P4)

In P3, press

to enter P4, remaining operations see section 5.4.8

4.5 Notes 4.5.1 The meter does not automatically recognize calibration solutions. Please make sure to use
correct solution according to the displayed reminders “0.19 mg/L ” or ” 19.00 mg/L “. 4.5.2 The calibration solution can be easily polluted, especially for low concentration solution (0.19
mg/l). Please pay attention to avoid contamination otherwise a large measuring error will occur. Replace the solution after 4~5 times of use. 4.5.3 When storing the electrode, just rinse it out and shake off excess water, then store dry with the electrode cap on. 4.5.4 When testing solutions in the order from low-concentration to high-concentration, the electrode does not need to be rinsed. Just shake off excess water or blot dry with filter paper. If it’s from high-concentration to low-concentration, you need to rinse the electrode with distilled or deionized water, otherwise it will affect the accuracy significantly. 4.5.5 Typically the meter needs to calibrated before use. But when requirement of accuracy is low, calibration is not necessary. Or try the electrode in the calibration solution, if the error is very small, calibration is not necessary. 4.5.6 According to the theory of Isothermal measurements of ion-selective electrode, the smaller
– 7 –

the difference of temperature between test solution and the calibration solution, the higher the measuring accuracy is. Typically, temperature difference between test solution and calibration solution should ±15°C; for high accuracy measurement, the temperatures should be the same. 4.5.7 When the meter is functioning abnormally, please set P4 as “ON” to restore the meter to factory default mode, and then perform calibration and measurements again.

5 pH Measurement

5.1 Preparing the Meter Connect the 201T-S pH Electrode to the meter; (pay attention to the inserting direction); loosen

the cap of the storage bottle of electrode; screw off the storage bottle; Press

to turn on

meter. The meter automatically switches to pH measurement mode. 5.2 Meter Calibration 5.2.1 Use distilled water to rinse off electrode and then dry it with clean tissue, then soak it into the
pH7.00 buffer solution, stir the electrode gently and let it stand still,

5.2.2 Press

to enter calibration mode, LCD displays C1, indicating the first point of

calibration; Wait for a stable reading and to

appear and stay, then press

again, the

LCD will show a blinking 7.00 pH, indicating the 1st point of calibration is completed; a stable pH value and a blinking C2 will be displayed, indicating the first point calibration has been finished and the meter enters the second point of calibration.

5.2.3 Rinse the pH electrode in distilled water and dry it, then soak it into the pH4.00 buffer solution, stir the electrode gently and let it stand still, waiting for a stable reading and a smiley

face icon to appear and stay, then again press

, the LCD will show a blinking 4.00 pH, a

stable pH value and a blinking C3 will be displayed, indicating the second- point calibration has been finished and the meter enters the third-point calibration. 5.2.4 Rinse the pH electrode in distilled water and dry it, then soak it into the pH 10.01 buffer solution, stir the electrode gently and let it stand still, waiting for a stable reading and a smiley

face icon to appear and stay, then again press

, the LCD will

show a blinking 10.01 pH. A stable reading and L M H calibration icons will be displayed, see picture (4-1), indicating the three-point calibration has been finished and the meter automatically returns to measuring mode.

5.2.5 Notes (a) This meter can perform one-point, two-point or three-point

automatic calibrations. After the first point calibration (see item 5.2.2.), press

to confirm

one-point calibration and return to measuring mode. The indication icon M standing for one-point calibration will appear on the lower left corner of LCD. When the requirement of accuracy is ±0.1pH, users can perform a one-point calibration with the pH7.00 buffer solution.
– 8 –

(b) After the second-point calibration, (see item 5.2.3.), press

to confirm two-point calibration

and return to measuring mode. The indication icon L and M for two-point calibration will appear on the lower left corner of LCD. User can choose pH4.00 and pH7.00 to calibrate if the measurement is within the acidity range and choose pH7.00 and pH10.01 to calibrate if within the alkalinity range. (c) Users should perform a three-point calibration to reach a more accurate measurement if the measuring range is wide, or if the electrode has been used for long or has aging phenomenon. If it’s a new electrode, a three-point calibration is necessary to adjust the meter’s slope to match the pH electrode.

5.3 Sample Measurement Soak the pH electrode into the sample solution after rinsing and drying, hold the electrode still. Take the readings when the smiley face icon shows up and stays on screen (indicates a stable reading). pH isothermal measurement principle According to the theory of Isothermal measurements, the smaller the difference of the temperatures between test solution and the calibration solution, the higher the measuring accuracy is. Typically, temperature difference between test solution and calibration solution should ±15°C; for high accuracy measurement, the temperatures should be the same.

5.4 Parameter Settings

5.4.1 pH measuring parameter setting (Chart (5-1))

Prompt Mark

Parameter Setting Items

P1

pH buffer solution series selection

P2

Purified water pH temperature compensation setting

P3

Ammonia added purified water pH temperature compensation setting

P4

Temperature unit setting

P5

Back light display time setting

P6

Auto power off setting

P7

Restore to Factory Default

Code SOL PU1 PU2
BL AC

Parameters
USA(U.S.A series) NIS (NIST series) CH (China series)
OFF-On
OFF-On
0-1-3-6min
0-10-20min
OFF-On

5.4.2 pH buffer solution series selection (P1)

(a) Press

key, meter enters P1 mode: see picture (4-2).

(b) Press

or

key to choose buffer solution series:

U.S.A series –1.68, 4.00, 7.00,10.01 and 12.45 pH NIST series — 1.68, 4.01, 6.86, 9.18 and 12.45 pH China series — 1.68, 4.00, 6.86, 9.18 and 12.46 pH

– 9 –

(c) Press

key to enter next parameter setting or press

key to confirm and return

to measuring mode. 5.4.3 Purified water pH temperature compensation setting (P2)

(a) Short press

key in the mode P1, the meter enters mode

P2, see picture (4-3).

(b) Press

or

to choose ON or OFF.

(c) Press

key to enter next parameter setting or press

key to confirm and return to measuring mode. (d) The default setting is ” OFF “. Note: A PU-1 icon will appear in the upper right corner of the LCD if purified water temperature compensation function is set by the user.

5.4.4 Ammonia added purified water pH temperature compensation setting (P3)

(a) Short press

key in mode P2 to enter mode P3, see picture (4-4).

(b) Press

or

key to choose ON or OFF

(c) Press

to enter next parameter setting or press

key to confirm and return to measuring mode. (d) The default setting is ” OFF “. Note: A PU-2 icon will appear in the upper right corner of the LCD if purified water temperature compensation function is set by the user.

5.4.5 Temperature unit / setting (P4)

(a) Short press

key in mode P3 to enter mode P4, see picture (4-5)

(b) Press

or

key to choose temperature unit: °C or °F

(c) Press

key to enter next parameter setting or press

key to confirm and return to measuring mode.

5.4.6 Backlight display time setting (P5)

(a) Short press

key in mode P4 to enter mode P5, see

picture (4-6).

(b) Press

or

key to choose the time of back light

auto power off: 0 min, 1 min, 3 min or 6 min. The back light display function will be closed if choosing a 0min.

(c) Press

key to enter next parameter setting or press

key to confirm and return to measuring mode. (d) The producer setting for P5 is 1min.
– 10 –

5.4.7 Auto power off time setting (P6)

(a) Short press

key in mode P5 to enter mode P6, see picture (4-7).

(b) Press

or

key to choose the time: 0min, 10min or

20min. The auto power off function will be closed if choosing 0min.

(c) Press

key to enter next parameter setting or press

key to confirm and return to measuring mode. (d) The producer setting for P6 is 10min.

5.4.8 Restore to producer setting (P7)

(a) Short press

key in mode P6 to enter mode P7, see

picture (4-8).

(b) (b) Press

or

key to choose “ON”, means that all

the pH measuring parameters have been restored to the factory default setting.

5.5 Notes 5.5.1 The frequency that you need to calibrate your meter depends on the tested samples,
performance of electrodes, and the requirement of the accuracy. For high- accuracy measurements ( ±0.02pH), the meter should be calibrated before test every time; For ordinaryaccuracy measurements (±0.1pH), once calibrated, the meter can be used for about a week or longer. In the following cases, the meter must be re-calibrated before testing: (a) The electrode hasn’t been used for a long time or a new electrode is connected. (b) After measuring strong acid (pH<2) or strong base (pH>12) solutions. (c) After measuring fluoride- containing solution and strong organic solution (d) There is a significant difference between the temperature of the test sample and the
temperature of the buffer solution that is used in the last calibration.
5.5.2 The soaking solution contained in the supplied protective bottle is used to maintain activation in the glass bulb and junction. Loosen the capsule, remove the electrode and rinse in distilled water before taking a measurement. Insert the electrode and tighten the capsule after measurements to prevent the solution from leaking. If the soaking solution is turbid or moldy, replace the KCL solution (Product# AI1107).
5.5.3 For calibration accuracy, the pH of the standard buffer solution must be reliable. The buffer solution should be refreshed often, especially after heavy use. We recommend 10-15 times of use before replacing the pH buffers.
– 11 –

5.5.4 Always keep the meter clean and dry; especially around the socket connecting the meter and electrode, otherwise it may lead to inaccurate measurements or a damaged probe/meter. To clean, dry them with medical cotton and rubbing alcohol if the connectors appear dirty.

5.5.5 The sensitive glass bulb at the front of the combination electrode should not come in contact with hard surfaces. Scratches or cracks on the electrode will cause inaccurate readings. Before and after each measurement, the electrode should be rinsed with distilled water and dried. If a sample sticks to the electrode or it’s contaminated, the electrode should be thoroughly cleaned using a soft brush with soap water and then rinsed with distilled water. After that, soak it in the KCL solution again for 6 hours.

5.5.6 How to renew the pH electrode when its response is getting slow: For general contaminants, soak the electrode in Apera’s cleaning solution (AI1166) or detergent water for 30 minutes. Then use a soft brush to remove the contaminants. Afterwards, soak the probe in Apera 3M KCL soaking solution (AI1107) for at least 1 hour. Rinse it off, then re-calibrate the tester before using.

For other special contaminants, refer to Table-8 Clean contaminated glass bulb and junction

Contamination

Cleaning Solutions

Inorganic metal oxide

Dilute acid less than 1mol/L

Organic lipid

Dilute detergent (weak alkaline)

Resin macromolecule Proteinic haematocyte sediment
Paints

Dilute alcohol, acetone, ether
Acidic enzymatic solution (saccharated yeast tablets)
Dilute bleacher, peroxide

(a) Every pH electrode gradually loses its sensitivity and will eventually fail. The lifespan period for the 201T-S pH Electrode is about 1-2 years depending on how it’s used and maintained.
(b) When an abnormal reading appears during calibration or measurement, please set P7 to “ON” to restore the meter to Factory Default settings, then re- calibrate the meter before measurement.
(c) Electrodes that have been used for a long period of time will become aged. Submerge the electrode in 0.1mol/L hydrochloric acid for 24 hours, then rinse the electrode in distilled water, then submerge it in 3M KCL soaking solution for 24 hours. The method to prepare 0.1mol/L hydrochloric acid: dilute 9mL hydrochloric acid in distilled or deionized water to 1000mL.

– 12 –

5.6 Self-diagnose Information When using the meter, the following icons might appear. This is the meter’s self-diagnose information, which helps users identify problems occurring with the meter or the electrode: a) A still display of -2.00 pH or 19.99 pH — this icon appears when the value has surpassed the
measurement range. There will also appear such signs when the electrode is not well connected with the meter or when the electrode is not immerged into the solution. This is a normal phenomenon. b) Err1 — Abnormal electrode zero potential (<-60mV or >60mV) c) Err2 — Abnormal electrode slope (< 85% or

105%) When “Err1” or “Err2” appears, the meter cannot work properly, please take the following steps to check and fix the problem: a) Check if the electrode bulb has air bubble, if has, please shake the electrode with force for a few times to eliminate the air bubble inside the sensor. b) Check the quality of buffer solution, if it is expired or contaminated. If so, please replace with new ones. c) Set the meter to factory default mode then recalibrate it. d) If the meter still cannot recover to the normal state after doing the above steps, please replace a new pH electrode.

6 ORP Measurement

6.1 Sample Test

6.1.1 Press

key to turn on, and short press

to switch to mV mode;

6.1.2 Connect the ORP probe (Product# AI1315, sold separately), soak it into sample solution, stir gently and let it stand still. Get the readings when the smiley face ion appears and stay on the screen.

7 Complete Kit
WS100 Fluoride/pH meter 201T-S 3-in-1 plastic pH/ATC electrode F501 Fluoride combination electrode pH standard buffer solution (pH4.00, pH7.00and pH10.01, 50mL) Fluoride calibration solution (0.19ml/L and 19mg/L, 50mL) Screw driver 3x Spare AA batteries Operation manual Carrying case
– 13 –

8 Warranty
We warrant this instrument to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), accidental damage, improper maintenance, or unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.

9 Appendix 1. Fluoride Concentration Conversion Table

mol/L 1 ×10-3 1 ×10-4 1 ×10-5

pF

mg/L

3.00

19.00

4.00

1.90

5.00

0.19

ppm 19.00 1.90 0.19