PACE 8889-0205 Arm Evac 200 Portable Fume Extractor User Manual
- June 15, 2024
- Pace
Table of Contents
PACE 8889-0205 Arm Evac 200 Portable Fume Extractor User Manual
MANUAL NO. 8881-0955 REV.A
APRIL 1999
Warning
Read the safety and applications information section in this manual thoroughly before installing and using your fume extraction system.
PACE Incorporated retains the right to make changes to specifications
contained herein at any time, without notice.
Contact your local authorized PACE Distributor or PACE Incorporated to obtain
the latest specifications.
The following are registered trademarks and/or service marks of PACE
Incorporated, Laurel Maryland U.S.A. and may only be used to identify genuine
PACE products or services:
Arm-Evac, Cir-Kit, ConducTweez, CRAFT, Flo-0-Sodr, Heat Wave, Hotspot, LapFlo, MST, Micro Portable, MicroChine, MiniChine, Mini-Wave, PACE, Pacenter, Pik- Tip, PETS, Prep-Set, ResisTweez, SensaTemp, Snap-Vac, Sodr-Pen, Sodr-X-Tractor, ST, StripTweez, Thermo-Drive, ThermoFlo, ThermoJet, ThermoPart, ThermoPik, ThermoTweez.
The following are trademarks and/or servicemarks of PACE Incorporated, Laurel Maryland U.S.A. and may only be used to identify genuine PACE products or services:
Sodrtek, Toolnet
GENERAL INFORMATION
This manual will provide the user with the basic knowledge to properly operate
and maintain the PACE Arm-Evac 250 and Arm-Evac 200 Fume Extraction Systems.
If you encounter any difficulty operating your system or have any questions,
call your local authorized PACE dealer or contact PACE Applications
Engineering directly at the United Kingdom Office at
Tel. (44) 01908 277666, Fax (44) 01908 277777, or PACE Product Management at
the U.S. Office at Tel. 301-490-9860,
Fax 301-604-8782.
Warning
Read the safety and applications information section in this manual thoroughly before installing and using your fume extraction system.
INTRODUCTION
The PACE Arm-Evac 250 and Arm-Evac 200 Fume Extraction Systems represent the
latest technology in fume extraction. They feature a heavy-duty, maintenance-
free brushless motor along with a filtration process that includes a pre-
filter for coarse particle removal, and a High Efficiency Particulate Air
(HEPA)/Gas Filler cartridge. The HEP NGas Filter is ideal for use in
applications where high efficiency particulate removal is required with the
additional benefit of a gas media blend to neutralize and adsorb noxious
gases. Optional filters are also available.
The Arm-Evac 250 Fume Extraction System also incorporates an innovative filter condition monitoring system, a self-calibration feature and an RS 232 computer interface for computerized communication.
PACE Arm-Evac 250/200 Fume Extraction Systems use specially designed extraction accessories that can be conveniently mounted on any workbench or tabletop. The Arm-Evac 250 and Arm-Evac 200 systems are designed to extract fumes from one or two work stations.
ENVIRONMENTAL SPECIFICATIONS
Ambient Operating Temperature: Storage Temperature:
Ambient Operating Humidity: Storage Humidity:
0°C to 50°C (32°F to 122°F)
-40°C to 125°C (-40°F to 257°F)
90% relative humidity maximum non-condensing
90% relative humidity maximum non-condensing
Central Filtration Unit
SAFETY AND APPLICATIONS
PACE Fume Extraction systems are designed to reduce the level of harmful contaminants from the work environment and to assist in the achievement of recommended health and safety requirements for local exhaust ventilation and contaminant removal. Be sure to follow all Application and Maintenance guidelines contained herein and precautions contained in other relevant product safety literature (i.e., Material Safety Data Sheets) provided with the substances and equipment producing the fumes to be extracted.
DISCLAIMER
PACE hereby disclaims all responsibility for any personal injury, property
damage, fine, citation or penalty imposed by any government or private entity
which results from any use, misuse or mis-application of this product, failure
of the user to regularly maintain the product according to the recommended
guidelines, or failure to adequately monitor fume extraction exhaust air and
the ambient workplace air for the presence of harmful levels of gases, fumes
and particulates.
Compliance with all applicable environmental and personnel safety regulations is the sole responsibility of the user and adequate self-monitoring of exhaust air released into the atmosphere or the workplace as well as monitoring of the ambient workplace air is strongly recommended. To insure continued effective performance, the following guidelines must be followed.
SAFETY
DANGERS
- PACE Furne Extractors are not intended to be used as a substitute for devices such as personal protective respirators which are designed to remove poorly adsorbed substances including carbon monoxide, methane, hydrogen, acetylene and other gases. Risk of serious injury or death, fire or explosion may result from improper use. If in doubt, consult with your industrial hygienist or PACE.
- Never use PACE Fume Extraction Systems to extract fumes from highly flammable liquids and gases such as Ether, Gasoline (Petrol) and others. Risk of serious injury, death, fire or explosion may result.
- High concentrations of airborne contaminants such as Methyl Ethyl Keytone, Cyclohexone, Ozone and others, when contacted with activated carbon, can undergo oxidation, decomposition or polymerization resulting in exothermic reactions or heat generation. PACE Fume Extraction Systems must not be used to extract fumes from these substances when high concentrations are present.
- Failure to comply with the application and maintenance guidelines, filter replacement schedules, monitoring recommendations and safety guidelines contained herein and in other relevant product safety literature (i.e., Material ,Safety Data Sheets) provided with the substances and equipment producing the fumes to be extracted could result in risk of serious injury, foe or explosion.
WARNINGS
- Do not attempt to clean any PACE fume extraction fillers for reuse. Cleaning the filters will severely damage the filter media and, consequently, the filter’s performance. This will result in unfiltered, contaminated air being returned to the working environment. Unfiltered air can also damage the motor pump.
- Use with inappropriate chemicals or substances, failure to provide regular maintenance or other misuse of your Fume Extraction System may result in contaminated air being recirculated into the work environment.
CAUTIONS
- Failure to supply the correct voltage to the unit will result in damage to the motor.
NOTES
- To ensure the highest level of performance, use only PACE replacement filters in your PACE Fume Extraction System.
- BE SURE the new filter cartridge is installed with the airflow arrow pointing downward.
FILTER CARTRIDGE NOTES
PACE Fume Extraction Systems are equipped with filters designed to capture
particulates and noxious gases and odors that are present in the air being
filtered. The Fume Extraction systems are equipped with combinations of pre-
filters for coarse particle removal, HEPA filters for fine particle removal,
and gas filter media to remove gaseous compounds. Over time, the HEPA filter
will gradually become clogged, impeding air flow through the system, and the
capacity of the gas filter will be reached, reducing its ability to continue
to neutralize and adsorb noxious gases and odors.
Filter Cartridges must be replaced at regular intervals to ensure that the
fume extraction system is operating effectively. These intervals will vary
depending on the type of work being performed, the level of use, and the
amount and composition of airborne contaminants produced.
The exhaust air stream from the central filtration unit as well as the ambient
air in the workplace should be monitored with appropriate and adequate
measuring/detection devices to assure compliance with all applicable Health
and Safety regulations.
Flammable vapors and gases (i.e. Isopropyl Alcohol) are removed by and
accumulate in the gas portion of the Filter Cartridge presenting a potential
fire hazard. Therefore, the user must exercise appropriate precautions when
extracting such fumes or when handling and disposing of filters containing
such flammable substances.
Follow all fire safety and personnel protection guidelines contained in the
Material Safety Data Sheet (MSDS) for the substance(s) producing the extracted
vapors and gases (fumes).
When disposing of used filters, extreme care must be taken to comply with all applicable environmental regulations. Carefully consult the MSDS supplied with the material(s) producing the fume. If in doubt, check with your local Environmental Authority.
Extreme care must be exercised when disposing of a used Filter Cartridges containjng potentially hazardous substances. When disposing of filters, appropriate and adequate personal protective equipment (i.e. gloves, respirators, plastic containment bags, etc.) must be used when hazardous substances such as asbestos, lead, radioactive or biohazard materials may be present. Consult the Material Safety Data Sheet (MSDS) of the materials generating the extracted fume.
INSTALLATION
GENERAL
The Ann-Evac 250 and Ann-Evac 200 have been designed with maximum installation
flexibility in mind. A number of configurations may be used. (See the
“Configurations” section.) The following are recommended guidelines for
setting up your system. If you have any specific questions that this section
does not cover please consult your PACE Representative or call PACE directly.
1. Placement
There are 3 major considerations that need to be addressed for proper placement of the central filtration unit.
- Exhaust Port: This is located on the back of the filtration unit, next to the power inlet. When positioning the unit, make sure that the exhaust port is 1101 blocked or that airflow from the exhaust port is not restricted in any way.
- Unit Location: The filtration unit should be positioned so that it will not block aisleways or impede normal operating traffic or material handling pathways. [t is important to position the unit to allow for proper clearance to remove the filter access cover and replace the filter as needed.
- Proximity to the work area: The filtration unit should be placed as close to the work area as possible. If using a flex hose, it is advantageous to keep the length of flex hose as short as possible. Do not kink or create any sharp bends in the flex hose, as this will reduce airflow.
2. Power
The Arm-Evac 250/200 will plug into a standard 110 volt or 230 volt electrical outlet, depending on the model you purchased. A line cord is provided with each system. Be sure the unit you have is rated for the intended power supply.
3. Self-Calibration Procedure
4. Filters
Before operating the unit, make sure the filters are correctly positioned. The
pre-filter is always located on top, and the direction of airflow is clearly
marked on the filter cartridge. Airflow is always in a downward direction for
the Arm-Evac 250/200.
5. Inlet Caps
If an inlet port is not connected to a collection device, it must be covered
with the inlet cap provided.
6. Accessories
The Ann-Evac 250/200 is designed lo be used with one or two collection accessories (i.e. flex arm, articulated artn, plenum, etc.). ·when using two collection devices, any two may be used together. This provides maximum flexibility in meeting the fume extraction needs in your work environment. Performance can only be guaranteed when using genuine PACE Fume Extraction collection accessories.
CONFIGURATIONS
Connecting the 45mm (1.75’1 and 50mm (2’1 Static-Safe Metal Flex Arms
- Attach the Static-Safe Metal Flex Arm (either size) at a convenient location on the workbench either permanently or with the bench mounting bracket (PIN 8886-0552).
ARM-EVAC200
As the system is used, the filter will eventually become clogged with
collected particulates. If air flow is insufficient to collect the fumes, the
filter(s) should be replaced. Refer to “Filter Replacement”, page 12.
- Turn the power switch to the On (“I”) position.
- Position the Arm-Evac accessory in a convenient and effective extraction location.
OPERATION
ARM-EVAC 250
Self-Calibration Feature
The Arm-Evac 250 contains advanced microprocessor-based sensing technology
which allows for self calibration of the filter condition monitor by
establishing clean filter air flow conditions (that is, 0% filter blockage)
and fully clogged filter flow conditions (that is, 100% filter blockage).
Further, this procedure insures that the air flow sensing process is
consistent with and particular to the accessory configuration array and
filters being used.
Self-Calibration Procedure
NOTE
This procedure must be performed on the Arm-Evac 250 before initial operation,
whenever a filter changeout occurs and whenever a change in system
configuration occurs.
DO NOT make any adjustments to the accessory configurations during calibration as this will disrupt the self-calibration process.
- The normal operating position of the Keyed Switch (item ®) is in a vertical position.
- TI1e Keyed Switch is supplied with 2 identical keys. To start the self- calibration procedure, insert the key into the switch. Tum the Keyed Switch 1/4 turn clockwise (from nom1al vertical position).
- Tum the Power Switch (item®) to the On (“l “) position (Green switch light
illuminates). An audible single beep followed by a double beep signals
initiation of the self-calibration procedure.
The LED lights will sequence from Green to
Amber to Red on the Filter Condition Gauge (item ®) several times. There is no need to adjust the 3 position Speed Controller (item CD); the self calibration mode over-rides the controller.
- The Self-Calibration procedure will require a running time of approximately 2 minutes in order to establishing clean filter air flow (0% filter blockage) levels for all three speed settings as well as fully clogged filter levels.
- When the far right LED remains illuminated, the self-calibration procedure is complete. To initiate normal operation, turn the Keyed Switch 1/4 tum counter-clockwise. The Red LED will turn off and the Green LED will illuminate and normal operation will resume. Adjust the 3 position Speed Controller (item
- as necessary. To assure tamperproof operation, the key can be removed from the switch.
- Daily operations can be initiated by simply turning the Power Switch (illuminates Green when On) to the “On” (” 1 “) position and setting the 3 position Speed Controller (item 0) to the appropriate position. A single beep will sound when the Arm-Evac 250 is turned on.
Arm-Evac 250 Filter Condition Display
The Arm-Evac 250 has a graphical Filter Condition Display which continuously
monitors filter condition. As the system filter(s) collect particulates, LEDs
on the Filter Condition Display will illuminate along the graphical curve of
the display.
Initially, a Green LED will illuminate (at bottom of graph). As the air flow
rate decreases (as the filter(s) start to become clogged), LEDs will
illuminate high on the graphical curve.
lllumination colors will darken to shades of Yellow and finally to Red as the
illumination reaches the top of the graphical curve.
When the first Red LED illuminates, it is indicating insufficient air flow.
Replacement of the filter(s) at this point is highly recommended. Refer to
“Filter Replacement”, page 12.
If the last Red LED (at the upper right of display) illuminates, an alarm will
sound. Immediately, turn system power off (Power Switch) and change filters.
CAUTION
Overheating may occur and cause motor damage if the system is operated with
completely clogged filters.
Optional Benchtop Controller for Arm-Evac 250
An optional Benchtop Controller (PIN 8884-2250) is available for the Arm-Evac
250 to enable convenient access to the three speed flow control switch ( refer
to item Cg), next page), and surveillance of the filter condition display
(refer to item G), next page).
Setup/Operation
Use the following procedure to set up and operate the Remote Control. Refer to
illustration below.
- Attach the end of the Benchtop Controller Cable identified as “250” to the connector located on the back of the Ann-Evac 250 identified as “Remote”. Refer to illustration below.
- Attach the end of the Benchtop Controller Cable identified as “REM” to the connector located on the Benchtop Controller (item ®).
- Position the Benchtop Controller i11 convenient location.
- Once the cables are connected, assume normal operating procedures.
NOTES
Once connected to the Arm-Evac 250, the Benchtop Controller will over-ride all
controls with the exception of the keyed calibration switch and Green lighted
Power Switch.
If the self-calibration procedure is initiated while the Benchtop Controller is connected, the LED Display on the Benchtop Controller will mimic the filter condition gauge (refer to item CD) as identified in the “Arm-Evac 250 Self- Calibration Procedure”.
Arm-Evac 250 RS232C Output for ISO 14000 Software
The Am1-Evac 250 has a microprocessor integral with its electronic control
circuitry which can transfer data to support ISO 14000 documentation
requirements. The output connector (marked “PC COMS “) on the back of the Arm-
Evac 250 and cable connection provides a corrununication link for transfer and
storage of data to a computer.
Contact your local PACE representative or contact PACE directly to obtain detailed operating instructions.
MAINTENANCE
GENERAL TIPS
- Before opening the Arm-Evac, always turn off the power source.
- Protect the Arm-Evac and accessories from dampness.
- Do not extract water vapor or steam.
- The Arm-Evac must not be used without filters. Impurities in the extracted air can damage the motor pump. Do not allow the Arm-Evac or ducting hoses to become blocked.
- Do not extract corrosive compounds without checking with PACE.
- Recalibrate the Arm-Evac 250 whenever a filter change or system configuration change occurs.
FILTER REPLACEMENT
Pre-Filter
This filter removes coarse particles from the air. In general, the pre-filter
must be changed on a monthly basis. However, depending on your particular
process, the filter may need to be changed more or less frequently. A visual
inspection on a weekly basis is recommended.
Main Filters
Fume extraction filters remove particulates from the air and capture noxious
odors and gases. Depending on your particular process, the filter cartridge
may need to be changed more or less frequently.
Replacement Procedure
-
Tum Power Switch to the Off (“O”) position.
-
Remove the Filter access cover.
-
Remove the old Filter cartridge ( or cartridges).
-
Install a new Filter (or filters).
On the Arm-Evac 250, insure that the
Pressure Tube (black, coiled cable inside filter compartment) does not become kinked when the new Filter is installed. -
Replace the Filter access cover.
-
Dispose of used Filters properly.
-
Arm-Evac 250 Systems Only: Recalibrate according to Self-Calibration Procedure (see Page 9).
-
Tum Power Switch to the On (” l “) position.
Important
BE SURE the new filter cartridge is installed with the airflow arrow pointing
downward and Pressure Tube does not become kinked.
FUSES
The fuse is located in a sliding tray on the back side of the Ann-Evac
250/200, just below the power receptacle. Each Ann-Evac is supplied with one
replacement fuse, also located in the sliding tray.
Replacement fuses are available for the Aon-Evac 250/200 systems in packages
of 5 fuses.
230 V AC systems (part numbers 8889-0200 and 8889-0250) come standard with a
3.15 amp fuse, part number 8884-9966-PS.
110 VAC systems (part numbers 8889-0205 and 8889-0255) come standard with an 8 amp fuse, part number 8884-9961-PS.
MOTOR
Fume Extractor Motors are maintenance-free, brushless single inlet fans which
do not require routine maintenance. Tt is important that fume extractor
filters are changed on a regular basis to protect the motor impellers from
pollutants.
ACCESSORIES
Monthly cleaning of the Ann-Evac 250/200 extraction accessories is
recommended. Arm-Evac accessories should be cleaned with a non-aggressive
cleaning solvent or detergent only. Make sure the Ann-Evac accessories are
completely dry before reuse.
REPLACEMENT FILTERS
REPLACEMENT FILTERS
Table 2: Replacement Filters
***** NOTE – The Arm-Evac 250/200 Pre-Filter and Combo HEPA/Gas Filter (items
3 and 4 in the table below) must be used together. NEVER use them separately!