Introduction
The 9760 power amplifier has been designed to complement the Time Electronics range of multifunction calibrators. The unit provides an output current suitable for driving even the most demanding analogue meters.
Operation is simple with connection via the front panel terminals. The reference calibrator (for example 5025C) connects to the input terminals. Then the output is connected to the meter. The user then presses the output button.
The amplifier has a gain of x10. So an input voltage of 1 V will produce an output of 10 V.
Due to the output resistance of the amplifier, the accuracy of the output is specified for loads greater than 600 Ω. If loads less than this are present a reduction of 1 mV per 10 mA of load current will apply
Features
- Compatible with 5051 and 5025 calibrators
- 60 V AC RMS, 100 mA
- 90 V DC, 100mA
- 0.02 % best accuracy
- ACV to 10kHz
- Short circuit protection: 160 mA
Operation
Safety Precautions
High Voltage: The instrument can output high voltage and users should be aware of the dangers involving serious electrical shock.
This instrument is rated as class 1, to protect against electric shock due to fault condition, the IEC C13 power cord supplied, connects the case to the protective ground provided by the local power utility company. An additional PE terminal is provided at the rear for use when the utility ground is unsatisfactory or does not exist. Please consult local regulations.
Mains (Line) Power Supply
The supply power is connected via a standard IEC Euro connector on the rear panel.
The standard voltage supply is 100V – 230V 50/60Hz. There are two protection
fuses mounted in the IEC connector assemble, both are T3.15A slow blow.
Front Panel
Operating Instructions
Ensure the output is off before the reference input, and output are connected.
- The reference input is connected to the lower terminals marked ‘Input’ (see page 7 for input limits).
- The ‘Output’ is connected from the two upper terminals.
- To turn the output on, push the output button.
The output button will illuminate to confirm the output is connected
The amplifier has a gain of x10. So an input voltage of 1V will produce an output of 10V.
Note:
Due to the output resistance of the amplifier the accuracy of the output is
specified for loads greater than 600Ω. If loads less than 600Ω are present a
reduction of 1mV per 10mA of load current will apply.
Example:
The amplifier sourcing 10V into a meter with 100Ω input impedance will result
in 100mA current being drawn.
Therefore, a reduction in output of 10mV will occur
Specifications
Technical Specifications
AC Voltage
Gain: x10
Input Range: 0.5 to 6 V RMS
Output Range: 5V to 60 V RMS
Maximum Output Current: 100 mA
Accuracy: 0.03% + 3 mV
Output Resistance: < 0.1 Q
Frequency Range: 45 Hz to 10 kHz
General Specifications
Short circuit protection: 160 mA
Switches/Indicators. : Output On/Off Switch, Output On indication.
Power Supply: 100 to 230 V AC 50/60 Hz
Dimensions: 430 x 230 x 90 mm .
Weight: 5 kg
Warranty and Servicing
Warranty
Time Electronics products carry a one-year manufacturer’s warranty as standard.
Time Electronics products are designed and manufactured to the highest standards and specifications to assure the quality and performance required by all sectors of industry.
Time Electronics products are fully guaranteed against faulty materials and workmanship. Should this product be found to be defective, please contact us using the below details.
Inform us of the product type, serial number, and details of any fault and/or the service required. Please retain the supplier invoice as proof of purchase.
This warranty does not apply to defects resulting from action of the user such as misuse, operation outside of specification, improper maintenance or repair, or unauthorized modification. Time Electronics’ total liability is limited to repair or replacement of the product.
Note that if Time Electronics determine that the fault on a returned product has been caused by the user, we will contact the customer before proceeding with any repair.
Product Registration
You can register your product at: www.timeelectronics.com/contact/product-
registration.
Registering your product will enable us to maintain a record of purchase for
your warranty.
You can also use the web form to provide feedback about our products and
services.
Calibration and Repair Services
Time Electronics offers repair and calibration services for all the products
we make and sell.
Routine maintenance by the manufacturer ensures optimal performance and
condition of the product. Periodic traceable or accredited calibration is
available.
Contacting Time Electronics
Online:
Please visit www.timeelectronics.com and
select Technical Support from the Contact links. From this page you will be
able to send information to the Time Electronics service team who will help
and support you.
By phone:
+44 (0) 1732 355993
By email:
mail@timeelectronics.co.uk
Returning Instruments
Prior to returning your product please contact Time Electronics. We will issue
a return
merchandise authorization (RMA) number that is to accompany the goods
returning.
Further instructions will also be issued prior to shipment. When returning
instruments, please
ensure that they have been adequately packed, preferably in the original
packing supplied.
Time Electronics Ltd will not accept responsibility for units returned
damaged.
Please ensure that all units have details of the service required and all
relevant paperwork.
Send the instrument, shipping charges paid to:
Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road,
Tonbridge, Kent, TN9 1RA.
United Kingdom.
Tel: +44(0)1732 355993
Fax: +44(0)1732 350198
Email: mail@timeelectronics.co.uk
Web Site:
www.timeelectronics.com
Disposal of your old equipment
- When this crossed-out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002/96/EC.
- All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities.
- The correct disposal of your old appliance will help prevent potential negative consequences for the environment and human health.
- For more detailed information about disposal of your old appliance, please contact your city office, waste disposal service or return to Time Electronics.