React Health Luna G3 Auto CPAP Machine User Guide

June 13, 2024
REACT HEALTH

React Health Luna G3 Auto CPAP Machine

React-Health-Luna-G3-Auto-CPAP-Machine-PRODUCT

Product Information

The product is a device designed for delivering air during treatment. It consists of a main device, a mask, a tube, a power adapter, and a power cord. The device has various button functions, including a home button for navigating menus, a knob for adjusting settings, and a start/stop button for controlling the delivery of air.

Button Functions

  • Home Button: Returns to the previous menu or main interface.
  • Knob: Adjusts device settings.
  • Start/Stop Button: Starts or stops delivering air.

Setting up the Device

  1. Connect one end of the tube to the assembled mask.
  2. Connect the other end of the tube to the air outlet of the device.
  3. If using heated tubing, connect the heated tubing joint to the air outlet of the device, and then insert the power plug of the heated tubing into the heated tubing port on the back of the device.
  4. Connect the power adapter to the DC inlet of the main device.
  5. Connect the power cord to the power adapter.
  6. Plug the power cord into a power outlet.

Adjusting Device Settings

  1. Power on the device to display the Main Interface on the screen.
  2. If using heated tubing, ensure it is connected correctly.
  3. Press the knob to access the Initial Setup Interface of the Patient Menu.
  4. Use the knob to select the option you want to adjust.
  5. Turn the knob to adjust the selected option.
  6. Press the knob to confirm your adjustment.
  7. Use the knob to switch between options and repeat steps 4-6 to adjust other options.
  8. Press the Home Button to exit the Setup Interface.

Button Functions

  • Home Button: Return to the previous menu or main interface.
  • Knob: Adjust device settings.
  • Start / Stop Button: Start / Stop delivering air.

Setting up the DeviceReact-Health-Luna-G3-Auto-CPAP-Machine-FIG-1
\(1\)

  • A. Connect one end of the tube to the assembled mask.
  • B. Connect one end of the tube to the air outlet of the device, as shown in the figureabove. Or Connect the heated tubing joint to the air outlet of the device, and then insert the power plug of the heated tubing into the heated tubing port on the back of the device.
  • C. Connect the power adapter to the DC inlet of the main device.
  • D. Connect the power cord to the power adapter.
  • E. Plug the power cord into the power outlet.

Adjusting Device SettingsReact-Health-Luna-G3-Auto-CPAP-Machine-FIG-1
\(2\) React-Health-Luna-G3-Auto-CPAP-Machine-FIG-1
\(3\) React-Health-Luna-G3-Auto-CPAP-Machine-FIG-1
\(3\)

Starting Treatment

After you finish adjusting device settings and return to the Main Interface, press the Start / Stop Button to start treatment.

Usage score

Usage score of Luna® G3 APAP (LG3600) / Luna® G3 CPAP (LG3500)
After each treatment, Luna® G3 APAP (LG3600) / Luna® G3 CPAP (LG3500) will provide a  Usage Score (for CPAP and AutoCPAP mode only) and Usage Summary on the screen forthe user to get a quick understanding of the usage for the recent one day (from 12:00 noon of previous day to 12:00 noon of the day). The total score is 100 points, with the below breakdown:React-Health-Luna-G3
-Auto-CPAP-Machine-FIG-1 \(6\)

Used time (60 points)
The more time of treatment the user received, the more points the user scores. If the device is used for 30 minutes, get 5 points, or for 1 hour, get 10 points. The user can score up to 60 points.

Used time (min) ≥360 330-359 300-329 60-89 30-59 0-29
Score 60 55 50 10 5 0

Mask seal (20 points)
If the user is wearing the mask well without extra leakage during treatment, then the user can score up to 20 points. If the average leakage is less than 10 L/min, it scores 20 points, and 5 points will be deducted for every 5 L/min increase in leakage.

Leak (L/min) <10 10-14 15-19 20-24 ≥25
Score 20 15 10 5 0

Respiratory events (10 points)
The fewer the respiratory events (apneas and hypopneas), the higher the score. If AHI is less than 5, it scores 10 points, when AHI ≥ 5 and for every 5 increase in AHI, 2 points is deducted.

AHI <5 5-9 10-14 15-19 20-24 ≥25
Score 10 8 6 4 2 0

CONTACT

Luna® G3 APAP (LG3600) / Luna® G3 CPAP (LG3500) Quick Operation Manual V1.0

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

REACT HEALTH User Manuals

Related Manuals