Multiparameter Waterproof Meter

pH, ORP, EC, TDS, Resistivity, Salinity, Seawater σ and Temperature

pH Features

  • Calibration
    • Up to a three-point calibration with five standard buffers and one custom buffer available
  • pH in mV option
    • Useful for diagnostics
  • GLP data
    • Offset, slope, date, time and buffers used
  • Automatically temperature compensated readings
  • pH or pH/ORP field replaceable sensors
    • Gel filled and maintenance free
    • Double junction for reduced contamination of reference cell

EC/TDS/Resistivity Features

  • Calibration
    • Single-point calibration from six standards
  • Temperature compensation
    • Automatic Temperature Compensation
    • Configurable temperature coefficient range from 0.00 to 6.00%/°C
    • Choice of reference temperatures at 20 or 25°C
    • Absolute conductivity can be displayed along with the temperature compensated value
  • Auto-ranging
  • Salinity readings
    • Practical Salinity Scale (PSU) based on conductivity calibration

The HI981954 is a waterproof portable logging multiparameter meter that monitors up to 9 deferent water quality parameters. It’s multi-sensor probe allows for the measurement of key parameters including pH, ORP, conductivity, and temperature. The probe transmits readings digitally to the meter, where data points can be displayed and logged. The complete system is simple to setup and easy to use.

  • Optional shockproof silicon rubber boot
    • Specially designed to protect your instrument from damage or impact

HI710034 Orange

Backlit Graphic LCD Display
The HI981954 features a backlit graphic LCD with on-screen help and the capability to display up to nine

parameters simultaneously.

The graphic display allows for the use of virtual keys to provide for an intuitive user interface.

Waterproof Protection

The meter is enclosed in an IP67 rated waterproof casing and can withstand immersion in water at a depth of 1 m for up to 30 minutes. The probe features an IP68 rating for continuous immersion in water.
Waterproof Protection

Quick Connect Digital Probe
The HI7698195 probe features a Quick Connect DIN connector that makes a waterproof connection with the meter.

Color Coded, Field Replaceable Sensors
Sensor replacement is quick and easy with field replaceable, screw type connectors that are color coded for easy sensor port identification.

Standard or Quick Calibration
Quick Calibration provides a speedy, single point calibration for pH and conductivity. Standard pH calibration options are available for calibrating up to three points from a selection of five standard butters and one custom buyer. Conductivity calibration is a single point from six standard selections or one custom standard.

Auto-sensor Recognition
The probe and meter automatically recognize the sensors that are connected. Any ports not used on the probe will not have the parameter displayed or be configurable.

Automatic Temperature Compensation
Integrated temperature sensor allows for automatic temperature compensation of pH and conductivity measurements.
Compensation

GLP Data
HI981954 includes a GLP feature that allows users to view calibration data and calibration expiration information at the touch of a key. Calibration data includes date, time, buyers/ standards used for calibration, and slope characteristics.
Slope Characteristics

Data Logging
The HI981954 allows users to store up to 45,000 continuous or log-on-demand samples with logging intervals from one second to three hours.
Data Logging

Intuitive Keypad

The fitted rubber keypad has dedicated keys for power, backlight, up/down arrows, help and alphanumeric characters. The meter also features two virtual soft keys that navigate the user through the configuration of each parameter, meter setup, and logging of data. The interface is intuitive for any user’s level of experience.
Intuitive Keypad

Dedicated Help Key
Contextual help is always available through a dedicated “HELP” key. Clear tutorial messages and directions are available on-screen to quickly and easily guide users through setup and calibration. The help information displayed is relative to the setting/option being viewed.

PC Connectivity
Logged data can be transferred to a Window’s compatible PC with the included HI920015 micro USB cable and HI9298194 software.

Long Battery Life
The display of the meter has a battery icon indicator to show the remaining power. The meter uses four 1.5V AA batteries that provide up to 360 hours of battery life.
Long Battery Life

Rugged Custom Carrying Case
The HI981954 meter, probe, and all accessories are supplied in a rugged carrying case designed to provide years of use. The inside compartment of the carrying case is thermoformed to securely hold and protect all of the components.

Probe and Sensors
The HI7698195 is a multiparameter pH/EC/Temperature probe for use with the HI981954 portable meter. It features a Quick Connect DIN that makes a waterproof connection with the meter. Sensors are automatically recognized by the probe and meter when connected. Any ports not used on the probe will not have the parameter displayed on the meter. Sensor replacement is quick and easy with field replaceable, screw type connectors that are color coded for easy sensor identification. The probe features a multistring-multiconductor shielded cable with 4m, 10m, 20m, and 40m lengths available. It’s rugged, waterproof design makes it ideal for field use.
Probe and Sensors

Specifications HI7698195
Sensor Inputs two (pH or pH/ORP, EC)
Sample Environment fresh, brackish, seawater
Waterproof Protection IP68
Operating Temperature -5 to 55°C
Storage Temperature -20 to 70°C
Maximum Depth 20 m (66’)
Dimensions (without cable) 342 mm (13.5”); 46 mm (1.8”) dia
Weight (without sensors)` 570 g (20.1 oz.)
Cable Specification multistrand- multiconductor shielded cable with internal

strength member rated for 68 kg (150 lb.) intermittent use
Wetted Materials| Body| ABS
Threads| Nylon
Shield| ABS / 316 SS
Temperature Probe| 316 SS
O-rings| EPDM

HI9828-25 Quick Calibration solution

Calibration Solution

Sensor Specifications HI7698194-0 HI7698194-1 HI7698194-3
Description pH sensor pH/ORP sensor EC sensor
Measurement Type pH, mV (pH) pH, mV (pH), ORP EC

Measurement Range 0.00 to 13.00 pH ; ±600.0 mV 0.00 to 13.00 pH; ±600.0 mV; ±2000.0    0.0 to 200.0 mS/cm;
| | | mV| 0.0 to 400 mS/cm (absolute)
Temperature Range| | -5 to 55°C| -5 to 55°C| -5 to 55°C
Color Code| | red| red| blue
| Tip| glass (pH)| glass (pH); Pt (ORP)| stainless steel electrodes AISI 316
| Glass Type| LT (low temperature)| LT (low temperature)| —
| Junction| ceramic| ceramic| —
| Body| PEI| PEI| ABS/epoxy
| Electrolyte| gel| gel| —
| Reference| double| double| —
Maintenance Solution| | HI70300 (storage solution)| HI70300 (storage solution)| none
Dimensions| | 118 x 15 mm| 118 x 15 mm| 111 x 17 mm
Depth| | 20 m (65’)| 20 m (65’)| 20 m (65’)
| Range| 0.00 to 14.00 pH / ±600.0 mV
| Resolution| 0.01 pH / 0.1 mV
pH / mV| Accuracy| ±0.02 pH / ±0.5 mV
| Calibration| automatic one, two, or three points with automatic recognition of five standard  buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
| Range| ±2000.0 mV
| Resolution| 0.1 mV
ORP| Accuracy| ±1.0 mV
| Calibration| automatic at one custom point (relative mV)
| Range| 0 to 200 mS/cm (absolute EC up to 400 mS/cm)
| manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from
| Resolution 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm; automatic mS/cm: 0.001 mS/cm from 0.000 to 9.999 mS/cm;
EC| 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm
| Accuracy| ±1% of reading or ±1 µS/cm whichever is greater
| Calibration| automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm,
| 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
| Range| 0.0 to 400.0 ppt (g/L) (the maximum value depends on the TDS factor)
| | manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L); automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L); automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
TDS| Resolution| 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
| Accuracy| ±1% of reading or ±1 ppm (mg/L) whichever is greater
| Calibration| based on conductivity or salinity calibration
| Range| 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm
Resistivity| Resolution| dependent on resistivity reading
| Calibration| based on conductivity or salinity calibration
| Range| 0.00 to 70.00 PSU
| Resolution| 0.01 PSU
Salinity| Accuracy| ±2% of reading or ±0.01 PSU whichever is greater
| Calibration| based on conductivity calibration
| Range| 0.0 to 50.0 σt, σo, σ,,
| Resolution| 0.1 σt, σo, σ,,
Seawater σ| Accuracy| ±1 σt, σo, σ,,
| Calibration| based on conductivity or salinity calibration
| Range| -5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K
| Resolution| 0.01°C; 0.01°F; 0.01K
Temperature| Accuracy| ±0.15°C; ±0.27°F; ±0.15K
| Calibration| automatic at one custom point
| Temperature Compensation| automatic from -5 to 55°C (23 to 131°F)
| Logging Memory| 45,000 records (continuous logging or log-on-demand of all parameters)
| Logging Interval| one second to three hours
Additional Specifications| PC Connectivity| via USB (with Hanna PC software)
| Environment| 0 to 50°C (32 to 122°F); RH 100% IP67
| Battery Type / Life| 1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight)
| Dimensions / Weight| 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”) / 400 g (14.2 oz.)
Ordering Information| HI981954 is supplied with HI7698195 multiparameter probe with 4m (13’) cable, HI7698194-1 pH/ORP sensor, HI7698194-3 EC sensor, HI7698295 short protective probe shield, HI9828-20 quick calibration solution, HI76981952 probe maintenance kit, HI7698290 calibration beaker, HI9298194 PC software, HI920015 micro USB cable, batteries (4), quality certificate, and instruction manual in a rugged carrying case with custom insert.

pH and ORP solutions begin on page 2.164; EC, TDS and salinity solutions begin on page 5.26
7Multiparameter www.hannainst.com.