COMPASS HEALTH TENS 7000 Muscle Stimulator and Pain Relief Device Instruction Manual

June 13, 2024
COMPASS HEALTH

COMPASS HEALTH TENS 7000 Muscle Stimulator and Pain Relief Device

COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-
PRODUCT

INTRODUCTION

Chapter 1: INTRODUCTION EXPLANATION OF PAIN

  • Pain is a warning system and the body’s method of telling us that something is wrong. Pain is important; without it abnormal conditions may go undetected, causing damage or injury to vital parts of our bodies.
  • Even though pain is a necessary warning signal of trauma or malfunction in the body, nature may have gone too far in its design.
  • Aside from its value in diagnosis, long-lasting persistent pain serves no useful purpose. Pain does not begin until the coded message travels to the brain where it is decoded, analyzed, and then reacted to.
  • The pain message travels from the injured area along the small nerves leading to the spinal cord.
  • Here the message is switched to different nerves that travel up the spinal cord to the brain.
  • The pain message is then interpreted, referred back and the pain is felt.

EXPLANATION OF TENS
Transcutaneous Electrical Nerve Stimulation is a non-invasive, drug-free method of controlling pain. TENS uses tiny electrical impulses sent through the skin to nerves to modify your pain perception. TENS does not cure any physiological problem; it only helps control the pain. TENS does not work for everyone; however, in most patients it is effective in reducing or eliminating the pain, allowing for a return to normal activity.

HOW TENS WORKS

There is nothing “magic” about Transcutaneous Electrical Nerve Stimulation (TENS). TENS is intended to be used to relieve pain. The TENS unit sends comfortable impulses through the skin that stimulate the nerve (or nerves) in the treatment area. In many cases, this stimulation will greatly reduce or eliminate the pain sensation the patient feels. Pain relief varies by individual patient, the mode selected for therapy, and the type of pain. In many patients, the reduction or elimination of pain lasts longer than the actual period of stimulation (sometimes as much as three to four times longer). In others, pain is only modified while stimulation actually occurs. You may discuss this with your physician or therapist.

IMPORTANT SAFETY INFORMATION

Read the instruction manual before operation. Be sure to comply with all “CAUTIONS” and “WARNINGS” in the manual. Failure to follow in- structions can cause harm to user or device.

INDICATIONS FOR USE

This device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work.

PRECAUTIONS

Chapter 2: CAUTIONS

  1. Federal law allows over-the-counter (no prescription) sales of this device.
  2. Do not use this device for undiagnosed pain syndromes until consulting a physician.
  3. Patients with an implanted electronic device, such as a cardiac pacemaker, implanted defibrillator, or any other metallic or elec- tronic device should not undergo TENS treatment without first consulting a doctor.
  4. Patients with heart disease, epilepsy, cancer or any other health condition should not undergo TENS treatment without first con- sulting a physician.
  5. Stimulation delivered by this device may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax or across the chest because it may cause a cardiac arrhythmia.
  6. Do not place electrodes on the front of the throat as spasm of the Laryngeal and Pharyngeal muscle may occur. Stimulation over  the carotid sinus (neck region) may close the airways, makebreathing difficult, and may have adverse effects on the heart rhythm or blood pressure
  7. Do not place electrodes on your head or at any sites that may cause the electrical current to flow trans-cerebrally (through the head).
  8. This device should not be used while driving, operating machinery, close to water, or during any activity in which involuntary muscle contractions may put the user at undue risk of injury.
  9. Turn the TENS off before applying or removing electrodes.
  10. Isolated cases of skin irritation may occur at the site of electrode placement following long term application. If this occurs, discontinue use and consult your physician.
  11. If TENS therapy becomes ineffective or unpleasant, stimulation should be discontinued until its use is re-evaluated by a physician.
  12. Keep this device out of the reach of children.
  13. TENS 7000 devices have no AP/APG protection. Do not use it in the presence of explosive atmosphere and flammable mixture.
  14. It is recommended not to exceed a current density of 2 mA /cm2,otherwise skin irritations or burns can occur. We recommend that:
    • Please use specific size electrodes provided by manufacturer to avoid skin irritations or burns.
    • For smaller electrodes, the maximum current setting of the waveform should be appropriately reduced.
    • Please place the electrodes carefully, ensure that the entire surface of the electrode has good contact with the skin.

CHAPTER 3: WARNINGS

  1. TENS devices have no curative value.
  2. TENS is a symptomatic treatment, and as such suppresses the sensation of pain, which would otherwise serve as a protective mechanism.
  3. TENS is not effective for pain of central origin (This includes headache).
  4. Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when TENS stimulation is in use.
  5. Caution should be used in applying TENS to patients suspected of having heart disease. Further clinical data is needed to show there are no adverse results.
  6. Electrodes should not be placed over the eyes, in the mouth, from the chest and the upper back or crossing over the heart, or internally.
  7. Do not use while sleeping.
  8. Do not use during pregnancy unless directed by your physician.
  9. Simultaneous connection of a PATIENT to a high frequency surgical device may result in burns at the site of the STIMULATOR electrodes and possible damage to the STIMULATOR.
  10. Operation in close proximity to a shortwave or microwave therapy device may produce instability in the STIMULATOR output.

CHAPTER 4: CONTRAINDICATIONS

  1. Do not use TENS over the carotid sinus (neck) region.
  2. TENS devices can affect the operation of demand type cardiac pacemakers.
  3. Do not use TENS device if you have heart disease without consulting your physician.
  4. Do not stimulate on the site that may cause current to flow trans- cerebrally – (through the head).
  5. Do not apply TENS for undiagnosed pain syndromes until etiology is established.

CHAPTER 5: ADVERSE REACTIONS
Skin irritation and electrode burns are potential adverse reactions. If skin irritation occurs, discontinue use and consult your physician.

CHAPTER 6: GENERAL DESCRIPTION

  • The TENS 7000 is a battery operated pulse generator that sends electrical impulses through electrodes to the body and reaches nervescausing pain.
  • The device is provided with two controllable output channels, each independent of each other. An electrode pair can be connected to each output channel.
  • The electronics of the TENS 7000 create electrical impulses whose intensity, duration, number per second and modulation may be altered with the controls / switches.
  • Press buttons are very easy to use and the large liquid crystal display showing the exact mode and values of parameters are very convenient for patients.

CONSTRUCTION

Chapter 7: CONSTRUCTION

FRONT

  1. LEAD CONNECTOR

  2. INTENSITY CONTROL (ON/OFF SWITCH)

  3. PANELCOVER

  4. LIQUID CRYSTAL DISPLAY

  5. MODE CONTROL

  6. SET CONTROL

  7. INCREMENT CONTROL

  8. DECREMENT CONTROL
    BACK

  9. BATTERY STRAP

  10. BATTERY CASE

  11. BELTCUP
    SIDE

  12. PROTECTIVE COVER

TECHNICAL SPECIFICATIONS

Chapter 8; TECHNICAL SPECIFICATIONS
The technical specification details of TENS 7000 are as follows:

| MECHANISM| TECHNICAL DESCRIPTION
---|---|---
01.| Channel| Dual, isolated between channels
02.| PulseAmplitude| Adjustable, 0-100 mA at 500 ohm load each channel.
03.| Wave Form| Asymmetrical Bi-Phasic Square Pulse
04.| Voltage| 0 – 50V (Load: 500 ohm)
05.| Power source| One 9 Volt Battery .
06.| Size| 10.1cm(L) x 6.1cm(W) x 2.45cm(H)
07.| Weight| 150 grams with battery.
08.| Pulse Rate| Adjustable , from 2 to 150 Hz, 1 Hz/step
09. Pulse Width| Adjustable , from 50 to 300 µs

microseconds ,10 µs /step

10.| Modes| B(Burst), N(Normal) , M(Modulation) , SD1 (Strength Duration), SD2
11.| Burst Mode| Burst rate: Adjustable , 0.5 – 5Hz Pulse width adjustable , 50~300µs Frequency fixed = 100 Hz
12| Normal Mode| The pulse rate and pulse width are adjustable . It generates continuous stimul­ation based on the setting value.
13.| Modulation Mode| Modulation mode is a combination of pulse rate and pulse width modulation.

The pulse rate and width are automatically varied in a cycle pattern. The pulse width is decreased by 50% from its original  setting in 0.5 second, then the pulse rate is decre­ased by 50% from its orignal setting in 0.5 second. Total cycle time is 1 second. In this mode, pulse rate (2~150Hz) and pulse width

(50-300µs) are fully adjustable.

14| SD1 Mode| The SD1 (Strength-Duration) mode consis­ts of automatic modulation intensity and

15

| | pulse width in 40% range. The intensity is always increasing while the pulse width is decreasing and vice versa. The intensity is decreased by 40% while the pulse width is increased by 40% in 5 seconds . In the next 5 seconds , the intensity is increased by 40% while the pulse width is decreased by 40%. Total cycle time is 10 seconds. Pulse rate (2~150Hz) and pulse width (50 ~300µs)

are fully adjustable .

---|---|---
SD2 Mode| The SD2(Strength-Duration) mode consists of automatic modulation intensity and pulse width in 70% range. The intensity is always increasing while the pulse width is decreasing and vice versa .The intensity is decrea­sed by 70% while the pulse width is incre­ased by 70% in 5 seconds . In the next 5 seconds , the intensity is increased by 70% while the pulse width is decreased by 70%. Total cycle time is 10 seconds. Pulse rate (2~150Hz) and pulse width (50 ~300µs)

are fully adjustable.

16| Timer| Adjustable , from 1 to 60 minutes or Contin­uous. Adjustable in 1minute each step from 1to 15 minutes and 5 minutes each from 15 to 60 minutes. Treatment time countdown

automatically.

17

18

| Patient Compliance Meter| This unit can store 60 sets of operation records. Total recorded time is 999 hours.
Low Battery Indicator| A low battery indicator will show up on the LCD when the battery is low.
19| Operating Condition| Temperature:0°~40°C Relative Humidity: 30%~75%

Atmosphere Pressure : 700Hpa~1060Hpa

20| Remark| There may be a +/-5% tolerance of all parame­ters and +/-20% tolerance of amplitude &

voltage

The waveforms of the 5 stimulation modes are as follows.COMPASS-HEALTH-
TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1
\(3\)

REPLACEABLE PARTS

Chapter 9: REPLACEABLE PARTS

The replaceable parts and accessories of TENS 7000 devices are as given below: Except leads, electrodes, battery and battery case cover, please do not try o replace the other parts of a device.

ACCESSORIES

Chapter 10: ACCESSORIES
Each TENS 7000 comes complete with standard accessories and the standard labels as given below:

  • DESCRIPTION Q’TY
  • 40 X 40 mm Adhesive Electrodes 4 pieces
  • Electrodes Leads 2 pieces
  • 9 V Battery, type 6F22 1 piece
  • Instruction Manual 1 piece
  • Carrying Case 1 piece

LABEL

  • Digital TENS Device
    • Model: TENS 7000™
    • Supply Voltage: 9V ==
    • Manufactured for: Compass Health Brands
    • Middleburg Heights, OH 44130
    • www.compasshealthbrands.com.

The label attached to the back of device contains important information about this device- model, sup- ply voltage and caution. Please do not remove.

GRAPHIC SYMBOLS

Chapter 11: GRAPHIC SYMBOLS

  • Type BF applied part
  • Do not insert the plug into AC power supply socket.
  • Timer
  • Low Battery Indicator
  • Increment
  • Decrement
  • Refer to Instruction Manual
  • Equipment capable of delivering output values in excess of 10mA r.m.s. or 10V r.m.s. averaged overany period of 5s.

OPERATING INSTRUCTIONS

Chapter 12: OPERATING INSTRUCTIONS

  1. Insert the 9V battery into the TENS 7000’s battery compartment.Make sure to remove the plastic seal on the 9V battery. Line up the positive and negative terminals on the battery with their cor- responding terminals in the TENS 7000. Make sure that both Intensity control (ON/OFF Switch) knobs are in the off position.
  2. Insert the lead wires into the lead wire sockets on top of the TENS 7000.
  3. Open the electrode package. Then insert each lead wire pin into the pig tail of the electrodes.
  4. Place electrodes on your body surrounding or directly on the area of pain.
  5. Slowly turn on the TENS 7000 by rotating the Intensity control (ON/OFF Switch) knobs.
  6. Select the mode and settings you want.
  7. Slowly increase or decrease the intensity by rotating the intensity control (ONN/OFF) knob clockwise to increase or counter clockwidse to decrease.
  8. After treatment, turn the TENS 7000 off by rotating the Intensity control (ON/OFF Switch) counter clockwise to the zero setting.

PARAMETER CONTROLS

Chapter 13: PARAMETER CONTROLS

PULSE WIDTH (DURATION)
Pulse width is how wide each pulse is. Typically, the higher the pulse width the stronger the stimulation feels. If set high enough, sometimes it will produce a contraction which is typically not what you want to elicit with a TENS unit. Various pulse widths are capable of stimulating different groups of nerve fibers.
PULSE RATE
Pulse rate is also known as pulses per second. It is the number of times a pulse occurs in one second. This setting can be adjusted to excite certain nerves to overcome accommodation effects and will feel the sensation of steady continuous stimulation

TREATMENT MODE
Normal or Conventional TENS offers the practitioners complete control over all the various treatment parameters of the instrument.  Burst Mode is analogous to the Low Rate TENS technique except that frequency individual pulses are replaced by individual “bursts” of 7-10 individual pulses. It is thus a combination of Conventional TENS and Low Rate TENS. In Burst Mode, the treatment frequency is fixed by the instrument and is not adjustable with the Frequency Rate control. Modulated Mode attempts to prevent nerve accommodation by continuously cycling the treatment intensity. When using Modulated Mode increase the intensity only when the unit is at the maximum intensity of the modulation cycle. If the intensity is increased during low-intensity period of the modulation cycle, the patient should increase the intensity slowly until the modulation cycle reaches the maximum to ensure a true maximum intensity output. Strength-duration modulation (SD1 & SD2) consists of alternating modulated amplitude and width so that one parameter is always decreasing while the other is increasing and vice versa. The amplitude decreases from the amplitude control setting and returns to that setting. The width decreases from the width control setting and returns to that setting.
TIME DURATION
The onset of pain relief should occur shortly after the intensity setting has been determined. However, in some cases, pain relief may take as long as 30 minutes to achieve. TENS units are typically operated for long periods of time, with a minimum of 20 – 30 minutes and in some post-operation protocols, as long as 36 hours. In general, pain relief will diminish within 30 minutes of the cessation of stimulation.

Chapter 14: ATTACHMENT OF ELECTRODE LEAD WIRES
The wires provided with the system are inserted into the jack sockets located on top of the device. Holding the insulated portion of the connector, push the plug end of the wire into one of the jacks (see drawing); one or two sets of wires may be used. After connecting the wires to the stimulator, attach each wire to an electrode. Use care when you plug and unplug the wires. Jerking the wire instead of holding the insulated connector body may cause wire breakage.
CAUTION
Do not insert the plug of the patient lead wire into any AC power supply socket.

LEAD WIRE MAINTENANCE

Chapter 15: LEAD WIRE MAINTENANCE
Clean the wires by wiping with a damp cloth. Coating them lightly with talcum powder will reduce tangling and prolong life.

Chapter 16: ELECTRODE OPTIONS
The electrodes are disposable and should be routinely replaced when they start to lose their adhesive nature. If you are unsure of your electrode adhesive properties, order replacement electrodes. Replacement electrodes should be approved for over-the-counter use Follow application procedures outlines in the electrode packaging to maintain optimal stimulation and to prevent skin irritation.

Chanter 17: ELECTRODE PLACEMENT
The placement of electrodes can be one of the most important parameters in achieving success with TENS therapy. Placing the electrodes at least 2″ apart but no more than 6″ apart will help with successful therapy. Every patient responds to electrical stimulation differently and their needs may vary from the conventional settings suggested here. If the initial results are not positive, speak to your physician about alternative stimulation settings and/or electrode placements. Once an acceptable placement has been achieved, mark down the electrodes sites and the settings, so the patient can easily continue treatment at home.
CONTIGUOUS PLACEMENT
This is the most common placement technique. It involves placing the electrodes alongside the area of localized pain site, in such a way as to direct the flow of current through or around the area of pain. In a single channel application, this would involve placing each pad on either side of the pain site if the pain is localized on a limb and deep within the tissue. Pad placement on the posterior and anterior aspects of the affected limb will allow the current to flow completely through the limb and thus through the endogenous pain site.

With a two channel application, you may either direct the current flow to cross through the pain site or, in what is called the “bracket” method allowing the current flow on either side of the painful area, generally through the nerve branches that feed into the pain site.
Chapter 18: TIPS FOR SKIN CARE
To avoid skin irritation, especially if you have sensitive skin, follow these suggestions:

  1. Wash the area of skin where you will be placing the electrodes, using mild soap and water before applying electrodes, and after taking them off. Be sure to rinse soap off thoroughly and dry skin well.
  2. Excess hair may be clipped with scissors; do not shave stimulacon arcan
  3. Wipe the area with skin preparation wipes. Let the area dry. Apply electrodes as directed.
  4. Many skin problems arise from the “pulling stress* from adhesive patches that are excessively stretched across the skin during application. To prevent this, apply electrodes from centre outward; avoid stretching over the skin.
  5. To minimize “pulling stress”, tape extra lengths of lead wires to the skin in a loop to prevent tugging on electrodes.
  6. When removing electrodes, always remove by pulling in the direction of hair growth.
  7. It may be helpful to rub skin lotion on electrode placement area when not wearing electrodes.
  8. Never apply electrodes over irritated or broken skin.

Chapter 19 : APPLICATION OF RE-USABLE SELFADHESIVE ELECTRODES
Application

  1. Clean and dry the skin at the prescribed area thoroughly with soap and water prior to application of electrodes.
  2. Insert the lead wire into the pin connector on the pre-wired electrodes.
  3. Remove the electrodes from the protective liner and apply the electrodes firmly to the treatment site. Make sure that the unit is turned off prior to applying the electrodes.

Removal

  1. Turn off the unit prior to removing the electrodes.
  2. Lift at the edge of the electrodes and peel; do not pull on the lead wires because it may damage the electrodes.
  3. Place the electrodes on the liner and remove the lead wire by twisting and pulling at the same time.

Care and Storage

  1. Between uses, store the electrodes in the resealable bag in a cool dry place.
  2. It may be helpful to improve repeated application by spreading a few drops of cold water over the adhesive and turn the surface up to air dry. Over Saturation with water will reduce the adhe- sive properties.

Important

  1. Do not apply to broken skin.
  2. The electrodes should be discarded and re-ordered from your physician when they are no longer adhering.
  3. The electrodes are intended for single patient use only.
  4. If irritation occurs, discontinue use and consult your physician.
  5. Read the instructions for use of self-adhesive electrodes before application.

Chapter 20 : ADJUSTING THE CONTROLS

  1. Panel Cover : A lid covers the controls for selecting mode and adjusting settings. Your medical professional may wish to set these controls for you and request that you leave the cover in place.COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(14\)

  2. Power On/Off: Switch and Intensity Controls: If both controls are in the off-position, the device is switched off.  By turning the controls clockwise, the appropriate channel isswitched on and the indicator of power (CH1 or CH2) will reveal on the LCD. The current strength of the impulses transmitted to the electro- des increases further when the control is turned clockwise. To reduce the current strength or switch the device off, turn the control counter clockwise to the required setting or off-position, respectively. The controls are protected by a cap to avoid unintentional change of intensity.COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(15\)

  3. Lead Connector: The connection of the electrodes is made with the two lead wires. The device must be switched off before connecting the cables. Both intensity controls must be at the Off position. Electrodes must be pressed firmly on the skin.COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(16\)

  4. Mode Control
    There are 5 modes available – Burst, Normal, Modulation, SD1 and SD2. The mode can be selected by pressing the “MODE” control. seI

  5. Set Control
    By pressing the “SET” control, you may enter the setting you intend to make adjustment. You may start to set the value by pressing the “Increment” and “Decrement” controls when the value is flashing.

  6. Increment Control
    This button controls the increase of settings. When pressing this button, the parameter will increase.

  7. Decrement Control
    This button controls the decrease of parameter. When pressing this button, the parameter will decrease.

  8. Timer
    The unit has a timer of 1-60 minutes and Continue. It can be adjusted by pressing the “Set” and “Increment” or “Decrement” controls. The treatment time will countdown automatically one minute by one minute. Its output will be shut off when time is up.

  9. Low Battery Indicator
    A low battery indicator will show up on the liguid crystal display when the battery need to be replaced as soon as possible. The unit may continue to operate for a few more hours depends on the setting intensity level.

  10. Steps to Set a New Program
    The settings can be adjusted according to the following steps.

    • a. Turn on the Intensity
      After the electrodes are placed firmly on skin and the lead wires are plugged in the socket of device, turn the on/off control clockwise. The liquid crystal display will be light up.

    • b. Select a Mode
      Select a mode by pressing the “MODE” control. The mode you select will show up on the top of liquid crystal display. There are 5 modes of your option, including -Burst. Normal, Modulation, SD1 and SD2

    • c. Set Pulse Width
      Pulse Width is adjustable from 50 us to 300 us. Press “SET” control to enter this menu, then press “Increment” or “Decre- ment” to adjust the setting. If no instructions regarding the pulse width are given in therapy, set the control to the suggested 70-120 us setting.

    • d. Set Pulse Rate
      Pulse rate is adjustable from 2Hz to 150 Hz. Press “SET” control to enter this menu, then press “Increment” or “Decre- ment” to adjust the setting. Unless otherwise instructed, turn the pulse rate control to the 70-120 Hz setting.

    • e. Set Timer
      Press “SET” to enter this setting. The treatment time is adjust- able from 1 to 60 minutes or Continue. Press “Increment” or “Decrement” control to adjust setting. Your settings will be stored in this unit eternally unless they are adjusted again. You can set the timer to “Continuous” mode by pressing the increment” control when it shows 60 minutes.

  11. Patient Compliance Meter
    This unit can store 60 sets of operation records. Total treatment time up to 999 hours can be stored.

    • Check & Delete Individual Records
      Press “Mode” control and turn on the power simultaneously. The LCD will show the number of records and operation time. Press the “Increment” and “Decrement” button to check each record. To delete a record, press “SET” control for 3 seconds.

    • Check and Delete Accumulative Record
      At the individual records menu, press “Mode” control to switch to accumulative record menu. Press the “SET” control first, then press the “MODE” control simultaneously for 3 seconds and all of the records will be deleted followed by a beeper sound.

  12. Check/Replace the Battery:
    Over time, in order to ensure the functional safety of TENS, changing the battery is necessary.

  13. Make sure that both intensity controls are switched to off position.

  14. Slide the battery compartment cover and open.

  15. Remove the battery from the compartment.

  16. Insert the battery into the compartment. Note the polarity indicated on the battery and in the compartment.

  17. Replace the battery compartment cover and press to close.

BATTERY INFORMATION PRECATIONS

Chapter 21; BATTERY INFORMATION PRECATIONS

  1. Remove battery if equipment is not likely to be used for some time.
  2. Please recycle the used battery in accordance with domestic regulation.
  3. Do not throw the used battery into fire. If you use rechargeable batteries, please follow the instructions.

RECHARGEABLE BATTERIES (NOT INCLUDED)
Prior to the use of a new unit, the rechargeable battery should be charged according to the battery manufacturer’s instructions. Be-  fore using the battery charger, read all instructions and cautionarymarkings on the battery and in this instruction manual. After being stored for 60 days or more, the batteries may lose their charge. After long periods of storage, batteries should be charged prior to use.

BATTERY CHARGING

  1. Plug the charger into any working 110 or 220/240V mains electr- ical outlet. The use of any attachment not supplied with the charger may result in the risk of fire, electric shock, or injury to persons.
  2. Follow the battery manufacturer’s instructions for charging time.
  3. After the battery manufacturer’s recommended charging time has been completed, unplug the charger and remove the battery.
  4. Batteries should always be stored in a fully charged state. To ensure optimum battery performance, follow these guidelines:
    • (a) Although overcharging the batteries for up to 24 hours will not damage them, repeated overcharging may decrease use- ful battery life.
    • (b) Always store batteries in their charged condition. After a battery has been discharged, recharge it as soon as possible. If the battery is stored more than 60 days, it may need to be recharged.
    • (c) Do not short the terminals of the battery. This will cause the battery to get hot and can cause permanent damage. Avoid storing the batteries in your pocket or purse where the termi- nals may accidentally come into contact with coins, keys or any metal objects.
    • (d) WARNINGS:
    • Do not attempt to charge any other types of batteries in your charger, other than rechargeable batteries made for your charger. Other types of batteries may leak or burst
    • Do not incinerate the rechargeable battery as it may explode!

Chapter 22: MAINTENANCE, TRANSPORTATION AND STORAGE OF THE DEVICE

  1. Non-flammable cleaning solution is suitable for cleaning the device.
    • Note: Do not smoke or work with open lights (for example, candles, etc.) when working with flammable liquids.
  2. Stains and spots can be removed with a cleaning agent.
  3. Do not submerge the device in liquids or expose it to large amounts of water.
  4. Return the device to the carrying box with sponge foam to en sure that the unit is well-protected before transportation.
  5. If the device is not to be used for a long period of time, remove the batteries from the battery compartment (acid may leak from used batteries and damage the device). Put the device and accesso- ries in carrying box and keep it in cool dry place.
  6. The packed TENS device should be stored and transported un der the temperature range of -20°C ~ + 60°C, relative humidity 20% ~ 95%, atmosphere pressure 500 hPa ~ 1060 hPa.

Chapter 23; SAFETY-TECHNICAL CONTROLS
For safety reasons, review the following checklist before using your TENS 7000.

  1. Check the device for external damage. deformation of the housing. damaged or defective output sockets.
  2. Check the device for defective operating elements. legibility of inscriptions and labels. make sure the inscriptions and labels are not distorted.
  3. Check the usability of accessories.
    • patient cable undamaged.
    • electrodes undamaged.
    • Battery is not corroded
    • Please consult your distributor if there are any problems with device and accessories.

Chapter 24 : MALFUNCTIONS
Should any malfunctions occur while using the TENS 7000, check whether the parameters are set to the appropriate form of therapy. Adjust the control correctly.

  • whether the cable is correctly connected to the device. The cables should be inserted completely into the sockets.
  • whether the LCD reveals the menu. If necessary, insert a new battery.
  • for possible damage to the cable. Change the cable if any dam- age is detected.
  • If there is any other problem, please return the device to your distributor. Do not try to repair a defective device.

Chapter 25; CONFORMITY TO SAFETY STANDARDS

  • The TENS 7000 devices are in compliance with IC 60601-1

WARRANTY

Chapter 26: WARRANTY

  • All TENS 7000 models carry a warranty of one year from the date of delivery. The warranty applies to the stimulator only and covers both parts and labor relating thereto.
  • The warranty does not apply to damage resulting from failure to follow the operating instructions, accidents, abuse, alteration or dis- assembly by unauthorized personnel.

TROUBLESHOOTING

CHAPTER 27: TROUBLESHOOTING
If your device does not seem to be operating correctly, refer to the chart below to determine what may be wrong. Should none of these measures correct the problem, the device should be serviced.COMPASS-HEALTH-TENS-7000-Muscle-
Stimulator-and-Pain-Relief-Device-FIG-1 \(30\) COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-
Pain-Relief-Device-FIG-1 \(31\)

CHAPTER 28: IMPORTANT INFORMATION REGARDING ELECTROMAGNETIC COMPATIBILITY (EMC)

  • This device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this device should be observed to verify normal operation in the configuration in which it will be used
  • Use of accessories other than those specified or provided by the manufacturer of this device could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • When the operating environment is relatively dry, strong electromagnetic interfer- ence usually occurs. At this time, the device may be affected as follows:
    • the device stops output:
    • the device turns off;
    • the device restarts;
  • The above phenomenon does not affect the basic safety and essential performance of the device, and the user can use it according to the instruction. If you want to avoid the above phenomenon, please use it according to the environment specified in the manual.COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(32\) COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(33\) COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(34\) COMPASS-HEALTH-TENS-7000-Muscle-Stimulator-and-Pain-Relief-Device-FIG-1 \(35\)

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

COMPASS HEALTH User Manuals

Related Manuals