APPLEDENTAL R-Smart Mini Apple Dental Endo Motor
- June 13, 2024
- APPLEDENTAL
Table of Contents
Endo Moto r
MODEL: R-Smart Mini
Product User Manual
For Dental Use Only
R-Smart Mini Apple Dental Endo Motor
Product Type: LED Cordless Endo Motor
Brand: DTS
Model: R-Smart Mini, K028
Producer Company: Guangzhou Reborn Endo Medical Instrument Co., Ltd.
Exporter Company: GUANGDONG JINME MEDICAL TECHNOLOGY CO. LTD.
Importer Company: METCO DENTAL A.Ş.
Adres: Alb. Faik Sözdener Cad. Çetintaş İş Merk.
No: 13/3 34710 Kadıköy İstanbul
Tel: 216 3457424 Fax: 216 3499368
e-mail: metco@metcodental.com
web: www.metcodental.com
Thank you for purchasing Endo Motor I .
General Introduction
Endo Motor I is an endodontic operation instrument Endo motor function is a
kind of endodontic micromotor to drive Ni-It instruments. It can assist the
dentists to shape more standard root-canal in the press of root-canal
treatment based on the micro-electron control technology. Endo motor I can
held in hand to operate and cordless charging. This unit contributes to
alleviate the dentist’s working intensity.
Please read this Operation Manual carefully before use for operation
instructions, care and maintenance. It is recommended to keep this Operation
Manual for future reference.
Contra-indications
- In cases where a patient has been fitted with an implanted heart pacemaker (or other electrical equipment) and has been cautioned against use of small electrical appliances (such as electric shavers,hair dryers, etc, )it is recommended not to use Endo Motoric .
- This instrument should not be used for severe curved root canal preparation. Do not use Endo Motor I for implant or any other dentist procedure outside endodontics
Warnings
- For your own safety, please wear personal protective equipment’s (gloves, glasses, mask).
- Do not use this unit when it’s’ charging.
- Do not disassemble or alter the motor handpiece.
- If you should notice battery fluid leak, deformation of the motor handpiece casing or partial discoloring, immediately stop use and contact your dealer.
- If you will not use the product for long period of time, remove the battery to avoid fluid leak.
- The system may present the possibility of malfunction when used in the presence of electromagnetic interference wave. Do not install system in the vicinity of the device which emits magnetic waves.
- Do not use in the presence of flammable products.
- This motor is equipped with an electronic circuit to limit file breakage. Nevertheless, files may still break due to higher torque conditions or to fatigue resistance.
Precautions
- The appliance must only be used with the manufacturer’s original accessories.
- Never use any other AC adapters.
- The RPM number indicates true instrument speed when the appropriate reduction rate is selected. Make sure the selected reduction matches the ratio of the contra angle handpiece.
- If the motor stops or runs too fast, discontinue use and call for assistance.
- After sterilization, make sure the motor has reached a temperature below 40°C before use.
- Do not introduce any oil into the micro motor.
- The system functions normally in the environment where the temperature is at 5-40°C (50-104°F), humidity at 10-85%RH, atmospheric pressure at 50-106kpa, and no moisture condensation in the Control Unit. Any use outside these limits may cause malfunction
- Do not use bent, damaged, deformed or non–ISO- injury due to its unexpected breaking or “flying off” conforming file. Using such a file may result in persona during rotation.
- Check the product before use by rotating it, paying attention to looseness. vibration, noise and temperature (heat generation). If any abnormal condition is found even slightly at that time, immediately stop use and contact your dealer.
- Always clean the shank of the file to be installed. Allowing dirt to enter the chuck could cause loss of concentricity and deterioration of chucking force.
- Before changing head or bur, turn off the power of the motor handpiece. Changing with the power kept on may cause unintended rotation by accidental touch of the ON/OFF button.
- Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrousoxide.
Notice
- Exercise sufficient care in using the product by giving patient safety first priority.
- Fully-charged rechargeable battery generally discharge gradually over time even the motor handpiece is not used It is recommended to recharge the battery just before use.
- When the motor handpiece stops automatically by detecting a low battery voltage, turning on the power again after a while may not detect the low voltage immediately. This is not a failure, but due to battery characteristics. Since the voltage drop does not coincide with the remaining battery capacity, consider it only as a yardstick.
- When discarding the contra angle head and motor handpiece. dispose of them as medical waste.
- Users are responsible for operational control, maintenance and inspection.
- Store the system in place where temperature is at-10. 50°C (14-122°F), humidity at 10-85%RH, atmospheric pressure at 50-106Kpa, and the system is not subject to air with dust, sulfur or salinity.
Features
- This unit can be used in independent endodontic shaping.
- 9 different programs can be preset according to requirement of operator.
- When rotary file reaches preset torque, the file will automatically withdraw until removal of resistance.
- This unit is powered by rechargeable Lithium battery, in order to expand life span of battery. The unit will be turned off automatically in 8 minutes without any uses.
- Contra angle, file clip and lip hook can be 135°C autoclaved
Main Specifications
< Main Unit>
Battery Voltage | DC3.7V |
---|---|
Battery capacity | 950mAh |
Charging timc Weight | 3 hours approx. |
500g
Packing size| 188x90x85mm
Rated input voltage | AC100-240V 50/60Hz |
---|
Output voltage
Output current| DC 5V/MICRO USB
1A
< Operation Conditions>
Environment temperature | + 5°C ~ + 40°C |
---|---|
Relative humidity | 10%-80%RH |
Atmospheric pressure | 50~106KPa |
Component Names
Before use, please refer to the packaging labels.
- Charger Base —1 pcs
- AC Adapter —1 pcs
- Main Unit —1 pcs
- Contra angle —1 pcs
- LED light —2pcs
- User manual —1 pcs
Installation Setting
Installation of Contra angle
(1)Connect endo file bracket to contra angle.
(2)Confirm contra angle.
(3)Connect together main unit and contra angle.
Notice
Please turn off this unit when connecting contra angle.
(4)Make sure that endo file hold is securely connected with file
Setting
Picture | Description |
---|---|
Power button |
Press power button for more than one second to turn on/off the power and OLED
panel light.
ON/OFF| Motor button
When the power is on, press ON/OFF button then the device start working. Press
this key again, the device stop working
| Select button
When motor is stop, press the Select button, to select the program of memory
from PRGI to PRG9. When motor is stop, press the select button more than 5
seconds to enter custom preset memory mode. Then, press the Select button
about one second to change GR or TQ or Rotation Model setting with the
UP/DOWN button.
| UP/DOWN button
When motor is running or stop, press the UP/DOWN button to increase or reduce
the speed. TO enter custom preset memory mode, Then, press the Select button
about one second to change GR or TQ or Rotation Model setting with the UP/DOWN
button.
Reversing| Reversing button
Press this key one time ,the device can stop working.
Continue to press this key, the device can enter
Reverse working Model.
---|---
| Forward Model–Auto reversing
If the load is removed after auto reverse rotation, it returns to the normal
rotation again.
| Reverse Model
Set the motor rotation to reverse
| AUTO Reciprocating Model
The motor will run reciprocating. The speed is 300rpm.
| Reciprocating Model
The motor will run reciprocating. The speed is 300rpm and the torque is
uncontrolled.
PRG 1-9| For 16: Contra angle. Speed 140-550rpm.
Torque 0.3-4Ncm
| Means low voltage. Charge immediately.
Main Directions For Use
(1)Preparation procedures
1.Please make sure that all accessories are securely connected (Refer to 4.
Installation and Setting).
2.Recheck if rotary file has been securely connected.
(2)Charge (if batteries are used)
1.Securely insert plug of AC adapter into a power supply.
2.Internal microcomputer checks voltage of battery and starts charging, if
necessary, If it starts charging, the blue LCD light of button mark. If
charging finished, the blue LCD button mark is all
●There is no need to turn on the power.
(3)Endo motor functioning
1. Connect contra angle to the main unit.
2. Push the ON/OFF button to start the operation.
Cleaning & Sterilization
(1)Sterilization
1.Contra angle and rotary file can be high pressure autoclaved. Autoclave
contra angle and rotary file after each use.
2.Don’t autoclave any parts (the main unit and charging base)other than those
that can be subjected to autoclave sterilization.
3.Autoclavable up to a max. 135°C. It’s recommended to autoclave for 20 min at
121°C or 15 min at 132°C.
Notice: Use AR sprays to clean contra angle before the autoclave process.
Please autoclave rotary file according to file manufacturer requirement.
Notice: Don’t use any solvent such as benzine and thinner for
sterilization.
(2)Cleaning
Wipe the main unit with neutral detergent if needed.
Trouble shooting
When trouble is found, check the following items again before asking repairs, If none of these are applicable or the trouble is not remedied even after action has been taken, a failure of this product is suspected. Contact your dealer.
Problems | Cause & Solution |
---|
Power hasn’t been
turned on.| The battery is empty.
Charge the battery
Pressing time isn’t enough.
Please long press the power button
Power button is broken, contact your dealer
The battery can’t
be charged| The adapter hasn’t been well connected
Please recheck the connection.
The adapter is broken.
Contact your dealer
The battery is broken.
Contact your dealer.
Disposing product
Please consult with dealer from whom you purchased it about waster despoil.
Warranty
Manufacturer warrants its products to the original purchaser against defects
in material and workmanship under normal practices of installation, use and
servicing. File clip, lip hook are expendable components, and are not covered
by
this warranty. This unit is guaranteed for its designated use against original
defects in materials and workmanship for a period of 12 months from date of
purchasing.