Blue-White WR10 APNTU Turbidity Analyzer Panel
Product Information
The APNTU Turbidity Analyzer Panel is a turn-key monitoring solution designed for clean water applications. It is used for measuring, displaying, and logging ultra-low turbidity in real-time. The analyzer panel features proprietary smart sensor technology, a touch screen display, and a data logging terminal. It is compliant with EPA 180.1 standards.
Key Features:
- Highly accurate measurement of ultra-low turbidity
- Real-time display and data-logging
- Touch screen display for easy operation
- Compliant with EPA 180.1 standards
Dimensions:
Dim | Inch | cm |
---|---|---|
A | 20.00 | 50.8 |
B | 36.00 | 91.4 |
C | 9.42 | 23.9 |
D | 7.80 | 19.8 |
E | 4.98 | 12.6 |
Safety Instructions:
- Read the guide and instruction manual before using
- Install and operate the analyzer only if you are qualified
- Wear eye protection when using chemicals
- Do not exceed maximum pressure to avoid damage
Product Usage Instructions
Pre-Installation:
- Ensure the location where the analyzer will be installed complies with existing codes and regulations.
- Make sure the mounting location is strong enough to support the weight of the analyzer and water. Note that the analyzer and mounts are not designed to support any other structures, piping, or equipment.
Wiring:
- To access the wiring terminal, open the enclosure by loosening the four screws using a screwdriver.
- The analyzer must be powered by a 100~240VAC 50/60Hz power supply. Use a shielded cable with 20-24 AWG (American Wire Gauge) for communication wiring.
Plumbing:
- Plan the incoming water supply and drainage piping before installing the analyzer.
- Additional components may be required to connect to the provided tubing.
Installation Requirements:
- Power Supply: 100~240VAC 50/60Hz
- Drainage: Connect the WR10 outlet tube (3/4 ID Tubing) to a discharge drain via gravity flow.
- Outlet Line: Connect 3/4 ID tubing to the outlet drain. This line must be diverted to drain.
- Wall Mount Space: The APNTU analyzer panel requires approximately 20 H x 36 W x 10 D inches of space for mounting.
- Ensure sufficient space is available.
- Wall Mount Weight: The panel weighs approximately 45 lbs (20kg). Use appropriate mounting hardware.
For more information or assistance, please contact customer service at customerservice@blue-white.com or call 714-893-8529.
Congratulations on your purchase on a APNTU Turbidity water analyzer!
- The APNTU Analyzer Panel is specifically designed as a ‘Turn-Key’ monitor-ing solution for clean water applications including drinking water networks, secondary water supply and decorative/swimming water applications.
- The APNTU series offers highly accurate, real-time measurement, display and data-logging of Ultra-Low Turbidity utilizing proprietary smart sensor technology, coupled with a touch screen display and data logging terminal. APNTU is compliant with EPA 180.1 standards.
- Reference the below dimensional drawings for installation.
APNTU Panel Dimensions
Safety instructions
- CAUTION
- Follow these instructions before installing your analyzer to avoid failure. Read this guide and Instruction Manual before using.
- CAUTION
- The APNTU is designed to be installed and operated by qualified personnel only. Please note that warranty coverage does not include damage due to misuse or improper installation.
- CAUTION
- Always wear eye protection when using chemicals.
- PRESSURE
- Do not exceed maximum pressure. High pressure and temperature will damage the system.
In the Box
- APNTU Analyzer Panel
- Pre-Mounted Struts
- Pre-mounted Flow Meter
- Turbidity Sensor
- 1/4” tubing OD
- 3/4” tubing ID
- Concrete Anchors
- Electrical Schematics
- Quick Start Guide
Tools Needed
- Phillips Screwdriver
- Adjustable wrench
Planning
Ensure location of Analyzer complies with existing codes and regulations.
Location of mounting must be strong enough to support weight of Analyzer and
water. Analyzer and mounts are not designed to support any other structures,
piping, or equipment.
Wiring
To access the wiring terminal, open the enclosure by loosening the four screws
with the screwdriver. The Analyzer must be powered by 100~240VAC 50/60Hz.
Communication wiring should be 20-24 AWG(American Wire Gauge) shielded cable.
Plumbing
Incoming water supply and drainage piping should
be planned before installing the Analyzer. Additional components may be
required to connect to the provided tubing.
Need more information. Please contact customer service with questions. customerservice@blue-white.com 714-893-8529
Scan QR code to download full Instruction Manual
Installation Requirements
- Power Supply: 100~240VAC 50/60Hz
- Water Supply: Inlet water pressure should be from 7.25 – 30 psi (0.05-0.2MPa) with an inlet feedwater line diameter of ¼-inch O.D. Tubing. The APNTU is pro-vided with an inlet Rotameter and flow regulating valve for sample water inlet flow control and limited pressure regulation. The range of inlet flow for the WR10 should be consistently maintained between 200 and 400 mL per minute.
- Inlet Water: Connect the ¼-inch inlet water tubing to the quick adapter provided. Consistent flow of 200-400mL is required.
- Drainage: The WR10 outlet tube (3/4” ID Tubing) located on the bottom of the WR10, should be connected to a discharge drain via gravity flow.
- Outlet Line: Connect 3/4” ID tubing to the outlet drain. This is the sample water outlet flow. This line must be diverted to drain.
- Wall Mount Space: The APNTU analyzer panel size is roughly 20” H x 36” W x 10” D in dimension. Please accommodate sufficient space for mounting. The panel is equipped with pre-installed rear unistrut for simple wall mounting.
- Wall Mount Weight: Approximately 45 lbs (20kg). Please use appropriate mounting hardware.
Wiring
- The APNTU analyzer has universal AC power supply equipment allowing users simply to plug the power supply into a 100~240V AC 50/60Hz power outlet for normal operation.
- The process of electrical connection to contact the 220V single-phase power supply, should be operated by personnel with an electrician’s license. Failure to operate according to the electrical code of practice may result in electric shock injury or even death.
Note : When in doubt regarding your electrical installation, contact a licensed electrician.
INSTALLATION
Step-By-Step
- Measure
- Use the provided drawing dimensions to determine best mounting location of panel.
- Measure wall and locate mounting holes. Panels should be mounted at an elevation to allow for easy access to control unit.
- Wall Mount Weight: 45 lbs
- Install Strut
- Use the provided strut and hardware to mount the panel.
- Strut to be mounted securely to wall using appropriate hardware.
- Strut spacing is on 15” centerline.
- If using provided anchor bolts, drill holes in concrete and clean thoroughly. Located strut and mount with appropriate tools.
- Check to make sure strut is secure and level
- Mount Panel
- Mounting panel will requires 2 or 3 persons.
- Mount panel to strut using provided hardware.
- Be sure panel is level before securing.
- Check spacing and panel before connecting electrical and piping.
- Connect Electrical and Tubing
- Use the provided electrical diagram to connect power and communication wiring.
- Consult with the electrical professionals before performing any electrical work
- Power Up System
- Check all electrical connections before powering on system.
- Check piping connections for leaks.
- System is activated by pressing Power button on side of control panel.
Start-Up and Operation
After the system is powered on (using the power button, located on the left
side of the TS10), a “Blue-White” introduction screen will appear and then
immediately proceed to Real Time Monitoring screen.
The system does not require “User Login” to view real time monitoring of the
system. “User Login” is only necessary when changing parameter settings or
performing calibrations. To access this now, press the icon and refer to
Instruction Manual for more instructions regarding login and user management.
Default password is “888888”
Real Time Monitoring
Upon start-up, the main interface screen will appear showing real-time monitoring of the system. The data detected by the Blue-White sensor will be displayed in real-time. For further programming instructions and a functional overview of each section, please refer to complete Instruction Manual.
Engineering Specifications
WARRANTY
LIMITED WARRANTY
- Your Blue-White product is a quality product and is warranted for a specific time from the date of purchase (proof of purchase is required).
- The product will be repaired or replaced at our discretion. Failure must have occurred due to a defect in material or workmanship and not as a result of the operation of the product other than in normal operation as defined in the product manual. Warranty status is determined by the product’s serial label and the sales invoice or receipt. The serial label must be on the product and legible. The warranty status of the product will be verified by Blue-White or a factory authorized service center.
- APNTU is warranted for 13 months from the date of purchase. Sensor is warranted for 6 months.
WHAT IS NOT COVERED
- Freight to the factory, or service center.
- Products that have been tampered with, or in pieces.
- Damage resulting from misuse, carelessness such as chemical spills on the enclosure, abuse, lack of maintenance, or alteration which is out of our control.
- Damage by faulty wiring, power surges, or acts of nature.
- Damage that occurs as a result of: meter misalignment, improper installation, over tightening, use of non-recommended chemicals, use of non-recommended adhesives or pipe dopes, excessive heat or pressure, or allowing the unit to support the weight of related piping
BLUE-WHITE does not assume responsibility for any loss, damage, or
expense directly or indirectly related to or arising out of the use of its
products.
Failure must have occurred due to a defect in material or workmanship and not
as a result of the operation of the product other than in normal operation as
defined in the manual.
Warranty status is determined by the product’s serial label and the sales
invoice or receipt. The serial label must be on the product and legible.
The warranty status will be verified by Blue-White or a factory-authorized
service center.
PROCEDURE FOR IN-WARRANTY REPAIR
- Warranty service must be performed by the factory or an authorized service center. Contact the factory or local repair center to obtain a RMA (Return Material Authorization) number.
- Decontaminate, dry, and carefully pack the product to be repaired. Please enclose a brief description of the problem and proof of purchase.
- Prepay all shipping and insurance costs. COD shipments will not be accepted. Damage caused by improper packaging is the responsibility of the sender. When In-Warranty repair is completed, the factory pays for return shipping to the dealer or customer.
P.N. #80000-188 APNTU QSG Rev. 1 2023425
Have Questions?
Contact us
+1 714-893-8529
customerservice@blue-white.com
blue-white.com/contact-us