HANNA HI749 Chromium VI Low Range Instruction Manual

June 12, 2024
HANNA

HANNA HI749 Chromium VI Low Range

HANNA-HI749-Chromium-VI-Low-Range-product

Product Information

The HI749 Chromium VI Low Range handheld checker is designed to accurately determine chromium(VI) levels in water. It features a single-button operation system and a large LCD display for easy reading. The auto shut-off feature helps to conserve battery life. The checker comes with various accessories, including reagent sets, a certified standard kit, a cuvette cleaning solution, and a battery set.

Specifications

  • Range: Low Range
  • Resolution: Not specified
  • Accuracy: Not specified
  • Light source: Not specified
  • Light detector: Not specified
  • Method: Not specified
  • Environment: Not specified
  • Battery type: 1.5V AAA battery
  • Auto shut-off: Yes
  • Dimensions: Not specified
  • Weight: Not specified

Product Usage Instructions

Battery Replacement

  1. Press and hold the ON/OFF button to turn the checker off.
  2. Turn the instrument upside down and use a screwdriver to unfasten the screw and remove the battery cover.
  3. Remove the old battery and replace it with a new 1.5V AAA battery, inserting the negative end first.
  4. Replace the battery cover, fasten, and tighten the screw.

Measurement Procedure
Ensure that the checker is suitable for your specific application and environment before use. Follow these steps to perform a measurement:

  1. Prepare the zero cuvette by following the provided instructions.
  2. Ensure that there is enough light to perform the measurement. If the “Light Low” warning appears, check the preparation of the zero cuvette.
  3. Place the sample cuvette in the cuvette holder.
  4. Press the ON/OFF button to turn on the checker.
  5. Insert the zero cuvette and close the cap.
  6. Wait for the measurement to be displayed on the LCD.
  7. Read and record the result.

Errors & Warnings
The checker provides clear warning messages and explanations for erroneous conditions or values outside the expected range. Take the recommended actions for each error or warning:

  • Light High: Excess ambient light is reaching the detector. Check the preparation of the zero cuvette.
  • Light Low: There is not enough light to perform a measurement. Check the preparation of the zero cuvette.
  • Inverted Cuvettes: The sample and zero cuvettes are inverted. Swap the cuvettes and repeat the measurement.
  • Under Range: A blinking 0 indicates that the sample absorbs less light than the zero reference. Check the measurement procedure and ensure you use the same cuvette for reference (zero) and measurement.
  • Over Range: The maximum concentration value displayed blinking indicates that the measured value is outside the limits of the method. Verify that the sample does not contain any debris, dilute if necessary, and repeat the measurement.

For further information, refer to the instruction manual provided by Hanna Instruments.

Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the Checker®HC handheld colorimeter. For more information about Hanna Instruments and our products, visit www.hannainst.com or e-mail us at sales@hannainst.com. For technical support, contact your local Hanna Instruments Office or e-mail us at tech@hannainst.com.

Preliminary Examination

Remove the Checker®HC handheld colorimeter and accessories from the packing material and examine it carefully. If you require any further information, please contact the Hanna Instruments technical support team.

Each HI749 is delivered in a case with a custom insert and is supplied with:

  • Sample cuvette and cap (2 pcs.)
  • Chromium VI Low Range reagent starter kit (reagents for 6 tests)
  • 1.5V AAA Alkaline battery (1 pc.)
  • Instruction manual

Note:
Save all packing material until you are sure that the Checker®HC handheld colorimeter works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.

General Description & Intended Use

  • HI749 Chromium VI Low Range handheld checker is designed to accurately determine chromium(VI) levels in water.
  • HI749 features a single-button operation system and is easy to use. The large LCD is easy to read and the auto shut-off feature assures the battery will not be drained.

Specifications

  • Range 0 to 300 ppb (as Cr(IV))
  • Resolution 1 ppb
  • Accuracy ±3 ppb ±5% of reading @ 25 °C (77 °F)
  • Light source Light Emitting Diode @ 525 nm
  • Light detector Silicon photocell
  • Method
    • Adaptation of the ASTM Manual of Water and Environmental
    • Technology, D1687-92, Diphenylcarbohydrazide Method.
    • The reaction between chromium(VI) and the reagent causes a purple tint in the sample.
  • Environment 0 to 50 °C (32 to 122 °F); max. 95% RH non-condensing
  • Battery type 1.5V AAA Alkaline
  • Auto shut-off After 10 minutes of non-use
  • Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
  • Weight 64 g (2.3 oz)

Functional Description & LCD Display

HANNA-HI749-Chromium-VI-Low-Range-fig- \(1\)

Measurement Procedure

  • Press the ON/OFF button to turn the checker on. All segments will be displayed for a few seconds, followed by “Add”, “C.1” with “Press” blinking.
  • Fill the cuvette with 10 mL of unreacted sample and replace the cap. Insert the cuvette into the checker and close the cap.
  • Press the ON/OFF button. When the display shows “Add”, “C.2” with “Press” blinking, the checker is zeroed.
  • Remove the cuvette, unscrew the cap and add the content of one packet of HI749-0 Chromium VI Low Range reagent. Replace the cap and shake vigorously for about 10 seconds.
  • Insert the cuvette into the checker and close the cap. Press and hold the ON/OFF button. The display will show the countdown prior to the measurement. Alternatively, wait 6 minutes and press the button.
  • When the timer ends the checker will perform the reading. The instrument displays the chromium VI concentration in ppb. The checker automatically turns off 10 minutes after reading.

Errors & Warnings

The checker shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. The information below provides an explanation of the errors and warnings, and the recommended action to be taken.

  • Light High: There is an excess amount of ambient light reaching the detector. Please check the preparation of the zero cuvette.
  • Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette.
  • Inverted Cuvettes: The sample and the zero cuvettes are inverted. Swap the cuvettes and repeat the measurement.
  • Under Range: A blinking “0” indicates that the sample absorbs less light than the zero reference. Check the measurement procedure and make sure to use the same cuvette for reference (zero) and measurement.
  • Over Range: Maximum concentration value displayed blinking indicates the measured value is outside the limits of the method. Verify that the sample does not contain any debris. Dilute the sample and repeat the measurement.
  • Battery Low: Battery level is too low for the checker to function properly. Replace the battery with a new one.
  • Drained Battery: The battery is drained and must be replaced. Replace the battery with a new one and restart the checker.

Tips for an Accurate Measurement

  • Ensure the sample does not contain any debris.
  • Whenever the cuvette is placed into the checker, it must be dry outside and free of fingerprints, oil and dirt.
  • Wipe the cuvette thoroughly with HI731318 microfiber cleaning cloth or a lint-free cloth prior to insertion.
  • Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
  • Do not let the reacted sample stand too long after reagent has been added, as accuracy will be affected.
  • Discard the sample immediately after the reading has been taken or the glass might become permanently stained.

Battery Replacement

  • To save the battery, the checker shuts down after 10 minutes of non-use.
  • A fresh battery lasts for a minimum of 5000 measurements. When the battery is drained, the instrument displays “bAd” then ”bAt”, and turns off.

To replace the battery, follow the next steps:

  1. Press and hold the ON/OFF button to turn the checker off.
  2. Turn the instrument upside down and use a screwdriver to unfasten the screw and remove the battery cover.
  3. Remove the old battery, replace it with a new 1.5V AAA battery, inserting the negative end first.
  4. Replace the battery cover, fasten and tighten the screw.

Accessories

Reagent Sets

  • HI749-25 Reagents for 25 Chromium VI Low Range tests

Other Accessories

  • HI749-11 Chromium VI Low Range certified standard kit
  • HI731225 Cuvette black cap for Checker® HC colorimeters (4 pcs.)
  • HI731318 Cloth for wiping cuvettes (4 pcs.)
  • HI731321 Glass cuvette and seal cap for Checker® HC colorimeters (4 pcs.)
  • HI740028P 1.5V AAA battery set (12 pcs.)
  • HI93703-50 Cuvette cleaning solution, 230 mL

Certification

  • All Hanna Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment. The product should not be treated as household waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources.
  • Disposal of waste batteries. This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.
  • Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to www.hannainst.com.

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the checker’s performance. For your and the checker’s safety do not use or store it in hazardous environments.

Warranty

HI749 Checker®HC is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the checker is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any product, make sure it is properly packaged for complete protection.

Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.