Newentor CO2 Monitor Indoor Air Quality Meters Instruction Manual

June 12, 2024
Newentor

Newentor-logo

Newentor CO2 Monitor Indoor Air Quality Meters

Newentor-CO2-Monitor-Indoor-Air-Quality-Meters-product

Product Information

Product Name Newentor CO2 Detector
Email (US) service@newentor.com
Email (UK) service.uk@newentor.com
Facebook Newentor
WhatsApp +8617512038920
Telegram NEWENTOR

Package List

Item No. Part Name Quantity
1 CO2 detector 1
2 Charger cable 1
3 User manual 1

CO2 Detector Appearance

Operating Buttons Function

Function Button Action (Press) Action (Hold)
NORMAL MODE Light & Snooze Display alarm time Activate time calendar

settings mode
UP| Max / Min data| Delete max/min data
DOWN| Manual real-time detection (In battery mode)| N/A
ALARM| Activate/Deactivate alarm| N/A
SETTING MODE| TIME| Setting alarm time| Save the change to the setting
ALERT| Activate/Deactivate the CO2 alert setting| CO2 alert setting

CO2 Value Introduction

CO2 ppm AQI Description
0~450 01 Good
451~1000 02 Unhealthy for sensitive people
1001~1500 03 Moderate for sensitive people
1501~2000 04 Unhealthy
2001~3000 05 Very unhealthy
3001~5000 06 Endangering health

CO2 DETECTOR FEATURES

Package List
ITEM NO. PART NAME FIGURE QTY
1 CO2 detector 1
2 Charger cable ****

1

| |
3| User manual| | 1

CO2 Detector Appearance

Front of the Main Unit

Newentor-CO2-Monitor-Indoor-Air-Quality-Meters-fig-
\(4\)

  • A1. “ ”Light Button
  • A2. Fan Icon
  • A3. Alert Icon
  • A4. CO2 Value
  • A5. CO2 Index
  • A6. Temperature
  • A7. Time
  • A8. Alarm
  • A9. CO2 Character
  • A10. Humidity
  • A11. Low Battery Indicator
  • A12. Date
  • A13. Day of Week

Back of the Main Unit

Newentor-CO2-Monitor-Indoor-Air-Quality-Meters-fig-
\(7\)

  • B1. Wall Mount Hole
  • B2. Operating Buttons
  • B3. CO2 Detection Port
  • B4. 3×AAA Batteries (not included)
  • B5. Fold-out Stand
  • B6. DC Interface (5V 0.5A)
Operating Buttons Function

Function


Button

| NORMAL MODE| SETTING MODE
---|---|---
Press| Hold| Press| Hold
**| Light &Snooze
TIME| Display alarm time| Setting alarm time| Save the change to the setting| Activate time calendar settings mode
UP| Max/Min data| Delete max/min data| Adjust values (Increase)| Adjust values (Increase continually)
DOWN| Manual real-time detection (In battery mode)| ℃/℉ switch| Adjust values (Decrease)| Adjust values (Decrease continually)
ALARM| Activate/ Deactivate alarm| | |
ALERT**| Activate/Deactivate the CO2 alert setting| CO2 alert setting| |

SWITCH ON

When switched on, the LCD full graphical display is shown for 2 seconds, followed by a “BI” sound, then the machine gets to work, detecting room temperature, humidity, and CO2 value.

With Power Supply:

  • Using an adapter higher than 5V 0.5A (package only with charger cable).
  • Plug the charger cable into the DC interface (B6).
  • The device enters a 60s countdown and then displays the current value.
  • It detects and updates the CO2 value every 6 seconds.

With Battery:

  • After inserting the batteries, the device enters 30s countdown and updates CO2 value 5 times.The CO2 value are updated 5 times in every 20 minutes.
  • When using the battery alone, press the “DOWN” button to enter 30s countdown immediately, followed by re-testing and updating CO2 value 5 times.

Note:

  • The countdown process mentioned above is a state of self-regulation of the CO2 sensor (The original sensor manufacturer: Sensirion) in order to make the detection data more accurate.
  • The detection port (B3) must never be covered. If the air circulation is poor, it may result in inaccurate CO2, temperature and humidity measurements.
  • It is recommended to use both batteries and power supply to avoid losing data in the event of a power outage.

CO2 VALUE INTRODUCTION

CO2 Index(A3)

  • From 400 to 650 ppm, “ ” lights up dark green.
  • From 651 to 900 ppm, “ ” lights up light green.
  • From 901 to 1650 ppm, “ ” lights up red.
  • From 1651 to 2400 ppm, “ ” lights up dark red.
  • Measuring range 400-5000 ppm, below 400 display 400 ppm, above 5000 display 5000 ppm.

CO2 Concentration Factor Indicator

CO2 ppm 0~450 451~1000 1001~1500 1501~2000 2001~3000 3001~5000
AQI 01 02 03 04 05 06
Good Moderate Unhealthy for sensitive people Unhealthy ****

Very unhealthy

| ****

Endangering health

CO2 measured value: carbon dioxide concentration in ppm (parts per million = 1 millionth = 1 ppm = 10-6 -equivalent to 0.001‰). The Federal Environment Agency recommends not staying in an environment where the CO2 value exceeds 1000 ppm for a long time.

CO2 ALERT FUNCTION

  • Press and hold the “ALERT” button for 3 seconds to enter the CO2 alert setting and press the “UP” and “DOWN” buttons to adjust the alert value.
  • Press the “ALERT” button to switch on/off the alert “ ” (A3).
    • When the alert is not on but the CO2 value has reached the alert preset value (default 1000 ppm), the icon “ ”(A2) for ventilation is displayed fixed, and the CO2 character (A9) flash.
    • When the alert is on and the value has reached the preset value, the alert tone lasts for 1 minute, while the icon “ ” is displayed fixed and the CO2 character flash. After one minute the alert is switched off and if the value still exceeds the preset value, the alert does not sound again, only the icon “ ” remains fixed and the CO2 character flashes as a warning.

TIME AND CALENDAR SETTING

  • Press and hold the “TIME” button for 3 seconds to enter time/date setting mode. Press the “UP” and “DOWN” buttons to change the values, press the “TIME” button to confirm, and go to the next steps. The sequence steps are as follows:
    • Year
    • DM/MD
    • Month
    • Day
    • 12/24 Hour Format
    • Hour
    • Minute
  • Power on default display: ℃, 24H, 1st January 2022, 0:00.

ALARM SETTING

Alarm Function

  • Press the “TIME” button to select the desired alarm time (ALARM 1 or ALARM 2).
  • Press and hold the “TIME” button for 3 seconds once you have chosen the desired alarm time. The hour indicator of the alarm time (AL 1 or AL 2) will flash.
  • Set the desired hour for the alarm with the “UP” and “DOWN” buttons.
  • Press the “TIME” button to store the setting. The minute indicator will flash on the display.
  • Set the desired minute for the alarm with the “UP” and “DOWN” buttons.
  • Press the “TIME” button to store the setting. The alarm time is set.
  • AL1 default 7:00, AL2 default 13:00.

Switching the Alarm On and Off
Press the “ALARM” button to switch the alarm on and off.

  • Press once: “ ” appears on the display, AL 1 is activated
  • press twice: “ ” appears on the display, AL 2 is activated
  • press three times: “ ” appear on the display, AL 1 and AL 2 are activated
  • press four times: “ ” disappear on the display, AL 1 and AL 2 are deactivated

Snooze Function and Stopping the Alarm
You will be woken by the alarm at the set time, and the sound will become more and more intense as the alarm continues.

  • Snooze function – Press the “ ” button (A1) to stop the alarm for approx. 5 minutes.
  • Switch off the alarm completely – Press any button on the main unit (apart from “ ” ) to stop the alarm. The alarm will go off again after 24 hours.
  • Automatic stop function- If you do not press any buttons, the alarm will automatically switch off after 2 minutes. The alarm will go off again after 24 hours.

TEMPERATURE AND HUMIDITY

  • Temperature measuring range: -9.9℃ — 60℃/14.18℉ — 122℉
  • Temperature accuracy: ±1℃(0~40℃); ±2℃ (others)
  • Humidity measuring range: 20% to 95% relative humidity
  • Humidity accuracy: ±5% with a measuring range of 40% to 80%, ±8% with a measuring range of 20% to 39%, and 81% to 95%.
  • Detection cycle: 30 seconds
    • When the alarm goes on, temperature detection is suspended.

MAX/MIN CO2/TEMPERATURE/HUMIDITY VALUES

  • Press the “UP” button to display maximum/minimum CO2 temperature and humidity values.
  • When you query the maximum and minimum values, press and hold the “UP” button for about 2 seconds to clear and re-record the maximum/minimum CO2 temperature and humidity.
  • Maximum and minimum values are automatically cleared every 0:00.

DISPLAY IN ℃ OR ℉

  • Press and hold the “DOWN” button for 3 seconds to select the temperature unit to display in ℃ or ℉.

BACKLIGHT

  • In power supply mode, the brightness can be adjusted “high – low – off”.
  • In battery mode, press “ ” to light up the backlight for 10 seconds.

LOW BATTERY INDICATOR

  • This indication does not exist in power supply mode.
  • In battery mode, the low voltage symbol “ ”(A11) appears when the battery is below 3.6V. When the low voltage symbol appears, the CO2 does not work anymore and the CO2 value position shows “LO” to indicate a battery change or a power supply connection.

Note:

  • When you replace the batteries all the settings will be lost.
  • Always exchange all batteries and only use those listed in the “Technical Data”.
  • Even if the base unit is powered by the power source, you should still remove all empty batteries present from the main unit.
  • Make sure that the poles of the batteries are correct (+/-) when inserting.
  • Clean the battery and device contacts if necessary before inserting.
  • Dispose of the old batteries in an environmental manner.

CALIBRATION

Calibration Way

  • With no power, press and hold the “TIME” and “DOWN” buttons at the same time (Don’t let go of your hand), followed by plugging in the charger or insert batteries and hold for 3 seconds, then the device enter calibration mode after the “BI” sound. The CO2 position displays the “CAL” character and the calibration time is 4 minutes.
  • If the calibration is successful, “PASS” is displayed. After the calibration is completed, please unplug the charger and remove the battery, and then plug in the charger or insert the battery again to use the device.
  • If the calibration fails, “Err” is displayed. After the calibration fails, please re-calibrate.

Calibration Environment

  • It should be in an environment with good air flow, such as a ventilated window (CO2 concentration is about 400 to 450 ppm).
  • Please ensure that there is no CO2 source in the immediate vicinity.
  • Avoid direct sunlight and rain.
  • CO2 is detected after calibration. Please try to avoid perfumes, paints, air fresheners, cigarette smoke, alcohol, chemical substances and so on while carrying out measurements since these can lead to incorrect measuring results.

Note:
When you feel that the CO2 value is displayed abnormally, please calibrate the device.

PRODUCT SPECIFICATION

  • Model: C1
  • Power Supply: 5V 0.5A (package only with charger cable)
  • Battery: 3 x AAA / 1.5V (not included)
  • CO2 display range: 400 to approx. 5000 ppm
  • Measurement error: CO2 accuracy ±(50 ppm +5% value)
  • Resolution: 1 ppm

PRECAUTIONS

  • Protect the product from dust and use it indoors only.
  • Do not use the product in the immediate vicinity of heaters, other heat sources or in direct sunlight.
  • Do not use the product in a damp environment and avoid splashing water.
  • This device should be kept out of the reach of children.
  • Do not operate the product outside the power limits specified in the technical data.
  • Do not use the product in areas in which electronic products are not permitted.
  • Do not place the product near interference fields, metal frames, computers, TVs, etc. Electronic devices and window frames can negatively affect the product’s functionality.
  • Do not drop the product and do not expose it to any major shocks.
  • Do not modify the product in any way. Doing so voids the warranty.
  • Route all cables so that there is no risk of tripping.
  • Do not bend or crush the cable.
  • Do not incinerate the battery or the product.

WARRANTY

  • The device has been manufactured in accordance with strict quality guidelines and has been inspected carefully before delivery.
  • Newentor guarantees that the warranty period of the device is 1 year from the date of purchase.

DECLARATION OF CONFORMITY
This CO2 detector complies with the basic requirements and the other relevant regulations of Directive EN55032. Products marked with this symbol meet all applicable community regulations of the European Economic Area.

DISPOSAL

  • Dispose of the packaging according to type. Old devices must not be disposed of with normal household waste. According to Directive 2012/19 / EU, the device must be disposed of properly at the end of its service life. Valuables contained in the device are recycled to avoid pollution of the environment. Take the old device to a collection point for electronic scrap or a recycling center. Before disposal, remove the batteries from the device and dispose of them separately. Batteries do not belong in household waste. Every consumer is legally obliged to dispose of batteries and rechargeable batteries at a collection point in their community, district or in retail. This means that all batteries and accumulators can be disposed of in an environmentally friendly manner.
  • Batteries and accumulators that contain harmful substances are marked with this symbol and chemical symbols (Pb for lead). For more information, contact your local waste disposal company or your local authority.
  • Dispose of the packaging according to type. Dispose of cardboard and cardboard as waste paper and foils in the recyclable waste collection.
DISCLAIMERS
  • The manufacturer reserves the right to change the specifications of the product.
  • Manufacturers and suppliers are not responsible for malfunctions where interference occurs.

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operation.

This device has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This device generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.

However, there is no guarantee that interference will not occur in a particular installation. If this device does cause harmful interference to radio or television reception, which can be determined by turning the device off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the device and receiver.
  • Connect the device to an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals