NATIONAL INSTRUMENTS SCXI-2000 Chassis Device
Product Information: SCXI-2000
The SCXI-2000 is a device manufactured by National Instruments. It is part of the SCXI system, designed for data acquisition and control applications. The device comes with a one-year warranty that covers defects in materials and workmanship, including parts and labor. The warranty period starts from the date of shipment.
This device is copyrighted by National Instruments Corporation, and reproduction or transmission of its documentation is prohibited without prior written consent.
The SCXI-2000 is a trademark of National Instruments Corporation, along with other trademarks such as ComponentWorksTM, CVITM, DAQCardTM, DAQPadTM, LabVIEWTM, ni.comTM, NI-DAQTM, PXITM, RTSITM, and VirtualBenchTM.
It is important to note that National Instruments products are not suitable for use in surgical implants or critical components in life support systems.
Product Usage Instructions
Chapter 1: Introduction to SCXI
In this chapter, you will find information on what you need to get started
with the SCXI system. It also provides guidance on identifying your SCXI
components.
Chapter 2: Installing SCXI Modules and Connecting the SCXI System
This chapter covers the installation process for the SCXI modules and
connecting the SCXI system to your DAQ device or computer. It includes
instructions for installing the necessary software, DAQ device, and SCXI
modules.
The chapter also provides guidance on connecting the SCXI system using
different interfaces:
- Using the Parallel Port
- Using a Plug-in DAQ Device or DAQCard
- Using an RS-232 or RS-485 Serial Port
Follow the instructions provided in this chapter to ensure a proper installation and connection of your SCXI system.
SCXI-2000
SCXITM
Getting Started with SCXI
Getting Started with SCXI
July 2000 Edition Part Number 320515F-01
Worldwide Technical Support and Product Information www.ni.com
National Instruments Corporate Headquarters 11500 North Mopac Expressway
Austin, Texas 78759-3504 USA Tel: 512 794 0100
Worldwide Offices Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02
757 00 20, Brazil 011 284 5011, Canada (Calgary) 403 274 9391, Canada
(Ontario) 905 785 0085, Canada (Québec) 514 694 8521, China 0755 3904939,
Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24, Germany 089
741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406,
Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456,
Mexico (D.F.) 5 280 7625, Mexico (Monterrey) 8 357 7695, Netherlands 0348
433466, New Zealand 09 914 0488, Norway 32 27 73 00, Poland 0 22 528 94 06,
Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085, Sweden 08 587
895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635
523545
For further support information, see the Technical Support Resources appendix.
To comment on the documentation, send e-mail to techpubs@ni.com
© Copyright 1993, 2000 National Instruments Corporation. All rights reserved.
Important Information
Warranty
SCXI devices are warranted against defects in materials and workmanship for a
period of one year from the date of shipment, as evidenced by receipts or
other documentation. National Instruments will, at its option, repair or
replace equipment that proves to be defective during the warranty period. This
warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not
to fail to execute programming instructions, due to defects in materials and
workmanship, for a period of 90 days from date of shipment, as evidenced by
receipts or other documentation. National Instruments will, at its option,
repair or replace software media that do not execute programming instructions
if National Instruments receives notice of such defects during the warranty
period. National Instruments does not warrant that the operation of the
software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory
and clearly marked on the outside of the package before any equipment will be
accepted for warranty work. National Instruments will pay the shipping costs
of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is
accurate. The document has been carefully reviewed for technical accuracy. In
the event that technical or typographical errors exist, National Instruments
reserves the right to make changes to subsequent editions of this document
without prior notice to holders of this edition. The reader should consult
National Instruments if errors are suspected. In no event shall National
Instruments be liable for any damages arising out of or related to this
document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS
OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED
BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO
THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE
LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR
INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. This limitation of the liability of National Instruments will apply
regardless of the form of action, whether in contract or tort, including
negligence. Any action against National Instruments must be brought within one
year after the cause of action accrues. National Instruments shall not be
liable for any delay in performance due to causes beyond its reasonable
control. The warranty provided herein does not cover damages, defects,
malfunctions, or service failures caused by owner’s failure to follow the
National Instruments installation, operation, or maintenance instructions;
owner’s modification of the product; owner’s abuse, misuse, or negligent acts;
and power failure or surges, fire, flood, accident, actions of third parties,
or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or
transmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in
whole or in part, without the prior written consent of National Instruments
Corporation.
Trademarks
ComponentWorksTM, CVITM, DAQCardTM, DAQPadTM, LabVIEWTM, National
InstrumentsTM, ni.comTM, NI-DAQTM, PXITM, RTSITM, SCXITM, and VirtualBenchTM
are trademarks of National Instruments Corporation.
ICP® is a registered trademark of PCB Piezotronics, Inc. Other product and
company names mentioned herein are trademarks or trade names of their
respective companies.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING
FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL
IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE
TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE
SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT
LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE
MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS,
SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF
ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC
SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS
ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS
THESE ARE HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION
WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS
(INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY
UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE
REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT
LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS
CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS’ TESTING PLATFORMS AND
BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS
IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED
BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY
RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL
INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED
IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE
DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Getting Started with SCXI
x
About This Manual
This guide is a beginner’s handbook to the installation, configuration, and use of the National Instruments Signal Conditioning eXtensions for Instrumentation (SCXI) chassis, modules, and associated data acquisition (DAQ) devices. This guide is an introduction to SCXI, not a complete reference. Advanced users should refer to the appropriate hardware and software manuals for further information.
Conventions Used in This Guide
AI device AT E Series devices bold DAQCard E Series devices DIO-24 DIO-96
The following conventions are used in this guide.
The symbol indicates that the following text applies only to a specific
product, a specific operating system, or a specific software version.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid
injury, data loss, or a system crash.
This icon denotes a warning, which advises you of precautions to take to avoid
being electrically shocked.
Refers to analog input devices that have AI in their names, such as the AT-AI-
16XE-10.
These AT devices have an E- toward the ends of their names, such as the AT-
MIO-16DE-10 and AT-AI-16XE-10.
Bold text denotes items that you must select or click on in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
Refers to the DAQCard-AI-16E-4 and DAQCard-AI-16XE-50.
Refers to the PC-DIO-24, DAQCard-DIO-24, and PCI-6503.
Refers to the PC-DIO-96/PnP, PCI-DIO-96, PCI-6508, PXI-6508, or DAQPad-6508.
© National Instruments Corporation
xi
Getting Started with SCXI
About This Manual
DIO device italic
Lab/1200 device MIO device MIO-16D monospace
NI-DAQ PXI Remote SCXI
SCXI analog input module
SCXI analog output module SCXI chassis SCXI communication module SCXI digital
module SCXI DAQ module
Refers to any digital I/O devices, such as the PC-DIO-96/PnP PCI-DIO-32HS, or
PXI-6533.
Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
Refers to the DAQCard-1200, DAQPad-1200, Lab-PC+, Lab-PC-1200, Lab-PC-1200AI,
SCXI-1200, and PCI-1200.
Refers to the multifunction I/O devices.
Refers to the AT-MIO-16DE-10.
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This
font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames and extensions, and code excerpts.
Refers to the NI-DAQ driver software unless otherwise noted.
Refers to PCI eXtensions for Instrumentation and is derived from the
CompactPCI Standard.
Refers to an SCXI configuration where either an SCXI-2000 chassis or an
SCXI-2400 remote communications module is connected to the serial port of your
computer.
Refers to the SCXI-1100, SCXI-1102/B/C, SCXI-1104/C, SCXI-1112, SCXI-1120/D,
SCXI-1121, SCXI-1122, SCXI-1125, SCXI-1126, SCXI-1140, SCXI-1141, SCXI-1142,
SCXI-1143, SCXI-1520, SCXI-1530/1531, and SCXI-1540.
Refers to the SCXI-1124.
Refers to the SCXI-1000, SCXI-1000DC, SCXI-1001, and SCXI-2000.
Refers to the SCXI-2400.
Refers to the SCXI-1160, SCXI-1161, SCXI-1162, SCXI-1162HV, SCXI-1163, and
SCXI-1163R.
Refers to the SCXI-1200.
Getting Started with SCXI
xii
www.ni.com
About This Manual
Related Documentation
As you read this manual, you may find it helpful to refer to the following
documents: · Your SCXI chassis, module, terminal block, connector-and-shell,
and
cable assembly user manuals or installation guides · Your DAQ hardware user
manual · Your National Instruments software manuals · Application notes
© National Instruments Corporation
xiii
Getting Started with SCXI
1
Introduction to SCXI
This chapter gives a brief overview of SCXI and helps you identify the
components of an SCXI system.
Figure 1-1 shows how you can use the flexible architecture of SCXI in three
ways–as a front-end signal conditioning system for plug-in DAQ devices, or as
an external DAQ system that attaches either directly to the computer parallel
port or an RS-232 or RS-485 serial port.
Figure 1-1a shows SCXI products as a front-end signal conditioning system for
plug-in DAQ devices. The SCXI chassis can house a variety of signal
conditioning modules. Analog input signals are conditioned and passed to a
single plug-in DAQ device for analog-to-digital conversion. The plug-in DAQ
device also controls and monitors the operation of digital I/O and analog
output modules in the SCXI system. You can monitor and control up to 3,072
signals with a single plug-in DAQ device.
Figure 1-1b shows an SCXI-1200 DAQ module with which you can condition and
digitize your signals in the SCXI itself. The 12-bit SCXI-1200 module in an
SCXI chassis digitizes the conditioned analog input signals and transmits the
digital data to the computer parallel port.
Figure 1-1c shows an SCXI system that uses an SCXI-2000 or SCXI-2400 to
transmit data digitized by an SCXI-1200 to the computer using an RS-232 or
RS-485 serial connection. If you use an RS-485, you can connect multiple SCXI
chassis within a single network. If you use an SCXI chassis without any analog
input modules, you do not need an SCXI-1200 module in that chassis.
© National Instruments Corporation
1-1
Getting Started with SCXI
Chapter 1 Introduction to SCXI
PC Plug-In DAQ Device
Conditioned Signals
SCXI Signal Conditioning
Chassis and Modules
SCXI-1001
SCXI
1140
SCXI
1140
SCXI
1140
SCXI
1140
SCXI MAINFRAME
a. Front-End Signal Conditioning for Plug-In DAQ Device
SCXI Signal Conditioning and
DAQ Modules
SCXI-1001
SCXI
1140
SCXI
1140
SCXI
1140
SCXI
1140
SCXI
1140
SCXI MAINFRAME
Parallel Port Link
SCXI-1200 12-Bit DAQ Module
b. External DAQ System
RS-485
RS-232 or RS-485
SCXI-1001
SCXI
1140
SCXI
1140
SCXI
1140
SCXI
1140
SCXI
1140
SCXI MAINFRAME
SCXI-1200 12-Bit DAQ Module
SCXI-2000 Chassis or SCXI-100X Chassis with
SCXI-2400 Module
c. Remote DAQ System Using Serial Connections Figure 1-1. SCXI Systems
Getting Started with SCXI
1-2
www.ni.com
Chapter 1 Introduction to SCXI
What You Need to Get Started
To set up and use your SCXI system, you will need the following:
One of the following chassis and your SCXI chassis user manual:
SCXI-1000 SCXI-1000DC SCXI-1001 SCXI-2000
SCXI modules and terminal blocks to use in your chassis SCXI-1200 DAQ module
or DAQ device (if using any analog input
SCXI modules)
SCXI cable assembly Your computer NI-DAQ driver software for PC compatibles or
Macintosh One of the following software packages:
LabVIEW for Macintosh LabVIEW for Windows LabWindows/CVI for Windows
ComponentWorks VirtualBench
Identifying Your SCXI Components
The basic components of the SCXI system are terminal blocks, modules, chassis,
cable assemblies, DAQ devices, and software, as shown in Figure 1-2. The
following sections briefly describe the components of an SCXI system and help
you to understand how they work together.
© National Instruments Corporation
1-3
Getting Started with SCXI
Chapter 1 Introduction to SCXI
Computer
Plug-in DAQ Device SCXI Cable Assembly
SCXI Chassis
Computer
Cable to Parallel or Serial Port
SCXI Chassis (with one SCXI-1200)
SCXI Module SCXI Terminal Block or
Terminal Block Extension (TBX)
SCXI Module SCXI Terminal Block or
Terminal Block Extension (TBX)
SCXI Chassis
Figure 1-2. SCXI System Components
The SCXI chassis houses the SCXI modules, supplying power and controlling the
SCXIbus. National Instruments offers four SCXI chassis: · SCXI-1000 4-slot
chassis · SCXI-1000DC 4-slot, DC-powered chassis · SCXI-1001 12-slot chassis ·
SCXI-2000 4-slot, RS-232/485 serial communication chassis
The rugged aluminum SCXI chassis contains an analog bus, a digital bus, and a
chassis controller that regulates bus operation, as shown in Figure 1-3. The
analog bus transfers analog signals from all resident analog input modules to
the DAQ device using a connection from one of the modules to the DAQ device.
The digital lines of the DAQ device communicate with the SCXI chassis
controller and manipulate the digital bus, which controls chassis operation.
Getting Started with SCXI
1-4
www.ni.com
Chapter 1 Introduction to SCXI
The SCXI-2000 chassis also includes a communication processor that you can use
to connect the SCXI system directly to the RS-232 serial port of a computer,
or to an RS-485 network for long-distance data acquisition.
SCXI-1000
Conditioned Signals to DAQ Device or Digitized Volatage
Values to Parallel or Serial Port
MSAINCFRXAMIE
Unconditioned Signals from Transducers
Analog and Digital Bus Backplane
SCXI Modules
Figure 1-3. SCXI Chassis Signal Routing
National Instruments manufactures a variety of SCXI modules for use in SCXI
chassis. Various analog input modules can multiplex, amplify, filter, and
isolate voltage and current signals. Some modules have excitation potentials
or currents and half-bridge completion networks for transducers such as strain
gauges or resistive temperature devices (RTDs). Digital modules isolate
digital signals through relays or optocouplers.
With the SCXI-1181 breadboard module, you can easily implement custom signal
conditioning circuits.
The SCXI-2400 module is a serial communication module with the same RS-232 and
RS-485 capabilities as the SCXI-2000 chassis. Use the SCXI-2400 in an RS-485
network for remote or long-distance data acquisition.
© National Instruments Corporation
1-5
Getting Started with SCXI
Chapter 1 Introduction to SCXI
Product SCXI-1100 SCXI-1102/B/C SCXI-1104/C SCXI-1112 SCXI-1120/D SCXI-1121
SCXI-1122 SCXI-1124 SCXI-1125 SCXI-1126 SCXI-1140 SCXI-1141 SCXI-1142
SCXI-1143 SCXI-1520 SCXI-1530 SCXI-1531 SCXI-1540 SCXI-1160
SCXI-1161 SCXI-1162 SCXI-1162HV
Table 1-1 lists the National Instruments SCXI modules and accessories. For further descriptions of National Instruments SCXI modules, refer to Chapter 3, Configuring Your SCXI Hardware and Software.
Table 1-1. SCXI Modules
Classification Analog input Analog input Analog input Analog input Analog
input Analog input Analog input Analog output Analog input Analog input Analog
input Analog input Analog input Analog input Analog input Analog input Analog
input Analog input Digital
Digital Digital input Digital input
Description 32-channel differential multiplexer/amplifier 32-channel amplifier module for low voltages 32-channel amplifier module for medium voltages 8-channel amplifier module for thermocouples 8-channel isolated amplifier 4-channel isolated amplifier with excitation 16-channel isolated multiplexer with excitation 6-channel isolated D/A converter Jumperless SCXI-1120 8-channel isolated frequency-to-voltage converter 8-channel simultaneously- sampling differential amplifier 8-channel programmable elliptic lowpass filter 8-channel programmable Bessel lowpass filter 8-channel programmable Butterworth lowpass filter 8-channel deluxe bridge with excitation 4-channel accelerometer module 8-channel accelerometer module 8-channel LVDT signal conditioner 16-channel electromechanical single-pole double-throw (SPDT) relay module 8-channel electromechanical SPDT relay module 32-channel optically- isolated digital input module 32-channel optically-isolated high-voltage digital input module
Getting Started with SCXI
1-6
www.ni.com
Chapter 1 Introduction to SCXI
Product SCXI-1163 SCXI-1163R SCXI-1180 SCXI-1181 SCXI-1200
SCXI-2400
Table 1-1. SCXI Modules (Continued)
Classification
Description
Digital output
32-channel optically-isolated digital output
Digital output
32-channel optically-isolated solid-state relay
Accessory
Feedthrough panel
Accessory
Breadboard
Analog, digital, Multifunction module with parallel port communication and timing I/O
Serial communication
RS-232/RS-485 serial communication interface
SCXI Terminal Blocks
SCXI terminal blocks attach directly to the front of an SCXI module and
provide a quick, convenient method for connecting I/O signals to your system.
There are two types of SCXI terminal blocks–direct-mount or TBX.
Direct-mount terminal blocks connect transducers to the screw terminals that
are in a fully shielded terminal block enclosure. Strain-relief clamps hold
the signal wires in place.
Terminal block extensions (TBXs) are DIN rail-mountable and attach to SCXI
modules with shielded cables. These TBX terminal blocks are functionally
equivalent to their direct-mount terminal block counterparts. Figure 1-4 shows
the terminal block options and Table 1-2 lists the modules with which they
work. Table 1-3 lists the TBX terminal blocks and the modules with which they
work.
Several terminal blocks have a temperature sensor for cold-junction
compensation with thermocouples. Terminal blocks can also perform additional
signal conditioning function, including high-voltage attenuation, AC/DC
coupling, bridge offset nulling, and shunt calibration.
© National Instruments Corporation
1-7
Getting Started with SCXI
Chapter 1 Introduction to SCXI
Direct-Mount SCXI
Terminal Block
SCXI Chassis
a. SCXI Direct-Mount Terminal Block TBX Terminal Block
Shielded Cable SCXI Chassis
b. TBX Terminal Block Figure 1-4. Terminal Block Options
Getting Started with SCXI
1-8
www.ni.com
Chapter 1 Introduction to SCXI
Module SCXI-1100, SCXI-1102/B/C
SCXI-1104/C SCXI-1112 SCXI-1120/D, SCXI-1121, SCXI-1125
SCXI-1122 SCXI-1124 SCXI-1126
Table 1-2. Direct-Mount Terminal Blocks
Terminal/ Connector
Block
SCXI-1300
Terminal Block or
CJC
Connector and Shell Sensor
Terminal block
Yes
Other Features —
SCXI-1303
Terminal block
Yes1
SCXI-1308 SCXI-1310 SCXI-1300 N/A
Terminal block
No
Connector and shell No
Terminal block
Yes
No terminal block
N/A
required
SCXI-1320
Terminal block
Yes
SCXI-1321
Terminal block
Yes
(for
SCXI-1121)
SCXI-1327
Terminal block
Yes1
SCXI-1328 SCXI-1330 SCXI-1338 SCXI-1322 SCXI-1325 SCXI-1305
Terminal block
Yes1
Connector and shell No
Terminal block
No
Terminal block
Yes1
Terminal block
No
Terminal block
No
SCXI-1320
Terminal block
Yes
SCXI-1327
Terminal block
Yes1
Isothermal, signal referencing, and open thermocouple detection
020 mA and 420 mA current input
Low-cost connector-and-shell assembly
—
Connect thermocouples directly using mini-thermocoupleconnectors,built-inCJC
sensor per channel
—
Offset-nulling and shunt calibration for strain gauges
100:1 attenuation of voltages to 250 Vrms2, switch configurable per channel
Isothermal design for thermocouples
Low-cost connector-and-shell assembly
020 mA and 420 mA current input
—
—
AC/DC coupling, signal referencing, switch-configurable per channel, BNC
connectors
— 100:1 attenuation of voltages to 250 Vrms2, switch configurable per channel
© National Instruments Corporation
1-9
Getting Started with SCXI
Chapter 1 Introduction to SCXI
Module SCXI-1140
SCXI-1141, SCXI-1142, SCXI-1143
SCXI-1160 SCXI-1161 SCXI-1162/ 1162HV, SCXI-1163/1163R SCXI-1180 SCXI-1181,
SCXI-1181K
SCXI-1200 SCXI-1520
SCXI-1530 SCXI-1531
Table 1-2. Direct-Mount Terminal Blocks (Continued)
Terminal/ Connector
Block SCXI-1301 SCXI-1304
SCXI-1305
SCXI-1310 SCXI-1304
SCXI-1305
SCXI-1310 SCXI-1324
— SCXI-1326
Terminal Block or
CJC
Connector and Shell Sensor
Other Features
Terminal block
No
—
Terminal block
No
AC/DC coupling, switch configurable per
channel
Terminal block
No
AC/DC coupling signal referencing,
switch-configurable per channel, BNC
connectors
Connector and shell No
Low-cost connector and shell assembly
Terminal block
No
AC/DC coupling,
switch-configurable per channel
Terminal block
No
AC/DC coupling, switch-configurable per
channel, BNC connectors
Connector and shell No
Low-cost connector-and-shell assembly
Terminal block
No
—
No terminal block–screw terminals in module
Terminal block
No
—
SCXI-1302 SCXI-1300 SCXI-1301 SCXI-1310 SCXI-1330 SCXI-1302 SCXI-1314
Terminal block
No
Terminal block
Yes
Terminal block
No
Connector and shell No
Connector and shell No
Terminal block
No
Terminal Block
SCXI-1310 N/A
Connector and shell No
No terminal block
No
required
— — — — — — Socketedquarter-bridgecompletionresistor per channel, two socketed shunt-calibration resistors per channel — Connect signals directly using BNC connectors
Getting Started with SCXI
1-10
www.ni.com
Chapter 1 Introduction to SCXI
Table 1-2. Direct-Mount Terminal Blocks (Continued)
Module
Terminal/ Connector
Block
Terminal Block or
CJC
Connector and Shell Sensor
Other Features
SCXI-1540
SCXI-1315
Terminal block
No
—
SCXI-1310
Connector and shell No
—
SCXI-2400
N/A
N/A
N/A
Serial communication module
1 The SCXI-1303, SCXI-1322, SCXI-1327, and SCXI-1328 use an onboard thermistor
for sensing cold-junction temperature. Other terminal blocks use an IC
temperature sensor.
2 300Vrms when used with the SCXI-1125
© National Instruments Corporation
1-11
Getting Started with SCXI
www.ni.com
1-12
Getting Started with SCXI
Term. Blks
TBX-1303
1100/ 1102/ B/C
1120/D 1121 1125
—
Table 1-3. TBX Terminal Blocks for SCXI
1122 —
1124 —
SCXI Modules
1140/ 1141/ 1142/ 1143
1160 1161
—
—
—
1162/ 1162HV/
1163/ 1163R
—
1180 —
TBX-1328
—
TBX-1325
—
TBX-1326
—
TBX-24F
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
CB-50
—
—
—
—
—
—
—
—
1 The TBX-1303 and TBX-1328 use an onboard thermistor for sensing cold- junction temperature.
1181/ 1181 K
— — — —
—
1200
CJC Sensor
Required Cables
—
1
SH96-96 or
R96-96
—
1
SH32-32-A
—
— SH48-48-A
—
— SH48-48-B
—
— User-supplied
wiring
—
NB1
Chapter 1 Introduction to SCXI
Chapter 1 Introduction to SCXI
Connector-and-Shell Assemblies
Connector-and-shell assemblies are low-cost alternatives to SCXI terminal
blocks and are less rugged, plastic assemblies that connect signals using
solder pins.
DAQ Plug-In Devices and PC Cards
You can connect the conditioned analog signals from an SCXI system directly to
a plug-in DAQ device or DAQCard. Using SCXI multiplexing capability, a single
plug-in DAQ device or DAQCard can measure up to 3,072 conditioned analog
signals.
You can use any of the following DAQ devices to control an SCXI system: ·
Multifunction I/O (MIO) and AI devices can control single-chassis or
multichassis SCXI systems consisting of any combination of analog and digital
SCXI modules. Use MIO and AI devices for systems that require high-speed,
hardware-controlled scanning. · You can use all E Series AT, PCMCIA, PCI, and
PXI bus devices except the 6110 and 6111 devices.
Lab/1200 Series devices can control single-chassis SCXI systems consisting of
any combination of analog and digital SCXI modules.
DIO devices can control single-chassis or multichassis SCXI systems consisting
of digital SCXI modules.
DAQ Modules
As an alternative to a plug-in DAQ device or DAQCard, you can use an SCXI DAQ module, such as the SCXI-1200. The SCXI-1200, which is functionally similar to the PCI-1200 and DAQCard-1200, can control single-chassis systems with any combination of analog and digital modules. The SCXI-1200 connects to the computer parallel port, or to the SCXI-2000 chassis or SCXI-2400 module, which in turn communicates with the computer using a serial port.
© National Instruments Corporation
1-13
Getting Started with SCXI
Chapter 1 Introduction to SCXI
SCXI Cable Assemblies
If you are using a plug-in DAQ device or DAQCard, connect it to the SCXI
chassis using one of the SCXI cable assemblies listed in Table 1-4.
Table 1-4. SCXI Cable Assemblies
Cable Assembly SCXI-1343
SCXI-1348
Device Connection
Screw-terminal adapter for custom configurations
DIO-32F
SCXI-1349
MIO E Series devices with 68-pin connector
SCXI-1353
MIO E Series devices with 100-pin connector
SCXI-1354
DAQCard-1200
SCXI-1355
AT-DIO-32HS, PCI-DIO-32HS
NB5 ribbon cable with PC-DIO-96/PnP optional SCXI-1351
R1005050 with optional SCXI-1351
PCI-DIO-96
Some cable assemblies have extra male breakout connectors with which you can
access unused DAQ device signals through the SCXI-1180 feedthrough panel or
the SCXI-1181 breadboard module. The SCXI-1200 comes with a 1 m parallel port
cable.
SCXI Multichassis System with Plug-In DAQ Device
For a multichassis SCXI system, that is cabled using an SCXI-1346 or
SCXI-1350, chassis are daisy-chained using multichassis adapters with shielded
or unshielded cables to a single plug-in DAQ device. You can control up to
eight SCXI-1000, SCXI-1000DC, or SCXI-1001 chassis with a single AI, MIO, or
DIO-24 device. Figure 1-5 shows a multichassis configuration using the
SCXI-1346 multichassis adapter.
Getting Started with SCXI
1-14
www.ni.com
Chapter 1 Introduction to SCXI
SCXI-1349 SCXI-1346 Multichassis Adapters
SH6868 Cables
Computer with MIO Device
SCXI Chassis
Figure 1-5. Multichassis SCXI System with Shielded Cabling
When using shielded cabling for a multichassis configuration, the maximum
allowable cable length from the DAQ device to the last chassis is 10 m.
Optional Software
The final component of an SCXI system is software. You can program your SCXI
system using NI-DAQ driver software, LabWindows/CVI, LabVIEW for Windows,
LabVIEW for Macintosh, or ComponentWorks.
NI-DAQ driver software ships with your National Instruments DAQ device,
SCXI-1200, SCXI-2000, or SCXI-2400. NI-DAQ has a library of functions that you
can call from your application development environment (ADE).
Alternatively, you can use LabVIEW for Windows, LabVIEW for Macintosh,
LabWindows/CVI for Windows, or ComponentWorks. LabVIEW and LabWindows/CVI are
innovative program development software packages for data acquisition and
control applications. LabVIEW uses graphical programming, whereas
LabWindows/CVI enhances ANSI C. Both packages include extensive libraries for
data acquisition, instrument control, data analysis, and graphical data
presentation.
© National Instruments Corporation
1-15
Getting Started with SCXI
Chapter 1 Introduction to SCXI
ComponentWorks is a collection of object linking and embedding (OLE) controls
and dynamic link libraries (DLLs) for acquiring, analyzing, and presenting
data within Visual Basic.
Whether you are using conventional programming languages, LabVIEW,
LabWindows/CVI, or ComponentWorks, your application uses the NI-DAQ driver
software, as illustrated in Figure 1-6.
Conventional Programming Environment (PC or Macintosh)
LabVIEW (PC or Macintosh)
LabWindows/CVI (PC)
Component Works (PC)
NI-DAQ Driver Software
DAQ or SCXI Hardware
Figure 1-6. The Relationship Between the Programming Environment, NI-DAQ, and
the DAQ Hardware
Getting Started with SCXI
1-16
www.ni.com
Chapter 1 Introduction to SCXI
Overview of SCXI Operation
This section describes on the SCXI analog input modules, and analog and
digital output modules used with DAQ devices.
Analog Input Modules
Table 1-1 lists the analog input modules. Analog SCXI modules can operate in
parallel or multiplexed mode. In parallel mode, the modules perform no signal
multiplexing and simply pass the conditioned signals directly to corresponding
channels on the DAQ device. Therefore, each module is directly connected to
its own DAQ device.
You will probably operate your analog SCXI modules in multiplexed mode. You
can multiplex thousands of conditioned analog input signals into a single DAQ
device. In this mode, connect the plug-in DAQ device to only one SCXI module
in a chassis. Digital lines on the DAQ device control the modules and chassis.
The DAQ device uses three or four digital output lines and the EXTSTROBE line
to program the SCXI system. In addition, a digital input line is used for
reading information back from the SCXI modules.
When using SCXI analog input modules, you can measure analog input signals by
doing single-channel readings or hardware-timed, multiple-channel scanning.
During single-channel readings, the DAQ device serially writes a digital
pattern to SCXI Slot 0, located in the chassis, indicating which SCXI module
to address. The DAQ device then writes a digital pattern to the module
indicating which input channel is to be read. This configures the module to
route the desired signal to the analog bus of the SCXI chassis. Ultimately,
the signal is routed to analog input channel 0 of the DAQ device. The DAQ
device then reads channel 0, as shown in Figure 1-7. NI-DAQ automatically
performs this low-level digital communication and signal routing when you call
the single-channel analog input functions.
© National Instruments Corporation
1-17
Getting Started with SCXI
Chapter 1 Introduction to SCXI
Analog Input Signals
SCXI Chassis
Module n
Multiplexing and
Conditioning Circuitry
Module 1
Multiplexing and
Conditioning Circuitry
Slot 0
Input Channel 0 Plug-In DAQ Device
DIO Lines (program the chassis
and modules)
Analog Bus
Figure 1-7. SCXI Signal Routing
With multichannel scanning, the DAQ device programs SCXI Slot 0 with a list of
modules and the number of channels to scan from each module. Each module in
the list is programmed with the channel on which to start the scan. The DAQ
device or module then begins the multichannel scan. The SCANCLK signal from
the AI or MIO device synchronizes the SCXI multiplexing with the internal
clock that triggers the A/D conversions on the DAQ device. SCXI Slot 0 enables
and disables the modules according to the preprogrammed list. In this way, the
system multiplexes channels from several modules to an analog input channel of
the DAQ device at very high rates. NI-DAQ contains high-level functions you
can use for multichannel scanning operations.
Getting Started with SCXI
1-18
www.ni.com
Chapter 1 Introduction to SCXI
Analog and Digital Output Modules
Digital SCXI modules include the SCXI-116X modules. Digital SCXI modules can
operate in either serial mode or parallel mode. In serial mode, the SCXI
system uses a maximum of four digital output lines and one digital input line
of the DAQ device to control and monitor multiple digital modules. The plug-in
device serially writes digital patterns to set the state of the digital output
lines or relays, or serially reads patterns from the modules to determine the
state of the digital input lines. In parallel mode, each SCXI digital I/O line
corresponds to a digital I/O line of the DAQ device.
The SCXI-1162, SCXI-1162HV, SCXI-1163, and SCXI-1163R can operate in parallel
mode. The SCXI-1160 and SCXI-1161 operate only in serial mode. Refer to
Chapter 3, Configuring Your SCXI Hardware and Software, for more information.
The SCXI-1124 is a 6-channel isolated analog output module. As with the
digital I/O modules, the SCXI-1124 is programmed serially by the DAQ device
using a maximum of four digital output lines and one digital input line.
© National Instruments Corporation
1-19
Getting Started with SCXI
2
Installing SCXI Modules and Connecting the SCXI System to the DAQ Device
This chapter describes how to install your SCXI hardware and how to connect
your SCXI system to your computer.
Installing Your Software
All SCXI system configurations require NI-DAQ software. Whether you are using
LabVIEW, LabWindows/CVI, ComponentWorks, or a third-party compiler, you must
install the NI-DAQ driver software and Measurement & Automation Explorer to
use your DAQ/SCXI system. 1. You receive NI-DAQ software with your DAQ device,
SCXI-1200,
SCXI-2400, or SCXI-2000. · If you are using LabWindows/CVI, install
LabWindows/CVI first.
Next, install NI-DAQ for PC compatibles and run the setup option, then select
the LabWindows/CVI option to install LabWindows/CVI DAQ support. · LabVIEW
includes all the NI-DAQ files you need to control the SCXI system. However,
NI-DAQ for PC compatibles also has a LabVIEW update option that installs newer
driver files in your system when the installation utility detects older files
on your hard disk. 2. If you already have installed Measurement & Automation
Explorer and the driver software and you received a newer NI-DAQ version,
install the newer version now.
© National Instruments Corporation
2-1
Getting Started with SCXI
Chapter 2 Installing SCXI Modules and Connecting the SCXI System to the DAQ
Device
Installing Your DAQ Device
If you are using SCXI with a plug-in DAQ device or DAQCard, install your DAQ
hardware into your computer now and configure your DAQ device with Measurement
& Automation Explorer. Refer to your DAQ device user manual for installation
instructions.
Installing Your SCXI Modules
Caution Before installing an SCXI module or removing a module from the
chassis, turn off power to the chassis.
Follow this procedure to install your SCXI modules: 1. To insert a module into
the chassis, line up the module with the chassis
device guides. Slide the module back until it connects smoothly to the SCXI
chassis backplane. Do not force the module. If the module does not connect
smoothly, check that it is aligned correctly. 2. Tighten the thumbscrews
finger-tight. Do not over-tighten. 3. Repeat steps 1 and 2 until you have
installed all of your modules.
Connecting the SCXI System to the DAQ Device or Computer
This section explains how to connect your SCXI to a DAQ device or a computer
using a parallel port, plug-in DAQ device or DAQCard, and an RS-232 or RS-485
Serial port.
Using the Parallel Port
Connect the parallel port cable to the computer parallel port. Connect the
other end of the cable to the SCXI-1200 and tighten the mounting screws on the
connectors to establish a firm connection.
Using a Plug-in DAQ Device or DAQCard
If you are using an SCXI module with a DAQ device or DAQCard, connect the SCXI
chassis to your DAQ device with an SCXI-134X or SCXI-135X cable assembly. In
certain instances, you can also use a 50-pin NB1 or 100-pin NB5 or NB6 ribbon
cable in conjunction with an SCXI-1351 cable assembly.
Getting Started with SCXI
2-2
www.ni.com
Chapter 2 Installing SCXI Modules and Connecting the SCXI System to the DAQ
Device
To connect the SCXI chassis to your DAQ device, you must connect the
appropriate SCXI cable assembly to the DAQ device and then connect the other
end to the rear connector of a single SCXI module. Use the following
procedure:
1. Determine to which SCXI module you will connect the SCXI cable assembly.
If all modules are configured for multiplexed or serial mode, you must cable
only one module to a DAQ device. Use the following rules to determine which is
the proper SCXI module to connect to the cable:
· For systems containing at least one analog SCXI module:
a. If you are using one or more SCXI-1140 modules, connect the cable assembly
to any one of the SCXI-1140 modules. The SCXI-1140 requires a counter signal
from the DAQ device to control the track-and-hold circuitry. Cable only one
SCXI-1140 to the DAQ device because other SCXI-1140 modules in the chassis can
access the track-and-hold counter signal through the SCXIbus in the chassis
backplane.
b. If you are not using an SCXI-1140 and are using one or more SCXI-110X
module, connect the cable assembly to any one of the SCXI-110X module. The
SCXI-110X module drives only two of the DAQ device analog input lines, leaving
the other lines available.
c. If you are not using any SCXI-1140 or SCXI-110X modules but are using other
analog modules, connect the cable assembly to any one of the analog modules.
· For systems containing only digital modules (no analog modules), connect the
cable assembly to any one of the digital modules.
· For systems containing both analog and digital modules, follow steps a
through c for connecting an analog module. Enable the MISO and SERDATOUT lines
of the module you connect to the DAQ device. For more information see Chapter
3, Configuring Your SCXI Hardware and Software.
· If the system contains any digital modules configured for parallel mode,
connect each of these digital modules to a separate DIO device. If the system
contains any analog modules configured for parallel mode, connect each of
these modules to a separate analog DAQ device.
© National Instruments Corporation
2-3
Getting Started with SCXI
Chapter 2 Installing SCXI Modules and Connecting the SCXI System to the DAQ
Device
2. Connect the connector at the bracket end of the SCXI cable assembly to the
rear signal connector of the module. Figure 2-1 shows the cable assembly
attached to the SCXI module and chassis.
Female Connector to Module (Rear panel omitted for clarity)
SCXI
Female Connector to DAQ Device
Male Breakout Connector (You can use the breakout connector, in conjunction
with a feedthrough panel, to access unused lines from the DAQ device. You do
not need to connect the breakout connector to any other module.) Rear of SCXI
Chassis
Figure 2-1. Attaching the SCXI Cable Assembly to the SCXI Module and Chassis
Getting Started with SCXI
2-4
www.ni.com
Chapter 2 Installing SCXI Modules and Connecting the SCXI System to the DAQ
Device
The DAQ device communicates with the SCXIbus through the connector at the
bracket end of the cable assembly. Therefore, you do not need to connect the
remaining modules to the SCXI cable assembly. Notice that the extra male
breakout connector of the SCXI cable assembly is usually not connected to
other modules. You can use this male breakout connector to connect
unconditioned signals from the DAQ device to an SCXI-1180 feedthrough panel,
SCXI-1181 breadboard module, or the SCXI-1351 one-slot cable extender.
In some cases, you can use an NB1 cable instead of the SCXI-1340 cable
assembly. The NB1 cable and the SCXI-1340 are functionally identical; however,
the SCXI-1340 ensures a more secure attachment than the NB1 cable, eliminating
the possibility of inadvertently loosening the cable. The SCXI-1340 also has
an extra male breakout connector that provides access to unconditioned signals
using an SCXI-1180 feedthrough panel or SCXI-1181 breadboard module.
Cable assemblies also have an integral rear panel that you can fasten to the
SCXI chassis with two screws. Using a cable assembly, you can remove a module
from or insert a module into the chassis without disconnecting the cable.
3. If you are using an SCXI-1180 feedthrough panel to access unused lines of
the DAQ device, install the feedthrough panel in the slot directly to the
right of the module cabled to the DAQ device. Attach the SCXI-1180 feedthrough
panel cable to the 50-pin male breakout connector of the SCXI cable assembly
connected to the adjacent module.
4. If you are using an SCXI-1181 breadboard module, install the breadboard
module in the slot directly to the right of the module that is connected to
the DAQ device. Attach an SCXI-1351 one-slot cable extender to the 50-pin male
breakout connector of the SCXI cable assembly connected to the adjacent
module. Attach the other end of the SCXI-1351 to the rear signal connector of
the SCXI-1181 breadboard module.
If you want to use more than 12 modules with a single DAQ device, you must use
more than one SCXI chassis. In this case, you must use multichassis adapters.
© National Instruments Corporation
2-5
Getting Started with SCXI
Chapter 2 Installing SCXI Modules and Connecting the SCXI System to the DAQ
Device
Using an RS-232 or RS-485 Serial Port
You can use the RS-232 port on your computer to control one chassis up to 100
ft from your computer. If your chassis is farther away or you are using
multiple chassis, you must use RS-485.
If you are using an SCXI-2400 module, insert the SCXI-2400 into the far left
slot of your SCXI-100X chassis and insert the SCXI-1200 module immediately
next to it. Attach the connector adapter that is shipped with the SCXI-2400
module to the rear of the SCXI-2400. Connect the SCXI-1200 to this connector
adapter using a 7 in. IEEE 1284 parallel port cable.
If you are using an SCXI-2000 chassis, insert the SCXI-1200 into the far left
slot of your SCXI chassis. Next, connect the 7 in. IEEE 1284 parallel port
cable to the back of the SCXI-1200 and the SCXI-2000 chassis.
Note If your SCXI chassis includes only analog output and digital I/O modules,
you do not need the SCXI-1200 module. The SCXI-2000 and SCXI-2400 can
communicate directly with analog output and digital I/O modules.
Next, connect the SCXI-2000 or SCXI-2400 to an RS-232 or RS-485 port of your
computer. · RS-232–You can connect one SCXI chassis directly to the standard
COM serial port of your computer. If your computer has a 9-pin communication
port, use the 2 m null modem cable included with the SCXI-2000/2400 to connect
the 9-pin RS-232 connector on the front panel of the SCXI-2000/2400 to the
serial port of your computer. · RS-485–Alternatively, you can connect the
RS-485 connectors on the back of the SCXI-2000 chassis or SCXI-2400 module to
an RS-485 network. For example, you can wire the RS-485 connector directly to
a port of the plug-in AT-485 adapter device. The RS-485 connection is a four-
wire connection. See your SCXI chassis user manual or SCXI-2400 User Manual
for more information on configuring an RS-485 network.
Getting Started with SCXI
2-6
www.ni.com
3
Configuring Your SCXI Hardware and Software
This chapter explains how to configure your SCXI hardware and National
Instruments software.
Module Configuration
You may need to configure your SCXI modules before using them in your
application. Some SCXI modules have a number of user-configurable jumpers that
you can access by removing a single screw on the back of the module, as shown
in Figure 3-1.
Rear of Module
Grounding Screw
© National Instruments Corporation
Figure 3-1. Removing the SCXI Module Grounding Screw
3-1
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Use this procedure to access the user-configurable jumpers. 1. Remove the
grounding screw shown in Figure 3-1. 2. Remove the cover by inserting a
flathead screwdriver in the slot at the
bottom of the module and gently pushing the screwdriver down until the cover
pops off, as shown in Figure 3-2.
Grounding Screw
Removable Cover
Top of Module
Front Connector
Figure 3-2. Removing the SCXI Module Cover
To replace the cover, follow these steps: 1. Align the cover with the slot at
the top of the module and snap the
bottom of the cover into place. 2. Replace the grounding screw, taking care
not to over-tighten it.
Analog Modules
The following sections describe each analog SCXI module and contain step-by-
step instructions for configuring the jumpers on each module. Follow the steps
in the procedure for your module, check the position of each jumper, and read
the jumper position description. Change the appropriate jumpers to configure
the module for your application.
Getting Started with SCXI
3-2
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Each analog module includes jumpers that configure the output referencing mode
of the module. Configure these jumpers according to the analog input mode of
the DAQ device. The recommended analog input mode of your DAQ device is as
follows:
· E Series AI and MIO devices default to differential (DIFF) mode. National
Instruments recommends that you leave your AI or MIO device configured for
DIFF inputs.
· The Lab/1200 devices default to referenced single-ended (RSE) mode. Change
its configuration to non-referenced single-ended (NRSE) mode for optimal noise
elimination. Because of the cabling, you cannot use the Lab and 1200 devices
in DIFF mode with SCXI.
Some analog SCXI modules include jumpers to configure module parameters such
as filter cut off frequency, gain, analog input referencing, and output
referencing. You must select the analog input referencing of a module
according to the type of signal you are measuring. This section begins with a
description of the main types of signals.
Ground-Referenced and Floating Signal Sources
Signal sources are either ground-referenced or floating.
A ground-referenced source is connected to the building ground and thus shares
a common ground with the DAQ device, assuming that your computer is connected
to the building power supply as well. Examples of ground-referenced signals
are the non-isolated outputs of any devices that plug into the building power
supply, such as signal generators and power supplies. Isolated outputs are not
ground-referenced sources; you should treat them as floating sources.
A floating source has an isolated ground-reference point that you cannot
assume to be at the same potential as any other ground reference. Examples of
this type of source include transformers, insulated thermocouples, optical
isolators, isolation amplifiers, and battery-powered devices.
In general, you should reference an input signal to ground at only one point.
Therefore, do not reference the input of a differential amplifier module to
ground if the signal source is already ground-referenced. If you are measuring
a floating source, on the other hand, reference the input signal to ground.
© National Instruments Corporation
3-3
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
SCXI-1100 Module
This module is a 32-channel differential-input multiplexer with an onboard
programmable-gain instrumentation amplifier (PGIA). The SCXI-1100 has 32
differential voltage or current input channels, an analog input range of ±10
V, jumper-selectable lowpass filters of 4 Hz and 10 kHz, and software-
selectable module gains of 1, 2, 5, 10, 20, 50, 100, 200, 500, 1,000, and
2,000. The SCXI-1100 works only in multiplexed mode and drives two analog
input channels on the DAQ device–channel 0 and channel 1.
The SCXI-1100 is factory-configured for use with an MIO or AI device in DIFF
mode, using ground-referenced signals, with no filtering. If this
configuration is suitable, do not reposition any jumpers. However, you must
adjust the configuration if any of the following conditions is true: · You are
cabling this SCXI-1100 directly to a DAQ device and you are
using a Lab or 1200 device, or you are using an AI or MIO device in SE mode. ·
Your input signals are floating or nonreferenced and you want to ground
reference the signals through a 100 k resistor installed in the module;
however, it is best to ground-reference the signals at the source or at the
terminal block. · You need hardware filtering and can accept slower scan
rates. · You are measuring current signals.
Read the rest of this section on the SCXI-1100 module if any of the previous
conditions applies to you.
The SCXI-1100 module has five user-configurable jumpers for configuring input
signal referencing, filtering, and PGIA output referencing. Leave all other
jumpers in their factory-default positions.
Use the following procedure to configure your SCXI-1100 module: 1. Configure
the input signal referencing mode.
For ground-referenced sources, leave jumper W1 in the factory-default
position. This is the preferred position.
Caution If you plan to use any floating thermocouples or other floating-source
transducers, National Instruments recommends that you leave jumper W1 in the
factory-default setting and either ground the negative lead of each transducer
to the chassis ground on your terminal block, or reference the transducer to
ground at the source. In either case, the transducer must not have a high
common-mode voltage. Ground your signal at
Getting Started with SCXI
3-4
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
only one location along the entire signal path to prevent saturation of the
amplifier and reduce common-mode noise.
Table 3-1. SCXI-1100 Input Signal Referencing Jumper Configurations
Jumper
Description
Configuration
1-2 Position–Ground-referenced source
W1
position (factory-default position).
1
2
3
2-3 Position–Floating (nonreferenced)
source position (connects an internal 100 k
1
resistor from CH- to ground). Read the text in
this section for the implications of using this
2
position.
3
If all of the sources are floating, you can configure jumper W1 to connect a
100 k resistor to the negative input of the amplifier to prevent saturation.
This reduces the input impedance, however, and usually increases settling time
and common-mode noise. Also, if W1 is connected to the 100 k resistor and any
of the sources are ground-referenced or have high leakage to ground, a ground
loop can result, causing DC offsets or noise.
2. Configure the filtering.
The SCXI-1100 has two lowpass, one-pole resistance-capacitor (RC) filters,
with bandwidths of 10 kHz and 4 Hz, positioned after the amplifier. You select
filtering by setting a single jumper to one of three positions–W2, W3, or W4.
Table 3-2 shows the SCXI-1100 filtering jumper configuration.
© National Instruments Corporation
3-5
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W2 W3 W4
Table 3-2. SCXI-1100 Filtering Jumper Configuration
Description
Configuration
W2–Full bandwidth; no filtering (factory-default position).
W3–10 kHz lowpass filter; maximum aggregate sampling rate is 6.7 kS/s.
W2 W3 W4
W4–4 Hz lowpass filter; maximum aggregate sampling rate is 2.9 S/s.
W2 W3 W4
W2 W3 W4
Position W2, the factory-default position, disables filtering. Position W3
selects the 10 kHz lowpass filter, which filters out signal components with
frequencies above 10 kHz. Position W4 selects the 4 Hz lowpass filter, which
filters out signal components with frequencies greater than 4 Hz, such as 60
Hz noise. For more information, refer to the Noise and Filtering section in
Chapter 4, Connecting and Using Accessories and Transducers.
Caution If you enable filtering, you cannot scan SCXI-1100 channels at high
rates. Because the filter is applied after the signal is multiplexed, the
sample rate or channel rate must be low enough for the filter to settle to the
required accuracy. Table 3-2 shows the maximum recommended aggregate sampling
rates for settling under worst-case conditions.
3. Configure the PGIA output referencing mode.
You use jumper W10 to select the output referencing mode of the SCXI-1100
PGIA. The correct output referencing mode depends on the input mode of the DAQ
device. Table 3-3 shows three configurations for jumper W10.
Getting Started with SCXI
3-6
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-3. SCXI-1100 PGIA Output Referencing Jumper Configuration
Jumper
Description
Configuration
A-R0R1–Parking position. Use this position for
W10
MIO or AI DAQ devices in DIFF input mode
AB
(factory-default position).
R0
R1
R2
AB-R0–Connects the PGIA reference to data acquisition analog ground, pins 1 and 2, of the rear signal connector. Use this position for DAQ devices with SE inputs, such as the Lab and 1200 devices, or MIO and AI devices in RSE mode.
AB R0 R1 R2
AB-R2–Select this position for AI or MIO devices in NRSE mode.
AB R0 R1 R2
4. Add current-loop receivers, if necessary. Process-current resistor packs
are available from National Instruments. Refer to Input Filtering and Current-
Loop Receivers in the SCXI-1100 User Manual for detailed information about
filter pads and current loop receivers.
SCXI-1102/B/C Module
The SCXI-1102/B/C is a 32-channel amplifier module designed for measuring
thermocouples and other low-bandwidth signals. Each of the 32 channels
includes input protection circuitry to 42 VACpeak and a software-selectable
gain of 1 or 100. The SCXI-1102, SCXI-1102/B, and SCXI-1102/C have lowpass
filters with cutoff frequencies of 2 Hz, 200 Hz, and 10 kHz respectively. The
SCXI-1102 works only in multiplexed mode and drives a single analog input
channel on the DAQ device (channel 0). The SCXI-1102/B/C is software
configurable and contains no jumpers.
© National Instruments Corporation
3-7
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
SCXI-1104/C Module
SCXI-1104 This module is for signal conditioning of low voltage and medium
voltage signals. The SCXI-1104 has 32 differential analog input channels. On
each channel, the SCXI-1104 has a three-pole lowpass filter with a 2 Hz cutoff
frequency to reject 60 Hz noise. Each channel also has a divide by 10
attenuator stage before the amplifier. You can multiplex the SCXI-1104 inputs
to a single output, which drives a single DAQ device channel.
SCXI-1104C This module is for signal conditioning of low voltage and medium
voltage signals. The SCXI-1104C has 32 differential analog input channels. On
each channel, the SCXI-1104C has a three-pole lowpass filter with a 10 kHz
cutoff frequency. Each channel also has a divide by 10 attenuator stage before
the amplifier. You can multiplex the SCXI-1104C inputs to a single output,
which drives a single DAQ device channel.
SCXI-1112 Module
The SCXI-1112 module is an 8-channel module designed for measuring temperature
using thermocouples. Each of the eight channels includes input protection
circuitry to 42 V, a lowpass 2 Hz noise filter, and a fixed gain of 100. The
SCXI-1112 works only in multiplexed mode. Use Measurement & Automation
Explorer to configure the SCXI-1112. The SCXI-1112 is software configurable
and contains no jumpers.
SCXI-1120/D Module
The SCXI-1120/D module is an 8-channel isolation amplifier. Features of the
SCXI-1120/D include eight isolated input channels, 250 Vrms common-mode
voltage range, jumper-selectable lowpass filtering, and individually
configurable jumper-selectable channel gains of 1, 2, 5, 10, 20, 50, 100, 200,
500, 1,000, and 2,000.
The SCXI-1120/D is factory configured for use with an MIO or AI device in DIFF
mode; each module channel is configured for a gain of 1,000 (input range of ±5
mV), and 4 Hz filtering (4.5 kHz for the SCXI-1120/D). If this configuration
is suitable for your application, do not reposition any of the
Getting Started with SCXI
3-8
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
jumpers. However, you must adjust the jumper configuration if any of the
following conditions is true:
· You are cabling this SCXI-1120/D directly to a DAQ device and you are using
a Lab or 1200 device, or you are using an AI or MIO device in SE mode.
· You need gain or bandwidth different from the default value on any channel.
· You are using digital modules in the same chassis and are cabling the
SCXI-1120/D to a DAQ device.
Read the rest of this section on the SCXI-1120/D module if any of the previous conditions applies to you.
The SCXI-1120/D has 42 user-configurable jumpers, many of which perform the same function but configure different channels. You can set the jumpers to configure gain, filtering, temperature sensor signal routing, and signal output referencing. Leave all other jumpers in the factory-default positions.
Use the following procedure to configure your SCXI-1120/D module:
1. Configure the gains.
Each input channel has two user-configurable gain stages. The first gain stage
provides gains of 1, 10, 50, and 100, and the second stage provides gains of
1, 2, 5, 10, and 20. Tables 3-4 and 3-5 show how to set up the gain for each
channel.
Table 3-4. Gain Jumper Allocation
Input Channel Number
First Gain Jumper
Second Gain Jumper
0
W1
W9
1
W2
W10
2
W3
W11
3
W4
W12
4
W5
W13
5
W6
W14
© National Instruments Corporation
3-9
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-4. Gain Jumper Allocation (Continued)
Input Channel Number
First Gain Jumper
Second Gain Jumper
6
W7
W15
7
W8
W16
The SCXI-1120D has an additional fixed pre-stage gain of 0.5.
The SCXI-1120 is shipped with the first-stage gain set to 100 (position A),
and a second-stage gain set to 10 (position D). The SCXI-1120D is shipped with
the first-stage gain set to 100 (position A), and the second-stage gain set to
20 (position E). To change the gain of your module, move the appropriate
jumper on your module to the position indicated in Table 3-5.
To determine the overall gain of a given channel on the SCXI-1120 use the
following formula:
Overall gain = First stage gain · Second stage gain
To determine the overall gain of a given channel on the SCXI-1120D use the
following formula:
Overall gain
=
1-2
·
First
stage gain · Second
stage gain
Table 3-5. Gain Jumper Positions
Gain
Setting
Jumper Position
First-stage
1
D
10
C
50
B
100
A (factory-default
setting)
Getting Started with SCXI
3-10
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-5. Gain Jumper Positions (Continued)
Gain
Setting
Jumper Position
Second-stage
1
A
2
B
5
C
10
D (factory-default
setting for the
SCXI-1120)
20
E (factory-default
setting for the
SCXI-1120D)
2. Configure the filtering.
SCXI-1120 Filter Jumpers
Two-stage filtering is also available on your SCXI-1120 module. The first
stage is located in the isolated section of the input channel, whereas the
second stage is located in the nonisolated section of your input channel. Two-
stage filtering eliminates the noise generated by the isolation amplifier,
producing a higher signal-to-noise ratio. Furthermore, two filter bandwidths
are available, 10 kHz and 4 Hz.
Table 3-6. SCXI-1120 Filter Jumper Allocation
First Filter Jumper
Second Filter Jumper
Input Channel Number
4 Hz (Factory Default)
10 kHz
4 Hz (Factory Default)
10 kHz
0
W17-A
W17-B
W25
W26
1
W18-A
W18-B
W27
W28
2
W19-A
W19-B
W29
W30
3
W20-A
W20-B
W31
W32
© National Instruments Corporation
3-11
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-6. SCXI-1120 Filter Jumper Allocation (Continued)
First Filter Jumper
Second Filter Jumper
Input Channel Number
4 Hz (Factory Default)
10 kHz
4 Hz (Factory Default)
10 kHz
4
W21-A
W21-B
W33
W34
5
W22-A
W22-B
W35
W36
6
W23-A
W23-B
W37
W38
7
W24-A
W24-B
W39
W40
Your SCXI-1120 is shipped in the 4 Hz position. Verify that both stages are
set to the same bandwidth to ensure that the required bandwidth is achieved.
Notice that one jumper block is available for each filter stage.
SCXI-1120D Filter Jumpers
Two-stage filtering is also available on your SCXI-1120D module. The first
stage is located in the isolated section of the input channel, whereas the
second stage is located in the nonisolated section of your input channel. Two-
stage filtering eliminates the noise generated by the isolation amplifier,
producing a higher signal-to-noise ratio. Furthermore, two filter bandwidths
are available, 22.5 kHz and 4.5 kHz.
Table 3-7. SCXI-1120D Filter Jumper Allocation
First Filter Jumper
Second Filter Jumper
Input Channel Number
4.5 kHz (Factory Default)
22.5 kHz
22.5 kHz
4.5 kHz (Factory Default)
0
W17-A
W17-B
W25
W26
1
W18-A
W18-B
W27
W28
2
W19-A
W19-B
W29
W30
Getting Started with SCXI
3-12
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-7. SCXI-1120D Filter Jumper Allocation (Continued)
First Filter Jumper
Second Filter Jumper
Input Channel Number
4.5 kHz (Factory Default)
22.5 kHz
22.5 kHz
4.5 kHz (Factory Default)
3
W20-A
W20-B
W31
W32
4
W21-A
W21-B
W33
W34
5
W22-A
W22-B
W35
W36
6
W23-A
W23-B
W37
W38
7
W24-A
W24-B
W39
W40
Your SCXI-1120D is shipped in the 4.5 kHz position. Verify that both stages
are set to the same bandwidth to ensure that the required bandwidth is
achieved. Notice that one jumper block is available for each filter stage.
3. Configure the temperature sensor signal routing, if necessary.
Jumper W41 controls the routing of the temperature sensor signal available on
the SCXI-1320 terminal block. Use position 1-2 (the factory-default position)
for MTEMP (multiplexed) mode. Position 2-3 for DTEMP (direct) mode is no
longer supported by the current version of the SCXI-1320 terminal block.
4. Configure the amplifier output referencing mode.
The signal output referencing is relevant only if you are connecting this
SCXI-1120/D module directly to a DAQ device.
Use jumper W46 to configure the various referencing modes of the SCXI-1120/D
instrumentation amplifier output. The correct output referencing mode depends
on the input mode of the DAQ device you use.
© National Instruments Corporation
3-13
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-8 shows the signal referencing configurations for jumper W46.
Table 3-8. SCXI-1120/D Signal Referencing Jumper Configuration
Jumper
W46
Description
B-R0R1–Parking position. Use this position for MIO or AI devices in DIFF input
mode. (factory-default position).
Configuration
B A R2 R1 R0
AB-R2–Select this position for AI or MIO devices in NRSE mode.
B A R2 R1 R0
AB-R0–Connects the PGIA reference to data acquisition analog ground, pins 1 and 2, of the rear signal connector. Use this position for DAQ devices with SE inputs, such as the Lab and 1200 devices, or MIO and AI devices in RSE mode.
B A R2 R1 R0
5. If you are using any digital SCXI modules in serial mode and want to
connect the SCXI-1120/D module to the plug-in DAQ device, configure the
SCXI-1120/D module to connect the MISO signal to the SERDATOUT pin (jumper
W43).
SCXI-1121 Module
This module is a 4-channel isolation amplifier with excitation. Major features
include four channels of isolated input, 1,500 Vrms isolation per channel,
current or voltage excitation, an analog input range of ±5 V, 240 Vrms common-
mode voltage, jumper-selectable lowpass three-pole RC filters with values of 4
Hz or 10 kHz, and individually configurable jumper-selectable channel gains of
1, 2, 5, 10, 20, 50, 100, 200, 500, 1,000, and 2,000.
The SCXI-1121 module has 45 user-configurable jumpers. Position the jumpers to
configure gain, filtering, excitation mode, bridge completion, and signal
output referencing. Leave all other jumpers in their factory-default
positions.
Getting Started with SCXI
3-14
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
The SCXI-1121 is factory-configured for use with an MIO or AI device in DIFF
mode. Each module channel is configured for a gain of 1,000 (an input range of
±5 mV), 4 Hz filtering, 3.33 V voltage excitation, and disabled bridge
completion. If this configuration is suitable for your application, do not
reposition any of the jumpers. However, you must adjust the jumper
configuration if any of the following conditions is true:
· You are cabling the SCXI-1121 module directly to a DAQ device and you are
using a Lab or 1200 device, or you are using an AI or MIO device in SE mode.
· You need a different gain or bandwidth (other than 1,000 and 4 Hz) on any
channel.
· Your transducers require current excitation or 10.0 V voltage excitation.
· Your transducers require half-bridge completion circuitry.
· You are using digital modules and are cabling the SCXI-1121 to a DAQ device.
Read the rest of this section on the SCXI-1121 module if any of the previous conditions applies to you.
Use the following procedure to configure your SCXI-1121 module:
1. Configure the gain.
You select the gain for each SCXI-1121 channel in two stages, each of which is
jumper controlled. The default gain for each channel is 1,000.
The total gain produced on a particular channel is equal to the product of the
first-stage and second-stage gains applied to that channel, as shown by the
following equation:
Total gain (channel n) = first-stage gain (channel n) · second-stage gain
(channel n)
The effective input range that you can measure with a particular gain is:
Input range(V)
=
——-±—-5—V——-total gain
a. Configure the first-stage gain selection.
Use jumpers W3, W19, W29, and W41 to configure the first-stage gain of
channels 0 through 3, respectively. Each jumper has four possible positions–A,
B, C, and D–each corresponding to a particular gain.
© National Instruments Corporation
3-15
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
b. Configure the second-stage gain.
Use jumpers W4, W20, W30, and W42 to configure the second-stage gain of
channels 0 through 3, respectively. Each of these jumpers has five possible
positions–A, B, C, D, and E–each corresponding to a particular gain.
The jumper settings and corresponding gain values are listed in Tables 3-9 and 3-10.
Table 3-9. SCXI-1121 First-Stage and Second-Stage Gain Jumper Configuration
Input Channel Number 0 1 2 3
First-Stage Gain Jumper W3 W19 W29 W41
Second-Stage Gain Jumper W4 W20 W30 W42
Table 3-10. SCXI-1121 Gain Jumper Configuration
Gain
Setting
Jumper Position
First-stage
1
D
10
C
50
B
100
A (factory setting)
Second-stage
1
A
2
B
5
C
10
D (factory setting)
20
E
2. Configure the filtering.
The filtering is also two-stage and is jumper-selectable for each channel. The
SCXI-1121 uses a two-stage filtering configuration to achieve a higher signal-
to-noise ratio. Filter bandwidth does not affect the maximum scanning rate.
The factory-default jumper positions configures each channel for 4 Hz
bandwidth.
Getting Started with SCXI
3-16
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
a. Configure the first-stage filtering.
Jumpers W5, W21, W31, and W43 select the first-stage filtering. Each of these
jumpers has two positions–A and B. Position A selects the first stage of 4 Hz
filtering, whereas position B selects the first stage of 10 kHz filtering.
b. Configure the second-stage filtering.
Jumpers W6 through W13 select the second-stage filtering, which you must
configure for the same frequency as that of the first stage. There are
actually only four jumpers, as in the first stage; W6 through W13 correspond
to eight positions, rather than eight jumpers. You can set each jumper to one
of two positions. Setting the jumper to an even number (for example, W6)
selects the second stage of 4 Hz filtering, whereas setting the jumper to an
odd number (for example, W7) sets the second stage of the same channel to 10
kHz filtering. Refer to Table 3-11 for the jumper and channel correspondence
for filtering.
Table 3-11. SCXI-1121 First-Stage and Second-Stage Filter Jumper Configuration
Input Channel Number
0
1
First-Filter Jumper
4 Hz (Factory-Default)
10 kHz
W5-A
W5-B
W21-A
W21-B
Second-Filter Jumper
4 Hz (Factory-Default)
10 kHz
W6
W7
W8
W9
2
W31-A
W31-B
W10
W11
3
W43-A
W43-B
W12
W13
3. Configure the excitation mode as current or voltage.
You can configure each SCXI-1121 excitation channel to produce voltage or
current excitation. Strain gauges require voltage excitation, whereas RTDs and
thermistors typically require current excitation. Each excitation channel has
two jumpers to select the appropriate mode–W14, W15, W22, W23, W34, W35, W46,
and W47. Table 3-12 summarizes the excitation mode jumper settings. Notice
that only position 1 is marked on the SCXI-1121 module. The factory-default
excitation mode for each channel is voltage.
© National Instruments Corporation
3-17
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumpers
W14 W15
Table 3-12. SCXI-1121 Excitation Mode Jumper Configuration
Excitation
Voltage Mode
Channel (Factory-Default Position)
Current Mode
0
W14
W15
W14
W15
123
123
123
123
W22 W23
1
W22
W23
W22
W23
123
123
123
123
W34 W35
2
W34
W35
W34
W35
123
123
123
123
W46 W47
3
W46
W47
W46
W47
123
123
123
123
Getting Started with SCXI
3-18
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
4. Configure the excitation level. You can select the level of excitation
applied to each excitation channel, again using two jumpers per channel–W16,
W24, W25, W26, W36, W37, W48, and W49. If you select voltage mode, you can
apply a potential difference of either 3.333 V or 10.00 V. If you select
current mode, you can apply an excitation current of either 0.150 mA or 0.450
mA. Refer to the transducer specifications to determine the necessary
excitation level. See Table 3-13 for the SCXI-1121 excitation level jumper
settings, and refer to Table 3-14 for the maximum load per excitation channel.
© National Instruments Corporation
3-19
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumpers
W16 W26
Table 3-13. SCXI-1121 Excitation Level Jumper Configuration
Excitation
3.333 V or 0.150 mA
Channel (Factory-Default Position)
10 V or 0.450 mA
0
W26
W26
W16
1
2
123 3
W16
1
2
123 3
W24 W25
1
W24
W25
W24
W25
123
123
123
123
W36 W37
2
W36
W37
W36
W37
123
123
123
123
W48 W49
3
W48
W49
W48
W49
123
123
123
123
Getting Started with SCXI
3-20
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-14. SCXI-1121 Maximum Load per Excitation Channel
Excitation Level
Maximum Load
3.333 V
28 mA
10 V 0.150 mA 0.450 mA
14 mA 10 k 10 k
5. Enable the half-bridge completion network, if necessary.
The SCXI-1121 includes a built-in half-bridge completion network to use with
half-bridge or quarter-bridge transducers. The network includes two 4.5 k
(±0.05%) resistors, connected in series, with a temperature coefficient of 5
ppm/°C. You can enable the network on an individual channel basis using two
additional jumpers per channel–W1, W2, W17, W18, W27, W28, W39, and W40. When
you enable the completion network, you automatically remove access to the
negative input of the amplifier to preserve overvoltage protection. See Table
3-15 for half-bridge completion network jumper information, and refer to
Figures 4-5, 4-6, and 4-7 for diagrams of standard bridge configurations.
© National Instruments Corporation
3-21
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-15. SCXI-1121 Half-Bridge Completion Network Jumper Configuration
Jumpers
Channel
Enable Network
Disable Network (Factory-Default Position)
0
W1
W2
W2
W2
AB
W1 123
AB
W1 123
1
W17 W18
W18
W18
AB
123
W17
AB
123
W17
2
W27 W28
W28
W28
AB
123
W27
AB
123
W27
3
W39 W40
W40
W40
AB
123
W39
AB
123
W39
Getting Started with SCXI
3-22
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
6. Configure the amplifier output referencing.
The signal output referencing is relevant only if you connect the SCXI-1121
module directly to a DAQ device.
Use jumper W33 to configure the various output referencing modes of the
SCXI-1121 amplifier. The correct output referencing mode depends on the input
mode of the DAQ device you use.
Table 3-16 shows the three possible signal referencing configurations.
Table 3-16. SCXI-1121 Amplifier Output Referencing Jumper Configuration
Jumper
W33
Description
A-R0R1–Parking position. Use this setting for MIO or AI devices in DIFF input
mode (factory-default position).
Configuration
A B R2 R1 R0
AB-R2–Select this position for AI and MIO devices in NRSE mode.
AB-R0–Connects the amplifier reference to data acquisition analog ground, pins
1 and 2 of the rear signal connector. Use this position for DAQ devices with
SE inputs, such as the Lab and 1200 devices, or MIO and AI devices in RSE
mode.
A B R2 R1 R0
A B R2 R1 R0
7. If you are using any digital SCXI modules in serial mode and will connect
the SCXI-1121 module to the plug-in DAQ device, configure the SCXI-1121 module
to connect the MISO signal to the SERDATOUT pin (jumper W38).
SCXI-1122 Module
This module has 16 isolated input channels and two isolated excitation
channels. The SCXI-1122 is a module for signal conditioning of strain gauges,
RTDs, thermistors, thermocouples, volt and millivolt sources, 4 to 20 mA
current sources, and 0 to 20 mA process-current sources. The SCXI-1122 can
operate in two modes. Use the two-wire scan mode with all 16 input channels
used for input or the four-wire scan mode with the eight upper channels
configured as sense leads for connecting input channels and
© National Instruments Corporation
3-23
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
the lower eight channels configured as current output channels. Other major
features of this module include software-programmable gain and filtering, and
the ability to account for voltage drop of long lead wires in strain-gauge
measurement. The SCXI-1122 input signals are multiplexed to a single output,
which drives a single DAQ device channel.
The SCXI-1122 has two jumpers. Jumper W2 connects a pullup resistor to the
SERDATOUT signal on the rear signal connector as shown in Table 3-17. Jumper
W1 configures the guard and the analog output ground and enables the NRSE
mode. Filter bandwidth and scanning mode, both two-wire or four-wire, is
software configured.
Use the following procedure to configure your SCXI-1122 module.
Note If the module is not connected to a DAQ device, the position of W2 is irrelevant.
Jumper
W2
1. When using a single chassis, set jumper W2 in position 1 on the SCXI-1122 that is connected to the DAQ device. When using multiple chassis, set jumper W2 to position 1 on only one of the SCXI-1122s that are cabled to the DAQ device. It does not matter which of the SCXI-1122s cabled to the DAQ device has the pullup connected. Set jumper W2 in position 3 on all of the other SCXI-1122 modules.
Table 3-17. SCXI-1122 Digital Signal Jumper Configurations
Description
Configuration
W2
1
Position 1 (pullup)–Use this setting for a single-chassis system. Connects a
2.2 k pullup resistor to the SERDATOUT line (factory-default position).
Position 3 (unmarked position, no pullup)–Use this setting for additional
cabled SCXI-1122 modules in a multichassis system. No pullup resistor is
connected to the SERDATOUT line.
W2
1
Getting Started with SCXI
3-24
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
2. The SCXI-1122 has one analog configuration jumper for grounding, shielding, and reference mode selection. Set jumper W1 as shown in Table 3-18.
Table 3-18. SCXI-1122 Analog Jumper Configuration
Description
A-R0R1–Parking position. Use this setting for MIO or AI devices in DIFF input
mode (factory-default position).
Configuration
B A
R0 R1 R2
W1
W1
AB-R0– Use this position for DAQ devices
W1
with SE inputs, such as the Lab and 1200
devices, or MIO and AI devices in RSE
mode.
B A
R0 R1 R2
Current-Loop Receivers
The SCXI-1122 includes provision for measuring current rather than voltage.
National Instruments offers an SCXI process-current resistor pack, which
consists of four 249 , 0.1%, 5 ppm, 1/4 W resistors. For more information on
how to install current loop receiver resistors, consult your SCXI-1122 User
Manual.
Scanning Techniques
Note The SCXI-1122 input multiplexer is composed of relays. Relays have a
certain life expectancy, as listed in Appendix A, Specifications, of the
SCXI-1122 User Manual.
To avoid mechanical wear on the relays when the SCXI-1122 is used for
continuous multichannel scanning, you should acquire n samples on a given
channel before proceeding to the next channel, as shown in Figure 3-3b. For
example, rather than performing 100 scans and taking a single sample from each
channel during each scan, as shown in Figure 3-3a, acquire 100 samples from a
channel then switch to the next channel and acquire a new set of samples, as
shown in Figure 3-3b. The recommended method, software-driven scanning, saves
relay life.
© National Instruments Corporation
3-25
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Scanned 100 times
CH0 (one sample) CH1 (one sample)
CH2 (one sample)
CH3 (one sample)
No 100 Scans Done?
Yes
Average the samples for each channel
a. Not Recommended
b.
CH0 x 100 samples
CH1 x 100 samples
CH2 x 100 samples
CH3 x 100 samples
Average the samples for each channel b. Software-Driven Scanning
b. Saves Relay Life.
Figure 3-3. Scanning Techniques
SCXI-1124 Module
The SCXI-1124 has six isolated DACs with voltage or current outputs. The
SCXI-1124 outputs are designed to generate DC or slowly-varying current or
voltage signals, not waveforms.
The SCXI-1124 works with National Instruments AI and MIO devices, Lab and 1200
devices, and DIO devices. You can control several SCXI-1124s in a single
chassis with one DAQ device and in combination with other SCXI module types.
Note If nothing is connected to the SCXI-1124 rear signal connector, the
position of all of the jumpers is irrelevant.
Getting Started with SCXI
3-26
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
If you cable this SCXI-1124 directly to a plug-in DAQ device, then use the
following procedure to configure your SCXI-1124:
1. Set jumper W1 for a single chassis or multichassis system, as shown in
Table 3-19.
Table 3-19. SCXI-1124 Chassis Jumper Configuration
Description
Pullup–Use this setting for a single-chassis system. Connects a 2.2 k pullup
resistor to the SERDATOUT line (factory-default position).
Configuration
W1 P
W1
NP
No Pullup–Use this setting for additional cabled SCXI-1124 modules in a multichassis system. No pullup resistor is connected to the SERDATOUT line (parking position).
W1 P
NP
2. If you are cabling this module directly to a DAQ device, set jumpers W1, W2, and W3 depending on the type of DAQ device you are using, as shown in Table 3-20.
© National Instruments Corporation
3-27
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-20. SCXI-1124 Device-Type Jumper Configuration
Jumper
Description
Configuration
W2 W3 W4
D–Use this setting for all three jumpers if you want to connect the SCXI-1124
to a DIO device (factory-default position).
M–Use this setting for all three jumpers if you want to connect the SCXI-1124
to an MIO or Lab/1200 device.
W2 D M
W3
D
D
M
M
W4
W2 D M
W3
D
D
M
M
W4
You can connect two types of DAQ devices to the SCXI-1124–DIO devices and MIO
devices. If you want to connect the SCXI-1124 to an MIO device, place jumpers
W2, W3, W4 in the M position. If you are using an SCXI-1124 as well as modules
with analog inputs, you should cable the MIO device to one of the analog input
modules. The SCXI-1124 cannot route analog signals from the SCXIbus to the MIO
device.
SCXI-1125 Module
The SCXI-1125 module is exactly like the SCXI-1120/D with the exception that
it is software configurable and does not have any jumpers. Use Measurement &
Automation Explorer to configure the gain and filter settings for each
channel. In addition, the SCXI-1125 only has support for the MTEMP
configuration of the terminal block onboard temperature sensor.
Getting Started with SCXI
3-28
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
SCXI-1126 Module
The SCXI-1126 is a signal conditioning module with eight isolated frequency
input channels. Each channel of the SCXI-1126 includes a frequency to voltage
conversion circuit with programmable input range, threshold, hysteresis, and
filtering. Each channel is individually isolated with a working common-mode
voltage of 250 Vrms between channels or channel to earth ground. The SCXI-1126
is software configurable and contains no jumpers.
SCXI-1140 Module
This module is an 8-channel simultaneously sampling differential amplifier.
The SCXI-1140 has an instrumentation amplifier that can track and hold on all
eight channels, an analog input range of ±10 V, and individual DIP-switch-
selectable channel gains of 1, 10, 100, 200, and 500.
DAQ Device Connections in Multiple-Module Systems
Note If you use an SCXI-1140 in a system that contains multiple SCXI modules
and one DAQ device, you must connect the SCXI-1140 to the DAQ device because
the SCXI-1140 requires a counter/timer signal from the DAQ device to control
the module track-and-hold circuitry. Other modules do not use this signal and
thus do not route the signal to the SCXIbus.
If your system contains more than one SCXI-1140, the system operates correctly
only when you have connected one of the SCXI-1140 modules to the DAQ device.
Additional SCXI-1140 modules in the chassis can access the counter/timer
signal using that module and the SCXIbus backplane. Other SCXI modules in the
same chassis operate normally.
User-Configurable Jumpers and DIP Switches
This module has 10 user-configurable jumpers and eight DIP switches. You can
use the jumpers and switches to configure input signal referencing,
instrumentation amplifier output referencing, grounding, shielding, and gain.
Leave all other jumpers in their factory-default positions.
The SCXI-1140 is factory-configured for use with an AI or MIO device in DIFF
mode measuring ground-referenced signal sources–each module channel is
configured for a gain of 1. If this configuration is suitable for your
application, do not adjust any of the jumpers or DIP switches.
© National Instruments Corporation
3-29
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
However, you must adjust the configuration if any of the following conditions
is true:
· Your input signals are floating or nonreferenced.
· You are cabling this SCXI-1140 directly to a DAQ device and you are using a
Lab/1200 device, or you are using an MIO or AI device in SE mode.
· You require a gain other than 1 on any channel.
· You are using digital modules and are cabling the SCXI-1140 to a DAQ device.
Read the rest of this section on the SCXI-1140 module if any of these conditions applies to you.
Use the following procedure to configure your SCXI-1140 module:
1. Configure the signal referencing modes. Table 3-21 lists the SCXI-1140
signal referencing configurations.
Table 3-21. SCXI-1140 Signal Referencing Jumper Configuration
Jumpers
Description
W1-W2 W3-W6 W7-W8
AB–Connects the instrumentation amplifier negative input to the front panel connector only (factory-default position). Use this setting to measure ground- referenced source signals.
BC–Connects instrumentation amplifier negative input to the front panel connector and to module analog ground through a 100 k resistor (for floating sources). Use this setting to measure floating signals with this channel.
Configuration
ABC
ABC
2. Configure the instrumentation amplifier output referencing mode.
The output referencing mode is relevant only if this SCXI-1140 is connected
directly to a plug-in DAQ device. Table 3-22 lists the SCXI-1140
instrumentation amplifier signal referencing configurations.
Getting Started with SCXI
3-30
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-22. SCXI-1140 Instrumentation Amplifier Jumper Configuration
Jumper
Description
Configuration
AB–Use this setting for MIO or AI devices in
W9
DIFF mode (factory-default position).
A
B C
BC–Use this position for DAQ devices with SE
inputs such as the Lab and 1200 devices, or MIO
A
and AI devices in RSE mode.
B
C
Jumper W12 selects various grounding and shielding options. The factory-
default position, CD, leaves the grounds disconnected and is suitable for most
applications. For more information, refer to the SCXI-1140 User Manual.
3. Configure the gain selection.
Use the eight 4-bit DIP switches, U12 through U19, to select the gains of
channels 0 through 7, respectively, as shown in Table 3-23. The factory-
default position sets the gain to 1; that is, all four bits of each DIP switch
are set to the off position (number side up). Select other gains by setting
one or more of the four bits of a particular DIP switch to the on (closed)
position as indicated in Table 3-24.
Table 3-23. SCXI-1140 Gain Switches for Each Channel
Channel
Use DIP Switch
0
U12
1
U13
2
U14
3
U15
4
U16
5
U17
© National Instruments Corporation
3-31
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-23. SCXI-1140 Gain Switches for Each Channel (Continued)
Channel
Use DIP Switch
6
U18
7
U19
Table 3-24. SCXI-1140 Switch Settings for Gain Selection
Switch
Desired Gain
1
ABCD
10
1 23 4
100
200
500
OFF
Closed Switches
None D C B A
Switch is shown in factory-default (gain = 1) position.
4. If you are using any digital SCXI modules in serial mode and will connect
the SCXI-1140 module to the DAQ device, configure the SCXI-1140 module to
connect the MISO signal to the SERDATOUT pin (jumper W11). See the SCXI-1140
User Manual for details.
SCXI-1141, SCXI-1142, and SCXI-1143 Modules
The SCXI-1141, SCXI-1142, and SCXI-1143 have eight lowpass filters with
differential input amplifiers. The filter types of the modules are elliptic,
Bessel, and Butterworth, respectively. You can use the SCXI-1141, SCXI-1142,
and SCXI-1143 for antialiasing applications as well as for general-purpose
amplification and filtering of signals.
The SCXI-1141, SCXI-1142, and SCXI-1143 have two user-configurable jumpers.
Jumper W1 is set at the factory for a single-chassis system, and jumper W2 is
set at the factory for a differential input DAQ device.
Note You need to change the SCXI-1141/1142/1143 jumper configuration only if
or when you are using a multichassis system, when using a DAQ device in SE
mode, or to change the default grounding and shielding configuration.
Getting Started with SCXI
3-32
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper W1, when set to position P, connects a 2.2 k pullup resistor to the SERDATOUT line. An open-collector driver drives the SERDATOUT line. An open- collector driver either actively drives low or goes to a high-impedance state, relying on a pullup resistor to make the signal line go high. If too many pullup resistors are attached to the SERDATOUT line, the drivers cannot drive the line low. To prevent this, set jumper W1 to position P on only one of the SCXI-1141/1142/1143 modules that are connected to the DAQ device in a multiple chassis system. It does not matter which of the SCXI-1141/1142/1143 modules that are cabled to the DAQ device has the pullup connected.
Tables 3-25 and 3-26 list the description and configuration of the configurable jumpers.
Table 3-25. SCXI-1141/1142/1143 Chassis Jumper Configuration
Jumper
W1
Description
Pullup–Use this position for a cabled module in a single-chassis system or the
first chassis in a multichassis system (factory-default position).
Configuration
W1 NP
P
No Pullup–Use this position for cabled modules in additional chassis in a multichassis system.
W1 NP
P
© National Instruments Corporation
3-33
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Table 3-26. SCXI-1141/1142/1143 Grounding and Shielding Jumper Configuration
Jumper
Description
Configuration
Parking position. Use this setting with DAQ W2 devices in DIFF mode (factory-
default
position).
W2
B A R2 R1 R0
Use this position for DAQ devices with SE inputs, such as the Lab and 1200 devices, DAQCard-700, or MIO and AI devices in RSE mode.
W2
B A R2 R1 R0
Connects the SCXIbus guard to the analog reference.
W2
B A R2 R1 R0
Select this position for MIO and AI devices in NRSE mode.
W2
B A R2 R1 R0
SCXI-1520 Module
The SCXI-1520 is an 8-channel module used to interface with strain gauges and
strain-gauge-based sensors.
You can set the bridge excitation voltage, hardware null setting, filter
cutoff frequency, or shunt calibration settings on a per-channel basis through
software control. You can set the gain of each input channel through LabVIEW
or NI-DAQ. An onboard EEPROM contains gain and offset calibration constants
for each channel. NI-DAQ and LabVIEW automatically use these constants to
correct for amplifier gain and offset errors when scaling data.
Getting Started with SCXI
3-34
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
You can use direct function calls to NI-DAQ to configure and calibrate the SCXI-1520 as well as scale data acquired through the SCXI-1520. Table 3-27 lists these functions and describes their use.
Table 3-27. NI-DAQ Functions Applicable to the SCXI-1520
Function SCXI_Set_Gain SCXI_Set_Excitation SCXI_Configure_Filter SCXI_Strain_Null SCXI_Calibrate_Setup SCXI_Configure_Connection SCXI_Calibrate
Action
Sets the analog input gain within the allowable range of X1 to X1000.
Sets bridge excitation voltage. Range is from 0 to 10 V in 0.625 V steps.
Sets filter cutoff frequency to one of five settings: 10 Hz, 100 Hz, 1 kHz, 10
kHz, or bypass.
Automatically zero-nulls the sensor-induced offset voltage of the specified
channel.
Enables or disables shunt calibration on a per channel basis.
Sets the bridge configuration to quarter, half, or full bridge.
Initiates a module self-calibration. Updates the gain and offset constants
stored in the EEPROM of the module. These constants are used by the scaling
function SCXI_Scale.
The SCXI-1314 terminal block provides a convenient method of connecting field
wiring from strain gauges and strain-gauge based sensors to the front signal
connector of the SCXI-1520. The terminal block includes a socketed quarter-
bridge completion resistor and two socketed shunt calibration resistors for
each channel.
SCXI-1530/1531 Module
The SCXI-1530 and SCXI-1531 are four- and eight-channel modules, respectively,
used for interfacing with Integrated Circuit Piezoelectric (ICP®) and other
current-excited accelerometers. They have BNC connectors for direct connection
to accelerometer cables. You can turn the 4 mA current excitation on or off.
You can set the bandwidth of the four-pole Bessel filter to 2.5, 5, 10, or 20
kHz. You can set the gain to 1, 10, or 100X. All channels are simultaneously
sampled and held to preserve timing and phase information. Excitation,
bandwidth, and gain are set
© National Instruments Corporation
3-35
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
on a per-channel basis without using jumpers. You configure the SCXI-1530/1531
using Measurement & Automation Explorer or through function calls to NI-DAQ.
SCXI-1540 Module
The SCXI-1540 module is an eight-channel module for interfacing to industry-
standard LVDTs, RVDTs, and resolvers. It can accommodate a 4-wire or 5-wire
connection to an LVDT, and it allows you to synchronize any number of channels
to a common frequency. You can set the excitation level at 1 Vrms or 3 Vrms.
You can set the excitation frequency at 2.5, 3.3, 5, or 10 kHz. You can set
the gain over the range 0.8X to 25X to handle a wide range of LVDT
sensitivities. Gain, level, frequency, and wire mode are set on a per-channel
basis without the use of jumpers. You configure the SCXI-1540 using
Measurement & Automation Explorer or through function calls to NI-DAQ.
Digital SCXI Modules
In addition to analog SCXI modules, National Instruments produces a variety of
digital modules–the SCXI-1160, SCXI-1161, SCXI-1162, SCXI-1162HV, SCXI-1163,
and SCXI-1163R. You can connect these modules to either an analog or a digital
DAQ device. The digital modules can work with analog modules in the same
chassis.
When you are working with both analog and digital modules, you must use an
analog DAQ device because it can accept both analog and digital signals. You
must also connect an analog module to the analog DAQ device because an analog
module can transfer both analog and digital signals from the SCXIbus to the
DAQ device.
The following sections describe each digital SCXI module and contain step-by-
step instructions for configuring the jumpers on each module. Find the section
about the module you want to configure. Check the position of each jumper and
read the jumper position description. Change the appropriate jumpers to
configure the module for your application.
Getting Started with SCXI
3-36
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Communication Modes
This section describes the serial and parallel communication modes.
Serial Mode
You can use any digital SCXI module in serial mode, in which you use a maximum
of five digital lines to serially transfer data to and from the DAQ device.
Either an analog or digital DAQ device can operate in serial mode, but an
analog device works only in serial mode. Also, if you want to control a number
of modules using any one DAQ device, you must configure all your digital
modules for serial mode. The software configuration utilities use the
multiplexed operating mode selection for digital modules to specify serial
mode.
Parallel Mode
The SCXI-1162, SCXI-1162HV, SCXI-1163, and SCXI-1163R modules can work in
parallel mode, in which you can use either 24 or 32 parallel lines to transfer
data to or from a digital DAQ device. You must connect a module configured for
parallel mode directly to a digital DAQ device. In parallel mode, you can use
only one module per digital DAQ device; you must connect any other modules in
the chassis, in either parallel or serial mode, to a different DAQ device. The
only exception to this rule is the DIO-96/PnP, which can work with two digital
modules in parallel mode or one module in parallel mode and any number in
serial mode. If you connect a digital module to the second half (pins 51
through 100) of the NB5 cable connected to the DIO-96/PnP, you must select
Parallel or Secondary operating mode in the Configuration window. No analog
DAQ devices except the MIO-16D devices work with digital modules in parallel
mode.
Note On the MIO-16D, the cable breaks out the 100-pin male I/O connector to
two 50-pin female connectors. One 50-pin connector is equivalent to the MIO-16
I/O connector; the other 50-pin connector is equivalent to the DIO-24 I/O
connector.
If you want to use your digital modules in serial mode, connect one module to
the MIO-16-equivalent half of the NB5 cable and use the MIO-16D as an analog
device.
If you want to use parallel mode with the SCXI-1162 or SCXI-1163, connect the
SCXI module to the PC-DIO-24-equivalent half of the cable and use the MIO-16D
as a digital device. If you use the DIO-24-equivalent half of the cable, you
must select the parallel (secondary) operating mode in Measurement &
Automation Explorer.
© National Instruments Corporation
3-37
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
SCXI-1160 Module
This module controls 16 isolated SPDT relays. The module has three user-
configurable jumpers for configuring the type of DAQ device connected to your
module. Leave all other jumpers in the factory-default positions.
The SCXI-1160 module is shipped configured for use with a DIO device.
Note If you are not connecting the SCXI-1160 module directly to a DAQ device, you do not need to change any jumper settings.
If you are connecting the SCXI-1160 directly to a DAQ device, set jumpers W2, W3, and W4 according to whether the module is connected to an analog device or a DIO device as shown in Table 3-28.
Table 3-28. SCXI-1160 Device-Type Configuration
Jumper
Description
Configuration
MIO
DIO–Use this setting to connect the SCXI-1160 to a DIO device (factory-default position).
DIO
MIO
DIO
W2
W3
W4
MIO
DIO
MIO
W2 W3 W4
MIO–Use this setting if you connect your SCXI-1160 to an AI or MIO device, or a Lab/1200 device.
DIO
MIO
DIO
W2
W3
W4
MIO
DIO
SCXI-1161 Module
The SCXI-1161 module controls eight SPDT relays. The module has three user-
configurable jumpers for configuring the type of DAQ device connected to your
module. Leave all other jumpers in the factory-default positions.
The SCXI-1161 module is shipped configured for use with a DIO device.
Note If you are not connecting the SCXI-1161 module directly to a DAQ device,
you do not need to change any jumper settings.
Getting Started with SCXI
3-38
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W3 W4
W5
If you are connecting the SCXI-1161 directly to a DAQ device, set jumpers W3, W4, and W5 according to whether the module is connected to an analog device or a DIO device as shown in Table 3-29.
Table 3-29. SCXI-1161 Device-Type Configuration
MIO
Description
DIO–Use this setting to connect the SCXI-1161 to a DIO device (factory-default
position).
DIO
Configuration
DIO
MIO
DIO
MIO
W3
W4
W5
MIO
MIO–Use this setting if you
DIO
MIO
DIO
MIO
DIO
connect your SCXI-1160 to an
W3
W4
W5
AI or MIO device or Lab /1200
device.
SCXI-1162 Module
This module has 32 optically-isolated digital input channels. Major features
include 450 Vrms isolation, 1,900 Vrms breakdown, TTL and CMOS compatibility,
and high-speed parallel data transfer to digital DAQ devices. The module has
four user-configurable jumpers for configuring the operating mode and the type
of DAQ device connected to your module. Leave all other jumpers in the
factory-default positions.
The SCXI-1162 module is shipped configured for use with a DIO device in serial
mode.
Note You need to adjust jumpers on this module only if you connect the module
directly to an analog DAQ device or use the module in parallel mode. If either
of these conditions applies to you, read the rest of this section.
© National Instruments Corporation
3-39
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Use the following procedure to configure your SCXI-1162 module:
1. Select parallel or serial mode.
Use jumper W2 to select whether the module transfers data in parallel to a
digital DAQ device (or the DIO-24 connector of an MIO-16D device) or serially
to either a digital or an analog DAQ device. For more information see Table
3-30. If you use parallel mode, you must also set the number of parallel lines
using jumper W1.
Table 3-30. SCXI-1162 Communication Mode Configuration
Jumper
W2 W6
Description
SER–Use this setting to configure the rear connector for serial communication
(factory-default position).
Configuration
W2 SER
PAR
PAR–Use this setting to configure the rear connector for parallel
W2
communication.
SER
W6
MIO
DIO
PAR PAR
2. Select 24 or 32 lines.
If jumper W2 is set to parallel mode, you can use jumper W1 to select whether
the module transfers 24 or 32 parallel lines of data to the device. If you are
using a DIO-24 or the digital connector (pins 51 through 100) of an MIO-16D,
you must select 24 lines. If you are using a DIO-32 or a DIO-96/PnP, select 32
lines. If you are using an analog DAQ device, set jumper W2 to the serial
position, which makes the position of W1 irrelevant. For more information see
Table 3-31.
Getting Started with SCXI
3-40
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W1
Table 3-31. SCXI-1162 Parallel Lines Jumper Configuration
Description
24–Use this setting to drive only 24 rear connector data lines when W2 is set
to PAR (factory-default position).
Configuration
24 32
32–Use this setting to drive all 32 rear connector
data lines when W2 is set to PAR.
24
32
3. Select your device type–either analog or digital.
If this module is connected directly to a DAQ device, jumpers W3, W4, and W6
specify to which type of device the module is connected. Set jumpers W3, W4,
and W6 to the DIO position if the module is connected to a DIO-24, DIO-32, or
DIO-96/PnP device. Set the jumpers to the MIO position if the module is
connected to an AI or MIO device or Lab/1200 device. If you are using the
SCXI-1162 in parallel mode, set jumper W6 to the PAR position. If the module
is not connected directly to a DAQ device, the positions of W3, W4, and W6 are
irrelevant.
© National Instruments Corporation
3-41
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
DIO
Jumper
W3 W4
W6
Table 3-32 shows the SCXI-1162 device-type jumper configurations.
Table 3-32. SCXI-1162 Device-Type Jumper Configuration
Description
Configuration
MIO–Use this setting to
configure the rear connector
W6
DIO
DIO
MIO
for an AI or MIO device or
Lab/1200 device in serial
W3
W4
mode (factory-default
MIO
MIO
position).
PAR
DIO–Use this setting to
configure rear connector for
W6
DIO
DIO
a DIO device in serial mode.
W3
W4
MIO
MIO
MIO
PAR
DIO
SCXI-1162HV Module
The SCXI-1162HV module consists of 32 optically-isolated, wide range, AC or DC
digital inputs. The SCXI-1162HV is a module used for sensing the presence of
AC or DC voltages.
To configure the SCXI-1162HV module, use the six user-configurable jumpers
(W2W7). (Jumper W1 is a reserved jumper and should remain unconnected.)
Note If you are controlling the SCXI-1162HV through the SCXIbus and are not
connecting a DAQ device directly to the rear connector of the SCXI-1162HV, the
positions of these jumpers are irrelevant.
Jumper W3, when set to position P, connects a 2.2 k pullup resistor to the
SERDATOUT line. An open-collector driver either actively drives low or goes to
a high-impedance state, relying on a pullup resistor to make the signal line
go high. If too many pullup resistors are attached to the SERDATOUT line, the
drivers cannot drive the line low. To prevent this, set jumper W3 to position
P on only one of the SCXI-1162HV modules cabled to the DAQ device in a
multichassis system. It does not matter which of the SCXI-1162HV modules
cabled to the DAQ device has the pullup connected.
Getting Started with SCXI
3-42
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W5
Table 3-33 lists the description and configuration of the user-configurable jumpers.
Table 3-33. SCXI-1162HV Jumper Configuration
Description
24–Use this setting to drive only 24 lines when W2 is set to PAR and you are
using a DIO-24 or an MIO-16D device (factory-default position).
32
Configuration
24
24
32
32–Use this setting to drive all 32 rear connector data lines when W2 is set to PAR and you are using a DIO-32 or DIO-96/PnP device.
S–Use this setting to configure the rear connector for
W6
serial communication (factory-default position).
S
P
S
P
P–Use this setting to configure the rear connector for parallel communication.
Use with DIO devices only.
DIO–Use this setting to configure the rear connector for W7 a DIO device
(factory-default position).
MIO
DIO
MIO
DIO
MIO–Use this setting to configure the rear connector for an MIO device.
Includes AI or MIO device or Lab/1200 device.
DIO–Use this setting to configure the rear connector for W4 a DIO device
(factory-default position).
DIO
MIO
DIO
MIO
MIO–Use this setting to configure the rear connector for an MIO device. Includes AI or MIO device or Lab/1200 device.
© National Instruments Corporation
3-43
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W3
W2
Table 3-33. SCXI-1162HV Jumper Configuration (Continued)
Description
Pullup–Use this setting for a single-chassis system or for the first chassis
in a multichassis system. Connects a pullup resistor to the SERDATOUT line
(factory-default position).
No Pullup–Use this setting for additional chassis in a multichassis system. No
pullup resistor is connected to the SERDATOUT line.
NP
NP
Configuration
P
P
DIO–Use this setting to configure SERDATOUT for
DIO devices (factory-default position).
PAR
MIO
DIO
MIO
MIO–Use this setting to configure SERDATOUT for
MIO device. Includes AI or MIO device or Lab/1200
PAR
device.
DIO
PAR–Use this setting to disconnect SERDATOUT line.
PAR
Use for parallel mode.
MIO
DIO
SCXI-1163 Module
This module is a 32-channel optically isolated digital output module with TTL
and CMOS compatibility, 450 Vrms isolation and 1,900 Vrms breakdown, and high-
speed parallel data transmission from digital DAQ devices. The module has five
user-configurable jumpers for configuring the operating mode and the type of
DAQ device connected to your module. Leave all other jumpers in the factory-
default positions.
The SCXI-1163 module is shipped configured for use with a DIO device in serial
mode.
Getting Started with SCXI
3-44
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
Note You need to adjust jumpers on this module only if you plan to connect the module directly to an analog DAQ device or if you want to use the module in parallel mode. If either of these conditions applies to you, read the rest of this section.
Jumper
Use the following procedure to configure your SCXI-1163 module:
1. Select your device type–either analog or digital.
If the module is connected to a DIO device, set jumpers W2 and W3 to the DIO
position, as shown in Table 3-34. If the module is connected to an analog
device, set W2 and W3 to the MIO position. If the module is not cabled
directly to a DAQ device, the positions of W2 and W3 are irrelevant.
Table 3-34 shows the SCXI-1163 device-type jumper configurations.
Table 3-34. SCXI-1163 Device-Type Jumper Configuration
Description
DIO–Use this setting to configure the rear connector for a DIO device in
serial mode (factory-default position).
Configuration
W2 DIO
W3 DIO
W2 W3
MIO–Use this setting to configure the rear connector for an AI or MIO or Lab/1200 device.
MIO
W2 DIO
MIO
W3 DIO
MIO
MIO
2. Select serial or parallel mode.
If you are using the module in parallel mode, set jumper W6 to the P position
and jumper W5 to the PAR position. For serial mode, set W6 to the S position
and jumper W5 to the MIO or DIO position, depending on the type of DAQ device
to which the module is connected. If you are using the module in parallel
mode, it must be directly connected to a digital DAQ device or the DIO-24
connector of an MIO-16D.
© National Instruments Corporation
3-45
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
Table 3-35 shows the SCXI-1163 mode jumper configurations.
Table 3-35. SCXI-1163 Mode Jumper Configuration
Description
S–Use this setting to configure the module in serial mode (factory-default
position).
Configuration
W6 S
W6 W5
P–Use this setting to configure the rear connector for a DIO device in parallel mode.
P
W6 S
P
DIO–Use this setting to configure the module in
serial mode (factory-default position).
W5
MIO
DIO
PAR
PAR–Use this setting to configure the rear
connector for a DIO device in parallel mode.
W5
MIO
DIO
PAR
MIO–Use this setting to configure the rear
connector for an MIO device.
W5
MIO
DIO
PAR
Getting Started with SCXI
3-46
www.ni.com
Chapter 3 Configuring Your SCXI Hardware and Software
SCXI-1163R Module
The SCXI-1163R is an SCXI module consisting of 32 optically-isolated solid-
state relays. The SCXI-1163R module switches loads up to 240 Vrms or VDC and
200 mA.
The SCXI-1163R operates in serial mode with MIO and AI devices, Lab/1200
devices, and DIO devices. You can control several SCXI-1163R modules in serial
mode in a single chassis with one DAQ device. Alternatively, you can control
24 of the solid-state relays in parallel mode with a DIO-24, or all 32 solid-
state relays in parallel mode with a DIO-32 or DIO-96/PnP.
Note You should change the jumper configuration from the factory-default
settings only if you plan to connect the SCXI-1163R directly to an MIO and AI
device, or if you plan to use the SCXI-1163R in parallel mode.
© National Instruments Corporation
3-47
Getting Started with SCXI
Chapter 3 Configuring Your SCXI Hardware and Software
Jumper
W2
Tables 3-36 and 3-37 show the SCXI-1163R jumper configurations.
Table 3-36. SCXI-1163R Device-T