UVT-LED Sensor
Installation and Operation Manual
Welcome to Sensorex UVT-LED!
Thank you for choosing our UVT-LED UV-transmittance sensor.
For over forty years, Sensorex has been the preeminent supplier of leading
edge technology and innovative products to measure and monitor the quality of
water. Today, we remain focused on the continual advancement of water quality
monitoring products and technology. The UVT-LED is the latest Sensorex
innovation and our first optical-based measurement system designed
specifically for simple, reliable UV- transmittance monitoring. The Sensorex
UVT-LED is the world’s first family of UV-T monitors that uses a UV-C LED
instead of a mercury-based lamp as a light-source. The use of a UV-C LED in
the UVT-LED ensures extremely stable readings in all conditions, over an
extended lifetime.
Getting to know your UVT-LED Sensors
UVT-LED-H
Used for benchtop and off-line field measurements. Unit comes standard with a
rechargeable internal battery and charging cord for cordless
operation.
UVT-LED-SW
Used for all open channel applications in which the unit is completely
submerged, or in-line applications utilizing a 1.5″ sanitary connection. Unit
is wiped using patented technology. Use in water with a higher concentration
of dissolved and suspended solids.
Warnings
| To avoid risk of damage to monitor, do not open the device. Refer servicing
to qualified personnel only.
---|---
| Do not place device in a location near heat sources, or in a place subject
to direct sunlight, excessive dust or mechanical shock.
| Do not operate sensor above 50°C because this can damage the UV Light source
| Monitor contains a UV LED emitting device.
Do not look at the probe of an operating monitor without wearing proper eye
and skin protection.
| Install unit so that power cord is protected from mechanical damage and can
be easily unplugged from device.
| Do not use the provided power/data cord if itis damaged. Use of a damaged
cord may result in damage to the device or fire.
| Do not pinch the power/data cord between heavy objects
| Device should only be used with provided 24V DC power supply or used with
asite supplied 24V DC source.
Introduction
UV-T is a measure of the amount of UV energy remaining after passing through a material (e.g., water sample or quartz) over a specific distance. UV Absorbance is a measure of the amount of UV light that is absorbed by a substance (e.g., water, microbial DNA, lamp envelope, quartz sleeve) at a specific wave- length (e.g., 254 nm). UV-Transmittance is related to UV Absorbance by the following equation (fora 1- cm path length): % UVT = 100 x 10
UV-Transmittance is typically;
- measured at 254 nm, corresponding to the peak output of mercury-based UV lamps
- measured over a path length of 1-cm
- represented as a percentage
UV-T has a strong effect on the dose delivery of a UV treatment system. As
UV-T decreases, the intensity throughout the reactor decreases, which reduces
the dose the treatment system delivers. UV treatment systems are sized to
deliver the required UV dose under specified UV-T conditions for the
application.
UV-T will typically vary over time due to changing concentrations of
compounds, seasonal effects and changes in biological activity of
microorganisms within the water source.
UV-T monitors are commonly used as an input to a dose control strategy of a UV
treatment system, helping to reduce over or under dosing. They are also used
to monitor water quality to help diagnose operational problems.
Specifications
Measurement Technique:| UV absorption method (single beam technique), reagent-
free
---|---
Measurement Wavelength
/Pathlength:| 254nm/1cm
Measurement Range:| 10-100% UV transmittance (UVT)
Accuracy:| 1.0%T@>50%T
Sample Flow Rate/
Maxium Pressure:| n/a, 150psig
Display:| LCD with LED backlight
Measuring Interval:| 60 seconds (UVT-LED-SW only)
User Maintenance:| Thr/month typical
Power Supply:| 20-28V DC, 5A, maximum 2000hm load, customer supplied (UVT-LED-
PW &-SW)
24V DC to charge battery (UVT-LED-H)
Output:| 0-100% UVT (UVT-LED-H-WD, UVT-LED-H-NA only), 4-20mA (UVT-LED-SW)
Temperature Range:| 32-122 °F (0-50 °C)
Weight:| UVT-LED-H – 0.78lbs (353g), UVT-LED-PW – 1.58lbs (717g), UVT-LED-SW
-1.78Ibs(807g)
Dimensions:| 7.60″H x 2.60″ cap x 1.25″ probe (19.1cm H x 6.4cm Dia) – UVT-
LED-H & -PW only,
9.4″H x 3.00″ cap x 1.25″ probe (23.9cm H x 7.6cm Dia)
IP Ingress Rating:| P66 housing, IP68 measurement head
Certifications:| CE
Warranty:| One Year Limited Warranty
Wetted Materials:| 55316 body and wiper holder, PPS top cap, PTFE Wiper Blade
Outline and Dimensions
UVT-LED-H
Overview (-SW)
The Sensorex UVT-LED-SW is designed for use in both open channel/submersion
UVT monitoring and in-line UVT monitoring of any type of water, including
wastewater and drinking water. This patented UV-T monitoring device includes
a UV-C LED light source, a UV sensor and wiper attached to a rotating
measuring head.
It is operated by submerging the sensor mounted to installation hardware into
a flowing water channel.
The water sample flows through the measurement head, in between the UV light
source and the UV sensor. The UV sensor measures the amount of UV light that
the UV sensor is exposed to, measuring the amount of UV light that passes
through the fluid sample. After the UV measurement of the water sample is
collected, a quartz cylinder having a known UVT value is moved in between the
UV light source and the UV sensor and another measurement is taken. By
comparing the UV measurement taken through the quartz cylinder and the UV
measurement taken through the fluid sample, a UVT value for the fluid is
determined. In addition a third measurement is taken with the UV light source
powered off providing a zero point calibration measurement for the UV sensor.
The sensor is calibrated by actuating with a magnet and placing in a sample of
DI water. Measurements are taken by actuating the sensor again with a magnet
and then measurements are automatic every 3 minutes and are output as 4-20mA
scaled to 0-100% UVT.
When the quartz cylinder is rotated into a position between the UV light
source and the UV sensor, a wiper, attached to the rotating mechanism, wipes
the windows in between the UV light source and the water sample and the UV
sensor and the water sample, thus removing any contaminants or buildup of
constituents from the water sample on the windows. See FIG. 1. Note that UVT-
LED-SW does not have a display but all other components match FIG 1.
First Time Setup and Preparation
Step 1 – Unpacking your UVT-LED-SW
Make sure you have all parts in the box. Review the Packing List included in
the product box for details.
Your UVT-LED-SW sensor is supplied with an alcohol cleaning swab and a 10mtr
cable (UVT-SW-001).
Step 2: Cable Assembly Preparation: A 10meter (33ft) cable assembly is
included with each UVT-LED-SW. Assemble the cable to the 1″ adapter of the
UVT-SW mounting assembly. After the assembly is completely mounted then
attach the sensor to the cable by aligning the connector parts and tightening
the knurled nut clockwise until it stops (FIG 1). 1 inch of the cable jacket
is stripped off for your convenience (see FIG 3).
If you want longer leads, remove as much of the gray cable covering to reveal
the length of wire you require. Next, strip the brown, white, and green wire
tips to desired length and tin with solder to prevent wirings from fraying.
Trim off braided shield and string filler even to cable jacket.
Step 3 – Powering your UVT-LED-SW
After the cable assembly has been connected to the UVT-LED-SW in step 2, wire
the stripped end of the cable BROWN, WHITE, GREEN wires to your 4-20mA input
device as shown in FIG 2 and 3.
Cleaning Procedure
Cleaning with Deionized water is recommended every 30days followed by
recalibration off-line.
When fouling of the quartz windows are suspected, The Sensorex UVT-LED-SW
should be cleaned according to the following procedure:
- Remove Sensorex UVT-LED-SW from the process.
- Fill up a small container with a 15-20% solution of phosphoric acid (e.g. Calcium Lime and Rust remover CLR) or food grade citric acid. Fill container high enough to completely submerge the measuring portion of the tip in the 15-20% solution of phosphoric acid or food grade citric acid (see FIG. 4).
- While the unit is still powered, allow unit to soak for 30 minutes.
- After 30 minute soak, rinse measuring tip with de-ionized water to rinse off excess cleaning solution.
- Fill container with isopropanol, high enough to fully submerge measuring tip.
- Submerge measuring tip in isopropanol for 10 minutes while unit is powered on.
- Using the UVT-LED cleaning swab (part #UVT0011), gently clean the quartz measuring windows in front of the sensor and the UV-C light source.
- Place unit in de-ionized water high enough to fully submerge measuring tip.
- Submerge measuring tip in de-ionized water for 10 minutes.
- Re-calibrate monitor according to recalibration procedure.
Calibration
Calibration should be done every 30 days after cleaning with deionized water.
It is also recommended to verify sensor accuracy weekly using a hand-held UVT-
LED-H sensor.
Step 1 – Place your UVT-LED-SW into de-ionized (DI) water.
Make sure it is inserted as shown in FIG 19 so that it is above the opening
for the wiper and LED.
Shake to remove air bubbles.
Step 2 – Hold Magnet over white dot sensor actuation location for at least 3
seconds as shown in FIG 19. Wait for wiper to click. Remove magnet.
At completion of calibration, UVT-LED-SW will beep 7x.
Taking Readings
Your UVT-LED-SW sensor will take readings every 3 minutes after actuation of
the wiper once the sensor is calibrated. You can take a reference reading at
any time. To do this:
Step 1 – Place your UVT-LED-SW into a sample of your water.
Make sure it is inserted as shown in FIG 18 so that it is above the opening
for the wiper and LED.
Shake to remove air bubbles.
Step 2 – Hold Magnet over white dot sensor actuation location as shown in FIG
5.
Wait for 1 second. Remove magnet. UVT-LED-SW will beep 1x after reading is
taken.
Installation
The UVT-LED-SW is designed for in-line and submersion mounting.
In abrasive solutions, install the sensor downstream of filtration to maximize
sensor lifetime.
In-line mounting hardware for the UVT-LED-PW is also compatible with the UVT-
LED-SW. Refer to pages 14-17 for in-line mounting options.
Submersion mounting hardware options include:
- UVT-SW-005: Complete installation kit that attaches to wall of channel. See below and following pages for instructions and FIGURE references.
- UVT-SW-004: Retrofit installation kit that attaches to existing installation brackets from other suppliers that require 114″ pipe. See FIG 32 for description of hardware.
A) Attach provided steel 90 degree elbow to pipe with sealing tape or paste
(see FIG 8
B) Atttach 1.5″ x 1″ reducing fitting to elbow with sealing tape or paste (see
FIG 9)
C) Attach UVT-SW cable assembly to reducer (1″ threads on both) and seal with
sealing tap or paste. See FIG 10
D Feed cable up through pipe and insert into top cap/cord grip assembly. Make
sure cord grip is loose when feeding cable through. See FIG11.
E Seal top cap onto pipe (make sure cord grip is loose). See FIG 12
F Tighten cord grip after applying some slack to cable. See FIG 13.
G Attach brackets x 2 to pipe. Make sure bolts are not too tight so the
brackets can be adjusted up or down the pipe to the correct height and
spacing. Spacing will depend on wall bracket height and distance apart.
H Attach wall mounting brackets to concrete wall in channel using customer
supplied concrete anchor bolts.
Note that assembly weighs 33lbs(15kg). Holes in wall mounting bracket are
0.65” (16mm) OD. See FIG 14 H)
Remove pins from brackets. See FIG. 15
I Slide pipe assembly onto the wall mounting brackets. See FIG. 16.
J Adjust pipe depth into water so that UVT-LED-SW measurement pathway is fully
submerged.
K Tighten bolts on pipe clamps to hold in position.
L Reinstall locking pins into wall mount brackets. See FIG. 17
M To remove sensor for maintenance, remove the 2 locking pins and slide the
pipe/assembly out of wall bracket channels.
UVT-SW-004: Retrofit installation kit that attaches to existing installation
brackets from other suppliers that require 1 14″ pipe. See FIG 18 for
description of hardware.
Refer to FIG 22 – 27 for assembly procedure. After assembling, attach to
existing clamps in place of 1.25″ pipe from other supplier.
In-line Installation
In abrasive solutions, install the sensor downstream of filtration to maximize
sensor lifetime.
The UVT-LED-SW can mount into a few different installation types:
A) Weld on $5316 sanitary flange adapter (UVT0009) to pipe.
Weld #4 to pipe –> Install #2 into groove on top of #4 –> Insert #1 (UVT-LED-
PW) into #4 –> Clamp in place with #3 by installing clamp around #4 and flange
of #1, then tighten wing nut of #3 until tight and no leaking occurs. (see
FIG. 19)
B) 2″ NPT threaded adapter (UVT0008 – SS316, UVT0021 – PVC).
Apply threaded sealing tape or paste to threads of #4 –> Thread #4 into #5
(customer supplied) –> Install #3 gasket into groove of #4. –> Insert #1(UVT-
LED-PW) into #4 —-> Clamp in place with #2 by installing clamp around #4 and
flange of #1 then tighten wing nut of #2 until tight and no leaking occurs
(see FIG. 20). FIG 21 shows part detail.
C) 3″ Sanitary tee kit (UVT0015) – refer to FIG. 12. 2″ Sanitary tee kit
(UVT0024) refer to FIG 22
Install #5 (x2) to grooves of #4 –> Install #6 (x2) to 4 with #5 (gasket) in-
between –> Use clamp #7 (x2) to hold #6 (x2) to #4. Install #3 into groove on
top of #4. Insert #1 into #4 . Clamp #2 around #1 and #4 and tighten using
wing nut of #2 until #1 is installed tightly (no leaks) into #4. Thread #8
(x2) into #6 (x2) using sealing tape or paste to avoid leaks. It is
recommended to install this into a bypass loop with valves on each side of the
assembly and a flow meter on the inlet side. See FIG 23.
D) 2″ NPT PVC tee kit (UVT0023) – refer to FIG. 12
Note: Use sealing tape or paste on all threaded connections to avoid leaks.
Thread #6 (x2) into #5 –>Thread #7 (x2) into #6 (x2) –> Thread #8 (x2) into #7
(x2). Thread #4 into #5.
Install #3 into groove on top of #4. Insert #1 into #4 . Clamp #2 around #1
and #4 and tighten using wing nut of #2 until #1 is installed tightly (no
leaks) into #4. It is recommended to install this into a by-pass loop with
valves on each side of the assembly and a flow meter on the inlet side. See
FIG 23.
See FIG 24 for all parts inlcuded in UVT0023. Suggested mounting orientation
is shown in FiG 25.
Troubleshooting (UVT-LED-SW)
3 beeps from sensor indicate a failure for the UVT-LED-SW. See chart below for faults and possible solutions.
Fault | Possible Solution |
---|---|
UVT-LED-SW will not switch on | – Check connections in power cabinet to ensure |
24V DC is being supplied.
– Place magnet over white dot sensor actuation location and wait for wiper to
actuate.
UVT reading is lower than expected| – During measurement an air bubble may
have been present in the measurement area. Ensure no air bubbles are present
and repeat the measurement.
– Follow cleaning procedure to remove materials that may foul the sensor.
– Unit needs to be recalibrated according to outlined calibration procedure.
UVT reading is higher than expected| – During calibration, an air bubble may
have been present in the measurement area, Recalibrate unit.
– Follow cleaning procedure on page 20.
UVT reading is not being transmitted to plant controls| – Check to make sure
the 4-20mA signal loop has an external power source. – Ensure signal wires are
connected per wiring diagram.
Analog scaling Is not correct (-PW)| – Scaling should be set up for 4mA = 0%
UVT and 20mA = 100% UVT.
UVT reading is ZERO| – During measurement an air bubble may have been present
in the measurement area. Ensure no air bubbles are present and repeat the
measurement.
– Follow cleaning procedure.
– Recalibrate unit.
– LED or sensor has stopped working. Return for repair.
– Water has high absorbance and is not allowing UV-C light to pass through.
Wiper Mechanism does not return to home position during cycle (-PW)| – Return
sensor to supplier for repair/replacement.
Calibration
Step 1- Place your UVT-LED-H into de-ionized (DI) water.
Make sure it is inserted as shown in FIG 26 so that it is above the opening
for UV light source. Shake the sensor to remove bubbles. Remove from water and
check for bubbles outside of the DI water. Re-insert into water if no bubbles
are seen.
Step 2 – Press and hold the “READ/CAL” button for at least 3 seconds as shown
in FIG 27 (Make sure LOW BATTERY indicator is not on before taking
measurement). See FIG 28
The display will flash blue 7x after calibrating. When the display reads
99.8%, calibration is complete. This should take approximately 10 seconds.
Taking Measurements
Step 1 – Place your UVT-LED-H into the water sample you want to measure the
for UVT.
Make sure it is inserted as shown in FIG 26 so that it is above the opening
for UV light source. Shake the sensor to remove bubbles. Remove from water and
check for bubbles outside of the DI water. Re-insert into water if no bubbles
are seen.
Step 2 – Reference measurment. Press the “READ/CAL” button and release when
the screen flashes 1x. This is done to ensure no bubbles are on the measuring
surface.
Step 3 – When the screen flashes 2x then the sensor is ready to take a new
reading (new readings can only be taken 30 seconds after the previous reading)
Note: if sensor is not measuring for 10min, it will automatically turn off
Cleaning
Using the Sensorex UVT-LED cleaning swab (part #UVT0011) gently clean the
quartz measuring windows in front of the sensor and the UV-C light source.
Place unit in de-ionized water high enough to fully submerge measuring tip.
Submerge measuring tip in de-ionized water tor for 10 minutes, according to
recalibration procedure.
Chargi ng Sensor – plug charging cable into sensor at connector at back of
sensor. Charge for at least 2 hours when battery is discharged. When not in
use, keep sensor attached to charger and plugged into wall.
Parts covered by this product instruction sheet include:
UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware
Ordering Information
Part Number | Description |
---|---|
UVT-LED-SW | Submersion Mount Process UV Transmittance Monitor (with wiper/auto |
cal.), includes 10m(33ft) detachable cable, cleaning swab
UVT-LED-H| Handheld, Off-Line Transmittance Monitor with power supply (outside
of North America), case, cleaning swab x 1
UVT0001| Camrying Case for UVT-LED-H (included with UVT-LED-HWD and UVT-
LED-H-NA kits)
UVT0005| Worldwide Power Supply
UVT0006| North American Power Supply
UVT-SW-001| Detachable Power/Communication Cable for UVT-LED-SW, 10m (33 ft)-
indluded with UvT-LED-sW
UVT-SW-002| Detachable Power/Communication Cable for UVT-LED-SW, 20m (66 ft)
UVT0011| Cleaning Swabs (pack of 50)
UVT0008| Installation Fitting, 2″ NPT 31655
UVT0009| Installation Fitting, Stainless Steel Pipe (requires welding)
UVT0015| Installation Kit, Sanitary Flow Cell (3″)
V10021| Installation Fitting for PVC Pipe (2″ NPT)
UVT0023| Installation Kit, PVC Flow Cell (2°) with connections for barbed
fitting with quick-disconnect
UVT0024| Installation Kit, Stainless Steel Flow cell (2′) with 1/4″ barbed
fittings
UVT0025| 1.5″ Sanitary Flange Cap, Stainless Steel
UVT-5W-004| UVT-LED-SW Installation Kit, retrofit to other supplier
installation kits that accept 1.25″ diameter pipe.
UVT-5W-005| UVT-LED-SW Installation Kit, Complete
Product Warranty
This UVT-LED product, as supplied and by Sensorex, or an authorized Sensorex
distributor, and delivered new, in the original carton is warranted by
Sensorex against manufacturing defects in materials and workmanship for a
limited warranty period of:
One (1) year on all components
The limited warranty commences on the date of shipment. To make a warranty
claim, the purchaser must con- tact Sensorex or distributor for problem
determination and receipt of an RMA number. The serial number must be
presented for warranty processing to begin.
Sensorex will repair, or replace, any part found to be defective, in our
opinion and at no charge as stipulated herein, with new or reconditioned parts
during the limited warranty period specified above. All replaced parts and
products become the property of Sensorex and must be returned to Sensorex.
Replacement parts and products assume the remaining original warranty, or
ninety (90) days, whichever is longer.
This limited warranty covers manufacturing defects in materials and
workmanship encountered in normal use of this product, and shall not apply to
the following, including but not limited to: wear and tear to quartz window
wiper blades, damage which occurs in shipment; delivery and installation;
applications and uses for which this product was not intended; altered product
or serial numbers; cosmetic damage or exterior finish; accidents, abuse,
neglect, fire, water, lightning or other acts of nature; use in applications
with water temperature and ambient air temperature over 100 degrees F, with
water pressure greater than 150 psi (PW only); use of products, equipment,
systems, utilities, services, parts, supplies, accessories, applications,
installations, repairs, external wiring not supplied or authorized by
Sensorex, or which damage this product or result in service problems;
incorrect supply voltage, fluctuations and surges in electrical supply;
customer adjustments and failure to follow operating instructions, cleaning
and maintenance instructions that are covered and prescribed in the
Operations and Maintenance manual.
There are no express warranties other than those listed and described above,
and no warranties whether expressed or implied, including, but not limited to,
any implied warranties of merchantability or fitness for a
particular purpose, shall apply after the express warranty periods stated
above, and no other express warranty or guaranty given by any person, firm or
corporation with respect to this product shall be binding on Sensorex.
Sensorex shall not be liable for loss of revenue or profits, failure to
realize savings or other benefits, or any other special, incidental or
consequential damages caused by the use, misuse or inability to use this
product, regardless of the legal theory on which the claim is based, and even
if Sensorex has been advised of the possibility of such damages. Nor shall
recovery of any kind against Sensorex be greater in amount than the purchase
price of the product sold by Sensorex and causing the alleged damage. Without
limiting the foregoing, purchaser assumes all risk and liability for loss,
damage or injury to purchaser and purchaser’s property and to others and
their property arising out of the use, misuse or inability to use this product
sold by Sensorex not caused directly by the negligence of Sensorex.
This limited warranty shall not extend to anyone other than the original
purchaser of this product either from Sensorex directly or through an
authorized Sensorex distributor, it is non-transferable and states the
exclusive remedy.
Parts covered by this product instruction sheet include: