JORVET J0558X Top Line Anesthesia System
The JorVet Top Line Anesthesia System pictured above shows optional accessories (J0558Y Bain Block and J0744SV Concentrator).
Thank you for using our JorVet J0558X and J0558Z Anesthesia System. We want you to get the most out of the machine and to have your experience in using it proceed with ease. Detailed instructions are provided along with the product’s description and features on how to correctly use it and take full advantage of all operations and functions.
TOP LINE ANESTHESIA SYSTEM MODELS
IMPORTANT! VAPORIZER NOT INCLUDED FOR BOTH MODELS J0558X
INTENDED USE
The JorVet Top Line Anesthesia System was designed for Veterinary Use Only. It is a lightweight, durable, mobile system with 3″ casters for maneuverability. The JorVet Top Line Anesthesia System includes an incorporated toggle to switch between rebreathing and nonrebreathing systems. This feature eliminates the need to remove the breathing circuits during use. There is a monitor shelf as well as a middle shelf to provide more workspace and keep the patient’s vitals and anesthesia controls in one area.
UNPACKING/INCLUDED ITEMS
JorVet Top Line Anesthesia System Assembly
- H-Frame Stand, Mounting Plate Assembly and Rebreathing Head, Vaporizer Inlet and Outlet Endcaps (Figure 1)
- Hardware Packet (Not pictured)
- 10 Second Card w/chain
- Black Absorber Canister Cover
- 4 Top Shelf Bolts (1/4″-20 x 1/2″ Flat)
- 4 Connecting Bolts (1/4″-20 x 1/2″ Button)
- 1 Allen Wrench (5/32″) u 2 Zip Ties
- 1 E-cylinder Wrench (Model J0558X Only)
- Assembly Quick Guide (Not pictured)
- Universal F-Circuit with 3L Bag (Figure 4)
- Non-rebreathing Circuit (Figure 5)
CIRCUIT DESCRIPTIONS, WARNINGS, AND NOTES
The F-Circuit is a rebreathing circuit for use with animals that weigh
approximately 15 lbs. (7kg) or more.
The Non-rebreathing Circuit is a circuit for use with animals that weigh
approximately 15 lbs. (7kg) or less.
WARNINGS
- Operator must read entire manual prior to operation and have proper training in the use of inhalant anesthetic equipment. Failure to do so may harm the animal.
- Do not connect unregulated compressed carrier gas to the Top Line Anesthesia System. Compressed oxygen, or another carrier gas, should be supplied at 50-55 psi. u Pop-off (APL) valve is closed for shipping. Completely open the pop-off valve prior to use. The pop-off valve should be kept open during normal operation and only closed when performing a leak check or using a ventilator.
- As a patient safety feature, the oxygen flush will not work with a non-rebreathing circuit. The oxygen flush is only piped to the rebreathing absorber.
- Always operate inhalant anesthetic equipment with waste gas scavenging systems. Failure to control waste anesthetic gas may cause harmful exposure to personnel.
- When using the oxygen concentrator, turn the anesthesia machine flowmeter on to allow flow. The concentrator alarm will sound if this step is not followed.
NOTE
- When using the oxygen concentrator with the J0558X model, the oxygen flush will not operate without the E-cylinder turned on. The oxygen flush is piped to the e-cylinder only.
- When using the oxygen concentrator, turn the concentrator flow to match the flow rate on the anesthesia machine.
- Do not overtighten the flowmeter knob. When the float is resting at the bottom of the flowmeter, the flowmeter is off. Overtightening will damage the needle valve and result in leakage.
- Anesthetic Vaporizer and carrier gas (oxygen) accessories are not included. Please follow the installation guides for these products.
DESCRIPTION OF CONTROLS/FEATURES
- Oxygen Flowmeter
- Flush Valve
- Accessory Mounting Area
- Inhalation Port
- Exhalation Port
- Mounting Area for E-cylinder Manifold
- Vaporizer Endcaps – Inlet and Outlet
- Accessory Mounting Area
- Switch – O2 Tank/02 Concentrator (J0558X Model Only)
- Common Gas Outlet
- Switch – Rebreathing/Non-rebreathing
- Pop-off (APL) Valve
- Rebreathing Bag Port
- Absorber Canister
- Absorber Canister Knob
- Manometer
PRODUCT OPERATION
REFER TO PRODUCT INSTALLATION GUIDE FOR ILLUSTRATION FOLLOW ALL STEPS ACCORDINGLY
- Remove all cartons from the main package.
- Remove the machine H-Frame base unit.
- Unpack and inspect remaining components for damage. Be sure to retain the mounting hardware contained inside package.
- Attach middle shelf to center post using 3 bolts.
- Attach center post with middle shelf attached to H-frame base with 4 bolts.
- Attach mounting plate assembly to center pole with 4 bolts.
- Note: Two people recommended for this step.
- Attach rebreathing head to mounting plate assembly using supplied bolts and allen wrench.
- Connect black supply line to the hose barb on the left side of the rebreathing head.
- Warning: Open pop-off valve prior to use.
- Install anesthetic vaporizer onto frame with vaporizer spacer block positioned on the front face of the mounting plate and tighten.
- Note : Two peopleare recommended for this step. Mounting hardware supplied witha vaporizer.
- Connect the inlet and outlet endcaps using a push and twist action to provide a leak free attachment.
- Note : This step can be performed prior to installing the vaporizer onto the frame if needed.
- Connect oxygen (carrier gas) supply source to the DISS Male fitting on the back of the flowmeter.
- Note: For Model, J0558X see next page for Concentrator Kit installation.
- Turn the oxygen (carrier gas) supply on.
- Connect the appropriate active or passive waste gas scavenging system to the 19mm outlet of the pop-off valve. Ensure the vaporizer has been filled per manufacturer’s instructions and is ready for use.
- Fill absorber canister to line indicated on the canister. Use supplied downtube cover (on-chain) when filling. Attach the absorber canister to the rebreathing head and tighten.
- Note: Absorbent not included.
- Leak test following instructions on 10 second pressure test card.
CONCENTRATOR KIT INSTALLATION
Stud Hardware Installed on Base Plate
- 4 3/8″-24 x 2″ Hex Bolt
- 8 3/8″ x 24 Hex Nut
- 12 3/8″ SS Washer
- 4 3/8″ Cushioning Washer
- In order as installed: Bolt, SS Washer, Base Plate, SS Washer, 2x Hex Nut, SS Washer, Cushioning Washer.
FOLLOW ALL STEPS ACCORDINGLY
- Unpack the concentrator and remove wheels by pulling away from (base) of the concentrator.
- Install stud hardware on base plate in correct holes based on where wheels were located on the concentrator.
- Set concentrator on studs sticking up from base plate.
- Note : Two people are recommended for this step.
- The single E-cylinder manifold will be pre-installed on the middle support shelf. Ensure the brass and black washer/seal is in place prior to installing the e-cylinder.
- Attach the two high-pressure, small green tubes from the anesthesia machine to the E-tank regulator by pushing the tubing into the fittings. Once resistance is felt, continue to push the tubing slightly further to secure connection. Gently pull on tubing to test for proper connection.
- Note: The E-cylinder must be turned on to operate the oxygen flush.
- Attach low pressure, large green tubing from the anesthesia machine to the concentrator by threading the green female fitting onto the male fitting on the concentrator.
- Optional : Use supplied cable tie to organize the green tubing.
WARNING
- When using the oxygen concentrator, flip the switch on the front of the anesthesia machine to the concentrator prior to turning the concentrator on.
- When using the oxygen concentrator, turn anesthesia machine flowmeter on to allow flow. The concentrator alarm will sound if this step is not followed.
- NOTE When using the oxygen concentrator, turn the concentrator flow to match the flowrate on the anesthesia machine.
ATTACHING CIRCUITS
F-CIRCUIT ATTACHMENT
For use with animals that weigh approximately 15 lbs. (7kg) or more.
- Connect the longer, single limb of the circuit to the inhalation valve.
- Connect the shorter, corrugated tube to the exhalation valve.
- Connect a rebreathing bag (appropriate to the patient size) to the port located below the popoff valve.
- Flip the Rebreathing/Non-rebreathing switch to Rebreathing when using this circuit.
- Note : It is not necessary to remove the non-rebreathing circuit.
NON-REBREATHING CIRCUIT ATTACHMENT
For use with animals that weigh approximately 15 lbs. (7kg) or less.
- Insert the adapter on the end of the long, clear tubing into the Common Gas Outlet (located above the Rebreathing/Non-rebreathing switch).
- Attach the corrugated tubing to the waste gas scavenging system.
- Confirm the bag bleed valve/slide valve is open on the non-rebreathing circuit.
- Flip the switch to non-rebreathing when using the non-rebreathing circuit. It is not necessary to remove the rebreathing circuit.
WARNING
- As a patient safety feature, the oxygen flush will not work with a non-rebreathing circuit. The oxygen flush is only piped to the rebreathing absorber.
GAS FLOW CHART
ROUTINE MAINTENANCE
Daily Testing
- Turn the flowmeter knob fully open and observe as the float ascends and descends freely. The float should rest completely at the bottom when knob is turned off.
- Inspect lines, connectors and accessories for wear prior to use.
- Perform a Leak Test (shown on bottom).
Cleaning
- Use a damp cloth to clean.
- DO NOT use alcohol.
- DO NOT spray cleaning agent directly onto the machine surface.
- Dilute Chlorhexidine or similar cleaning and disinfecting agent can be used.
BEFORE EACH USE, “LEAK TEST” THE MACHINE, AND ALSO ENSURE THAT THE WASTE GASES HAVE A CLEAR WAY THROUGH YOUR EVACUATION SYSTEM.
- Close the pop-off valve and create a tight seal of the patient circuit with your hand.
- Push the oxygen flush button, or turn on oxygen flow meter, until the bag is fully distended and/or the manometer reads 30-40 cm of H2O pressure.
- Release the oxygen flush button and/or ensure the oxygen flowmeter is off; float is at the bottom of the tube. Observe the manometer (or the bag if your machine does not have a manometer) for 20-30 seconds. The manometer needle should remain steady. If the manometer needle drops, the bag deflates, or if you hear a hissing sound, you have a leak.
- Check hoses, bag, vaporizer inlet and outlet, any fittings or seals at the top and/or bottom of your CO2 absorbent container for leaks (a soapy water solution can be used for leak test). When the pressure remains constant, with the oxygen turned off, your machine can be considered leak-free on the low-pressure side.
- Re-open the pop-off valve to your standard setting.
- Squeeze the bag (with your hand still over the patient circuit) to ensure the gases have a clear way out through your scavenging system.
TROUBLESHOOTING
PROBLEM | CAUSE ANALYSIS | SOLUTION |
---|---|---|
FLOWMETER WILL NOT TURN OFF |
Inspect needle valve for wear. Check adjustments.
|
Replace needle valve or make adjustments.
ANIMALS WILL NOT GO TO SLEEP/STAY ASLEEP
| 1. Inspect system for leak
2. Inspect breathing circuit and bag for leak or loose connections.
3. Check active waste gas system function/ attenuation.
4. Confirm vaporizer has an adequate amount of agent.
| 1. Tighten any loose connections (vaporizer endcaps, tubing connections).
2. Replace breathing circuit or bag. Fix loose connections.
3. Ensure active waste gas system function/ attenuation is working properly
4. Fill vaporizer with correct amount of agent.
FOR ASSISTANCE, CALL 1.800.525.5614
SPECIFICATIONS
- Weight 45 lbs./20.5 kg
- Dimensions 21″W x 21″D x 53″H (53mm x 53mm x 135mm)
- Vaporizer Endcaps 23mm tapered
- Flowmeter 0.2-4 LPM
- Oxygen (Carrier Gas) Inlet DISS Male
- Frame Aluminum
WARRANTY SERVICE
Limited Warranty Jorgensen Laboratories warrants to the original purchaser that the products, not including accessories, shall be free from defects in materials and workmanship under normal use, if maintained in accordance with Jorgensen Laboratories’ guidelines and used according to its labeling, for the period specified in the manual. Warranty Period 5 Years from the Invoiced Date of Purchase THIS LIMITED WARRANTY, IS IN LIEU OF AND EXCLUDES ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, BY OPERATION OF LAW OR OTHERWISE, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. This warranty is void if the product has been altered, misused, damaged by neglect or accident, tampered with, not properly maintained, not installed in strict compliance with applicable codes and ordinances, or repaired by persons not authorized by Jorgensen Laboratories. This warranty does not cover normal wear and tear and maintenance items and specifically excludes accessory items and any other equipment used with the product. Limitation of Remedies Jorgensen Laboratories’ only obligation under this limited warranty is the repair or replacement of the product. THIS IS THE EXCLUSIVE REMEDY. Jorgensen Laboratories shall not be liable for and hereby disclaims any direct, incidental, consequential or special damages or delays, including but not limited to loss of use, downtime, lost business, revenues and profits.
Warranty Procedure To obtain warranty service, contact Jorgensen Laboratories:
- 800.525.5614
- info@jorvet.com
- Jorgensen Laboratories Inc. 1450 N. Van Buren Ave. Loveland, CO 80538 Professional Veterinary Equipment Supplier Jorgensen Laboratories, Inc. Loveland, CO 80538 800.525.5614
- fax 970.663.5042
- www.jorvet.com