RICE LAKE IS6V2 Battery Charger Instructions

June 12, 2024
RICE LAKE

RICE LAKE-LOGO

RICE LAKE IS6V2 Battery Charger

Product Information

The IS6V2 battery charger is an encased power adapter that provides charging power at 7.35 V and 2 amps while charging and 7.0 V once the battery is fully charged. It features two LEDs that indicate AC power status and charge trip point. The charger has a dual mode (charge/float) to provide rapid battery charging while preventing overcharging. It is also designed to optimize battery life by maintaining the float voltage of 7.0 V to prevent battery discharge during normal AC operation.

LED Descriptions

LED LED Color Description
Power LED Green States the charger has power from an AC source and is

continually on
Charger LED (Multi-State LED)| Green| Charging status: Charging
| Green (Flashing)| Charging status: Float Charging
| | Charging status: Full Charge

Specifications

Power|
---|---
Input Voltage Range| 100 – 250 VAC, 50/60 Hz, 75 VA
Voltage Output| 7.35 VDC Charge, 7.0 VDC Float
Maximum Output Current| 2.0 A
Compliance|
August 10, 2022| www.ricelake.com
PN| 195930 Rev E

Product Usage Instructions

  1. Disconnect the 882IS battery pack cable from the 882IS indicator.
  2. Remove the battery pack from the indicator tilt stand and take it to a non-hazardous area.
  3. Remove the lockout device and battery cable from the battery.
  4. Plug the connector from the battery charger into the connector on the battery pack.
  5. Connect the charger power cord to a properly grounded outlet supplying the correct input voltage.
  6. Charge the battery. Note that it will take a minimum of 12 hours to charge a dead battery.
  7. The charger LED switches from a blue color to a green color when the unit reaches approximately 70% charged. At that time, the charger switches over to a float charge mode.
  8. The battery is considered fully charged when the charger LED is flashing green.
  9. To keep the battery at full capacity, it is recommended to leave the battery connected to the charger in float charge mode until ready to use. The battery can remain on the charger in float charge mode indefinitely without damaging it.

882IS Intrinsically Safe Indicator

IS6V2 Battery Charger Instructions

The IS6V2 battery charger instruction sheet contains important safety and operating instructions. Before using the battery charger, please read all instructions and cautionary information on the battery charger and the product using the battery.
Manuals and additional resources are available from the Rice Lake Weighing Systems website at www.ricelake.com
Warranty information can be found on the website at www.ricelake.com/warranties

CAUTION:

  • Do not expose the charger to rain or moisture.
  • Do not remove the cover.
  • Disconnect the charger from AC power before maintenance or cleaning.

WARNING:

  • Only use Rice Lake Weighing Systems approved charger for Intrinsically Safe indicator batteries.
  • Because hydrogen gas may be present, disconnect unit from AC power before connecting or disconnecting output connectors.

IMPORTANT: Connections to the battery pack must be made in a non- hazardous area before connecting the cable to the 882IS indicator.

RICE LAKE-IS6V2-Battery-Charger-1

Introduction

The IS6V2 battery charger is an encased power adapter that provides charging power at 7.35 V and 2 amps while charging and 7.0 V once the battery is fully charged.

Two LEDs indicate AC power status and charge trip point.

Table 1.  LED Descriptions

LED LED Color LED Description
Power LED Green States the charger has power from an AC source and is

continually on
Charger LED (Multi-State LED)| Blue| Charging status: Charging
Green| Charging Status: Float Charging
Green (Flashing)| Charging Status: Full Charge

The dual mode (charge/float) provides rapid battery charging while preventing hazardous overcharging. In addition, the charger is ideal for optimizing battery life as it will maintain a float voltage of 7.0 V to prevent battery discharge during normal AC operation.

Battery Charging and Operation

CAUTION: Re-charge batteries in a non-hazardous location.

  1. Disconnect the 882IS battery pack cable from the 882IS indicator.
  2. Remove the battery pack from the indicator tilt stand and take to a non-hazardous area.
  3. Remove the lockout device and battery cable from the battery.
  4. Plug the connector from the battery charger into the connector on the battery pack.
  5. Connect the charger power cord to a properly grounded outlet supplying the correct input voltage.
  6. Charge the battery.

NOTE:

  • It will take a minimum of 12 hours to charge a dead battery.
    The charger LED switches from a blue color to a green color when the unit reaches approximately 70% charged. At that time,the charger switches over to a float charge mode. The battery is considered fully charged when the charger LED is flashing green.

  • To keep the battery at full capacity, it is recommended to leave the battery connected to the charger, in float charge mode, until ready to use. The battery can remain on the charger in float charge mode indefinitely without damaging the battery. For more information on the IS6V2 Battery, see PN 194893.

Specifications

Power

  • Input Voltage Range: 100 – 250 VAC, 50/60 Hz, 75 VA
  • Voltage Output: 7.35 VDC Charge, 7.0 VDC Float
  • Maximum Output Current: 2.0 A

Environmental
Operating Temperature: 14°F to 104°F (-10°C to 40°C)

Compliance

Rice Lake Weighing Systems Specifications subject to change without notice.

230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319

www.ricelake.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

RICE LAKE User Manuals

Related Manuals