Interface 4850 Battery Powered Bluetooth Weight Indicator Instruction Manual
- June 12, 2024
- Interface
Table of Contents
4850 Battery Powered Bluetooth Weight Indicator
Operating Manual
Released June 23, 2023
DISPLAY & KEYPAD DETAILS
This instrument uses a 6-digit LCD (Liquid Crystal Display) with adjustable LED backlight. The Table below summarizes the display annunciators.
Symbol
| Display Indication
---|---
| Displays when the instrument reading is at Zero.
P
| The instrument is in Peak Hold mode.
lb
| The displayed force reading is in pounds (lbf).
kg
| The displayed force reading is in kilograms (kgf).
g
| The displayed force reading is in grams (gf).
oz
| The displayed force reading is in ounces (ozf).
| The displayed force reading is in Newtons (N).
N
| The displayed reading is negative (minus).
| Displays whenever the instrument reading is at rest, i.e., not in motion.
| When blinking, the battery needs to be replaced soon.
Notes about the backlight:
- If no keys are pressed within 60 seconds, the backlight will switch off to save battery life.
- You can wake up the backlight by pressing any key.
The keypad comprises five (5) function keys.
Marking | Keypad Function |
---|---|
Units | Selects the displayed unit of measure, e.g., lbf, kgf and N. |
Peak | Selects the peak hold display mode of the instrument, e.g., live |
data, positive peak and negative peak. (This function must first be enabled
using the F36 parameter)
Zero| Zeroes the force display reading.
Data| Sends the displayed data to the serial communication port.
On/Off| Press and hold for a few seconds to turn the instrument ON.
Press and hold for about five seconds to shut the instrument OFF.
Note: This unit automatically zeroes when powered ON, regardless of the applied load, its polarity, or its stability.
INSTALLATION & OVERVIEW
Remember that the installer is ultimately responsible to assure that a
particular installation will be and remain safe and operable under the
specific conditions encountered.
The instrument must be properly configured and calibrated prior to use.
Installation
Find a suitable location for the instrument and use the included bracket to
mount the unit to a wall or table. Use this handy guide for mounting the
bracket to a wall or table:
The INSTRUMENT ships with a force sensor cable which plugs into the
appropriate connector.
Use the following color codes to connect the force sensor.
Force Sensor Cable Leads
Color | Function | Color | Function | |
---|---|---|---|---|
Black | – Excitation | White | – Signal | |
Red | + Excitation | Green | + Signal |
The INSTRUMENT features one full duplex RS-232 serial port, designed for
connection to a computer or a serial printer. The same port may be also used
as a simplex, RS-232 port de-signed for connection to a remote display.
DSUB9 Connector
Pin No. | Wire Name |
---|---|
2 | RXD |
3 | TXD |
5 | Ground |
The INSTRUMENT requires 4-AA batteries to operate. To install the batteries, remove the plastic battery cover from the rear panel. Observe proper direction (polarity) of the batteries. Replace the battery cover.
The INSTRUMENT may also be powered by the included AC wall adaptor.
Getting Started
- Press and hold the ON/OFF key on the instrument for two seconds.
After a brief initialization period, the instrument will revert to a zero (“0”) force reading.
Your instrument is now ready for configuration and calibration.
INSTRUMENT CONFIGURATION
Configuration Menus
The Model 4850 contains three (3) menus to configure the instrument:
Setup (“F”) Menu – Configures all metrologically-related parameters including
calibration proce- dures.
User (“A”) Menu – Configures COM2 communication parameters and other misc.
parameters, e.g., automatic turn off.
COM1 (“C”) Menu – Configures COM1 communication parameters.
The configuration menus are laid out in the following vertical arrangement:
- Top [Menu selection] level
- Parameter level
- Selection level (or function level, e.g., span calibration)
Please review the following chart to get a feel for how to navigate among the various menus and parameters.
Entering the Setup (“F”) Configuration Menu
- Switch off the instrument by pressing and holding down the ON/OFF key for about 5 sec- onds.
- Press and hold down the ON/OFF key (about 20 seconds) until the screen shows “-F-”.
- Scroll down using the PEAK (down) key to reach the parameter level. The instrument shows “F 1”.
- Move from one “F” menu parameter to the next by using the DATA (left) or ON/OFF (right) keys. For example, to go from F1 to F2, press the ON/OFF key. To go from F2 back to F1, press the DATA key.
- Once you have arrived at the proper “F” menu parameter, e.g.,” F 1”, press the PEAK (down) key once to arrive at the selection level. The instrument displays the current parameter setting.
- If there is a selection list, scroll thru the available parameter settings, use the DATA (left) or ON//OFF (right) keys. Otherwise, use the arrow keys to adjust the displayed value to the new value.
- Once the setting you want is displayed on the screen, press the ZERO (set) key to save this value and revert up to the parameter level, e.g.,” F 1”.
To save all parameter settings, it is imperative to exit all menus in the following manner: At the parameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The instru- ment displays ‘SAvE’ and then automatically powers off.
Setup (“F”) Menu Descriptions
This section provides more detailed descriptions of the selections found in
the Setup Menu
Chart. Factory-set defaults are shown in bold with a checkmark; (√).
CODE/NAME | DESCRIPTION | SELECTION LIST |
---|
F1
Graduations
| Sets the number of full-scale graduations, i.e., capacity ÷ division.
DO NOT CHANGE
| Key-in
100 – 50000
50000 Ö
F2
Sampling Rate
| Sets the sampling rate in Hertz (measurements per second).| Key-in
1 – 1200
100 Ö
F6
Digital Filter
| Sets the number of measurements to be averaged to affect the speed and stability of the readings. A lower number provides a faster re- sponse. Choose the value that works best for your application.| 0 to 12
8 Ö
F8
Calibration Unit
| Sets the force unit of measure to be used in the calibration process. “1” = unit is lbf “2” = unit is kgf
“3” = unit is ozf “4” = unit is gf
“6” = unit is kN (kilonewton) “7” = unit is kip (kilopound)
| 1 Ö 2
3 4
6 7
F9
Display Divisions
| Sets the interval value. Use together with F10.| 1 Ö
2
5
F10
Decimal Pt.
| Sets the decimal point value. Use together with F9.| 0 Ö 0.0
0.00 0.000
0.0000 00
F15
Span Calibra- tion – Negative
| Places instrument into live negative (-) span calibration mode. Scroll- ing
down with the PEAK key one level begins the procedure.| Press PEAK key to
begin sequence
F16
Zero Calibra- tion
| Places instrument into live zero-calibration mode. Scrolling down with the
PEAK key one level begins the procedure.| Press PEAK key to begin
sequence
F17
Span Calibra- tion – Positive
| Places instrument into live positive (+) span calibration mode. Scroll- ing
down with the PEAK key one level begins the procedure.| Press PEAK key to
begin sequence
F18
View Calibra- tion Data
| Activates the function that allows you to view calibration values.
Scrolling down with the PEAK key one level begins the procedure.| Press
PEAK key to begin sequence
F19
Key-in Zero
| Allows you to key-in a known zero calibration value. Scrolling down with
the PEAK key one level begins the procedure.| Press PEAK key to begin
sequence
F20
Key-in Span
| Allows you to key-in known span calibration values. Scrolling down with the
PEAK key one level begins the procedure.| Press PEAK key to begin
sequence
F21
Factory Reset (US)
| This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. It will not overwrite any previously saved calibration data.
USE WITH CAUTION!| Press the PEAK key twice to exe- cute
F22
Factory Reset (EU)
| This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. It will not overwrite any previously saved calibration data.
USE WITH CAUTION!| Press the PEAK key twice to exe- cute
F23
Full Factory Reset
| This sub-menu will reset all instrument parameters to the default set-
tings. It will not overwrite any previously saved calibration data. USE WITH
EXTREME CAUTION!| Press the PEAK key twice to exe- cute
F25
Set Point Function
| Sets the function of the set points and relay outputs.| 0 to 10
0 Ö
F26
Calibration Mode
| Sets the instrument calibration mode.
“no” = Live, “yE5” = Key-In (mV/V)
| no Ö
yE5
F27
Key-in Calibra- tion
| Places instrument into key-in (mV/V) calibration mode. Scrolling down with
the PEAK key one level begins the procedure.| Press PEAK key to begin
sequence
F28
Reference Calibration
| Places instrument into reference (mV/V) calibration mode. Scrolling down with the PEAK key one level begins the procedure.
Factory Use Only. Password protected.
| Press PEAK key to begin sequence
F30
Application
| Sets one special application.
“0” = None (Gross/Net), “2” = Remote Display, “5” = Peak Hold, “
| 0 2 5Ö
F31
Quick Cal
| Sets the calibration stability level. Enable for hydraulic sensor calibra- tion.
“no” = Disabled “YES” = Enabled
| No Ö
YES
F35
Default Units Mode
| Sets the force unit of measure to be displayed when the instrument is switched on.
“1” = Pounds (lbf), “2” = Kilograms (kgf), “3” = ounces (ozf), “4” = grams (gf), “5” = Newtons (N)
“6” = Kilonewtons (kN) “7” = Kilopounds (kips)
| 1Ö 2
3 4
5 6
7
F36
Default Peak Hold Mode
| Sets the peak hold mode to be activated when the instrument is first switched on.
“rEAL” = Live, “hoLd P” = Positive Peak, “hoLd U” = Negative Peak
| rEALÖ hoLd P hoLd U
Entering the User (“A”) or COM (“C”) Menu
NOTE: Follow the same steps to enter the COM (“C”) Menu– just substitute
“C” for “A” below
- Switch off the instrument by pressing and holding down the ON/OFF key for about 5 sec- onds.
- Press and hold down the ON/OFF key (about 20 seconds) until the screen shows “-F-”.
- Press the ON/OFF (right) key once. The screen displays “-A-”.
- Scroll down using the PEAK (down) key to reach the parameter level. The instrument shows “A 1”.
- Move from one “A” parameter to the next by using the DATA (left) or ON/OFF (right) keys. For example, to go from A1 to A2, press the ON/OFF key. To go from A2 back to A1, press the DATA key.
- Once you have arrived at the proper “A” menu parameter, e.g. “A 1”, press the PEAK (down) key once to arrive at the selection level. The instrument displays the current parameter setting.
- If there is a selection list, scroll thru the available parameter settings, use the DATA (left) or ON/OFF (right) keys. Otherwise, use the arrow keys to adjust the displayed value to the new value.
- Once the setting you want is displayed on the screen, press the ZERO (set) key to save this value and revert up to the parameter level, e.g. “A 1”.
- To save all parameter settings, it is imperative to exit all menus in the following manner: At the parameter level, press the UNITS (up) key twice (two times) to exit the User Menu. The instrument displays ‘SAvE’ and then automatically powers off.
User (“A”) and COM (“C”) Menu Descriptions
This section provides more detailed descriptions of the selections found in
the User Menu Chart.
Factory-set defaults are shown in bold with a checkmark; (√).
CODE/NAME | DESCRIPTION | SELECTION LIST |
---|
A1/C1
Baud Rate
| Selects the baud rate for data transmission through the serial port.| 600, 1200, 2400,
4800, 9600 Ö,
19200, 38400,
57600, 115200
A2/C2
Data Bits, Parity and Stop Bits
| Selects the number of data bits and parity of serial transmission. “8_n_1” = 8 data bits with no parity bit and one stop bit “8_n_2” = 8 data bits with no parity bit and two stop bits “7_n_1” = 7 data bits with no parity bit and one stop bit “7_n_2” = 7 data bits with no parity bit and two stop bits “7_E_1” = 7 data bits with even parity bit and one stop bit “7_E_2” = 7 data bits with even parity bit and two stop bits “7_o_1” = 7 data bits with odd parity bit and one stop bit “7_o_2” = 7 data bits with odd parity bit and two stop bits| 8_n_1 Ö
8_n_2
7_n_1
7_n_2
7_E_1
7_E_2
7_o_1
7_o_2
A3/C3
Serial Port Mode
| Selects the mode of the serial port: Refer to Serial Port Information for more details.
“0” = Demand Full Duplex “1” = Continuous Full Duplex “2” = Auto Print
| 0 Ö
1
2
A6/C6
Output String
| Selects fixed output string for serial port. Refer to Serial Port Infor- mation for more details.
“0” = String Format 1 (Condec Demand) “1” = String Format 2 (Condec Continuous) “2” = Text Print Ticket
“3” = Text Print Ticket with MP-20 Auto Label Feed
| 0 Ö (A6)
1
2 Ö (C6)
3
A8
ID Number
| Selects ID Number for serial port.
Factory Use Only. Password protected.
| Press PEAK key to begin sequence
A10
Auto Power Off
| Selects the automatic power off time, expressed in minutes of inactiv- ity
(keys and force measurement change). Pressing the PEAK key to scroll down one
level begins the sequence.| Key-in 0 to 30 30 Ö
A12
Backlight Brightness
| Selects the brightness of the LCD backlight. Selections are in % of full brightness. Intended to help save battery life.| 0 (OFF)
20 50
75 100 Ö
A22
Low Battery Auto Power Off
| Selects the automatic power off time of the instrument after it enters a low
battery condition. Expressed in minutes. Pressing the PEAK key to scroll down
one level begins the sequence.| Key-in 0 to 99 02 Ö
A23
Audible Key Feedback
| Selects function of the audible key feedback (beeper).
“no” = Disabled “YES” = Enabled
| no
YES Ö
INSTRUMENT CALIBRATION
Calibration Overview
There are two ways to calibrate the instrument:
- Live calibration: You will be calibrating an actual load sensor to the instrument using live test loads. You can have up to seven positive calibration points and up to seven negative calibration points. These calibration points are denoted as C1 through C7.
- mV/V calibration: You will be calibrating the instrument using a load cell simulator (cal- ibrator). Previously established mV/V input signal values – and the corresponding force values – are required. You can have up to six positive calibration points and up to six negative calibration points. These calibration points are denoted as C1 through C6. The absolute value of each subsequent calibration point should be higher than the last, e.g., the C2 value should be greater than the C1 value, etc.
Live Calibration Overview
Live calibration comprises two main steps: zero (F16) and span (F17). A third
calibration menu for negative span (F15) is also available for systems
operating in both tension and compres- sion.
What follows is the recommended live calibration sequence. Any deviations from
this may cause unwanted behavior:
- Place the force sensor onto your test fixture, with no other forces applied.
- Navigate to F17 (span calibration) and press the PEAK (down) key; indicator prompts for the first calibration point, C1.
- Key in number 111111 and press the ZERO (set) key; indicator will show “FIT” momentarily and then resume its prompt for the first calibration point, C1.
- Follow the F17 procedure (see next section) for up to seven span calibration points.
- Optional: repeat steps 2-4 for negative span calibration (F15)
- Mount the force sensor onto its final fixture.
- Navigate to F16 and perform the zero calibration.
Notes:
- You may see dashes appear temporarily on the screen during live calibration. This simply means that the instrument is processing data.
- To disable the positive calibration slope, simply enter a zero value (e.g., “0.0”) for C1 while in F17.
- To disable the negative calibration slope, simply enter a zero value (e.g., “0.0”) for C1 while in F15.
Live Span Calibration Instructions (F17)
- While in the Setup mode, scroll to “F 17”, and then scroll down once using the PEAK (down) key. The indicator will briefly display ‘C 1’ and then prompt you to enter the data for the first calibration point (C1). The previously saved force value will be displayed with one digit blinking.
- Use the four directional keys to adjust the displayed value to 111111 and press the ZERO (set) key; indicator will show “FIT” momentarily and then automatically record the fixture ref- erence point. The previously saved force value will be displayed with one digit blinking.
- Apply the first test load to the force sensor.
- Use the four directional keys to enter in the actual positive force value, e.g., 1000.0 lbf. Increase the flashing digit by pressing the UNITS key. Decrease the flashing digit by press- ing the PEAK key. Pressing the DATA key or the ON/OFF key will change the position of the flashing digit.
- Press the ZERO (set) key to save the value. The indicator briefly displays ‘End C1’ and then moves to the next calibration point (C2).
- Repeat steps 3 through 5 to enter data for the remainder of the calibration points. You need not enter data for all seven calibration points. To cease entering additional calibration points, simply enter zero for the new value, e.g., 0.0 lbf.
- At the conclusion of C7 (or the last calibration point), the instrument will show “-donE” and reverts to the top setup menu level, i.e.: “-F-”.
If the calibration was not successful, one of the following error messages will appear.
- “Err0” – The calibration test load or the keyed-in load is larger than the full capacity of the instrument. Change the calibration test load or check the input data.
- “Err1” – The calibration test load or the keyed-in load is smaller than 1% of the full ca- pacity of the instrument. Change the calibration test load or check the input data.
- “Err2” – There is not enough signal from the force sensor to complete the calibration process. Most common causes include incorrect force sensor wiring, a mechanical ob- struction or a faulty (damaged) force sensor.
Take the indicated action to correct the problem, and then perform a new calibration. To save all calibration settings, you must exit all menus in the following manner: At the pa- rameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The in- strument displays ‘SAvE’ and then automatically powers off.
Live Negative Span Calibration Instructions (F15)
For negative calibration data, you will not see a minus sign on the screen.
You need not be con- cerned as all data entered is presumed to be a negative
value, e.g., 1000.0 lbf is really -1000.0 lbf.
- While in the Setup mode, scroll to “F 15”, and then scroll down once using the PEAK (down) key. The indicator will briefly display ‘C 1’ and then prompt you to enter the data for the first calibration point (C1). The previously saved force value will be dis- played with one digit blinking.
- Use the four directional keys to adjust the displayed value to 111111 and press the ZERO (set) key; indicator will show “FIT” momentarily and then automatically record the fixture reference point. The previously saved force value will be displayed with one digit blinking.
- Apply the first test load to the force sensor.
- Use the four directional keys to the actual positive force value, e.g., 1000.0 lbf. Increase the flashing digit by pressing the UNITS key. Decrease the flashing digit by pressing the PEAK key. Pressing the DATA key or the ON/OFF key will change the position of the flashing digit.
- Press the ZERO (set) key to save the value. The indicator briefly displays ‘End C1’ and then moves to the next negative calibration point (C2).
- Repeat steps 3 through 5 to enter data for the remainder of the calibration points. You need not enter data for all seven calibration points. To cease entering additionalcalibration points, simply enter zero for the new value, e.g., 0.0 lbf.
- At the conclusion of C7 (or the last calibration point), the instrument will show “-donE” and reverts to the top setup menu level, i.e.: “-F-”.
To save all calibration settings, you must exit all menus in the following manner: At the pa- rameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The in- strument displays ‘SAvE’ and then automatically powers off.
Live Zero Calibration Instructions (F16)
-
While in the Setup mode, scroll to “F 16”, and then scroll down once using the PEAK (down) key. The display will momentarily show “C 0” followed by a value. This value is the internal A/D count and can prove useful when trying to troubleshoot setup
problems. If you see an ‘N’ on the screen, this indicates that the displayed value is neg- ative (minus). -
Mount the force sensor onto its final fixture, assure a no-load condition and then press the PEAK key to zero the display reading. Do NOT skip this step!
-
Quickly press the ZERO (set) key to save the zero-point value. The display will show “SET” and “EndC0” momentarily, and then revert up to F16.
To save all calibration settings, you must exit all menus in the following manner: At the pa- rameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The in- strument displays ‘SAvE’ and then automatically powers off.
Underneath the display you may find down arrows which have been activated. For each down arrow displayed, add 1 million to the displayed count. (Max value is 4.8 million counts)
Key-in mV/V Calibration Instructions (F27)
You will be entering data for up to six positive calibration points and up to
six negative calibra- tion points. These calibration points are denoted as C1
through C6. The absolute value of each subsequent calibration point should be
higher than the last.
Each calibration point requires entry of two values:
A force value, e.g., 200 lbf
A corresponding mV/V input signal value, e.g., 1.0125 mV/V.
- While in the Setup mode, scroll to “F 27”, and then scroll down once using the PEAK (down) key. The indicator will briefly display ‘Pos C’ and then prompt you to enter the data for the first positive calibration point (C1). The previously saved force value will be displayed with one digit blinking.
- Use the four directional keys to enter in the actual positive force value, e.g., 200 lbf.
- Press the ZERO (Set) key to save the force value. The indicator will prompt you to enter the corresponding mV/V input signal. The previously saved mV/V value will be dis- played with one digit blinking.
- Use the four directional keys to enter in the actual mV/V value.
- Press the ZERO (Set) key to save the mV/V value. The indicator briefly displays ‘End C1’ and then moves to the next positive calibration point (C2).
- Repeat steps 2 through 5 to enter data for the remainder of the positive calibration points. You need not enter data for all six positive calibration points. To cease en- tering additional calibration points, simply enter zero for both values, e.g., 0 lbf and 0.0000 mV/V.
- The indicator will briefly display ‘Neg C’ and then prompt you to enter the data for the first negative calibration point (C1). The previously saved force value will be displayedwith one digit blinking.
- Use the four directional keys to enter in the actual negative force value, e.g., 200 lbf. For negative calibration data, you will not see a minus sign on the screen. You need not be concerned as all data entered is presumed to be a negative value, e.g., 200 lbf is really -200 lbf.
- Press the ZERO (Set) key to save the force value. The indicator will prompt you to enter the corresponding mV/V input signal. The previously saved mV/V value will be dis- played with one digit blinking.
- Use the four directional keys to enter in the actual negative mV/V value. For negative calibration data, you will not see a minus sign on the screen. You need not beconcerned as all data entered is presumed to be a negative value, e.g., 1.0125 mV/V is really -1.0125 mV/V.
- Press the ZERO (Set) key to save the value. The indicator briefly displays ‘End C1’ and then moves to the next negative calibration point (C2).
- Repeat steps 8 through 11 to enter data for the remainder of the negative calibration points. You need not enter data for all six negative calibration points. To cease en- tering additional calibration points, simply enter zero for both values, e.g., 0 lbf and 0.0000 mV/V.
- When all values are entered successfully, the display will show “End Cx” momentarily before reverting to F27.
To save all calibration settings, you must exit all menus in the following manner: At the pa- rameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The in- strument displays ‘SAvE’ and then automatically powers off.
ADVANCED OPERATION
Peak Hold Mode
This mode is used to capture peak forces measured during a specific process.
These peak force values are held (frozen) on the screen. A common application
is determining the breaking point of a part or assembly. The instrument
captures both positive and negative peak forces. Peak forces are not retained
when the instrument is powered OFF.
This mode of operation is enabled at the Factory [the F30 parameter = “5”].
Use the F36 param- eter setting to select the default (power on) peak hold
mode.
-
Push the PEAK key to activate the peak mode you want to use (repeat as needed):
– “hoLd P” = Positive Peak; the “P” annunciator is ON.
– “hoLd U” = Negative Peak; the “P” annunciator is ON
– “LiVE” = Live Measurement (peak forces not captured); the “P” annunciator is OFF -
Apply force to the item under test. The measured peak force is displayed on the screen.
-
To reset all peak values to zero, press and hold the PEAK key for about 3 seconds un- til the display shows “Clr P”. This can be done in any peak hold mode.
SERIAL PORT INFORMATION
SERIAL PORT MODES
DEMAND DUPLEX MODE
The Demand Duplex Mode provides a two-way serial transmission mode. In this
mode, the output information is transmitted on demand; either by pressing the
DATA key on the instrument’s front panel or upon receiving a recognized
command from a host de-
vice (i.e., computer).
NOTE: Ensure that your cabling has a crossover (null modem) and contains
the proper handshaking lines.
CONTINUOUS DUPLEX MODE
The Continuous Duplex Mode provides a two-way serial transmission mode. In
this mode, the output information is transmitted continuously making it a
popular choice for remote displays and other remote devices requiring a
constant data stream. The trans-
mission automatically occurs at the end of each display update. The instrument
will re- act upon receiving a recognized command from a host device.
RECOGNIZED HOST COMMANDS (applies to both demand and continuous duplex modes)
ASCII code (Hex) | Symbol | Action by the instrument |
---|---|---|
50 or 70 | P or p | Transmit the displayed force data through the serial port. |
5A or 7A | Z or z | Zero the instrument |
43 or 63 | C or c | Change the displayed unit of measure |
46 or 66 | F or f | Transmit the firmware version through the serial port. |
AUTO PRINT MODE
The Auto Print Mode provides a one-time serial transmission once a non-zero,
stable condition is achieved.
OUTPUT STRINGS
TEXT PRINT TICKET
The Text Print Ticket is designed specifically for a serial printer.
GROSS 1000.0 lb
STRING FORMAT 1 (Condec Demand String)
String Format 1 is designed for two-way communication.
NOTE: Additional unit is available:
NT = Newtons
oz = ounce
g
lb = pounds
kg = kilograms
STRING FORMAT 2 (Condec Continuous String)
String Format 1 is designed for one-way communication.
NOTE: Additional unit is available:
N = Newtons
O = oz
G= grams
L= pounds
K = grams
SPECIFICATIONS
Instrument Specifications
Enclosure: Aluminum/ABS enclosure (IP54)
Display: 0.8”, 6-digit LCD with LED backlight
Upgraded A-to-D converter (AD-02):
Max. Signal Input: ±4.5 mV/V
Resolution: Up to 4,800,000 Internal Counts @ 4.5 mV/V
Conversion Rate: 1 to 1200 Hz, selectable
Serial Port: Full Duplex RS-232C
Operating temperature: 14°F to 104°F (-10°C to 40°C)
Battery Power
– 4 AA alkaline
– User replaceable
– 85+ continuous hours of operation under typical operating conditions
ERROR MESSAGES
| Batteries need to be replaced.
---|---
Err 99| Setup menus blocked. Place internal calibration jumper back to its
original position.
No-ad| A signal from the load sensor has not been detected.
TROUBLESHOOTING
Issue / Recommendation
“Low Battery” icon blinks on the digital readout, then the instrument powers
off.
Replace the batteries.
Instrument turns off on its own.
The instrument has a power conservation feature, set to automatically power
off the instrument after 30 minutes of non-use. If your needs require a
different setting, call Interface Inc.Tech Support or installer.
Display is erratic or No-Ad message.
– A battery may be fully depleted. If so, this condition can cause erratic
displays. Power off the instrument and replace the battery.
– Check the testing machine load cell for any obstructions or foreign debris
– Loose or faulty connection between the instrument and the load sensor
– Electromagnetic Interference (EMI). Call Interface Inc. Tech Support or
installer
Warranty
All instrument products from Interface Inc., (‘Interface’) are warranted
against defective material and workmanship for a period of (1) one year from
the date of dispatch. If the ‘Interface’ product you purchase appears to have
a defect in material or workmanship or fails during normal use within the
period, please contact your Distributor, who will assist you in resolving the
problem. If it is necessary to return the product to ‘Interface’ please
include a note stating name, company, address, phone number and a detailed
description of the problem. Also, please indicate if it is a warranty repair.
The sender is responsible for shipping charges, freight insurance and proper
packaging to prevent breakage in transit. ‘Interface’ warranty does not apply
to defects resulting from action of the buyer such as mishandling, improper
interfacing, operation outside of design limits, improper repair or
unauthorized modification. No other warranties are expressed or implied.
‘Interface’ specifically disclaims any implied warranties of merchantability
or fitness for a specific purpose. The remedies outlined above are the buyer’s
only remedies. ‘Interface’ will not be liable for direct, indirect, special,
incidental or consequential damages whether based on the contract, tort or
other legal theory. Any corrective maintenance required after the warranty
period should be performed by ‘Interface’ approved personnel only.
Interface Inc.
7418 East Helm Drive, Scottsdale, Arizona 85260 USA
www.interfaceforce.com