HANNA HI98509 Checktemp1 Digital Thermometer Instruction Manual
- June 12, 2024
- HANNA
Table of Contents
INSTRUCTION MANUAL
HI98509 Checktemp® 1
HI98509 Checktemp1 Digital Thermometer
Dear Customer,
Thank you for choosing a Hanna Instruments ® product. Please read this
instruction manual carefully before using the instrument. For more information
about Hanna Instruments and our products, visit
www.hannainst.com or e-mail us at
sales@hannainst.com.
For technical support, contact your local Hanna Instruments office or e-mail
us at tech@hannainst.com.
Preliminary Examination
Remove the HI98509 Checktemp® 1 and accessories from the packing material and
examine it carefully. If you require any further information, please contact
Hanna Instruments technical support team at
tech@hannainst.com.
HI98509 is supplied with: 1.5V AAA battery (3 pcs.), meter stand, and
Instruction manual with instrument quality certificate printed on the back.
General Description & Intended Use
HI98509 is a high-accuracy thermometer connected to a stainless steel probe by
a 1 m (3.3’) flexible, silicone cable. The probe incorporates an NTC
thermistor sensor, providing accurate temperature measurements compliant with
HACCP requirements. Lightweight and easy to clean the sharp-tip penetration
probe, is ideal for checking the temperature of semi-solid products such as
fruit, vegetable, and cheese. This probe can also handle measurements in
liquid, air, and frozen materials.
Specifications
Range | −50.0 to 150.0 °C ; −58.0 to 302.0 °F |
---|---|
Resolution | 0.1 °C (−50.0 to 150.0 °C); 0.1 °F (−58.0 to 199.9 °F) / 1 °F |
(above 200 °F)
Accuracy| ±0.2 °C (−30.0 to 120.0 °C); ±0.3 °C (outside: −50.0 to −30.0 °C and
120.0 to 150.0 °C)
±0.5 °F (−22.0 to 199.9 °F); ±1 °F (outside: −58.0 to −22.0 °F and 200.0 to
302.0 °F)
Probe| AISI 316 stainless steel with 1 m (3.3’) silicone cable
Battery type / life| 1.5V AAA / approximately 2 years of use
Auto shut-off| 8 min. (default), 60 min. or OFF
Environment| −30 to 50 °C (−22 to 122 °F); IP65
Dimensions| 107 x 59 x 17 mm (4.2 x 5.3 x 0.7“)
Weight| 130 g (4.6 oz.)
Certification| EN13485; suitability: storage
Climatic environment: E; accuracy class: 0.5
Operation
Press the ON/OFF button to turn the thermometer on.
CAL Check™ automatically verifies the thermometer’s circuit accuracy at 0.0
°C.
During start up, the calibration of the instrument’s circuit is checked and
the result is briefly displayed to show the calibration result.
“CAL” and “-0-” indicate that the instrument performs within specifications.
“Err” displayed during measurement indicates that the user has to wait for the
reading to stabilize.
Note: The CAL Check system can only check the electronic circuit
calibration and does not check the sensor of the Checktemp 1. It is
recommended to check the sensor by performing an ice test or comparative test
against a reference thermometer.
Meter Setup
Remove the battery cover to find the SET button located inside.
Press SET button once for temperature unit selection and twice for auto shut-
off selection. With current temperature unit displayed:
- Press the ON/OFF button to select “°C” or “°F”. Press the SET button once to save.
- Press the ON/OFF button to navigate Auto shut-off options: 8 minutes (“d08”, default), 60 minutes (“d60”), OFF (“d–”). Press the SET button once to save.
Battery Replacement
When the battery symbol is displayed blinking, or the thermometer does not turn on, batteries need to be replaced. To replace the batteries:
- Remove the four screws on the back of the instrument to open the battery compartment.
- Remove the old batteries.
- Insert three new 1.5V AAA batteries in the battery compartment while paying attention to the correct polarity. Do not mix old and new Alkaline batteries.
- Close the battery compartment using the four screws.
Note: Only use the battery type specified in this instruction manual. Old batteries should be disposed of in accordance with local regulations.
Certification
All Hanna Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. The product should not be
treated as household waste.
Instead, hand it over to the appropriate collection point for the recycling of
electrical and electronic equipment, which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not
dispose of them with other household waste. Hand them over to the appropriate
collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative
consequences for the environment and human health. For more information,
contact your city, your local household waste disposal service, or the place
of purchase.
Recommendations for Users
Before using this instrument, make sure that it is entirely suitable for your
specific application and for the environment in which it is used. Any
variation introduced by the user to the supplied equipment may degrade the
instrument’s performance. For your and the instrument’s safety do not use or
store the instrument in hazardous environments.
Warranty
HI98509 Checktemp 1 is warranted for a period of one year against defects in
workmanship and materials when used for its intended purpose and maintained
according to instructions. This warranty is limited to repair or replacement
free of charge. Damage due to accidents, misuse, tampering, or lack of
prescribed maintenance is not covered. If service is required, contact your
local Hanna Instruments office. If under warranty, report the model number,
date of purchase, serial number, and the nature of the problem. If the repair
is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments office, first obtain
a Returned Goods Authorization (RGA) number from the Technical Service
department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products withoutadvance notice.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner,
Hanna Instruments Inc., Woonsocket,
Rhode Island, 02895, USA.
IST98509 09/21
References