Ossila T2002 Solar Cell I-V Test System User Manual
- June 4, 2024
- Ossila
Table of Contents
Ossila T2002 Solar Cell I-V Test System
Safety
Use of Equipment
The Ossila Solar Cell I-V Test System is designed to be used as instructed. It is intended for use under the following conditions:
- Indoors in a laboratory environment (Pollution Degree 2).
- Altitudes up to 2000m.
- Temperatures of 5°C to 40°C; maximum relative humidity of 80% up to 31°C.
The unit is supplied with a 24 VDC power adapter with a power cord for the
country of purchase, in accordance with European Commission regulations and
British Standards. Use of any other electrical power cables, adaptors, or
transformers is not recommended
Hazard Icons
The following symbols can be found at points throughout the rest of the
manual. Note and read each warning before attempting any associated operations
associated with it:
Hazard warning labels used in this manual.
Associated Hazard
Electrical shock
General Hazards
Before installing or operating the Ossila Solar Cell I-V Test System there are
several health and safety precautions which must be followed and executed to
ensure safe installation and operation.
Power Cord Safety
Emergency power disconnect options: use the power cord as a disconnecting method and remove from wall. To facilitate disconnect, make sure the power outlet for this cord is readily accessible to the operator.
Servicing
If servicing is required, please return the unit to Ossila Ltd. The warranty will be invalidated if:
- Modification or service has taken place by anyone other than an Ossila engineer.
- The Unit has been subjected to chemical damage through improper use.
- The Unit has been operated outside the usage parameters stated in the user documentation associated with the Unit.
- The Unit has been rendered inoperable through accident, misuse, contamination, improper maintenance, modification, or other external causes.
Health and Safety – Servicing
Servicing should only be performed by an Ossila engineer. Any modification or alteration may damage the equipment, cause injury, or death. It will also void your equipment’s warranty.
Requirements
details the power requirements for the Solar Cell I-V Test System, and the
minimum computer specifications for the Ossila Solar Cell I-V software.
Solar Cell I-V Test System requirements.
Power | 24 VDC |
---|---|
Operating System | Windows 10 or 11 (32-bit or 64-bit) |
CPU | Dual Core 2 GHz |
RAM | 2 GB |
Available Hard Drive Space | 192 MB |
Monitor Resolution | 1680 x 1050 |
Connectivity | USB 2.0 |
Ethernet (requires DHCP)
Unpacking
Packing List
The standard items included with the Ossila Solar Cell I-V Test System are:
- Ossila Solar Cell I-V Test System.
- 24 VDC power adapter.
- USB-B cable.
- USB memory stick pre-loaded with the user manual, USB drivers, quality control data, and software installer.
- Resistor test device.
Damage Inspection
Examine the components for evidence of shipping damage. If damage has occurred, please contact Ossila directly for further action. The shipping packaging will come with a shock indicator to show if there has been any mishandling of the package during transportation.
Specifications
The Solar Cell I-V Test System specifications
Ossila Solar Cell I-V Test System specifications.
Voltage range | ±10 mV to ±10 V |
---|---|
Voltage accuracy | ±10 mV offset |
Voltage resolution | 170 μV |
Current range | ±10 nA to ±200 mA (5 ranges) |
Current accuracy | ±10 nA (at 20 μA range) |
Current resolution | 0.1 nA (at 20 μA range) |
Substrate Size | 20 mm x 15 mm |
25 mm x 25 mm
75 mm x 25 mm
Substrate Compatibility| T2002B, T2003B – S211
T2002E, T2003E – S2006 T2002F, T2003F – S241, S251
Overall Dimensions (Automated)| Width: 150 mm
Height: 55 mm
Depth: 300 mm
Overall Dimensions (Manual)| Source Measure Unit
Width: 125 mm; Height: 55 mm; Depth: 185 mm Test Board
Width: 105 mm; Height: 40 mm; Depth: 125 mm
(T2002F – Width: 100 mm; Height: 40 mm; Depth: 150 mm)
System Components
The Solar Cell I-V Test System is comprised of 2 items: the Solar Cell I-V
Test System (Figure 7.1 or Figure 7.2) and the Ossila I-V Curve software
Installation
- Install the Ossila Solar Cell I-V software on your PC.
- Run the file ‘Ossila-Solar-Cell-IV-Installer-vX-X-X-X.exe’ on the USB memory stick provided.
- Follow the on-screen instructions to install the software.
- Connect the 24 VDC power adaptor to the power socket on the rear of the unit.
- Connect the unit to your PC using the provided USB-B cable, or an Ethernet cable if preferred.
Note : The Ossila Solar Cell I-V software can also be downloaded from ossila.com/pages/software-drivers
Measurement Types
The Solar Cell I-V software can perform 3 different types of measurements.
Each measurement type can be selected using the tabs at the top of the window.
The available measurements are:
- Characterisation (Section 9.1.1).
- Lifetime (Section 9.1.2).
- Stabilised Current (Section 9.1.3).
Each measurement type requires several settings to be selected before it can
be performed. Settings that are shared between all measurements are detailed
in Section 9.3. Measurement-specific settings are detailed in Sections 9.4,
9.5, and 9.6.
Characterisation
The Characterisation tab performs current-voltage (I-V) measurement and
analysis of solar cells. The analysis calculates the following properties:
- Power conversion efficiency (PCE)
- Fill factor (FF)
- Short-circuit current density (Jsc)
- Open-circuit voltage (Voc)
- Shunt resistance (Rsh)
- Series resistance (Rs)
- Maximum power (Pmax)
Lifetime
The Lifetime tab tracks PCE, FF, Jsc, and Voc over time by performing periodic
I-V measurements and analysis. Between I-V measurements, the solar cell can be
held at short-circuit, open-circuit, or maximum power.
Stabilised Current
The Stabilised Current tab lets you measure the evolution of the
photogenerated current at specific voltages.
Quick Start Guide
-
Start the Ossila Solar Cell I-V software. The window shown in Figure 9.1 will open.
-
Choose a measurement type as described in Section 9.1.
-
Place your sample in the device holder.
-
Place the device holder beneath your solar simulator.
-
Set the appropriate settings in the software (explained in more detail in Sections 9.4 – 9.8).
-
Open the shutter of your solar simulator.
-
Click the ‘Measure’ button.
I. For each pixel, measurements are performed using the chosen measurement settings.
II. This process is repeated until all pixels have been measured. -
If automatic saving is turned on, the measurement data and settings will then be saved.
Shared Software Settings
The settings in these sections are shared between all measurement types.
System Settings
System Address
-
Select the COM port or IP address of the connected unit you intend to use (USB and Ethernet connection respectively).
-
This box will be populated automatically with the addresses of any units connected to the computer.
Voltage Source -
Select which SMU channel of the Source Measure Unit the test board is connected to.
-
‘SMU 1’ will be automatically selected when pixel switching is set to ‘Automated’.
Current Range -
Select the range of currents to be used for the measurement.
-
This defines the upper limit and accuracy of current measurements that can be performed by the unit. The values for each range are given in Table
-
Automatic range selection will start on the lowest current range and automatically switch to higher ranges if the current increases above the maximum for a range.
Maximum Current | Accuracy | Precision | Resolution |
---|---|---|---|
±200 mA | ±500 μA | 10 μA | 1 μA |
±20 mA | ±10 μA | 1 μA | 100 nA |
±2000 μA | ±1 µA | 100 nA | 10 nA |
±200 μA | ±100 nA | 10 nA | 1 nA |
±20 μA | ±10 nA | 1 nA | 100 pA |
Sampling Rate
- Select the number of samples to be taken for each data point.
I. A higher number of samples per point will improve the accuracy and precision of the measurement. However, this will increase the time taken for the measurement to be performed.
Device Details
Pixel Switching
-
Select whether changing the connected pixel is done manually or automatically by the system.
Pixels to Test (Automated Switching) -
Select which pixels to measure.
I. The pixel numbers are labelled on the device holder.
Number of Pixels (Manual Switching)
- Set the number of individual solar cell pixels in the device being measured.
Pixel Area
-
Set the area in cm2 of each pixel in the device.
Inverted Device -
Set whether the device to be measured is inverted.
I. This option should be on if the anode of your device connects to the ‘cathode’ pins in the device holder.
Characterisation Settings
Measurement Settings
Start Voltage
-
Set the voltage in volts at which to start the current-voltage measurement.
-
This can be set between -10 V and +10 V.
End Voltage -
Set the voltage in volts at which to end the current-voltage measurement.
-
This can be set between -10 V and +10 V.
Voltage Increment -
Set the step size in volts for changing the voltage during current-voltage measurement.
Voltage Settle Time -
Set the time in seconds between applying a voltage and measuring the current.
-
This has a maximum of 10 seconds.
Illumination -
Set the illumination intensity (in mW.cm-2) being used during the measurement.
Hysteresis I-V -
This option performs a reverse current-voltage measurement after the forward current-voltage measurement has completed.
-
This reverses the set start and end voltages and uses the same voltage increment and settle time as the forward measurement.
Lifetime Settings
Measurement Settings
Start Voltage
-
Set the voltage in volts at which to start the current-voltage measurement.
-
This can be set between -10 V and +10 V.
End Voltage -
Set the voltage in volts at which to end the current-voltage measurement.
-
This can be set between -10 V and +10 V.
Voltage Increment -
Set the step size in volts for changing the voltage during current-voltage measurement.
Voltage Settle Time -
Set the time in seconds between applying a voltage and measuring the current.
-
This has a maximum of 10 seconds.
Illumination -
Set the illumination intensity in mW.cm-2 being used during the measurement.
Hysteresis I-V -
Set whether to perform a reverse current-voltage measurement after the forward current-voltage measurement has completed.
-
This reverses the set ‘start’ and ‘end’ voltages and uses the same voltage increment and settle time as the forward measurement.
Lifetime Parameters
Duration
-
Set the total duration in hours of the lifetime measurement.
Interval -
Set the time interval in minutes between performing repeat current-voltage measurements of the device.
Hold Voltage -
Set the voltage that all pixels will be held at between measurements.
-
This can be set as:
Short-Circuit – hold at 0 V. -
Maximum Power – hold at the average maximum power point determined from most recent current-voltage curve.
-
Open-Circuit – hold at the average open-circuit voltage determined from the most recent current-voltage curve.
Note : As the voltage source is a single channel, the hold voltage will be the same for all pixels being tested.
Stabilised Current Settings
Measurement Settings
Voltage
-
Set the voltage to apply to the sample for the measurement.
-
This can be set between -10 V and +10 V.
Total Time -
Set the total length of the measurement in minutes.
Measurement Interval -
Set the time between each current measurement in seconds.
Saving and Loading Settings
Save Settings
-
Saves the current settings as a profile that can be loaded quickly for use at another time.
-
When clicked, you will be prompted to name the settings profile.
-
If the name is already in use, you will be asked if you wish to overwrite the previous profile.
-
The name cannot contain the characters: \ / : * ? “ < > |
-
The settings profile will be added to the drop-down box using the given name.
Settings Profiles -
Select a saved settings profile from the drop-down box.
-
The settings fields will be populated with the saved values.
-
Settings profiles can be deleted by selecting the profile, and then clicking the red ‘delete’ button next to the drop-down box.\
Saving Results
Save After Measurement
- Set whether the measurement data will be saved after the measurement has completed.
Warning: Automatic saving can be turned off for lifetime measurements. However, manual saving is unavailable for lifetime measurements, so you will not be able to save your data if it is turned off.
- The program allows for data to be saved automatically and manually once the measurement is complete.
- For automatic saving, the ‘Saving’ fields must be filled in before the measurement can start, these are detailed below.
- For all measurements, a save directory must be specified. This can be done either by:
- Manually typing the directory into the ‘Save Directory’ field,
- Copy and pasting from your file explorer,
- Clicking the file icon button, which will open a dialog box to allow the selection of a folder to save to.
- All output files are comma separated variable (.csv) files.
Save Directory
-
Set the directory in which to create the data files.
-
This can be filled in by:
-
Manually typing the directory into the ‘Save Directory’ field,
-
Copy and pasting from your file explorer,
-
Clicking the ‘Select Directory’ button, which will open a dialog box to allow the selection of a folder to save to.
-
Device Label
-
Set the name of the device being tested.
-
This is used to label the files for I-V data and measurement settings.
-
This field cannot contain the following characters: I. \ / : * ? “ < > |
Save Data Format -
All data is saved as .csv (comma separated value) files.
-
The figures below show the files that are created when saving data for each of the measurements.
Controls
Measure
-
Clicking this button will start the measurement using the chosen settings.
-
This button cannot be clicked if the software has not detected the test system.
Abort -
Stops a measurement that is currently in progress.
Plot Controls
Position Readout
Whilst the mouse cursor is over the plot in the Characterisation and Stabilised Current tabs, the x and y position of its location are displayed in the bottom-right of the plot,
Plot Display Controls
By default, the axes of the plot will automatically scale to display all the
data within it. The view can be controlled manually using the following mouse
controls:
- Left/Middle click and drag – pan the axes.
- Right click and drag – scale the axes (left-right for x-axis, up-down for y-axis).
- Scroll wheel – scale the axes centred on the cursor location.
A specific axis can be controlled by using these controls on the axis labels. The axes can be reset by clicking the ‘A’ button in the bottom-left of the plot
Test Devices
The system is shipped with a test device that can be used to check the
calibration of the system. They have resistors arranged in the geometry of the
substrate pixels, and the appearance of the test device will depend on the
substrate system being used
Taking a Measurement
- Plug in and switch on the system.
- Allow at least 30 minutes for the system to warm up.
- Place the test device in the device holder with the resistors facing up for S211 and S2006 and facing down for S241/S251.
- Start the Solar Cell I-V software and enter the following settings
- These settings can be used with any current range except for the 20 μA range. For this range the start and end voltages must be lowered to -2 V and 2 V respectively.
- The ‘Pixels to Measure’ checkboxes (Automated systems) or ‘Number of Pixels’ (Manual systems) should match the device configuration you have.
- Click the ‘Measure’ button.
- The system should measure straight line resistor responses from -100 μA.cm-2 to 100 μA.cm-2 (or -20 μA.cm-2 to 20 μA.cm-2 for the 20 μA current range) as shown in Figure 10.3.
- To check the calibration of the system, use the I-V data to calculate the measured resistance at -10 and 10 V (-2 and 2 V for the 20 μA current range).
- Resistance can be calculated using: R = V / I
- For the 200 mA current range the calculated resistance should be between 98 and 102 kΩ (within 2% of the resistor value).
- For all other ranges the calculated resistance should be between 99 and 101 kΩ (within 1% of the resistor value).
Troubleshooting
Most of the issues that may arise will be detailed here. However, if you
encounter any issues that aren’t in this list, please contact us by email at
info@ossila.com,
and we will respond as soon as possible.
Installation and Setup
Problem | Possible Cause | Action |
---|---|---|
No power | The power supply may not be connected properly. | Ensure the system |
is firmly plugged into the power supply, and that the plug is connected to
both the adaptor and a working power socket.
| The power supply adaptor has a fault.| Contact Ossila for a replacement
power supply adaptor.
Software does not start| The wrong version of Windows is installed on the
computer.| Install the software on a computer with Windows Vista or newer.
| The software has not installed properly.| Try reinstalling the software.
Cannot connect to the system via USB| The USB cable may not be connected
properly.| Ensure the USB cable is firmly plugged in at both ends.
| The USB cable may not be connected to a working USB port.| Try connecting
the unit to a different USB port on the computer.
| The USB cable is defective.| Try using a different USB-B cable, and contact
Ossila if necessary.
Cannot connect to the system via network| The MAC address of the unit is not
registered with the internal network.| Register the system on the network
using the MAC address obtained via a USB connection (see Source Measure Unit
manual).
| The Ethernet cable may not be connected properly.| Ensure the Ethernet cable
is firmly plugged in at both ends.
| The Ethernet cable is defective.| Try using a different Ethernet cable.
Error Messages and Warnings
Message | Description |
---|---|
Current compliance reached | The measured current is greater than the set |
current limit.
Error communicating with system| The software is unable to connect to the
system.
No device holder detected| The device holder is not connected to the system
properly.
Voltage increment cannot be zero| The voltage increment is set to 0 V.
Start and end voltage cannot be equal| The start and end voltages of are set
to the same voltage.
No save directory or device label entered| The save directory and/or the
device label fields are empty.
Settings profile not found| The given settings profile does not exist or is
open in other software.
Error loading settings| There is a problem with the settings profile
preventing it from being loaded.
Error deleting profile| The given settings profile does not exist, the
software does not have the necessary permissions to delete it, or it is open
in other software.
No data to save| There is no measurement data in memory to save to file.
Error saving data| The software does not have the necessary permissions to
access the given file path, or the file is already open in other software.
Error saving settings| The software does not have the necessary permissions to
access the given file path, or the file is already open in other software.
Error creating data directories| The software could not create the directories
for save data files.
Error creating data files| The software could not create the files for saving
measurement data.
Related Products
Related Consumables
ITO Coated Substrates
Our range of ITO substrates for OPV, OLED, and sensing applications.
Product codes: S111 / S101 / S211 / S281 / S171
FTO Coated Substrates
Designed to be used as transparent electrodes for thin-film
photovoltaics.
Product codes: S301 / S302 / S303 / S304
Flat Tip Tweezers
Provides a good substrate grip without scratching.
Product code: C121
Substrate Cleaning Rack
Holds 20 substrates for a variety of processing techniques.
Related Equipment
Spin Coater
Product high-quality coatings without any substrate warping. Perfect for busy
labs with limited space.
Product code: L2001A3
UV Ozone Cleaner
For removing contamination on the surface of samples, providing you with
ultraclean surfaces within minutes.
Product code: L2002A2
Syringe Pump
High-precision, programmable single and dual syringe pumps for automatic
dispensing of solutions.
Product codes: L2003S1 / L2003D1
Source Measure Unit
Source voltage, measure current, get data. Simplify and accelerate your data
collection!
Documents / Resources
|
Ossila T2002 Solar Cell I-V Test
System