Overview
This thermometer is a working radiation thermometer (hereinafter referred to
as a product) that can determine its surface temperature by measuring the
infrared energy radiated from the surface of the target body.
Warning
Before using the product, please read all safety instructions.
Table 1 Symbol
symbol | meaning | symbol | meaning |
---|---|---|---|
Danger. |
Important information.
See manual| ****| The product symbol is required by the WEEE Directive
(2002/96/EC).
The affixed label indicates that electrical/electronic products must not be
discarded as household waste. Product category:
According to the equipment type in the WEEE Directive Appendix 1, the product
is classified as a category 9 “monitoring and control instrument” product. Do
not dispose of this product as unsorted municipal waste.
Safety notice
Warnings indicate conditions and operations that are dangerous to the user. Be
careful Or the condition and operation of the equipment under test causing
damage.
The symbols used in the product and this manual are listed in Table 1
warning
To prevent eye damage or personal injury:
- Before using the product, please read the “Safety Instructions”.
- Do not use if the product works abnormally.
- Please use the product only for the designated purpose, otherwise the protection provided by the product may be weakened.
- Before using the product, check the case. If the product is damaged, do not use it.Check for cracks or missing plastic parts.
- Please refer to the emissivity information for the actual temperature. Reflecting objects will cause the measuredThe temperature is lower than the actual temperature. These objects pose a risk of burns.
- Do not use optical tools (such as binoculars, telescopes, microscopes, etc.) to look directly laser. Optical tools may focus the laser light and damage the eyes
- Do not look directly at the laser. Do not aim the laser directly at people or animals or from reflective surfaces Indirect exposure.
- Please replace the battery when the low battery indicator is displayed to prevent incorrect measurement.
- Do not use the product around explosive gas, steam or in a humid environment.
- Please use the product only in accordance with the relevant instructions, otherwise you may be exposed to dangerous laser shots.
Maintain
Note To avoid damage to the product, do not put the thermometer at high
temperature On or near objects.
How to replace the battery
To install or replace the 6F22 9V laminated battery, open the battery
compartment And replace the battery as shown on page 14
How to clean the product
Use a sponge or soft cloth moistened with soapy water to clean the product
shell. Wet with Wipe the surface carefully with a cotton swab.
The cotton swab can be moistened with water.
Standards and Agency Certification
Protection level ( IP) | according to IEC 60529, The protection level is IP 54 |
---|---|
Drop test | 3 m |
Vibration and shock | IEC 60068-2-6 2.5 g, 10 to 200 Hz, IEC 60068-2-27, 50 g, |
11 ms
Electromagnetic compatibility (EMC)| EN 61326-1:2006 EN 61326-2:2006
meets the ———– EN/IEC 61010-1:
Laser safety ———– EN 60825-1 2
Executive standard ———– Q/S20150116CN
Specification
temperature range | Please check the logo <400 °C | Please check the logo >400°C |
---|
Accuracy
Geometric correction ambient temperature
23°C±2°C| ≥0 °C: ±1.5 °C Or reading
±1.5 %, Take the larger value
(32 °F: +3 °F Or reading +1 5 % Take the larger value )
≥ -10 °C to <0 °C: ±2 °C
(≥14 °F to <32 °F: +4 °F) < -10 °C: ±3 °C
(<14 °F: ±6 °F)| ≥ °C: ±1 °C Or reading
±1 % , Take the larger value
(≥32 °F: ±2 °F Or reading± 1%, Take the larger value)
≥ -10 °C to <0 °C: ±2 °C (14 °F to <32 °F: ±4 °F) < -10 °C: ±3 °C
(<14 °F: ±6 °F)
Response time (95 %)| <500 ms (Reading 95 % )| <300 ms ( Reading95 % )
Spectral response| 8 to 14 Micron
Emissivity| 0.10 to 1.00
Temperature Coefficient| ±0.1 °C/ °C Or ±0.1% of reading / °C, whichever is
greater
Optical resolution| 12:1
(Energy is 95 °A Calculate when)
---|---
display resolution| 0.1°C (0.2°F)
Repeatability (Reading %)| Reading± 0.8 % or ± 1.0 °C (2 °F) Take the larger
value
power supply| 6F22 9V Laminated battery
Battery Life| 10 hours when the laser and backlight are on
weight| 169 g (With battery)
size| 95 x 43 x 155 mm (Length x width x height)
Operating temperature| 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature| -20 °C to 60 °C (-4 °F to 140 °F) , ( No battery)
Working humidity| 10 % to 90 % RH Non—condensing @ 30 °C (86 °F)
Working altitude| altitude 2000 m
Storage altitude| altitude 12000 m
Note: After the setting is completed, you need to pull the trigger to confirm, the instrument returns to the measurement interface, Pull the trigger this time to start measuring
- Open the battery cover and install 9V battery
- Hold the trigger and start measuring
COMMON OBJECT EMISSIVITY FROM (1)
Material Name| Specification| Emissivity| Material Name| Specification|
Emissivity
---|---|---|---|---|---
Aluminum| Oxidation| 0.20-0.40| Human skin| | 0.98
Polishing| 0.02-0.04| Graphite| Oxidation| 0.20-0.60
Copper| Oxidation| 0.40-0.80| Plastic| Transparency
0.5mm| 0.95
Polishing| 0.02-0.05
Gold| | 0.01-0.10| Rubber| | 0.95
Lron| Oxidation| 0.60-0.90| Plastic| | 0.85-0.95
Steel| Oxidation| 0.70-0.90| Concrete| | 0.95
Asbestos| | 0.95| Cement| | 0.96
Plaster| | 0.80-0.90| Soil| | 0.90-0.98
Asphalt| | 0.95| Putty| | 0.89-0.91
Pottery| | 0.95| Brick| | 0.93-0.96
COMMON OBJECT EMISSIVITY FROM (2)
Material Name| Specification| Emissivity| Material Name| Specification|
Emissivity
---|---|---|---|---|---
Wood| | 0.90-0.95| Marble| | 0.94
Charcoal| Powder| 0.96| Textile| | 0.90
Lacquerware| | 80- 0.950| Paper| Various Colors| 0.94
Mat| 0.97
Carbon plastic| | 0.90| Sand| | 0.90
Soap bubbles| | 0.75-0.80| Soil| | 0.92-0.96
Water| | 0.93| Graver| | 0.95
Snow| | 0.83-0.90| Glass| Tableware| 0.85-0.92
Ice| | 0.96-0.98| Textile| | 0.95