APPA 506 Digital Multimeter Bluetooth User Manual
- June 10, 2024
- APPA
Table of Contents
APPA 506 Digital Multimeter Bluetooth
Product Information
The 506/506B is a digital multimeter with Bluetooth connectivity. It comes
with a user manual and a limited warranty.
The meter is designed to measure voltage, frequency, resistance, continuity,
diode, capacitance, and temperature. It has a 40,000 count dual display and
true rms readings for accurate measurements of distorted sine waves and other
waveforms. The meter has various safety features such as CAT III and CAT IV
equipment designations to protect against transients from the primary supply
level. It also has push buttons for selecting measurement functions and
ranges, enabling/disabling high-frequency reject mode, auto-hold mode, and
Bluetooth mode. The meter can measure up to 1000V for AC and DC voltage.
Product Usage Instructions
Before using the 506/506B digital multimeter, read the safety information and operating instructions carefully. Use the meter only as specified in the user manual to avoid impairing the meter’s protection.
Basic Measurements
To measure AC and DC voltage:
- Turn on the meter by pressing the rotary switch.
- Select the voltage measurement function by pressing the yellow function button.
- Select the voltage range by pressing the range button or enter auto range mode by pressing the range button for more than one second.
- Connect the test leads to the input terminals for voltage measurements.
- Turn the rotary switch to select the desired voltage range.
- Read the measurement from the dual display.
Note: The meter can measure up to 1000V for AC and DC voltage.
To measure millivolts (mV), turn the rotary switch to mV position.
Push Buttons
The push buttons on the meter allow you to:
- Select measurement function
- Select measurement range
- Enable/disable high-frequency reject mode
- Enable/disable Bluetooth mode
- Enable/disable auto-hold mode
- Select peak-hold mode (MAX or MIN)
- Enter menu function in pointer position
- Cancel current menu function
Power On Options
When turning on the meter, you can execute the following options:
- Display the firmware version by pressing the function button.
- Display all LCD segments by pressing the A HOLD button.
Note: Follow the cautionary guidelines to avoid damaging the meter or equipment under test.
Read First
Safety Information
Understand and follow operating instructions carefully. Use the meter only as specified in this manual; otherwise,the protection provided by the meter may be impaired.
WARNING
This identifies hazardous conditions and actions that could cause BODILY HARM
or DEATH. To avoid possible danger, follow below guidelines.
-
Use the meter only as specified in this manual or the protection by the meter might be impaired.
-
Never operate the meter with the cover removed or the case open.
-
To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator.
-
Use caution with voltages above 30VAC rms, 42VAC peak, or ±30VDC. These voltages pose a shock hazard.
-
When using test leads or probes, keep your fingers behind the finger guards.
-
Remove test lead from meter before opening the battery door or meter case.
-
Always use proper terminals, switch position, and range for measurements.
-
Do not apply more than the rated voltage, as marked on meter, between terminals or between any terminal and earth ground.
-
Do not use the High-Frequency Rejection (Low Pass Filter) option to verify the presence of hazardous voltages.
Voltages greater than what is indicated may be present. First, make a voltage measurement without the filter to detect the possible presence of hazardous voltage. Then select the filter function. -
To avoid possible electric shock or personal injury, never attempt an in-circuit current measurement where the open circuit potential to earth is greater than 1000V.
-
Replace the fuse as soon as the indicator (FUSE) appears.
-
Only replace the blown fuse with the proper rating as specified in this manual.
-
Do not use the meter around explosive gas, vapor or dust.
-
To reduce the risk of fire or electric shock do not expose this product to rain or moisture.
-
Do not attempt a current measurement when the open voltage is above the fuse protection rating. Suspected open voltage can be checked with voltage function.
-
Never attempt a voltage measurement with the test lead inserted into the A input terminal.
CAUTION
This identifies conditions and actions that could DAMAGE the meter or
equipment under test. To avoid possible damage, follow below guidelines.
-
Disconnect the test leads from the test points before changing the position of the function rotary switch.
-
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
-
Always use proper terminals, switch position, and range for measurements.
-
Do not use the LoZ mode to measure voltages in circuits that could be damaged by this mode’s low impedance.
-
Replace the fuse as soon as the indicator (FUSE) appears.
-
Never connect a source of voltage with the function rotary
switch in resistance, diode, continuity, and capacitance position. -
Never set the meter in current function to measure the voltage of a power supply circuit in equipment that could result in damage the meter and the equipment under test.
Electrical Symbols
Error Message
Error Message
ProbE
| Test probe alert. Displayed when the test probes are in the A or mA
terminal and the selected rotary switch position does not correspond to the
terminal being used.
FUSE| Fuse broke. Replace fuse as soon.
Er| Meter error. Have meter serviced.AC (Alternating Current)
The Meter Description
Front panel illustration :
- Auto backlight sense point
- 40,000 count dual display
- Push buttons
- Rotary switch
- Input terminal for voltage, frequency, resistance,continuity,diode, capacitance and temperature measurements.
- Return terminal for all measurements.
- Input terminal for 0 to 400mA current measurements.
- Input terminal for 0 to 10A current measurements.
Push Buttons
Push Buttons
Function (Yellow)| Select measurement function.
RANGE| Select measurement range. Press > 1 sec to enter auto range mode.
HFR| Enable/Disable the High Frequency Reject mode in the AC
measurements.
Bluetooth
|
Enable/Disable the Bluetooth mode.
A- HOLD| Enable/Disable the Auto-Hold mode.
P-HOLD
| Enable the Peak-Hold mode in the AC or DC measurements. In this mode, press
button to select Peak-Hold MAX or MIN. Press > 1 sec to disable the Peak-Hold
mode.
ENTER| Enter menu function in pointer positon.
CANCEL| Cancel current menu function.
°C / °F| Select degrees Celsius or degrees Fahrenheit.
Power On Options
When turn the power on, press the function button to execute the below
options.
Power On Options
ENTER| Display the firmware version.
A HOLD| Display all LCD segments.
Making Basic Measurements
WARNING
- When connecting the test leads to the DUT (Device Under Test), connect the common test lead before connecting the live lead. When removing the test leads, remove the test live lead before removing the common test lead.
The following sections describe how to take measurements with the meter.
Measuring AC and DC Voltage
This meter have true rms readings, which are accurate for distorted sine waves and other waveforms (with no dc offset) such as square waves, triangle waves, and staircase waves. The ranges of measuring voltage are 40mV, 400mV, 4V, 40V, 400V and 1000V. To select the mV range, turn the rotary switch to mV position.
For best accuracy when measuring the DCmV, touch the probe tips together and read the DC offset. If necessary, you can use the relative () mode to automatically subtract this value.
Make dB measurement
The meter is capable of displaying voltage as a dB value, either relative to 1
milliwatt (dBm), a reference voltage of 1 volt (dB). A dBm measurement must
use a reference impedance (600) to calculate a dB value based on 1 milliwatt.
A dB measurement uses a 1 volt reference voltage to compare the present
measurement against. Define as below :
To use dB or dBm function, turn the rotary switch to ACV or ACmV position. Then move the blink cursor of menu to dB or dBm position, and press the ENTER button to enter function. Press the CANCEL button to exit function.
Measuring Voltage in LoZ Mode
Caution
- Do not use the LoZ mode to measure voltages in circuits that could be damaged by this mode’s low impedance.
To eliminate ghost voltages, the meter’s LoZ mode presents a low impedance across the leads to obtain a more accurate measurement. The ranges of measuring LoZ voltage are 400V and 1000V. In this mode, meter will automatic measure input signal which is AC or DC and determine range. To use the LoZ mode, turn the rotary switch to LoZ position.
Measuring AC and DC Current
WARNING
- Always use proper terminals, switch position, and range for measurements.
- To avoid possible electric shock or personal injury, never attempt an in-circuit current measurement where the open circuit potential to earth is greater than 1000V.
- Only replace the blown fuse with the proper rating as specified in this manual.
Caution
- Replace the fuse as soon as the indicator (FUSE) appears.
To measure current, you must break the circuit under test, thenplace the meter
in series with the circuit.
The ranges of measuring current are 40mA, 400mA, 4A and 10A. AC current is
displayed as an rms value. Insert the black lead into the COM terminal. For
currents less than 400 mA, insert the red lead into the mA terminal. For
currents above 400 mA, insert the red lead into the A terminal.
You can press the function (yellow) button to select current measurement function.
Measuring Frequency
The meter measures the frequency of a voltage or current signal by counting
the number of times the signal crosses a threshold level each second. This
function only can be operated in AC voltage and current measurements. The
ranges of measuring frequency are 400Hz,4kHz, 40kHz and 100kHz. If a reading
shows as 0 Hz or is unstable, the input signal may be below or near the
trigger level. The detail of frequency trigger level refer to the electrical
specifications. To use the frequency function, press function (yellow) button
to select measurement function.
Make High Frequency Rejection Measurement
WARNING
- Do not use the High Frequency Rejection (Low Pass Filter) option to verify the presence of hazardous voltages. Voltages greater than what is indicated may be present. First, make a voltage measurement without the filter to detect the possible presence of hazardous voltage. Then select the filter function.
The High Frequency Rejection mode equip a low pass filter in the AC measurements. The cut-off frequency (-3dB point) of low pass filter is 800Hz. To use the HFR mode, press the HFR button to equip a low pass filter in the AC measurements.
Make Peak-Hold Measurement
The Peak-Hold mode records wave peak maximum and minimum input values.
Response time of Peak-Hold is 10us.
To use the Peak-Hold mode, press the P-HOLD button to enable the Peak-Hold mode in the AC or DC measurements. In this mode, press P-HOLD button to select peak MAX or MIN value. Press > 1 sec to disable the Peak-Hold mode. For square wave, use the peak-hold mode in the DC measurements.
Make AC+DC Measurement
When input signal is ac and dc combinations: AC over DC or DC over AC, the
meter is capable of displaying one AC+DC (rms) value combined. Define as below
:
To use the AC+DC function, turn rotary switch stop in DCV, DCmV or A position, then press function (yellow) button to select measurement function.
Measuring Resistance
CAUTION
- To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all high voltage capacitors before measuring resistance.
The ranges of measuring resistance are 400, 4k, 40k, 400k, 4M, and 40M. The
test leads may be add 0.1 to 0.2 of error to resistance measurements. To test
the leads, touch the probe tips together and read the resistance of the leads.
For best accuracy, you can use the relative () mode to automatically subtract
this value.
High-resistance (>10M) readings are susceptible to electrical noise. To smooth
out most noisy readings, enter the MAX/MIN recording mode; then step to the
average (AVG) reading.
Continuity Check
CAUTION
- To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all high voltage capacitors before testing continuity.
The continuity check features a buzzer that sounds as long as a circuit is
complete. The buzzer allows you to quick continuity checks without watching
the display.
When measuring resistance is less than threshold, the buzzer sounds. You can
setup the threshold in setup mode. The continuity threshold is default 30.
To use continuity check, turn the rotary switch to resistance position, then
press the function (yellow) button to select measuring mode.
Testing Diodes
CAUTION
- To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all highvoltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, silicon controlled rectifiers (SCRs), and other semiconductor devices. This function tests a semiconductor junction by sending a current through the junction, then measuring the junction’s voltage drop. A good silicon junction drops between 0.5V and 0.8V. For forward-bias readings on any semiconductor component, place the red test lead on the component’s positive terminal and place the black lead on the component’s negative terminal. In a circuit, a good diode should still produce a forward-bias reading of 0.5V to 0.8V. For reverse-bias readings on any semiconductor component, the meter still can measure. In a circuit, a good diode should still produce a reverse-bias reading of -0.5V to -0.8V. The display shows “OL” if the diode is open or short. To use diode test, turn the rotary switch to resistance position, then press the function (yellow) button to select measuring mode.
Measuring Capacitance
CAUTION
- To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all highvoltage capacitors before measuring capacitance. Use the dc voltage function to confirm that the capacitor is discharged.
The ranges of measuring capacitance are 40nF, 400nF, 4uF, 40uF, 400uF, 4mF and 40mF. To improve the accuracy of measurements less than 1000nF, you can use the relative () mode to subtract the residual capacitance of the leads. To use capacitance measurement, turn the rotary switch to resistance position, then press the function (yellow) button to select measuring mode.
Measuring Temperature
The meter measures the temperature of a K-Type thermocouple. You can press the function (yellow) button to choose degrees Celsius (°C) or degrees Fahrenheit (°F). Display ranges are -200°C to +1200°C and -328°F to +2192°F. Readings outside of these ranges show “OL” on the display. When there is no thermocouple connected, the display also shows “OL”. To use temperature measurement, turn the rotary switch to temperature position, then press the function (yellow) button to select measuring mode.
Auto-Hold
When measuring, you can press the A-HOLD button to start the Auto-Hold mode.
In this mode, the meter holds reading and shows it on the secondary display.
If the difference between new reading and hold reading is bigger than 5d (3¾-
digit mode), and new reading is also stable, then meter automatically holds a
new reading on the secondary display. When reading is smaller than Auto-Hold
limit, or reading is OL, the Auto-Hold mode is not working.
Function | Limit |
---|---|
V, A, LoZ, Hz, Cap | 1% of range |
Others | No limit |
To exit Auto-Hold mode, press the A-HOLD button again. If you don’t want to use the Auto-Hold mode, you can disable it in the setup mode. When Auto-Hold mode is disable, the hold mode is not update any new reading.
Maximum / Minimum Record
When measuring, you can record the maximum, minimum and average value of
reading.
To use maximum/minimum record mode, move the blink cursor of menu to MAX, MIN
or AVG position, and press the ENTER button to enter mode. In this mode, the
meter records each data to compare the maximum and minimum value. Also, meter
calculate the average of reading. You can move the blink cursor of menu to
MAX, MIN or AVG position, and press the ENTER button to select result on the
secondary display.
When maximum/minimum record mode running, if you wants to pause recording,
press the A-HOLD button. Press again to continue.
To exit maximum/minimum record mode, press the CANCEL button.
Relative
When measuring, you can use the relative () mode to subtract the offset. To
use relative () mode, move the blink cursor of menu to position, and press the
ENTER button to enter mode. In this mode, meter records the presently reading
as reference and shows it on the secondary display. The relative () mode
subtract reference from each reading, and shows result on the main display. In
relative () mode, the minimum range is relative () range. For example: The
ranges of measuring resistance are 400, 4k, 40k, 400k, 4M, and 40M. If you use
the relative () mode in 4k range, then the minimum range is the 4k range. You
can use range is 4k to 40M, cannot change to 400. To exit relative () mode,
press the CANCEL button.
Relative %
When measuring, you can use the relative (%) mode to calculate the relative
percent value. The relative percent value is define as below:
To use relative (%) mode, move the blink cursor of menu to % position, and
press the ENTER button to enter mode. In this mode, meter records the
presently reading as reference and shows it on the secondary display. The
relative (%) mode calculate the relative percent value from each reading, and
shows result on the main display.
To exit relative (%) mode, press the CANCEL button.
Memory Save / Load
When measuring, you can save the reading to memory and load it from memory.
The meter can store maximum 1000 data in memory. The recorded data amount
shows on the secondary display.
To use memory save / load mode, move the blink cursor of menu to MEM position,
and press the ENTER button to enter mode. In this mode, you can operate the
below options:
Memory Options
Memory Options
A- SAVE
|
You can operate the Auto-Save mode to automati-cally save new reading. When you use the probes to measure a new reading, the meter will automati-cally save it. In some case, the Auto-Save mode will not work. For example, the reading is smaller than the limit (refer Auto-Hold), or the reading is OL. Press ENTER button to start Auto-Save mode, press CANCEL button to exit.
SAVE| Press ENTER button to save a new reading to memory.
LOAD
| You can press the ENTER button to review data from memory. Press UP or DOWN
button to select data. Press the CANCEL button to return.
CLR| Press ENTER button to clear all data from memory.
MAX| Press ENTER button to review the maximum data from memory.
MIN| Press ENTER button to review the minimum data from memory.
To exit memory save / load mode, press the CANCEL button.
Data Logger
You can record a lot of reading to memory in a long time, then analyze and
plot graph. The meter can store maximum 40,000 data in memory. The recorded
data amount shows on the secondary display.
The record rate can be set from 1 sec and 600 sec. The error of timer is less
than 3 seconds per hour.
To use data logger, move the blink cursor of menu to LOG position, and press
the ENTER button to enter mode. In this mode, you can operate the below
options :
Data Logger Options
SAVE
| Press ENTER button to start data logger. The logger automatically records at regular intervals. You can press ENTER button to pause data
logger, press again to continue. To stop data logger, press CANCEL button to return.
LOAD
| You can press the ENTER button to review data from memory. Press UP or DOWN
button to select data. Press the CANCEL button to return.
CLR| Press ENTER button to clear all data from memory.
RATE
| You can setup the record rate of logger. Press UP or DOWN button to select
rate. Press the CANCEL button to return.
MAX| Press ENTER button to review the maximum data from memory.
MIN| Press ENTER button to review the minimum data from memory.
To exit data logger, press the CANCEL button.
Auto Power Off
If you don’t operate the rotary switch or buttons for a specified time, the
meter will turn off automatically to save the power of batteries. The default
APO timer is 10 minutes. In setup mode, you can change the APO timer.
Backlight
The backlight is automatically turned on at dark environment. The backlight
mode is default auto. In setup mode, you can set the backlight mode.
Buzzer
The meter equip a 2kHz tone buzzer. Valid button press: Beep once. And invalid
button press: Beep twice. In setup mode, you can turn on or off the buzzer.
But the buzzer in continuity check cannot be turn off.
Display Resolution
This meter have two display resolution: normal resolution (3¾-digit mode) and
high resolution (4¾-digit mode). The normal resolution is set to default. You
can setup the resolution in setup mode.
Setup
To use meter setup mode, move the blink cursor of menu to SETUP position, and
press the ENTER button to enter function. Press the UP or DOWN button to
select item; press the LEFT or RIGHT to select options. When you complete
setup, press the CANCEL button to exit setup mode.
Setup Options
APO| 1min to 30min, or OFF
b.Lit| Backlight mode: Auto, ON or OFF
bEEP| Buzzer ON or OFF
A.Hold| Auto-Hold mode ON or OFF
Cntin| Continuity threshold: 10Ω to 50Ω
diGit| Display digit: Lo or Hi
TEMP| Temperature default unit: °C or °F
RESET| Press ENTER to reset all setup options.
IR Communications
You can use the IR (infrared) communication link and WinDMM software to
transfer the real-time data of meter to PC. In addition, the meter allows the
user to log to internal memory and connect to the computer later for download.
For detailed information, refer to the WinDMM Installation Guide or the on-
line help.
Bluetooth (506B only)
The meter uses low-power Bluetooth v4.0 wireless technology to transfer the
real-time data. You can use the RF communi-cation to link to android or iOS
devices. The RF communication range: Open air up to 10 m
Maintenance
Verify the meter’s operation by measuring a known voltage. If in doubt, have
the meter serviced. Do not attempt to repair this meter. It contains no user
serviceable parts. Repair or servicing should only be performed by qualified
personnel. To maintain best accuracy, calibrate meter once a year.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use
abrasives or solvents.
Replace Batteries & Fuse
WARNING
- Remove test lead from meter before opening the battery door or meter case.
- Replace the fuse as soon as the indicator (FUSE) appears. · Only replace the blown fuse with the proper rating as specified in this manual.
- Battery Type : 4 x 1.5V IEC LR6 or size AA
- Fuse1 Type : 440mA, 1000V IR 10kA Fuse
- Fuse2 Type : 11A, 1000V IR 20kA Fuse When the battery low indication shows on the display, replace the batteries soon. To save the power of batteries, you can disable the backlight and buzzer in setup mode.
General Specifications
- MAX Voltage between any Terminal and Earth Ground: 1000Vrms
- Fuse Protection for mA inputs: 440mA, 1000V IR 10kA
- Fuse Fuse Protection for A inputs: 11A, 1000V IR 20kA Fuse
- Display: 4,000/40,000 counts, over range to 110%.
- Over Range Indication: OL
- Measuring Rate: 10 samples per second
- Power Requirements: 4 x 1.5V IEC LR6 or size AA
- Battery Life: 50 hours typical with alkaline (with backlight off).
- Operating Ambient : -10°C to 30°C (< 85% RH),30°C to 40°C (< 75% RH), 40°C to 50°C (< 45% RH)
- Storage Temperature: -20°C to 60°C, 0% RH to 80% RH (batteries not fitted)
- Temperature Coefficient : 0.1 x (Specified Accuracy) / °C, < 18°C or > 28°C
- Operating Altitude: 6561.7ft (2000m)
- Calibration Cycle: 1 time per year.
- Weight: 465g including battery.
- Dimensions (H x W x L): 52 x 83 x 188 (mm) with holster.
- RF Communications : 2.4 GHz ISM Band, open air 10m
- Safety: Complies with EN 61010-1 CAT IV 600V, CAT III 1000V
CAT | Application Field |
---|---|
I | The circuits not connected to mains. |
II | The circuits directly connected to Low-voltage installation. |
III | The building installation. |
IV | The source of the Low-voltage installation. |
EMC: EN 61326-1 Pollution Degree : 2 Shock Vibration : Per MIL-PRF-28800F for
a Class 2 instrument Drop Protection : 5ft (1.5m)
Indoor Use
Electrical Specifications
- Accuracy is ± (% of reading + number of digits) at 18°C to 28°C (< 80% RH)
- For specifications in the 4¾-digit mode, multiply the number of digits by 10.
- For the best measurements, with relative () mode to compensate for offsets.
Voltage
Function | Range | Accuracy |
---|
AC
| 40.00mV [1]
400.0mV [1]
4.000V
40.00V
400.0V [1]
1000V [2]
| Sine Wave: 0.5%+2d for 40Hz to 70Hz [3] 1.5%+4d for 70Hz to 1kHz [3] 3.0%+4d for 1kHz to 5kHz [3]
5.0%+20d for 5kHz to 100kHz [4] [5]
DC
| 40.00mV| 0.03%+3d
400.0mV
4.000V
40.00V
400.0V
1000V
| ****
0.03%+1d
AutoV LoZ| 400.0V
1000V
| 2.0%+4d for ACV 40Hz to 1kHz 2.0%+4d for DCV
[1] The bandwidth is 40Hz to 5kHz [2] The bandwidth is 40Hz to 1kHz [3] Below
10% of range, add 2d to accuracy. [4] Below 10% of range, add 10d to accuracy.
< 50kHz.
Input Impedance : 10MΩ, < 100pF
LoZ Input Impedance: 3kΩ
Bandwidth : 40Hz to 100kHz
Minimum Resolution: 10uV
CMRR / NMRR (Common / Normal Mode Rejection Ratio) :
VAC : CMRR > 60dB at DC, 50Hz / 60Hz
VDC : CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
Current
Function | Range | Accuracy |
---|
AC
| 40.00mA
400.0mA
4.000A [1]
10.00A [1][2]
| Sine Wave: 0.8%+2d for 40Hz to 70Hz [3] 2.0%+4d for 70Hz to 1kHz [3]
2.0%+4d for 1kHz to 10kHz [4]
DC
| 40.00mA
400.0mA
4.000A
| ****
0.2%+1d
10.00A [2]| 0.2%+2d
[1] The bandwidth is 40Hz to 1kHz [2] When > 10A, accuracy is
unspecified and maximum
measuring time is 30 sec.
[3] Below 10% of range, add 2d to accuracy. [4] Below 10% of range, add 10d to accuracy.
Input Impedance: < 2Ω at mA inputs, < 0.1Ω at A inputs.
Bandwidth : 40Hz to 10kHz
Minimum Resolution: 10uA
Maximum Measuring Time :
1 minutes at A inputs, 10 minutes at mA inputs.
Rest time is 20 minutes minimum.
AC and DC Additional Specifications
Function | Range | Accuracy |
---|---|---|
AC+DC | **** |
Same as V & A
| AC accuracy + 1.0%
HFR| AC accuracy + 1.0% for 40Hz to 400Hz
Peak-Hold| 3.0%+200d for 40Hz to 1kHz [1]
[1] For square wave, the accuracy is unspecified.
The Cut-Off Frequency of HFR: 800Hz (-3dB point)
Attenuation Characteristic of HFR: Approx. -24dB
AC Conversion Type:
The AC conversion type is ac-coupled, true rms responding,calibrated to the
sine wave input. For non-sine wave add the following Crest Factor corrections:
- For Crest Factor of 1.4 to 2.0, add 1.0% to AC accuracy.
- For Crest Factor of 2.0 to 2.5, add 2.5% to AC accuracy.
- For Crest Factor of 2.5 to 3.0, add 4.0% to AC accuracy.
Frequency Counter
Function | Range | Accuracy |
---|---|---|
400.0Hz | 0.1Hz | 1d (3¾-digit mode) |
5d (4¾-digit mode)
4.000kHz| 1Hz
40.00kHz| 10Hz
100.0kHz| 100Hz
Minimum Sensed Frequency : 5Hz
Frequency Counter Sensitivity
Function
| ****
Range
| Sensitivity (Peak to Peak)
---|---|---
5 to 10k Hz| 10k to 100k Hz
mV
| 40.00mV| 10mV| 10mV
400.0mV| 40mV| 100mV
V
| 4.000V| 0.4V| 1V
40.00V| 4V| 10V
400.0V| 40V| ****
Unspecified
1000V| 400V
mV
| 40.00mA| 10mA| ****
Unspecified
400.mA| 40mA
A
| 4.000A| 1A
10.00A| 4A
Resistance
Function | Resolution | Accuracy |
---|---|---|
400.0Ω | 0.1Ω | 0.2%+2d |
4.000kΩ | 1Ω | **** |
0.2%+1d
40.00kΩ| 10Ω
400.0kΩ| 100Ω
4.000MΩ| 1kΩ| 1.0%+1d
40.00MΩ| 10kΩ| 2.0%+20d
- Maximum Open Circuit Voltage: Approx. 2.5V
- Maximum Short Test Current: Approx. 0.1mA
Continuity Check
Function | Resolution | Accuracy |
---|---|---|
400.0Ω | 0.1Ω | 0.2%+2d |
- Maximum Open Circuit Voltage: Approx. 2.5V Maximum Short Test Current : Approx. 0.1mA
- Continuity Threshold: Adjustable 10 to 50 , default 30. Continuity Indicator : 2kHz Tone Buzzer
Testing Diodes
Function | Resolution | Accuracy |
---|---|---|
400.0Ω | 0.1Ω | 1.5%+2d |
- Maximum Open Circuit Voltage : Approx. ±2.5V
- Maximum Short Test Current : Approx. ±1mA
Capacitance
Function | Resolution | Measuring Time | Accuracy |
---|---|---|---|
40.00nF | 10pF | 1 sec | 0.9%+20d |
400.0nF | 100pF | 1 sec | 0.9%+10d |
4.000uF | 1nF | 1 sec | **** |
0.9%+2d
40.00uF| 10nF| 1 sec
400.0uF| 100nF| 1 sec
4.000mF| 1uF| 4 sec| 0.9%+10d
400.0uF| 10uF| 8 sec| 0.9%+20d
Temperature
Function | Resolution | Accuracy |
---|
-200°C
to
+1200°C
| ****
0.1°C
| ****
1.0%+30d
-328°F
to
+2192°F
| ****
0.1°F
| ****
1.0%+54d
[1] Does not include error of the thermocouple probe. [2] Accuracy specification assumes ambient temperature
stable to ± 1°C. For ambient temperature changes of
± 2°C, rated accuracy applies after 1 hour.
Limited Warranty
This Meter is warranted to the original purchaser against defects in material and workmanship for 3 year from the date of purchase. During this warranty period, manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.
APAC
MGL APPA Corporation
cs.apac@mgl-intl.com
Flat 4-1, 4/F, No. 35,
Section 3 Minquan East Road,
Taipei, Taiwan
Tel: +886 2-2508-0877
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>