Truenat 601430005 COVID-19 Packinsert User Guide

June 10, 2024
Truenat

Truenat-LOGO

Truenat 601430005 COVID-19 Packinsert

Truenat-601430005-COVID-19-Packinsert-PRODUCT

Product Information

The product is a Chip-based Real Time Duplex PCR Test for COVID-19. The kit is designed to detect the E and Orf1a gene and human RNase P as a full process internal positive control (IPC) for proper swab collection, nucleic acid extraction, and PCR. The product is available in different quantities, including 20T, 25T, 100T, and 200 T.

  • Target Sequence: E and Orf1a gene and human RNase P
  • Full Process Internal Positive Control: Human RNase P gene

The product is intended for professional use only and requires protective clothing and disposable gloves when handling samples and performing sample extraction.

Product Usage Instructions

  1. Put on protective clothing and disposable gloves before handling samples.
  2. Collect the swab sample from the patient’s nose, throat, or both as per the standard procedure.
  3. Place the swab into the provided tube containing the sample collection buffer.
  4. Close the tube tightly with the cap and mix the contents well by vortexing or pipetting up and down several times.
  5. Extract the nucleic acid from the sample using the recommended method.
  6. Prepare the PCR reaction mix as per the kit instructions.
  7. Add the extracted nucleic acid to the PCR reaction mix and run the PCR reaction on a Real-Time PCR instrument.
  8. Interpret the results as per the instrument’s manual and the kit instructions.
  9. Dispose of the used swab, tubes, and gloves properly as per the biohazardous waste disposal guidelines.

INTENDED USE

  • Truenat® COVID-19 (REF 601430005 / 601430020 / 601430025 / 601430050 / 601430100 / 601430200) is a Chip-based Real Time Duplex Reverse Transcription Polymerase Chain Reaction (RT PCR) test for the semi-quantitative detection of SARS CoV-2 RNA in human oropharyngeal and nasopharyngeal swab specimen and aids in detection and confirmation of SARS CoV-2 infection and diagnosis of COVID-19. The test detects the E and Orf1a genes of the virus. Truenat® COVID-19 runs on
  • Truelab® Real Time Quantitative micro PCR Analyzers. Truenat® COVID-19 is an in vitro diagnostics test meant for professional use only.

INTRODUCTION

  • SARS CoV 2 is the causative agent for corona virus disease 2019 or COVID-19 in Humans. SARS CoV 2 is a Beta Corona Virus, one of the four genera of Corona Viruses.

  • Coronaviruses are enveloped non-segmented positive sense RNA viruses belonging to the family coronaviridae and the order Nidovirales and broadly distributed in humans and other mammals. The common signs of COVID-19 infection include respiratory symptoms, fever, cough, shortness of breath and breathing difficulties. In more severe cases, infection can cause pneumonia, severe acute respiratory syndrome, kidney failure and even death.
    Early and correct identification of infection with SARS CoV-2 is important for effective isolation, treatment and case management of COVID-19. In line with WHO recommendations, molecular diagnostics are currently the method of choice for such virus detection and differentiation. However, molecular tests for COVID-19 have so far been restricted to centralized reference laboratories as they require skilled manpower and elaborate infrastructure. Also the turnaround time for results could take a few days.Truenat-601430005-COVID-19
-Packinsert-FIG-1

  • The Truelab® Real Time micro PCR System enables decentralization and near patient diagnosis of and monitoring of COVID-19. This is enabled by making the real time PCR technology rapid, simple, robust and user friendly, thereby offering “sample to result” capability even at resource limited settings. This is achieved through a combination of lightweight, portable, mains / battery operated Truelab® Real Time micro PCR Analyzers and Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Device and room temperature stable Truenat® micro PCR chips and Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Kits so that even the peripheral laboratories with minimal infrastructure and minimally trained technician can easily perform these tests routinely in their facilities and report PCR results in less than an hour. Moreover, with these devices PCR testing can also be initiated in the field level, on site. Truenat® COVID-19 is a disposable, room temperature stable, Chip-based Real Time Duplex PCR test with dried MgCl in reaction well and 2 freeze dried RT PCR reagents in microtube for performing Real

  • Time RT-PCR test for viral infection and runs on the Truelab® Real

  • Time micro PCR Analyzer. It requires only six (6) µL of purified

  • RNA to be added to the reaction well for the analysis. The intelligent chip also carries test and batch related information. The Truenat® COVID-19 chip also stores information of used test to prevent any accidental re-use of the chip.

  • NOTE : Truelab®/ Truenat® / Trueprep® / Truepet® / STABILYSE® are all trademarks of Molbio Diagnostics Private Limited.

  • The Truelab® Real Time micro PCR Analyzer is protected by the following patents and patents granted: IN 2313/CHE/2007 (Patent No. 281573), WO2009/047804 and corresponding claims of any foreign counterpart(s) thereof.

  • The Truenat® micro PCR chip is protected by the following patents and patents pending: IN 2312/CHE/2007, WO 2009/047805 and corresponding claims of any foreign counterpart(s) thereof.

PRINCIPLE OF THE TEST

  • Truenat® COVID-19 works on the principle of Real Time Reverse Transcription Polymerase Chain Reaction (RT PCR) based on Taqman chemistry. The RNA from the patient sample is first extracted using Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Device and Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Kit. The Truenat® COVID-19 chip is placed on the chip tray of the Truelab® Real Time micro PCR Analyzer. Six (6) µL of the purified RNA is then dispensed using the provided micropipette and tip into the microtube containing freeze dried PCR reagents, including reverse transcriptase (RT) and allowed to stand for 30-60 seconds to get a clear solution. No mixing by tapping, shaking or by reverse pipetting should be done. Six (6) µL of this clear solution is then pipetted out using the same pipette and tip and dispensed into the reaction well of the Truenat® COVID-19 chip and the test is inserted in the Truelab® Real Time Quantitative micro PCR
  • Analyzer where the RNA is first converted into complementary DNA (cDNA) by the RT enzyme and further thermal cycling takes place. A positive amplification causes the dual labeled fluorescent probe in the Truenat® COVID-19 Chip-based Real Time PCR test to release the fluorophores in an exponential manner which is then captured by the built-in opto-electronic sensor and displayed as amplification curve on the analyzer screen, on a real time basis during the test run. The Cycle threshold (Ct) is defined as the number of amplification cycles required for the fluorescent signal to cross the threshold (i.e. exceed the background signal). Ct levels are inversely proportional to the amount of target nucleic acid in the sample (i.e. the lower the Ct level the greater is the amount of target nucleic acid in the sample). In the case of negative samples, amplification does not occur and a horizontal amplification curve is displayed on the screen during the test run.At the end of the test run, E and Orf1a gene “DETECTED” or “NOT DETECTED” result is displayed and in positive cases, semi-quantitative result is also displayed on the screen. Based on the detection of the internal positive control (IPC), the validity of the test run is also displayed. The IPC is a full process control that undergoes all the processes the specimen undergoes – from extraction to amplification thereby validating the test run from sample to result. Absence of or shift of IPC Ct beyond a pre-set range in case of negative samples invalidates the test run. While IPC will co-amplify in most positive cases also, in some specimen having a high target load, the IPC may not amplify, however the test run is still considered valid. The results can be printed using the Truelab® micro PCR printer or transferred to the lab computer /or any remote computer via Wifi network or 3G/GPRS network. Upto 20,000 results in Truelab® Uno Dx/Duo/Quattro can be stored on the analyzer for future recall and reference.
TARGET SELECTION
  • The target sequence for this kit is E and Orf1a gene and human RNase P. Detection of the human RNase P gene serves as a full process internal positive control (IPC) for proper swab collection, nucleic acid extraction and PCR.

CONTENTS OF THE Truenat® COVID-19 KIT

A. Individually sealed pouches, each containing a

  1. Truenat® COVID-19 micro PCR chip.
  2. Microtube with freeze dried RT PCR reagents.
  3. DNase & RNase free pipette tip.
  4. Desiccant pouch.

B. Package Insert.

| 601430005| 601430020| 601430025| 601430050| 601430100| 601430200
---|---|---|---|---|---|---
| 5T| 20T| 25T|  | 100T| 200T

CONTENTS OF THE Trueprep® AUTO Universal Sample Pre-treatment Pack (for Extraction with Trueprep® AUTO/AUTO v2)

  • A. Lysis Buffer (contains lysis cum transport medium).
  • B. Disposable transfer pipette (graduated).
  • C. Package Insert.

| 60205AB05| 60205AB20| 60205AB25| 60205AB50| 60205AB100| 60205AB200
---|---|---|---|---|---|---
| 5T| 20T| 25T|  | 100T| 200T

CONTENTS OF THE Trueprep® AUTO Transport Medium for Swab Specimen Pack (for Extraction with Trueprep® AUTO/AUTO v2)

  • A. Transport Medium for Swab specimen tubes (contains transport medium).
  • B. Package Insert.

| 60206TS05| 60206TS20| 60206TS25| 60206TS50| 60206TS100| 60206TS200
---|---|---|---|---|---|---
| 5T| 20T| 25T|  | 100T| 200T

CONTENTS OF THE STABILYSE® Prep Free pack (for Rapid PCR Protocol)

  • A. Collection and Lysis Medium Tube
  • B. Package Insert.

| 90101PF05| 90101PF20| 90101PF25| 90101PF50| 90101PF100| 90101PF200
---|---|---|---|---|---|---
| 5T| 20T| 25T|  | 100T| 200T

STORAGE AND STABILITY

Truenat® COVID-19 is stable for two (2) years from the date of manufacture if stored between 2-30oC. It is also stable for upto one (1) month at temperatures up to 45°C. Avoid exposure to light or elevated temperatures (above recommended levels). Do not freeze.
Trueprep® AUTO Universal Sample Pre-Treatment Pack, Trueprep® AUTO Transport Medium for Swab Specimen Pack and STABILYSE® Prep Free Pack are stable for two (2) years from the date of manufacture if stored between 2-40oC. It is also stable for one (1) month at temperatures upto 45oC. Do not freeze.

MATERIALS REQUIRED BUT NOT PROVIDED WITH THE KIT

Truelab® Real Time micro PCR Workstation (REF 623010001 / 633010001 / 643010001 / 653010001) consisting of

  1. Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Device (REF603041001/603042001).
  2. Truelab® Uno Dx/Truelab® Duo/Truelab® Quattro Real Time micro PCR Analyzer (REF603021001/603022001/603023001).
  3. Truelab® micro PCR Printer (REF 603050001).
  4. Truepet® SPA fixed volume precision micropipette – 6 µl (REF 604070006).
  5. Truelab® Microtube Stand (REF 603070001).
  6. Dry Bath (for Rapid PCR protocol)
  • Also required additionally are: Trueprep® AUTO Universal Sample Pre-treatment Pack (REF60205AB05 / REF60205AB20 / REF60205AB25 / REF60205AB50 /REF60205AB100 / REF60205AB200), Trueprep® AUTO Transport Medium for Swab Specimen Pack (REF60206TS05 / REF60206TS20 / REF60206TS25 /REF60206TS50 / REF60206TS100 / REF60206TS200), STABILYSE® Prep Free Pack (REF 90101PF05 / 90101PF20 / 90101PF25 / 90101PF50 / 90101PF100 / 90101PF200), Trueprep® AUTO Universal Cartridge Based Sample Prep Kit (REF60203AR05 / REF60203AR25 / REF60203AR50 / REF60203AR100) or Trueprep®
  • AUTO v2 Universal Cartridge Based Sample Prep Kit (REF60207AR05 /REF60207AR25 / REF60207AR50 / REF60207AR100), Truenat® Positive Control Kit – Panel I (REF 801010008), Powder free disposable gloves, waste disposal container with lid.

SPECIMEN PREPARATION FOR EXTRACTION WITH Trueprep® AUTO/AUTO v2

Swab specimen:

  • Oropharyngeal or nasopharyngeal swab specimen must be collected as per standard procedures using a standard nylon flocked swab. Insert the swab with specimen into the Transport Medium for Swab Specimen Tube provided and mix well by repeatedly twirling the swab in the buffer solution. Gently break the handle of the nylon swab at the break point, leaving the swab containing the specimen in the Transport Medium for Swab Specimen Tube. Tightly close the cap of the Transport Medium for Swab Specimen Tube (Refer to the package insert of Trueprep® AUTO Transport Medium for Swab Specimen Pack for further details). Dispose off the remaining part of the swab after use, as per the section on “Disposal and Destruction” (Section 19).

Sample Storage and Transportation:

  • Transport Medium for Swab Specimen decontaminates the specimen and makes it ready for storage / transportation / extraction. The specimen in this form is stable for up to three (3) days at 40ºC and one(1) week at 30ºC.

Nucleic acid extraction:

  • Transfer 500 µL from the Transport Medium for Swab Specimen Tube into the Lysis Buffer Tube for oropharyngeal or nasopharyngeal swabs for further procedure (Refer to the package insert of Trueprep® AUTO Universal Sample Pre-treatment pack for further details) with the Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Device and Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Kit (Refer to the User Manual of Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Device and the package insert of Trueprep® AUTO/AUTO v2 Universal Cartridge Based Sample Prep Kit for details). Dispose off the
  • Transport Medium for Swab Specimen Tube with cap, lysis buffer tube with cap and transfer pipette after use, as per the section on “Disposal and Destruction” (Section 19).
SPECIMEN PREPARATION FOR RAPID PCR PROTOCOL

Swab specimen:

  • Oropharyngeal or nasopharyngeal swab specimen swab specimen must be collected as per standard procedures using a standard nylon flocked swab. Insert the swab with specimen into the provided Collection and Lysis Medium Tube and mix well by repeatedly twirling the swab in the buffer solution. Dispose off the swab after use, as per the section on “Disposal and Destruction” (Section 19). Tightly close the cap of the tube.

Sample Storage and Transportation:

  • The media is used as a medium for sample collection and lysis for Prep Free Swab Specimen. The specimen in this form is stable for up to two (2) days at 30ºC and one(1) week at 2ºC – 8ºC.
  • Sample processing : Set the dry bath at 104oC. Once the temperature reaches to the set temperature [+/- 2oC] place the solution in the Collection and Lysis Medium Tube in dry bath and incubate for 5 minutes at 104oC. Remove the tube from the dry bath and allow to stand at Room Temperature for at least 3 minutes (Refer to the package insert of STABILYSE® Prep Free pack for further details). Proceed for the further test procedure as per the Section 16. Dispose off Collection and Lysis Medium Tube with cap after use, as per the section on “Disposal and Destruction” (Section 19).

SAFETY PRECAUTIONS

  1. For in vitro diagnostic use only.
  2. Bring all reagents and specimen to room temperature (20 – 30OC) before use.
  3. Do not use kit beyond expiry date.
  4. Carefully read the User Manuals, package inserts and Material Safety Data Sheets (MSDS) of all the components of the Truelab® Real Time micro PCR System before use.
  5. All materials of human origin should be handled as though potentially infectious.
  6. Do not pipette any material by mouth.
  7. Do not eat, drink, smoke, apply cosmetics or handle contact lenses in the area where testing is done.
  8. Use protective clothing and wear disposable gloves when handling samples and while performing sample extraction.
PROCEDURAL PRECAUTIONS
  1. Check all packages before using the kit. Damage to the packaging does not prevent the contents of the kit from being used. However, if the outer packaging is damaged the user must confirm that individual components of the kit are intact before using them.
  2. Do not perform the test in the presence of reactive vapours (e.g. from Sodium hypochlorite, acids, alkalis or aldehydes) or dust.
  3. While retrieving the Truenat® COVID-19 micro PCR chip, microtube and the DNase & RNase free pipette tip from the pouch, ensure that neither bare hands nor gloves that have been used for previous tests run are used.

PROCEDURAL LIMITATIONS

  1. Optimal performance of this test requires appropriate specimen collection, handling, storage and transport to the test site.
  2. Though very rare, mutations within the highly conserved regions of the target genome where the Truenat® assay primers and/or probe bind may result in the under-quantitation of or a failure to detect the presence of the concerned pathogen.
  3. The instruments and assay procedures are designed to minimize the risk of contamination by PCR amplification products. However, it is essential to follow good laboratory practices and ensure careful adherence to the procedures specified in this package insert for avoiding nucleic acid contamination from previous amplifications, positive controls or specimens.
  4. A specimen for which the Truenat® assay reports “Not Detected” cannot be concluded to be negative for the concerned pathogen. As with any diagnostic test, results from the Truenat® assay should be interpreted in the context of other clinical and laboratory findings.

CLEANING AND DECONTAMINATION

  1. Spills of potentially infectious material should be cleaned up immediately with absorbent paper tissue and the contaminated area should be decontaminated with disinfectants such as 0.5% freshly prepared Sodium hypochlorite [10 times dilution of 5% Sodium hypochlorite (household bleach)] before continuing work.
  2. Sodium hypochlorite should not be used on an acid-containing spill unless the spill-area is wiped dry first. Materials used to clean spills, including gloves should be disposed off as potentially bio-hazardous waste e.g. in a biohazard waste container.

TEST PROCEDURE

(Please also refer the Truelab® Real Time Quantitiative micro PCR Analyzer user manual)

  1. Switch on the Truelab® Analyzer .

  2. Select User and enter password.

  3. For Truelab® Uno Dx, select the test profile for “COVID-19” to be run from the Profiles Screen on the Analyzer screen. For Truelab® Duo/Quattro, select the Bay (Idle1/2) for Duo and (Idle1/2/3/4) for Quattro from the Status Screen to view the Profiles Screen. Select the test profile for “COVID-19” to be run from the Profiles Screen on the Analyzer screen.

  4. Enter the patient details as prompted in the Truelab® Analyzer screen.

  5. Press Start Reaction.

  6. For Truelab® Uno Dx, Press the eject button to open the chip tray. For Truelab® Duo/Quattro, the chip tray opens automatically on tapping the “Start Reaction” button.

  7. Open a pouch of Truenat® COVID-19 and retrieve the micro PCR chip, microtube and DNase & RNase free pipette tip.

  8. Place the Truenat® COVID-19 chip on the chip tray without touching the white reaction well. The reaction well should be facing up and away from the Analyzer. Gently press the chip to ensure that it has seated in the chip tray properly.

  9. Place the microtube containing freeze dried RT PCR reagents in the microtube stand provided along with the Truelab® Real Time micro PCR workstation after ensuring that white pellet of dried PCR reagents remains at the bottom of the microtube. Remove the microtube cap and dispose it off as per the section on “Disposal and Destruction” (Section 19). Using the filter barrier tip provided in the pouch, pipette out six (6) µL of the purified RNA from the Elute Collection Tube (for Extraction with Trueprep® AUTO/AUTO v2)/Collection and Lysis Medium Tube (for Rapid PCR protocol) into the microtube. Allow it to stand for
    30- 60 seconds to get a clear solution. Do not mix it by tapping, shaking or by reverse pipetting. Using the same filter barrier tip, pipette out six (6) µL of this clear solution and dispense into the centre of the white reaction well of the Truenat® COVID-19 chip. Take care not to scratch the internal well surface and not to spill elute on the outside of the well. Dispose off the microtip as per the section on “Disposal and Destruction” (Section 19).

  10. For Truelab® Uno Dx, slide the chip tray containing the Truenat® COVID-19 Chip-based Real Time PCR test loaded with the sample into the Truelab® Analyzer. Press Done on the “Please Load Sample” Alert message. For Truelab® Duo/Quattro, select “YES” at the Please load Sample prompt. Chip tray will close automatically and the reaction will start.

  11. Read the result from the screen.

  12. After the reaction is completed, for Truelab® Uno Dx, push the Eject button to eject the chip tray. For Truelab® Duo/Quattro, tap the “Open/Close Tray” button to eject the chip tray.

  13. Take out the Truenat® COVID-19 Chip-based Real Time PCR test at end of the test and dispose it off as per the section on “Disposal and Destruction” (Section 19).

  14. Turn on Truelab® micro PCR printer and select print on the screen for printing out hard copy of t he results. Test results are automatically stored and can be retrieved any time later. (Refer to the Truelab® Analyzer manual).

  15. Switch off the Truelab® Analyzer.

1RESULTS & INTERPRETATIONS

Three amplification curves are displayed on the Truelab® Real Time micro PCR Analyzer screen to indicate the progress of the test. Both the target and the internal positive control (IPC) curves will take a steep, exponential path when the fluorescence crosses the threshold value in case of positive samples. The time taken (Ct) of the specimen will depend on the number of virus copies in the sample. The curve will remain horizontal throughout the test duration and the IPC curve will take an exponential path in case of negative samples. In case the IPC curve remains horizontal in a negative sample, the test is considered as Invalid. At the end of the test run, the results screen will display “DETECTED” for Positive result or “NOT DETECTED” for Negative result. The result screen would also display the viral load as “HIGH (Ct<20)”,“MEDIUM (20≤Ct<25)”, “LOW (25≤Ct<30)” or “ VERY LOW (Ct ≥ 30)” for positive specimen. The result screen also displays the validity of the test run as “VALID” or “INVALID”. Invalid samples have to be repeated with fresh specimens from the sample preparation stage. While IPC will co-amplify in most positive cases also, in some specimen having a high target load, the IPC may not amplify, however, the test run is still considered valid.

Detec tion Channel Result Interpretation Action
Orf1a E RNase P  
+ + +/- SARS CoV-2 POSITIVE Report Positive
+ +/- SARS CoV-2 POSITIVE Report Positive
+/- SARS CoV-2 PRESUMPTIVE POSITIVE Repeat after 48-72

hours
| | +| SARS CoV-2 NEGATIVE| Report Negative
| | | INVALID| Collect new swab and repeat

QUALITY CONTROL PROCEDURES
  • To ensure that the Truelab® Real Time micro PCR Analyzer is working accurately, run positive and negative controls from time to time. Truenat® Positive Control Kit – Panel I (REF 801010008), The containing Positive Control and Negative Control must be ordered separately. It is advisable to run controls under the following circumstances:
  • Whenever a new shipment of test kits is received.
  • When opening a new test kit lot.
  • If the temperature of the storage area falls outside of 2-30o C.
  • By each new user prior to performing testing on clinical specimen.

DISPOSAL AND DESTRUCTION

  1. Submerge the used content such as remaining part of the swab, Transport Medium for Swab Specimen Tube with cap, lysis buffer tube with cap, transfer pipettes, Collection and Lysis Medium Tube (used for Rapid PCR Protocol), Truenat® COVID-19 chip, microtube, microtube cap, pipette tips etc. in freshly prepared 0.5% sodium hypochlorite solution for 30 minutes before disposal as per the standard medical waste disposal guidelines.
  2. Disinfect the solutions and/or solid waste containing biological samples before discarding them according to local regulations.
  3. Samples and reagents of human and animal origin, as well as contaminated materials, disposables, neutralized acids and other waste materials must be discarded according to local regulations after decontamination by immersion in a freshly prepared 0.5% of sodium hypochlorite for 30 minutes (1 volume of 5%sodium hypochlorite for 10 volumes of water).
  4. Do not autoclave materials or solutions containing sodium hypochlorite.
  5. Chemicals should be handled in accordance with Good Laboratory Practice and disposed off according to the local regulations.

SPECIFIC PERFORMANCE CHARACTERISTICS

Performance parameters

  1. Analytical Sensitivity:

Evaluation of the analytical sensitivity of Truenat® COVID-19 assay, in comparison to TaqMan SARS CoV-2 qRT-PCR VRDL assay was performed. Sample with low Ct value (ID 54049) was used for this study. An aliquot of VTM of sample ID 54049 was extracted using Trueprep® AUTO Universal Cartridge Based Sample Prep Device (As per manufacturer protocol). RNA was diluted 10 fold and six dilutions were made from Trueprep® AUTO elute. These dilution series were run on Truenat® COVID-19 chips as well as Taq Man SARS CoV-2 qRTPCR systems in parallel. Observed Ct values are given in below table.

  Truenat ® COVID-19 Taqman SARSCoV2 [VRDL]
Sample ID Orf1a E Gene
E Gene RNAseP
54049 D1 19.67 19.17
54049 D1 20.14 19.11
54049 D1 19.60 19.00
54049 D2 23.17 22.17
54049 D2 23.29 22.40
54049 D2 23.20 22.33
54049 D3 26.29 25.80
54049 D3 26.29 26.50
54049 D3 26.00 25.80
54049 D4 29.67 29.14
54049 D4 30.00 29.20
54049 D4 30.83 29.50
54049 D5 33.20 31.50
54049 D5 32.28 32.33
54049 D5 33.00 32.13
54049 D6 ND ND
54049 D6 ND ND
54049 D6 ND ND

Conclusion: Truenat® assay detected up to dilution 10-5(D5) from undiluted sample, with valid Ct value.

Linearity & PCR Efficiency:

Using the dilution series from Trueprep® AUTO elutes run on Truenat® COVID-19, log linear curve was plotted to check the linearity of Ct values on Truenat® COVID-19 test.Truenat-601430005-COVID-19-Packinsert-
FIG-4

Figure 1: Linearity and PCR efficiency on Truenat® COVID-19. Y axis indicate Ct values and X axis is arbitrary log numbers indicating dilutions.
Conclusion: Slope of the curve is Orf1a ( -3.303) and E gene ( -3.278). Assay was observed to be linear over the range of dilutions tested and PCR efficiency was found to be 100.8 % for Orf1a and 101.87% for E gene.

Analytical Sensitivity test with samples of low viral load:
Evaluation of analytical sensitivity of Truenat® COVID-19 multiplexed PoC test, in comparison to TaqMan SARS-CoV-2 rRT-PCR VRDL assay was performed. Five Samples with low viral load (low viral copy number) for Orf1a / E gene was used for this study. An aliquot of virus transport medium (VTM) containing sample was extracted using Trueprep® AUTO and 12 replicates of each sample was run on Truenat® COVID-19 and compared against TaqMan SARS CoV-2 qRT-PCR VRDL assay.
Conclusion: All replicates of 5 low viral load samples were detected by Truenat® COVID-19, Orf1a assay.
Limit of Detection (LoD):
The limit of detection (LoD) was estimated using AccuPlex™ SARS-CoV-2 Verification Panel (Seracare, 0505-0168), as per CLSI EP-17A guidelines. Briefly, six dilutions of quantified material were done in Trueprep® AUTO Transport Medium for Swab Specimen Pack. Each dilution was extracted 24 times using Trueprep® AUTO Universal Cartridge Based Sample Prep Device and tested on the Truenat® COVID-19 chip. Probit regression was used to estimate LoD at 95% positivity, as indicated below.Truenat-601430005-COVID-19-Packinsert-
FIG-5

LoD of Orf1a and E gene was estimated to be 480 and 487 genome copies/mL respectively and in transport medium for swab specimen, with 95% CI interval of 410 – 628 and 419-631 respectively.

Precision:

To evaluate reproducibility and repeatability of the assay, ten clinical elutes representing High, Medium and Low Ct values {Sample IDs:54053(Ct:20.75), 54716(Ct:25.8),54724(Ct:14.71),54745(Ct:29.43),54803(Ct:14.11),54897 (Ct:22.83),54918(Ct:24.06),54935(Ct:20.43),55055(Ct:14.60),55113(Ct:24.29} were run on devices used in this evaluation. Following table depicts the Precision analysis. Ct values for E gene and Orf1a are given, with observed standard deviation and % CV.

 | ID:| 54053| ID:54| 716| ID:54| 724| ID:54| 745| ID:54| 803
---|---|---|---|---|---|---|---|---|---|---
Device ID| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene
TLDU1306| 27.50| 26.33| 25.80| 24.50| 14.71| 13.67| 29.43| 29.00| 14.11| 12.60
TLDU0431| 28.13| 26.43| 26.80| 25.00| 14.30| 13.50| 30.14| 29.17| 13.50| 12.60
TLDU0381| 28.14| 26.22| 26.35| 24.60| 14.50| 13.43| 31.50| 29.33| 14.17| 12.57
TLDU0366| 26.75| 25.80| 26.00| 24.43| 14.17| 13.33| 28.67| 29.00| 13.86| 12.60
Mean| 27.63| 26.20| 26.24| 24.63| 14.42| 13.48| 29.94| 29.13| 13.91| 12.59
STDEV| 0.659| 0.277| 0.439| 0.255| 0.236| 0.143| 1.204| 0.158| 0.305| 0.015
%CV| 2.38| 1.06| 1.67| 1.03| 1.64| 1.06| 4.02| 0.54| 2.19| 0.12
 | ID: 54| 897| ID:54| 918| ID:54| 935| ID: 55| 055| ID:55| 113
---|---|---|---|---|---|---|---|---|---|---
Device ID| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene| Orf1a| E Gene
TLDU1306| 22.83| 22.27| 24.06| 21.29| 20.43| 19.43| 14.60| 13.60| 24.29| 23.00
TLDU0431| 23.13| 22.14| 22.40| 21.00| 20.17| 19.50| 15.00| 13.83| 24.10| 23.13
TLDU0381| 23.00| 22.00| 23.60| 21.00| 21.25| 19.80| 15.14| 13.50| 24.71| 23.14
TLDU0366| 22.71| 21.80| 23.00| 20.80| 20.60| 19.60| 14.67| 13.43| 24.43| 23.13
Mean| 22.92| 22.05| 23.27| 21.02| 20.61| 19.58| 14.85| 13.59| 24.38| 23.10
STDEV| 0.185| 0.201| 0.722| 0.202| 0.460| 0.161| 0.259| 0.175| 0.257| 0.067
%CV| 0.81| 0.91| 3.10| 0.96| 2.23| 0.82| 1.74| 1.28| 1.05| 0.29

Conclusion: The test was found to be reproducible with percent coefficient of variation less than 5% which is well below the accepted 10%, across samples and between devices.

Clinical Sensitivity:

  • Clinical sensitivity was tested by running 40 confirmed positives samples of SARS-COV-2; representing high [12], medium [13] and low Ct [15] value samples for testing and comparison with both the systems using three lots of Truenat® COVID-19. All positive samples were detected by all three lots of Truenat® COVID-19 assay.

Specificity and cross-reactivity:

  • The specificity of the test was evaluated using a panel of clinical samples, including COVID-19 negatives and other respiratory disease positive ones. The panel included; H1N1 (38 samples: 8 positive and 30 Negative), Severe Acute Respiratory Illness (SARI) (30 samples), and also blood samples (3 from SARS CoV-2 positive cases and 3 from SARS CoV-2 negative cases) and confirmed COVID-19 negatives (30 swab samples) were used.
  • Egene and Orf1a was not detected in any of above specimens, indicating specificity and no cross reactivity to other common respiratory pathogens.

Clinical evaluation:

  • Clinical evaluation of Truenat® COVID-19 was performed at the State VRDL lab at Bangalore Medical College. Totally, 155 confirmed SARS CoV-2 positives and 323 confirmed RT-PCR negative samples were tested. Comparison of results from Real Time PCR (BioRad CFX96) and Truenat® COVID-19 is summarized as follows:
  SARS CoV 2 real-time PCR (n=478)
Positive Negative

Truenat ®

COVID-19

| Positive| 155| 0| 155
Negative| 0| 323| 323
Total| 155| 323| 478

All positive and negatives were correctly detected on Truenat® COVID-19 assay indicating 100% sensitivity, specificity and 100% overall concordance to reference gold standard assay.

SYMBOL KEYS

Truenat-601430005-COVID-19-Packinsert-
FIG-6

  • Molbio Diagnostics Private Limited
  • Registered Office & Manufacturing Unit – I:
  • Plot No. L-46, Phase II D, Verna Industrial Estate, Verna, Goa – 403 722, INDIA Manufacturing Unit – II:
  • Plot No. L-42, Phase II B, Verna Industrial Estate, Verna, Goa – 403 722, INDIA www.molbiodiagnostics.com
  • Email: [email protected] (Sales Enquiries)
  • [email protected] (Feedback and Customer Support)
  • Qarad EC-REP BV, Pas 257, 2440 Geel, Belgium

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals