ATHENA TC-HOOD Tissue Culture Flow Hood with Accessories User Manual
- June 11, 2024
- Athena
Table of Contents
ATHENA TC-HOOD Tissue Culture Flow Hood with Accessories
INTENDED USE
The Athena Flow Hood and its accessories are intended to be used for the
purposes of facilitating plant tissue culture. Designed for professional use,
the product is not intended to be used by individuals with reduced sensory,
physical, or mental capabilities, or by individuals with lack of experience or
knowledge. Not suitable for any other purpose than defined herein.
IMPORTANT SAFETY INSTRUCTIONS
When using electrical furnishing, basic precautions should always be followed,
including the following: Read all instructions before using (this furnishing).
DANGER – To reduce the risk of electric shock: Unplug the power plug
before cleaning, and wipe the dust on the surface with a damp cloth when
cleaning. Be careful not to let water drops into the internal parts, or loosen
the connector.
DANGER!
Read the manual before use, and use only as directed. Improper use may cause
injury or other damage. Keep appliances and cords away from children. Never
drape cord over the edge of the counter, never use the power outlet below the
counter, and never use it with an extension cord. To reduce the risk of
electric shock, fire, or injury, plug the power cord into a grounded
electrical outlet that is easily accessible. Disconnect the power before
cleaning, performing any maintenance, or replacing the filter. To ensure
proper operation of the Flow Hood, the filter should be replaced every 3
months of regular use, when contamination is evident, or when airflow is
noticeably reduced. Failure to do so may result in decreased product
performance or damage to the unit. Dispose of used filters in accordance with
local ordinances and regulations. The operating fans present a risk of injury
by contact. Ensure to keep your hands clear. Note that the fan blades may
continue to rotate for some time after the operation ceases. Confirm that
rotating fan blades have come to a complete stop before beginning routine
maintenance work, such as replacing the HEPA filter or fuses. Failure to do so
may result in bodily injury. Visually inspect the unit and cord prior use, to
check for loose parts or damage. Do not use if damaged.
WARNING
-
Keep a clean, unobstructed area behind the Flow Hood.
-
Close the filter door when not in use.
-
Do not expose to rain or snow.
-
Max load for both outlets may not exceed 10 amps.
-
Use of a power supply or charger not recommended or sold by the product manufacturer may result in risk of fire or injury.
-
Store flow hood in a cool, dry place.
-
Keep in an upright position while transporting.
-
Do not place where ambient room air vents may disrupt the Flow Hood airflow.
-
Not for use in high temperature or humidity environments.
-
Not for use in high wind areas; always shelter the Flow Hood for maximum effectiveness.
-
The product doesn’t provide with a work surface. A secondary working surface shall be provided which is able to support an additional 33 lbs and 0.7m³ working area.
-
The open flame of the alcohol burner presents a burn and fire hazard. Never leave the alcohol burner unattended while it is lit. When using, ensure a fire extinguisher is accessible.
-
Alcohol burners produce small amounts of carbon monoxide and other toxic combustion by-products. Use only in a well-ventilated area.
-
Use only ethyl alcohol or isopropyl alcohol (with an alcohol content of 90% or higher) with the alcohol burner. Use of other fuels may result in serious injury. Read and follow the safety instructions on the alcohol container.
-
The alcohol burner may become hot during use and cause burns. Do not touch any part of the burner with your hands until it has fully cooled.
-
The scalpel and its blades are sharp and present a danger of serious injuries and severe bleeding.
Do not touch the blades on the sharp edges. Store in a safe and dry area, away from children. -
Prior to disposing of used blades, wrap in a protective coating to prevent injury.
CLEANING
Wipe with a microfiber cloth and water.
Overview
HEPA Filter | True H13 Captures particles greater than 85% of 0.5μm |
---|---|
Pre-Filter | G3 Captures particles greater than 85% of 0.5μm |
Filter Life | 3 months of regular use |
Average Face Velocity | 0.5 M/s / 98 fpm |
Working Area | 2.58ft3 / 0.7m3 |
Power | 85 – 264VAC 50w 1.3A |
AC Output AC Output | 125V 15A (max load for both outlets may not exceed 10 |
amps)
125V 15A (max load for both outlets may not exceed 10 amps)
Dimensions (L x W x H)| 9.80” x 24” x 15.7” / 250mm x 609mm x 398mm
Weight| 33 lbs / 15 kg
Power Cord| 125V 15A NEMA 5-15P / C13
- A: Foam Toolbox Tray
- B: SOP
- C: Toolbox Lid
- D: Pre-Filter
- E: HEPA Filter
- F: Latch
- G: Filter Door
- H: Carry Handles
- I: AC Output 1
- J: AC Output 2
- K: Power Input
- L: Fuse Door
- M: Air Intake
Instructions
WARNING
-
Place the Flow Hood on a flat, stable surface.
Note: Allow a minimum of 4” between the wall and the rear air intake vents (FIG A). -
Plug in the power cord to the POWER INPUT connector.
Note: Flow Hood and Autoclave power cords are the same, can use with either product. -
Open the filter cover to power on the fans and LED lighting (FIG B).
Note: Close the filter cover when not in use. -
Open each latch to unlock and open the toolbox lid. Rotate until the hinges stop (FIG C).
-
Follow the procedure printed on the toolbox lid to begin.
WARNING: Unplug the power cord before doing any maintenance.
FILTER MAINTENANCE
Both the pre-filter and the HEPA filter should be replaced every 3 months of regular use, when contamination is evident, or when airflow is noticeably reduced.
- Open the toolbox lid.
- Remove the foam toolbox tray by lifting up the straps on each side (FIG D).
- Pull out the air filters using the tab placed on top (FIG E).
- Replace with new filters ensuring they face the correct direction per the tab’s location.
- Replace the filter maintenance card that came with the replacement filter packaging. Record the date the filters were replaced (FIG F).
- Ensure the vent intake on the rear of the flow hood is kept clean.
FAN MAINTENANCE
There are four fans located between the two air filters. These fans should be
inspected and cleaned during each filter change. Replacement fans are also
available.
- Remove the front HEPA air filter for access to the fan compartment (FIG G).
- Use a Phillips head screwdriver to remove the two screws that hold the wiring plate along the bottom of the fans (FIG H).
GROUNDING INSTRUCTIONS
This product is for use on a nominal 120-volt circuit and has a grounding plug
that looks like the plug illustrated in sketch A ( Fig 1). Make sure that the
product is connected to an outlet having the same configuration as the plug.
No adapter should be used with this product.
-
Use a Phillips head screwdriver to unscrew the fan (FIG I).
-
Gently remove the fan and unplug the connectors by pushing down on the clip (FIG J).
Note: Keep the wiring harness in place to prevent damage. -
The fan may be cleaned of any dust and debris build-up or completely replaced.
-
Plug in the connector and reinstall the screws for the fan.
-
Repeat this process for each fan.
FUSE REPLACEMENT
- Use a small flathead screwdriver to gently pry off the fuse door and remove the fuse (FIG K).
- Replace the fuse with a 5 x 20mm 10A fuse.
- Snap the fuse door back on.
FLOW HOOD TROUBLESHOOTING
- The LED light turns on but the fans do not turn on.
- Ensure the fan connectors are attached (See Fan Maintenance).
- Replace any malfunctioning fan.
- The fans turn on but the LED light does not turn on.
- Remove the foam tool molding.
- Use a Phillips head screwdriver to remove the four screws at each corner of the bottom cover.
- Remove the bottom cover from the toolbox.
- Ensure the light connector is attached to the power cord.
- The Flow Hood LED light and fans do not turn on.
- Ensure the source of power is adequate.
- Test the Autoclave power cord on the Flow Hood. They use the same power cord.
Accessories
- A: Scalpel Blades (50)
- B: 3 mL Pipettes (3)
- C: Forceps (1)
- D: Scalpel Handle (1)
- E: Culture Vessels (6)
- F: 10-Pack Box Roots
- G: 10-Pack Box Shoots
- H: Mixing Vessel (1)
- I: Chemical Bottles (2)
- J: Spray Bottle (1)
- K: Working Surface (1)
- L: Alcohol Swabs (30)
- M: Utility Vessels (3)
- N: Power Cord (1)
- O: Alcohol Burner (1)
- P: Utility Vessel Lids (3)
- Q: Pre-Filter (1)
- R: HEPA Filter (1)
Warranty + Support
Warranty; Limitations. Athena warrants that the Autoclave and replacement
parts (“Goods”) shall be free from defects in material and workmanship under
proper and normal and recommended use for a period of 1 year from the date of
shipment. Goods shall be considered defective only if a defect exists that
materially impairs the value of the Goods to Customer and does not comply with
Athena’s specifications (or, if none, does not comply with normal industry
tolerances or standards). Goods shall not be defective if the Goods comply
with Customer’s instructions or if the Goods have been modified, have been
subject to misuse or neglect or if the Goods are not operated in a typical
fashion. In accordance with the terms of this limited warranty (“Warranty”),
Athena will, at its option, either replace any such defective Goods or parts
thereof or issue Customer a credit for the purchase price of the defective
Goods, provided that written notice of the defect is received by Athena within
30 days of the appearance of such defect and within the 1-year warranty
period. If notice is not given by the required deadlines set forth in this
paragraph, any claim for breach of warranty is waived and Athena shall not be
liable. Upon receipt of notice of a possible defect in the Goods, Athena shall
issue instructions for their disposition. No Goods shall be returned to Athena
without its prior consent. Transportation charges for return and replacement
under the Warranty will be borne by Athena if the Goods are defective. Under
no circumstances will Athena’s liability exceed the purchase price of the
defective Goods.
DISCLAIMER OF WARRANTIES: To the extent consistent with applicable law,
Athena Ag, Inc. makes no other warranties express or implied, of
merchantability or of fitness for a particular purpose or otherwise, that
extend beyond the statements made on the product’s label. No agent of Athena
Ag, Inc. is authorized to make any warranties beyond those contained herein or
to modify the warranties contained herein. To the extent consistent with
applicable law, Athena Ag, Inc. disclaims any liability whatsoever for
special, incidental, or consequential damages resulting from the use or
handling of this product. LIMITATIONS OF LIABILITY: To the extent consistent
with applicable law, the exclusive remedy of the user or buyer for any and all
losses, injuries, or damages resulting from the use or handling of this
product, whether in contract, warranty, tort, negligence, strict liability or
otherwise, shall not exceed the purchase price paid, or at Athena Ag Inc.’s
election, the replacement of product.
- 1-YEAR MANUFACTURER
- WARRANTY
READ ALL INSTRUCTIONS BEFORE USING, OR SERVICING, THIS EQUIPMENT. KEEP THIS MANUAL IN A LOCATION THAT IS READILY AVAILABLE TO USERS AND SERVICE TECHNICIANS.
WARNING: THIS PRODUCT CONTAINS ACCESSORIES THAT MAY CAUSE INJURY OR HARM IF USED IMPROPERLY. English (Original Instructions)
For technical support and troubleshooting please call or email: 844-333-1818 support@athenaag.com www.AthenaAg.com