AND AD-8526 Ethernet Converter Instruction Manual
- June 4, 2024
- AND
Table of Contents
AND AD-8526 Ethernet Converter
The AD-8526 Ethernet Converter can connect the RS-232C interface of a weighing instrument to the Ethernet (LAN) port of a computer that is not equipped with an RS-232C interface. By using the AD8526, weight data from a weighing instrument can be managed with a computer connected to the network. The weighing instrument can also be controlled by the connected computer
Unpacking the AD-8526
AC adapter labels
Affix them to the AC adapter and the connector end of the cable.
IP address label (5 labels)
Write a preset IP address and affix it to the weighing instrument
IP address protective cover
(5 labels) Affix it on the IP address label
-
Note: Only the following cable is included if specified when ordering:
DIN 7-pin to D-Sub 9-pin cable (AX-KO1786-200) -
If not specified, the following two communication cables are included:
D-Sub 25-pin to D-Sub 9-pin cable (AX-KO1710-200), D-Sub 9-pin to D-Sub 9-pin cable (AX-KO2466-200)
Specification
1. Specification | 3. RS-232C serial interface |
---|---|
Operating environment: -10°C to +40°C | Connector: |
AC adapter: Confirm that the adapter type is correct for the local voltage and power receptacle type| Transmission form: Asynchronous, bi-directional
Data format
| Baud rate: 600, 1200, 2400*, 4800,
9600, 19200bps
Power consumption: Approx. 11VA (supplied to the AC adapter)| Data bits:| 7
or 8 bits
Dimensions:| 113(W) x 60(D) x 38(H) mm| Parity:| Even, Odd, None
Net weight:| Approximately 250g| Stop bits:| 1 bit* or 2 bits
2. Ethernet interface|
*: Factory settings
Connector:| RJ45
Protocol:| TCP/IP
Weighing Instrument Cables
Weighing instrument | Interface option | Communications cable (Length 2 m) |
---|
GX, GF, GX-K, GF-K, GP, GR, HR,
MC
| None (D-Sub 25-pin, standard accessory)| AX-KO1710-200
EK- i , EW- i , FC- i , FC-S i , GH, HR- i|
None (D-Sub 9-pin, standard accessory)| AX-KO2466-200
AD-4212C| None (D-Sub 9-pin, standard accessory)| AD4212C standard accessory
HV-G, HV-WP, HW-G, HW-WP| None (DIN 7-pin, standard accessory)| AX-KO1786-200
FG| OP-03 (DIN 7-pin)| AX-KO1786-200
FG-L, FG-M| OP-23 (DIN 8-pin)| AX-KO1786-200
Connecting the Weighing Instrument to a Computer
Caution
- Contact your network manager before connecting the AD-8526 to a network. The AD-8526 may generate a network error. A&D assumes no responsibility for any errors that may be caused.
- Affix the AC adapter labels to the AC adapter and the connector end of the cable.
- Set the IP address and subnet mask to the AD-8526 one at a time. The factory default IP address is 172.16.100.2. Do not duplicate an IP address.
- The IP address cannot be returned to the factory setting. We recommend that you write the IP address on the supplied label.
Preparation
-
Please download each software from the A&D website https://www.aandd.jp/products/software/software.html below.
(1) Setting procedure “WinCT-Plus” instruction manual
(2) IP address setting software “Device Installer”
(3) Data acquisition software “WinCT-Plus” -
Step 1 Connect the AC adapter to the AD-8526.
-
Step 2 Connect the AD-8526 and the computer with a cross cable directly or connect them with a hub and straight cables.
-
Step 3 Input the IP addresses and subnet masks of the AD-8526 and computer. For the setting procedure, refer to the “WinCT-Plus” instruction manual. Communication errors may occur if set incorrectly.
-
Step 4 Write the IP address on the IP address label and affix it on the AD-8526 (and the weighing instrument if necessary) where you can see the address easily.
-
Step 5 Install the RsMulti data acquisition software to download the “WinCT-Plus” instruction manual.PDF from the A&D website. https://www.aandd.jp/products/software/software.html
-
Step 6 Connect the accessory RS-232C cable between the RS-232C interface of the weighing instrument and the AD-8526’s port
Data Communication Software
The RSMulti data acquisition software can be launched on a Windows PC. Menu: [Start] [Program] [A&D WinCT-Plus] [RsMulti]. Refer to the Manual PDF file in the A&D WinCT-Plus for details on the operation.
- This software can acquire data from multiple devices connected by LAN or RS-232C.
- The software can control these devices with commands.
- The software can acquire data transmitted from devices. Example: When the PRINT key is pressed on the balance, data is output and is acquired with the computer.
- Stored data can be used with Excel. (Install Microsoft Excel before use.)
Example
IP address | Subnet mask | |
---|---|---|
Computer | 172.16.100.1 |
255.255.0.0
GX-600| 172.16.100.2
GP-30K| 172.16.100.3
Two balances are connected.
A sample window of RsMulti
Documents / Resources
| AND
AD-8526 Ethernet
Converter
[pdf] Instruction Manual
AD-8526, Ethernet Converter, AD-8526 Ethernet Converter
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>