The Pain-Free Blood Glucose Monitoring System
User Instructions for the Genteel® Lancing Device
PETTEST / GENTEEL PARTS
(1) Contact Tip
(2) Nozzle
(3) Nozzle Slots
(4) Lancet
(5) Shuttle
(6) Front End Tabs
(7) Front End| (8) Activation Button
(9) Hole In Activation Button
(10) Vacuum Tube
(11) Push Shaft
(12) Vacuum Indicator
(13) Push Cap
(14) Piston Assembly
---|---
GATHER SUPPLIES NEEDED
With clean hands, gather supplies needed for testing:
(A) PetTest/Genteel Lancing Device
(B) Contact Tip Card (contains 6 Contact Tips)
(C) Advocate Twist Top or Butterfly Touch Lancets
(D) Alcohol (not included) to clean Contact Tip
(E) Clean, damp cloth (not included) for cleaning Genteel’s surface
(F) Meter and test strips
TESTING SITES FOR DRAWING BLOOD
Cats have one testing site while dogs have up to six to choose from.
PREPARE GENTEEL FOR LANCING
PREPARATION : Clean the surface of Genteel and other testing supplies.
Verify Contact Tip is securely affixed to Nozzle.
4.1 Prime Genteel: Press the Push Cap to prime the PetTest/Genteel,
device clicks if done correctly.
4.2 Remove Nozzle: Twist counterclockwise and pull gently to remove.
4.3 Insert a Lancet: While the PetTest/Genteel is primed, insert either a
Blue Butterfly or PetTest Maroon lancet into Shuttle and the #1 Blue Contact
Tip to start the test. Carefully twist off the protective cover from lancet
and put aside.
(The PetTest Maroon Twist Top lancets are 6 steps more aggressive than the
Blue Butterfly Touch lancets.)
Note: You want to use the thickest Contact Tip that gets the right amount
of blood with little to no discomfort. Contact Tips going from thickest to
thinnest:
1 Blue #2 Yellow, #3 Green, #4 Clear, #5 Orange, #6 Violet.
NOTE: Different test sites may require a different Contact Tip.
4.4 Replace Nozzle: Line up the Nozzle Slots with the Front End Tabs.
Gently push together and rotate clockwise into place.
PERFORM BLOOD DRAW
Wash hands with soap and water. Clean test site. Have your meter and test
strip ready to wick up your blood sample.
5.1 Place Genteel on skin: Pick your test site from the diagram in 3.0 at
the top right. Make sure you have a seal between Contact Tip and skin by
gently and flatly holding the device perpendicular to the skin. Do not press
Genteel forcibly into the skin, as this can cause pain during testing.
NOTE: Hold Genteel like a pen, with index finger on the Activation
Button.
5.2 Press Activation Button: Using gentle pressure, hold down the
Activation Button for at least 4 seconds, or until you see a blood drop form.
NOTE: With your finger over the Activation Button, the plunger will rise
slowly to the halfway point, but not fully extend (you should not see the red
Vacuum Indicator).
5.3 Release Activation Button: BEFORE pulling Genteel away from your
skin, lift your finger off the Activation Button. Pull Genteel away from the
skin. If done out of order, blood drop may smear.
NOTE: Hold Activation Button longer to get a larger blood drop. If no
blood is produced, choose the next thinnest Contact Tip in order depending on
the lancet and body site chosen.
5.4 Test Blood Sugar: Apply the meter’s test strip to the blood drop and
follow the instructions for the test strips and meter.
5.5 Remove Lancet: Prime Genteel again (see 4.1) and remove Nozzle (see
4.2).
Carefully insert lancet needle into its protective cap. Once needle is safely
in its protective cap, remove the lancet.
5.6 Lancet Disposal: Safely dispose of the used lancet according to local
laws for Sharps disposal.
CLEANING GENTEEL
6.1 Surface cleaning: Before every use, wipe the outside of the device
with a clean damp cloth.
IMPORTANT: If the Push Shaft rises past its halfway point—while holding
the Activation Button, this may be a vacuum leak caused by dirt or other
contaminants inside Genteel. If you experience this vacuum loss symptom—
perform (6.2). If symptom persists—perform (6.3). If cleaning doesn’t resolve
vacuum leak, call customer support +1
866-373-2824.
6.2 Quick Clean (inside):
6.2.1 Remove Piston Assembly: De-prime Genteel by pressing Activation
Button. Twist Piston Assembly base counter-clockwise until you can pull it out
of the Vacuum Tube.
6.2.2 O-ring Wipe: Wipe the O-ring on the Piston Assembly with your clean
finger to remove surface dirt or other contaminants.
6.3 Deep Clean (inside):
6.3.1. Remove Piston Assembly: Repeat step 6.2.1
6.3.2. Flush Vacuum Tube: Run fast moving, warm water through the Vacuum
Tube and out the Nozzle while repeatedly pressing the Activation Button.
6.3.3. Flush Piston Assembly: While running the Piston Assembly under
fast-moving, warm water repeat 6.2.2.
6.3.4. Dry : Let Genteel air dry thoroughly. Do not reassemble Genteel
with moisture in the Vacuum Tube.
6.3.5. Reassemble: Put Piston Assembly back inside the Vacuum Tube and
twist base clockwise until it locks into place securely. Verify device
functions as intended.
Distributed By: Diabetic Supply of Suncoast, Inc.
PO Box 2102, Vega Alta, PR 00692
1-866-373-2824 /
www.ShopPetTest.com
(24 hours a day / 7 days a week)
Follow us on:
Facebook.com/PetTest
STOP! If you need troubleshooting assistance or more help
with your Genteel blood draw experience,
PLEASE DO NOT WAIT TO CONTACT US.
WE ARE HERE TO HELP!
For more information visit