PRO-ALARM-SS-PCB Smart Start Module Kit for aPro Alarm Analyzers
Instruction Manual SKU PRO-
ALARM-SS-PCB
PRO-ALARM-SS-PCB Smart Start Module Kit for aPro Alarm Analyzers
If you have any questions about this equipment, please contact Technical
Support at:
Nuvair
1600 Beacon Place
Oxnard, CA 93033 USA
Phone: +1.805.815.4044
Fax: +1.805.486.0900
Email:info@Nuvair.com
Hours: Monday through Friday
8:00 AM to 5:00 PM Pacific Time
WARNING
This Operation Manual contains important safety information and should always
be available to those personnel operating this equipment. Read, understand,
and retain all instructions before operating this equipment to prevent injury
or equipment damage.
Every effort was made to ensure the accuracy of the information contained
within. Nuvair, however, retains the right to modify its contents without
notice.
Under Nuvair’s system of continuous improvement, certain components may be
updated or changed as higher quality or more efficient parts and assemblies
become available.
Nuvair will make every effort to update manuals as parts and functional
aspects change. However, the look or location of components on your product
may differ from those in this manual if improvements have been made that do
not affect functionality or operational procedures.
Units pictured may also be equipped with different options than those on your
product. In this case, the basic operational and maintenance guidelines will
still apply.
If you have problems or questions after reading the manual, stop and call
Nuvair at +1.805.815.4044 for information.
Introduction
This manual will assist you in the installation of the SMART START Module (SKU
PRO ALARM-SS-PCB) into existing handheld or panel mount Nuvair Pro Alarm
analyzer powered externally by either DIN Rail or wall plug-in. This manual
will also help customers understand the function of Nuvair Pro Alarm analysers
equipped with a factory-installed SMART START Module.
The Nuvair Pro Alarm analyzer series are gas analysers with alarms (visual
and/or audible) that activate when user-defined set points (low and/or high)
are violated.
Be sure to read the entire manual.
1.1 Symbol Conventions
This manual uses certain words and symbols to call your attention to
conditions, practices or techniques that may directly affect your safety. Pay
particular attention to information introduced by the following symbols or
words:
SYMBOL | MEANING | DESCRIPTION |
---|---|---|
DANGER | Indicates an imminently hazardous situation, which if not |
avoided, will result in serious personal injury or death.
| CAUTION| Indicates a potentially hazardous situation, which if not
avoided, could result in serious personal injury or death.
| WARNING| Indicates a potentially hazardous situation, which if not
avoided, may result in minor or moderate injury. It may also be used to alert
against unsafe practices.
| NOTICE| Notifies people of installation, operation or maintenance
information which is important but not hazard-related.
Module Description
As originally designed, the power to Nuvair Pro Alarm analysers is manually
controlled by an On/Off button (Figure 1) on the front of the analyzer panel.
Some customers want to have analysers power up automatically when the
integrated power equipment—like a compressor—turns on. To avoid having to
power on analysers after turning on connected equipment, Nuvair developed the
SMART START Module. Adding this printed circuit board (PCB) to your existing
Pro Alarm analyzer will automatically power the analyzer on when external
power equipment is energized.
The SMART START Module can be installed on the following on Pro Alarm analysers originally equipped with these power configurations:
- Wall plug-in
- DIN power
The SMART START Module is not compatible with any Pro Alarm analyzer powered
by battery. Any internal power source (i.e., battery) must be removed and
replaced with either DIN Rail or wall plug-in power.
By installing the SMART START Module, the analyzer will automatically power on
when external equipment power is detected. After three (3) seconds the On/Off
button will return to its normal function.
The SMART START Module can be purchased separately and installed by existing
Pro Alarm analyzer owners, or the analyzer can be returned by the customer to
the Nuvair production facility in California (USA) for installation.
WARNING
This manual describes how a customer can install the SMART START Module to
existing Pro Alarm handheld or panel mount analysers. To complete this
installation, the customer must have a rudimentary knowledge of, access to,
and the ability to use and understand:
- Electronic components
- Small hand tools
- Solder
- Soldering iron, and
- Electrical tape.
DANGER
Failure to properly install the SMART START Module will result in the Pro
Alarm analyzer not functioning properly. An improperly operating analyzer will
not alarm and will not alert users that dangerous gas conditions exist.
Exposure to dangerous gases will result in significant injury or death.
2.1 Module Compatibility
The SMART START Module is compatible with many Nuvair Pro Alarm analyzers
power by wall plug-in or DIN power. This installation manual covers the
installation of the SMART START Module on the following analyzers:
- Pro CO2 Alarm Handheld (9616)
- Pro CO2 Alarm Panel Mount (9617)
- Pro CO Alarm Handheld (9625)
- Pro CO Alarm Panel Mount (9617)
- Pro He Alarm Handheld (9627)
- Pro He Alarm Panel Mount (9628)
- Pro H2O Alarm Panel Mount (9607)
- Pro N2 Alarm Handheld (9613)
- Pro N2 Alarm Panel Mount (9614)
- Pro O2 Alarm Handheld (9611)
- Pro O2 Alarm Panel Mount (9462)
- Pro O2 Alarm Panel Mount Remote (9460)
The SMART START Module is also compatible with the following multi gas and specialty container analysers. Due to the complexity of these unique and specialized Pro Alarm analysers, Nuvair does not offer written guidance to customers on how to install the SMART START Module. Customers are advised to return the following analysers to the Nuvair production facility in California (USA) for installation of SMART START Module/s:
- Pro 4 Warn Alarm (9604)
- Pro Trio Multigas Alarm (9603)
- Pro CO Analyzer with Low Pressure Alarm (9621.5)
- Pro CO Analyzer with Low Pressure & High Temperature Alarms (9621)
- Pro Trimix Alarm (9609-LB)
- Pro He-O2 Trimix (9602)
- Pro Duo Alarm (DUO – all versions)
Installation of the SMART START Module
The images shown in this manual are of a handheld Nuvair Pro Alarm analyzer. Within the context of a SMART START Module installation, the panel mount Nuvair Pro Alarm analyzer version is the same as the handheld version except for the aluminum overlay faceplate.
3.1 Remove Faceplate Screws
- Power off internally powered analyzer with On/Off button.
- Disconnect power supply from externally powered analyser.
- Remove any additional external parts, such as the flow adapter cap, external relay connections, etc.
- Remove four (4) faceplate screws using a small Phillips screwdriver (Figure 2).
- Separate faceplate from back cover.
3.2 Remove PCB and Display Screen from Face plate
- Unplug red and black power wire (Molex connector) from faceplate PCB.
- Set back cover and battery (if any) aside.
- Place faceplate face-up (with display screen facing you) on a flat, clean, dry, nonmetallic surface (Figure 3).
- Remove two (2) bottom screws below display screen using a small flat blade screwdriver and a 1/4 in nut driver.
3.3 Replace Screws
- Replace the two (2) 1/2 in screws (just removed) and install the 3/4 in screws supplied in your SMART START Module Kit.
- Make sure the two (2) white plastic spacers located between the analyzer PCB and faceplate stay in place as you replace both screws (Figure 4).
3.4 Reinstall Flange Nuts
Use a 1/4 in nut driver to reinstall the two (2) white plastic flange nuts
(Figure 5).
WARNING
Do not overtighten flange nuts. Damage to analyzer PCB or screws will occur.
WARNING
After securing analyzer board to plastic assembly, ensure analyzer control
buttons operate freely and do not bind to cover.
3.5 Locate PCB Soldering Connections
- With the PCB and display screen properly oriented, locate the left screw and flange nut you just reinstalled.
- To the right of the flange nut are two (2) posts that you will solder the SMART START Module power supply to (Figure 6).
WARNING
If you do not have rudimentary knowledge of electronics, which includes
possession of and no-how of how to properly solder sensitive electronic
connections, STOP! Package up all parts and contact Nuvair for assistance.
DANGER
Your failure to properly install the SMART START Module will result in the Pro
Alarm analyzer from functioning properly. An improperly operating—or non-
operating analyzer will not alarm and will not alert users that dangerous
compressor gas conditions exist. Exposure to dangerous gases will result in
significant injury or death.
3.6 Solder SMART START Module PowerSupply Wires
Using a soldering iron and solder, solder green and yellow SMART START
Module power supply wires to analyzer PCB posts (Figure 7).
3.7 Connect SMART START Module
- Using existing screw holes, place SMART START Module on over screws.
- Using a 1/4 in nut driver, screw the two (2)white plastic flange nuts supplied with your SMART START kit onto screw.
- Connect green and yellow power supply to SMART START Module.
- Connect red and black wire jumper as pictured (Figure 8).
WARNING
Do not overtighten flange nuts. Damage to analyzer PCB or SMART START Module,
or screws will occur.
Modifying Internal Power Pro Alarm Versions
If your Pro Alarm analyzer is equipped with an internal power supply (lithium [Li-ion] battery), the battery must be removed, and the power connection to the battery stubbed off (Figure 9).
- Disconnect battery from external power supply.
- Remove battery from case.
- Stub off open battery connector with shrink tubing or electrical tape.
- Connect external battery supply wiring to open connector on SMART START Module labeled “IN” (Figure 10).
- From the connector labeled “OUT” connect to the original power connector on the green PCB.
- Plug green and yellow connector in SW position on SMART START Module (Figure 11).
- Reassemble.
Additional Parts
SKU | Identification |
---|---|
BPM 6 | Pro Analyzer Flow Adapter Cap |
5463K369 | Hose Barb Fitting for Pro Analyzer Flow Adapter Cap |
51075K246 | Tubing for Nuvair Pro Analyzer Flow Adapter Cap |
PC | Wall Socket for SP-C2S3A Charger |
120-052 | 12-volt Power Supply for Nuvair Pro Alarm Analyzers without Battery |
Nuvair Standard Warranty
Unless otherwise specified, materials (parts and equipment) supplied by Nuvair
are covered by a one (1) year warranty period from the date of delivery to
customer. “Date of delivery” is determined by shipping or proof of receipt
documentation.
Should the customer note any flaws and/or defects, it must be reported in
writing to Nuvair within two (2) months of discovery or the warranty shall be
rendered null and void.
Nuvair shall repair or replace materials it acknowledges to be faulty during
the warranty period.
In replacing faulty material, Nuvair shall not be liable for any other
expenses sustained by the dealer or its customer, such as presumed damage
(present or future), lost earnings or fines.
Warranty only covers flaws and faults that occur where materials are used
properly in compliance with the instructions contained in the User Manual and
where periodic maintenance is carried out. The warranty does not cover faults
caused by improper use of materials, exposure to atmospheric conditions (rain,
etc.) or damage during transport.
Materials subject to wear and those subject to periodic maintenance are not
covered by warranty and are to be paid for by the customer in full. Routine
and unscheduled maintenance must be carried out in compliance with User Manual
instructions. If required work is not covered by the User Manual or if
technical assistance is required, please contact Nuvair directly by email or
phone. Nuvair cannot be held liable for any delays or failure to execute work.
Use of genuine OEM (original equipment manufacturer) parts and lubricants (if
applicable) are recommended to maintain a product warranty. If a product
failure is deemed a result of using nonOEM part and/or lubricant, the warranty
will be considered null and void.
Nuvair is not liable for any damage or malfunctions caused by work carried out
on materials by unauthorized personnel.
Warranty is rendered null and void if materials are tampered with or if work
(other than routine maintenance) is carried out by personnel who have not been
authorized by Nuvair.
Faulty equipment due to flaws in design, workmanship, or materials used shall
be repaired or replaced free of charge by Nuvair at its California production
facility. Transportation costs, delivery of spare parts, and/or any materials
subject to normal wear, are the responsibility of the customer.
Should on-site warranty work be required on the customer’s premises, Nuvair
repair personnel travel, accommodations, and per diem costs of shall be the
responsibility of the customer.