Nuvair SS-PCB-GEN2 Failsafe Module Kit Installation Guide
- June 9, 2024
- Nuvair
Table of Contents
Nuvair SS-PCB-GEN2 Failsafe Module Kit
WARNING
This Operation Manual contains important safety information and should always
be available to those personnel operating this equipment. Read, understand,
and retain all instructions before operating this equipment to prevent injury
or equipment damage.
Every effort was made to ensure the accuracy of the information contained
within. Nuvair, however, retains the right to modify its contents without
notice.
Under Nuvair’s system of continuous improvement, certain components may be
updated or changed as higher quality or more efficient parts and assemblies
become available.
Nuvair will make every effort to update manuals as parts and functional
aspects change. However, the look or location of components on your product
may differ from those in this manual if improvements have been made that do
not affect functionality or operational procedures.
Units pictured may also be equipped with different options than those on your
product. In this case, the basic operational and maintenance guidelines will
still apply.
If you have problems or questions after reading the manual, stop and call
Nuvair at +1.805.815.4044 for information.
Introduction
This manual will assist you in the installation of the FAILSAFE Module (SKU
SS-PCB-GEN2) into existing handheld or panel mount Nuvair Pro Alarm analyzer
powered externally. This manual will also help customers understand the
function of Nuvair Pro Alarm analyzers equipped with a factory-installed
FAILSAFE Module.
The Nuvair Pro Alarm analyzer series are gas analyzers with alarms (visual
and/or audible) that activate when user-defined set points (low and/or high)
are violated.
Be sure to read the entire manual.
Symbol Conventions
This manual uses certain words and symbols to call your attention to conditions, practices or techniques that may directly affect your safety. Pay particular attention to information introduced by the following symbols or words:
SYMBOL | MEANING | DESCRIPTION |
---|---|---|
DANGER | Indicates an imminently hazardous situation, which if not avoided, |
will result in serious personal injury or death.
CAUTION| Indicates a potentially hazardous situation, which if not avoided,
could result in serious personal injury or death.
WARNING
| Indicates a potentially hazardous situation, which if not avoided, may
result in minor or moderate injury. It may also be used to alert against
unsafe practices.
NOTICE| Notifies people of installation, operation or maintenance
information which is important but not hazard-related.
Module Description
As originally designed, most Nuvair Pro Alarm analyzers are:
- Manually controlled by an On/Off button on the front of the analyzer control panel.
- Equipped with a relay that can activate or disable external devices.
Some customers with a compressor-integrated analyzer want the compressor to shut down automatically if analyzer power is interrupted or the analyzer goes into alarm. In response to customer requests, Nuvair developed the FAILSAFE Module. Adding the FAILSAFE printed circuit board (PCB) to your existing Pro Alarm analyzer will automatically disable compressor power if:
- Analyzer power is interrupted, or
- Analyzer goes into alarm.
The FAILSAFE Module can be installed on the following on Nuvair Pro Alarm analyzers originally equipped with these power configurations:
- Wall plug-in
- DIN power
The FAILSAFE Module is not compatible with any battery-powered Pro Alarm
analyzer.
By installing the FAILSAFE Module, the analyzer will automatically power on
when compressor power is turned on.
The FAILSAFE Module can be purchased separately by existing Pro Alarm analyzer
owners, or the analyzer can be returned by the customer to the Nuvair
production facility in California (USA) for installation.
WARNING
This manual describes how a customer can install the FAILSAFE Module to
existing Pro Alarm handheld or panel mount analyzers. To complete this
installation, the customer must have a rudimentary knowledge of, access to,
and the ability to use and understand:
- Electronic components
- Small Phillips screwdriver
- Solder, and
- Soldering iron.
DANGER
Failure to properly install the FAILSAFE Module will result in the Pro Alarm
analyzer not functioning properly. An improperly operating analyzer will not
alarm and will not alert users that dangerous gas conditions exist. Exposure
to dangerous gases will result in significant injury or death.
Module Compatibility
The FAILSAFE Module is compatible with many Nuvair Pro Alarm analyzers power by wall plug-in or DIN power. This installation manual covers the installation of the FAILSAFE Module on the following analyzers:
- Pro CO2 Alarm Handheld (9616)
- Pro CO2 Alarm Panel Mount (9617)
- Pro CO Alarm Handheld (9625)
- Pro CO Alarm Panel Mount (9617)
- Pro He Alarm Handheld (9627)
- Pro He Alarm Panel Mount (9628)
- Pro H2O Alarm Panel Mount (9607)
- Pro N2 Alarm Handheld (9613)
- Pro N2 Alarm Panel Mount (9614)
- Pro O2 Alarm Handheld (9611)
- Pro O2 Alarm Panel Mount (9462)
- Pro O2 Alarm Panel Mount Remote (9460)
The FAILSAFE Module is also compatible with the following multigas and specialty container analyzers. Due to the complexity of these unique and specialized Pro Alarm analyzers, Nuvair does not offer written guidance to customers on how to install the FAILSAFE Module. Customers are advised to return the following analyzers to the Nuvair production facility in California (USA) for installation of FAILSAFE Module/s:
- Pro 4 Warn Alarm (9604)
- Pro Trio Multigas Alarm (9603)
- Pro CO Analyzer with Low Pressure Alarm (9621.5)
- Pro CO Analyzer with Low Pressure & High Temperature Alarms (9621)
- Pro Trimix Alarm (9609-LB)
- Pro Duo Alarm (DUO – all versions)
Installation of the FAILSAFE Module
WARNING
If you do not have rudimentary knowledge of electronics, which includes the
ability to properly solder sensitive electronic connections, STOP! Package up
all parts and contact Nuvair for assistance.
DANGER
Your failure to properly install the FAILSAFE Module will result in the Pro
Alarm analyzer from functioning properly. An improperly operating analyzer
will not alarm and will not alert users that dangerous compressor gas
conditions exist. Exposure to dangerous gases will result in significant
injury or death.
Step 1: Gain access to the Pro Analyzer enclosure and remove Pro Analyzer circuit board (pictured).
Step 2: Remove FAILSAFE parts from shipping bags.
Supplied screws and spacers.
FAILSAFE circuit board.
Step 3: Solder green and blue “switch” (SW) wires onto Pro Analyzer circuit board.
Step 4: Connect circuit boards together with provided screws and spacers (6 pieces total).
Step 5: Connect wires to FAILSAFE Module as shown.
Assembled unit.
Additional Analyzer Parts
SKU: Identification
- BPM6 Pro Analyzer Flow Adapter Cap
- 5463K369 Hose Barb Fitting for Pro Analyzer Flow Adapter Cap
- 51075K246 Tubing for Nuvair Pro Analyzer Flow Adapter Cap
- PC Wall Socket for SP-C2S3A Charger
- 120-052 12-volt Power Supply for Nuvair Pro Alarm Analyzers without Battery
Nuvair Standard Warranty
Unless otherwise specified, materials (parts and equipment) supplied by Nuvair
are covered by a one (1) year warranty period from the date of delivery to
customer. “Date of delivery” is determined by shipping or proof of receipt
documentation.
Should the customer note any flaws and/or defects, it must be reported in
writing to Nuvair within two (2) months of discovery or the warranty shall be
rendered null and void.
Nuvair shall repair or replace materials it acknowledges to be faulty during
the warranty period.
In replacing faulty material, Nuvair shall not be liable for any other
expenses sustained by the dealer or its customer, such as presumed damage
(present or future), lost earnings or fines.
Warranty only covers flaws and faults that occur where materials are used
properly in compliance with the instructions contained in the User Manual and
where periodic maintenance is carried out. The warranty does not cover faults
caused by improper use of materials, exposure to atmospheric conditions (rain,
etc.) or damage during transport.
Materials subject to wear and those subject to periodic maintenance are not
covered by warranty and are to be paid for by the customer in full. Routine
and unscheduled maintenance must be carried out in compliance with User Manual
instructions. If required work is not covered by the User Manual or if
technical assistance is required, please contact Nuvair directly by email or
phone. Nuvair cannot be held liable for any delays or failure to execute work.
Use of genuine OEM (original equipment manufacturer) parts and lubricants (if
applicable) are recommended to maintain a product warranty. If a product
failure is deemed a result of using non-OEM part and/or lubricant, the
warranty will be considered null and void.
Nuvair is not liable for any damage or malfunctions caused by work carried out
on materials by unauthorized personnel.
Warranty is rendered null and void if materials are tampered with or if work
(other than routine maintenance) is carried out by personnel who have not been
authorized by Nuvair.
Faulty equipment due to flaws in design, workmanship, or materials used shall
be repaired or replaced free of charge by Nuvair at its California production
facility. Transportation costs, delivery of spare parts, and/or any materials
subject to normal wear, are the responsibility of the customer.
Should on-site warranty work be required on the customer’s premises, Nuvair
repair personnel travel, accommodations, and per diem costs of shall be the
responsibility of the customer.
Warranties are nontransferable.
Nuvair
Phone +1.805.815.4044
Fax +1.805.486.0900
Address 1600 Beacon Place
Oxnard, CA 93033 USA
Email info@nuvair.com
Website www.Nuvair.com
Rev. 01.2023
SKU SS-PCB-GEN2