KANOMAX 3888 Handheld Particle Counter 6 Channel User Guide

June 9, 2024
KANOMAX

3888 Handheld Particle Counter 6 Channel
User Guide

3888 Handheld Particle Counter 6 Channel

Include

  • Main unit
  • USB cable
  • CD-ROM: USB driver, Remote software, User manual
  • Li-ion battery x 2 (installed in the unit)
  • Zero filter with tubing
  • Inlet protective cap
  • AC adapter

Functions

Note: Please do not cover or insert anything into the threaded exhaust port on back of the instrument
General setting
Opens the menus to set the unit, calendar, data, sound, language and to display instrument information
Measurement setting
Opens the menu to set the mode, alarm, save, remote and temperature/humidity
Preset
Opens the menu to set or select up to 18 Presets
History
Views the historical data or measurements
Error
Tap the Error key to display an error description
Differential
Opens the Differential display on the Main screen
Purging
Prior to using the instrument in a clean environment such as cleanrooms, it is recommended to purge the instrument with the provided zero count filter.

  1. Attach the provided zero count filter to the inlet
  2. Turn the power on and press the Start button. Perform a continuous measurement for approximately 15 minutes until the measuring count value is suitable for 10 seconds or more.
  3. After measurement, remove the zero count filter from the inlet

Measurement

  1. Select measurement mode Measurement modes
    Repeat
    Repeats the measurement of configured sampling time and cycle
    Single
    Performs a measurement of set sampling time only once
    Continuous
    Continues a measurement until pressing the Stop button
    Calculation
    Calculates the average, standard deviation, maximum, and minimum after measurements
    ISO
    Performs measurements in accordance with ISO 14644
    GB
    Performs measurements in accordance with Chinese National Standard

  2. Select measurement mode Set up location and sampli Location
    Sets measurement locations by number Ranges: from 1 to 99
    Sampling time
    Sets the sampling time of 1 cycle Ranges: from 6 sec to 99 min 59 sec

  3. Start measurement ① Go to main screen by pressing Home button
    ② Press start button to start measurement * Measurement starts after 10 seconds warm up time

General Setting

Select on the Main screen to set the unit, calendar, data, language, and sound
Unit
Select one of the following 3 options of units to reflect to the readings displayed on the Main screen and Measurement results.
Calendar
Set the date and time
Data
Confirms the information and perform operation on the entire measured data
Language
Selects a language from English, Japanese, Chinese, Spanish
Sound
Sets the operation sound ON/OFF
Information
Displays the instrument information Latest calibration date and firmware version