RadonAway 28003 Fan Selection Diagnostic Aid Kit Instructions

June 9, 2024
RadonAway

RadonAway 28003 Fan Selection Diagnostic Aid Kit

RadonAway-28003-Fan-Selection-Diagnostic-Aid-Kit-
PRODUCT

INTRODUCTION

This diagnostic procedure is intended to assist in system design and fan selection for subslab depressurization (suction) radon mitigation systems. It provides a simple method by which one may derive information about the natureof the material below a concrete slab floor. It requires only one slab penetration and can be easily performed unassisted by one person in a few minutes. For maximum information it should be used in conjunction with communications testing; however, where this is not feasible or desired it may be used as a stand-alone test.

KIT CONTENTS

  • One (1) Dwyer 0-100” magnehelic gauge (magnehelic) with accessories
  • One (1) Dwyer magnehelic carrying case with accessories
  • One (1) set of instructions

OTHER REQUIRED EQUIPMENT & MATERIALS
Vacuum Cleaner with 1 1/4” hose and suction wand. It is recommended that a canister-type vacuum cleaner capable of approximately 50” of water column suction be used (Eureka Mity Mite 3 HP or equivalent). The vacuum cleaner should use standard 1 1/4” vacuum hose.

ASSEMBLY

MAGNEHELIC & CARRYING KIT
See Dwyer instructions accompanying 0-100″ magnehelic. Using 1/8″ NPT pipe plugs provided plug both high and low pressure tap holes on back of magnehelic. Attach the magnehelic to the metal plate provided with the carrying kit using No. 6-32 machine screw also provided with the carrying kit.

PRESSURE CONNECTIONS
There are two 1/8″ NPT brass pipe adaptor plugs to rubber tubing (small brass nipples) provided with the magnehelic. Insert one into low pressure tap hole on side of magnehelic. High pressure tap hole should be left open.

VACUUM CLEANER CONNECTION
The second 1/8″ NPT brass pipe adaptor plug to rubber tubing (small brass nipple) is to be inserted in the vacuum line of the vacuum cleaner that is to be used. This is most easily accomplished by drilling a 3/8″ hole in one of the vacuum cleaner “wands”. The brass nipple may then be screwed tightly into the hole (do not overtighten). To an airtight seal silicone caulk may be used.

GNOSTIC TESTING

REFERENCE SUCTION

Connect one end of rubber tubing supplied with carrying kit to the brass nipple on the low pressure side of the magnehelic. Connect the other end·of rubber tubing to the brass nipple on the vacuum line of the vacuum cleaner. Turn on the vacuum cleaner. To determine
accomplished by placing one’s hand over the opening). Note the suction indicated on the
Note: This reference suction should be determined each time a diagnosis is performed since the maximum suction of any given vacuum cleaner will vary with many factors. To maintain consistency it is recommended that a new clean vacuum cleaner bag be used.

TESTING THE SUBSLAB MATERIAL
Connect the vacuum line of the vacuum cleaner to a penetration in the concrete slab as shown in the figure below. It is important to ensure a good seal between the vacuum line and the slab.
Turn on the vacuum cleaner. Make sure there is no leakage of outside air into the vacuum line and that all the air is coming from the subslab. Record the suction indicated on the magnehelic

VACUUM CLEANER SUCTION TESTRadonAway-28003-Fan-Selection-Diagnostic-Aid-
Kit-FIG 1

ANALYSIS

Calculate the percentage of maximum vacuum cleaner suction that is encountered at the slab equation is:

Percentage of Maximum Suction= (PHole/PMax)x 100

Having obtained this percentage the chart on the next page may be used to provide guidance on likely subslab material, communication and reasonable fan choices.
Note: The chart starts at 25%. This is because at open flow the magnehelic will register roughly 25% of PMax-) The chart is intended to provide general guidelines on fan selection. Actual fan selection will be influenced by other factors such as cost, availability, appearance, building restrictions, etc.

POTENTIAL PITFALLS

One must be somewhat careful in drawing conclusions and designing the mitigation system based on this diagnostic procedure. Other factors may cause the results to be misleading. These include:

  1. Slab condition
  2. Foundation structure such as subslab strip footers
  3. Variations in subslab permeability across the slab
  4. Air sources such as drain fields
  5. Ground water

Communications testing can often detect the presence of some of these conditions. Good system design should deal with all factors and allow the mitigator to price the system appropriately.

WARRANTIES AND REPRESENTATIONS

Subject to applicable consumer protection legislation, RadonAway, Inc. warrants that this test kit will be free from defective materials and workmanship for a period of (1) year from the date of purchase. Warranty is contingent on use in accordance with the instructions provided. This warranty does not apply where repairs or alterations have been made or attempted by others; or the equipment has been abused or misused. Warranty does not include damage in shipment unless the damage is due to the negligence of RadonAway, Inc. To make a claim under these limited warranties, you must return the defective item to RadonAway, Inc. with a copy of the purchase receipt. All other warranties, expressed or written, are not valid.
This diagnostic aid kit is intended for use by trained Radon professionals. These instruction
are the sole responsibility of the user. RadonAway makes no representations or guarantees as to the accuracy of such interpretation.

INSTRUCTIONS FAN SELECTION DIAGNOSTIC AID KIT

FAN SELEC TION CHARTRadonAway-28003-Fan-Selection-Diagnostic-Aid-Kit-FIG
2 PERCENTAGE OF MAXIMUM SUCTION OF VACUUM

3 Saber Way, Ward Hill MA 01835 | Radonaway.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

RadonAway User Manuals

Related Manuals