KENWOOD VT-175 2-Channel AC Voltmeter Instruction Manual
- June 9, 2024
- KENWOOD
Table of Contents
KENWOOD VT-175 2-Channel AC Voltmeter Instruction Manual
GENERAL
The Model VT-175 is dual-channel electronic voltmeter sets having two
independent A C voltmeters in a single cabinet. Two-pointer meter allows easy
dual measurements and inter-channel comparisons.
MODE L VT-175 is wideband, high sensitive voltmeter for use anywhere voltage
measurements are made, with a remote control unit. The easy-to-use voltmeter
provides frequency response as wide as 5Hz to 1 MHz, measurable range from 1mV
to 300V, and input impedance of 10MO.
FEATURES
-
a . Two power source circuits, positive and negative, are provided for very stable DC bias and recover from any overload in a very short time.
-
b. Small-sized compact construction is easy to carry and large-sized two-pointer meter is easy to read.
-
c. Four-scale dial plate and red and black pointers allow quick, and accurate voltage reading.
-
d. Number of adjusting trimmers are minimized with amplifiers that are precisely calculated in open loop gains and are negative feedbacked through metalfilm resistors as accurate as 1% Amplifier characteristics therefore are from secular change.
-
e. Input impedance is as high as 10 M O on al l ranges and its parallel capacitance less than 45 pF
-
f. Indirect attenuator switching by relay and FE T (field-effect transistor) switch provides higher reliability, signal-to-noise ratio, and inter-channel crosstalk than most previous direct method of rotary switch.
-
g.- Ranges can be remotely selected using the RT-6 1
(option). -
h. Two channel range setting can be made either individually or together (in interlocked manner).
-
i. A ground mode selector may float measuring circuit above ground.
SPECIFICATIONS
[Meter Section]
-
Measurable voltages : 1 mV-300V in 12 ranges :
1 mV, 3mV , 10 mV, 30mVr 100mV,
300mV, 1V, 3 V, 10 V , 30 V , 100V and
300V. full scales. -
d B : -80~+50dB(0dB = 1V) .
-
dBm : – 80~+52dB m (0dBm= 1 mW,600n).
-
Error : Within ± 3% of full scale at 1kHz.
-
Frequency response : ± 10% at 5Hz— 1 MHz,
± 5 % at 10Hz-500kH z and
± 3 % at 20 Hz-200kH z as
± 2 % at 30Hz~100kH z referenced to 1 kHz response. -
Input impedance : 10MO ± 5%, with less than 45pF parallel capacitance.
-
Durable input voltage: 500V(D C +A C peak) at 1V—300V range 100V(DC+A C peak) at 1mV~300m V range
-
Stability : Within ± 0.5% of full scale for i 10% line voltage fluctuation.
-
Residual voltage : Less than 20 ^ V with input shorted on 1 mV range
-
Crosstalk, Individual: Less than -80dB with other input terminated with 600O
-
Inetrlock : Less than -50dB with other input terminated with 600O.
[Amplifier Section]
- Gain : Approx. 60dB.
- Output voltage : 1Vrms(Ful l scale) ±20%
- Output impedance: 600X1+ 20%
- Distortion : Less than 1% at full scale. (Rated by signal-to-noise ratio in 1 mV and 1V ranges.)
- Signal-to-noise ratio: Over 40dB at full scale.
- Frequency response: Within ± 3dB at 5 Hz —500kHz
[Environmental]
-
Coefficient: ± 0.08%/ °C
-
Temperature (within specification) : 10~40°C
(ful l operation) : 0~50°C -
Relative humidity : Less than 80%
[Power Supply Section]
-
Line voltage : 100,120, 220V A C + 10%,
216 – 250V AC , 50/60 Hz -
Power consumption: Approx. 7W
-
[Dimensions] W x 190(212) H x 238(268) Dmm.
Values in ( ) include protrusions -
[Net weight] 3 kg
[Accessories ]
-
Power cable : 1 pc.
-
Input cable : BNC cord, 2pes.
-
Replacement fuse : 0.3A 2pes.
0.5 A 2pes. -
Instruction manual: 1 copy.
FUNCTIONAL CONTROLS
Figure 1 – Front Panel View
REFER- ENCEy«. | IDENTIFICA- TION | DESCRIPTION |
---|---|---|
1 | Power-on indicator (LED) | |
2 | ON OFF | on-offswitch |
3 | RANGE-CH 1/2 | Channel 1 and 2 atten- uator selector |
4 | INPUT 1 | channel 1 input connector |
5 | CH 2 SELECTOR | Individual-interlock range setting selector |
6 | CH2INPUT | Channel 2 input connector |
7 | RANGE-CH 2 | Channel 2 attenuator selector |
8 | GND MODE | Negative input floating switch |
9 | Dual-pointer meter | |
10 | Meter zero-adjustCH1(left), CH2(right) |
Figure 2 – Rear Panel View
REFERE-NCE fb. | IDENTIFICA- TION | DESCR IPTION |
---|---|---|
11 | GND | Casing grounding post |
12 | CH 2 OUTPUT | Channel 2 monitor output connector |
13 | Line vol tage selector and fuse holder | |
14 | Power cable connector | |
15 | CH 1 OUTPUT | Channel 1 monitor output connector |
16 | CONT. INPUT | Remote control mounting hole |
OPERATION
Set-up
-
Press the ON/OF F Pushswitch 2 in. The Power-on indicator 1, a light-emitting diode, wil l light, indicating that power has been applied.
-
Plug the signal calbes into the CH 1 INPU T and CH 2 INPUT connectors 4 and 6
-
CH 2 SELECTOR 5 , with the switch released( ), takes single mode in which Ranges C H 1 and C H 2 are changed over uging Attenuators CH 1/2 3 and CH2 7 , respectively.’ . ‘ •
With the switch pressed ( ), the SELECTO R takes inter lock mode in which both Ranges C H 1 and CH 2 are simultane-ously changed over using Attenuator CH 1/2 -
In the interlook mode, Attenuator CH2 7 is no longer operable to select a Range. Select either single or interlook mode according to your particular measurement.
-
Set the RANGE-C H 1/2 selector 3 and the RANGE – C H 2 selector 7 to “300V/ To measure an A C voltage superimposed on a DC voltage, be sure to set the selectors to the positions before connecting the cables to points to be measured. r erwise , the DC voltage causes a high surge, which could burn voltmeter.
-
- GND MODE ® is effected with the switch released
( ) r in which both CH 1 and CH 2 are connected to the grounding terminal of the casing, taking GND state, with the switch pressed ( ), the grounding of CH 1 and CH 2 is floated to OPE N state in which you move the earth point to the signal source side. Normally use the device in the GND state.
- GND MODE ® is effected with the switch released
-
Connect the cables to points to be measure.
-
Turn the selectors 3 and 7 for proper ranges until each Meter pointer switngs over one third of the fullscale.
-
OUTPUT S CH 2 12 and CH 1 15 provide output signals by amplifying input signals in such a degree as the signal amplitude of a meter fullscale becomes 1 Vrm s no matter which range is set.
-
Read the Meter.
-
Voltage Scales See Figures3
There are two black voltage scales : a scale ® graduated 0 to 10 and a scale © 0 to 3. When the RANG E selector 3 is at “1 V,fo r example, the division 10 on the scale ® indicates 1V. At the “300 m V ” position, the division 3 on the scale ® indicates 300 mV. Similarly , the other RANG E selector positions show their full scale values. -
dB Scale
In general, the dB values are expressed in dBV which is a unit referenced by OdB equal to 1 V. The division 10 at the scale A corresponds to 0 dB on the dB scale, which is a voltage ratio scale. Since the RANG E selector has 12 range positions in steps of 10 dB, the voltage ratio of 1 m V to 300V is 110 dB attenuation.
Assume a reference voltage level on the scale A be 1V with the RANG E selector at “1 V,” a given voltage can be read as low as -60dB(1mV)b y turning the selector downward. Further, as the scale A allows reading to —20dB (0.1 mV) , you can continuously measure the voltage ratio as high as -80dB(0. 1 mV to 1V) . The read of -80d B means a signal-to-noise ratio of around 10dB. Also, the dB scale allows continuous measuring up to +50dB( 1 to 300V) by turning the selector upward -
dBm Scale
In general, voltmeters have a reference level division of OdBm equal to 0.775V ( 1 mW power) induced across a 800O resistance load. Therefore, the dBm scale is available to measure a power level referenced to 0 dBm, with the impedance of the given power circuit being 60012 pure resistance. Where measured across specific resistance loads, for example, 10kO , other than the 600O load, the levels are sometimes expressed in dBs.
Figure 3 – Meter Scale Graduation
How to Use Remote Control
The RT-6 1 option, a remarkable feature of the VT-175 , provides a remote control capability. To use the option, remove the case and the rear-panel blind plate, install the connector supplied with the RT-61 , then plug the RT-6 1 into the 16-pin connector on the print circuit board. For details, refer to the RT-6 1 Instruction Manual.
APPLICATIONS
The basic use of your voltmeter is to measure sinusoidal wave voltages as an A C voltmeter. In addition, it provides a wide variety of applications as described below.
Amplifier Gain Measurement
Voltmeter is capable of measuring an amplifier gain, which is a signal magnification from point A to B in Figure 4 with a signal generator connected to the amplifier input. If the measured signal levels at points A and B are a and b dB, respectively, then the gain is (b-a ) dB.
Figure 4 – Amplifier Gain Measurement
Also the method is applicable for measuring an open loop gain of each
amplifier circuit with the negativefeedback signal leaves applied. Further,
the method c be used for measi”th e frequency responses of given circuits by
changing the signal generator frequency.
Voltmeter is capable of directly reading a gain between two points as having
two independent voltmeters. A s an example, let us calculate the open loop
gain (from point A to B ) of the negative-feedback amplifier shown in Figure
5. Assume that the measured level at point A be+ 1.5dB on the -60d B range of
the channel 1 and that of point B -4d B on the -MOdB range of the channel 2.
The gain from point A to B is then ( + 10dB-4dB)-(-60dB+1.5dB ) = 64.5dB.
Figure 5 – Negative Feedback Amplifier Gain Measurement
Voltage Calculations of Some Special Waveforms
An voltmeter of absolute-mean value indication type, reads root-mean-square
values of sinusoidal wave inputs.
Also it deflects the pointers in proportion to the absolutemean value of a
given input wave.
If the form factor ( = root-mean-square value/absolutemean value) of the input
wave and the crest factor ( = peak value/root-mean-square value) are known ;
then the root-mean-square value and peak value can be calculated as follows.
Assume the Meter read x .
- Absolute-mean value=
- Root-mean-square value = (Absolute-mean value) x (For m factor).
- Peak value= (Root-mean-square value) x (Crest factor).
For rectangular waves, their form factor and crest factor are unity (1).
- Aboslute-mean va!ue=
- Root-mean-square value = 0.9 x .
- Peak value= 0.9x .
For sawtooth waves, their form factor is and the crest factor
- Absolute-mean value=
- Root-mean-square value
- Peak value=
Ways To Better Dual-Channel Voltmeter Operation
The two input channels and two meter pointers arranged specifically in
Voltmeter are most useful for measuring an electronic equipment having two
signal sources.
F or example, Voltmeter is far more convenient for measuring stereo left and
right channel characteristics than using two units units of usual single-
channel voltmeter.
- **Frequency response
Figure 6 – Stereo Amplitier Frequency Response Measurement
**CH2 SELECTO R swich to pressed (interlock position) as most stereo
amplifiers are not deviated vertually in the left and right channel levels.
Set the G ND MOD E switch to “GND.” If the given stereo amplifier output has
no connector common to the left and right channels, then turnt he switch to
“OPE N “an d connect the GND post on the rear panel of voltmeter to that of
the stereo amplifier. ‘
- Crosstalk characteristic
Figure 7 – Stereo Amplifier Crosstalk Measurement
Crosstalk measurement is needed if there is a large difference between the left and right channel output levels. CH 2 SELECTO R Pushbutton to release(individual position). Set the RANGE-C H 1/2 selector and RANGE – C H 2 selector to proper posi-tions for ease of read. Set the GND MOD E switch for no adverse effect by hum as explained in above Example (1), the “Frequency response.”
MAINTENANCE
Replacing the fuse (see Figure8)
REPLACIN G FUS E
When a fuse goes out, the device is no longer operable.
If gone out, check for a trouble cause, remedy it, remove the cap of the fuse
holder on the rear panel using a minus screw driver and take it out and
replace the fuse with a new one.
When replacing fuse, be sure to check the capacity of a new fuse for a specified value as follows:
- A C 100V, 120V : 0.5A
- A C 220V, 240V : 0.3A
CHANGEOVER O F VOLTAGE IN SWITCHING REGULATOR
Remove the fuse holder on the rear side of the device set, using a minus surew
driver, adjust your preferred voltage indication with • mark and plug the
holder in. For changing 100V or 120V to 220V or 240V, change the fuse of 0.5A
to that of 0.3A.
Figure 8 – Replacing the fuse
ALIGNMENT
Voltmeter was presicely aligned at the factory. However it can be aligned through the adjust holes located on the side with leaving is housed in the casing, if required. In realignment, firs t adjust the line to the rated voltage, use a precisely calibrated measuring instruments, and proceed as follows.
Preset Controls on Side Panel (see Figure9)
The following controls, placed on the Main PC board, are to be preset on the right-hand side panel.
IDENTIFICA-TION | DESCR IPTION |
---|---|
TC101 | First Attenuator preset trimmer capacitor |
VR 101 | First Attenuator preset variable resistor |
VR 102 | Meter Amplifier gain preset variable resistor |
- Before turning the ON/OF F switch “ON,” adjust the Meter zero-adjust screws (right for channel 1 and left for channel 2) on the front panel until the pointers correctly indica “0 “
- Set the CH 2 SELECTO R pushbatton to release (individual position).
- Connect a voltmeter calibrator 1 kHz (o r 400 Hz ) output to the INPU T connector®.
- Set the RANGE-C H 1/2 selector® on your VT-165 to the “30mV ” range and set the voltmeter calibrator output voltage to 30mV.
- Adjust VR102 until the pointer swings to the fullscale.
- In turn, set the RANGE-CH1/ 2 selector® to the “10 V range and set the voltmeter calibrator output voltage to 10V.
- Adjust VR101 until the pointer swings to the fullscale.
- Disconnect the voltmeter calibrator and connect a wideband signal generator to the CH 1 INPU T connector®.
- Set the signal generator frequency to 1 kHz and set the RANGE-C H 1/2 selector @ on voltmeter to the ” I V range.
- Adjust the signal generator output level until the pointer swings to the fullscale.
- Change the signal generator frequency from 1 kHz to 50 kHz.
- Adjust TC101 until the pointer swings to the fullscale.
- Repeat Steps 3 through 10.
- For channel 2 alignment, similary proceed with Steps 3 through 13.
Flgure 9 – Pres Controls on Side Panel
CAUTIONS FOR USE
- Your voltmeter wil l work just when the ON/OF F switch is pressed in. For accurate measurements, wait around five minutes for warm-up.
- Avoid placing your voltmeter where magnetic field and electric field are too strong.
- The input signal cable other than supplied should be low-capacitance shielded cable, or coaxial cable. For measurements at higher than audio frequencies, terminate the input connectors with 50 0 or 600O, resistor.
- The continuous maximum input voltage allowable for your voltmeter is 100V (DC+ACpeak) with the RANG E selector at “-60dB ” to “-10dB” . Ful l care should be observed in measuring high voltages.
- Be careful of a line noise and similar smal l noises as your voltmeter is a highly sensitive voltmeter.
- Do not leave your VT-175 at any of high temperature and humidity places for a long period of time.
- As having two channel independent voltmeter, your voltmeter is involved in selection of grounding point on some equipment udner measurement. Set the GN D MOD E switch to “OPEN ” and connect the GN D post of your voltmeter to a point where noise effect is minimized.
- The ON/OF F switch of this unit is designed to act on the secondary side of the power transformer.
The primary side of the transformer remains energized even if the ON/OF F switch is set to OFF . Therefore, if the unit is to be unused for a long time, the power cord should be unplugged from the receptacle outlet.