RIGOL HDO1000 Series Digital Oscilloscope User Guide
- June 9, 2024
- Rigol
Table of Contents
- Safety Requirement
- Document Overview
- General Inspection
- Product Overview
- To Prepare for Use
- Connecting to Power
- Touch Screen Gestures
- Accessing the Built-in Help System
- Parameter Setting Method
- Replacing the Fuse
- Remote Control
- More Product Information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RIGOL HDO1000 Series Digital Oscilloscope User Guide
Guaranty and Declaration
Copyright
© 2022 RIGOL TECHNOLOGIES CO., LTD. All Rights Reserved.
Trademark Information
RIGOL®is the trademark of RIGOL TECHNOLOGIES CO., LTD.
Software Version
Software upgrade might change or add product features. Please acquire the latest software version from RIGOL website or contact RIGOL to upgrade the software.
Notices
- RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
- RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company’s sole decision.
- Information in this publication replaces all previously released materials.
- Information in this publication is subject to change without notice.
- RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing, use, or performance of this manual, as well as any information contained.
- Any part of this document is forbidden to be copied, photocopied, or rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or require ent when using our products or this manual, please contact RIGOL.
E-mail: service@rigol.com
Website: http://www.rigol.com
Safety Requirement
General Safety Summary
Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly
- Only the exclusive power cord designed for the instrument and authorized for use within the destination country could be used
- Ensure that the instrument is safely grounded
- Observe all terminal ratings.
- Use proper overvoltage protection.
- Do not operate without covers.
- Do not insert objects into the air outlet.
- Use the proper fuse.
- Avoid circuit or wire exposure.
- Do not operate the instrument with suspected failures.
- Provide adequate ventilation
- Do not operate in wet conditions.
- Do not operate in an explosive atmosphere.
- Keep instrument surfaces clean and dry
- Prevent electrostatic impact.
- Handle with caution.
WARNING
Equipment meeting Class A requirements may not offer adequate protection to broadcast nser vices within residential environment.
Safety Notices and Symbols
Safety Notices in this Manual:
WARNING
Indicates a potentially hazardous situation or practice which, if not avoided,
will result in serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not avoided,
could result in damage to the product or loss of important data.
Safety Notices on the Product
-
DANGER
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately. -
WARNING
It calls attention to an operation, if not correctly performed, could result in potential injury or hazard. -
CAUTION
It calls attention to an operation, if not correctly performed, could result in damage to the product or other devices connected to the product.
Safety Symbols on the Product
-
Hazardous Voltage
-
Safety Warning
-
Protective Earth Terminal
-
Chassis Ground
-
Test Ground
Measurement Category
This instrument can make measurements in Measurement Category I.
WARNING
This instrument can only be used for measurements within its specified
measurement categories.
Measurement Category Definitions
- Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable. Thus, you must know the transient withstand capability of the equipment.
- Measurement category II is for measurements performed on circuits directly connected to low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
- Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring (including cables, bus-bars, junction boxes, switches and socket-outlets) in the fixed installation, and equipment for industrial use and some other Safety Requirement 2 Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. equipment. For example, stationary motors with permanent connection to a fixed installation.
- Measurement category IV is for measurements performed at the source of a low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Ventilation Requirement
This instrument uses a fan to force cooling. Please make sure that the air inlet and outlet areas are free from obstructions and have free air. When using the instrument in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
CAUTION
Inadequate ventilation may cause an increase of temperature in the instrument,
which would cause damage to the instrument. So please keep the instrument well
ventilated and inspect the air outlet and the fan regularly.
Working Environment
Temperature
- Operating: 0℃ to +50℃
- Non-operating: -30℃ to +60℃
Humidity
-
Operating Below +30℃: ≤90% RH (without condensation) +30℃ to +40℃: ≤75% RH (without condensation) +40℃ to +50℃: ≤45% RH (without condensation)
-
Non-operating:
Below +60℃: ≤90% RH (without condensation)
WARNING
To avoid short circuit inside the instrument or electric shock, never operate
the instrument in a humid environment.
Altitude
- Operating: below 3 km
- Non-operating: below 15 km
Protection Level Against Electric Shock
ESD ±8kV
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
WARNING
Ensure that no overvoltage (such as that caused by a bolt of lightning) can
reach the product. Otherwise, the operator might be exposed to the danger of
an electric shock. Installation (Overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is
applicable to equipment measurement terminals connected to the source circuit.
Among these terminals, precautions are done to limit the transient voltage to
a low level. Installation (overvoltage) category II refers to the local power
distribution level which is applicable to equipment connected to the AC line
(AC power).
Pollution Degree
Pollution Degree 2
Pollution Degree Definition
- Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs. The pollution has no effect. For example, a clean room or air conditioned office environment.
- Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment.
- Pollution Degree 3: Conductive pollution or dry nonconductive pollution that becomes conductive due to condensation occurs. For example, sheltered outdoor environment.
- Pollution Degree 4: The pollution generates persistent conductivity caused by conductive dust, rain, or snow. For example, outdoor areas.
Safety Class
Class 1 – Grounded Product
Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct
sunlight for long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
- Disconnect the instrument from all power sources.
- Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. Avoid having any water or other objects into the chassis via the heat dissipation hole. When cleaning the LCD, take care to avoid scarifying it.
CAUTION
To avoid damage to the instrument, do not expose it to caustic liquids.
WARNING
To avoid short-circuit resulting from moisture or personal injuries, ensure
that the instrument is completely dry before connecting it to the power
supply.
Environmental Considerations
The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC.
The equipment may contain substances that could be harmful to the environment or human health. To avoid the release of such substances into the environment andm avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information. You can click on the following link https://int.rigol.com/services/services/declaration to download the latest version of the RoHS & WEEE certification file.
Document Overview
This manual gives you a quick overview of the front and rear panel, user interface as well as basic operation methods of DHO1000 series.
TIP
For the newest version of this manual, download it from RIGOL official website (http://www.rigol.com).
Publication Number
QGA32103-1110
Software Version
Software upgrade might change or add product features. Please acquire the latest version of the manual from RIGOL website or contact RIGOL to upgrade the software
Format Conventions in this Manual
-
Key
The front panel key is denoted by the menu key icon. For example, indicates the “DEFAULT” key. -
Menu
The menu item is denoted by the format of “Menu Name (Bold) + Character Shading” in the manual. For example, Setup indicates the “Setup” sub-menu under the “Utility” function menu. You can click or tap Setup to access the “Setup” menu. -
Operation Procedures
The next step of the operation is denoted by “>” in the manual. For example, > Storage indicates first clicking or tapping and then clicking or tapping Storage. The front/rear panel connector is denoted by “Brackets + Connector Name (Bold)”, for example, [AUX OUT].
Content Conventions in this Manual
DHO1000 series digital oscilloscope includes the following models. Unless otherwise specified, this manual takes the 4-channel model DHO1104 as an example to illustrate the basic operation methods of DHO1000 series.
Model | Max. Analog Bandwidth | Analog Channels |
---|---|---|
DHO1072 | 70 MHz | 2 |
DHO1074 | 70 MHz | 4 |
DHO1102 | 100 MHz | 2 |
DHO1104 | 00 MHz | 4 |
DHO1202 | 200 MHz | 2 |
DHO1204 | 200 MHz | 4 |
General Inspection
-
Inspect the packaging
If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. -
Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative. -
Check the accessories
Please check the accessories according to the packing lists. If the accessories are damaged or incomplete, please contact your RIGOL sales representative
Recommended Calibration Interval
RIGOL suggests that the instrument should be calibrated every 18 months.
Product Overview
DHO1000 series digital oscilloscope is designed to meet the requirements for the design, debug, and test of the mainstream oscilloscope market. Adopting the brand new chipset “Centaurus” developed by RIGOL, this series achieves a fast waveform capture rate of 1,500,000 wfms/s with the Ultra Acquire mode, 100 M pts memory depth, 12-bit resolution, all combined with excellent noise floor performance and vertical accuracy to meet your requirements for more accurate measurements.
Front Panel Overview
Figure 4.1 Front Panel (4-channel Model)
Figure 4.2 Front Panel (2-channel Model)
- 10.1” Capacitive Touch Screen
- Multipurpose Knobs
- Measure Key
- Cursor Key
- Analyse Key
- Touch Lock Key
- Common Tools Keys
- Quick Action Key (Self-defined function)
- Trigger Controls
- Vertical Controls
- Horizontal Controls
- Probe Compensation Signal Output Terminal/Ground Terminal
- Analog Channel Input Terminals
- USB HOST Port
- Power Switch
Rear Panel Overview
Figure 4.3 Rear Panel
- Trigger Output Connector
- External Trigger Input Connector
- 10 MHz Reference Clock Input Connector
- 10 MHz Reference Clock Output Connector
- HDMI Connector
- USB DEVICE Port
- LAN Port
- AC Power Cord Connector
- Fuse
- Security Lock Hole
User Interface Overview
Figure 4.4 User Interface
- Waveform View
- Run State Label
- Horizontal Timebase Label
- Sample Rate & Memory Depth Label
- Horizontal Delay Label
- Trigger Label
- Function Toolbar
- Result Sidebar
- Notification Area
- Split-screen Display
- Math Labels
- Channel Labels
- Function Navigation Icon
To Prepare for Use
Tilting the Oscilloscope for Easier Viewing
Flip out the tabs under the oscilloscope’s rear feet to tilt the oscilloscope to stabilize it for easier operation and viewing. Flip in the tabs for storage or transporting when not using the instrument. Please see the figure below.
(a). Flip out tabs
(b). Flip in tabs
Connecting to Power
The power requirements of the oscilloscope are 100-240 V, 50-60 Hz. Please use
the power cord provided in the accessories to connect the oscilloscope to the
AC power source, as shown in the figure below.
WARNING
To avoid electric shock, ensure that the instrument is correctly grounded.
Turn-on Checkout
After the instrument is connected to the power source, press the power switch at the lower-left corner of the front panel to power on the instrument. During the startup process, the instrument performs a series of self-tests. After the self-test, the splash screen is displayed.
- Restart: Click or tap > Restart. Then a prompt message “Are you sure to reboot?” is displayed. Click or tap OK to restart the instrument.
- Shutdown:
- Click or tap > Shutdown. Then a prompt message “Are you sure to shutdown?” is displayed. Click or tap OK to shut down the instrument.
- Press the power switch and a prompt message “Are you sure to shutdown?” is displayed. Click or tap OK to shut down the instrument.
- Press twice to directly shut down the instrument.
- Press for three seconds to directly shut down the instrument.
TIP
You can also click or tap > Utility > Setup and select “Switch On” in “Power Status” item. In this way, the instrument powers on once connected to power.
Setting the System Language
This oscilloscope supports multiple languages. You can click or tap > Utility > Setup > Language to select the system language
Connecting Probes
RIGOL provides passive probes for DHO1000 series. For specific probe models, please refer to DHO1000 series Data Sheet. For detailed technical information of the probes, please refer to the corresponding Probe User Guide
Connect the Passive Probe
- Connect the BNC terminal of the probe to an analog channel input terminal of the oscilloscope on the front panel as shown in the figure below.
- Connect the ground alligator clip or spring of the probe to the circuit ground terminal, and then connect the probe tip to the circuit point to be tested.
Figure 5.3 Connecting the Passive Probe
After you connect the passive probe, check the probe function and probe
compensation adjustment before making measurements. For details, please refer
to Function Inspection and Probe Compensation.
Function Inspection
-
Press the front-panel and a prompt message “Restore default settings?” is displayed. Click or tap OK to restore the instrument to its factory default settings.
-
Connect the ground alligator clip of the probe to the “Ground Terminal” as shown in Figure 5.4 .
-
Use the probe to connect the input terminal of CH1 and the “Compensation Signal Output Terminal” of the probe, as shown in Figure 5.4 .
Figure 5.4 Using the Compensation Signal -
Set the probe ratio based on the attenuation of the probe, and then click or tap > Auto.
-
Observe the waveform on the display. In normal condition, you should see a square waveform similar to the waveform shown in the figure below.
Figure 5.5 Square Waveform Signal -
Test the other channels in the same way. If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure described in Probe Compensation. If you do not see the waveform, perform these steps again.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
Probe Compensation
When used for the first time, the oscilloscope probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected. A non compensated or poorly compensated probes can introduce measurement inaccuracy or errors. The compensation procedure is as follows:
- Perform step 1, 2, 3, and 4 in Function Inspection.
- Check the displayed waveforms and compare them with waveforms shown in the figure below.
Figure 5.6 Probe Compensation
- Use the probe compensation adjustment tool provided in the accessories to adjust the low-frequency compensation adjustment hole on the probe until the displayed waveform is consistent with the “Perfectly compensated” waveform shown in the figure above.
Touch Screen Gestures
The instrument’s large capacitive touch screen makes operation and configuration easy and flexible. The highly sensitive user interface designed for touch has a strong waveform display capability, bringing an extraordinary user experience. The actions supported by the touch screen controls include tapping, pinching & stretching, as well as dragging.
Tap
Use one finger to tap the symbol or characters on the screen slightly, as shown in Figure 6.1 . With the Tap gesture, you can perform the following operation:
- Tap the menu displayed on the screen to operate on the menu.
- Tap the function navigation icon at the lower-left corner of the touch screen to enable the function navigation.
- Tap the displayed numeric keypad to set the parameters.
- Tap the virtual keypad to set the label name and the filename.
- Tap the close button at the upper-right corner of the message box to close the prompt window.
- Tap other windows on the touch screen and operate on the windows
Figure 6.1 Tap Gestures
Drag
Use one finger to select the object, and then drag the object to a destination place, as shown in the figure below. With the drag gesture, you can perform the following operation:
- Drag the waveform to change its position or scale.
- Drag the window controls to change the position of the window (e.g. numeric keypad).
- Drag the cursor to move the cursor.
- Drag the trigger cursor to change the trigger level.
- In multi-window display, drag one of the displayed windows to change its position on the display
Figure 6.2 Drag Gesture
Pinch & Stretch
Pinch or stretch two points on the screen with two fingers to zoom in or out the waveform. To zoom in the waveform, first pinch the two fingers and then stretch the fingers; to zoom out the waveform, first stretch the two fingers, and then pinch the fingers together, as shown in the figure below. With the pinch & stretch gesture, you can perform the following operation:
- Pinching & stretching in the horizontal direction can adjust the horizontal time base of the waveform.
- Pinchin g &stretching in the vertical direction can adjust the vertical scale of the waveform.
Figure 6.3 Pinch & Stretch Gesture
Accessing the Built-in Help System
The built-in help system provides instructions for all the front-panel keys and their
corresponding menu keys. Click or tap > Help to access the “Help”
menu.
After that, you can get its help information by clicking on or tapping the
link for the desired item.
Parameter Setting Method
For this instrument, you can use the knobs and touch screen to set parameters. The common parameter setting methods are as follows:
- Method 1: Some parameters can be adjusted by rotating the front panel knobs.
- Method 2: Click or tap the input field of a specified parameter, then a virtual keypad is displayed. Complete the parameter setting with the keypad.
Input a Value
When setting or modifying a parameter, you can input an appropriate value with the numeric keypad.
- Input a value or unit in the keypad using the touch screen.
- You can also rotate the multipurpose knob (1 or 2) to move the cursor and select the desired value and unit. Then press the knob to input the selected value or unit.
Figure 8.1 Numeric Keypad
After you input all the values and select the desired units, the numeric keypad is turned off automatically. This indicates that you have completed parameter setting. Besides, after you have input the values, you can also press “OK” directly to close the numeric keypad. At this time, the unit of the parameter is the default unit. In the numeric keypad, you can perform the following operations:
- Delete the parameter value that has been input;
- Set the parameter value to a maximum or minimum value;
- Set the parameter to a default value.
- Clear the parameter input field.
Replacing the Fuse
If you need to replace the fuse, please use the correct fuse (AC 250 V, T3.15 A; 5.2 mm×20 mm) and follow the steps shown below (see Figure 9.1 ).
- Power off the instrument and remove the power cord.
- Insert a small straight screwdriver into the slot at the power socket and pry out the fuse holder gently.
- Remove the fuse.
- Insert the proper fuse into the fuse holder.
- Re-insert the fuse holder into the power socket.
WARNING
To avoid electric shock, please make sure that the instrument is powered off
and disconnected from the power before replacing the fuse. Also, please make
sure the fuse is consistent with the required fuse rating.
Remote Control
The instrument can be remotely controlled by the following three methods:
-
User-defined Programming
Users can program and control the instrument by using the SCPI (Standard Commands for Programmable Instruments) commands. For details about the SCPI commands and programming, refer to Programming Guide. -
PC Software
Users can use the PC software to send commands to control the instrument remotely. RIGOL Ultra Sigma is recommended. You can download the software from RIGOL official website (http://www.rigol.com).
Operation Procedures:
- Set up communication between the instrument and PC.
- Run Ultra Sigma and search for the instrument resource.
- Open the remote command control panel to send commands
Web Control
This instrument supports Web Control. Connect the instrument to the network, then input the IP address of the instrument into the address bar of the browser of your computer. The web control interface is displayed. Click Web Control to enter the web control page. Then you can view the display of the real-time interface of the instrument. Through the Web Control method, you can migrant the device control to the control terminals (e.g. PC, Mobile, iPad, and other smart terminals) to realize remote control of the instrument. When you first log in to the Web Control, the user name is “admin” and password is “rigol”.
This instrument can be connected to the PC via the USB, LAN, or GPIB interface to set up communication and realize remote control through the PC. The remote control can be realized by using SCPI (Standard Commands for Programmable Instruments) commands.
CAUTION
Before connecting the communication cable, please turn off the instrument to
avoid causing damage to the communication interfaces.
More Product Information
- Obtain the Device Information Click or tap > Utility > About to obtain the information of the instrument, such as the model, serial number, and firmware version number.
- View the Option Information and Install the Option The instrument is installed with the trial versions of the options before leaving factory. The trial time starts from the time when you power on the instrument for the first time, and the trial time is about 2,160 minutes. Open the “Utility” menu, and then click or tap Option to view the options currently installed on the oscilloscope and their information. For details, refer to descriptions in this series User Guide.
For more information about the instrument, refer to the relevant manuals by logging in to the official website of RIGOL (http://www.rigol.com) to download them.
- DHO1000 User Guide: introduces the functions of the instrument and the operation methods, remote control methods, possible failures and solutions in using the instrument, specifications, and order information.
- DHO1000 Programming Guide: provides detailed descriptions of SCPI commands and programming examples of the instrument.
- DHO1000 Data Sheet: provides the key features and technical specifications of the instrument.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
References
- 示波器-数字示波器-信号发生器-频谱分析仪-直流电源-普源精电科技-RIGOL TECHNOLOGIES, Co. LTD
- Environmental Declaration-RIGOL Technologies, Co. Ltd.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>