RIGOL DL3000 Programmable DC Electronic Load User Guide

May 15, 2024
Rigol

RIGOL DL3000 Programmable DC Electronic Load User Guide

RIGOL DL3000 Programmable DC Electronic Load User
Guide

Guaranty and Declaration

Copyright
© 2017 RIGOL TECHNOLOGIES, INC. All Rights Reserved.

Trademark Information
RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC.

Publication Number
QGJ01100-1110

Notices

  • RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
  • RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company’s sole decision.
  • Information in this publication replaces all previously released materials.
  • Information in this publication is subject to change without notice.
  • RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing, use, or performance of this manual, as well as any information contained.
  • Any part of this document is forbidden to be copied, photocopied, or rearranged without prior written approval of RIGOL.

Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2008 standard and the ISO14001:2004 standard. Other international standard conformance certifications are in progress.

Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL.

General Safety Summary

  1. Only the exclusive power cord designed for the instrument and authorized for use within the local country could be used.
  2. Ensure that the instrument is safely grounded.
  3. Observe all terminal ratings.
  4. Use proper overvoltage protection.
  5. Do not operate without covers.
  6. Do not insert objects into the air outlet.
  7. Use the proper fuse.
  8. Avoid circuit or wire exposure.
  9. Do not operate the instrument with suspected failures.
  10. Provide adequate ventilation.
  11. Do not operate in wet conditions
  12. Do not operate in an explosive atmosphere.
  13. Keep instrument surfaces clean and dry.
  14. Prevent electrostatic impact.
  15. Handle with caution.

Safety Notices and Symbols

Safety Notices in this Manual:

RIGOL DL3000 Programmable DC Electronic Load - Safety Notices in this
Manual

Safety Terms on the Product:

DANGER
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could result in potential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed, could result in damage to the product or other devices connected to the product.

Safety Symbols on the Product:

RIGOL DL3000 Programmable DC Electronic Load - Safety Symbols on the
Product

Care and Cleaning

Care
Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time.

Cleaning
Clean the instrument regularly according to its operating conditions.

  1. Disconnect the instrument from all power sources.
  2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. When cleaning the LCD, take care to avoid scarifying it.

RIGOL DL3000 Programmable DC Electronic Load - Cleaning
warning

Document Overview

This manual introduces some basic information that you should know when you use DL3000 for the first time. It contains the following contents: out-of-box inspection method, product overview, turn-on checkout, and remote control overview.

Format Conventions in this Manual

(1) The key on the front panel is denoted by the format of “Key Name (Bold) + Text Box” in the manual. For example,  denotes the ON/OFF  key.
(2) Use the screen shot to indicate the key. For example, denotes the Power key.

Content Conventions in this Manual

The number of channels and the rated values of input parameters for DL3000 are listed in the following table. Unless otherwise specified, this manual takes DL3031A as an example to illustrate the functions and operation methods of the DL3000 series.

RIGOL DL3000 Programmable DC Electronic Load - Content Conventions in this
Manual

General Inspection

  1. Inspect the packaging
    If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests.
    The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument.

  2. Inspect the instrument
    In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.

  3. Check the accessories
    Please check the accessories according to the packing lists. If the accessories are damaged or incomplete, please contact your RIGOL sales representative.

Product Overview

DL3000 is a cost-effective programmable DC electronic load with high performance. With a user-friendly interface and superb performance specifications, DL3000 provides various interfaces for remote communication to meet your diversified test requirements. It can be widely used in various industries, such as automotive electronics, aerospace, and fuel cells.

For descriptions of the front panel, refer to Figure 1; for descriptions of the rear panel, refer to Figure 2; and for descriptions of the main interface (display screen), refer to Figure 3.

RIGOL DL3000 Programmable DC Electronic Load - Figure
1

Figure 1 Front Panel

RIGOL DL3000 Programmable DC Electronic Load - Figure
2

Figure 2 Rear Panel

RIGOL DL3000 Programmable DC Electronic Load - Figure
3

Figure 3 Main Interface

Turn-on Checkout

To Connect to Power Supply

Two kinds of AC power inputs are supported by DL3000 series electronic load. When the DL3000 series is connected to different power sources, the AC selector setting on its rear panel is different, as shown in Table 1.

Table 1 AC Input Power Specification and AC Selector Setting

RIGOL DL3000 Programmable DC Electronic Load - Table 1 AC Input Power
Specification and AC Selector Setting

Please strictly follow the steps below to connect the DL3000 electronic load to the power supply.

  1. Check the input power
    Ensure that the AC power to be connected to the load conforms to the AC input power requirement specified in Table 1.

  2. Check the AC selector
    Ensure that the AC selector setting (115 or 230) on the rear panel of the load matches the actual AC input power (For the match relationship, refer to Table 1).

  3. Check the fuse
    When leaving the factory, the load has installed a fuse that conforms to the destination country standard. Please refer to the “fuse rating” on the rear panel of the load to ensure that the fuse matches the actual AC input power.

  4. Connect the AC power
    Plug the specified power cord available in the accessories into the AC power supply connector of the instrument, and then connect the instrument to the properly grounded AC power.

RIGOL DL3000 Programmable DC Electronic Load - Connect the AC
power

To Power on the instrument

After the instrument is connected to the power source, press the Power key at the left bottom of the front panel to power on the instrument. When the instrument is turned on, it will undergo a self-test. If the instrument passes the self-test, the welcome interface is displayed; otherwise, self-test failure information will be displayed. At this time, please contact RIGOL distributors.

RIGOL DL3000 Programmable DC Electronic Load - To Power on the
instrument

Tip
After powering off the electronic load, please wait for at least 1 s before you power it on again.

To Use the Built-in Help System

The built-in help system provides help information for any key on the front panel (except parameter input area) and menu keys, which are convenient for you to get the descriptions of the function keys or menus, as shown in Figure 4.

Figure 4 Help Information Interface

  1. Get the built-in help information
    Press , and then “Help” will be displayed at the top of the interface, being highlighted. At this time, press the corresponding function key or menu key to enter the corresponding help information interface.

  2. Page up/down operation
    If the help information is displayed in several pages, press Previous to go to the previous page or press Next to go to the next page. You can also use the arrow keys or the knob to page up/down the help information.

  3. Exit the current help information interface
    When the help information is displayed, press OK to exit the current help information interface and return to the previous interface.

Parameter Setting Method

Most parameters can be set by operating the keys on the front panel. The common setting methods are listed below. The setting method for certain parameters is different from the methods below, refer to the relevant chapters of this manual for the further explanation.

Method 1: Use the numeric keys

  1. In the main interface, press the specified menu key to switch the parameter focus; in the guide interface, use the arrow keys to switch the parameter focus.
  2. Enter a value by using the numeric key. While entering a name, press to delete the unwanted character if necessary.
  3. When setting resistance, press the specified menu key and select “Ω” or “kΩ” to be the unit.
    When setting period/width, press the specified menu key and select “ms” or “s” to be the unit.
    When setting frequency, press the specified menu key and select “Hz” or “kHz” to be the unit.
    When setting other parameters, press OK to confirm the input.

Method 2: Use the knob or the arrow keys

  1. In the main interface, press the specified menu key to switch the parameter focus.
  2. Press the Left/Right arrow key to move the cursor to a desired position.
  3. Press the Up/Down arrow key or rotate the knob to modify the value.

Remote Control

DL3000 electronic load can be connected to the PC via the USB, LAN, GPIB, or RS232 interface to build communication and realize remote control. The remote control can be realized by using SCPI (Standard Commands for Programmable Instruments) commands. DL3000 electronic load supports two ways of remote control: user-defined programming and PC software (e.g. RIGOL Ultra Sigma).

When the instrument is in remote control, the icon is displayed on the interface. The keypads on the front panel are locked. At this time, you can press Local to exit the remote mode.

More Product Information

  1. Obtain the device information
    Press to enter the system utility function interface, and then press System Info. Select the “system information” tab to obtain the device information, including the manufacturer, device model, software and hardware version number, FPGA version number, Boot version number, and the number of boot times, calibration date, and product serial number.

  2. Check the option installation status
    Press to enter the system utility function interface. Then, press Option to view the installation status of all the options.

For more information about this instrument, refer to the relevant manuals by logging in to the official website of RIGOL to download them.
DL3000 User’s Guide: introduces the functions of the instrument and the operation methods, remote control methods, possible failures and solutions in using the instrument, the technical specifications, and order information;
DL3000 Programming Guide: provides detailed descriptions of SCPI commands and programming instances of the instrument.
DL3000 Datasheet: provides the main features and technical specifications of the instrument.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Rigol User Manuals

Related Manuals