BLUE-RAY BSST-0100 BlueSwan Pipette Controller Instruction Manual
- June 9, 2024
- BLUE-RAY
Table of Contents
BSST-0100 BlueSwan Pipette Controller
Instruction Manual
Safety Precautions
Before using the BlueSwan pipette controller for the first time, please read this entire operation manual carefully. To guarantee problem free, safe operation of the BlueSwan pipette controller, it is essential to observe the following:
- When using infectious, radioactive, toxic and other solutions which may pose health risks, please observe the appropriate safety precautions.
- Do not use the BlueSwan pipette controller in a potentially explosive environment or with potentially explosive, flammable liquids.
- The BlueSwan pipette controllers operate on a small but powerful lithium-ion polymer battery. Misuse or abuse of the lithium-ion polymer battery may cause damage or injury through fire, electric shock, or chemical leakage.
- Do not incinerate the lithium-ion polymer battery or expose it to temperatures higher than 60°C.
- Do not short-circuit, puncture, crush, disassemble, damage, force over-discharge (reversal) or modify the battery.
- Do not expose the battery to water or moisture.
- Only use the battery specified in this manual.
- Only use the specified battery charging adaptor to charge the battery. Using an unauthorized battery charging adaptor might cause a short-circuit or fire on the BlueSwan pipette controller.
- Do not use a leaking battery.
- If charging is not completed within the specified time period (see Section 3.3 Recharging the Battery), unplug the charger and stop charging immediately.
- The charger temperature and battery temperature rise with extended periods of use. Care should be taken to avoid burns.
- If fluid from the battery enters your eye, immediately rinse the eye with plenty of fresh water and seek medical advice. If fluid from the battery makes contact with your skin or clothing, wash the area thoroughly with water.
- The BlueSwan pipette controller can’t be used with solvents whose vapors attack silicone, acrylonitrile butadiene styrene (ABS), polypropylene (PP), nitrile butadiene rubber (NBR), polyvinylidene fluoride (PVDF).
- Please observe the relevant regulations to dispose of the battery.
- All repairs should be carried out by authorized service personnel only.
General Description
The BlueSwan is an elegant, easy to use and reliable pipette controller. Its
ergonomic design makes it easy to handle and operate. It also offers precise
control of aspiration and dispensing of liquid.
2.1 Features
- Light-weight and ergonomic design for easier handling
- Continuous speed control by wheel and push buttons
- Blow-out and Gravity dispense selection
- Powerful and stable lithium-ion polymer battery
- Reliable mini-USB plug for charger adapter
- Three-color LED battery indicator shows the battery status
- Can be operated during the recharging period
- Self-inverted design
- Liquid and vapor protective hole
- UV-resistant housing
- Allow fast removal of nose cone for easier access to filter
- Compatible with 1 ~ 100 ml plastic and glass serological pipettes
- Removable battery cover
2.2 Product Overview
1 | Battery indicator | 2 | Protective hole |
---|---|---|---|
3 | Filter | 4 | Pipette holder |
5 | Nose cone | 6 | Battery cover |
7 | Lithium-ion polymer battery | 8 | Speed control wheel |
9 | Aspiration button | 10 | Dispensing button |
11 | Mini-USB socket | 12 | Charger adapter |
13 | Mini-USB plug |
2.3 Materials
Component | Materials |
---|---|
Housing, battery cover, wall mount | Acrylonitrile Butadiene Styrene (ABS) |
Aspiration button, dispensing button, nose cone | Polypropylene (PP) |
LED cover | Polycarbonate (PC) |
Membrane filter | Polytetrafluoroethene (PTFE) |
Pipette holder | Silicone |
Getting Started
3.1 Unpacking
Open the BlueSwan pipette controller package and confirm that all items listed
below are included:
- BlueSwan pipette controller
- Lithium-ion polymer battery
- Wall mount
- Operation manual
- Charger adapter
- Spare membrane filter, 0.2 μm
If there are any items missing, damaged, or the package contains incorrect
items, please contact Blue-Ray’s distributor or sales representative
immediately.
3.2 Battery Installation
After ckecking the pipette controller package, please follow the steps below
to install the battery (Figure 2).
- Open the battery cover by sliding the cover outward.
- Connect the battery cable to the unit. Be aware of the orientation of the cable connector.
- Insert the battery into the battery compartment. Place the battery cable inside first.
- Align the battery cover with cover slot first. Close the battery cover by sliding the battery cover inward.
3.3 Recharging the Battery
Please charge the lithium-ion polymer battery 4 hours before first-time
operation.
When the battery indicator turns red (low battery), follow the steps below to
recharge the battery (Figure 3).
- Plug the mini-USB jack of the charger adapter into the mini-USB socket in the base of the unit.
- Plug the charger adapter into an external power source.
- The battery indicator will start blinking red during charging period.
- After the battery is fully charged, the battery indicator will turn blue.
The charging period takes approximately 4 hours.
Note
- Charge the battery at least once every 3 month to avoid over-discharge which might damage the internal battery cells, even you did not use the BlueSwan pipette controller during this period.
- The BlueSwan pipette controller can be operated during the recharging period.
- Only use the original manufacturer’s battery and charger adapter (see Appendix D: Order Information).
3.4 Battery Indicator
The LED battery indicator explains the battery status for users (Table 1).
Table 1. Battery indicator
Use Status | Battery Indicator | Battery Status |
---|---|---|
Operating | Blue | Fully charged |
Purple | Half charged | |
Red | Almost empty | |
Off | Empty | |
Charging & operating during charging | Blinking red | Partial charged |
Continuous blue | Fully charged |
3.5 Starting Pipetting
3.5.1 To Aspirate
- Carefully hold the pipette and securely insert it into the pipette holder in the nose cone.
- Select the pipetting speed by rotating the speed control wheel.
- Press the aspirating button. The flow rate is regulated by the pressure applied to the aspiration button.
3.5.2 To Dispense
- In “Gravity mode”, press the dispensing button lightly (4 ~ 5 mm). The liquid in the pipette is dispensed by gravity only.
- In “Blow-Out mode”, press the dispensing button (> 5 mm). The flow rate is regulated by the pressure applied to the dispensing button.
Note
- Speed control should depend on the pipette used.
- Please do not use the BlueSwan pipette controller without the protective membrane filter.
- Replace the filter if the liquid pass it.
3.6 Wall Mount Installation
The BlueSwan pipette controller is supplied with a wall mount that can be
attached to a wall with 2 stickers (provided). The wall mount can hold the
nose cone of the pipette controller and it keeps the pipette controller in a
vertical position. Please follow the procedure below to perform installation
(Figure 4).
- Adhesive tape works best on smooth surfaces, e.g. metal, glass or plastic surfaces. Clean with isopropyl rubbing alcohol by wiping gently.
- Remove the liner and press the wall mount directly onto the wall. Press along the entire wall mount for 30 seconds. The harder you press, the stronger the adhesive bond will be.
- Wait 24 hours before hanging the pipette controller on the mount.
3.7 Self-inverted Design
The pipette controller can be self-inverted on the bench without a holder or
stand as shown in Figure 5 below.
3.8 Protective Hole
The liquid and vapor that get into the filter will flow in the tube and spurt
from the protective hole. This design can prevent the parts in the BlueSwan
pipette controller from corrosion and contamination.
Preventative Maintenance
4.1 Replacing Filter/Pipette Holder
The hydrophobic PTFE filter and pipette holder will require occasional
replacement, particularly when aspirating speed is decreased. Rotate the nose
cone clockwise and remove it. You can squeeze the top of the pipette holder
and then remove the pipete holder and filter together. Replace the wet, dirty
filter/worn pipette holder with a new one. Reassemble the device.
4.2 Cleaning and Disinfection
Before cleaning, disconnect the unit from the charging adaptor.
The BlueSwan Pipette controller can be cleaned with most common laboratory
disinfectants, such as 60% Isopropanol, 70% ethanol or mild detergent. The
PTFE filter, polypropylene nose cone and silicone pipette holder can be
autoclaved (121°C, 1.2 kg/cm² for 20 min).
Troubleshooting
To ensure of the product’s quality and performance, the pipette controller is designed to be robust and long-lasting. If the problems cannot be resolved with the solutions below, please contact the authorized distributors.
Symptom | Possible Cause | Solution |
---|---|---|
Reduced suction efficiency. | Filter blocked, dirty or wet. | Clean or replace |
filter.
Low battery.| Recharge battery.
Pipette drips.| Pipette not securely inserted in pipette holder.| Insert
pipette further.
Pipette holder/filter inserted incorrectly.| Reinsert the pipette holder and
filter carefully.
Pipette holder or pipette damaged.| Replace pipette holder or pipette.
No function even though the battery is fully charged.| Battery is aged.|
Replace battery.
Pipette loose.| Pipette holder damaged.| Replace pipette holder.
Appendix A: Technical Specifications
Pipette Compatibility | 1 – 100 ml plastic/glass pipette |
---|---|
Max Aspirating and Dispensing Speed | >7.5 ml/sec. |
Dispensing Mode | Blow-out/Gravity |
Speed Control Wheel | Continuous; 4- indicates Fast/Slow |
Number of Dispenses | Approx. 4500 times (with a 25 ml pipette) |
Charging Time | Approx. 4 hours at 100% charging |
Charger Adaptor | Input: AC 100 – 240V 50/60 Hz |
Selectable power plug: US/EU/UK/AU
Output: DC 5V/2.0 A
Power Jack| Mini-USB
Filter| Hydrophobic 25 mm diameter, 0.2 pm
Battery| Lithium-ion polymer battery 3.7 V/1100 mAh
Autoclavable| Nose cone, pipette holder and filter only
Operating Temperature| 5 – 40°C
Operating Humidity| RH: 0 – 85%
Weight| 180 g, including a battery
Certification| Complies with CE, EN61010-1, EN55011, EN61326-2-6 (Class A)
RoHS compliant
Appendix B: Warranty
The BlueSwan pipette controllers are covered by a warranty for one year
against defects in materials and workmanship. This period begins from the date
of purchase, and within this period all defective parts will be replaced at no
charge by the manufacturer. The warranty does not cover defects caused by
excessive wear and tear or damage due to shipping, accident, abuse, misuse,
problems with electrical power, or usage not in accordance with product
instructions, or if other than original spare parts supplied by the
manufacturer have been used. Each BlueSwan pipette controller is tested and
documented by the manufacturer before shipping.
Serial No.: ——————-
Date Purchased: ———–
Supplier: ——————–
Appendix C: CE Declaration
BLUE-RAY BIOTECH CORP.
4F., No. 31, Sec. 2, Chang-An E. Rd.,
Zhong-Shan Dist., Taipei City 10456,
Taiwan (R.O.C.)
Declaration of Conformity
Product Name: BlueSwan
Model Names: BSST-0100
All models comply with the following European standards:
EMC: EN55011/EN61326-2-6 (Class A)
Safety: EN61010-1
To the best of my knowledge and belief, these units conform to these
standards.
Name: Jimmy Kuo
Position: Quality Assurance Manager
Issue Date: 2016. 7. 18
Appendix D: Order Information
Order No. | Description |
---|---|
BS01-1000 | BlueSwan pipette controller, US plug |
BS01-2000 | BlueSwan pipette controller, EU plug |
BS01-3000 | BlueSwan pipette controller, UK plug |
BS01-4000 | BlueSwan pipette controller, AU plug |
Blue-Ray BIOTECH CORP.
4F., No. 31, Sec. 2, Chang-An E. Rd., Zhong-Shan Dist.,
Taipei City 10456, Taiwan (R.O.C.)
P/N: 401-pcbr00-10
Copyright © 2016 Blue-Ray Biotech Corp.
All Rights Reserved.
Ver. 1.0
Read User Manual Online (PDF format)