EXTECH PH90 Waterproof pH-Temperature Meter User Manual
- June 8, 2024
- EXTECH
Table of Contents
Waterproof pH/Temperature Meter
Model PH90
USER MANUAL
Introduction
Thank you for selecting the Extech Model PH90 waterproof pH/Temperature meter. This instrument is equipped with a flat surface pH electrode (model pH95) for reliable test results. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.extech.com) to check for the latest version of this User Guide, Product Updates, Product Registration, and Customer Support.
Features
- Rugged, replaceable, flat surface pH electrode (pH95) for quick pH measurements
- LCD display shows pH and temperature measurements simultaneously
- Calibration automatically recognizes buffer solution
- PTS (percent of slope) feature alerts user when electrode replacement is necessary
- ATC (automatic temperature compensation)
- Waterproof design (IP57); meter can float in water
- Auto power off after 10 minutes of inactivity
- Easy to replace pH electrode
Description
- Battery compartment cap
- LCD Display
- Power and CAL button
- HOLD and MAX/MIN button
- °C/°F button
- Electrode collar
- Electrode (PN: PH95; replaceable)
CAUTIONS
- This device contains small objects that can be swallowed; do not allow children or pets to handle or play with this device.
- Do not allow children or pets to handle or play with the batteries and packing material, these items pose a hazard.
- If this device is to be stored unused for an extended period of time, remove the batteries.
- Expired or damaged batteries can cause cauterization on contact with skin. Always use suitable hand protection.
- Please ensure that the batteries are not short‐circuited.
- Do not dispose of batteries in a fire.
Overview
pH Overview
pH is a unit of measure (ranging from 0 to 14pH) indicating the degree of
acidity or alkalinity of a solution. Solutions with a pH lower than 7 are
considered acidic, solutions with a pH higher than 7 are known as bases, and
solutions with a pH of exactly 7 are considered ‘neutral’.
The pH scale is logarithmic so, for example, if sample ‘A’ is 1 pH lower than
Sample ‘B’, then Sample ‘A’ is 10 times more acidic than Sample B. A
difference of 1 pH represents a ten‐fold difference in acidity.
Getting Started
- For new meters, remove the battery cap (1) at the top of the meter and install 2 batteries observing correct polarity.
- Remove the cap from the bottom of the instrument (7) to expose the flat surface electrode and reference junction.
- Before first use, rinse the electrode with clean tap water and wipe dry.
- The sponge should be kept moistened at all times to ensure accuracy.
Replacing Electrodes
The PH90 is shipped with a flat surface pH electrode attached (7). Electrode
life is limited, and is dependent on (among other factors) frequency of use
and care. If the electrode needs to be replaced, follow these steps for
removing and connecting electrodes. Replacement pH electrodes (part number
pH95) are available from Extech Instruments.
- Unscrew the electrode collar (6) clockwise, and remove it completely.
- Pull the pH electrode module out from the tester.
- Carefully, plug a new electrode module into the tester socket.
- Replace and tighten the electrode collar to ensure a good seal.
Automatic Electrode Recognition
When the meter is powered, it recognizes the type of electrode that is
connected and displays the appropriate unit of measure. Attach the electrode
before powering the meter; never power the meter without an electrode
attached.
Powering the Meter
The batteries are already installed in the meter. Press the power button (3)
to power ON or OFF. The displayed battery indicator lets the user know the
status of the battery strength. The Auto Power off feature shuts the meter off
automatically after 10 minutes of inactivity to preserve battery life.
Operation
Understanding the PH90 Display
When an electrode is placed in a solution, the main display indicates the pH
reading while the lower display shows temperature.
pH Calibration
A two point calibration with a buffer of 7 and 4 or 10 (whichever is nearest
to the expected sample value) is always recommended. For best accuracy, always
calibrate at the sample temperature.
- Check that the pH icon is displayed on the LCD.
- Dip the electrode into the pH7 buffer solution. Stir gently and wait for a stable reading.
- Press and hold the CAL button (3) to enter calibration mode.
- The LCD will display the CAL icon and then flash 7.00.
- When calibration is complete the meter will stop flashing and indicate SA and END.
- Rinse the electrode with clean tap water and wipe dry.
- Dip the electrode in pH4 (or pH10) buffer solution.
- Stir gently and wait until the display is stable.
- Press and hold the CAL button to enter calibration mode.
- The LCD will display the CAL icon and then flash 4.00 or 10.00.
- When calibration is complete the meter will stop flashing and indicate % (percent of slope), SA and END.
- After the calibration of pH4 or pH10, the display will indicate the slope percentage (PTS), indicating the status of the electrode.
- If the PTS is < 70% or > 130% the electrode must be replaced. A slope of 100% is ideal.
Note: The SA icon will not appear if the calibration fails.
Note: Calibrate with a pH 7 buffer first and then follow with a pH 4 or
pH 10 calibration.
pH Measurements
After calibration (detailed earlier), rinse the electrode with clean tap water
and wipe dry. Dip the electrode into the sample solution to be measured. Stir
gently and wait for a stable reading. Automatic temperature compensation is
always active when measuring pH (ATC icon shown on LCD).
Temperature Units Selection ( °C/ °F)
Press and hold the C/F button to toggle the displayed temperature units of
measure.
Data Hold
Momentarily press the HOLD/MAX‐MIN button (4) to enter the HOLD mode; the
displayed reading will freeze. The icon HOLD will be displayed on the LCD
while in this mode. Press this button again to exit the Data Hold mode and
return to the measurement mode. The HOLD icon will disappear and the reading
will return to the real‐time value.
MAX/MIN
Press and hold the HOLD/MAXMIN button (4) until the MAX or MIN icon flashes on
the display. Momentary presses of this button will now toggle the MAX (highest
reading shown) and MIN (lowest reading shown) modes. Press and hold this
button to exit the MAX/MIN mode and return to the measurement mode (the
MAX/MIN icons will disappear). Note that the meter will not Auto Power OFF
while in the MAX/MIN mode.
Factory Default RESET Procedure
To restore the meter to the factory default settings, power on the meter and
press and hold these two buttons while powering the meter OFF.
Battery Replacement
1. Unscrew the battery compartment cap counterclockwise to access the battery
compartment.
2. Replace the two (2) 2032 batteries observing correct polarity.
3. Replace the battery compartment cap securely.
** Never dispose of used batteries or rechargeable batteries in household
waste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or
wherever batteries are sold.
Disposal:** Do not dispose of this instrument in household waste. The user
is obligated to take end‐ of‐ life devices to a designated collection point
for the disposal of electrical and electronic equipment.
Specifications
Display | pH | Temp |
---|---|---|
Range | 0 to 14.00 pH | 0 to 90.0°C |
(32 to 194°F)
Accuracy| ± (0.01pH + ldgt)| ± 0.5°C + 1 dgt
(± 0.9°F + 1 dgt)
Resolution| 0.01 pH| 0.1°F/C
ATC
(Auto temperature compensation)| 0 to 90°C
(32 to 194°F)| n/a
Calibration points| pH 4.00, 7.00, and 10.00
Power| 2 x 3V Lithium batteries (CR2032)
Dimensions| 36 x 170 x 36mm (1.4 x 6.7 x 1.4″)
Weight| 85g (3oz.)
Copyright © 2015‐2016 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in
any form
ISO‐9001 Certified
www.extech.com
PH90‐en‐GB_V1.4 6/16
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>