midmark 1000 Mobile Dental Delivery System Installation Guide

June 8, 2024
Midmark

Midmark 1000® Installation

Applies to Models:
8000-001, -002, -003, -004,
-005, -006, -006S, -007,
-007S, -008, -008S

1000 Mobile Dental Delivery System

WARNING
Equipment not suitable for use in the presence of a flammable anesthetic mixture.
CAUTION
Exercise care when transporting over thresholds and other low lying objects.
Failure to do so can lead to injury.

Note
The high-speed and low-speed handpieces are located in the back compartment of the animal health mobile dental delivery system.

Step 1: Close water needle valve for handpieces that do not use water, like most low-speed handpieces and polishers.| Step 2: Locate cart in area of operation. Place foot control on floor. Remove, fill with distilled water
and replace water bottles.| Step 3: Connect power cord to cart. Plug cord into appropriate wall outlet. On/Off switch is located on unit above power cord plug.
Note: Perform checks on next page before  operating instruments.
---|---|---

Step 4: Set-up Scaler…
A) Make sure Midmark power switch is OFF.
B) Screw scaler tip on and tighten moderately with tool provided.
C) Plug handpiece into hose connection, place in designated handpiece holder.
D) Turn Midmark ON.
E) Set scaler intensity knob to lowest position in the color range that corresponds to the tip color.
F) Remove from holder, depress foot pedal and adjust the water control knob until a steady FINE mist is produced.
G) Scaler is ready to operate.

Equipment Alert
The Satelec scaler accessory is manufactured by ACTEON North America. Refer to ACTEON’s literature for detailed information on operation and maintenance. For additional information go to: www.acteonsupport.com.

Step 5: Install Prophy Angle on Low-Speed Handpiece.
A) Turn chuck release ring counter clockwise to unlock.
B) Install Prophy angle. Turn chuck release ring clockwise to lock.

Step 6: Install Bur in High-Speed Handpiece.

Equipment Alert
Handpieces will not work properly if not installed as shown.

Step 7: Install handpieces.

Note
The air adjustment valve that corresponds to the ultrasonic scaler (optional) must be OPEN in order to operate the scaler with the foot pedal.

EQUIPMENT ALERT
Check only one hand piece at a time. Do not adjust regulator over 65 PSI.

Step 8: Check handpiece air flow.
A) Remove handpiece from holder.
B) Depress foot control.
C) Regulate air pressure by adjusting the corresponding valve.
D) Repeat for each handpiece.
Note: If the air adjustment valve is fully open and pressure is too low, then increase main air regulator pressure until the recommended pressure is reached.

Air Regulator Adjustment
To Adjust Regulator…
A) Pull adjustment knob up to unlock.
B) Rotate adjustment knob clockwise to increase pressure, counterclockwise to decrease pressure.
C) Push adjustment knob down to lock.

EQUIPMENT ALERT
Check only one hand piece at a time.
Step 9: Check handpiece water flow.
A) Remove handpiece from holder.
B) Depress foot control.
C) Regulate water pressure by adjusting the corresponding needle valve.
D) Repeat for each handpiece that uses water.

© 2020 Midmark Corp.
60 Vista Drive Versailles, OH 45380 USA
1.800.643.6275
1.937.526.3662
www.midmark.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals