rossmax SB220 Fingertip Pulse Oximeter Instruction Manual
- June 9, 2024
- RossMax
Table of Contents
rossmax SB220 Fingertip Pulse Oximeter
Introduction
Rossmax Fingertip Pulse Oximeter SB220 is used to measure arterial oxygen
saturation (% SpO2) of hemoglobin and pulse rate, an important indicator of
your respiratory function. It is non-invasive device intended for spot-check
of adult and pediatric whose age is over 3 at home, hospital and clinics.
Attention: Consult the accompanying documents. Please read this manual
carefully before use. Please be sure to keep this manual.
Error code for your reference
SENSOR ERROR:| Sensor cannot be detected, return the device to your local
distributor or service centre.
---|---
MEASURE ERROR:| Signals cannot be detected, turn the device off and measure
again.
Name/ Functions of each part
- OLED display
- Pulse strength
- Pulse search icon
- Remind icon
- Battery indicator
- Power On/Off Button
- Battery compartment
- Pulse rate icon
- SpO2 icon
Installing Batteries
- Use thumb to slide battery cover out.
- Insert or replace 2 “AAA” sized batteries down with the correct electrical polarity.
You need to replace the batteries when:
- L ow battery icon appears on display
- The function button is pressed and nothing appears on display.
Caution:
Batteries may leak or explode if used or disposed of improperly. Remove
batteries if the device will be stored for long time. Do not use different
types or brands of batteries at the same time. Do not mix fully charged and
partially charged batteries at the same time. The device will automatically
shut down in low battery.
Attaching the lanyard
- Insert the narrow end of the lanyard through the holder.
- Draw the other end of the lanyard through the loop at the narrow end and tighten.
How to use
-
Open the clip; press the Power On/Off button as 1.
-
Information of software version appears. Insert one finger(left hand middle finger is recommended), nail side up, into the finger opening of the pulse oximeter as 2.
Note: If no finger insert, the device will auto shut off after 8 seconds. -
T he display shows , pulse oximeter begins its measurement.
Note: Make sure the finger is lying flat. Do not shake and keep body steady during measurement. -
Your SpO2 and pulse rate values will appear on the screen after few seconds as 3.
Note: -
Don’t remove your finger until the measurement is completed.
-
While pulse strength is low, the reading will flicker.
-
Press button shortly to reverse the display upside down.
Note:
- The SpO2 sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a position there between.
- Make sure the optical path is free from any optical obstacles like rubberized fabric.
- Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, direct sunlight and etc.
- Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
- The device has a visual and audio signal when the measurement of SpO2 is lower than 90%.
Cautionary Notes
- This device is to be operated by trained personnel only.
- The device has intended only for spot-checking, but not medical result evaluation.
- This device is designed to determine the percentage of ar-terial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
- Do not apply the pulse oximeter on the same arm as a blood pressure cuff, arterial catheter or infusion line(s).
- Excessive light, such as sunlight or direct home lighting.
- Not steady at the site of application (e.g. trembling).
- Moisture in the device.
- Improperly applied device.
- Finger is too large or too small to fit into the device.
- Poor pulse quality.
- Venous pulsations.
- Anemia or low hemoglobin concentrations.
- Cardio green and other intravascular dyes.
- Carboxyhemoglobin.
- Methemoglobin.
- Dysfunctional hemoglobin.
- Artificial nails or fingernail polish.
- On fingers with anatomical changes, oedemas, scars or burns.
- Using the device for long periods may cause pain for people with circulatory disorders. Reposition the device at least once every 4 hours to allow the patient’s skin to breath and to check patient’s condition regularly.
- Do not use the device near flammable or explosive gas mixtures.
- Do not use the device during an MRI or CT scan, be used no closer than 30 cm (12 inches) to any part of the SpO2 device, including cables specified by the manufacturer.
- The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
- This device is a precision electronic instrument and must be repaired by qualified technical professionals. Field re-pair of the device is not possible. Do no attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.
- Do not overextend the device’s spring.
- A functional tester cannot be used to access the accu-racy of a pulse oximeter monitor.
- Do not self-diagnose or self-medicate on the basis of the measurements without consulting your doctor. In particular, do not start taking any new medication or change the type and/or dosage of any existing medication without prior approval.
- Do not look directly inside the housing during the measurement. The red light and the invisible infra-red light in the pulse oximeter are harmful to your eyes.
- This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who has responsibility for their safety or they receive instructions from this person on how to use the device. Children should be supervised around the device to ensure they do not play with it.
- Neither of the displays for the pulse wave and pulse bar allows the strength of the pulse or circulation to be evaluated at the measurement site. Rather, they are exclusively used to display the current signal variation at the measurement site and do not enable reliable diagnostics for the pulse.
- The maximum temperature of sensors which the user will touch might reach 43°C when operating in the 40°C environment.
- The safety for user use in high-temperature environment is measuring for 10 minutes, and turn it off for 20 minutes before measure again.
- The oximeter is calibrated in the factory before sale.
There is no need to calibrate during its life cycle.
Cleaning
- Please clean the surface of the device before using. Wipe the device with medical alcohol (70% isopro-pyl alcohol)first, and then let it dry in air or clean it by dry clean fabric. When cleaning the device with water, the water temperature should be lower than 60°C
- Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time use.
- The best storage environment of the device is – 25°C to 70°C ambient temperature and not higher than 90% relative humidity.
Note:
- Do not sterilize, autoclave or immerse this device in liquid. Do not pour or spray any liquids onto the device.
- Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl alcohol.
Maintenance
Recommends user to return this device to the man-ufacturer perform the following checks every 24 months.
- Inspect the equipment for mechanical and functional damage or deterioration.
- Ensure all user interface keys and accessories function normally.
Note: Manufacturer use Index2 SpO2 simulator to ver- ify operation of the pulse oximeter equipment.
Troubleshooting
Symptoms | Check points | Corrections |
---|---|---|
SpO2 or pulse rate cannot displayed | Applied finger improperly. | Place the |
finger prop- erly and try again.
SpO2 is too low to detect| Try again; go to consult with your physician if you
are sure the device works well.
SpO2 or pulse rate are not dis- played stably| Applied finger improperly.|
Place the finger prop- erly and try again.
Finger is shaking or body is moving.| Keep body steady
No display when button is pressed| Batteries run down| Replace with new
batteries
Batteries not inserted correctly.| Re-insert batteries
The display disappears suddenly| The device will auto power off when it gets
no signal.| Normal
Low battery| Replace with new batteries
Note: I f the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.
Specification
SpO2
Measuring range| 35%~99%, (the resolution is 1%)
Accuracy| 70%~99%: ±2%, Below 35~69%: unspecified
Optical Sensor| The wavelength of red LED is 660 nm and Infrared LED is
905/880 nm with maximum optical output power of 4 mW/sr.
Pulse
Measuring range| 30 bpm~250 bpm (the resolution is 1 bpm)
Accuracy| ±3 bpm
Power source| AAA × 2 (Alkaline)
Battery life| Continually for 16 hours with two alkaline batteries
Operating Condition| Temperature: 5°C~40°C (41°F ~ 104°F), Relative Humidity:
15-95% (non condensing), Atmospheric pressure: 700hPa ~ 1060hPa,
Attitude: -1,280 to 12,000 feet (-390m to 3,658m)
Storage / Transportation Condition| Temperature: -25°C~+70°C(-13°F ~ 158°F),
Relative humidity: 15-90%(non condensing), Atmospheric pressure: 700hPa ~
1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m)
Dimensions| 63.5(L) × 34(W) × 35(H) mm
Weight| About 37g (without the batteries)
Standards| IEC60601-1-2, Class B, IEC60601-1, Type BF, ISO80601-2-61,
IEC60601-1-11
| Type
BF applied parts
IP Classification| IP22: Protection against harmful ingress of water and
particulate matter
|
Attention
| Serial number
| Manufacturer
| Date of manufacture
| |
Authorized representative in the European Community
Note:
- A description of the effect on displayed and transmitted SpO2 and pulse rate:
- Data averaging: 4 seconds for SpO2; 8 seconds for pulse rate.
- Data update delay: Less than 2 seconds.
Note:
- The product is with a service life of 5 years in the course of regular use.
- I f any serious incident (e.g. death) has occurred in relation to the device should be reported to the dealer, manufacturer, and the competent authority of the Member State in which the user and/or patient is established.
- The text is subject to change without further notice.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger- tip pulse oximeter as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter / W| Separation distance according
to frequency of transmitter / m
---|---
150 kHz to 80 MHz , d=[3.5/V1]√P| 80 MHz to 800 MHz , d=[3.5/E1]√P| 800 MHz to
2,5 GHz , d=[3.5/E1]√P
0.01| 0.12| 0.12| 0.23
0.1| 0.37| 0.37| 0.74
1| 1.17| 1.17| 2.33
10| 3.7| 3.7| 7.37
100| 11.67| 11.67| 23.33
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and
SYSTEMS that are not LIFE- SUPPORTING and are specified for use only in a
shielded location
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Finger-tip pulse oximeter system should assure that it is used in such an
environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment-guidance
Conducted RF IEC 61000-4-6| 3 Vrms
150 kHz to 80 MHz| 3V| Portable and mobile RF communications equipment should
be used no closer to any part of the EQUIPMENT or SYSTEM including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter. Interference may occur in the
vicinity of equipment marked with the following symbol.
Radiated RF IEC 61000-4-3| 3 V/m
80 MHz to 2.5 GHz| 3V/m
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Finger-tip pulse oximeter system should assure that it is used in such an
environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment- guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±6 kV contact
±8 kV air| ±6 kV contact
±8 kV air| Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30
%.
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power sup- ply lines
±1 kV for input/ output lines| ±2 kV for power sup- ply lines| Mains power
quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| ± 1kV differential mode
± 2kV common mode| ± 1kV differential mode
± 2kV common mode| Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interrup-tions and voltage variations on power supply
input lines IEC 61000-4-11| <5% UT(>95% dip in UT) for 0,5 cycle
40% UT(60% dip in UT) for 5 cycles
70% UT(30% dip in UT) for 25 cycles
<5% UT(>95% dip in UT) for 5 s| <5% UT(>95% dip in UT) for 0,5 cycle
40% UT(60% dip in UT) for 5 cycles
70% UT(30% dip in UT) for 25 cycles
<5% UT(>95% dip in UT) for 5 s| Mains power quality should be that of a
typical commercial or hospital environment. If the user of the EQUIPMENT or
SYSTEM requires continued operation during power mains interruptions, it is
recommended that the EQUIP- MENT or SYSTEM be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 3 A/m| 3 A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical com- mercial or hospital environment.
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the Finger-tip pulse
oximeter should assure that it is used in such an environment.
Emission test| Compliance| Electromagnetic environment-guidance
CE emissions CISPR11| Group 1| The Finger-tip pulse oximeter uses RF energy
only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Finger-tip pulse oximeter is suitable for
use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions IEC 61000-3-3| Complies
Warranty Card
This instrument is covered by a 2 year guarantee from the date of purchase, batteries and accessories are not included. The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming date of purchase or the receipt. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.
- Customer Name: __
- Address: ___
- Telephone: _____
- E-mail address: ___
- Product Information: __
- Date of purchase: _____
- Store where purchased:
WARNING: The symbol on this product means that it’s an electronic product
and following the European directive 2012/19/EU the electronic products have
to be dispose on your local recycling centre for safe treatment.
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006, Málaga, Spain
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>