KaWe EUROLIGHT Ophthalmoscope Pocket Set Two Handles User Manual
- June 3, 2024
- KaWe
Table of Contents
- KaWe EUROLITE Ophthalmoscope Pocket Set Two Handles
- Symbols used in the application manual and on the packaging
- Purpose and intended use
- Safety rules/contraindications
- Installation and maintenance
- Preparing for operation
- Hygienic preparation
- Manual cleaning/disinfection
- Transportation and storage
- Maintenance and Service
- Accessories and Spare parts
- Disposal
- Annex
- Warranty & Guarantee
- Documents / Resources
KaWe EUROLITE Ophthalmoscope Pocket Set Two Handles
-
We would like to thank our valued customers for their decision to purchase a KaWe product. Our ophthalmoscopes are characterized by high quality and durability. KaWe ophthalmoscopes meet the requirements of Regulation (EC) 2017/745 (European Regulation on Medical Devices) and are classified as medical devices class I.
-
Please read this manual carefully, adhere to it strictly, and keep it for future reference if necessary. Before using the product, carefully study operation rules, especially care requirements.
If you have any questions concerning the device’s operation, please contact the service department (the manufacturer’s address is indicated on the back).
KaWe ophthalmoscopes are available in two models: -
EUROLITE: Models with metal handles and plastic button
-
PICCOLIGHT: Models with plastic handles and button
-
Images are shown on the cover of this application manual. The difference between the two models are described in this appli-cation manual. Other options (battery operation, light sources, etc.) are not discussed in detail, as they do not have significant differences in service.
Symbols used in the application manual and on the packaging
- ATTENTION!
- Please follow this User’s Manual
- Item No.
- Manufacturer’s number, lot number
- Manufacturer
- Temperature range limits
- Separate collection of electrical / electronic equipment and rechargeable batteries
- CE conformity mark
- GOST-R certification of goods imported into Russia
- Medical product
Purpose and intended use
KaWe ophthalmoscopes are diagnostic instruments and are primarily used for optical examination of the fundus by direct ophthalmoscopy. In this method, the central elements of the visual organ, such as optic disc, the vascular bundle, macula (Makula lutea) are examined. The ophthalmoscope allows exa- minations without the use of mydriatics. However, in this case, the patient‘s field of vision is limited.
- Target patient groups: Patients of all ages and genders who require visual examination.
- The product must only be used by authorized specialized personnel!
- Intended use is compliance with all instructions in this ap-plication manual as well as basic rules and directives. The ma-nufacturer does not accept any liability for damage resulting from any other use.
Safety rules/contraindications
- Before each use, please check whether the ophthalmoscope is operating properly. Do not use the device in case any damage is detected.
- Ensure the ophthalmoscope is protected from moisture.
- Avoid direct contact of the ophthalmoscope with the visual organ.
- If the light is on for a long time, the areas around the lamp may become hot.
- KaWe ophthalmoscopes are not intended for use during surgery.
- Use only KaWe MedCentether 5000 wall-mounted station for power supply of KaWe EUROLITE E25 head.
- Do not use the device in areas subject to strong magnetic fields (MRI systems).
- When handling ophthalmoscopes – as with all energy-consuming diagnostic devices – special electromagnetic compatibility (EMC) precautions must be observed.
- The performance of ophthalmoscopes can be affected by mobile high-frequency communication devices.
- Only original spare parts and accessories from KaWe may be used.
- In case of a malfunction, a large amount of heat can be generated, resulting in an increased risk of fire. In case of strong heat, turn off the handle.
Installation and maintenance
When handling the instrument, it is necessary to take into ac-count that the
examiner must bring his eye as close as possible to the instrument. Due to the
lateral illumination of the patient‘s fundus area and the approach to a
maximum distance of 2-3 cm, diagnostics can be carried out without drugs.
Duration of operation
The devices are not intended for continuous operation, the ma-ximum duration
of one use is 1 minute with a 10 minutes pause between uses.
KaWe EUROLIGHT E10/ E15/ E16
- Light outlet (dustproof)
- Blend wheel (except E10 model)
- Inspection window
- Correction lens wheel with 18 lenses (± 20 diopters)
- Correction value indication (+) black; (-) red
- Plug connector
- The light regulator on/off
- Handle with battery
- End cover
- a) Place the ophthalmoscope head on the handle with the battery (8) and secure it with the screw. Consider the location of the holes on the head for securing the screw.
b) Insert batteries or accumulators into the handle. Observe the polarity and the lamp voltage (2.5 or 3.5 V)! For detailed information see Section 8: Maintenance and Service.
c) Switching on is carried out by pressing the white switch fixing the on position and moving the light regulator to the left (7). To switch it off, move the light regulator to the initial position 0 (until it clicks!). Brightness can be adjusted smoothly.
- Light outlet (dustproof)
- Blend wheel (except E30 model)
- Inspection window
- Correction lens wheel with 18 lenses (± 20 diopters)
- Correction value indication (+) black; (-) red
- Clic connector
- The light regulator on/off
- Handle with battery
- End cover
-
a) Place the ophthalmoscope head on the handle with battery (8) and secure it until it clicks. Observe the position of the cutout.
b) Insert batteries or accumulators into the handle. Observe the polarity and the lamp voltage (2.5 or 3.5 V)! For detailed information see Section 8: Maintenance and Service -
c) Switching on is carried out by pressing the white switch fixing the on position and moving the light regulator to the left (7). To switch it off, move the light regulator to the initial position 0 (until it clicks!). Brightness can be adjusted smoothly.
- Closing blend
- Blend wheel
- Filter 4000 K (F)
- Filter arm
- Polarizing filter for lighting (P)
- Attachment lens wheel
- Correction value indication (+) green; (-) red
- Correction lens wheel 71 Lenses (-36 to +35 diopters)
- Inspection window
- Polarizing filter for inspection window
Attention: When the ophthalmoscope is not operated, close the inspection window (9) with a blend (1) to prevent dust from entering the head!
Blend| Explanations|
---|---|---
| Gap
Small circle
Big circle
Fixing point
Green filter (without red color)
|
Attachment lens wheel position| Diopter range
---|---
| -12 to +11 diopters
-13 to -36 diopters
+12 to +35 diopters
E35 head installation
see KaWe EUROLIGHT E30/E36
E25 head: KaWe MedCenter 5000 installation
The KaWe EUROLITE E25 head can only be used with the KaWe MedCenter 5000 wall station.
- Insert the E25 head into the handle of the KaWe MedCenter 5000 device.
- Note that the centering hole must be lined up with the support on the retaining screw.
- c) Tighten the retaining screw lightly.
- d) To remove the head, unscrew the retaining screw and carefully pull out the head.
- Observe the KaWe MedCenter 5000 manual instructions.
- Light outlet (dustproof)
- Blend wheel (except E50 model)
- Inspection window
- Correction lens wheel with 18 lenses (± 20 diopters)
- Correction value indication (+) black; (-) red
- Thread
- On/off slide switch
-
a) Install batteries into the handle (2 pcs. AA 1.5V minion). Observe the polarity. For detailed information see Section 8: Maintenance and Service
b) Screw the head onto the handle (6). -
Move the slide switch (7) to turn the lamp on and off .
-
Battery and lamp replacement is possible after unscrewing the head. For detailed information see Section 8: Maintenance and Service
Preparing for operation
- Periodically check that all parts are in good condition -replace defective parts.
- Check the smooth running of moving parts (e.g. correction lens /blend wheel, plug connector).
- Keep all electrical contacts clean.
- Check the battery’s charge.
- After a successful inspection, the ophthalmoscope is ready for use.
Hygienic preparation
Keep the ophthalmoscope in its packaging to protect it from dirt and dust.
Manual cleaning/disinfection
Protect the device from moisture!
- Do not place the ophthalmoscope in liquids, make sure that no liquid gets into the casing.
- Clean the outside of the ophthalmoscope with a damp, soft, lint-free cloth. If necessary, wipe the light outlet glass with a cotton swab slightly moistened with alcohol.
- Do not press on the glass.
- Verwenden Sie keine scheuernden Reinigungsmittel!
- Do not use abrasive cleaning agents. For disinfection, the ophthalmoscope can be wiped with an alcohol-based surface disinfectant. Use a slightly damp cloth.
Automatic cleaning
Not applied.
Sterilization
Not applied.
Transportation and storage
Ophthalmoscopes are to be stored and transported protected from dust, moisture
and contamination.
Environment parameters
| ****
Temperature
| Relative humidity| Air pressure
---|---|---|---
Operation| +10°C to
+35°C
| 30% to 90%| 800 GPa to
1060 GPa
Storage| ****
-10°C bis
+55°C
| ****
10% to 95%
| 700 GPa to
1060 GPa
Trans- portation| 500 GPa to
1060 GPa
Maintenance and Service
The end of the useful life is usually determined by wear and tear. Damaged parts are to be sorted out and replaced immediately. If the light intensity decreases, check for the following possible causes:
- Battery charge; replace or recharge if necessary.
- Clean or replace the lamp.
- If the above reasons are excluded, repair or replacement may be required. In this case, please, contact a specialized dealer.
Battery replacement
Check the batteries charge regularly. Low battery voltage (depending on
consumption) leads to a light intensity decrease and the possible appearance
of LED lamps flashing. In both cases, it’s necessary to replace the batteries.
When replacing, use only new, high-quality alkaline batteries (or fully
charged batteries).
-
Besides, observe the manuals’ requirements for use of the batteries.
-
Installation of batteries for
KaWe EUROLITE, 2 pcs. medium size 1.5V (type C) -
Installation of batteries for KaWe EUROLITE
-
Installation of batteries for KaWe PICCOLIGHT, 2 pcs. minion type 1.5V (type AA)
- Unscrew the retaining screw on the handle and remove the ophthalmoscope head from the handle.
- Unscrew the cap ring (1).
- Pull out the lamp holder (2) on the bottom of the head and replace the lamp.
- Clean the glass bulb (3) of the new lamp with alcohol. The glass bulb must be clean and free from fingerprints (grease free).
- Install the lamp holder and lamp back into the head.
- Screw the cap ring.
KaWe EUROLIGHT E15/ E16/ E25/ E30/ E35/ E36
- Remove the ophthalmoscope head from the handle.
- Models E15/E16:
- Unscrew the retaining screw on the handle.
- Models E25/E30/E35/E36: Push down on the wickered sleeve.
- Remove the lamp at the bottom of the head.
- Clean the glass bulb of the new lamp with alcohol. The glass bulb must be clean and free from fingerprints (grease free).
- Insert the lamp into the head until it stops.
KaWe PICCOLIGHT E50/ E56
- Unscrew the ophthalmoscope head from the handle.
- Remove the lamp at the bottom of the head.
- Clean the glass bulb of the new lamp with alcohol. The glass bulb must be clean and free from fingerprints (grease free).
- Insert the lamp into the head until it stops.
Accessories and Spare parts
Accessories and spare parts can be found in the general catalog of our company at https://www.kawemed.com/download/broschueren/ or can also be viewed at your specialist dealer.
Disposal
Defective and/or recyclable electrical/electronic devices and rechargeable batteries must be transferred to a designated waste collection point for recycling.
Instructions for service personnel and patients
In case of problems during the device use, it is necessary to immediately
contact the manufacturer and the competent authorities of the re-levant member
state of the Union, in which the service personnel or the patient are located.
Annex
Warranty & Guarantee
In case of proper handling and adherence to the rules laid down in this
application manual, as well as the use of original KaWe spare parts, the
statutory warranty period of two years from the date of sale is valid
(excluding lighting equipment, batteries and accumulators). LED lamps are
warranted for five years from the date of sale.
For other questions and if repair is required, contact your speci-alized
dealer.
Contact information
-
Please contact your specialist dealer or directly the manufacturer (see back for address).
-
Basic UDI-DI
-
4030155KaWe0102YT
-
QM-1-176I / B-27182 / 2021-03
-
KIRCHNER & WILHELM GmbH + Co. KG Eberhardstraße 56
71679 Asperg, Germany
Phone: +49 7141 68188-0 -
Fax: +49 7141 68188-11
-
All information is without guarantee and subject to change.
Documents / Resources
| KaWe
EUROLIGHT Ophthalmoscope Pocket Set Two
Handles
[pdf] User Manual
EUROLIGHT Ophthalmoscope Pocket Set Two Handles, EUROLIGHT, Ophthalmoscope
Pocket Set Two Handles, Pocket Set Two Handles, Two Handles
---|---
References
- Broschüren – KaWe Medizintechnik
- KaWe Medizintechnik - Kompetenz und Qualität seit 1890
- Broschüren – KaWe Medizintechnik
- KaWe Medizintechnik - Kompetenz und Qualität seit 1890
- Download – KaWe Medizintechnik
- Broschüren – KaWe Medizintechnik
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>