embio diagnostics BELD Research Lab Starters Kit User Manual
- June 3, 2024
- embio diagnostics
Table of Contents
embio diagnostics BELD Research Lab Starters Kit
Materials Supplied in kit
- BELD Device
- Disposable electrodes
- Biosensors
Equipment required (not provided)
- Incubator (30-37οC)
- Micropipettes (20-200 μL, 100-1000 μL)
- Stomacher bags
- LX broth (225 mL / sample) [Biomerieux, ref. 42626]
- Agar Listeria according to Ottaviani and Agosti – Api Listeria Rapid Kit
Sample preparation and handling
SAMPLE TRANSPORT AND STORAGE
For sample transport please store samples in a cool box between 1 and 8οC.
Samples should be tested as soon as possible aPer collecQon, preferably no
later than 24 h aPer arrival at the laboratory and no later than 36 h aPer
sampling (EN ISO 7218 Standard). The Qme duraQon before analysis should be
recorded and wri[en in the analysis report.
TEST PROCEDURE
- Add a test por)on of 25 g or 25 mL to 225 mL of selec)ve enrichment broth (LX broth) to obtain a tenfold dilu)on.
- Homogenize.
- Incubate the suspensions at 30oC for 22-24h (Figure 1).
- Test the samples mixing 20 μL of the suspensions with 20 μL of the biosensors (see Realiza)on of BRaD LMO Test below).
- In case of Posi9ve result, proceed to confirma)on tests (Figure 2).
Realization of BRaD LMO test
Connect the screen-printed electrode with the device and turn on the device. Be sure that the electrodes are aligned with the connectors.
Turn on the Bluetooth functionality and location services on your mobile phone, open the application and login
Press the search command and connect the application to the device.
Tap the connected device and fill in the appropriate fields (select your Sample and Sample level).
START the measurement and then add 20μL (volume can be adjusted according to the application) of the sample onto channels 1-8 with a mul>channel or single- channel automatic pipette.
ATTENTION, if the lines on the chart do not appear aSer pressing the start buOon do not add the sample. Stop the measurement and start it again (a). Ensure that the lines are visible in the chart and they all start at 0 mV before adding the sample (b). In case lines do not start at 0 mV (c) stop the measurment and start it again.
Once the first minute has passed, immediately start adding 20μL of the biosensors onto the same channels as above hence the final volume in each channel will be 40μL.
After the END of the measurement, the result will appear on your mobile screen and the data will be uploaded automatically.
For each new measurement press Restart and then press EDIT (pencil icon found in the top right hand corner of the screen) to add new informa>on for the next sample. You will now be ready to start the next measurement.
Warnings and precautions
- For in vitro diagnostic use.
- Do not use the biosensors beyond the expiration date.
- Use appropriate precautions in the collection, handling, storage, and disposal of samples.
- Use of Nitrile, Latex (or equivalent) gloves is recommended when handling samples.
- Test electrodes are for one use only.
- The user should not open the tube containing the biosensors exposing it to the ambient environment until is ready for immediate use.
- If the biosensor solution contacts the skin or eye, flush with copious amounts of water.
- To obtain accurate results, the Package Insert instructions must be followed.
- Inadequate or inappropriate sample collection, storage, and transport may yield false test results.
- Sample collection and handling procedures require specific training and guidance.
- Use the appropriate Fixed Volume Pipette in accordance with test procedures.
- To obtain accurate results, the test should not be used inside a laminar flow hood or in a heavily ventilated area.
- Testing should be performed in an area with adequate ventilation.
- Dispose of containers and unused contents in accordance with local regulatory requirements.
- Wear suitable protective clothing, gloves, and eye/face protection when handling the contents of this kit.
- Wash hands thoroughly after handling.
Troubleshooting
Faults | Actions |
---|---|
User enters false Credentials | Error message appears. User must enter the |
right credentials.
User forgot his password| Can press “Forgot Password?” bu[on to restore the
password.
User (email) already exists| Error message appears. Must use a different
password.
Wrong password. a) Password does not meet the requirements (e.g. short length)
b) Confirm password does not match.| Error message appears. Must enter Password
that meets the requirements.
Wrong information accidentally entered| Completed measurements can be found in
the results tab. Edit is available.
---|---
Bluetooth is off| Turn on Bluetooth
Location Permission disabled| User must navigate to app permissions and enable
Location.
No results found| Check if there are acQve BELD devices nearby. Beld device
must not be connected to another mobile app.
Cannot connect to BELD device| a) BELD Must not run on low battery b) BELD
device must be in close proximity (<2m) to the mobile phone and no obstacles
located between them c) If a and b does not fix the problem then restart both
the BELD and mobile app.
Accidentally wrong parameters| a) Sample code and Sample Level can be edited
aPer the measurement has finished. b) If the sample is incorrect, the
measurement must run again (each sample corresponds to a different algorithm)
c) Measurement Qme must be >3 minutes.
The chart remains empty aPer the start bu[on is pressed| a) Stop measurement
and check if BELD is sQll connected. If it is connected press start again. If
it is disconnected or pressing start does not fix the problem then start again
with a new
measurement.
Result is stuck in “Uploading”| a) Check your internet connection b) If no
result appears in 2 minutes, the measurement must be taken from the beginning.
User enters wrong email data| Ιf “share via email screen” is sQll open a new
email can be sent, otherwise the measurement must be taken from the beginning.
User enters wrong phone number| Ιf “share via email screen” is still open a
new sms can be sent, otherwise the measurement must be taken from the
beginning.
Email or SMS has not arrived| Email and phone must be in the correct form.
Possibility of accessing the user’s data| Only non-sensitive data is stored
and encrypted in the database. Communication via the application is also
encrypted via SSL.
Kit storage and stability
Store the kit at 4oC temperature. On the day of the test let the biosensors
and samples warm up to room temperature. Cold samples can lead to erroneous or
invalid results.