Ustar Biotechnologies U202025-1 COVID-19 Rapid PCR Test Instruction
Manual
[INTENDED USE]
The GenuIN COVID-19 Rapid PCR Test is a single-use test kit intended to detect
SARS-CoV-2 that causes COVID-19.
This test is for home use with self-collected anterior nasal swab specimens in
individuals aged 14 years and older (self-collected) or individuals aged 2~13
years (collected by an adult) with suspected COVID-19. The assay is only used
for auxiliary diagnosis of SARS-CoV-2 infection.
This assay applies isothermal amplification and the nucleic acid lateral flow
assay to detect and identify specific sequences of SARS-CoV-2 RNA. This assay
is similar to a PCR test in that it utilizes a molecular amplification
technology for the detection of SARS-CoV-2 viral RNA. SARS-CoV-2 viral RNA is
generally detectable in anterior nasal swab specimens during the acute phase
of infection. In general, in vitro testing for SARS-CoV-2 infection is most
accurate in the first week after symptom onset. Individuals diagnosed with
COVID-19 within 3 months are recommended to first seek medical advice of
testing because SARS-CoV-2 RNA is detectable in upper respiratory tract
specimens even for weeks after the onset of symptoms.
Persons who test positive with the GenuIN COVID-19 Rapid PCR Test should seek
follow up
care with their physician or healthcare provider as additional testing and
public health reporting may be necessary. Positive results do not rule out
bacterial infection or co-infec- tion with other viruses. Persons who test
negative and continue to experience COVID-19 like symptoms of fever, cough
and/or shortness of breath may still have SARS-CoV-2 infection and should seek
follow up care with their physician or healthcare provider.
COVID-19 is an acute respiratory infectious disease. All individuals infected
COVID-19 are a source of infection.
Based on the current epidemiological investigation, the incubation period is 1
to 14 days, mostly 3 to 7 days. The main manifestations include fever, fatigue
and dry cough. Nasal congestion, runny nose, sore throat, myalgia and diarrhea
are found in a few cases
It is difficult to differentiate between COVID-19 and respiratory infections
caused by other
-Take out all components from the package.
[TEST PRINCIPLE]
The assay applies cross priming amplification (CPA) and nucleic acid lateral
flow to qualitatively detect the specific sequence of SARS-CoV-2 RNA (genes
ORF1ab and N). With the self-driven heating module, the assay can perform
isothermal amplification reactions through specific amplification primers,
probes, RNA reverse transcriptase, and DNA polymerase with high strand
displacement activity. The viral RNA was first transferred to DNA with RNA
reverse transcriptase and then amplified with DNA polymerase. The assay
consists of a module preloaded with nucleic acid amplification reagent and
corresponding lateral flow strip.
The paper strip is used for rapid detection of nucleic acid amplification
products by chromatographic double antibody sandwich method. Biotin and FAM
probe are modified at both ends of one specific primer in the reaction system.
For positive samples, due to capillary forces, reported nucleic acid fragment
flow from sample pad side to detection pad, then the FAM end of nucleic acid
will combine with colloidal gold particles and binds specifically to the T-line
with secondary antibody. In the process of chromatography, the other end is
captured by streptavidin (line C), so line C with T line color at the same
time. For negative samples, only one end of modified biotin is captured by
streptavidin (line C) during chromatography, so only C line is colored. If
there is an operation error, only T, or both T and C are not colored, the
result is regarded as invalid.
[MATERIALS PROVIDED
No. | Materials Provided | Quantity |
---|---|---|
1 | Disposable sampling swab | 1 |
2 | Sample tube |
(Prefilled with lysis buffer)
| 1
3| Dripper| 1
4| Buffer A (Prefilled in bubbles)| 2
5| Test cassette| 1
6| Zip-lock bag| 1
Note:
- Do not mix materials from different
- Do not reuse the materials of the
[STORAGE CONDITIONS & VALIDITY PERIOD]
- Usage: the kit should be used at 2~30℃.
- Storage: the kit should be stored at 2~8℃.
- Validity period: 6 Refer to the package for the production date and expiry date. 4.Transportation: The kit performance will not be affected by temperature 2~30℃ transpor- tation within 15 days.
[SAMPLE COLLECTION AND HANDLING]
- Sample types: anterior nasal swabs
- Sample storage
The samples should be sent for testing as soon as possible after collection.
[TEST PROCEDURE]
Note: Read this IFU carefully before using.
-
Prepare for test
-Take out all components from the package. -
Tear off the seal of the sample tube
-
Swab Both Nostrils
-Take out the swab and hold with handle end.
-Tilt head back and gently insert swab tip(about 18mm) until it is fully inside your nostril.
-Roll the swab tip 5~10 times around the inside walls of your nostril.
The swab tip should be touching the walls of the nostril as you rotate.
-Repeat swab step in the other nostril.
Insert swab into the sample tube and make sure the swab tip is fully immersed
in the lysis buffer.
-Stir the swab and press the tube bottom for 15 times to release the sample fully.
-Dispose of the swab into the ziplock bag.
-Put on the dripper.
- Load Sample and Run Test
-Take out the test cassette and place it on a clean flat surface with the side of sample hole facing up.
-Open the cap of sample hole.
-Press the sample tube lightly to add only one drop lysis buffer into the sample hole.
-Open one bubble of buffer A and add all of it into the sample hole.
-Cover the cap of test cassette.
-Connect the cassette with power supply by the USB Type-C cable and connector (output voltage: 5V).
-If the blue light is on, the cassette is working; otherwise, please change the power supply.
-Do not move the cassette when the test is running.
-Wait 35 minutes
-Pull out the Type-C connector and open the cap of sample hole.
-Open another bubble of Buffer A and add all of it into the hole, then close with the cap again.
-The result can be read in 5 min
Reading the result
Read the test result according to the notice provided on the test cassette.
Note: Please make sure to check the test result within 30 minutes after adding
the second bubble of Buffer A.
[INTERPRETATION OF RESULTS]
The result of COVID-19 test is shown in the “Test” area of result window. The
kit contains internal control material to confirm the lateral flow worked. The
result of internal control is
shown in the “Control” area of result window
Positive: If a clear visible line is seen in “Test” area of the result window, the nucleic acid of SARS-CoV-2 is detected in the sample, namely the result is positive. In positive results, sometimes the lines in “control” area could be light. In this case, the results are still valid.
Negative:
If there is only one line presented in “Control” area and no visible line is seen in “Test” area, the nucleic acid of SARS-CoV-2 is not detected in the sample, namely the result is negative.
Invalid:
If there is no visible line in both areas of the result window, test is failed and the result is invalid.
[LIMITATIONS]
-
Persons who test positive with the GenuIN COVID-19 Rapid PCR Test should seek follow
up care with their physician or healthcare provider as additional testing and public health reporting may be necessary. Positive results do not rule out bacterial infection or co-infec- tion with other viruses. Persons who test negative and continue to experience COVID-19 like symptoms of fever, cough and/or shortness of breath may still have SARS-CoV-2 infection and should seek follow up care with their physician or healthcare provider. -
Inappropriate collection, transport and treatment of samples or low titer of pathogens may yield false negative
-
This product has been validated for current known Future mutations of target nucleic acid may yield false negative result.
-
Interference or amplification inhibitors other than those presented in [PERFORMANCE CHARACTERISTICS] 3.2 Interfering substances may yield false negative result.
-
Users are advised not to eat, drink or take unnecessary nasal drugs before self-testing.
-
Positive results do not rule out co-infection of bacteria or other
[PERFORMANCE CHARACTERISTICS]
- Limit of detection 3000 copies/mL
- Precision
- Intra-batch precision
Through the analysis on the results of repeatability samples from the same batch within 20 days, the detection rates of medium positive samples, weak positive samples and negative samples are all 100%, and the coincidence rate of negative samples is 100% (n ≥ 20).
- Within-batch precision
Through the analysis of the 20-day test results of the same operator in different batches, the 20-day detection rates of medium positive samples were all 100%, the detection rate of weak positive samples was 100%, and the coincidence rate of negative samples = 100% (n ≥ 20).
- Specificity
- Cross-reactivity
Cross-reactivity (organisms tested in the absence of SARS-CoV-2)
The specificity of the assay was evaluated in cross-reactivity testing using 25
commensal organisms. The cross-reactivity testing confirmed that none of the 25
organisms were
cross reactive with the GenuIN COVID-19 Rapid PCR Test at the concentrations
tested. The 25 commensal organisms are listed below:
Bacteria and viruses| Test concentration| Bacteria and viruses| Test
concentration
---|---|---|---
Bordetella pertussis| 10⁶ bacteria/mL| Staphylococcus aureus| 10⁶
bacteria/mL
Respiratory Syncytial virus| 10⁶ copies/mL| Chlamydia pneumoniae| 10⁶
ccu/mL
Adenovirus| 10⁶copies/mL| Rhinovirus| 10⁶ copies/mL
Influenza A virus| 10⁶ copies/mL| Bocavirus| 10⁶ copies/mL
Influenza B virus| 10⁶ copies/mL| Coronavirus NL63| 10⁶ copies/mL
Parainfluenza virus| 10⁶ copies/mL| Mycoplasma pneumoniae| 10⁶ ccu/mL
Epstein-Barr virus| 10⁶ copies/mL| Coronavirus 229E| 10⁶ copies/mL
Coronavirus OC43| 10⁶ copies/mL| Coronavirus HKU1| 10⁶ copies/mL
Legionella pneumophila| 10⁶ bacteria/mL| MERS coronavirus| 10⁶
copies/mL
Streptococcus pneumonia| 10⁶ bacteria/mL| SARS CoV-1| 10⁶ copies/mL
Human Metapneumovirus| 10⁶ copies/mL| Candida albicans| 10⁶
bacteria/mL
Klebsiella pneumonia| 10⁶ bacteria/mL| Staphylococcus epidermidis| 10⁶
bacteria/mL
Haemophilus influenzae| 10⁶ bacteria/mL| /| /
- Interfering substances
The following substances have no effects on test results of this product.
- Endogenous interfering substances: human plasma (1% v/v), human mucoprotein (1% v/v), Human genome DNA (300ng).
- Exogenous interfering substances: Phenylephedrine, Methotrexate, Sodium
chloride, Beclomethasone, Dexamethasone, Flunisolide, Triamcinolone acetonide, Budesonide, Mometasone,Fluticasone,Histamine hydrochloride, Interferon α, Zanamivir, Oseltamivir, Peramivir, Arbidol, Lopinavir, Ritonavir, Mupirocin, Levofloxacin, Azithromycin, Ccephalosporin, Minocycline, Tobramycin, Azelastine, Vitamin A ointment, D-panthenol ointment, Biotin
- Clinical performance
Clinical performance of GenuIN COVID-19 Rapid PCR Test was determined by testing 109 positive and 508 negative specimens. 95.413% of individuals with positive real-time PCR tests were tested positive by GenuIN COVID-19 Rapid PCR Test. 99.803% of individuals with negative real-time PCR tests were tested negative by GenuIN COVID-19 Rapid PCR Test. And the total coincidence rate was 99.028%
RT-PCR | Total | |
---|---|---|
Positive | Negative | |
GenuIN COVID-19 Rapid PCR Test | Positive | 10 4 |
Negative | 5 | 50 7 |
Total | 10 9 | 50 8 |
Relative Sensitivity: 95.413% (95%CI: 89.619% ~ 98.494%)
Relative Specificity: 99.803% (95%CI: 98.908% ~ 99.995%)
Accuracy: 99.028% (95%CI: 97.895% ~ 99.642%)
[PRECAUTIONS]
Please make sure to read this Instruction for Use carefully before use.
This product is for auxiliary diagnosis in vitro only. It shall not be used as
the only basis for confirmed diagnosis.
Please integrate other test methods or clinical symptoms for comprehensive
consideration.
- Testing
This product is for disposable use. Please use it in accordance with this instruction.
- Do not squeeze or press the test cassette while operating
- Do not move the test cassette while testing.
- Read test results within 30 minutes after adding the second bubble of buffer A, otherwise it may yield false positive result.
- Operation
The sample and other components contacted with the sample (e.g. used fragment of the tabletop, timer surface) may be a source of infection even if the test is negative and should be disinfected. Hands also should be washed or disinfected. Throughout the testing process, users are advised to wear masks and gloves to prevent infection.
This test kit is a disposable in vitro diagnostic product. Any experimental wastes such as test cassettes, gloves, unused samples or reagent, etc., which have potential biological hazards, should be disposed of in accordance with biological safety regulations, environ- mental protection regulations or medical waste regulations.
- If the reagent mistakenly enters the eyes or mouth or sprays on the skin, please wash with clean water and seek help from doctors if
- Please make sure there is no liquid or other matters affixed on the outside surface of the test cassette before test.
- Proper sample collection and sample handling are essential for correct
- Storage and use
- This product must be stored at the condition as required in this
- Do not use the expired
- Before test, please make sure there is no rift on the test cassette or leakage of the
- Please do not use this product for purposes not described in this
- Please store the product properly and prevent children from touching
- Disposal
- The pollution level of this kit is PD1 (No pollution or only dry non-conductive pollution).
- The used product should be sealed in the zip-lock bag and be disposed of as the medical
- Caution
FCC Compliance Statement:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
- this device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired
changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
—Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 18 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference,
- this device must accept any interference received, including interference that may cause undesired
[EXPLANATION OF SYMBOLS]
GenuIN Biotech LLC
Address: 2270 Kraft Drive, Suite 1311, Blacksburg, VA 24060, USA Customer
service: +01-540-605-9777
CAUTION – Investigational device. Limited by Federal law to investigational
use.
[INSTRUCTION VERSION AND MODIFICATION DATE]
Approved on April 25, 2022. Version: A0