GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- June 3, 2024
- GARNET
Table of Contents
- CHAPTER 1 – INTRODUCTION
- CHAPTER 2 – FEATURES
- CHAPTER 3 – WIRING DIAGRAMS
- CHAPTER 4 – DISPLAY PROGRAMMING
- CHAPTER 5 – SERIAL INTERFACE
- CHAPTER 6 – TROUBLESHOOTING GUIDE
- CHAPTER 7 – SPECIFICATIONS
- CHAPTER 8 – SERVICE & WARRANTY INFORMATION
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
CHAPTER 1 – INTRODUCTION
Congratulations on purchasing the Garnet Instruments SEELEVEL Access™ Data Portal. The SEELEVEL Access™ compliments the SEELEVEL Annihilator™ 806-B, 806-Bi, or SEELEVEL Special™ 808-P2 and SeeLeveL ProSeries II 810-PS2 gauges by providing an additional volume readout in the cab of your truck.
In addition to providing readout of the tank level, the SEELEVEL Access™ provides a 4-20 mA analog output proportional to the fluid volume displayed. This analog output can be used to communicate the tank level to other pieces of equipment such as fleet management systems or Electronic Logging Devices (ELD).
The full scale value of the analog output can be set using the buttons on the back of the display, no additional equipment is required for calibration.
The SEELEVEL Access™ also contains a serial RS-232 interface which allows fleet management or ELD systems to gather fluid volume data from the gauge. The interface is full duplex and contains security features to prevent unauthorized access.
The SEELEVEL Access™ display has been designed to withstand the vibration and shock encountered in mobile applications. While the 808-P2 and 810-PS2 operate on internal batteries, (12 volt truck power is used to operate the back light and external alarms), the SEELEVEL Access display operates on 12V truck power.
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- Remote Display & Data Portal](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-Remote-Display-Data-Portal.jpg)
CHAPTER 2 – FEATURES
The SEELEVEL Access™ has been uniquely designed for specific applications and
with specific features:
Standard SEELEVEL Access Features
-
The signal between the 806-B, 806-Bi, 808-P2 or 810-PS2 display and the SEELEVEL Access™ is digitally encoded so the signal line can be connected using a standard 7 pin trailer plug.
-
The display operates on 12 volt truck power, and draws less than 150 mA.
-
All-digital design (except the 4-20 mA output) eliminates reading drift or degradation, ensuring long-term accuracy under all operating conditions.
-
Operation from -40 °C to +60 °C (-40 °F to +140 °F) ambient temperature.
-
Easy installation and servicing with a one-year limited warranty.
Additional SEELEVEL Access T-DP0301-A Features -
An analog 4-20 mA output, with 4 mA corresponding as zero displayed volume, and 20 mA corresponding to the full scale displayed volume programmed into the remote display.
-
An RS-232 serial interface available to connect to a a variety of ELD or fleet management systems.
-
The SEELEVEL Access™ provides an easy-to-read LED display inside a compact, edge-view enclosure, optimized for top-of-dash or overhead console mounting. The display is housed in an aluminum enclosure 2.7″ wide x 1.1″ high x 3.4″ deep (68 mm wide x 29 mm high x 87 mm. deep).
-
A dimmer button switch enables the operator to control brightness.
-
Simple 6 wire electrical installation – 12V power (red), ground (black), gauge signal (yellow), analog output (white/blue), serial receive (purple) and serial transmit (grey).
CHAPTER 3 – WIRING DIAGRAMS
THE SEELEVEL Access™ has been designed for easy installation with your 806-B, 806-Bi, 808-P2 or 810-PS2 series SEELEVEL™ gauge. Installation instructions are available online at www.garnetinstruments.com. The SEELEVEL Access™ Remote Display is easy to install:
808-P2 Wiring Diagram
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- 808-P2 Wiring Diagram](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-808-P2-Wiring-Diagram.jpg)
810-PS2 Wiring Diagram
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- 810-PS2 Wiring Diagram](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-810-PS2-Wiring-Diagram.jpg)
806-B Wiring Diagram
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- 806-B Wiring Diagram](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-806-B-Wiring-Diagram.jpg)
806-Bi Wiring Diagram
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- 806-Bi Wiring Diagram](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-806-Bi-Wiring-Diagram.jpg)
CHAPTER 4 – DISPLAY PROGRAMMING
The SEELEVEL Access™ display shows the tank level by repeating the information shown on the 806-B, 806-Bi, 808-P2 or 810-PS2 gauge. The 4-20 mA analog output is calculated from the display level with a 4 mA output corresponding to a display level of zero, and a 20 mA output corresponding to the full scale level programmed into the SEELEVEL Access™ display.
For example, if the full scale is programmed to be 500.0, then a display value of 400.0 will result in an analog output of 16.80 mA. The display will recognize the decimal location and adjust the output accordingly, so in this example a display value of 400 will also result in an analog output of 16.80 mA.
To set the full scale level:
- Determine the maximum volume that can be displayed and pick a full scale amount that is equal to or greater than this volume.
- Press both the NEXT MENU and UP/ENTER buttons on the back panel, the display will show ACAL. Release both buttons.
- The display will show the existing calibration with the left digit bright. Press the UP/ENTER button to change the bright digit. Press the NEXT MENU button to go to the next digit.
- Set all 4 digits, then press NEXT MENU again to set the decimal point, it will be bright to indicate that it is selected. Press the UP/ENTER button to select either x.xxx, xx.xx, xxx.x or no decimal. For best accuracy of the analog output, try to use all 4 digits such as 500.0 instead of just 500.
- After the decimal as been set, press NEXT MENU again, the display will show Stor. Press UP/ENTER to store the calibration and exit the setting menu. The display will continue to show Stor for a moment and then will show donE for a second. Then normal operation resumes.
- If you do not want to store the calibration, press NEXT MENU again and the display will show Abrt. Press UP/ENTER to abort which exits the calibration menu without saving.
- If you press NEXT MENU again from the Abrt display, the menu will return to the beginning with the left digit selected by being bright.
- If the full scale calibration is below 103, the display will be unable to calculate a valid calibration, and will show cErr (calibration error) after a few seconds. The existing calibration will be retained, and the display will return to normal operation.
To view the existing calibration:
- Press either the NEXT MENU or UP/ENTER button (but not both) on the back panel, the display will show the existing full scale analog calibration while the button is held down. Release the button to return to normal operation.
To test the analog output:
- While either the NEXT MENU or UP/ENTER button on the back panel is pressed, the display will show the full scale calibration and the analog output will go to full scale (20 mA). This can be used to test or calibrate the equipment connected to the analog output.
- While the display is in the calibration mode (entered by pressing both the NEXT MENU and UP/ENTER buttons) the analog output will be at 4mA.
CHAPTER 5 – SERIAL INTERFACE
SeeLeveL ELD Portal Format and Signal Format
- The supported signal format is bidirectional serial (separate TX and RX lines), RS232 voltage levels, 9600 baud, 8 bit, no parity, 1 stop bit.
- All messages obey the following format: [start sequence] [total number of bytes in message] [message ID] [payload – optional] [CRC] [stop sequence]
- All multi-byte parameters are transferred big-endian (MSB first)
- Start sequence: [0xFE][0xFE][0x24]
- Total number of bytes in message (1 byte)
- Message ID (1 byte)
- Payload (optional depending on the message)
- CRC (1 byte) = direct sum of all preceding bytes, truncated to 1 byte
- Stop sequence: [0xFF][0xFF][0x2A]
SeeLeveL Query Message (ELD - > SeeLeveL)
- Value: 0x00
- Allows the ELD to query the SeeLeveL device
- [0xFE][0xFE][0x24][0x09][0x00][0x29][0xFF][0xFF][0x2A]
SeeLeveL Query Response (SeeLeveL - > ELD)
- Value: 0x01
- SeeLeveL responds with model ID (1 byte), H/W Rev (1 byte), S/W Rev (2 bytes), alarm capability (1 byte), and SN support (1 byte). If the SeeLeveL device supports a unique serial number, it will follow (8 bytes in length).
- Example: SeeLeveL model ID = 0x01, hardware rev = `E’ (0x45), software major rev = 0x05, minor rev = 0x09, no alarm capability = 0x00 (0x01 = alarm capable), serial number supported = 0x01 (serial number not supported = 0x00), and a serial number = 0x0102030405060708:
- [0xFE][0xFE][0x24][0x17][0x01][0x01][0x45][0x05][0x09][0x00][0x01] [0x01][0x02][0x03][0x04][0x05][0x06][0x07][0x08][0xB1][0xFF][0xFF] [0x2A]
SeeLeveL Handshake Demand Message (SeeLeveL - > ELD)
- Value: 0x02, 1 byte payload
- ELD must respond with the proper coded response in order to either begin or continue the transfer or broadcast of liquid levels from the SeeLeveL device. Handshake demands will be broadcast at random times.
- Example:
- [0xFE][0xFE][0x24][0x0A][0x02][0x3E][0x6A][0xFF][0xFF][0x2A]
ELD Handshake Response (ELD - > SeeLeveL)
- Value: 0x03, 1 byte payload
- To calculate the response, the payload from the SeeLeveL Handshake Demand Message is used as an address/offset to fetch the contents of the lookup table:
- Response to example above: ·
- [0xFE][0xFE][0x24][0x0A][0x03][0x85][0xB2][0xFF][0xFF][0x2A]
- Contact Garnet Instruments at 1-800-617-7384 or at info@garnetinstruments.com to set up the appropriate working relationship. Once this is established, the handshake response table will be provided.
Send Liquid Level Message (ELD - > SeeLeveL)
- Value: 0x04, no payload
- SeeLeveL responds with a single liquid level or a handshake demand message.
- [0xFE][0xFE][0x24][0x09][0x04][0x2D][0xFF][0xFF][0x2A]
Begin Liquid Level Broadcast Message (ELD - > SeeLeveL)
- Value: 0x05, no payload
- SeeLeveL responds with a liquid level or with a handshake demand message.
- [0xFE][0xFE][0x24][0x09][0x05][0x2E][0xFF][0xFF][0x2A]
Stop Liquid Level Broadcast Message (ELD - > SeeLeveL)
- Value: 0x06, no payload
- SeeLeveL will cancel any further broadcast of liquid level.
- [0xFE][0xFE][0x24][0x09][0x06][0x2F][0xFF][0xFF][0x2A]
SeeLeveL Query Alarm Liquid Level Message (ELD - > See LeveL)
- Value: 0x07, no payload
- SeeLeveL will respond with a liquid alarm level response or an error response if the alarm function is not supported.
- [0xFE][0xFE][0x24][0x09][0x07][0x30][0xFF][0xFF][0x2A] See LeveL Liquid Alarm Level Response (SeeLeveL -> ELD)
- Value: 0x08, 7 byte payload
- SeeLeveL responds with liquid alarm level (4 bytes = unsigned int32), number of digits to the right of the decimal (1 byte), alarm type (1 byte; high = 0x01, low = 0x00), and whether liquid level is currently in alarm (1 byte; alarm active = 0x01, no alarm = 0x00).
- Example: liquid alarm level = 347.56, alarm type = low level, alarm active:
- [0xFE][0xFE][0x24][0x10][0x08][0x00][0x00][0x87][0xC4][0x02][0x00] [0x01][0x86][0xFF][0xFF][0x2A]
SeeLeveL Query Alarm Status Message (ELD - > SeeLeveL)
- Value: 0x09, no payload
- SeeLeveL will either respond with current alarm status or an error response if the alarm function is not supported.
- [0xFE][0xFE][0x24][0x09][0x09][0x32][0xFF][0xFF][0x2A]
SeeLeveL Query Alarm Status Response (SeeLeveL - > ELD)
- Value: 0x0A, 1 byte payload
- SeeLeveL responds with current alarm status (1 byte; alarm active = 0x01, no alarm = 0x00).
- Example: alarm active:
- [0xFE][0xFE][0x24][0x0A][0x0A][0x35][0xFF][0xFF][0x2A]
SeeLeveL Error Response (SeeLeveL - > ELD)
- Value: 0x0F , 1 byte payload
- SeeLeveL issues this response if a command/message is not supported. Payload = unsupported message code.
- Example: ELD has previously issued a SeeLeveL Query Alarm Liquid Level Message (0x07) to a SeeLevel device which does not support alarms:
- [0xFE][0xFE][0x24][0x0A][0x0F ][0x07][0x40][0xFF][0xFF][0x2A]
SeeLeveL Liquid Level Report Message (SeeLeveL - > ELD)
- Value: 0x10, 6 or 7 byte payload, depending on whether alarms are supported
- SeeLeveL transmits liquid level (4 bytes = unsigned int32), number of digits to the right of the decimal (1 byte), optical error status (1 byte), and alarm status (currently active = 0x01, not in alarm state = 0x00). The alarm status field is optional and is not transmitted by a SeeLevel device which does not support alarms. Optical error status: no light = 0x00, low light level = 0x01, sunlight = 0x02, no error = 0x10. In the event that the optical error status is NOT in error, the liquid level/ number of digits to the right of the decimal is ignored.
- Example: liquid level = 1,083.1, no optical error, alarms not supported.
- For the liquid level, the first 4 bytes of the payload represent the hex value of the level, not the BCD value.
- [0xFE][0xFE][0x24][0x0F][0x10][0x00][0x00][0x2A][0x4F][0x01][0x10] [0xC9][0xFF][0xFF][0x2A]
Broadcast:
- Done after every data reception (good or bad) or no signal timeout for the 808P2 and 810PS2 gauges. Done after every 8 successful data transmissions for the 806B/806Bi gauges.
- Every 25 broadcasts (approximately 20 seconds) a handshake request is sent to allow continued broadcasts.
- On power up, if broadcasts are enabled, a handshake request is sent to allow broadcasts.
- Does NOT need a handshake to stop broadcasts.
- If a handshake request is not correctly responded to, the broadcasts are stopped.
- The Begin and Stop Broadcast requests are not explicitly responded to, the starting or stopping of the broadcast is the confirmation.
ELD requests needing handshake confirmation:
- Start broadcast and Send liquid level
- The handshake request is done every time one of these requests is received. The handshake must be responded to within 500ms, or else the response is considered invalid and an error message will be sent from the SeeLeveL to the ELD for late responses.
Handshake format:
- Request from ELD is received by SeeLeveL
- SeeLeveL responds with handshake request
- ELD sends handshake response
- SeeLeveL sends reply to original ELD request if handshake response is correct.
The alarms are not currently supported. In the future, if they are:
- The content of Message 0x08 is alarm set point, high or low level alarm, and current alarm status.
SeeLeveL Query of Handshake Frequency During Broadcast Message (ELD -> SeeLeveL)
- Value: 0x2D
- This asks for the handshake frequency, response is shown below.
- [0xFE][0xFE][0x24][0x09][0x2D][0x56][0xFF][0xFF][0x2A]
Handshake Frequency During Broadcast Response (SeeLeveL - > ELD)
- Value: 0x2E, 1 byte payload
- The frequency can range from 1 to 126 broadcasts per handshake request. The number is shown in hex from 0x02 to 0x7F (total number of transmissions per handshake, including the handshake).
- Message format, frequency is 20 (0x14):
- [0xFE][0xFE][0x24][0x0A][0x2E][0x14][0x6C][0xFF][0xFF][0x2A]
CHAPTER 6 – TROUBLESHOOTING GUIDE
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- TROUBLESHOOTING GUIDE](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-TROUBLESHOOTING-GUIDE.jpg)
Accuracy:
The analog output has an accuracy of ± 0.25% of the full scale value, so any
output value should be within 0.05 mA of the “ideal” value. There are no user
adjustments that can be made to alter the accuracy.
As with any digital system, there are round off and truncation errors inherent in the mathematical process. However, since the SEELEVEL Access™ utilizes a 10 bit digital to analog convertor, it has sufficient accuracy to allow the full resolution of the truck gauge to be realized. Note that the truck gauge sending the data has a resolution of only 8 bits (1/3″ systems).
CHAPTER 7 – SPECIFICATIONS
![GARNET T-DP0301-A SEELEVEL ACCESS Data Portal and Remote Display User Manual
- SPECIFICATIONS](https://manuals.plus/wp- content/uploads/2022/11/GARNET-T-DP0301-A-SEELEVEL-ACCESS-Data-Portal-and- Remote-Display-User-Manual-SPECIFICATIONS.jpg)
CHAPTER 8 – SERVICE & WARRANTY INFORMATION
The warranty will only apply if the warranty has been registered online from
the Garnet Instruments registration web page.
Go online to
garnetinstruments.com/support/ and
select “Register Warranty”.
DISCLAIMER OF WARRANTY ON HARDWARE
Garnet Instruments warrants equipment manufactured by Garnet to be free from
defects in material and workmanship under normal use and service for a period
of three years from the date of sale from Garnet or an Authorized Dealer. The
warranty period will start from the date of purchase or installation as
indicated on the warranty card. Under these warranties, Garnet shall be
responsible only for actual loss or damage suffered and then only to the
extent of Garnet’s invoiced price of the product. Garnet shall not be liable
in any case for labor charges for indirect, special, or consequential damages.
Garnet shall not be liable in any case for the removal and/or reinstallation
of defective Garnet equipment. These warranties shall not apply to any defects
or other damages to any Garnet equipment that has been altered or tampered
with by anyone other than Garnet factory representatives. In all cases, Garnet
will warrant only Garnet products which are being used for applications
acceptable to Garnet and within the technical specifications of the particular
product. In addition, Garnet will warrant only those products which have been
installed and maintained according to Garnet factory specifications.
LIMITATION ON WARRANTIES
These warranties are the only warranties, expressed or implied, upon which
products are sold by Garnet and Garnet makes no warranty of merchantability or
fitness for any particular purpose in respect to the products sold. Garnet
products or parts thereof assumed to be defective by the purchaser within the
stipulated warranty period should be returned to the seller, local
distributor, or directly to Garnet for evaluation and service. Whenever direct
factory evaluation, service or replacement is necessary, the customer must
first, by either letter or phone, obtain a Returned Material Authorization
(RMA) from Garnet Instruments directly. No material may be returned to Garnet
without an RMA number assigned to it or without proper factory authorization.
Any returns must be returned freight prepaid to: Garnet Instruments, 286 Kaska
Road, Sherwood Park, Alberta, T8A 4G7. Returned warranted items will be
repaired or replaced at the discretion of Garnet Instruments. Any Garnet items
under the Garnet Warranty Policy that are deemed irreparable by Garnet
Instruments will be replaced at no charge or a credit will be issued for that
item subject to the customer’s request.
If you do have a warranty claim or if the equipment needs to be serviced, contact the installation dealer. If you do need to contact Garnet, we can be reached as follows:
CANADA Garnet Instruments 286 Kaska Road Sherwood Park, AB T8A 4G7 CANADA email: info@garnetinstruments.com
UNITED STATES Garnet US Inc. 5360 Granbury Road Granbury, TX 76049 USA email: infous@garnetinstruments.com
References
- Garnet Instruments – Liquid management solutions, your way.
- Support – Garnet Instruments
- Garnet Instruments – Liquid management solutions, your way.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>