AQUAMASTER Combo Pen P110 Pro User Manual

June 3, 2024
AQUAMASTER

AQUA MASTER Combo Pen P110 Pro

AQUAMASTER-Combo-Pen-P110-Pro

Overview

AQUAMASTER-Combo-Pen-P110-Pro

  1.  Cap
  2.  On/off button
  3.  Celsius/Fahrenheit & Calibration button
  4.  Mode button
  5.  Display
  6.  Electrode
  7. Screw cap
  8.  Protective cover

Operation and use

Soak the meter for 12 hours in tap water before first use.

  1.  Remove the protective cover.
  2.  Gently clean the electrode with distilled water and dry it off.
  3.  Press the On/off button to start the meter.
  4.  Insert the meter upright into the solution up to the edge and stir gently. Wait a few seconds for the correct value to appear on the display.
  5.  Press the C/F button to switch between Celsius and Fahrenheit.
  6.  Press the MODE button to switch between pH and EC.
  7.  Rinse the electrode with water after use and press the On/off button to turn off the meter.

Our electrodes are designed for dry storage. Therefore, storage in liquid is not required for the electrode.

Calibrating the pH meter

Watch the How-to videos at Aqua Master Tools’ website.

  1.  Remove the protective cover.
  2.  Gently clean the electrode with distilled water and dry it off.
  3.  Press the On/off button to start the meter.
  4.  Insert the meter upright into the solution up to the edge and stir gently. Wait a few seconds for the correct value to appear on the display.
  5.  Press the C/F button to switch between Celsius and Fahrenheit.
  6.  Press the MODE button to switch between pH and EC.
  7.  Rinse the electrode with water after use and press the On/off button to turn off the meter.

Our electrodes are designed for dry storage. Therefore, storage in liquid is not required for the electrode.

Calibrating the pH meter
Watch the How-to videos at Aqua Master Tools’ website.

Calibrating the EC meter
Watch the How-to videos at Aqua Master Tools’ website.

  1.  Press the On/off button to start the meter.
  2.  Press the MODE button to switch to EC on the display.
  3.  Insert the meter in an upright position in a calibration solution (µS/cm 1413) EC 1.4 (make sure this is 25 ºC or 77 ºF) and stir gently.
  4. Wait until the value on the display is stable.
  5. Press and hold the C/F button for 5 seconds. Then release the button. The value EC 1.4 will now blink on the display. As soon as the meter switches back to the standard mode, calibration is complete.
  6.  Rinse the meter with clean water.
  7. The meter is now ready for use.

When is calibration required?

  •  If the electrode has been replaced.
  •  If the meter was last calibrated over a month ago.
  •  If the meter is used on a regular basis.
  •  If high precision is needed.

Replacing the electrode
Watch the How-to videos at Aqua Master Tools’ website.

  1.  Properly unscrew the sealing ring in the lower part of the meter.
  2.  Remove the defective/old electrode by gently pulling it.
  3.  Carefully place the new electrode with the recess in the correct position so that the connectors properly connect and do not damage the meter or electrode.
  4.  Check that the rubber seal is properly placed.
  5.  Tightly tighten the sealing ring.
  6.  Calibrate the meter as described in this manual.
  7.  The meter is now ready for use.

Specifications

AQUAMASTER-Combo-Pen-P110-Pro

Replacing the batteries
If the battery indicator at the top right of the display is activated, or if the display dims, please replace the batteries. To do so, unscrew the top and replace the 3 x 1.5 Volt AG-13 button cell batteries.

Warning or issues

  •  If your meter is not working properly, watch the troubleshooting videos on the product page of our website aquamastertools.com
  •  Always make sure that there are no air bubbles present in the solution during calibration. This could affect the measurements.
  •  If the EC value does not decrease to 0, insert the electrode into ethyl alcohol and shake gently for 5 to 10 seconds.
  •  Make sure the calibration solution is 25 ºC when calibrating the meter. Failure to do so may affect the result of the calibration.
  •  Never store the electrode in distilled water.
  •  DO NOT immerse the electrode in oil, proteins or suspended solids that leave a layer on the glass bulb.
  •  DO NOT soak in RO (Reverse Osmosis), distilled or de-ionized water. Pure water alters the chemistry of the reference liquid so that it no longer works.
  •  DO NOT place a cold electrode in a hot liquid (or vice versa). Sudden changes in temperature may break the glass and permanently damage the pen. Broken glass is not covered by the warranty.
  •  Make sure not to bump the glass electrode into anything, because this might break the glass bulb on the outside of the internal glass tube. Broken glass is not covered by the warranty.
  •  Avoid placing the meter near ballasts or any other electrical appliances. Force fields may affect the readings.
  •  If during calibration the pH/EC value slowly keeps rising or dropping and does not stabilize, please replace the batteries and calibrate the meter. Bad batteries or low power levels may influence the measurement results.

Warranty

This meter is warranted from all defects in material and manufacturing for a period of one year from the date of purchase. The electrode is warranted from all defects in material and manufacturing for a period of six months from the date of purchase. If during this period any parts need reparation or replacement and the damage is not due to incorrect operation by the user, please send the parts to the dealer or to us and the repair will be free of charge. Please refer to the exclusion warning.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>