Hillrom Golvo 9000 Mobile Patient Lifts Instruction Manual

September 25, 2024
Hillrom

Hillrom Golvo 9000 Mobile Patient Lifts

Specifications

  • Product Name: Golvo 9000 Mobile Lift
  • Models: Golvo 9000, Golvo 9000 LowBase
  • Product Numbers: 2000045 (Golvo 9000), 2000049 (Golvo 9000 LowBase)
  • Unique Design: Excellent aid for daily transfer of adults and children
  • LowBase Model: Extra low base for use in limited space situations

Product Usage Instructions

Fitting Liko Accessories
Individual fitting of Liko slings and other Liko lifting accessories is crucial for optimal performance and safety. Ensure proper fitting according to patient’s needs

Patient and Caregiver Roles
In this context, the person being lifted is referred to as the patient, and the assisting person is the caregiver. Always ensure the presence of at least one caregiver during patient transfer. Intended Uses
The Golvo 9000 Mobile Lift is designed for transferring patients between devices, floor lifting, horizontal lifting, supporting patient limbs, ambulating patient, bathing patient, toileting patient, weighing patient, and transferring patients from cars. FAQ (Frequently Asked Questions)
Q: Can the Golvo 9000 be used without a caregiver present?
A: No, a caregiver must always be present during patient transfers using the Golvo 9000.

Symbol Description

These symbols can be found in this document and/or on the product.

Safety Instructions

Intended use

  • Transferring patients (adult or children) between devices (e.g., within the room), floor lifting, horizontal lifting, supporting patient limbs, ambulating patient, bathing patient, toileting patient, weighing patient and transferring patients from car. Intended for use in following environments: health care, intensive care, emergency ward, rehabilitation, habilitation.
  • This product is not intended to be used by the patient alone. Lifting and transferring a patient shall always be performed with the assistance of at least one caregiver. This product is used as a means to perform the lift but is not in contact with the patient; therefore we do not discuss or describe various patient conditions in this instruction for use. Contact your Hill-Rom representative for support and advice.
  • Certain environments and conditions can limit the correct us of the mobile lifts, including:
  • Thresholds, unlevel floor surfaces, various obstacles, and extra-thick carpets. These environments and conditions can cause the wheels of the mobile lift not to roll as intended, possible imbalance in the mobile lift, and increased exertion by the caregiver. If you are uncertain that your care environment fulfills the requirements for correct use of the mobile lift, please contact your Hill-Rom representative for further advice and assistance.

Unbalanced lifting poses a tipping risk and may damage the lift equipment! Never leave a patient unattended during a lifting situation!

Before use, make sure that

  • the lift is assembled in accordance with the assembly instructions;
  • the lifting accessories are properly attached to the lift;
  • the battery has been charged for at least 6 hours;
  • you have read the instructions for use for the lift and lifting accessories;
  • personnel using the lift are informed of the correct operation and use of the lift.

Before lifting, always make sure that

  • the lifting accessory is selected appropriately, in terms of type, size, material and design, with regard to the patient’s needs;
  • the lifting accessories are not damaged;
  • the lifting accessories are correctly attached to the lift;
  • the lift strap is not twisted or worn and can move in and out of the lift;
  • the lifting accessories hangs vertically and can move freely;
  • the lifting accessory is correctly and safely applied to the patient in order to prevent injuries;
  • the sling bar latches are intact. Missing or damaged latches must always be replaced with new ones;
  • the sling’s strap loops are correctly connected to the sling bar hooks when the sling straps are extended but before the patient is lifted from the underlying surface.
  • Incorrect attachment of sling to slingbar may cause severe injury to the patient.
  • Prod. No. 2000045 and 2000049 have been tested by an accredited testing institute
  • No modification of this product is allowed.
  • Use of this equipment adjecent to other equipment should be avoided because it could result in improper operation, if such use is neccessary, observe and verify that other equipment are operating normally.
  • Particular care must be observed when using strong sources of potential disturbance, such as diathermy, etc, so  that diathermy cables are not positioned on or near the lift. If you have questions, please consult the responsible
  • assistive device technician or the supplier.
  • The lift may not be used in areas where flammable mixtures may occur, for example, in areas where flammable goods are stored.

This Caution notice is found on the Battery

  • CAUTION! NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL
  • DO NOT SHORT CIRCUIT
  • USE THE SPECIFIED CHARGER ONLY
  • MAY EXPLODE IF DISPOSED IN FIRE
  • This Caution notice is found on the Control box:
  • CAUTION! NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL

Definitions

  1. Lift Strap
  2. Sling bar with latches
  3. Retractable armrests
  4. Parking panel for sling bar
  5. Optional; Quick reference guide
  6. Optional: Holder for Quick Reference Guide
  7. Product decal
  8. Mast with inbuilt motor
  9. Front wheels
  10. Base
  11. Rear wheels with brakes
  12. Motor for base-width adjustment
  13. Control box with;
    • Emergency stop
    • Electrical Emergency Lowering
    • Electrical Emergency lifting
    • Battery charger indicators
    • Information display
  14. Battery
  15. Hand control
  16. Handles
  17. Emergency Lowering (mechanical)
  18. Lift arm
  19. Extra low base (Golvo LowBase)

Technical Data

  • Maximum load: 200 kg (440 lbs)

  • Please note: The base width may be adjusted

  • with weight up to 140 kg ( 308 lbs).

  • Material: Anodized aluminium

  • Weight: 9000: 38,8 kg kg (83,3 lbs) 9000 LowBase: 42,2 kg (93,0 lbs)
    Heaviest removable part: 9000: 23,4 kg (51,6 lbs)  9000

  • LowBase: 23,4 kg (51,6 lbs)

  • Wheels: Front: 75 mm* (3 in) twin wheels.

  • *LowBase: 46 mm (1.8 in) twin wheels.

  • Rear: 75 mm (3 in) twin wheels with brakes.

  • Turning diameter Golvo 9000: 1330 mm (52 in)

  • Golvo 9000 LowBase: 1330 mm (52 in)

  • Emergency lowering device: Mechanical and electrical

  • Lifting Speed (no load) 37 mm/s (1.5 in/s) or 27 mm/s ( 1.1 In/s)

  • Lifting interval: 1245 mm (49 in) (height-adjustable)

  • Sound level: 52.8 dB(A)

  • Protection class: IP X4

  • Operating forces, controls:

  • Hand control: 5 N

  • Electrical data: 24 V

  • Intermittent power:

  • Int. Op 5/95, active operation max 2 min. Only 5% of a given length of time may be active, but

  • no more than 2 min.

  • Battery options: Lead-acid gel, valve-regulated battery 24 V 2.9 Ah Prod. No. 2006106. 2,8 kg (6.2 Ibs.)

  • Li-ION battery 25.6 V, 3.3 Ah Prod. No. 2006110. 1,4 kg (3.1 Ibs.)

  • New batteries provided by supplier!

  • Battery charger: Internal charger, 100-240 V AC, 50-60 Hz,

  • max 400 mA.

  • Lift motor: 24 V, 7.5 A

  • Base motor 24 V, 5.5 A

  • Surrounding functional environment:

  • Temperature: +10°C to +40°C, (50° F to 104° F)

  • Humidity: 20% to 80% at 30°C non-condensing,

  • Atmospheric pressure: 700hPa to 1060hPa.

  • The device is intended for use indoors

  • Type B, in accordance with the electrical shock protection class.

  • Class II equipment.

Dimensions

Measurements

 | A|  |  |  |  | C|  | D|  |  |  |  |  |  |  |  |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---
Model|  |  |  | B| B1| B2| B3| B4|  |  |  |  | D2| E| F| F1| G| Lmax| Lmin| M| O
 | max.| min.|  |  |  |  |  |  | max.| min.| max.| min.|  |  |  |  |  |  |
Golvo 9000| 2090| 1455|  | 1185| 870| 480| 480| –| 1020| 735| 907| 623| 0| 1100| 105| 25| 940| 1816| 571| 436| 391
__

Golvo 9000

LowBase

| __

2090

| __

1455

|  | __

1185

| __

870

| __

480

| __

480

| __

600

| __

1028

| __

745

| __

907

| __

623

| __

0

| __

1100

| __

60 /

105

| __

22

| __

940

| __

1816

| __

571

| __

436

| __

391

Model| A|  | B| B1| B2| B3| B4| C| D| D2| E| F| F1| G| Lmax| Lmin| M| O
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---
max.| min.| max.| min.| max.| min.
Golvo 9000| 82.3| 57.3| 46.6| 34.2| 18.9| 18.9| | 40.2| 28.9| 35.7| 623| 0| 43.3| 4.1| 1.0| 37| 71.5| 22.5| 17.2| 15.4
Golvo 9000 Low-

Base

| __

82.3

| __

57.3

| __

46.6

| __

34.2

| __

18.9

| __

18.9

| __

23.6

| __

40.5

| __

29.3

| __

35.7

| __

623

| __

0

| __

43.3

| 2.4 /

4.1

| __

0.9

| __

37

| __

71.5

| __

22.5

| __

17.2

| __

15.4

The lifting interval 1245 mm (49 in) is height-adjustable, see ”Lift strap adjustment”, Operation chapter.
Note: When changing to other lifting accessories check that the lift still achieves desired lifting height.

  • Reference measurement according to Standard EN ISO 10535:2006.

EMC Table

Guidance and manufacturer’s declaration – electromagnetic emissions

The mobile lift is intended for use in the electromagnetic environment specified below. The customer or the user of the mobile lift should assure that it is used in such an environment.

”Essential performance according to the manufacturer: The hoist shall not move unintentionally while being submitted to disturbances.”

Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The mobile lift uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| **** The mobile lift is suitable for use in all establishments inclu- ding domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Complies
Voltage fluctuations/ flicker emis- sions

IEC 61000-3-3

| Complies
Guidance and manufacturer’s declaration – electromagnetic immunity

The mobile lift is intended for use in the electromagnetic environment specified below. The customer or the user of the mobile lift should assure that it is used in such an environment.

”Essential performance according to the manufacturer: The hoist shall not move unintentionally while being submitted to disturbances.”

Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2| +/- 8 kV contact

+/- 15 kV air

| +/- 8 kV contact

+/- 15 kV air

| +/- 8 kV contact

+/- 15 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

Electrical fast transient / Burst IEC 61000-4-4| +/- 2 kV for power supply lines

+/- 1 kV for input/output lines

| +/- 2 kV for power supply lines

n/a. for input/output lines

| Mains power quality should be that of a typical commercial or hospital environment.
Surge

IEC 61000-4-5

| +/- 1 kV differential mode

+/- 2 kV common mode

| +/- 1 kV differential mode n/a. for common mode| Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short inter- ruptions and voltage variations on power supply input lines

IEC 61000-4-11

| 0 % UT for 0,5 cycle, at 0, 45,

90, 135, 180, 225, 270 and

315 degrees

0 % UT for 1 cycle, at 0 degrees

70 % UT for 25 cycles at 50 Hz and 30 cycles at 60 Hz, at 0 degrees

0 % UT for 250 cycles at 50 Hz and 300 cycles at 60 Hz.

| 0 % UT for 0,5 cycle, at 0,

45, 90, 135, 180, 225, 270

and 315 degrees

0 % UT for 1 cycle, at 0 degrees

70 % UT for 25 cycles at 50 Hz and 30 cycles at 60 Hz, at 0 degrees

0 % UT for 250 cycles at 50 Hz and 300 cycles at 60 Hz.

| Mains power quality should be that of a typical commercial or hospital environment. If the user of the [Equipment or System] requires continued operation during power mains interruptions, it is recom- mended that the [Equipment or Sys- tem] be powered from an uninter- ruptible power supply or battery.
Power frequency (50/60 Hz) magnetic field

IEC 61000-4-8

| 30 A/m| Complies| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity

The mobile lift is intended for use in the electromagnetic environment specified below. The customer or the user of the mobile lift should assure that it is used in such an environment.

”Essential performance according to the manufacturer: The hoist shall not move unintentionally while being submitted to disturbances.”

Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6

Radiated RF IEC 61000-4

| 6 Vmrs

150 kHz to 80 MHz

10 V/m

80MHz to 2,7GHz

| 6 Vmrs


10 V/m

| Portable and mobile RF communications equipment should be used no closer to any part of the mobile lift, including cables, than the recommended separation distance cal- culated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

  • d = 1,2 P
  • d = 1,2 P 80 MHz to 800 MHz
  • d = 2,3 P 800 MHz to 2,7 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b

Interference may occur in the vicinity of equipment marked with the following symbol.

NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec- ted from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. An electromagnetic site survey should be considered to assess the electromagnetic environment due to fixed RF transmitters. If the measured field strength in the location in which the Golvo™ 9000 lift is used exceeds the applicable RF compliance level above, the mobile lift should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the mobile lift.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the mobile lift

The mobile lift is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the mobile lift can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the mobile lift as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)| Separation distance according to frequency of transmitter

(m)

150 kHz to 80 MHz

d = 1,2 P

| 80 MHz to 800 MHz

d = 1,2 P

| 800 MHz to 2.7 GHz

d = 2,3 P

0.01| 0,12| 0,12| 0,24
0.1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec- tion from structures, objects and people.

Assembly

Before assembly, make sure you have the following parts and tools:

  •  if mast with control box and hand control, sling bar with latches, 2 M6 screws
  • Armrest
  • Base, with width adjustment motor
  • Battery
  • Tools: 4, 5 mm hex keys
  • Bag containing: charger cable, charger connector cable
  • Instructions for use

Lock the wheels on the base before starting to assemble the lift.

  1. Position the lift mast between the two black plastic plugs on the base cross member.

    • Then push the mast forwards as in the figure above so that
    • it hooks onto the cross member.
  2. Screw the two provided M6 screws into the upper holes on the lift mast.
    No screws in the lower holes!

  3. Place the armrest part in the attachment on the lift mast, starting with the lower groove.
    Lower and load the armrest until it hooks onto the upper groove on the armrest part. Do not remove any of the pre-assembled M8 screws completely, but it may be necessary to loosen them.

  4. Secure the armrest by tightening the twopre-assembled M8 screws.

  5. Connect cables to the control box, see illustration. Make sure plugs are fully seated.

  6. Connect the battery and secure it to the control box bracket. A click sound can be heard when the battery is installed correctly.

  7. Hang the hand control on the handle.

  8. Optional accessories: Quick reference guide to be placed in Holder for Quick reference guide on the mast.

  9. Reset the emergency stop by turning the button clockwise.

  10. A) Connect the extension cable for the charging cable to the control box.

    • Insert the extension cable in the tension clip underneath the control box.
    • Connect the charging cable to the extension cable.
  11. Place the charger cable on the hook provided on the mast after completed charging.

IMPORTANT! Always charge the battery before using the lift the first time. Charge the battery until charging is completed, see “Charging the Battery” for more information and instructions

After assembly and charging, ensure that:

  • The motion of the lift arm corresponds to the buttons on the hand control/operation panel.
  • The emergency lowering functions are working (mechanical and electrical).
  • The wheel brakes are working.
  • The base-width adjustment works.
  • The battery is fully charged.
  • service interval is activated! Push following buttons on the handcontrol simultaneously: Up / Down , until a audio signal (single beep) is heard = service interval activated.
  • (Alternatively use the push buttons simultaneously for emergency lifting up and down on the control box)

Disassembly

  1. Begin by removing the sling bar or other accessory that is fitted to the lift.

  2. Remove the armrest holder as described below:

    • Do not remove, just loosen the two M8 screws in the armrest holder on either side of the lift mast.
    • Remove the armrest holder using two screwdrivers. Place screwdrivers inside on top of handlebar and pull up simultaneously as illustrated.
  3. Loosen the cables from the control box, see assembly, Remove the mast as described below:
    When the mast has been detached from the base, it must be supported to prevent it from falling.

    • Unscrew both the safety screws in the upper holes on themast.
    • Then screw the safety screws into the lower holes on the lift mast. This releases the mast from the base, and it can now be removed.

Operation

Handcontrol operation and indicators
Operate the lift using the push buttons on the handcontrol. For raising and lowering: Directional arrows show the direction of movement (up/down) The lifting and base movement stops as soon as the push button is released.
Indicator: 1 – 4

  1. Overload (Kg/Ibs) light “flashes yellow”, too much load is applied to the lift.

  2. Green light, battery power (100 – 50 %), Ok! – will constantly light up green when charger is connected to mains.

  3. Yellow light, battery power (50 – 25 %), battery needs charging

  4.  Yellow light, battery power (Less than 25 %), battery needs charging.
    A buzzer will sound when pressing a push button.
    Note! If the buzzer sound starts during an ongoing lift complete the lift and charge the lift afterwards!

  5.  Light “flashes yellow” and a buzzer will sound when pressing a push button. Charge the lift immediately! The remaining battery power can only make the lift arm go down.
    Note! Please see chapter “Charging the Battery” for more information.

Control Box operation and information

  1. Emergency Stop button

    • Activate: Push the red button
    • Reset: Turn the red button clockwise.
  2. UP (Arrow), Electrical emergency lifting.

  3. DOWN (Arrow), Electrical emergency lowering.
    Operation of the push buttons 2 and 3 are done by pressing with a narrow object into the circle mark above each (Arrow).
    The Actuator movement stops as soon as the push button is released.

  4. “ON” – lights up green when the charger is connected to mains.*

  5. “CHARGE” – lights up yellow constantly during charging and will turn off when charging is completed.
    Display Pop-up information:

    • Battery power (100 – 50 %) Ok!
    • Battery power (50 – 25 %) Battery needs charging.
    • Battery power (Less than 25 %) battery needs charging.
    • A buzzer will sound when pressing a push button.
    • Note! If the buzzer sound starts during an ongoing lift complete the lift  and charge the lift afterwards!
    • Charge the lift immediately! A buzzer will sound when pressing a  push button. The remaining battery power can only make the lift arm go down.
    • The lift is connected to the mains.
    • Short circuit warning!
    • check cables and connections. Warning is shown until repaired!
    • Overload!
    • Too much load is applied to the lift.
    • Service needed; contact Hill-Rom.
  6. Information display:

    • Activate information display by a “quick press” at a UP, push button.

Li-ION battery – specific information
Sleep mode! the sleep mode will be activated in a Li-ION battery if not in use or charged in one week or more. The sleep mode switches off the battery and it´s electronics to save power. The battery will stay in sleep mode until the battery is set back to operation mode again.
How to set the Li-ION battery back in operation mode; Charge the battery, when the “CHARGE” indicator, 5 is lit the battery has been set back in operation mode and ready for use. Note! We recommend to charge the battery until charging is completed, see “Charging the Battery” for more information and instructions.
Delay! a delay to indicators for current battery power at the control box and hand control occurs if the emergency stop function is activated and restored, see 1 above.

  • To activate the emergency stop: Push the red Emergency Stop button on the control box.
  • To reset the emergency stop: Turn the button clockwise.
  • Electrical emergency lowering / lifting Use a narrow object to press into the circle mark above each (Arrow),
  • See chapter “Operation” for more information.
  • Do not use sharp objects, since this may cause damage on the control box!
  • Mechanical Emergency Lowering Emergency lower by moving the handle up and down. Repeat the motion until the patient being lifted is on a firm surface; draw down the sling bar manually and continue to pump the handle until the sling bar is low enough to enable the sling’s strap loops to be unhooked.

Locking the Wheels
The rear wheels can be locked to prevent rotating and turning. The locking/unlocking of the wheels is done with the foot.
Locked wheels during lifting can increase the risk of tipping.
NOTE: When lifting, the wheels should be unlocked so that the lift can be moved to the patient’s center of gravity. The wheels should be locked, however, if there is a risk of the lift rolling into the patient, for instance, when lifting from the floor.

Lift strap adjustment

After mechanical emergency lowering

Restoring the lifting level
If the lift strap has been lengthened due to using the emergency lowing function, the lift interval height will be lower than previously. To restore maximum lifting height, the lift strap must be reset to its original length.

Do as follows

  1. Remove any load / weight from the strap above the emergency lowering device. Do this by either placing the sling bar above the lift arm or by another person holding the sling bar up so that the strap hangs loose.
  2. Lower and raise the handle with your left hand. At the same time tension the strap by turning the knob (a) clockwise with your right hand. Repeat this procedure until the red mark on the strap is just above the emergency lowering device.

Adjustment of the lift interval level
If it is necessary to reach a lower level with the sling bar,  this can be arranged by lengthening the lift strap using the mechanical emergency lowering device. Do not lengthen the strap more than necessary as the highest obtainable lifting level is also affected.
An example of when it would be advantageous to lower the lift interval is when lifting from the floor using a sling whose strap loops do not reach up to the sling bar hooks when the lift is in its lowest position.

Do as follows
Push down the red emergency lowering handle at the same time as loading the sling bar (pull down the sling bar with your other hand). This extends the strap and the sling bar is lowered. Repeat until the required strap length is obtained.

Armrest
To use the armrest you need to rotate it from the (vertical) rest position up to the (horizontal) support position. The armrest have two purposes: to help the patient feel more secure and facilitate for the caregiver when moving the lift.
When using the lift to transfer a patient between rooms, the armrest should be set in the support position!

Parking the sling bar
When the lift is not in use or is being moved without  a load, it can be beneficial to place the sling bar in the parking panel.
The parking panel is intended for parking Universal SlingBars 350, 450 and 600 (all models).
When the sling bar is parked in the parking panel, the lift should not be raised since this could be dangerous and cause personnel injury or damage to the lift if the sling bar should release and swing out from the panel.

Installation of Latches
After installation, ensure that the spring loaded clip is taut against the sling bar and moves freely in the sling bar hook.

Lift correctly!

  • Before each lift, make sure that:
  • the Sling loops at opposite sides of the Sling are at the same height
  • all the Sling loops are fastened securely in to the Slingbar hooks
  • the Slingbar is level during the lift, see Figure 1.
  • If Slingbar is not level (see Figure 2) or if the sling loops is wrongly attached to the Slingbar (see Figure 3) lower the user to a firm surface and adjust according to the Instruction for use of Sling in use.
  • An improper lift can be uncomfortable for the user and cause damage to the lift equipment! (see Figure 2 and figure 3).

Position of the Lift when Lifting

Transfer from car
Ensure that the lift used for car transfers is always stored and charged indoors. Outdoor operation should be kept to a minimum. Using the lift in harsh conditions such as rain, snow or extreme cold may impact the performance of the lift immediately.

Prerequisites
Transfer from car is applicable from the front and rear with the patient in a seated position. The caregiver must assess that transfer from the car is achievable. Considerations should be made regarding the status of the patient (status/condition, height and weight, position and accessibility), the relative size of the car and position of the lift equipment. The ground surface of the transfer area should be level, hard and smooth and free from gravel, debris, ice and potholes. Two caregivers are recommended for this task. A gurney or a wheel chair should be in the immediate vicinity of the car when performing the lift. Recommended slings for car transfer are the Universal Sling (models 000 & 002) and the High Back Sling (models 200/210/25/26). For correct sling application consult the sling instruction for use.

Do as follows

  1. Apply the sling according to the sling instruction for use, a HandySheet or Tube can be used to reduce friction. One caregiver may assist from inside the car if necessary.
  2.  Position the lift as perpendicular as possible to the car, with the car door open. Keep the lift arm outside the car and the wheels of the lift unlocked.
  3. Attach sling loops to the sling bar. Ensure the correct application of sling loops to the sling bar. Raise the lift to apply more tension on the sling. Rotate the patient toward the door opening and guide the legs of the patient out of the car. Apply friction reducing devices if necessary.
  4. One caregiver must guide the sling bar and patient out of the car and ensure that the head of the patient is guided safely while exiting the car. The other caregiver will need to raise the lift while simultaneously pulling the lift backwards. Notice the correct grip on the sling bar to avoid pinch injuries to the caregiver as the sling bar is guided out of the car. Avoid placing hands between the sling bar and the door frame of the car.
  5. Removal of the patient from the car is complete, contin-ue the transfer to a wheel chair or gurney.

Charging the Battery

Charger information

  1. ON” – lights up green when the charger is connected to mains.
  2. CHARGE” – lights up yellow constantly during charging and will turn off when charging is completed.
  • NOTE! Charging a deep discharged Li-ION battery
  • When charging a deep discharged Li-ION battery the charger will start charging at a low charging rate to protect the battery.   normal charging state. The orange light of the control box will not be turned on as the operation is analogue. It is therefore not possible to see that the charging has started, however, only at a low level.
  • When the low rate charging is completed the charger will automatically switch to normal charge rate and the “CHARGE” indicator will light up orange and will turn off when charging is completed.
  • Charging with the control box internal charger (standard)
  • Plug the charger cable into mains (100-240 VAC), see charger information  1 – 2 above. The battery is fully charged after about 6 hours and the  charger disconnects automatically, the yellow “CHARGE” diode turn of.
  • For maximum battery life, batteries must be charged regularly.
  • We recommend charging after each use or every night.
  • Never charge batteries in a wet area!
  • NOTE! If the charger cable is stretched out it should be replaced to avoid the risk of the cable getting caught and tear. NOTE! The lift cannot be used when the charger cable is plugged into a wall socket.
  • NOTE! If the yellow “CHARGE” diode at the control box continuous to be lit after 8 hours, discontinue charging and replace the battery with a new one.
  • NOTE! A damaged battery shall be replaced and contact with leaking fluids shall be avoided.

Alternative charging procedures

  • Wall mounted charger accessory or table charger housing accessory:
  • Loosen the holder for the charger cable. Remove the battery pack from the control box by loosening the locking device on top of the battery pack. See chapter “Assembly”.

Charger information

  • “ON” – lights up green when the charger is connected to mains.
  • “CHARGE” – lights up yellow constantly during charging and will turn off when charging is completed.
  • Alt. A. Place the battery pack on the wall mounted charger. Plug the charger cable into mains (100-240 VAC)  check that both “ON” and “CHARGE” on the charger lights up.
  • Alt. B. Place the battery pack on the charger in the table charger housing. Plug the charger cable into mains (100-240 VAC) check that both “ON” and “CHARGE” on the charger lights up.

Maximum Load

  • Different maximum loads may apply to different products on the assembled lift unit, sling bar, sling and any other accessories used. For the assembled lift unit, including accessories, the maximum load is always the lowest maximum load rating for any of the components. A Golvo™ mobile lift that is approved for 200 kg  (440 lbs) can be equipped with a lifting accessory that is approved for 300 kg (660 lbs). In this case, the maximum load of 200 kg (440 lbs.) applies to the assembled lift unit. Study the markings on the lift and lifting accessories or contact your Hill-Rom representative if you have any questions.

Recommended Lifting Accessories

  • Using lifting accessories other than those approved can entail a risk.
  • Generally recommended sling bars and accessories for Golvo mobile lifts are described below.
  • When changing a sling bar or other lifting accessories, the highest possible lifting height of the lift is affected. Before changing lifting accessories you should always ensure that the lift, after change, can fulfil the desired lifting height in order to manage the lifting situations for which the lift is to be used.
  • For additional guidance in selecting a sling, study the Instruction for use for the respective sling models. There you will also find guidance for combining Liko™ sling bars with Liko slings.
  • Contact your Hill-Rom representative for advice and information on Liko’s product range. * this product is also available in a version with Quick-Release Hook.

Quick-Release Hook
Liko’s Quick-Release Hooks are a system for quick change of lifting accessories on Liko’s mobile and stationary lifts. The GolvoTM mobile lift must be equipped with Q-link in order to be used with the Quick-release Hook.
The Quick-Release Hook Universal fits the Universal Bars 350, 450 and 600 (prod. no. 3156074 – 3156076).
Quick-Release Hook TDM fits Sling Cross-bar 450 and 670 (prod. no. 3156021 and 3156018) and Universal TwinBar 670 (prod. no. 3156077).
Contact your Hill-Rom representative for more information.

Stretchers

  • Golvo mobile lift can be used for horizontal lifting in combination with
  • Liko OctoStretch Prod. No. 3156056
  • LikoStretch Mod 600 IC Prod. No. 3156065B
  • FlexoStretch Prod. No. 3156057
  • Contact your Hill-Rom representative for more information.

Optional Components for Use with Lift

LikoScale™ device

  • for weighing a patient in combination with Golvo™ mobile lift.

  • LikoScale™ 350, Max 400 kg (880 lbs.) Prod. No. 3156228

  • LikoScale™ device only for use in France:

  • LikoScale™ 350, Max 400 kg (880 lbs.) Prod. No. 3156228FR

  • LikoScale™ 350 is certified according to the European Directive

  • NAWI 2014/31/EU (Non-Automatic Weighing Instruments).

  • LikoScale™ devices only for use in the United states and Canada:

  • LikoScale™ 200, Max. 200 kg (440 Ibs.) Prod. No. 3156225

  • LikoScale™ 400, Max. 400 kg (880 Ibs.) Prod. No. 3156226

  • Contact your Hill-Rom representative for more information.

Recycling instructions

Old batteries are to be deposited at the nearest recycling station or given to personnel authorized by Hill-Rom.
Hillrom evaluates and provides guidance to its users on the safe handling and disposal of its devices to aid in the prevention of injury, including, but not limited to: cuts, punctures of the skin, abrasions, and any required cleaning and disinfection of the medical device after use and prior to its disposal. Customers should adhere to all federal, state, regional, and/or local laws and regulations as it pertains to the safe disposal of medical devices and accessories.
If in doubt, the user of the device shall first contact Hillrom Technical Support for guidance on safe disposal protocols.

Simple Troubleshooting

Cleaning and Disinfection

Safety recommendations
Cleaning and disinfecting procedures for Liko™ Mobile lifts. This instructions do not replace the facility’s own cleaning and disinfection policies.

  • Wear protective equipment according to manufacturer’s instruction and per facility protocol throughout the cleaning operations, such as: rubber gloves, goggles, apron, face mask and shoe covers.
  • Unplug mains (AC power source) before cleaning and disinfection.
  • Never clean the lift by pouring water on it, steam cleaning it, or by using a high-pressure jet.
  • Refer to the recommendations made by the cleaning and disinfecting product manufacturer.

Equipment

  • Protective equipment (such as: rubber gloves, goggles, apron, face mask and shoe covers) as recommended by the facility protocol and manufacturers instructions
  • Clean buckets
  • Cloths for washing and drying
  • Soft brush
  • Warm water

To find Cleaning / Disinfectants compatible or not compatible for use on Liko’s products, follow the Application of commonly used Cleaning / Disinfectants on Liko products” in this document.

Cleaning instructions

  1.  Unplug mains (AC power source) before cleaning and disinfection.
  2. Clean the lift with a cloth moistened with warm water and a neutral cleaning agent approved by your organization. A soft brush may be used to remove stains and resistant dirt.
  3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all areas run the lift into the highest and lowest positions and extend the base width adjustment entirely out. Remove the Battery to have access behind the battery. Lower the sling bar to get access to the whole lift strap. Fully extend the lift strap by using the emergency lowering. After cleaning the lift strap, and before you raise the sling bar, make sure the lift strap has dried. After the emergency lowering device is used, the lifting level needs to be restored; see ”Operation” in this document.
  4. Pay special attention to the following areas:
    • Lift strap
    • Sling bar (different designs)
    • Armrest
    • Mechanical emergency lowering
    • Hand control
    • Handles
    • Control box
    • Emergency stop
    • Battery
    • Wheels

Disinfection Instructions

  1. For the use of suitable disinfectants see ”Application of commonly used Cleaning / Disinfectants on Liko products” in this document.
  2. Use the choice of disinfectant according to the manufacturer’s instructions and repeat the work step as in ”Cleaning instructions”.
  3.  Remove traces of disinfectant after disinfection. Wipe off the lift with a cloth moistened with clean water starting from the top and working down. The cloth shall not be so damp that it drips.
  • The lift may not be cleaned with CSI or equivalent.
  • The hand control may not be cleaned with Viraguard or equivalent.
  • The control box may not be cleaned with Anioxy Spray or equivalent.
  • The lift strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.

Application of commonly used Cleaning / Disinfectants on Liko products


Chemical class

| ****

Active ingredient

| ****

pH

| *Cleaners / Disinfectant )| **

Manufacturer *)

| May not be used on the following items:
---|---|---|---|---|---
Quaternary ammonium chloride| Didecyl dimethyl ammonium chloride = 8.704% Alkyl dimethyl benzyl ammonium chloride = 8.19%| 9.0 – 10.0 in use| Virex II (256)| Johnson/Diversey| Foot rest for Sabina™ and Roll-On™
Quaternary ammonium chloride| Alkyl dimethyl benzyl ammonium chloride = 13.238% Alkyl dimethyl ethylbenzyl ammonium chloride = 13.238%| 9.5 in use| HB Quat 25L| 3M|
Accelerated Hydrogen Peroxide| Hydrogen Peroxide 0.1 -1.5%

BenzylAlcohol: 1-5%

Hydrogen Peroxide 0.1 -1.5%

BenzylAlcohol: 1-5%

| 3| Oxivir Tb| Johnson/Diversey| The lift straps for Golvo™ and ceiling lifts
Phenolic| Ortho-Phenylphenol = 3.40%

Ortho-Benzyl-para-Chlorophenol = 3.03

| 3.1 +/- 0.4 in use| Wexcide| Wexford Labs|
Bleach| Sodium hypochlorite| 12.2| Dispatch| Caltech| The lift straps for Golvo™ and ceiling lifts
Alcohol| Isopropyl alcohol = 70%| 5.0 – 7.0| Viraguard| Veridien| Hand controls for all lifts
Quaternary ammonium| n-Alkyl dimethyl benzyl ammonium chlorides = 0.105%

n-Alkyl dimethyl ethylbenzyl ammonium chlorides = 0.105%

| 11.5 – 12.5| CSI| Central Solutions Inc.| Viking™, Liko M220™, Liko M230™, Uno™, Sabina™, Golvo™,

LikoLight™, Roll-On™, Likorall™, Multirall™

Benzyl-C12-18-alkyldimethylammonium, chlorides| Benzyl-C12-18-alkyldimethylammonium, chlorides (22 %) 2-Phenoxyethanol (20 %)

Tridecylpolyethylenglycolether (15 %)

Propan-2-ol (8 %)

| approx 8.6 in use| Terralin Protect| Shülke| Foot rest for Sabina™ and Roll-On™
Organic peroxide (type E, solid)| Magnesium monoperoxyphtalate hexahydrate (50-100%) Anionic surfactant (5-10%)

Nonionic surfactant (1-5%)

| 5.3 in use| Dismozon Pur| Bode| The lift straps for Golvo™ and ceiling lifts
Ethanol| Hydrogen peroxyde (2.5-10 %) Lauryldimethylamine oxid (0-2.5 %) Ethanol (2.5-10 %)| 7| Anioxy-Spray WS| Anios| Control box for all mobile lifts
Troclosene sodium| Adipic acid 10-30% Amorphous silica < 1%

Sodium Toluene sulphonate 5-10 % Troclosene sodium 10-30 %

| 4-6 in use| Chlor-Clean| Guest Medical Ltd| The lift straps for Golvo™ and ceiling lifts

Inspection and Maintenance

For trouble-free use, certain details should be checked each day the lift is used:

  • Inspect the lift and check to make sure that there is no external damage.
  • Check the sling bar attachment.
  • Check the lift strap for wear and to ensure the strap is not twisted.
  • Check the functionality of the latches.
  • Check the integrity of the lifting motion and the base-width adjustment.
  • Check that the lifting interval level is set correctly and that the emergency lowering is working as it should (both the elec-trical and the mechanical).
  • Charge the battery each day the lift is used and make sure the charger works.

When necessary, clean the lift with a moist cloth and check that the wheels are free from dirt. Find more detailed information regarding cleaning and disinfection of your Liko™ product in chapter; Cleaning and Disinfection.
The lift should not be exposed to running water.

Service
A periodic inspection of the lift should be carried out at least once per year.
Periodic inspection, repair and maintenance may be performed only in accordance with the Liko service manual by personnel authorized by Hill-Rom and using original Liko spare parts.\

Service activities are not allowed with the patient in the lift.

Service Agreement
Hill-Rom offers the opportunity to enter into service contracts for the maintenance and regular inspection of your  Liko product.

Expected Life Time
The product has an expected service life of 10 years when correctly handled, serviced and periodically inspected in accordance with Liko’s instructions.
Parts listed below are subject to wear and tear and have specific expected life time:

  • Handcontrol, expected life time 2 years,
  • Battery, expected life time 3 years.

Transport and Storage

  • During transportation, or when the lift is not to be used for a long time, the emergency stop should be activated.
  • The environment where the lift is transported and stored should have a temperature of -10°C to +50°C (14°F to 122°F),
  • 20-90% humid., pressure 700-1060 hPa.
  • The environment where batteries is transported and stored should have a temperature of -10°C to +40°C (14°F to 104°F) , 20-80% humid., pressure 700-1060 hPa.

Product Changes

  • Liko products undergo continuous development, which is why we reserve the right to make product changes without prior notice. Contact your Hill-Rom representative for advice and information about product upgrades.

Design and Quality by Liko in Sweden

  • The management system for both manufacturing and development of the product is certified in accordance with ISO9001 and its equivalent for the medical device industry, ISO13485. The management system is also certified in accordance with the environmental standard ISO14001.

Notice to Users and/or Patients in EU

  • Any serious incident that has occurred in relation to the device, should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.

www.hillrom.com

Liko AB
Nedre vägen 100
975 92 Luleå, Sweden
+46 (0)920 474700
Liko AB is a subsidiary of Hill-Rom Holdings Inc.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>