HORIBA 300-NH4-C Gel Filled Ammonium Ion Sensor Cartridge Instruction Manual
- September 29, 2024
- HORIBA
Table of Contents
300-NH4-C Gel Filled Ammonium Ion Sensor Cartridge
“`html
Specifications:
- Model: 300-NH4-C
- Type: Gel-filled Ammonium Ion Sensor Cartridge
- Dimensions: 16mm x 110mm
- Weight: 15mm
- Ammonium Detection Range: 100 mg/L – 1000 mg/L
- pH Range: pH 2 – pH 7
- Compliance: RoHS, EN IEC 63000, BS EN IEC 63000
Product Usage Instructions:
1. Safety Precautions:
Ensure to follow all safety precautions mentioned in the manual
before using the product.
2. Firmware Check:
Check and update the firmware of the ion sensor head or smart
handheld meter to the latest version before using an ion sensor
cartridge.
3. Tips for Accurate Measurement:
-
Add ISA (Internal Standard Addition) to both standard solutions
and sample solutions in a volume ratio of standard solution and ISA
as 50:1. -
Recalibrate the sensor cartridge every few hours.
-
Prepare standard solutions with a concentration relationship of
low concentration standard solution < sample solution < high
concentration standard solution for calibration. -
Ensure to match the temperature of the standard solution during
calibration with the temperature of the sample solution during
measurement.
FAQ:
Q: How can I obtain the latest firmware for the ion sensor head
or smart handheld meter?
A: You can visit the following website to download the latest
firmware and find instructions on how to update it:
Download Center
Q: How should I handle interfering ions when using the
product?
A: Confirm the effects of interfering ions and take measures to
avoid them as their influence varies depending on the sample
solution’s composition. Refer to the website mentioned in the
manual for details on main interfering ions.
“`
CODE:GZ0000714260A © 2023 -2024 HORIBA Advanced Techno, Co., Ltd.
May,2024
300-NH4-C
URL http://www.horiba.com/patent
·
·
pH pH pH 100 mg/L 1
300-I WQ-300
1.0.0
1.0.01.0.11.1.0
https://www.horiba.com/jpn/water-quality/updateinformation/
501 1/10 Web
O
601-8551 2 http://www.horiba-adt.jp
0120-37-6045 9:00 12:0013:00 17:00
0.18 18,000 mg/L NH4+1×10-5 1 mol/L NH4+
0 50°C
1
±5%
pH 2
pH5 8100 mg/L NH4+25°C 50 62 mV10 100 mg/L NH4+25°C
ABS
O
FPM
16 mm x 110 mm
15 mm
110 mg/L 100 mg/L
2 pH
NH4+ 100 mg/L 1000 mg/L GHS
1 1 1 1 1 1 1
1. 2.
O
1 400
Web https://www.horiba.com/water_quality_300_ion_c_info/
2
CODE:GZ0000714260A © 2023 -2024 HORIBA Advanced Techno, Co., Ltd.
May,2024
Gel-filled Ammonium Ion Sensor Cartridge 300-NH4-C Instruction Manual
Patent marking
This product is protected under one or more of the patents found at the
following address: http://www.horiba.com/patent
Conformable standards
This equipment conforms to the following standards:
RoHS: EN IEC 63000 9. Monitoring and control instruments including industrial
monitoring and control instruments
RoHS: BS EN IEC 63000 9. Monitoring and control instruments including
industrial monitoring and control instruments
Authorised representative in EU and UK
HORIBA Europe GmbH Hans-Mess-Str.6, D-61440 Oberursel, Germany
HORIBA UK Limited Kyoto Close, Moulton Park, Northampton, NN3 6FL, UK
China regulation
()
(Pb) (Hg) (Cd) (Cr (VI)) (PBB)
(PBDE)
SJ/T 11364 : GB/T
26572 :
GB/T 26572
Caution on use
Safety Precautions
CAUTION
Precautions for chemicals
Handle the solutions used for calibration and measurements with care. Always
wear protective gloves and eye protection when handling solutions. · Harmful
in contact with skin and eyes. Rinse immediately with
tap water. · If accidentally swallowed, consult a doctor immediately.
Points of concern Do not give an excessive shock to the sensor cartridge.
Cracked or chapped sensor cartridge can not be used. Replace such sensor
cartridge with a new one. Immerse the sensor up to the sensor body under the
nut. Use with the plug facing up and the membrane facing down as shown in the
figure. The pH range is generally smaller when the concentration of the ions
to be measured is lower. If the pH of the sample solution has exceeded the
measurable pH range, accurate measurement may not be possible. Do not allow
the liquid junction to dry out. If liquid junction is dried, measurement value
may drift. Do not use (store) the sensor cartridge in any place where the
operation temperature (storage temperature) is out of the specified range. Do
not strongly place ion selective membrane against hard surface. Do not poke
ion selective membrane with sharp object. Make sure the membrane has no dirt
or air bubbles on it. Before the first use or if the sensor cartridge is less
responsive, perform conditioning of the sensor cartridge before calibration.
Soak a sensor cartridge in 100 mg/L or more standard solution for 1 hour.
Always use new standard solution for calibration. Do not return a standard
solution to its original container after it has been used once.
Check the firmware
In an ion sensor head (300-I) and a smart handheld meter (WQ-300) with the
following firmware versions, there is no option to select an ion sensor
cartridge. Then, it is necessary to update the firmware of the ion sensor head
or the smart handheld to the latest version. Make sure to check before using
an ion sensor cartridge.
Ion sensor head:
1.0.0
Smart handheld meter: 1.0.0, 1.0.1, 1.1.0
Refer to the instruction manual for smart handheld meter on how to check the firmware. It is possible that these firmwares will be updated from time to time as new functions and improvements are made. Check the following website for how to obtain the latest firmware and how to update the firmware. https://www.horiba.com/sgp/water-quality/support/ download-center/
Tips for accurate measurement Add the ISA to both standard solutions and
sample
solution. The volume ratio shall be the standard solution
and the ISA = 50:1. Recalibrate sensor cartridge every few hours. Prepare
standard solutions which have the following
concentration relationship for calibration.
Standard solution with low concentration < Sample
solution < Standard solution with high concentration Match the temperature of
the standard solution in
calibration and the temperature of the sample solution in
measurement.
2 Miyanohigashi-cho, Kisshoin, Minami-ku, Kyoto, 601-8551, Japan http://www .horiba-adt.jp
For any questions regarding this product, please contact your local agency, or inquire from the following website. http://global.horiba.com/contact_e/index.htm
Confirm the effects of interfering ions and take measure to avoid them because
the influence of interfering ions varies depending on the composition of the
sample solution. For the main interfering ions, refer to the website written
in “Reference”.
Specifications and parts description
Parts name
Rubber cap
Plug
O-ring Nut
Sensor body Guard tube
Liquid junction Temperature
sensor
Ion selective membrane
Protective cap
Name
Q’ty
Instruction manual (this book) 1 booklet
GHS label
1 pc
Installation procedure
1. Remove the rubber cap of the plug. 2. Connect the slot of the sensor
cartridge with the marker
of the sensor head and push into the sensor head.
Sensor head
Plug
Nut
3. Tighten the nut firmly until the nut is totally tight on the sensor head.
Specifications
Measurement range
Operation temperature range Accuracy1 pH range2
Potential gradient
0.18 mg/L to 1 mol/L
to 18,000 NH4+)
mg/L
NH4+
(1×10-5
mol/L
0°C to 50°C (Sensor cartridge part)
±5%
pH5 to pH8 (100 mg/L NH4+ (25°C))
50 mV to 62 mV (10 mg/L to 100 mg/L NH4+ (25°C))
Sensor body
Polycarbonate/Silicone rubber
Guard tube
Polycarbonate
Nut
ABS
Parts material
Plug
Polyethylene terephthalate/ Brass/Copper alloy
O-ring Membrane Liquid junction
Fluororubber (FPM) Polyvinyl chloride Porous sintered polyethylene
Temperature sensor
NTC/Polyolefin
Protective cap Polyethylene
Wetted parts dimensions Outer diameter 16 mm x Length 110 mm
Liquid junction height
Approximately 15 mm (from the tip of sensor cartridge)
: Wetted material 1: The measurement accuracy is the difference between the measured value and the assigned value of the 100 mg/L standard solution after the two points calibration using 10 mg/L and the 100 mg/L standard solutions. The temperature in the calibration and in the measurement is 25°C. 2: The ion concentration value enclosed in () is the measurable pH tolerance range.
Packaged contents
Name
Q’ty
NH4+ sensor cartridge
1 pc
100 mg/L standard solution 1 pc
1000 mg/L standard solution 1 pc
Ionic strength adjuster
1 pc
Syringe
1 pc
2
Note
Push the plug into the sensor head keeping connecters dry. Avoid using a
rusted plug. In the absence of O-ring or in the case where the nuts is not
tight enough, liquids might flow into the connection space and cause
malfunction.
Maintenance and Storage
Rince the sensor cartridge with pure water and wipe off the moisture. Attach
the protective cap.
Replace sensor cartridge at the period whichever occurs first; approx. one
year after purchase or approx. 400 hours of use, depending on sample solution,
measurement frequency, and measurement environment.
Use up standard solution and ISA before the expiration dater displayed on the
bottle.
Store the standard solution and ISA in a tightly closed bottle in a cool, dark
place.
Disposal procedure
When disposing of this product, follow the related laws and/ or regulations of
your country/region. It is possible to separate the cable part and the
electrode part by cutting the cable (Figure below). The cable part is
classified as plastic. The electrode part is mainly composed of metal, but
contains hard plastic.
Parts name (Refer to specifications and parts material
of the sensor cartridge)
Disposal classification
Sensor part (Sensor body, Guard cap, Plug, Nut)
Plastics (Containing metal)
Bottle
Plastics
Instruction manual, package
Papers
Disposal classifications depend on each country or region.
Reference
Refer to the meter’s instruction manual or the website below.
https://www.horiba.com/water_quality_ 300_ion_c_info_en/