Specifications:

  • Product Name: Gel-filled Potassium Ion Sensor Cartridge
    300-K-C

  • Model: 300-I WQ-300

  • Dimensions: 16mm x 110mm

  • pH Range: 2 to 14

  • Potassium Ion Concentration Range: 100 mg/L to 1000 mg/L

Product Usage Instructions:

Patent Information:

This product is protected under one or more patents. For more
information, visit http://www.horiba.com/patent.

Safety Precautions:

Before using the product, ensure to check the firmware
compatibility. Update the firmware of the ion sensor head or smart
handheld meter if necessary.

Tips for Accurate Measurement:

  • Add the ISA to both standard solutions and sample solution in a
    volume ratio of 50:1.

  • Recalibrate the sensor cartridge every few hours.

  • Prepare standard solutions with low, medium, and high
    concentrations for calibration.

  • Match the temperature of the standard solution during
    calibration with the temperature of the sample solution during
    measurement.

FAQ:

Q: How often should I recalibrate the sensor cartridge?

A: It is recommended to recalibrate the sensor cartridge every
few hours to ensure accurate measurements.

Q: Where can I find information on interfering ions and how to

avoid their effects?

A: For information on interfering ions and measures to avoid
their effects, refer to the website mentioned in the References
section of the user manual.

“`

CODE:GZ0000714258A © 2023 -2024 HORIBA Advanced Techno, Co., Ltd.
May,2024
300-K-C

URL http://www.horiba.com/patent

·
·

pH pH pH 100 mg/L 1

300-I WQ-300
1.0.0
1.0.01.0.11.1.0
https://www.horiba.com/jpn/water-quality/updateinformation/

501 1/10 Web

O

601-8551 2 http://www.horiba-adt.jp

0120-37-6045 9:00 12:0013:00 17:00

0.39 39,000 mg/L K+1×10-5 1 mol/L K+

0 50°C

1

±5%

pH 2

pH5 11100 mg/L K+25°C 50 62 mV10 100 mg/L K+25°C

ABS

O

FPM

16 mm x 110 mm

15 mm

110 mg/L 100 mg/L
2 pH

K+ 100 mg/L 1000 mg/L GHS

1 1 1 1 1 1 1

1. 2.

O

1 400

Web https://www.horiba.com/water_quality_300_ion_c_info/

2

CODE:GZ0000714258A © 2023 -2024 HORIBA Advanced Techno, Co., Ltd.
May,2024
Gel-filled Potassium Ion Sensor Cartridge 300-K-C Instruction Manual

Patent marking
This product is protected under one or more of the patents found at the following address: http://www.horiba.com/patent
Conformable standards
This equipment conforms to the following standards:
RoHS: EN IEC 63000 9. Monitoring and control instruments including industrial monitoring and control instruments
RoHS: BS EN IEC 63000 9. Monitoring and control instruments including industrial monitoring and control instruments

Authorised representative in EU and UK
HORIBA Europe GmbH Hans-Mess-Str.6, D-61440 Oberursel, Germany
HORIBA UK Limited Kyoto Close, Moulton Park, Northampton, NN3 6FL, UK
China regulation

()

(Pb) (Hg) (Cd) (Cr (VI)) (PBB)

(PBDE)

SJ/T 11364 : GB/T
26572 :
GB/T 26572

Caution on use

Safety Precautions

CAUTION
Precautions for chemicals
Handle the solutions used for calibration and measurements with care. Always wear protective gloves and eye protection when handling solutions. · Harmful in contact with skin and eyes. Rinse immediately with
tap water. · If accidentally swallowed, consult a doctor immediately.

Points of concern Do not give an excessive shock to the sensor cartridge. Cracked or chapped sensor cartridge can not be used. Replace such sensor cartridge with a new one. Immerse the sensor up to the sensor body under the nut. Use with the plug facing up and the membrane facing down as shown in the figure. The pH range is generally smaller when the concentration of the ions to be measured is lower. If the pH of the sample solution has exceeded the measurable pH range, accurate measurement may not be possible. Do not allow the liquid junction to dry out. If liquid junction is dried, measurement value may drift. Do not use (store) the sensor cartridge in any place where the operation temperature (storage temperature) is out of the specified range. Do not strongly place ion selective membrane against hard surface. Do not poke ion selective membrane with sharp object. Make sure the membrane has no dirt or air bubbles on it. Before the first use or if the sensor cartridge is less responsive, perform conditioning of the sensor cartridge before calibration. Soak a sensor cartridge in 100 mg/L or more standard solution for 1 hour. Always use new standard solution for calibration. Do not return a standard solution to its original container after it has been used once.
Check the firmware
In an ion sensor head (300-I) and a smart handheld meter (WQ-300) with the following firmware versions, there is no option to select an ion sensor cartridge. Then, it is necessary to update the firmware of the ion sensor head or the smart handheld to the latest version. Make sure to check before using an ion sensor cartridge.

Ion sensor head:

1.0.0

Smart handheld meter: 1.0.0, 1.0.1, 1.1.0

Refer to the instruction manual for smart handheld meter on how to check the firmware. It is possible that these firmwares will be updated from time to time as new functions and improvements are made. Check the following website for how to obtain the latest firmware and how to update the firmware. https://www.horiba.com/sgp/water-quality/support/ download-center/

Tips for accurate measurement Add the ISA to both standard solutions and sample
solution. The volume ratio shall be the standard solution
and the ISA = 50:1. Recalibrate sensor cartridge every few hours. Prepare standard solutions which have the following
concentration relationship for calibration.
Standard solution with low concentration < Sample
solution < Standard solution with high concentration Match the temperature of the standard solution in
calibration and the temperature of the sample solution in
measurement.

2 Miyanohigashi-cho, Kisshoin, Minami-ku, Kyoto, 601-8551, Japan http://www .horiba-adt.jp

For any questions regarding this product, please contact your local agency, or inquire from the following website. http://global.horiba.com/contact_e/index.htm

Confirm the effects of interfering ions and take measure to avoid them because the influence of interfering ions varies depending on the composition of the sample solution. For the main interfering ions, refer to the website written in “Reference”.
Specifications and parts description
Parts name
Rubber cap

Plug

O-ring Nut

Sensor body Guard tube

Liquid junction Temperature sensor

Ion selective membrane

Protective cap

Name

Q’ty

Instruction manual (this book) 1 booklet

GHS label

1 pc

Installation procedure
1. Remove the rubber cap of the plug. 2. Connect the slot of the sensor cartridge with the marker
of the sensor head and push into the sensor head.
Sensor head

Plug

Nut
3. Tighten the nut firmly until the nut is totally tight on the sensor head.

Specifications

Measurement range

0.39 mg/L to 39,000 mg/L K+ (1×10-5 mol/L to 1 mol/L K+)

Operation temperature range
Accuracy1
pH range2

0°C to 50°C (Sensor cartridge part)
±5% pH5 to pH11 (100 mg/L K+ (25°C))

Potential gradient

50 mV to 62 mV (10 mg/L to 100 mg/L K+ (25°C))

Sensor body

Polycarbonate/Silicone rubber

Guard tube

Polycarbonate

Nut

ABS

Parts material

Plug

Polyethylene terephthalate/ Brass/Copper alloy

O-ring Membrane Liquid junction

Fluororubber (FPM) Polyvinyl chloride Porous sintered polyethylene

Temperature sensor

NTC/Polyolefin

Protective cap Polyethylene

Wetted parts dimensions Outer diameter 16 mm x Length 110 mm

Liquid junction height

Approximately 15 mm (from the tip of sensor cartridge)

: Wetted material 1: The measurement accuracy is the difference between the measured value and the assigned value of the 100 mg/L standard solution after the two points calibration using 10 mg/L and the 100 mg/L standard solutions. The temperature in the calibration and in the measurement is 25°C. 2: The ion concentration value enclosed in () is the measurable pH tolerance range.

Packaged contents

Name

Q’ty

K+ sensor cartridge

1 pc

100 mg/L standard solution 1 pc

1000 mg/L standard solution 1 pc

Ionic strength adjuster

1 pc

Syringe

1 pc

2

Note
Push the plug into the sensor head keeping connecters dry. Avoid using a rusted plug. In the absence of O-ring or in the case where the nuts is not
tight enough, liquids might flow into the connection space and cause malfunction.

Maintenance and Storage
Rince the sensor cartridge with pure water and wipe off the moisture. Attach the protective cap.
Replace sensor cartridge at the period whichever occurs first; approx. one year after purchase or approx. 400 hours of use, depending on sample solution, measurement frequency, and measurement environment.
Use up standard solution and ISA before the expiration dater displayed on the bottle.
Store the standard solution and ISA in a tightly closed bottle in a cool, dark place.
Disposal procedure
When disposing of this product, follow the related laws and/ or regulations of your country/region. It is possible to separate the cable part and the electrode part by cutting the cable (Figure below). The cable part is classified as plastic. The electrode part is mainly composed of metal, but contains hard plastic.

Parts name (Refer to specifications and parts material
of the sensor cartridge)

Disposal classification

Sensor part (Sensor body, Guard cap, Plug, Nut)

Plastics (Containing metal)

Bottle

Plastics

Instruction manual, package

Papers

Disposal classifications depend on each country or region.

Reference
Refer to the meter’s instruction manual or the website below. https://www.horiba.com/water_quality_ 300_ion_c_info_en/

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals