GIMA LD-582 Digital Blood Pressure Monitor Instruction Manual
- September 23, 2024
- GIMA
Table of Contents
LD-582 Digital Blood Pressure Monitor
“`html
Product Information
Specifications:
- Main Body
- Display
- Air Connector
- Tube plug
- Air Hose
- Cuff
- D-ring
Product Usage Instructions:
Battery Installation:
-
Open the battery cover and then open the button battery
cover. -
Install one ‘CR2025’ button battery into the button battery
compartment. -
Close the button battery compartment cover.
-
Install four ‘AA’ type batteries into the battery compartment
as indicated, ensuring correct polarity. -
Close the battery compartment cover.
Assembly the Storage Case:
-
Press button ‘PRESET’, and the year number will flash.
-
Use buttons ‘MEM1’ or ‘MEM2’ to adjust the number, then press
‘START’ for confirmation. -
Follow similar steps to set month, date, and time.
-
Press ‘START’ to finish the setup.
Reminders Clearance:
- When the device is in standby mode, press ‘PRESET’ twice to
enter alarm 01 mode.
Correct Posture:
-
Sit beside a table with your arm supported by the table during
measurement. -
Sit upright with a straight back.
-
Ensure the cuff on the upper arm is not crossed and is at heart
level. -
Keep your feet flat on the ground.
-
If lying down, look at the ceiling, stay calm, and avoid moving
your neck or body during measurement.
FAQ
Q: What do I do if the display shows nothing after pressing the
START button?
A: If nothing appears on the display after pressing START, it
may indicate worn batteries. Ensure correct battery installation
and replace if needed.
Q: How do I set the date and time on the device?
A: Press PRESET to adjust the year, use MEM1 or MEM2 to change
numbers, then press START for confirmation. Follow similar steps
for month, date, and time settings.
“`
PROFESSIONAL MEDICAL PRODUCTS
DIGITAL BLLOD PRESSURE MINITOR INSTRUCTION MANUAL Parts and Components
1. Main Body 2. Display 3. Air Connector 4. Tube plug 5. Air Hose 6. Cuff 7. D-ring
8. Button ‘PRESET’ 9. Button’MEM1′ 10. Button e’MEM2′ 11. Button ‘ START’ 12. Storage Case 13.4xAA Batteries(Optional) 14.AC Adapter(Optional)
SYMBOLS Symbols
START
Manufacturer
Meaning
Authorized Representative in the European community
WEEE disposal
Medical Device complies with Directive 93/42/EEC
Keep in a cool, dry place
Follow instructions for use
Type BF applied Part Stand by Lot number Product code Date of manufacture
Importer
GENERAL This instruction manual is intended to assist the user for safe and efficient operation of the automatic digital blood pressure monitor (hereinafter: device) model LD-582.The device must be used in accordance with the procedures described in the manual. It is important to read and understand the entire manual, especially the section <IMPORTANT SAFETY INSTRUCTIONS
.This device is intended for the non-invasive measurement of systolic and diastolic arterial blood pressure and pulse rate in adults (age 15 and above). CAUTION: 1. Do not use this device on infants or persons who cannot express their intentions 2. The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on elder children. 3. The patient is an intended operator. But persons who suffer from arrhythmia.diabetes, cardiovascular problems or who have had a stroke should consult your doctor before using the device. PRINCIPLE OF OPERATION This device adopts the oscillometric technology with Fuzzy Algorithm to measure the arterial blood pressure and pulse rate. The cuff is wrapped around the arm and automatically inflated by the air pump. The sensor of the device catches weak fluctuation of the pressure in the cuff produced by extension and contraction of the artery of the arm in response to each heartbeat. The amplitude of the pressure waves is measured, converted in millimeters of the mercury column, and is displayed by digital value. ATTENTION: This device can not provide reasonable accuracy if used or stored the temperature, humidity or altitude beyond the range stated in the section
of this manual. NEW TECHNOLOGIES USED Fuzzy Algorithm is the processing algorithm, taking into account the specialty of individual heartbeats, which provides higher accuracy of measurement. Software version:V1.1 IMPORTANT SAFETY INSTRUCTIONS It is necessary to know that arterial blood pressure is subjected to sharp fluctuations. The level of the arterial blood pressure depends on many factors.Generally arterial blood pressure is lower in summer and higher in winter. Arterial blood pressure changes with atmospheric pressure and is affected considerably by many factors, e.g.physical loads, emotional excitability, stress, meals, etc.Medicines, drinking, smoking affect greatly the level of an individual’s blood pressure.Blood pressure does vary with age and individuals, and it is recommended to write down the readings from blood pressure records daily, then you can check with your doctor to find out what is a “normal blood pressure measurement” for you. Please read the instruction manual carefully before using this device, especially <Important safety instructions>, it can help you use the device correctly and safely!Please keep the instruction manual for future use. For specific information aboutyour own blood pressure, consult your physician. WARNINGS · Consult your physician if you suffer from illnesses prior to using the device. · The device is not suitable for persons who have electrical implants.If you had a mastectomy
(breast amputation) do not use this blood pressure monitor on the arm on the side of the mastectomy. · Pregnant women should only measure their own blood pressure in consultation with their doctor, since the readings may be changed with pregnancy. · Do not service or maintain the cuff while in use with patient. · Do not use this blood pressure monitor on any arm where intravascular access or therapy (such as an intravenous drip or a blood transfusion), or an arteriovenous shunt (A-V shunt) is present. The temporary interference to blood flow by the blood pressure measurement could result in injury. · Do not use the device with other medical electrical(ME) equipment simultaneously. · Do not use the device in the area the HF surgical equipment, MRI, or CT scanner exists, or in the oxygen rich environment. · Do not use a mobile phone or other devices that emit electromagnetic fields, near the device. This may result in incorrect operation of the device. · Never use any accessories or parts from other manufacturers, Using such accessories or parts could cause a hazardous situation for the user or damage to the device.
· Do not modify this equipment without authorization of the manufacturer. ·
The batteries used in this device may present a fire or chemical burn hazard
if mistreated.
Do not disassemble, heat or incinerate. · Keep equipment away from fire and
heat sources to prevent fire or explosion. · Please keep the unit out of reach
of infants, children or pets, since inhalation or swallowing
of small parts can be dangerous or even fatal. · Please pay attention that the
continuous CUFF pressure due to connection tubing kinking
will cause a harmful injury. · Do not use an extension cord with this device.
· The air tube or the AC adapter cable may cause accidental strangulation in
infants. · Do not put the air tube around your neck- danger of suffocation!A
device should never be
left unattended when plugged in. · Do not reach for a corded device that has
fallen into water. Unplug immediately. · It is quite normal that two
measurements taken in quick succession may produce
significantly different results, because too frequent and consecutive
measurements could cause disturbances in blood circulation and injuries
Cautions · Use this device under the right environmental conditions as
indicated in this user manual. lf not, this could affect the performance,
lifetime of the device and measurement results. · Only use this device for its
intended purpose as described in this user manual. · Do not confuse self-
monitoring with self-diagnosis. This device allows you to monitor your blood
pressure. Do not begin or end medical treatment based on the measurement
results. Always consult your physician for treatment advice. · Do not take any
therapeutic measures on the basis of a self-measurement.Never change
prescribed medication without consulting your physician. Consult our physician
if you have any questions about your blood pressure. · If you are taking
medication, consult your physician to determine the most appropriate time to
measure your blood pressure. · Consult the physician if measurement errors
occur in children or persons with arrhythmia. · The pulse display is not
suitable for monitoring the frequency of cardiac pacemakers. · Common
arrhythmias (such as atria or ventricular premature beats or atrial
fibrillation) and peripheral artery disease /arteriosclerosis can affect the
accuracy of this blood pressure monitor. Please consult your physician how to
best use this blood pressure monitor if you suffer from any of these
conditions. Blood pressure measurement is not suitable in cases of serious
arteriosclerosis (hardening of the arteries). · The effectiveness of this
blood pressure monitor has not been established in pregnant women. · Always
check the device and cuff before you use it. Do not use the device or cuff if
one of them is damaged, because this may cause injury. · This device is not
intended for use on extremities other than the arm or for functions other than
obtaining a blood pressure measurement · Do not attach the cuff on the same
arm on which other monitoring medical electrical equipment is attached
simultaneously, because this could cause temporary loss of function of those
simultaneously-used monitoring medical electrical equipment. · Never attach
the cuff on injured skin, an injured arm or an arm under medical treatment as
this can cause further injury. · Do not forcibly crease the arm cuff or the
air tube excessively. · Do not press the air tube while taking a measurement.
· Do not use the device in case of existing polyester or nylon material
allergies. · This device is not suitable for continuous monitoring during
medical emergencies or operations. · This device cannot be used with HF (High
Frequency) surgical equipment at the same time. · This device is not washable.
Never immerse the device in water and do not rinse it under the tap. · This
device should keep dry to prevent from moisture. · The equipment is not AP/APG
equipment and is not suitable for use in the presence of a flammable
anesthetic mixture with air, with oxygen or nitrous. · To avoid measurement
errors, do not use the device near strong electromagnetic fields, radiated
interference signal or electrical fast transient/burst signal. For example
magnets, radio transmitters, microwave ovens. · If this device was stored in
ow temperature, eave it in room temperature for at least 1 hour. · Repeated
measurements with an interval of 3 minutes are recommended, so you can
calculate the average to get a more accurate measurement. An internal of
3minutes can also ensure that the operation of the device does not result in
prolonged impairment of the circulation of the blood. · Atherosclerosis
patients may require longer interval (10-15minutes) as elasticity of patient’s
vessels decreases significantly with the disease.10-15minutes interval is also
applicable for patients suffering from diabetes for a long period of time. ·
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution. · Connecting electrical equipment to mso effectively leads to
creating a ME system,and can result in a reduced level of safety
CLASSIFICATION · ME EQUIPMENT not intended for use in an oxygen rich
environment or in the presence of flammable mixers. · Internally powered
equipment (without adaptor), Class ll equipment (with adapter). · Type BF
applied part, recognize he cuff as applied part
BATTERY INSTALLATION
1. Open the battery cover and then open the button battery
cover, then install one `CR2025’button battery into the button
battery compartment;
2. Close the button battery compartment cover;
3. Install four `AA’ type batteries into the battery compartment
as indicated. Make sure that the polarity is correct;
4. Close the battery compartment cover.
· Inbuilt button battery for keeping the date/time
uninterruptedly during changing the batteries(4x AA
batteries).If the new batteries are installed into the device,the
date and time displays’01/01’and’00:00′ icon in the LCD, it
indicates that you need change the new button battery.
· Replace the batteries when the replacement indication
` ‘ appears in the display or nothing after START button
is pressed:
· Batteries in this kit are intended to check work capacity of
the device and the life-span of the batteries can be shorter
than the recommended
· Replace all batteries simultaneously, and don’t use
rechargeable batteries;
· Only same type batteries are allowed to be used together;
· If the device is to be unused for long time, please take out
the batteries. Don’t leave the worn batteries in the device;
· When the low battery indication’ ‘appears in the display
or nothing after START button is pressed;
· Batteries in this kit are intended to check work capacity of
the device and the life-span of the batteries can be shorter
than the recommended;
· Replace all batteries simultaneously, and don’t use
rechargeable batteries;
· Only same type batteries are allowed to be used together;
· If the device is to be unused for long time, please take out
the batteries.Don’t leave the worn batteries in the device;
· When the low battery indication’
‘ flashes on the LCD
during measurement, it reminds that the user will change all
the batteries but can be used currently,;
· when the low battery indication’ `in the LCD and at
the same time the buzzer beeps for 4 times continuously, i
indicates that the user need change all batteries at once
USE AC POWER ADAPTER Besides batteries you can use AC power adapter as the power supply. The AC power adapter is optional for this device for sale. The AC adapter is specified as a part of the blood pressure monitor. · Insert the AC adapter cord into the jack on the right side of the monitor. · Insert the AC adapter plug into the outlet. · To remove the AC adapter, disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor’s jack.
Caution · When using optional AC adapter, the AC adapter must comply with the
requirements of
standard lEC60601-1. · To avoid possible damage of the monitor, use only the
exclusive AC adapter that can be
purchased from authorized dealers. Other adapter may damage the blood pressure
monitor. · The AC adapter is used as an isolating means, the AC adapter plug
shall insert into the outlet
nearby the operator, make it easy to disconnection the device from the outlet.
· lf long time work, remove the plug after the adapter cools, and prevent
bums. · Plug the AC adapter into the appropriate voltage outlet. Do not use in
a multi-outlet plug. · Not to position the blood pressure monitor to make it
difficult to operate the disconnection
device( adaptor). Note: The monitor is designed not to draw power from the
batteries when the AC adapter in use.
Optional AC adapter technical feature: Model: YS5M-0600600 Input: 100-240V
50/60Hz Output voltage: 6V+5% Output current: 600 mA Output plug polarity: <->
inner
ASSEMBLY THE STORAGE CASE
1. Three hooks of storage case aim at the concave of device respectively; 2.
Push the storage case upwards; 3. To fill tightly with the plug. SETTING DATE
AND TIME The function provides accurate measuring time for each measurement.
To get accurate date and time, the user should preset the date and time
correctly before the first use of this device. The operation procedure for
presetting Date/Time is as follows: 1. When the device is connected to power
supply at first time, the display will show as Fig.1:
2. Press button PRESET’, and the year number flashes; 3. Press button
MEM1’or MEM2′ to subtract or add the number, and press button
START’
for confirmation; 4. When the year setup is finished, the month number will
flash automatically as Fig.3.Please
follow the same instruction as above tc set month.date and time: 5. Press
button START’ to finish setup. lf you want to change the date and time,please repeat procedure 2, 3, 4. FUNCTION OF REMINDERS SETTING REMINDERS This monitor has 3 reminder alarms. You can set 3 different reminder alarms within a 24 hours period. 1.When the device stands by, press button
PRESET’ two times to enter into alarm 01 mode,
the display will show as Fig.4; 1. Press button MEM1’or’MEM2′ the display will show as Fig.5 and at the same time the hour number flashes; 2. Press button
MEM1’or’MEM2′ again to subtract or add the number, and press button
START’ for confirmation. 3. When the hour number setup is finished, the minute number will flash automatically.Please follow the same instruction as above to set minute number; 4. Press button
START’ for confirmation. 6.When
the device stands by, press button PRESET’ three and four times respectively to enter into alarm 02 and 03 mode. Repeat the above. Annotation: When the alarm is on under device stands by, the icon
‘ flashes
on LCD and goes with beep for 1 minute,Press the button START’ to turn off the alarm. When the alarm is on during measurement, the icon’
flashes on LCD
for 1 minute without beep. Under this situation, if you press the button
START’,it will stop both the icon ‘ ‘ flashing and measurement. REMINDERS CLEARANCE 1. When the device stands by, press button
PRESET’ two
times to enter into alarm 01 mode,
then press button MEM1′ for at least 5 seconds, the display will show as Fig.7 which means the alarm 01 is removed. 2. When the device stands by, press button
PRESET’ three and four times respectively to enter into alarm 02 and
03 mode . Repeat the above process to remove the alarm 02 and alarm 03.
AMBIENT TEMPERATURE DISPLAY AND ADJUSTMENT This monitor can display the
ambient temperature and the unit °C and °F can be adjustable.°C mode display
in the LCD when the first time used. 1. When the device stands by, press
button PRESET’ five times to enter into the temperature adjustable mode , then press button
MEM1 to turn into F mode and press button `START’ for
confirmation. 2. Press button’MEM2 to convert °F mode into °C mode.
Annotation: When under the mode of function reset,if without any operation in
1 minute, the device will automatically return standby mode.
Please keep quiet for 5-10 minutes and avoid eating, drinking, alcohol,
smoking,exercising and bathing before taking a measurement, All these factors
will influence the measurement result. · Remove any garment that fits closely
to your upper arm. · Always measure on the same arm(normally left). ·
Measurements should be taken regularly at the same time of each day, as the
blood
pressure varies even during the day. · Any effort to support the arm during
measurement may increase the measured blood
pressure. · Make sure, you are in a comfortable, relaxed position with leg-
uncrossed, feet flat on the
floor, back and arm supported, middle of the cuff at the level of the right
atrium of the heart and do not move or constrict your muscles and talk during
measurement, Use a cushion to support your arm if necessary, Keep position in
normal use. · If the arm artery lies lower or higher than the heart, a false
reading will be obtained. · A loose or open cuff causes false readings. · With
repeated measurements, blood accumulates in the arm which can lead to false
reading. · Consecutive blood pressure measurements should be repeated after 1
minute pause or after the arm has been held up in order to allow the
accumulated blood to flow away
CORRECT POSTURE 1. Sit beside the table and let the table support your arm as
you take the measurement. 2. Sit upright with your back straight. 3. Make sure
that the cuff on the upper arm no cross, and is at approximately the same
level as the heart. 4.Make sure that your feet lie on the ground and no cross.
5.You may lie on your back and take a measurement.Look at the ceiling, keep
calm, and don’t move your neck or body during the measurement.
ASSEMBLY THE CUFF 1. Insert the edge of the cuff approximately 5 centimeters
into the D-ring as shown in figure. 2.Put the cuff on the left upper arm with
the tube pointing to the direction of palm.lf measurement on your left arm is
difficult, you can use right arm for measurement. In this case, it is
necessary to know that the readings may differ about 5-10 mmHg between left
arm and right arm. 3. Wrap cuff around your upper arm with the lower edge of
the cuff approximately 2-3 centimeters above the elbow. The mark
MEMORY CLEARANCE After a measurement is finished or when the device stands by, hold down button’MEM1’or’MEM2′ for at least 5 seconds,the display will show `CLR’ which means the stored reading for’MEM1’or’MEM2′ is removed.
ERROR AND LOW BATTERYINFORMATION
INDICATION
POSSIBLE REASON
CORRECTION METHODS
The cuff is put on wrongly or Make sure that cuff is put on the tube plug is inserted too correctly and the tube plug
loosely.
is inserted tightly and repeat
Movement of arm/hand or the measurement.Repeat the talking during measurement. measurement with following The cuff is not inflated to completely recommendations
necessary pressure.
of manual. Repeat the measurement with pumping cuff to higher
pressure.
The batteries are weak
Replace all 4 batteries with new ones.
TROUBLESHOOTING
SYMPTOM
CHECK POINT
REMEDY
No display when connect the power The batteries have run down.The Replace all the batteries with new
polarity of battery is wrong.The ones.Install the batteries correctly.
contact of battery compartment Clean the battery terminals with dry
is polluted.
cloth.
Inflation stops and reinflate later.
The automatic inflation for ensuring See
The readingis extremely low or high. Is the cuff at the same level as Make sure that your posture is right. the heart?Is the cuff wrapped Wrap the cuff correctly.Relax during right?Did you strain your arm measurement.Keep quiet and silent during measurement?Did you talk during the measurement. or move your arm(or hand) during measurement?
Pulse rate is too low or too high.
Did you talk or move your arm(or Keep quiet and silent during the
hand) during measurement?Did measurement.Take measurement
you make measurement right after again after resting for more than
exercise?
5minutes.
The batteries are run down soon. Faulty batteries are used.
Use alkaline batteries of known manufacturers.
CARE,STORING,REPAIR AND RECYCLING 1. It’s necessary to protect this device against high moisture, direct sunlight,shock, solvent, alcohol and gasoline. 2. Remove the batteries if the device is being stored for a long time, and keep the batteries far away from children. 3. Keep the cuff away from sharp objects and don’t extend or twist the cuff. 4. This device is not washable. Never immerse the device in water and do not rinse it under the tap. Use only soft and dry cloth to clean the device. 5. Do not serve or maintain the cuff and the device when in use with patient. 6. The cuff is sensitive and must be handled with care. You can clean the cuff with damp cloth for daily maintenance.To avoid cross infection when sharing the cuff, you can sterilize the fabric cover ofthe cuff with tampons moistened by 3% solution of hydrogen dioxide. After long use there will be a partial discoloration on the fabric surface of the cuff. Do not laundry the cuff as well as ironing with a hot flatiron. WARNING: Under no circumstances may you wash the inner bladder! 7. Since neither the device nor batteries are household waste, follow your local recycling rules and dispose them at an appropriate collection site. 8. Do not open the device, or delicate electrical components as an intricate air unit could be damaged. lf you can not fix the problem using the troubleshooting instruction, please request service from your dealer. WARNING: Do not repair the device without manufacturer’s authorization. Do not carry out maintenance when using the device. Caution: Generally, we recommend the device should be inspected every 2 years and utilize the manometer mode to verify the accuracy of the manometer at least at 50mmHg and 200mmHg after maintenance and repair. Please contact your dealer for maintenance.
6.Press the button START’ to return to standby mode. Please rest for at least 3 minutes for another measurement. The device keeps unused for3 minutes, the device will be return to standby mode automatically. 7. lf irregular heartbeat was detected during the measurement, LCD display the icon’ ‘ to remind users of heartbeat irregularity. NOTE: The device will inflate to a higher pressure automatically in case the inflation pressure is not sufficient to determine measurement result. Attention: We recommend contacting your physician if you see the
‘indicator frequently. WHO CLASSIFICATION INDICATION Standards for
assessment of high or low blood pressure, regardless of age, have been
established by World Health Organization(WHO) as show in the chart as below
The indicator display a segment, based on the current data, corresponding to
the WHO classification. For example, if your blood press is 145mmHg (systolic
pressure), 88mmHg(diastolic pressure), according to the world health
organization standard, your blood pressure level is Mild Hypertension. Note:lf
the systolic blood pressure and the diastolic blood pressure fall into
different categories, the higher value should be taken for classification.
FUNCTION OF MEMORY MEMORY RECALL 1. LD-582 can store 60 sets of readings each
in’ ‘and `, and will automatically calculate the average value of the latest 3 readings for MEM1’and
MEM2′ respectively. When the memory is full (60 sets
of readings are stored), the oldest reading will be replaced by new one.
Memory will not clear away even if power supply is removed; 2. After a
measurement is finished or when the device stands by,the user can press button
MEM1’or button
MEM2′ to recall memory.Press button’MEM1′ or `MEM2′, the
display will show the average value of the latest 3readings as Fig.14;
SPECIFICATIONS
Model
LD-582
Size
158(L)x120(W)x127(H)mm
Weight
Approximately 490g without batteries
Measuring method
Oscillometry
Extreme Pressure/ cuff pressure
290mmHg
Measuring range
40 to 180mmHg(DIA,diastolic pressure) 60 to 260mmHg(SYS,systolic pressure) 40 to 160 beats/minute(PUL,pulse rate)
Measuring accuracy
± 3 mmHg for static pressure ± 5% of the reading for the pulse rate
Inflation
Automatic by the pump
Rapid deflation
Automatic electronic valve
Batteries
Optional component, 4″AA”x1.5V
Adapter
Optional component,6V.600mA
Memory
2 Users with 60 sets of memory each
Operation temperature and humidity, air pressure +10°C to+ 40°C,85% and below 800hPa to 1060hPa
Transport and storage temperature and humidity, -20°C to +50°C,85% and below 500hPa to
air pressure
1060hPa
Upper arm circumference
Applicable for arm circumference 22-32 cm (standard cuff)
Complete kit
Main body, storage case, cuff, 4 x AA batteries (Optional), 1 x CR 2025 button battery, adapter (Optional), instruction manual
Pollution Degrees
Degrees 2
Over voltage category
Category ll
High Altitudes(m)
2000m
Fuse
1A6V 2.1mm1.45mm0.81mm
MANUFACTURER’S DECLARATION Compliance information for each EMC test
Electromagnetic Emission (Home Healthcare Environment)
Emission test (lEC60601-1-2:2014)
Compliance
Conducted and radiated RF emissions
CISPR 11 Group1 Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC Complies 61000-3-3
Compliance information for each EMC test Declaration -Electromagnetic Immunity (Home Healthcare Environment)
Immunity test
IEC 60601 test level
Compliance level
Conducted RFIEC 61000-4-6
3V 150 kHz to 80 MHZ 6V in 3V 150 kHz to 80 MHZ 6V in ISM and amateur radio bands ISM and amateur radio bands between 0.15 MHz and 80 MHz between 0.15 MHz and 80 MHz
Radiated RFEC61000-4-3
10 V/m
10 V/m
80 MHz to 2.7GHz also meet 80 MHz to 2.7GHZ also meet
the requirement of table 9 of IEC the requirement of table 9 of lEC
60601-1-2:2014.
60601-1-2:2014.
Electrostatic discharge (ESD) +8kV contact
+8kV contact±2 kV, ±4 kV, ±8 kV,
EEC 61000-4-2
+2 kV, +4 kV, ±8 kV, +15 kV air ±15 kV air
Electrical fast transient/burst +2kV for power supply lines +2kV for power supply lines
IEC 61000-4-4 Surge IEC 61000-4-5
+0.5kV,+ 1 kV line(s) to lines + 0.5kV, ± 1 kV line(s) to lines
Voltage dips,short interruptions 0% Ut,0.5 Cycle at 0°, 45°, 90° 0% Ut,0.5 Cycle at 0°, 45°, 90°
and voltage variations on power 135°, 180°225°, 270° and 315° 135°, 180°225°, 270° and 315°
supply input lines IEC 61000- 0% Ut,1 Cycle and 70% Ut 0% Ut,1 Cycle and 70% Ut
4-11
25/30 cycles sigle phase: at 0° 25 cycles sigle phase: at 0° 0%
0% Ut, 250/300 cycles
Ut, 250 cycles
Power frequency (50/60Hz) 30 A/m
30 A/m
magnetic field IEC 61000-4-8
NOTE: The EUT is the a.c. mains voltage prior to application of the test level.The following phenomenon is still fulfill the requirement of basic safety and essential performance. * UT:230V~/50Hz, The pressure of the EUT is deviation the normal value but the value is still more than 10psi when flow is 4.5l/min. **UT:230V~/50Hz, The EUT stop working when adding 0% UT, but the EUT can restore its normal mode automatically. · Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally. · Portable RF communications equipment(including peripherals such as antenna cables and external antennas) should be used no closer than 30cm(12 inches to any part of this device, including cables specified by the manufacturer. Otherwise,degradation of the performance of this equipment could result. · Under the test condition specified in immunity, the product can provide the basic safety and essential performance.If the essential performance is lost or degraded, additional measures are necessary,such as reorienting or relocating the device.
QUALITY GUARANTEE MODEL
Warranty Period
Two-years from purchasing date
Purchasing date
Purchasing shop
Name:
Telephone:
Address:
Customer
Name:
Telephone:
Address:
1. Warranty for this automatic digital blood pressure monitor is 24 months from the date of purchase. The 24 months warranty excludes the monitor cuff. The cuff is warranted for 12 months. 2. The warranty obligations are prescribed for by warranty certificate for buyer. 3. The addresses of organizations for guarantee maintenance are present in the warranty certificate. WARNING Do not modify this equipment without authorization of the manufacturer. All major maintains on the device must be performed by an authorized service center or distributor. No use-serviceable parts inside, before servicing to authorized representative or manufacturer! DECLARATION: When technical information for user or service personnel requirements is not in the scope of confidentiality of the Company, the Company committed to provide information disclosure in accordance with procedure, including circuit diagrams and parts lists, and other related type technology information that do not involve commercial secrets may be disclosed, Access to information channels and procedures,please contact your dealer or manufacturer.
REQUIRING RECORD
Date
TROUBLE
SERVICE MAN
Guarantee Regulation
1. During warranty period the repair could be made at any BPM repair department. 2. The following things not belong to warranty range: (1) Operating BPM different from procedures or instructions of the manual. (2) The body is damaged artificially. (3) Self-repairing or modifying the monitor construction in any way. (4) Breakdown due to corrosion of battery leakage. (5) Problem which occurs under natural calamity and other force maieures.
PERIODIC SAFETY CHECKS If you use the device with power adapter, preventive inspection and maintenance to be performed including the frequency of such maintenance. Every time before use, please check the adapter, once damaged, never to use. Please clean the plug of adapter plug at least once a year. Too much dust on plug may cause the fire. The manufacturer reserves the right to make technical changes without notice in the interest of progress.Prior notices will not be given in case of any amendments within this manual. The mentioned trademarks and names are owned by the corresponding companies.
GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.
3. Press again,the display will show 01′, which means the latest reading, then turns to another screen to show readings and measuring time as Fig.15; 4.Press again, the display will show
02’, which means the second to the
latest reading…
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment.
Made in China
LD-582 (GIMA 32778) Gima S.p.A. Via Marconi,1 20060 Gessate (Mi) – Italy
gima@gimaitaly.com – export@gimaitaly.com www.gimaitaly.com
M32778-M-Rev.0.07.24