FLUKE 287,289 True-rms Digital Multimeters User Manual
- September 17, 2024
- FLUKE
Table of Contents
287,289 True-rms Digital Multimeters
“`html
Specifications:
- Model: 287/289 True-rms Digital Multimeters
- Release Date: June 2007, Rev. 2, 3/09
- Warranty: Lifetime Limited Warranty
- Manufacturer: Fluke Corporation
- Country of Origin: USA/Netherlands
Product Usage Instructions:
Introduction:
XWWarning: Read Safety Information before using this Meter.
The descriptions and instructions in this manual apply to the
model 289 and model 287 True-rms Digital Multimeters (hereafter
referred to as the Meter). The model 289 appears in all
illustrations.
Contact Information:
To contact Fluke, call:
- USA: 1-888-993-5853
- Canada: 1-800-363-5853
- Europe: +31 402-675-200
- Japan: +81-3-3434-0181
- Singapore: +65-738-5655
- Worldwide: +1-425-446-5500
Visit Fluke’s website at: www.fluke.com
Register your Meter at: http://register.fluke.com
Safety Information:
The Meter complies with:
-
XWWarning: To avoid possible electric shock or personal injury,
follow guidelines provided in the manual. -
WCaution: To avoid damage to the Meter or equipment under test,
follow safety guidelines.
Symbols:
Table 1 lists and describes the symbols used on the Meter and in
this manual.
FAQ:
Q: How do I register my Meter with Fluke?
A: You can register your Meter by visiting the Fluke website at
http://register.fluke.com and following the instructions
provided.
Q: What should I do before connecting the Meter in a
circuit?
A: Always remember to place the Meter in series with the circuit
to ensure safe usage and accurate readings.
“`
®
287/289
True-rms Digital Multimeters
June 2007, Rev. 2, 3/09
© 2007, 2008 Fluke Corporation. All rights reserved. Specifications subject to
change without notice. All product names are trademarks of their respective
companies.
Users Manual
Lifetime Limited Warranty
Each Fluke 20, 70, 80, 170, 180 and 280 Series DMM will be free from defects
in material and workmanship for its lifetime. As used herein, “lifetime” is
defined as seven years after Fluke discontinues manufacturing the product, but
the warranty period shall be at least ten years from the date of purchase.
This warranty does not cover fuses, disposable batteries, damage from neglect,
misuse, contamination, alteration, accident or abnormal conditions of
operation or handling, including failures caused by use outside of the
product’s specifications, or normal wear and tear of mechanical components.
This warranty covers the original purchaser only and is not transferable.
For ten years from the date of purchase, this warranty also covers the LCD.
Thereafter, for the lifetime of the DMM, Fluke will replace the LCD for a fee
based on then current component acquisition costs.
To establish original ownership and prove date of purchase, please complete
and return the registration card accompanying the product, or register your
product on http://www.fluke.com. Fluke will, at its option, repair at no
charge, replace or refund the purchase price of a defective product purchased
through a Fluke authorized sales outlet and at the applicable international
price. Fluke reserves the right to charge for importation costs of
repair/replacement parts if the product purchased in one country is sent for
repair elsewhere.
If the product is defective, contact your nearest Fluke authorized service
center to obtain return authorization information, then send the product to
that service center, with a description of the difficulty, postage and
insurance prepaid (FOB Destination). Fluke assumes no risk for damage in
transit. Fluke will pay return transportation for product repaired or replaced
in-warranty. Before making any non-warranty repair, Fluke will estimate cost
and obtain authorization, then invoice you for repair and return
transportation.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A
PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR
ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,
INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. AUTHORIZED RESELLERS
ARE NOT AUTHORIZED TO EXTEND ANY DIFFERENT WARRANTY ON FLUKE’S BEHALF. Since
some states do not allow the exclusion or limitation of an implied warranty or
of incidental or consequential damages, this limitation of liability may not
apply to you. If any provision of this warranty is held invalid or
unenforceable by a court or other decision-maker of competent jurisdiction,
such holding will not affect the validity or enforceability of any other
provision.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
5/07
Table of Contents
Title
Page
Lifetime Limited Warranty………………………………………………………………………………….. ii Introduction
…………………………………………………………………………………………………….. 1 Contacting Fluke
……………………………………………………………………………………………… 1 Safety Information
……………………………………………………………………………………………. 1 Hazardous
Voltage…………………………………………………………………………………………… 3 Symbols
…………………………………………………………………………………………………………. 4 Features
…………………………………………………………………………………………………………. 5
Understanding the Push Buttons ……………………………………………………………………. 5 Using Auto Repeat
………………………………………………………………………………………. 6 Understanding the Display
……………………………………………………………………………. 7
Bar Graph ………………………………………………………………………………………………. 8 Status Bar Elements
………………………………………………………………………………… 8 Page
Area………………………………………………………………………………………………. 9 Softkey Labels
………………………………………………………………………………………… 9 Adjusting Display
Contrast………………………………………………………………………… 9 Understanding the Rotary Switch
…………………………………………………………………… 10 Using the Input
Terminals……………………………………………………………………………… 11 Controlling Meter
Power……………………………………………………………………………………. 12
i
287/289
Users Manual
Powering the Meter On and Off Manually ……………………………………………………….. 12 Battery
Level Indicator …………………………………………………………………………………. 12 Automatic Power-Off
……………………………………………………………………………………. 12 Battery Saver Mode
…………………………………………………………………………………….. 12 Controlling the Backlight
………………………………………………………………………………. 13 Selecting the Range
………………………………………………………………………………………… 13 Understanding Function Menus
…………………………………………………………………………. 13 Input AlertTM
Feature………………………………………………………………………………………… 15 Using the Info Button
……………………………………………………………………………………….. 15 Hold and AutoHold
Mode………………………………………………………………………………….. 15 Measuring Crest Factor
……………………………………………………………………………………. 16 Capturing Minimum and Maximum Values
………………………………………………………….. 16 Capturing Peak Values
…………………………………………………………………………………….. 18 Low Pass Filter (Model 289
only)……………………………………………………………………….. 20 Making Relative
Measurements…………………………………………………………………………. 21 Making
Measurements……………………………………………………………………………………… 22 Measuring AC Voltage
…………………………………………………………………………………. 22 Using LoZ for Voltage Measurements (Model
289 only) ……………………………………. 23 Making dB Measurements
……………………………………………………………………………. 23 Measuring DC Voltage
…………………………………………………………………………………. 25 Measuring AC and DC Signals
……………………………………………………………………… 26 Measuring
Temperature……………………………………………………………………………….. 28 Using the Y Function (Model 289
Only) ……………………………………………………… 31 Testing for Continuity
…………………………………………………………………………………… 31 Using Conductance for High Resistance
Tests………………………………………………… 34 Measuring Capacitance
……………………………………………………………………………….. 35 Testing Diodes
……………………………………………………………………………………………. 36 Measuring Current
………………………………………………………………………………………. 38
ii
Contents (continued)
Measuring Frequency…………………………………………………………………………………… 42 Measuring Duty cycle
…………………………………………………………………………………… 43 Measuring Pulse
Width…………………………………………………………………………………. 45 Changing Meter Setup Options
………………………………………………………………………….. 47 Resetting Meter Setup
Options………………………………………………………………………. 47 Setting Display Contrast
……………………………………………………………………………….. 47 Setting the Meter’s
Language………………………………………………………………………… 47 Setting Date and
Time………………………………………………………………………………….. 48 Setting Backlight and Auto Off
Timeouts …………………………………………………………. 48 Setting a Custom dBm Reference
………………………………………………………………….. 48 Disabling and Enabling the Beeper
………………………………………………………………… 48 Enabling and Disabling the Smoothing Mode
…………………………………………………… 49 Using Other Setup
Options……………………………………………………………………………. 49 Using Memory
…………………………………………………………………………………………………. 49 Storing Individual Measurement
Data……………………………………………………………… 49 Naming Saved Data
…………………………………………………………………………………….. 49 Viewing Memory Data
………………………………………………………………………………….. 50
Viewing Snapshot and Summary Data ……………………………………………………….. 50 Viewing Trend
Data …………………………………………………………………………………. 50 Zooming in on Trend Data
………………………………………………………………………… 51 Deleting Stored Measurement Data
……………………………………………………………….. 51 Recording Measurement Data
…………………………………………………………………………… 52 Setting up a Recording Session
…………………………………………………………………….. 53 Setting the Event Threshold Value
…………………………………………………………………. 54 Starting a Recording Session
………………………………………………………………………… 54 Stopping a Recording Session
………………………………………………………………………. 54 Using Communications
…………………………………………………………………………………….. 55 Error Messages
……………………………………………………………………………………………….. 56
iii
287/289
Users Manual
Maintenance …………………………………………………………………………………………………… 57 General
Maintenance…………………………………………………………………………………… 57 Testing the
Fuses………………………………………………………………………………………… 57 Replacing the Batteries
………………………………………………………………………………… 59 Replacing the Fuses
……………………………………………………………………………………. 59 Test Lead
Storage……………………………………………………………………………………….. 59
In Case of Difficulty ………………………………………………………………………………………….. 61 Service and
Parts ……………………………………………………………………………………………. 62 General Specifications
……………………………………………………………………………………… 66 Detailed
Specifications……………………………………………………………………………………… 67
AC Voltage Specifications…………………………………………………………………………….. 68 AC Current
Specifications …………………………………………………………………………….. 69 DC Voltage Specification
……………………………………………………………………………… 70 DC Current
Specifications…………………………………………………………………………….. 71 Resistance Specifications
…………………………………………………………………………….. 72 Temperature Specifications
………………………………………………………………………….. 72 Capacitance and Diode Test
Specifications…………………………………………………….. 73 Frequency Counter Specifications
…………………………………………………………………. 74 Frequency Counter Sensitivity
………………………………………………………………………. 75 MIN MAX, Recording, and Peak Specifications
……………………………………………….. 76 Input Characteristics
……………………………………………………………………………………. 77 Burden Voltage (A, mA,
µA)………………………………………………………………………….. 78
iv
List of Tables
Table
Title
Page
1. Symbols………………………………………………………………………………………………………………… 4 2. Push Buttons …………………………………………………………………………………………………………. 5 3. Display Features ……………………………………………………………………………………………………. 7 4. Rotary Switch Positions…………………………………………………………………………………………… 10 5. Input Terminals………………………………………………………………………………………………………. 11 6. Battery Level Indicator…………………………………………………………………………………………….. 12 7. Trend Data Display…………………………………………………………………………………………………. 51 8. Recording Display ………………………………………………………………………………………………….. 53 9. Stopped Recording Display ……………………………………………………………………………………… 55 10. Error Messages ……………………………………………………………………………………………………… 56 11. Replacement Parts …………………………………………………………………………………………………. 62 12. Accessories …………………………………………………………………………………………………………… 65
v
287/289
Users Manual
vi
List of Figures
Figure
Title
Page
1. Push Buttons …………………………………………………………………………………………………………. 5 2. Display Features ……………………………………………………………………………………………………. 7 3. Rotary Switch ………………………………………………………………………………………………………… 10 4. Input Terminals………………………………………………………………………………………………………. 11 5. Function Menu……………………………………………………………………………………………………….. 14 6. MIN MAX Record Display………………………………………………………………………………………… 17 7. Peak Record Display ………………………………………………………………………………………………. 18 8. Low Pass Filter ………………………………………………………………………………………………………. 20 9. Relative Mode Functions …………………………………………………………………………………………. 21 10. AC Voltage Measurements………………………………………………………………………………………. 22 11. dBm Display ………………………………………………………………………………………………………….. 23 12. DC Voltage Measurements ……………………………………………………………………………………… 25 13. AC and DC Display…………………………………………………………………………………………………. 26 14. Temperature Measurement ……………………………………………………………………………………… 28 15. Resistance Measurement………………………………………………………………………………………… 30 16. Continuity Indicator…………………………………………………………………………………………………. 31 17. Continuity Testing…………………………………………………………………………………………………… 32
vii
287/289
Users Manual
18. Conductance Measurement …………………………………………………………………………………….. 34 19. Capacitance Measurement ……………………………………………………………………………………… 35 20. Diode Testing………………………………………………………………………………………………………… 37 21. Current Measurement Setup……………………………………………………………………………………. 40 22. Current Measurement Circuit Connection ………………………………………………………………….. 41 23. Functions Allowing Frequency Measurement …………………………………………………………….. 42 24. Frequency Display …………………………………………………………………………………………………. 43 25. Duty Cycle Measurements ………………………………………………………………………………………. 44 26. Duty Cycle Display …………………………………………………………………………………………………. 45 27. Pulse Width Measurements …………………………………………………………………………………….. 46 28. Testing the Current Fuses……………………………………………………………………………………….. 58 29. Test Lead Storage …………………………………………………………………………………………………. 59 30. Replacing Batteries and Fuses ………………………………………………………………………………… 60 31. Replaceable Parts………………………………………………………………………………………………….. 64
viii
Introduction
XWWarning
Read “Safety Information” before using this Meter.
The descriptions and instructions in this manual apply to the model 289 and
model 287 True-rms Digital Multimeters (hereafter referred to as the Meter).
The model 289 appears in all illustrations.
Contacting Fluke
To contact Fluke, call: USA:
1-888-993-5853 Canada :
1-800-363-5853 Europe : +31 402-675-200
Japan: +81-3-3434-0181 Singapore : +65-738-5655 Anywhere in the world:
+1-425-446-5500
Visit Fluke’s web site at: www.fluke.com
Register your Meter at: http://register.fluke.com
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Safety Information
The Meter complies with:
· ANSI/ISA 82.02.01 (61010-1) 2004
· UL 61010B (2003)
· CAN/CSA-C22.2 No. 61010-1-04 · IEC/EN 61010-1 2nd Edition Pollution Degree 2
· EMC EN 61326-1
· Measurement Category III, 1000V, Pollution Degree 2
· Measurement Category IV, 600 V, Pollution Degree 2
In this manual, a Warning identifies hazardous conditions and actions that
could cause bodily harm or death. A Caution identifies conditions and actions
that could damage the Meter, the equipment under test, or cause permanent loss
of data.
XWWarning
To avoid possible electric shock or personal injury, follow these guidelines:
· Use this Meter only as specified in this manual or the protection provided
by the Meter might be impaired.
· Do not use the Meter if it is damaged. Before you use the Meter, inspect the
case. Look for cracks or missing plastic. Pay particular attention to the
insulation surrounding the connectors.
1
287/289 Users Manual
· Make sure the battery door is closed and latched before operating the Meter.
· Remove test leads from the Meter before opening the battery door.
· Inspect the test leads for damaged insulation or exposed metal. Check the
test leads for continuity. Replace damaged test leads before you use the
Meter.
· Do not apply more than the rated voltage, as marked on the Meter, between
the terminals or between any terminal and earth ground.
· Never operate the Meter with the cover removed or the case open.
· Use caution when working with voltages above 30 V ac rms, 42 V ac peak, or
60 V dc. These voltages pose a shock hazard.
· Use only the replacement fuses specified by the manual.
· Use the proper terminals, function, and range for measurements.
· Avoid working alone. · When measuring current, turn off circuit power
before connecting the Meter in the circuit. Remember to place the Meter in
series with the circuit.
2
· When making electrical connections, connect the common test lead before
connecting the live test lead; when disconnecting, disconnect the live test
lead before disconnecting the common test lead.
· Do not use the Meter if it operates abnormally. Protection may be impaired.
When in doubt, have the Meter serviced.
· Do not operate the Meter around explosive gas, vapor, or dust.
· Use only 1.5 V AA batteries, properly installed in the Meter case, to power
the Meter.
· When servicing the Meter, use only specified replacement parts.
· When using probes, keep fingers behind the finger guards on the probes.
· Do not use the Low Pass Filter option to verify the presence of hazardous
voltages. Voltages greater than what is indicated may be present. First, make
a voltage measurement without the filter to detect the possible presence of
hazardous voltage. Then select the filter function.
· Only use test leads that have the same voltage, category, and amperage
ratings as the meter and that have been approved by a safety agency.
· Use proper protective equipment, as required by local or national
authorities when working in hazardous areas.
· Comply with local and national safety requirements when working in hazardous
locations.
WCaution
To avoid possible damage to the Meter or to the equipment under test, follow
these guidelines:
· Disconnect circuit power and discharge all high-voltage capacitors before
testing resistance, continuity, diodes, or capacitance.
· Use the proper terminals, function, and range for all measurements.
True-rms Digital Multimeters Hazardous Voltage
· Do not remove batteries while the Meter is turned on or a signal is applied
to the Meter’s input jacks.
· Before measuring current, check the Meter’s fuses. (See “Testing the Fuses”
in the Users Manual on the accompanying CD.)
· Do not use the LoZ mode to measure voltages in circuits that could be
damaged by this mode’s low impedance (3 k). (Model 289 only)
Hazardous Voltage
To alert you to the presence of a potentially hazardous voltage, when the
Meter detects a voltage 30 V or a voltage overload
(OL), the Ysymbol is displayed.
3
287/289 Users Manual
Symbols
Table 1 lists and describes the symbols used on the Meter and in this manual.
Table 1. Symbols
Symbol B F
X
E
R P
CAT III
~
Description
Symbol
Description
AC (Alternating Current or Voltage) DC (Direct Current or Voltage) Hazardous voltage
I
Fuse
T
Double Insulated
W
Important Information; refer to manual
Battery (Low battery when shown on the display)
J
Earth ground
Continuity test or continuity beeper tone Conforms to European Union directives
$
;
N10140
Conforms to relevant Canadian and US standards
Conforms to relevant Australian standards
Underwriters Laboratory listed product
s
Inspected and licensed by TÜV Product Services
IEC Measurement Category III CAT III equipment is designed to protect against transients in equipment in fixed equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.
CAT IV
IEC Measurement Category IV CAT IV equipment is designed to protect against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service.
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
4
Features
Tables 2 through 5 briefly describe the Meter’s features.
Understanding the Push Buttons
The 14 push buttons on the front of the Meter activate features that augment
the function selected using the rotary switch, navigate menus or control power
to Meter circuits. The buttons shown in Figure 1 are described in Table 2.
Figure 1. Push Buttons
est02.emf
True-rms Digital Multimeters Features
Table 2. Push Buttons
Button
O
Function Turns the Meter on or off.
12 Selects sub-functions and modes related 34 to the rotary switch function.
Cursor buttons select an item in a menu, adjust display contrast, scroll through information, and perform data entry.
H
Freezes the present reading in the display and allows the display to be saved. Also accesses AutoHold.
R M
Switches the Meter range mode to manual and then cycles through all ranges. To
return to autoranging, press the button for 1 second.
Starts and stops MIN MAX recording.
I
Displays information about the present function or items on the display at the moment the info button is pressed.
G
Switches the display backlight between off, low, and high.
5
287/289 Users Manual
Using Auto Repeat
For some menu selections, holding down a softkey or cursor button will
continuously change (or advance) a selection until the button is released.
Normally, each press of a button causes the selection to change once. During
some selections, the selections will change faster if the button is held for
two or more seconds. This is helpful when scrolling through a list of
selections, such as a list of stored measurements.
6
Understanding the Display
Display features shown in Figure 2 are described in Table 3 and the following
sections.
7 8 9 10
11 12
6
8:10pm
123.45 VAC
06/13/07
13
5
REL
HOLD AutoSAVE
4
123.45 mVAC
14
Crest Factor
3
5 mVDC Auto Range
0
100
200
300
400
500 VAC
15
2
60.000 Hz 16
AutoHOLD Save
1
Setup
Figure 2. Display Features
est01.eps
True-rms Digital Multimeters Features
Table 3. Display Features
Item Function
A Softkey labels
B Bar graph
C Relative
D Minus sign E Lightning bolt
F Remote communication
G Battery level
H Time
I Mode annunciators
Indication
Indicates the function of the button just below the displayed label.
Analog display of the input signal (See the “Bar Graph” section for more
information).
Indicates the displayed value is relative to a reference value.
Indicates a negative reading.
Indicates hazardous voltage present at the Meter’s input.
Indicates activity over the communication link.
Indicates the charge level of the six AA batteries.
Indicates the time set in the internal clock.
Indicates the Meter’s mode.
7
287/289 Users Manual
Table 3. Display Features (cont.)
Item Function
Indication
J Minimeasurement
Displays the lightning bolt (when necessary) and the input value when the primary and secondary displays are covered by a menu or pop-up message.
K Date
Indicates the date set in the internal clock.
L Beeper
Indicates the Meter’s beeper is enabled (not associated with the continuity beeper).
M Units
Indicates the units of measure.
N Auxiliary Units
Indicates unitless measurements like Crest Factor.
O Range indicator
Indicates the range the Meter is in and the ranging mode (auto or manual).
P Secondary display Displays secondary measurement information about the input signal.
8
Bar Graph
The analog bar graph functions like the needle on an analog meter, but without
the overshoot. The bar graph updates 30 times per second. Because the graph
updates faster than the digital display, it is useful for making peak and null
adjustments and observing rapidly changing inputs. For frequency, duty cycle,
pulse width, dBm, and crest factor functions, the bar graph represents the
amplitude of the input signal (volts or amps) and not the value in the primary
display. The bar graph is not shown for capacitance, temperature, LoZ, AC+DC,
AC over DC, peak, or min max functions.
For dc voltage, dc current, and all relative percent modes, a zero-centered
bar graph is displayed. For dc voltage and current, the bar graph range is the
maximum of the selected range. For relative percent mode, the bar graph goes
to ±10 %.
The number of lit segments indicates the measured value and is relative to the
full-scale value of the selected range. In the 50 Vac range, for example, the
major divisions on the scale represent 0, 5, 10, 15, 20, 25, 30, 35, 40, 45,
and 50 Vac. An input of 25 Vac turns on segments up to the middle of the
scale.
For off-scale values, f appears to the right of the normal bar graph. For the
zero-center bar graph, a e appears at the left end of the bar graph for
negative off-scale values and fappears on the right end for positive off-scale
values.
Status Bar Elements
The status bar at the top of the Meter’s display contains indicators for
battery level, time of day, mini-measurement display, present date, and beeper
on/off icon.
The mini-measurement display shows the measurement value of the primary
function, if it is not already shown in the page area of the display. For
example, when the display is frozen for a HOLD, the mini-measurement display
continues to show the input signal (Live) measurement and a mini z. In
addition, the minimeasurement display will flash, if z (for inputs above 30
volts) would normally appear in the primary display but is obscured. To warn
of the possibility of blowing a current fuse, the minimeasurement display also
flashes when current measurements exceed the maximum continuous current levels
(see specifications).
Page Area
The page area of the display is where the main meter content is displayed. The
primary display (upper half of the page area) is where the most important
value of the selected function is shown. The secondary display contains the
bar graph and values that may be measured in addition to the primary function
value. For example, with frequency measurement selected in Vac, the frequency
value will appear in the primary display with the ac voltage value in the
secondary display.
Softkey Labels
Labels for the four function softkeys (F1 through F4) appear in the bottom row
of the display. These labels will change based on the function and/or menu
selection.
Adjusting Display Contrast
When not selecting items on a menu or inputting data, pressing 7 increases
display contrast and pressing 8 decreases it.
True-rms Digital Multimeters Features
9
287/289 Users Manual
Understanding the Rotary Switch
Select a primary measurement function by positioning the rotary switch to one
of the icons around its perimeter. For each function, the Meter presents a
standard display for that function (range, measurement units, and modifiers).
Button choices made in one function do not carry over into another function.
The model 289 offers two additional functions: low ohms (Y) and low impedance
(LoZ) ac volts. Each position shown in Figure 3 is described in Table 4.
Figure 3. Rotary Switch
est03.emf
10
Table 4. Rotary Switch Positions
Switch Position
Function
L
AC voltage measurement using a low input impedance (model 289 only)
V
AC voltage measurements
T AC millivolt measurements
U
DC and ac+dc voltage measurements
N
DC millivolts, ac+dc millivolt, and temperature measurements
S
Resistance, continuity, and conductance measurements
P
Diode test and capacitance measurements
A
X Y
AC, dc and ac+dc amps and milliamps measurements
AC, dc and ac+dc microampere measurements up to 5,000 µA
Resistance measurements with 50e range (model 289 only)
Using the Input Terminals
All functions except current use the W and COM input terminals. The two
current input terminals (A and mA/µA) are used as follows:
Current from 0 to 400 mA, use the and COM terminals.
Current between 0 and 10 A use the A and COM terminals.
True-rms Digital Multimeters Features
Table 5. Input Terminals
Terminal
Description
Input for 0 A to 10.00 A current (20VA
A
overload for 30 seconds on, 10 minutes off),
frequency, and duty-cycle measurements.
Input for 0 A to 400 mA current measurements, frequency, and duty cycle.
COM Return terminal for all measurements.
Input for voltage, continuity, resistance, diode test, conductance, capacitance, frequency, temperature, period, and dutycycle measurements.
Figure 4. Input Terminals
est04.emf
11
287/289 Users Manual
Controlling Meter Power
The Meter is powered by six AA batteries and controlled through a front panel
power switch and internal circuits designed to help conserve battery power.
The following sections describe several techniques for controlling Meter
power.
Powering the Meter On and Off Manually
With the Meter off, press O to turn on the Meter. Pressing O while the Meter
is on causes it to turn off.
Note Collected data is retained when the Meter is turned off while in record,
MIN MAX record, or Peak record modes. When the Meter is next turned on, the
dispay shows the collected data in stopped mode. Pressing the softkey labeled
Save will save the data.
Battery Level Indicator
The battery level indicator in the upper left-hand corner of the display
indicates the relative condition of the batteries. Table 6 describes the
various battery levels the indicator represents.
12
Table 6. Battery Level Indicator
Indication
Battery Capacity
B C D E F[1]
Full capacity ¾ capacity ½ capacity ¼ capacity Almost empty (less than one day)
[1] When critically low, a “Replace batteries” pop-up message appears 15 seconds before the Meter shuts down.
The Meter will display a “Batteries low” message whenever the battery level will not support a selected function.
Automatic Power-Off
The Meter automatically turns off if the rotary switch is not moved or a
button is not pressed for 15 minutes (default). Pressing O will turn the Meter
back on after it is powered off automatically. To change the timeout period or
completely disable automatic power-off, refer to “Setting Backlight and Auto
Off Timeouts” later in this manual.
Battery Saver Mode
If Auto off is enabled (set to a time period), and MIN MAX record, Peak
record, Recording, or AutoHold is enabled, the Meter will enter a battery-
saver mode if a push-button is not pressed or the rotary switch is not moved
for a set period of time. For the
recording mode, the time period is five minutes. For MIN MAX, Peak and
AutoHold modes, the time period is the same time the Auto Off feature is set
for. See the “Setting Backlight and Auto Off Timeouts” section later in this
manual. Battery saver mode conserves battery power by shutting down circuits
not necessary for the selected function, including the display. However, the
LED surrounding the power button (O) will continue to flash to indicate the
Meter is still collecting data.
The Meter “wakes up” from battery-save mode under the following conditions:
· A button is pressed
· The rotary knob is moved
· A lead is removed or inserted into a current input jack.
· The Meter changes range
· IR Communications begin
These condictions only awaken the Meter and does not change the Meter’s
function or mode of operation.
Controlling the Backlight
If viewing the display becomes difficult in low-light conditions, press G to
activate the LCD backlight. The backlight button cycles the backlight through
three states: low, high, and off. The Meter displays a message if the battery
level will not support the backlight operation.
To conserve battery life, a user-adjustable timeout controls how long the
backlight stays on. The default timeout is 5 minutes. To change the timeout,
refer to “Setting Backlight and Auto Off Timeouts” later in this manual.
True-rms Digital Multimeters Selecting the Range
Selecting the Range
The Meter’s selected range is always displayed above the righthand end of the
bar graph, as the range indicator. Pressing R switches the Meter between
manual and autoranging. It also cycles through the Meter ranges when manual
ranging is enabled.
Note You cannot use R in conductance, diode test, LoZ, Low Ohms, and
temperature functions. These functions all use a fixed range. In autorange,
the Meter selects the lowest range to display the highest available precision
(resolution) for the input signal. If Manual range is already enabled, press
and hold R for one second to enter the autoranging mode. If autorange is
enabled, press R to enter manual ranging. Each additional press of R sets the
Meter to the next higher range, unless it is already in the highest range, at
which point the range switches to the lowest range.
Understanding Function Menus
Each primary measurement function (rotary switch position) has a number of
optional sub-functions or modes accessed by pressing the softkey labeled Menu
(F1). A typical menu is shown in Figure 5.
13
287/289 Users Manual
VAC REL Peak, CF
Menu
Hz, %, mS dBm dBV
REL
REL%
Close
Figure 5. Function Menu
est05.eps
Menu selection is indicated by the filled-in black square (hereafter the menu
selector) to the left of a menu item. Use the four front-panel cursor buttons
(5 6 7 8) to position the menu selector next to a menu item. As the menu
selector moves between menu items, the four softkeys and their labels change
to reflect the available functions and/or modes available for the selection
menu item.
The example menu in Figure 5 shows the REL (Relative) function as the current
selection. The function selected when the menu opens is the function selected
the last time the menu was used. To get to the Hz menu item from the REL item,
press 6 once, followed by one press of 7. As the menu selector moves between
the menu items, the softkey labels change to indicate each softkey’s function.
Once the desired function or mode appears in one of the softkey labels, press
the appropriate softkey to activate it. The pop-up menu closes and the display
changes to reflect the selection just made. Pressing the softkey labeled Close
closes the pop-up menu, leaving the Meter in the state it was in before
pressing the Menu softkey.
14
In most cases, the softkeys revealed by the menu selection act like toggles.
The example menu shown in Figure 5 shows REL, REL%, and Close softkeys. In
this example, the Meter is not in the relative mode, so pressing the softkey
labeled REL would activate, or toggle, the relative mode. If, on the other
hand the Meter is already in the relative mode, pressing the same softkey
would disable the relative function.
In some cases, pressing a function that can not be used with other functions
appearing in the menu turns off the previously selected function. For example,
in Figure 5, if the Meter is already in the relative function, pressing REL%
causes the Meter to turn off relative and display relative percent.
In cases where multiple modes have been selected, selecting the first (top-
left) menu item always turns off all other functions and modes, and returns
the Meter to the primary function selected by the rotary switch. For example,
assume that the Meter is setup for frequency (Hz) and is displaying in
relative mode as selected through the menu in Figure 5. Moving the menu
selector to the menu item labeled VAC and pressing the softkey labeled VAC,
clears both frequency and relative selections, leaving the Meter in volts ac
only.
Menu selections are remembered for each rotary switch position. For example,
selecting REL for the volts ac position causes REL to be selected the next
time the menu is opened in volts ac, even though in the interim, Hz,%,ms was
selected from a similar menu for the millivolts ac function.
Up to two columns of four items each are displayed at any one time. If more
than eight menu items are available for a primary function, g appears in the
lower right-hand corner of the page area of the display, indicating more menu
items are available. With the menu selector on one of the items in the left
column,
press 5 to scroll the screen horizontally and reveal the off-screen menu
items. Conversely, with the menu selector on an item in the right-hand column,
press 6 to reveal the off-screen menu items.
Input AlertTM Feature
XW Warning
To avoid circuit damage and possibly blowing the Meter’s current fuse, do not
place the probes across (in parallel with) a powered circuit when a lead is
plugged into a current terminal. This causes a short circuit because the
resistance through the Meter’s current terminals is very low.
If a test lead is plugged into the mA/µA or A terminal, but the rotary switch
is not set to the correct current position, the beeper warns you by making a
chirping sound and displays “Leads connected incorrectly”. This warning is
intended to stop you from attempting to measure voltage, continuity,
resistance, capacitance, or diode values when the leads are plugged into a
current terminal.
Using the Info Button
While operating the Meter, more information about a selected function, a
front-panel button, or a menu item may be necessary. Press I to open an
information window that lists topics covering the functions and modifiers that
are available at the time the button is pressed. Each topic provides a brief
explanation on a Meter function or feature.
True-rms Digital Multimeters Input AlertTM Feature
The information revealed through I is not meant to replace the more detailed
information found in this manual. Function and feature explanations are brief
and only meant to refresh a person’s memory.
The number of information topics displayed at any one time may exceed the
display area. Use the softkeys labeled Next and Prev to move from topic to
topic. Use the softkey labeled More or 7 and 8 to scroll through the
information a full screen at a time.
Pressing the softkey labeled Close, or I will close the information window.
Hold and AutoHold Mode
To freeze the display for any function, press H. Only the minimeasurement
display and hazardous voltage icon (z) continue to indicate the actual input.
The battery level indicator is also active. The Meter’s softkeys are relabeled
for saving the frozen reading or activating the AutoHold mode.
If H is pressed while MIN MAX record, peak record, or a recording session is
in progress, the display freezes but the data acquisition continues in the
background. Pressing H again updates the display to reflect data that was
acquired during the hold.
Pressing the softkey labeled AutoHOLD activates AutoHold if the Meter is not
in the Peak, MIN MAX, or Record modes. AutoHold operation monitors the input
signal and updates the display and, if enabled, sounds the beeper, whenever a
new stable measurement is detected. A stable measurement is one that does not
vary more than a selected adjustable percentage (AutoHold threshold) for at
least one second. The Meter filters
15
287/289 Users Manual
out open lead conditions so the Meter leads can be moved between test points
without triggering a display update.
Note
For temperature measurements, the AutoHold threshold is a percent of 100
degrees. The default AutoHold threshold is 4% of 100 degrees, or 4 degrees
Celsius or Fahrenheit.
Pressing H while in AutoHold mode, forces the Meter’s display to update with
the present measurement, just as if a stable measurement had been detected.
To set the AutoHOLD Threshold Value, press the softkey labeled Setup to access
the setup menu. Using the cursor buttons, move the menu selector next to the
menu item labeled Recording and press the softkey labeled Recording to open
the recording setup screen. Using the cursor buttons, move the menu selector
next to the menu item labeled Event Threshold for AutoHOLD and then press the
softkey labeled Edit. Press 7 or 8 to scroll through the AutoHold threshold
values. With the desired value selected, press the softkey labeled Close.
Measuring Crest Factor
Crest factor is a measure of signal distortion and is calculated as a signal’s
peak value over its rms value. This is an important measurement when looking
at power quality issues.
The Meter’s crest factor function is only available for the ac measurements:
Vac, mVac, Aac, mAac, and µAac. With the Meter in one of the ac measurement
functions, press the softkey labeled Menu. Next, move the menu selector next
to the menu item labeled Peak,CF and press the softkey labeled CF. The
16
crest factor value is displayed in the primary display while the ac
measurement appears in the secondary display. Frequency, duty cycle, and pulse
width are not allowed during crest factor measurements.
Capturing Minimum and Maximum Values
The MIN MAX Record mode captures minimum, average, and maximum input values.
When the input goes below the recorded minimum value or above the recorded
maximum value, the Meter beeps and records the new value. The Meter stores the
elapsed time since the recording session was started at the same time. The MIN
MAX mode also calculates an average of all readings taken since the MIN MAX
mode was activated.
This mode is for capturing intermittent readings, recording minimum and
maximum readings unattended, or recording readings while equipment operation
precludes watching the Meter. The MIN MAX mode is best for recording power
supply surges, inrush currents, and finding intermittent failures.
Response time is the length of time an input must stay at a new value to be
captured as a possible new minimum or maximum value. The Meter has a 100
millisecond MIN MAX response time. For example, a surge lasting 100
milliseconds would be captured but one lasting only 50 milliseconds may not be
captured at its actual peak value. See the MIN MAX specification for more
information.
The true average value displayed is the arithmetic mean of all readings taken
since the start of recording (overloads are discarded). The average reading is
useful for smoothing out unstable inputs, calculating power consumption, or
estimating the percentage of time a circuit is active.
Note
For input signals that are noisy or change rapidly, turn on the Smoothing mode
to display a steadier reading. See the “Enabling and Disabling the Smoothing
Mode” section later in this manual.
To extend battery life during MIN MAX recording, the Meter will enter a
battery saver mode. See the “Setting Backlight and Auto Off Timeouts” section
for more information on the battery saver mode.
To activate the MIN MAX mode, press M. As shown in Figure 6, the Meter
displays e at the top of the measurement page, and the MIN MAX start date and
time along the bottom of the page. In addition, the recorded maximum, average,
and minimum values appear in the secondary display with their respective
elapsed times.
True-rms Digital Multimeters Capturing Minimum and Maximum Values
8:10pm
06/07/07
Min Max
119.81 VAC
Maximum
Auto Range
127.09 VAC
500 VAC
00:03:17
Average
119.50 VAC 01:10:09
110.23 Minimum
VAC 00:59:59
Start : 06/07/07 7:00 pm
Restart
Stop
Figure 6. MIN MAX Record Display
est42.eps
To stop a MIN MAX recording session, press M or the softkey labeled Stop. The summary information in the display freezes, and the softkeys change function to allow saving the collected data. Pressing M again or the softkey labeled Close exits the MIN MAX record session without saving the collected data.
Note
Turning the rotary switch before saving the MIN MAX recording data will cause
all the accumulated data to be lost.
To save the MIN MAX screen data, the MIN MAX session must be ended by pressing
the softkey labeled Stop. Next, press the softkey labeled Save. A dialog box
opens where the default saved name can be selected or another name assigned.
Press
17
287/289 Users Manual
the softkey labeled Save to store the MIN MAX screen data. MIN MAX can not be
continued at this point. Press the softkey labeled Close to exit the MIN MAX
mode. Pressing the softkey labeled Restart while MIN MAX is running stops the
MIN MAX session, discards all MIN MAX data, and immediately starts a new MIN
MAX recording session.
Capturing Peak Values
Peak record is almost the same as MIN MAX record explained earlier in this
manual. The significant difference between the two recording functions is the
shorter response time for peak recording: 250 µs. With this short response
time, the actual peak values of a sinusoidal signal are measurable. Transients
are more accurately measured using the peak record feature. To activate the
peak mode, press the softkey labeled Menu. Move the menu selector next to the
menu item labeled Peak,CF or Peak. Press the softkey labeled Peak to start the
peak recording session.
18
8:10pm
06/07/07
Peak
119.8 VAC
Auto Range
Peak Max 168.2 V
500 VAC 00:03:17
118.9 Average
VAC
Peak Min -173.9 V
01:10:59 01:10:09
Start : 06/07/07 7:00 pm
Restart
Stop
Figure 7. Peak Record Display
est43.eps
As shown in Figure 7, the primary display shows the “live” measurement present
on the Meter’s inputs. In the secondary area of the display, the maximum and
minimum peak values as well as the average value are shown along with their
respective time stamps. The time stamp next to the average value indicates the
elapsed time of the peak recording session. The peak recording session start
time is shown along the bottom of the page area of the display.
When the peak value of the input signal goes below the recorded minimum value
or above the recorded maximum value, the Meter beeps and records the new
value. At the same time, the elapsed time since the peak recording session was
started is stored as the recorded value’s time stamp.
Pressing the softkey labeled Stop ends the peak recording session. The summary
information in the display freezes and the softkeys change function to allow
saving the collected data. Pressing the softkey labeled Close exits the peak
recording session without saving the collected data.
Note
Turning the rotary switch before saving the peak recording data will cause all
the accumulated data to be lost.
To save peak screen data, the peak capture session must be ended by pressing
the softkey labeled Stop. Next press the softkey labeled Save. A dialog box
opens where the default saved name can be selected or another name assigned.
Press the softkey labeled Save to store the Peak screen data. Peak capture can
not be continued at this point. Press the softkey labeled Close to exit the
Peak capture mode.
Pressing the softkey labeled Restart while the peak recording session is
running stops the session, discards all peak recorded data, and immediately
starts a new peak record session.
When viewing saved records, snapshot peak records look the same as a stopped
peak records. Therefore, use the elapsed time (average value time stamp) to
identify one record from another.
To extend battery life during peak record, the Meter enters a battery-saver
mode after a period of time set for the Auto Off feature. See the “Setting
Backlight and Auto Off Timeouts” section for more information on the battery
saver mode.
True-rms Digital Multimeters Capturing Peak Values
19
287/289 Users Manual
Low Pass Filter (Model 289 only)
The Meter is equipped with an ac low pass filter. When measuring ac voltage,
or Vac frequency, press the softkey labeled Menu to open the function menu,
and move the menu selector to the l item. Next, press the softkey labeled l to
toggle the low pass filter mode on (l displayed) and off.
XWWarning
To avoid possible electric shock or personal injury, do not use the Low Pass
Filter option to verify the presence of hazardous voltages. Voltages greater
than what is indicated may be present. First, make a voltage measurement
without the filter to detect the possible presence of hazardous voltage. Then
select the filter function.
The Meter continues measuring in the chosen ac mode, but now the signal passes
through a filter that blocks unwanted voltages above 1 kHz, as shown in Figure
8. The low pass filter can improve measurement performance on composite sine
waves that are typically generated by inverters and variable frequency motor
drives.
Note In Low Pass Mode, the Meter goes to manual mode. Select ranges by
pressing R. Autoranging is not available when the Low Pass Filter is enabled.
20
1 kHz 100 Hz
Figure 8. Low Pass Filter
aom11f.eps
Making Relative Measurements
The Meter displays calculated values that are based on a stored value when set
to relative and relative percent mode. Figure 9 shows the functions for which
the two relative modes are available. In addition, the two relative modes are
available in frequency, duty cycle, pulse width, crest factor, and dB.
Figure 9. Relative Mode Functions
est29.eps
To activate the relative or relative percent modes while in one of the functions shown in Figure 9, press the softkey labeled Menu. Move the menu selector to the menu item labeled REL. Next, press either the softkey labeled REL or REL%. The measurement value at the time that either Rel or Rel % is enabled, is stored as the reference value and displayed in the
True-rms Digital Multimeters Making Relative Measurements
secondary display. The present or “Live” measurement moves to the secondary
display and the primary display indicates the difference between the present
measurement and the reference value in measurement units for REL and as a
percentage for REL %. When relative percent is enabled, the bar graph is a
zerocentered bar graph that indicates the percentage difference. The bar
graph’s range is limited to ±10 %, but the display goes to ±999.9 %. At 1000 %
or more, the display indicates OL. When the reference value is 0, the Meter
displays OL. With the exception of dB measurements, ranging is set to manual
and can not be changed. Both auto and manual ranging is possible when making
relative dB measurements. When relative is enabled during dBm or dBV
measurements, the displayed units change to dB. In relative or relative
percent mode, the softkey label for F3 indicates REL or REL%, depending on
which of the two modes is not presently selected. The F3 button acts as a
toggle, switching the Meter between the two modes. Moving the rotary switch
between V and mV while in relative dBm or dBv mode does not disable the dB
measurement. This allows continuous measurements over a wide range of input
voltage.
21
287/289 Users Manual
Making Measurements
The following sections describe how to take measurements with the Meter.
Measuring AC Voltage
The Meter displays ac voltage measurements as rms (root mean square) readings.
The rms value is the equivalent dc voltage that would produce the same amount
of heat in a resistance as the measured voltage. True-rms readings are
accurate for sine waves and other wave forms (with no dc offset) such as
square waves, triangle waves, and staircase waves. For ac with dc offset,
refer to the “Measuring AC and DC Signals” section later in this manual.
Rotate the Meter’s rotary switch to V or T and set up the Meter to measure ac
volts as shown in Figure 10. The Meter’s ac volts function offers a number of
modes to provide more details about an ac signal. Pressing the softkey labeled
Menu opens a menu of items that can be used to modify the basic ac voltage
measurement. Refer to the appropriate section in this manual to learn more
about each menu item. To clear all modes and return to the basic volts ac
measurement, press the softkey labeled Menu. Move the menu selector to the
item labeled VAC. Press the softkey labeled VAC to clear all functions and
modes.
22
08:10pm
03 / 13/ 06
123.45VAC
Auto Range
0
100
200
300
400
500 VAC
Menu
Save
Setup
Switch Box
est07.eps
Figure 10. AC Voltage Measurements
Using LoZ for Voltage Measurements (Model 289 only)
W Caution
Do not use the LoZ mode to measure voltages in circuits that could be damaged
by this mode’s low impedance (3 k).
To eliminate ghost voltages, the Meter’s LoZ function presents a low impedance
across the leads to obtain a more accurate measurement.
To make a LoZ measurement, set the rotary switch to L. The Meter displays the
ac voltage in the primary display and the dc voltage in the secondary display.
During LoZ measurements, the Meter’s range is set to 1000 volts in the manual
ranging mode.
In LoZ, both R and M are disabled. There are no additional modes for this
function and the softkey labeled Menu is therefore disabled as well.
Making dB Measurements
The Meter is capable of displaying voltage as a dB value, either relative to 1
milliwatt (dBm), a reference voltage of 1 volt (dBV) or a user-selectable
reference value. See the “Setting a Custom dBm Reference” section later in
this manual.
8:10pm
True-rms Digital Multimeters Making Measurements
06/13/07
41.83 dBm
Auto Range
0
100
200
300
400
500 VAC
123.45 VAC
1000 Reference
Menu
Save
Ref
Figure 11. dBm Display
Setup
est08.eps
To set the Meter to display values in dBm, set the rotary switch to V or T and
press the softkey labeled Menu. Move the menu selector to the menu item
labeled dBm. Press the softkey labeled dBm. The dBm, Hz menu selection
replaces the secondary display (123.45 VAC in Figure 11) with the frequency
measurement. All voltage measurements are displayed as a dBm value, as shown
in Figure 11.
A dBm measurement must use a reference impedance (resistance) to calculate a
dB value based on 1 milliwatt. When set to 600 (default), the reference
impedance is not displayed during a dBm measurement. When set to something
other than 600 , the reference impedance is displayed just above the softkey
labels.
23
287/289 Users Manual
To select another reference value, press the softkey labeled Ref to display a
message box with the current reference value. Pressing 7 or 8, scrolls through
the nine predefined references: 4, 8, 16, 25, 32, 50, 75, 600, and 1000. Set
the reference by pressing the softkey labeled OK. To add a custom reference
impedance, see the “Setting a Custom dBm Reference” section later in this
manual. A dBV measurement uses a 1 volt reference voltage to compare the
present measurement against. The difference between the two ac signals is
displayed as a dBV value. The reference impedance setting is not part of a dBV
measurement. To make a dBV measurement, position the rotary switch to V or T
and place the Meter leads on the voltage to be measured. Next, press the
softkey labeled Menu. Move the menu selector to the menu item labeled dBV and
press the softkey labeled dBV. The Meter displays the voltage in dBV. To exit
the dBV or dBm function, press the softkey labeled Menu followed by the
softkey labeled dBV or dBm, respectively. Selecting one of the other modifiers
such as ms, %, or CF also cancels dBV or dBm.
24
Measuring DC Voltage
The Meter displays dc volts values as well as their polarity. The bar graph
for dc voltage measurements is a zero-centered bar graph. Positive dc voltages
cause the bar graph to fill to the right of center while negative dc voltages
fill left of center.
To measure a dc voltage with the Meter, rotate the rotary switch to the U or N
position as shown in Figure 12.
The Meter’s dc volts function offers a number of modes to provide more details
about a dc signal. Pressing the softkey labeled Menu opens a menu of items
that can be used to modify the basic dc voltage measurement. Refer to the
appropriate section in this manual to learn more about each menu item.
To clear all modes and return to the basic volts dc measurement, press the
softkey labeled Menu. Move the menu selector to the item labeled VDC. Press
the softkey labeled VDC to clear all functions and modes.
True-rms Digital Multimeters Making Measurements
8:10pm
06/13/07
9.752 VDC Auto Range -50 -40 -30 -20 -10 0 10 20 30 40 50 VDC
Menu
Save
Setup
V
est09.eps
Figure 12. DC Voltage Measurements
25
287/289 Users Manual
Measuring AC and DC Signals
The Meter is capable of displaying both ac and dc signal components (voltage
or current) as two separate readings or one AC+DC (rms) value combined. As
shown in Figure 13, the Meter displays ac and dc combinations three ways: ac
displayed over dc (AC,DC), dc displayed over ac (DC,AC), and ac combined with
dc (AC+DC). Select one of these three displays using the Function and Mode
menu.
With the rotary switch set to U, N, A, or X, press the softkey labeled Menu.
Move the menu selector to the menu item labeled AC+DC. At this point, three
different softkey labels
indicate AC+DC (F1), AC,DC (F2), and DC,AC (F3). Press the softkey that
presents these two signals as needed.
While in any of the three AC+DC modes, peak measurements, frequency, duty
cycle, and period measurements are not allowed. In addition to these modes,
MIN MAX, relative, and relative % are not allowed in AC,DC or DC,AC modes.
Note
The bar graph is not displayed while the Meter is in any of the three AC+DC
modes.
AC over DC
8:10pm
06/13/07
DC over AC
8:10pm
06/13/07
8:10pm
AC + DC
06/13/07
123.45 VAC Auto Range 500 VAC
23.45 VDC
23.45 VDC Auto Range 500 V
123.45VAC
125.66 V AC + DC Auto Range 500 V
23.45 VDC
Menu
Save
Setup
Menu
Save
Setup
Menu
Save
Setup
Figure 13. AC and DC Display
est30.eps
Both manual and autoranging is available when using the ac+dc modes. The same range is used for both ac and dc signals.
However, while in autorange, up-ranging occurs when either the ac or dc signal exceeds the present range. Down-ranging occurs
26
only when both the ac and dc signals drop below 10% of the present range. For
AC+DC, ranging is controlled by the underlying values of the ac and dc signals
and not by the sum of the AC+DC calculation.
To exit the AC+DC mode, press the softkey labeled Menu and select the default
mode for the selected function. For dc volts and dc millivolts functions, move
the menu selector to VDC and press the softkey labeled VDC. For the current
functions, move the menu selector to the AC,DC menu item and press either the
AC or DC softkey.
True-rms Digital Multimeters Making Measurements
27
287/289 Users Manual
Measuring Temperature
XW Warning
To avoid the potential for fire or electric shock, do not connect the
thermocouple to electrically live circuits. The Meter uses an 80BK-A
Integrated DMM Temperature Probe or other type-K temperature probe for
measuring temperature. To measure temperature, set up the Meter as shown in
Figure 14. Press the softkey labeled Menu and move the menu selector to the
menu item labeled Temp. Press the softkey labeled F for temperature in
Fahrenheit or C for Celsius.
Note A Meter classified as “SI” will not have an F selection. The primary
display normally shows temperature or the message “Open Thermocouple”. The
open thermocouple message may be due to a broken (open) probe or because no
probe is installed into the input jacks of the Meter. Shorting the W terminal
to the COM terminal will display the temperature at the Meter terminals.
Note R is disabled when the Meter is in the Temperature function.
28
8:10pm
06/13/07
26.5 C
Menu
+8.0 Offset
Save
Offset Setup
80BK Type K Thermocouple
Probe
Vent or
Pipe
est17.eps
Figure 14. Temperature Measurement
To input a temperature offset value, press the softkey labeled Offset to open
a message box with the present offset value. Use 6 and 5 to position the
cursor over one of the digits or the polarity sign. Use 7 and 8 to scroll
through the numbers for each digit in the offset or switch between a + or
offset. With the desired value displayed, press the softkey labeled OK to set
the temperature offset. When set to something other than 0.0, the offset value
is shown in the secondary display.
Measuring Resistance
W Caution
To avoid possible damage to the meter or to the equipment under test,
disconnect circuit power and discharge all high-voltage capacitors before
measuring resistance.
The Meter measures resistance (opposition to current flow) in ohms (). This is
accomplished by sending a small current out through the test leads to the
circuit under test.
To measure resistance, set the Meter’s rotary switch to S and
set up the Meter as shown in Figure 15.
True-rms Digital Multimeters Making Measurements
Keep the following in mind when measuring resistance.
Because the meter’s test current flows through all possible paths between the
probe tips, the measured value of a resistor in a circuit is often different
from the resistor’s rated value.
The test leads can add 0.1 to 0.2 of error to resistance measurements. To test
the leads, touch the probe tips together and read the resistance of the leads.
To remove lead resistance from the measurement, hold the test lead tips
together and press the softkey labeled Menu. Next, move the menu selector to
the menu item labeled REL and press the softkey labeled REL. Now all future
displayed readings indicate the resistance at the probe tips.
The Meter’s resistance function includes modes to help with resistance
measurements. Pressing the softkey labeled Menu opens a menu of items that can
be used to modify the basic resistance measurement. Refer to the appropriate
section in this manual to learn more about each menu item.
29
287/289 Users Manual
8:10pm
06/13/07
5.67
Auto Range
0
100
200
300
400
500
Menu
Save
Setup
In-Circuit Resistance Measurements
Circuit Power
OFF
Isolating a Potentiometer
1 3
2
Disconnect
1
2
3
Isolating a Resistor
Disconnect
Figure 15. Resistance Measurement
est11.eps
30
Using the Y Function (Model 289 Only)
W Caution
To avoid damaging the circuit under test, be aware the Meter sources current
up to 10 mA at an open circuit voltage up to 20 volts.
To measure low resistances with the Meter, position the rotary switch to Y.
This function has a single range and R is therefore disabled when the Meter is
in the Y function.
Only the relative and relative percent functions work with the Y function.
Press the softkey labeled Menu to access these two functions.
Testing for Continuity
W Caution
To avoid possible damage to the meter or to the equipment under test,
disconnect circuit power and discharge all high-voltage capacitors before
testing continuity.
Continuity is the presence of a complete path for current flow. The continuity
function detects intermittent opens and shorts lasting as short as 1 ms. The
Meter uses three indicators for the absence and presence of continuity: a
resistance reading, an open/short indicator, and a beeper.
The resistance reading is simply an ohms function measurement. However, for
continuity transitions that are very short, the slow
True-rms Digital Multimeters Making Measurements
measurement response of the Meter will not appear in the digital display.
Therefore, the continuity function uses a graphical indicator for the presence
or absence of continuity. Figure 16 shows the short and open continuity
indication.
Short
Open
Figure 16. Continuity Indicator
est36.eps
To perform a continuity test, position the rotary switch to S and
set up the Meter as shown in Figure 17. Press the softkey labeled j. In
continuity, a short means a measured value less
than 8 % of full scale for the 500 range and less that 4 % for other
resistance ranges.
Note
The Meter operates in manual range only while the continuity function is
selected.
31
287/289 Users Manual
For in-circuit tests, turn circuit power off. Beep on short
8:10pm
06/13/07
5.67 Continuity
0
100
200
300
400
500
Beep on SHORT
Menu
Save
Setup
OFF (open)
Beep on open
8:10pm
06/13/07
OL
Continuity
0
100
200
300
400
500
Beep on OPEN
Menu
Save
Setup
ON (closed)
ON (closed)
OFF (open)
Figure 17. Continuity Testing
est13.eps
32
To change whether the beeper sounds on shorts or opens, press the softkey
labeled Menu. Move the menu selector to the menu item labeled Beeper and press
the softkey labeled Short/O…. This beeper selection, Beep on Short or Beep on
Open, is displayed just above the continuity indicator. The continuity beeper
is always enabled when continuity mode if first entered.
To enable or disable the beeper for continuity, press the softkey labeled
Menu. Move the menu selector to the menu item labeled Beeper and press the
softkey labeled Beeper. The status of the continuity beeper is displayed to
the right of the resistance reading with j when enabled and i when disabled.
This setting is independent of the Meter’s beeper enable/disable setting in
the setup menu.
Toggle between the continuity and ohms functions by pressing softkey F3, which
is always labeled with the alternate function.
True-rms Digital Multimeters Making Measurements
33
287/289 Users Manual
Using Conductance for High Resistance Tests
Conductance, the inverse of resistance, is the ability of a circuit to pass
current. High values of conductance correspond to low values of resistance.
The unit of conductance is the Siemens (S). The meter’s 50 nS range measures
conductance in nanosiemens (1 nS = 0.000000001 Siemens). Because such small
amounts of conductance correspond to extremely high resistance, the nS range
is used to measure the resistance of components up to 100,000 M, or
100,000,000,000 (1 nS = 1,000 M).
To measure conductance, position the rotary switch to S and
set up the Meter as shown in Figure 18. Move the menu selector to the menu
item labeled Ohms,nS,j and press the softkey labeled nS. There is normally a
residual conductance reading with the test leads open. To ensure accurate
readings, press the softkey labeled Menu. Move the menu selector to the menu
item labeled REL and press the softkey labeled REL to subtract the residual
value with the test leads open.
Note R is disabled when the Meter is measuring conductance.
34
8:10pm
06/13/07
26.55 nS
Manual Range
0
10
20
30
40
50 nS
Menu
Save
Setup
est14.eps
Figure 18. Conductance Measurement
Measuring Capacitance
W Caution
To avoid possible damage to the meter or to the equipment under test,
disconnect circuit power and discharge all high-voltage capacitors before
measuring capacitance. Use the dc voltage function to confirm that the
capacitor is discharged.
Capacitance is the ability of a component to store an electrical charge. The
unit of capacitance is the farad (F). Most capacitors are in the nanofarad
(nF) to microfarad (µF) range.
The Meter measures capacitance by charging the capacitor with a known current
for a known period of time, measuring the resulting voltage, and then
calculating the capacitance.
True-rms Digital Multimeters Making Measurements
8:10pm
06/13/07
26.52 Auto Range 100 uF
Menu
Save
Setup
+ + + + + + + +
- est15.eps
Figure 19. Capacitance Measurement
35
287/289 Users Manual
To measure capacitance, position the rotary switch to P and set up the Meter
as shown in Figure 19. If the display doesn’t already indicate the Meter is
measuring capacitance, press the softkey labeled Menu. Next, move the menu
selector to the menu item labeled Diode,Cap and press the softkey labeled Cap.
Note
To improve measurement accuracy of small value capacitors, press Menu and move
the menu selector to the menu item labeled REL. With the test leads open,
press the softkey labeled REL to subtract the residual capacitance of the
Meter and leads.
Testing Diodes
W Caution
To avoid possible damage to the meter or to the equipment under test,
disconnect circuit power and discharge all high-voltage capacitors before
testing diodes.
Use the diode test to check diodes, transistors, silicon controlled rectifiers
(SCRs), and other semiconductor devices. The test
sends a current through a semiconductor junction, and then measures the
junction’s voltage drop. A typical junction drops 0.5 V to 0.8 V.
To test a diode out of a circuit, position the rotary switch to P and set up
the meter as shown in Figure 20. If the display doesn’t already indicate the
Meter is in the Diode Test function, press the softkey labeled Menu. Next,
move the menu selector to the menu item labeled Diode,Cap and press the
softkey labeled Diode.
If the beeper is enabled during diode test, it will beep briefly for a normal
junction and sound continuously for a shorted junction, below 0.1 V. See the
“Disabling and Enabling the Beeper” section to disable the beeper.
In a circuit, a similar diode should still indicate a forward-bias reading of
0.5 V to 0.8 V; however, the reading can vary depending on the resistance of
other pathways between the probe tips.
Note
R and MIN MAX are disabled when the Meter is setup for diode test.
36
8:10pm
06/13/07
.567 VDC
Manual Range
0
1
2
3
4
5 VDC
Menu
Save
Setup
Typical Reading
Forward Bias
True-rms Digital Multimeters Making Measurements
8:10pm
06/13/07
OL VDC
Manual Range
0
1
2
3
4
5 VDC
Menu
Save
Setup
Reverse Bias
Figure 20. Diode Testing
est16.eps
37
287/289 Users Manual
Measuring Current
XW Warning
To avoid damage to the Meter and possible injury, never attempt an in-circuit
current measurement where the open-circuit potential to earth is greater than
1000 V.
W Caution
To avoid possible damage to the Meter or to the equipment under test, check
the meter’s fuses before measuring current. See the Maintenance section later
in this manual. Use the proper terminals, function, and range for your
measurement. Never place the probes across (in parallel with) any circuit or
component when the leads are plugged into the current terminals.
Current is the flow of electrons through a conductor. To measure current, you
must open the circuit under test, then place the meter in series with the
circuit.
Note
When measuring current, the display will flash when the input current exceeds
10 amps for the A terminal
and 400 mA for the mA/µA terminal. This is a warning
that current is approaching the fuse’s current limit.
To measure ac or dc current, proceed as follows:
1. Turn off power to the circuit. Discharge all high- voltage capacitors.
38
2. Insert the black lead into the COM terminal. Insert the red lead in an
input appropriate for the measurement range.
Note
To avoid blowing the Meter’s 440 mA fuse, use the mA/µA terminal only if you
are sure the current is less than 400 mA.
3. If you are using the A terminal, set the rotary switch to A. If you are
using the mA/µA terminal, set the rotary switch to X for currents below 5000
µA (5 mA), or A for currents above 5000 µA. See Figure 21 for test lead
connections and function selection. Refer to the “Input Alert Feature” section
for information on the alerts the Meter uses when leads are not used correctly
for current measurements.
4. As shown in Figure 22, open the circuit path to be tested. Touch the red
probe to the more positive side of the break; touch the black probe to the
more negative side of the break. Reversing the leads will produce a negative
reading, but will not damage the meter.
5. Turn on power to the circuit; then read the display. Be sure to note the
measurement unit given at the right side of the display (µA, mA, or A).
6. Turn off power to the circuit and discharge all high-voltage capacitors.
Remove the Meter and restore the circuit to normal operation.
Note
While in a current measurement function, the Meter will stay in the selected
AC or DC current measurement mode when switching between n and
. Whenever switched to one of the current
measurement functions, the Meter will default to the last current type
selected (AC or DC).
True-rms Digital Multimeters Making Measurements
39
287/289 Users Manual
8:10pm
06/13/07
2.5527 AAC
Auto Range
0
1
2
3
4
5 AAC
Menu
Save
Setup
8:10pm
06/13/07
19.783 mAAC
Auto Range
0
1
2
3
4
5 AAC
Menu
Save
Setup
8:10pm
06/13/07
4.863 uAAC
Auto Range
0
1
2
3
4
5 AAC
Menu
Save
Setup
Figure 21. Current Measurement Setup
est18.eps
40
8:10pm
06/13/07
mAAC
Circuit Power: OFF to connect meter. ON for measurement. OFF to disconnect meter.
True-rms Digital Multimeters Making Measurements
Total current to circuit
Current through one component
Figure 22. Current Measurement Circuit Connection
est19.eps
41
287/289 Users Manual
W Caution
Placing the probes across (in parallel with) a powered circuit when a lead is
plugged into a current terminal can damage the circuit you are testing and
blow the Meter’s fuse. This can happen because the resistance through the
Meter’s current terminals is very low, so the Meter acts like a short circuit.
The following are tips for current measurements:
A current meter drops a small voltage across itself, which might affect
circuit operation. You can calculate this burden voltage using the values
listed in the specifications under Burden Voltage (A, mA, µA).
The Meter’s current functions have a number of modes which provide more
details about a current signal. Pressing the softkey labeled Menu opens a menu
of items that can be used to modify the basic current measurement. Refer to
the appropriate section in this manual to learn more about each menu item.
To clear all modes and return to the basic ac or dc current measurement, press
the softkey labeled Menu. Move the menu selector to the item labeled AC,DC.
Press the softkey labeled AC to clear all functions and modes and make basic
ac current measurements, or DC for basic dc current measurements.
Measuring Frequency
Frequency is the number of cycles a signal completes each second. The meter
measures the frequency of a voltage or current signal by counting the number
of times the signal crosses a threshold level within a specified period of
time.
42
Figure 23 highlights the functions that allow frequency measurements.
est21.eps
Figure 23. Functions Allowing Frequency Measurement
The Meter autoranges to one of five frequency ranges: 99.999 Hz, 999.99 Hz,
9.9999 kHz, 99.999 kHz, and 999.99 kHz. Figure 24 shows a typical frequency
display. Pressing R controls the input range of the primary function (volts or
amps) and not the frequency range. To measure frequency, rotate the switch to
one of the primary functions allowing frequency measurements highlighted in
Figure 23. Press the softkey labeled Menu and move the menu
selector to the menu item labeled Hz,%,ms. Next press the softkey labeled Hz.
8:10pm
06/13/07
60.050Hz
Auto Range
0
100
200
300
400
500 VAC
123.45 VAC
Menu
Save
Setup
Figure 24. Frequency Display
est22.eps
As shown in Figure 24, the frequency of the input signal is displayed in the primary display. The volts or amps value of the signal is shown in the secondary display. The bar graph does not indicate frequency but indicates the volts or amps value of the input signal.
True-rms Digital Multimeters Making Measurements
Selection between a rising trigger c or falling trigger dedge is performed by
pressing the softkey labeled c d. This softkey toggles the trigger setting
between the two selections.
The following are tips for frequency measurements:
If a reading shows as 0 Hz or is unstable, the input signal may be below or
near the trigger level. You can usually correct these problems by manually
selecting a lower input range, which increases the sensitivity of the meter.
If a reading seems to be a multiple of what you expect, the input signal may
be distorted. Distortion can cause multiple triggerings of the frequency
counter. Selecting a higher voltage range might solve this problem by
decreasing the sensitivity of the meter. In general, the lowest frequency
displayed is the correct one.
Measuring Duty cycle
Duty cycle (or duty factor) is the percentage of time a signal is above or
below a trigger level during one cycle, as shown in Figure 25.
The duty-cycle mode is optimized for measuring the on or off time of logic and
switching signals. Systems such as electronic fuel injection systems and
switching power supplies are controlled by pulses of varying width, which can
be checked by measuring duty cycle.
43
287/289 Users Manual
Measures Positive Pulse
Trigger Level
30% Above Trigger Level
100%
Measures Negative Pulse
Trigger Level 70% Below Trigger Level 100%
Figure 25. Duty Cycle Measurements
44
est28.eps
To measure duty cycle, position the rotary switch on one of the functions allowing frequency measurements shown in Figure 23. Press the softkey labeled Menu and move the menu selector to the menu item labeled Hz,%,ms. Next press the softkey labeled %.
8:10pm
123.45 VAC
06/13/07
49.75 % Duty Cycle
Auto Range
0
100
200
300
400
500 VAC
59.756 Hz
Menu
Save
Setup
Figure 26. Duty Cycle Display
est24.eps
As shown in Figure 26, the duty cycle percentage is shown in the primary display while the signal frequency appears in the secondary display. The mini- measurement display indicates the
True-rms Digital Multimeters Making Measurements
volts or amps value of the input signal. The bar graph tracks the volts or
amps value of the signal and not the duty cycle value.
The pulse polarity is displayed to the right of the duty cycle value. J
indicates a positive pulse and K indicates a negative pulse. To change the
polarity being measured, press the softkey labeled J K. The polarity indicator
changes to the opposite polarity.
For 5 V logic signals, use the 5 V dc range. For 12 V switching signals in
automobiles, use the 50 V dc range. For sine waves, use the lowest ac or dc
range that does not result in multiple triggering. A manually-selected lower
input range will often measure better than the AUTO-selected input range.
Measuring Pulse Width
The pulse width function measures the amount of time a signal is high or low,
as shown in Figure 27. The measured waveform must be periodic; its pattern
must repeat at equal time intervals.
The meter measures pulse width from 0.025 ms to 1250.0 ms ranges.
To measure pulse width, position the rotary switch to one of the functions
allowing frequency measurements shown in Figure 23. Press the softkey labeled
Menu and move the menu selector to the menu item labeled Hz,%,ms. Next, press
the softkey labeled ms.
45
287/289 Users Manual
Trigger Level
46
Measure Positive Pulse Width
Pulse Width
1 Period = Frequency
Measure Negative Pulse Width
Trigger Level Pulse Width
Period Figure 27. Pulse Width Measurements
est27.eps
The primary display indicates the input signals pulse width in milliseconds.
The frequency of the signal is displayed in the secondary display. The mini-
measurement display indicates the volts or amps value of the input signal. The
bar graph tracks the volts or amps value of the signal and not the pulse width
value.
The pulse width polarity is displayed to the right of the duty cycle value. J
indicates a positive pulse width and K indicates a negative pulse. To change
the polarity, press the softkey labeled J K. The polarity indicator changes to
the opposite polarity.
Changing Meter Setup Options
The Meter has a number of preset features such as date and time formats,
backlight and battery save mode timeouts, and the displayed language. These
variables are referred to as Meter setup options. Many setup options affect
general Meter operations and are active in all functions. Others are limited
to one function or group of functions.
Access to the setup options is always available through the softkey labeled
Setup. Information about the Meter, such as serial number, model, for example.
is also accessed through the setup menu.
Resetting Meter Setup Options
The Meter’s setup options can be reset to default values through the setup
menu. Open the setup menu by pressing the softkey labeled Setup. Position the
menu selector next to the menu item labeled Reset and press the softkey
labeled Setup. A message will appear asking to confirm the reset action. Press
the softkey labeled OK to perform the reset.
True-rms Digital Multimeters Changing Meter Setup Options
Note
A setup reset also resets the temperature offset and dBm reference to their
default value.
In addition to resetting the setup variables, pressing the softkey labeled
Meter will also clear all saved measurement screens, MIN MAX screens, peak
screens, and recording records. The Meter’s clock is reset to a default value
as well.
Setting Display Contrast
Meter display contrast can be adjusted through the Meter’s setup menu. Open
the setup menu by pressing the softkey labeled Setup and position the menu
selector next to the menu item labeled Contrast. Pressing the softkey labeled
- (F1) increases display contrast, while the softkey labeled (F2) decreases
contrast.
Contrast can also be set through the 7 and 8 buttons when not being used to move between menu selections.
Setting the Meter’s Language
The Meter comes from the factory with the display language set to English. To select another language, open the setup menu by pressing the softkey labeled Setup. Move the menu selector next to the menu item labeled Display. Next, press the softkey labeled Format (F2) to open the format menu. If not already selected, move the menu selector to the left of the menu item labeled Language and press the softkey labeled Edit. The currently selected language becomes highlighted and f appears to the right of the language. Use 7 and 8 to scroll through the available languages, then press the softkey labeled OK to set the Meter’s display language. Press the softkey labeled Close to return to normal Meter operation.
47
287/289 Users Manual
Setting Date and Time
The Meter’s internal clock is used in the display and for timestamping
recorded measurements. To change the date and time as well as the display
format, press the softkey labeled Setup. Position the menu selector next to
the menu item labeled Display. To set the date and time, press the softkey
labeled Date/Time to open the date/time menu. Next, position the menu selector
next to either the Set Date item or Set Time item and press the softkey
labeled Edit. Using 5 and 6, position the cursor on the date or time element
to adjust. Use 7 and 8 to change the selected date or time element value.
Press OK to complete the action.
Setting Backlight and Auto Off Timeouts
The Meter’s backlight and auto off features use timers to determine when to
turn off the backlight, when to automatically turn the Meter off or enable the
battery saver mode. To set these timeouts, press the softkey labeled Setup and
position the menu selector next to the menu item labeled Instrument. Position
the menu selector next to the menu item labeled Auto Backlight Timeout or Auto
Power Off and then press the softkey labeled Edit. Use 7 and 8 to adjust the
time to one of the preset values. Press OFF to disable the timeout feature.
Press the softkey labeled OK to set the selected time. Press the softkey
labeled Close to return to normal Meter operation.
The battery-saver mode is used when the Meter is performing a recording
session or during MIN MAX, Peak record and AutoHold. The battery-saver mode
powers down circuits that are not involved in the operation of these recording
sessions, including the display. For the record mode, the timeout period is
set to five minutes and is enabled only when the Auto Power Off
48
timeout is set to a value other than Off. For MIN MAX, Peak, and AutoHold, the
timeout is the time period set for Auto Off.
Setting a Custom dBm Reference
To add a custom dBm reference value, press the softkey labeled Setup and
position the menu selector next to the menu item labeled Instrument. Next
press the softkey labeled Instrument and position the menu selector next to
the menu item labeled dBm Reference. Next, press the softkey labeled Edit. Use
5 and 6 to position the cursor on a specific digit. Press 7 and 8 to increment
or decrement the digit. With the desired reference displayed, press the
softkey labeled OK to add this value to the dBm reference list. Only one
custom value is allowed. Press the softkey labeled Close to return to normal
Meter operation.
Disabling and Enabling the Beeper
The Meter’s beeper alerts users to the presence of messages, operator errors
such as incorrect lead connections for the selected function, and newly sensed
values for MIN MAX and Peak recording. Although the beeper is also used for
the continuity function, controlling the beeper for that function is not done
through this setup option. See the “Testing for Continuity” section for
information on the continuity beeper.
To enable or disable the Meter’s beeper, press the softkey labeled Setup and
position the menu selector next to the menu item labeled Instrument. Next
press the softkey labeled Instrument and position the menu selector next to
the menu item labeled Beeper. Press the softkey labeled Edit to move the
cursor to the on or off selection. Use 7 and 8 to switch the beeper on or off.
The status of the beeper is indicated in the display’s status bar (see item 12
in Figure 2).
Enabling and Disabling the Smoothing Mode
When an AC input signal is noisy or changes rapidly, the smoothing mode may
display a steadier reading. To enable or disable the smoothing mode, press the
softkey labeled Setup and position the menu selector next to the menu labeled
Instrument. Next press the softkey labeled Instrument and position the menu
selector next to the menu item labeled Smoothing. Press the softkey labeled
Edit to move the cursor to the on or off selection. Use 7 and 8 to switch the
smoothing mode on or off.
Using Other Setup Options
Additional setup options hold information about the Meter as well as some
general Meter functions. The Meter Info selection lists the serial number,
model number, firmware version, calibration date, and calibration counter.
Operator name, company name, site name, and contact information are also
displayed when this information has been loaded into the Meter from FlukeView®
Forms software.
The Calibration selection allows a qualified calibration technician to enter a
password that allows the Meter to be calibrated. See the 287/289 Calibration
Information document to calibrate the Meter.
The Secure Erase option allows user-accessible memory to be erased as required
by Homeland Security regulations. Meter calibration is not lost when this low-
level erase is performed.
As new Meter features are created, the latest version of software can be
downloaded to the Meter from Fluke’s support web page using the Software
Update option.
True-rms Digital Multimeters Using Memory
Using Memory
The Meter has memory for storing individual measurements, measurements
collected over a specified duration, and measurement events.
All stored data can be viewed on the Meter or downloaded to a PC through the
Meter’s infrared (IR) communication link using FlukeViewTM Forms. See the
“Using Communications” section for more information on communicating with a PC
through FlukeView Forms software.
Storing Individual Measurement Data
For all measurement functions, a snapshot of the screen data is saved by
pressing the softkey labeled Save. Except for the minimeasurement in the
status bar, the display freezes and the Save menu appears. Two choices allow
either saving the data under a previously selected name or choosing another
name by pressing the softkey labeled +Name. See the “Naming Saved Data”
section later in this manual. The displayed data is stored along with the date
and time of day the save was performed.
For MIN MAX and Peak, the displayed summary data can be stored at any time by
pressing the softkey labeled Save, thus preserving a snapshot of the session
at that moment.
Naming Saved Data
The Meter has a list of eight preset names under which measurement data is
saved. Multiple records can be saved using the same name. For example, one
preset name is Save. The first time a save operation is performed with that
name, Save-1 is used to name the record in memory. The next time the Save name
is used, the number increments to 2 and the record is
49
287/289 Users Manual
saved under the name of Save-2. The auto-incrementing number can be reset to 1
by positioning the menu selector next to the save name and then pressing the
softkey labeled Reset #.
To save a screen shot, a recording session, or MIN MAX or Peak record session,
press the softkey labeled Save. To select the name from the preset list, press
+Name. To save to the same name as before but with the next number, press the
softkey labeled Save. This second method makes it easy to save a series of
measurements simply by pressing the Save softkey twice for each save
operation.
When selecting the name for a save operation, position the menu selector next
to the desired name using the cursor buttons. Next press the softkey labeled
Save.
Viewing Memory Data
Viewing data stored in the Meter’s memory is performed through the save menu.
Press the softkey labeled Save. Position the menu selector next to the menu
item labeled View Memory and press the softkey labeled View.
Note
To view data stored in memory, the Meter must not be recording or performing a
MIN MAX or peak record session.
The Meter separates stored data into four different categories: Measurement,
MIN MAX, Peak, and Recording. Use the cursor buttons to position the menu
selector next to the desired saved data category and then press the softkey
labeled View. The Meter displays the last record saved for the selected data
category.
50
If there are previously stored records, press the softkey labeled Prev or 5 to
page back through previously stored records. Press the softkey labeled Next or
6 to page in the other direction. Press Close to return to normal Meter
operation.
Viewing Snapshot and Summary Data
After selecting the MIN MAX, Peak, or Measurement category described in the
Viewing Memory Data section above, pressing View only displays the information
stored at the time a Save was performed. The display is reconstructed from
this data when viewed.
Viewing Trend Data
For the Recording category, the interval and event data stored during a
recording session are viewed on the Meter through a trend-plot view, similar
to a strip-chart recorder. See the “Recording Measurement Data” section later
in this manual for an explanation of interval and event data.
After selecting the recording category described in the Viewing Memory Data
section above and pressing View, the recording session’s summary screen is
displayed (See Table 9). Press the softkey labeled Trend to display the
recorded data in a trend-plot view. Table 7 shows the trend view along with a
description of each of its components.
To look at data stored in the individual records that make up the trend, move
the cursor to any point along the plot by pressing 6 or 5. The value and
timestamp of the minimum, maximum, and end-of-record values of the selected
record are displayed at the bottom of the cursor. All data contained in a
record can only be viewed on a PC running FlukeView Forms.
Note X-axis time labels are shown in elapsed time while the timestamp shown
under the cursor is an absolute time
Table 7. Trend Data Display
1
2
34
01/16/08 09:43:17am
x1
40
30 VAC
20
10
0
0:15 0:30 0:45 1:00 1:15 1:30
HH:MM
23.456 VAC 08:13:05 6/28/07
6
7
5
est35.eps
Item A B C D
Description Cursor Start date and time Trend line Zoom level.
True-rms Digital Multimeters Using Memory
E
Elapsed time. Units in hours and minutes, or minutes and seconds.
F Time scale legend (HH:MM or MM:SS)
G
Measured value and timestamp of selected record.
Zooming in on Trend Data
While viewing trend data, pressing 7 or 8 zooms in or out respectively on the
data around the cursor. Each press of 7 reduces the x-axis time period by one
half to reveal more details. Each press of 8 doubles the time period until all
the recorded data is displayed. The zoom level is displayed in the upper-right
hand corner of the display. X1 indicates the trend of the complete recording
period is displayed. X2 is one-half the recording time. X3 is one-fourth the
recoring time. This magnification can continue until the x-axis time period is
one second.
Deleting Stored Measurement Data
Deleting data stored in the Meter’s memory is performed through the save menu.
Press the softkey labeled Save. Use the softkey labeled Prev and Next to
select an item for deletion.
The Meter separates stored data into four different categories: Measurement, MIN MAX, Peak, and Recording. Use the cursor buttons to position the menu selector next to a saved data category and then press the softkey labeled View to view the item.
Pressing the softkey labeled Delete All will delete all stored data under the selected saved data category. Or press the softkey labeled View. After accepting a confirmation message, use the
51
287/289 Users Manual
softkeys labeled Prev and Next to select an item for deletion. Next, press the
softkey labeled Delete. A message asking to confirm the deletion will appear
before anything is deleted from memory.
Recording Measurement Data
The Meter’s record feature collects measurement information over a user-
specified duration. This collection of information is called a recording
session. A recording session is made up of one or more measurement records.
Each record contains measurement summary information covering the duration of
the record.
Each record contains the minimum, maximum, and average value detected over the
record’s duration. In addition to measurement values, time stamps are also
captured and saved with each record. The time stamps consist of the record
start time, the time the maximum value was detected, the time the minimum
value was detected, and the record end time.
Some of a records data can be viewed through the Meter’s Viewing Trend Data
function. Viewing all the data that makes up a record can only be viewed on a
PC running FlukeView Forms software.
There are two measurement record types that are captured during a recording
session: interval and event. An interval record covers a user-specified
interval. An event record has a duration determined by the activity of the
measured signal and can interrupt an interval record. Even if an interval
record is interrupted, a record will end and a new interval record will begin
when the scheduled interval time expires.
52
Event records are triggered by the measured signal varying more than an
adjustable percentage of the value measured at the start of the record. This
adjustable percent is called the Event Threshold for recording. In addition to
the values and time stamps mentioned above, an event record also stores
whether the signal was stable or unstable during the event record duration. To
be classified as stable, the measured signal’s value must stay within the
selected percentage of the start value for at least one second. Measured
signals that exceed the percentage threshold in less than one second are
classified as unstable. See the “Setting the Event Threshold Value” section
later in this manual.
Note
For temperature measurements, the AutoHold threshold is a percent of 100
degrees. The default AutoHold threshold is 4% of 100 degrees, or 4 degrees
Celsius or Fahrenheit.
A record ends when one of the following occurs:
· The start of a new interval record.
· A range overload, causing the Meter to change range.
· A non-ranging overload, when in manual range or the highest range.
· The measured value changes more than 4 % of the measured value at the start
of the record.
· The recording session terminates.
A recording session termination can be caused by one of the following:
· Recording session duration expiring.
· Manually stopping the recording session.
Setting up a Recording Session
Before starting a recording session, setup the Meter for the measurements to
be recorded. If needed, change the event threshold value (see the “Setting the
Event Threshold Value section later in this manual). Press the softkey labeled
Save to open the save menu. Using the cursor buttons, move the menu selector
next to the menu item labeled Record and press the softkey labeled Record to
open the configuration display.
There are two variables in setting up a recording session: Recording session
duration, and sample interval duration. Both variables affect the recording
length and number of intervals recorded. These two variables may interact, in
that setting one variable may adjust the other variable to fit the recording
session within the available memory. The percentage of memory available at the
beginning of a recording session is displayed below the duration and sample
interval settings. Option values can be adjusted as follows:
The sample interval can be set from one second to 99 minutes and 59 seconds.
Recording session duration can be set from one minute to 99 days 23 hours 59
minutes.
True-rms Digital Multimeters Recording Measurement Data
Table 8. Recording Display
8:10pm
123.45 VAC
06/13/07
Recording
1
1.75 VAC
Auto Range
0
100
200
300
400
500 VAC
Start Time: 06/10/07 07:23:55
2
6
Remaining Time: 2 Hrs 26 mins Interval Samples: 47 Events:7
3
5
Reference: 121.70 VAC
4
Stop
est31.eps
Item
Description
A Recording session in progress icon.
B Time and date when recording session started.
C Time remaining until recording session stops.
D Total number of event records recorded so far.
E Reference value for relative measurements.
F Total number of interval records recorded so far.
53
287/289 Users Manual
The Meter allocates memory in such a way as to guarantee capturing all of the
user-specified sample intervals. Event records will also be captured until the
Meter detects the allocated memory has been used up. At this point, events are
not recorded, but the event counter continues to advance to indicate the total
number of events that occurred. A plus (+) sign appears after the event count
to indicate this condition.
Note
The maximum number of recorded sample intervals is 10,000. The maximum number
of recorded events is 15,000 minus the number of sample intervals. These
maximum numbers are proportionally decreased when the available memory is low.
To change either of the two recording variables, use the cursor buttons to
position the menu selector next to the desired menu item and press the softkey
labeled Edit. Use 5 6 7 and 8 to move between and set each digit of the
selected variable.
If the battery level is anything but full, a message appears at the bottom of
the record menu reminding you of the battery level before you start the
recording session.
Setting the Event Threshold Value
Press the softkey labeled Setup to access the setup menu. Using the cursor
buttons, move the menu selector next to the menu item labeled Recording and
press the softkey labeled Recording to open the recording setup screen. Using
the cursor buttons, move the menu selector next to the menu item labeled Event
Threshold for Recording and then press the softkey labeled Edit. Press 7 or 8
to scroll through the event threshold
54
values. With the desired value selected, press the softkey labeled Close.
Starting a Recording Session
Once the variables are set, press the softkey labeled Start, at which point s
appears in the display and the green LED surrounding the power button (O)
flashes. Table 8 shows the recording display and describes the information
displayed.
The Menu, Setup, Reference, and Temperature Offset softkey functions are not
available while the Meter is recording. This ensures that measurements within
a recording session are consistent.
To extend battery life while recording, the Meter may enter a battery-saver
mode five minutes after a push-button is pressed or IR communications activity
ends. If the Auto Power-Off timeout is set to never, the battery-saver mode is
disabled.
Stopping a Recording Session
The recording session will continue until the allocated memory is used, the
batteries expire, the rotary switch is moved, a probe is inserted or removed
from the A or mA/µA jacks, or the session is terminated by pressing the
softkey labeled Stop.
Table 9 shows the display and describes the displayed information after
stopping a recording session.
After stopping a recording session, choose to save the recording session, view
the trend data (see the “Viewing Trend Data” section), or close the recording
session. If the session wasn’t saved before pressing the Close softkey, the
data is lost.
Table 9. Stopped Recording Display
1
8:10pm
REL
123.45 VAC
06/13/07
Stopped Recording
1.75 VAC
Auto Range
0
100
200
300
400
500 VAC
Start Time : 06/10/07 07:23:55
2
6
Duration : 3 days 14 Hrs 47 mins Interval Samples: 135 Events:53
3
5
Reference: 121.70 VAC
4
Save
Trend
Close
est34.eps
Item A B C D E F
Description
Stopped Icon indicating the recording session is stopped. Time and date when
the recording began. Length (duration) of time recording session was in
progress. Number of event records detected. Reference value for relative
measurements. Number of interval records detected.
True-rms Digital Multimeters Using Communications
Using Communications
You can use the IR communication link and FlukeView Forms software to transfer
the contents of a meter’s memory to a PC. When using a PC-to-meter IR
(infrared) communication link, refer to the FlukeView Forms Installation Guide
or the on-line help.
Note The Meter will log in real-time mode to a connected computer running
FlukeView Forms. In addition, the Meter allows the user to log to internal
memory and connect to the computer later for download. FlukeView Forms allows
you to place the data into standard (default) or customized forms. The forms
display the data in table and graph form, as well as show user comments. You
can use these forms to satisfy ISO-9000 and other documentation requirements.
55
287/289 Users Manual
Error Messages
Table 10 list some of the error messages the Meter may display and the
conditions that may be causing the error.
Table 10. Error Messages
Message
Conditions
Leads connected incorrectly.
Lead in A or mA/µA jack but rotary switch not in corresponding A/mA or µA position.
Leads in both A and mA/µA jacks.
Rotary switch set to measure current but no lead in either A or mA/µA jack.
Open Thermocouple
Thermocouple wire is open or thermocouple junction is corroded.
No Thermocouple connected to the Meter’s input.
Batteries low function unavailable.
The function selected requires a higher battery level to operate within specifications.
Error: Date and Time need to be Batteries were left out too long and the Meter’s date and time were lost. reset.
Not enough memory for operation.
When starting a recording session or saving screen data, the Meter does not have enough memory to store the information.
Batteries critically low, replace Battery is too low to make measurements within stated specifications. The Meter will turn
now.
off within 15 seconds after this message appears to preserve the Meter’s date and time.
56
Maintenance
XWWarning
To avoid electrical shock or personal injury, repairs or servicing not covered
in this manual should be performed only by qualified personnel as described in
the 287/289 Service Information.
General Maintenance
Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives, isopropyl alcohol, or solvents.
Dirt or moisture in the terminals can affect readings and can falsely activate
the Input Alert feature. Clean the terminals as follows:
1. Turn the Meter off and remove all test leads.
2. Shake out any dirt that may be in the terminals.
3. Soak a clean swab with mild detergent and water. Work the swab around in
each terminal. Dry each terminal using canned air to force the water and
detergent out of the terminals.
True-rms Digital Multimeters Maintenance
Testing the Fuses
As shown in Figure 28, with the Meter in the S function, insert
a test lead into the W jack and place the probe tip on the other end of the
test lead against the metal of the current input jack. If the “Leads Connected
Incorrectly” message appears, the probe tip has been inserted too far into the
amps input jack. Back the lead out a bit until the message disappears and
either OL or a resistance reading appears in the Meter’s display. The
resistance value should be between 0.00 and 0.50 for the A jack and 10.00
±0.05 k for the jack.
XWWarning
To avoid electrical shock or personal injury, remove the test leads and any
input signals before replacing the battery or fuses. To prevent damage or
injury, install only Fluke specified replacement fuses with the amperage,
voltage, and speed ratings shown in Table 11.
57
287/289 Users Manual
8:10pm
06/13/07
10.000 k
Menu
Replace F1 if reading is OL
8:10pm
06/13/07
0.50
Menu
Replace F2 if reading is OL
Figure 28. Testing the Current Fuses
est33.eps
58
Replacing the Batteries
Refer to Figure 30 and replace the batteries as follows:
1. Turn the Meter off and remove the test leads from the terminals.
2. Remove the battery door assembly by using a standardblade screwdriver to
turn the battery door screw one-half turn counterclockwise.
3. Replace the batteries with 1.5 volt AA batteries (NEDA 15A IEC LR6).
Observe proper polarity.
4. Reinstall the battery door assembly and secure it by turning the screw
one-half turn clockwise.
Replacing the Fuses
Referring to Figure 30, examine or replace the Meter’s fuses as follows:
1. Turn the Meter off and remove the test leads from the terminals.
2. Remove the battery door assembly by using a standardblade screwdriver to
turn the battery door screw one-half turn counterclockwise.
3. Remove the fuse by gently prying one end loose, then sliding the fuse out
of its bracket.
4. Install only Fluke specified replacement fuses with the amperage, voltage,
and interrupt ratings shown in Table 11.
5. Reinstall the battery door assembly and secure it by turning the screw
one-half turn clockwise.
True-rms Digital Multimeters Maintenance
Test Lead Storage
Figure 29 shows the proper method for storing the test leads with the Meter.
Figure 29. Test Lead Storage
est41.eps
59
287/289 Users Manual
F2 11 A
F1 0.44 A
Figure 30. Replacing Batteries and Fuses
est32.eps
60
In Case of Difficulty
If the Meter does not seem to work properly:
1. Check that all batteries are installed with the correct polarity.
2. Examine the case for damage. If damage is detected, contact Fluke. See the
“Contacting Fluke” section earlier in this manual.
3. Check and replace (as needed) the batteries, fuses, and test leads.
4. Review this manual to verify correct operation.
5. If the Meter still does not work, pack it securely and forward it, postage
paid, to the location provided by the appropriate Fluke contact. Include a
description of the problem. Fluke assumes no responsibility for damage in
transit.
A Meter under warranty will be repaired or replaced (at Fluke’s option) and
returned at no charge. See the registration card for warranty terms.
True-rms Digital Multimeters In Case of Difficulty
61
287/289 Users Manual
Service and Parts
Replacement parts and accessories are shown in Tables 11 and 12 and Figure 31.
To order parts and accessories, refer to the “Contacting Fluke” section.
Item
1
2
3 4 5 6
7 8 9 10 11
12
Knob
Skin
Keypad O-Ring Case Top Screw, Phillps
Mask, LCD
LCD Module
Shock Absorber
Spring Detent
RSOB Housing, Upper
Shield
Upper Lower
Table 11. Replacement Parts Description
Qty.
Fluke Part/Model Number
1
2798434
1
2798418 (289) 2798429 (287)
1
2578234
1
2740185
1
2578178
5
2743764
1
2760673 (289) 2798407 (287)
1
2734828
3
2793516
1
2723772
1
2578283
1
2578252
1
2578265
62
True-rms Digital Multimeters Service and Parts
Table 11. Replacement Parts (cont.)
Item
Description
Qty. Fluke Part/Model Number
13 RSOB Housing, Lower
1
2578290
14 Case Bottom
1
2578184
15 Shock Absorber, Battery Compartment
1
2793525
16 Battery Contact, Negative
2
2578375
17 Battery Contact, Positive 18 WFuse (F1), 0.440 A, 1000 V, FAST, Interrupt rating 10 kA 19 WFuse (F2), 11 A, 1000 V, FAST, Interrupt rating 20 kA
1
2578353
1
943121
1
803293
20 Battery, 1.5 V NEDA 15C/15F or IEC R6S
6
376756
21 Battery Door Assembly (includes tilt bail)
1
2824477
22 Screw, Phillips
7
853668
23 TL71 Right-Angle Test Lead Set
1
TL71
24 Alligator Clips, one black and one red
2
1670652 (Black) 1670641 (Red)
25 Manual, Manual Pack, Fluke 287/289
1
2748851
26
287/289 Users Manual CD [1]
1
2748872
WTo ensure safety, use exact replacement only.
[1] The Users and Getting Started manuals are available through www.Fluke.com. Click on Support and then Product Manuals.
63
287/289 Users Manual
1 2
3 4 5
6
7
4 PL
8
2 PL 9
10
11 12
13
64
14
15 18
19 20
21
23
16 25
17
22
6 PL
Figure 31. Replaceable Parts
24 26
est40.eps
Table 12. Accessories
Item
Description
AC72
Alligator Clips for use with TL75 test lead set
AC220
Safety Grip, Wide-Jaw Alligator Clips
80BK-A
80BK-A Integrated DMM Temperature Probe
TPAK
ToolPak Magnetic Hanger
C25
Carrying Case, Soft
TL76
4 mm Diameter Test Leads
TL220
Industrial Test Lead Set
TL224
Test Lead Set, Heat-Resistant Silicone
TP1
Test Probes, Flat Blade, Slim Reach
TP4
Test Probes, 4 mm diameter, Slim Reach
Fluke accessories are available from an authorized Fluke distributor.
True-rms Digital Multimeters Service and Parts
65
287/289 Users Manual
General Specifications
Maximum voltage between any Terminal and Earth Ground: 1000 V W Fuse
Protection for mA or µA inputs ………….0.44 A (44/100 A, 440 mA), 1000 V FAST
Fuse, Fluke specified part only W Fuse Protection for A input………………………..11 A,
1000 V FAST Fuse, Fluke specified part only Battery Type ………………………………………………….6
AA Alkaline batteries, NEDA 15A IEC LR6 Battery Life ……………………………………………………100
hours minimum. 200 hours in Logging mode Temperature
Operating ……………………………………………………-20 °C to 55 °C Storage……………………………………………………….-40
°C to 60 °C Relative Humidity …………………………………………..0 % to 90 % (0 °C to 37 °C), 0
% to 65 % (37 °C to 45 °C), 0 % to 45 % (45 °C to 55 °C) Altitude Operating
……………………………………………………3,000 m Storage……………………………………………………….10,000 m Temperature
Coefficient…………………………………0.05 X (specified accuracy) /°C (<18 °C or >28 °C)
Vibration………………………………………………………..Random Vibration per MIL-PRF-28800F Class 2
Shock ……………………………………………………………1 meter drop per IEC/EN 61010-1 2nd Edition Size
(HxWxL) …………………………………………………8.75 in x 4.03 in x 2.38 in (22.2 cm x 10.2 cm x
6.0 cm) Weight …………………………………………………………..28.0 oz (871 g) Safety Standards US
ANSI……………………………………………………..Complies with ANSI/ISA 82.02.01 (61010-1) 2004
CSA……………………………………………………………CAN/CSA-C22.2 No 61010-1-04 to 1000 V Measurement
Category III and 600 V
Measurement Category IV, Pollution Degree 2 UL………………………………………………………………UL 61010
(2003) CE European……………………………………………….IEC/EN 61010-1 2nd Edition Pollution
Degree 2
66
True-rms Digital Multimeters Detailed Specifications
Electromagnetic Compatibility Standards (EMC) European
EMC…………………………………………….EN61326-1 Australian EMC ……………………………………………; N10140 US FCC
………………………………………………………FCC CFR47: Part 15 CLASS A
Certifications …………………………………………………UL, CE, CSA, ; (N10140), s
Detailed Specifications
Accuracy: Accuracy is specified for a period of one year after calibration, at
18 °C to 28 °C (64 °F to 82 °F), with relative humidity to 90 %. Accuracy
specifications are given as: ±( [ % of reading ] + [ number of least
significant digits ] ). Accuracy specification assumes ambient temperature
stable at ±1 °C. For ambient temperature changes of ±5 °C, rated accuracy
applies after 2 hours. To obtain full accuracy in DC mV, Temperature, Ohms and
Lo (50) Ohms, let the meter stabilize 20 minutes after using LoZ. True-rms: AC
mV, AC V, AC A, AC mA, and AC A specifications are ac-coupled, true rms, and
are specified from 2 % of range to 100 % of range, except 10 A range is
specified from 10 % to 100 % of range. Crest Factor: Accuracy is specified
with AC crest factor 3.0 at full-scale, increasing linearly to 5.0 at half-
scale, except the 1000 V range, where it is 1.5 at full scale, increasing
linearly to 3.0 at half-scale and 500 mV and 5000 µA, where it is 3.0 at 80 %
of full scale, increasing linearly to 5.0 at half-scale. For non-sinusoidal
waveforms add ±(0.3 % of range and 0.1 % of reading). AC Floor: When the input
leads are shorted together in the ac functions, the Meter may display a
residual reading up to 200 counts. A 200 count residual reading will cause
only a 20 count change for readings at 2 % of range. Using REL to offset this
reading may produce a much larger constant error in later measurements. AC+DC:
AC+DC is defined as ac2 + dc2
67
287/289 Users Manual
AC Voltage Specifications
Function
Range
Resolution
AC mV AC V dBV
Low pass filter [4]
50 mV [1] 500 mV 5 V [1] 50 V [1] 500 V [1] 1000 V -70 to -62 dB [3] -62 to -52 dB [3] -52 to -6 dB [3] -6 to +34 dB [3] 34 to 60 dB [3]
0.001 mV 0.01 mV 0.0001 V 0.001 V 0.01 V 0.1 V 0.01 dB 0.01 dB 0.01 dB 0.01 dB 0.01 dB
L [4]
1000 V
0.1 V
20 to 45 Hz 1.5 % + 60 1.5 % + 60 1.5 % + 60 1.5 % + 60 1.5 % + 60 1.5 % + 60
3 dB 1.5 dB 0.2 dB 0.2 dB 0.2 dB
2 % + 80
2 % + 80
45 to 65 Hz 0.3 % + 25 0.3 % + 25 0.3 % + 25 0.3 % + 25 0.3 % + 25 0.3 % + 25
1.5 dB 1.0 dB 0.1 dB 0.1 dB 0.1 dB
2 % + 40
2 % + 40
Accuracy 65 Hz to 10 kHz
0.4 % + 25 0.4 % + 25 0.6 % + 25 0.4 % + 25 0.4 % + 25 0.4 % + 25
2 dB 1 dB 0.1 dB 0.1 dB 0.1 dB 2 % +10 -6 % -60 [2] 2 % + 40 [6] [1] Below 5 %
of range, add 20 counts. [2] Specification increases linearly from -2 % at 200
Hz to -6 % at 440 Hz. Range is limited to 440 Hz.
[3] dBm (600 ) is specified by adding +2.2 dB to the dBV range values. [4] 289 only. [5] Add 2.5 % above 65 kHz. [6] Range is limited to 440 Hz. See Detailed Specifications introduction for additional information.
10 to 20 kHz 0.7 % + 40 0.7 % + 40 1.5 % + 40 0.7 % + 40 Not Spec’d Not Spec’d
2 dB 1 dB 0.2 dB 0.2 dB Not Spec’d
Not Spec’d
Not Spec’d
20 to 100 kHz 3.5 % + 40 [5] 3.5 % + 40 3.5 % + 40 [5] 3.5 % + 40 Not Spec’d
Not Spec’d 3 dB 2 dB 0.8 dB 0.8 dB Not Spec’d
Not Spec’d
Not Spec’d
68
AC Current Specifications
Function
Range
Resolution
AC µA [3]
500 µA
0.01 µA
AC mA [3]
5000 µA 50 mA
0.1 µA 0.001 mA
AC A [2]
400 mA
5 A 10 A [1]
0.01 mA 0.0001 A 0.001 A
[1] 10 A range (10 % to 100 % of range). [2] 20 A for 30 seconds on, 10 minutes off. >10 A not specified. [3] 400 mA continuous; 550 mA for 2 minutes on, 1 minute off. [4] Verified by design and type tests.
See Detailed Specifications introduction for additional information.
20 to 45 Hz 1 % + 20 1 % + 5 1 % + 20 1 % + 5 1.5 % + 20 1.5 % + 5
True-rms Digital Multimeters Detailed Specifications
Accuracy
45 to 1 kHz
1 to 20 kHz
0.6 % + 20
0.6 % + 20
0.6 % + 5
0.6 % + 10
0.6 % + 20
0.6 % + 20
0.6 % + 5 0.8 % + 20 0.8 % + 5
1.5 % + 10 3 % + 40 [4] 3 % + 10 [4]
20 to 100 kHz [4] 5 % + 40 5 % + 40 5 % + 40 5 % + 40 Not Spec’d Not Spec’d
69
287/289 Users Manual
DC Voltage Specification
Function DC mV DC V [1] L [1]
Range
50 mV [3] 500 mV 5 V 50 V 500 V 1000 V 1000 V
Resolution
0.001 mV 0.01 mV 0.0001 V 0.001 V 0.01 V 0.1 V 0.1 V
DC[2] 0.05 % + 20 [4] 0.025 % + 2 [5] 0.025 % + 2 0.025 % + 2
0.03 % + 2 0.03 % + 2
1 % + 20
Accuracy AC over DC, DC over AC, AC + DC [2]
20 to 45 Hz 45 Hz to 1 kHz 1 to 20 kHz 20 to 35 kHz
1.5 % + 40
5 % + 40
1.5 % + 40
5 % + 40
2 % + 80
0.5 % + 80
1.5 % + 40 1.5 % + 40
5 % + 40 5 % + 40
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
[1] Add 20 counts in dual display ac over dc, dc over ac or ac+dc. [2] AC+DC ranges are specified from 2 % to 140 % of range, except 1000 V is specified from 2 % to 100 % of range. [3] When using the relative mode (REL Q) to compensate for offsets. [4] Add 4 counts/10 mV AC in dual display ac over dc, dc over ac or ac + dc [5] Add 10 counts/100 mV AC in dual display ac over dc, dc over ac or ac + dc.
70
True-rms Digital Multimeters Detailed Specifications
DC Current Specifications
Function DC µA [4]
Range 500 µA
Resolution 0.01 µA
DC[1][3] 0.075 % + 20
Accuracy AC over DC, DC over AC, AC + DC [1] 20 to 45 Hz 45 Hz to 1 kHz 1 to 20 kHz 20 to 100 kHz[5]
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
DC mA [4] DC A [2]
5000 µA 50 mA 400 mA 5 A 10 A
0.1 µA 0.001 mA 0.01 mA 0.0001 A 0.001 A
0.075 % + 2 0.05 % + 10 [6] 0.15 % + 2 0.3 % + 10 0.3 % + 2
1 % + 5 1 % + 20 1 % + 5 1.5 % + 20 1.5 % + 10
0.6 % + 5 0.6 % + 20 0.6 % + 5 0.8 % + 20 0.8 % + 10
0.6 % + 10
0.6 % + 20
1.5 % + 10 3 % + 40[5] 3 % + 10[5]
5 % + 40 5 % + 40 5 % + 40 Not Spec’d Not Spec’d
[1] AC+DC ranges are specified from 2 % to 140 % of range. [2] 20 A for 30 seconds on, 10 minutes off. >10 A not specified. [3] Add 20 counts in dual display ac over dc, dc over ac or ac+dc. [4] 400 mA continuous; 550 mA for 2 minutes on, 1 minute off. [5] Verified by design and type tests. [6] Temperature coefficient: 0.1 X (specified accuracy)/ °C (<18 °C or > 28 °C)
71
287/289 Users Manual
Resistance Specifications
Function Resistance
50 [1][3] 500 [1] 5 k [1] 50 k [1]
Range
500 k
5 M
30 M
50 M
50 M up to 100 M
Conductance
100 M up to 500 M 50 nS [2] [1] When using the relative mode (REL Q) to compensate for offsets.
[2] Add 20 counts above 33 nS in 50 nS range. [3] 289 only.
Resolution 0.001 0.01
0.0001 k 0.001 k 0.01 k 0.0001 M 0.001 M 0.01 M 0.1 M 0.1 M 0.01 nS
Accuracy 0.15 % + 20 0.05 % + 10 0.05 % + 2 0.05 % + 2 0.05 % + 2 0.15 % + 4
1.5 % + 4 1.5 % + 4 3.0 % + 2 8 % + 2 1 % + 10
Temperature Specifications
Temperature
Resolution
Accuracy [1,2]
-200 °C to +1350 °C
0.1 °C
1 % + 10
-328 °F to +2462 °F
0.1 °F
1 % + 18
[1] Does not include error of the thermocouple probe. [2] Accuracy specification assumes ambient temperature stable to ±1 °C. For ambient temperature changes of ±5 °C, rated accuracy applies after 2 hours.
72
Capacitance and Diode Test Specifications
Function Capacitance
1 nF [1] 10 nF [1] 100 nF [1]
Range
Resolution 0.001 nF 0.01 nF 0.1 nF
1 µF
0.001 µF
10 µF
0.01 µF
100 µF
0.1 µF
1000 µF
1 µF
10 mF
0.01 mF
100 mF
0.1 mF
Diode Test
3.1 V
0.0001 V
[1] With a film capacitor or better, using relative mode (REL ) to zero residual.
True-rms Digital Multimeters Detailed Specifications
Accuracy 1 % + 5 1 % + 5 1 % + 5 1 % + 5 1 % + 5 1 % + 5 1 % + 5 1 % + 5 2 % +
20 1 % + 20
73
287/289 Users Manual
Frequency Counter Specifications
Function
Range
Resolution
Frequency (0.5 Hz to 999.99 kHz, pulse width >0.5 µs)
Duty Cycle [1][2] Pulse Width [1][2]
99.999 Hz 999.99 Hz 9.9999 kHz 99.999 kHz 999.99 kHz 1.00 % to 99.00 % 0.1000 ms
0.001 Hz 0.01 Hz 0.0001 kHz 0.001 kHz 0.01 kHz 0.01 % 0.0001 ms
1.000 ms
0.001 ms
10.00 ms
0.01 ms
1999.9 ms
0.1 ms
[1] For rise times <1 µs. Signals centered around trigger levels. [2] 0.5 to 200 kHz, pulse width >2 µs. Pulse width range is determined by the frequency of the signal.
Accuracy 0.02 % + 5 0.005 % + 5 0.005 % + 5 0.005 % + 5 0.005 % + 5 0.2 % per kHz + 0.1 % 0.002 ms + 3 counts 0.002 ms + 3 counts 0.002 ms + 3 counts 0.002 ms + 3 counts
74
True-rms Digital Multimeters Detailed Specifications
Frequency Counter Sensitivity
Input Range
Approximate Voltage Sensitivity (rms sine wave) [1] 15 Hz to 100 kHz
AC Bandwidth [2]
Approximate DC Trigger Levels
DC Bandwidth [2]
50 mV
5 mV
1 MHz
5 mV & 20 mV
600 kHz
500 mV
25 mV
1 MHz
20 mV & 60 mV
1 MHz
5 V
0.25 V
700 kHz
1.4 V & 2.0 V
80 kHz
50 V
2.5 V
1 MHz
0.5 V & 6.5 V
1 MHz
500 V
25 V
300 kHz
5 V & 40 V
300 kHz
1000 V
50 V
300 kHz
5 V & 100 V
300 kHz
Input Range
Approximate Current Sensitivity (rms sine wave)
15 Hz to 10 kHz
AC Bandwidth
Approximate DC Trigger Levels
DC Bandwidth
500 µA
25 µA
100 kHz
5000 µA
250 µA
100 kHz
50 mA 400 mA
2.5 mA 25 mA
100 kHz
NA
NA
100 kHz
5 A
0.25 A
100 kHz
10 A
1.0 A
100 kHz
[1] Maximum input = 10 x range (1000 V maximum, 2 x 107 V-Hz product maximum). Noise at low frequencies and amplitudes may affect accuracy. [2] Typical frequency bandwidth with full scale (or maximum 2 x 107 V-Hz product) rms sine wave.
75
287/289 Users Manual
MIN MAX, Recording, and Peak Specifications
Function
Nominal Response
Accuracy
MIN MAX, Recording Peak
200 ms to 80% (dc function) 350 ms to 80 % (ac function) 250 µS (peak) [1]
Specified accuracy ±12 counts for changes >425 ms in duration in manual range.
Specified Accuracy ±40 counts for changes >1.5 s in duration in manual range.
Specified accuracy ±100 counts [2] up to 5,000 count (full range) reading. For
higher peak reading (to 12,000 counts), specified accuracy ±2 % [3] of
reading.
Crest Factor
350 ms to 80 %
For periodic waveforms from 50 to 440 Hz ± (4 % + 1 count).
[1] For repetitive peaks; 2.5 ms for single events. Peak not specified for 500 µA DC, 50 mA DC, 5 A DC. [2] 200 counts in 500 mV AC, 500 µA AC, 50 mA AC, 5 A AC. [3] 3 % in 500 mV AC, 500 µA AC, 50 mA AC, 5 A AC.
76
True-rms Digital Multimeters Detailed Specifications
Input Characteristics
Function L
Overload Protection [1]
Input Impedance
Common Mode Rejection Ratio (1 k unbalance)
1000 V
10 M <100 pF >120 dB at dc, 50 Hz or 60 Hz
mF V
1000 V [2]
10 M <100 pF >120 dB at dc, 50 Hz or 60 Hz
Normal Mode Rejection
60 dB at 50 Hz or 60 Hz >60 dB at 50 Hz or 60 Hz
K
1000 V
10 M <100 pF (ac-coupled)
60 dB, dc to 60 Hz
L
1000 V
3.2 k <100 pF (ac-coupled)
Not specified
Not specified
Function
Overload Protection [1]
Open Circuit Test Voltage
Full Scale Voltage
Typical Short Circuit Current
To 500 k 5 M or 50 nS 500 5 k 50 k 500 k 5 M 50 M 500 M
e
1000 V [2]
5 V dc
550 mV
<5 V
1 mA 100 µA 10 µA 1 µA 0.3 µA 0.3 µA 0.3 µA
50e
1000 V [2]
20 V decreasing to 2.5 V
500 mV
G
1000 V [2]
5 V dc
3.1 V dc
[1] Input is limited to the product of a V rms sinewave times frequency of 2 x 107 V-Hz.
10 mA 1 mA
[2] For circuits <0.5 A short circuit. 660V for high energy circuits.
77
287/289 Users Manual
Burden Voltage (A, mA, µA)
Function mA, µA
A
Range
500 µA 5000 µA 50.000 mA 400.00 mA 5.0000 A 10.000 A
Burden Voltage
102 µV/ µA 102 µV/ µA 1.8 mV/mA 1.8 mV/mA
0.04 V/A 0.04 V/A
78
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>